Upload
hephaestus-vulcan
View
218
Download
0
Embed Size (px)
Citation preview
8/3/2019 Film Screen Combinations
1/36
8/3/2019 Film Screen Combinations
2/36
Unit III
Creating the Image
8/3/2019 Film Screen Combinations
3/36
Chapter 23
Film/Screen Combinations
8/3/2019 Film Screen Combinations
4/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 4
Objectives
Relate emission spectra of intensifying
screens to specific types of radiographic
film Explain radiographic film/screen
combination relative speed numbering
systems
8/3/2019 Film Screen Combinations
5/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 5
Objectives
Calculate relative speed conversions
from one film/screen combination to
another
8/3/2019 Film Screen Combinations
6/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 6
Objectives
Describe methods of measuring
resolution:
Basic description of line pairs permillimeter
Line spread function
Modulation transfer function
8/3/2019 Film Screen Combinations
7/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 7
Objectives
Relate film/screen contrast to latitude
Determine appropriate film/screen
combinations for various clinicalsituations
8/3/2019 Film Screen Combinations
8/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.8
Film/Screen Combinations
Best combination: Film and screens are
designed to complement each other to
produce the highest image quality with
the lowest patient dose.
Film and screens must match each
other to achieve diagnostic quality
images.
8/3/2019 Film Screen Combinations
9/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.9
Film/Screen Combinations
If screens are designed to emit a light of
a certain wavelength, then the film must
be designed to enhance sensitivity to
the same wavelength.
Mismatching of film and screens
often increases patient dose. *
8/3/2019 Film Screen Combinations
10/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.10
Emission Spectra
Different elements used in intensifying
screens emit a different color light on the
electromagnetic spectrum after interaction
with x-ray photons (pg. 34) Rare earth elements emit a green light. The
film matching must be sensitive to green
emitted light. Blue light sensitive film will
not match green light emitting screens. *
8/3/2019 Film Screen Combinations
11/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.11
Emission Spectra
Green sensitive film is not sensitive to
the blue-violet light as blue sensitive
film, But is sensitive to the entire range
of phosphor emissions including yellow-
green wavelengths.*
8/3/2019 Film Screen Combinations
12/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.12
Emission Spectra
Selection of film/screen combinationsfor clinical situations should be basedon the combined qualities of both
screen and film. (See Clinical Choicespgs. 338 339)
Angio
GI & GU Ab,skeletal, chest, peds
8/3/2019 Film Screen Combinations
13/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.13
Emission Spectra
Clinical selection cannot beaccomplished by evaluating film andscreens separately.
Qualities to be considered are speed,resolution, contrast, and latitude.*
Film/screen combination relationships
are the same as the interrelationships ofthese factors for film.
8/3/2019 Film Screen Combinations
14/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.14
Emission Spectra
The most common decision that must
be made when selecting film/screen
combinations is image resolution versus
patient dose.*
8/3/2019 Film Screen Combinations
15/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.15
Characteristics
The most important characteristics
of film/screen combinations are:
Speed Resolution
Contrast
Latitude
8/3/2019 Film Screen Combinations
16/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.16
Speed
The speed of an imaging system
depends on:
Silver halide phosphor/crystal size* Silver halide phosphor/crystal layer
thickness*
Efficiency of the crystal/phosphor in
emitting photons (screens) or capturingphotons (film)*
8/3/2019 Film Screen Combinations
17/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.17
Speed
Relative speed numbers (RS) are used
to rate film/screen combinations not
units.*
The numbers are based on a relative
value of 100 for calcium tungstate IS
with medium contrast and latitude blue
sensitive film. This was know as parspeed.
8/3/2019 Film Screen Combinations
18/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.18
Speed
It has been suggested that the base RS
100 be calibrated equal to 1.28 mR to
produce film speed point (OD 1.00 +
base + fog)
8/3/2019 Film Screen Combinations
19/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.19
Speed
Relative speed numbers are usually
established at 70-80 kVp with 80
preferred.* When kVps below 70 or
above 100 are used, Relative speed
may be less consistent.
Looking at table 23-2 you can see that
relative speed can be calculated by theformula:
8/3/2019 Film Screen Combinations
20/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.20
Speed
This makes the relationship between
the relative speed number and
film/screen sensitivity inversely
proportional.
8/3/2019 Film Screen Combinations
21/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.21
Speed
Sensitivity in mR=128
RS
8/3/2019 Film Screen Combinations
22/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.22
Speed
Relative speed numbers are accurateenough to be used by technologist toconvert exposure technique settings
from one situation to another. The formula for converting mAs from
one relative speed to another is:
mAs1 RS2 mAs2 RS1
=
8/3/2019 Film Screen Combinations
23/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.23
Speed
Relative speeds combination of 400 is
the most common compromise between
patient dose and resolution in the
hospital settings.*
Higher RS reduce patient dose,
decrease latitude, and decrease
resolution.
8/3/2019 Film Screen Combinations
24/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.24
Resolution
Recorded detail, sharpness, and
resolution are measured as:
Line pairs/mm (lp/mm)*
Measures the minimum size and space
between objects that can be visualized on
the final image
8/3/2019 Film Screen Combinations
25/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED.25
Resolution
Modulation Transfer Function* MTF provides the best measurement of the
resolving ability of a screen/film combination. Itmeasures the information lost between the subjectand the image receptor. It is determined by thelight diffused between screen and film. It is ascreen function.
Line spread function (LSF)*
Measures the ability of a film/screen system toaccurately represent or measure the boundaries ofan image.*
8/3/2019 Film Screen Combinations
26/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 26
Contrast
Is primarily the contrast and function of
the film.
Intensifying screens also exhibitcontrast. Rare earth phosphors contrast
is slightly higher than calcium tungstate.
8/3/2019 Film Screen Combinations
27/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 27
Latitude
Primarily dependent upon film latitude
and is directly related to contrast.
8/3/2019 Film Screen Combinations
28/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 28
Latitude
Latitude and contrast are inversely
related. Narrow latitude film exhibit a high scale of
contrast.
Wide latitude film exhibit a low scale of
contrast.
High speed film/screen systems tend to have
lower resolution (decrease detail)*
8/3/2019 Film Screen Combinations
29/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 29
Selecting Film/Screen Combinations
The selection of appropriate film/screen
combinations requires consideration of
the various film and screen
characteristics.
8/3/2019 Film Screen Combinations
30/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 30
Clinical Choices
Film and intensifying-screen
manufacturers publish film/screen
combination charts that compare all
possible combinations of their films and
intensifying screens.
8/3/2019 Film Screen Combinations
31/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 31
Angiography
Higher speed rare earths provide
several advantages*
Decreased patient dose Allows for shorter exposure times
Maintenance of resolution
Quantum mottle is less of a problem.
8/3/2019 Film Screen Combinations
32/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 32
Gastrointestinal and Urologic
Radiography
High subject contrast due to use of
contrast media
Allows for use of faster film/screencombination*
8/3/2019 Film Screen Combinations
33/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 33
Abdominal Radiography
Low subject contrast
Makes slower film/screen combination
preferable*
8/3/2019 Film Screen Combinations
34/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 34
Skeletal Radiography
Extremities
Fine/detail screens or slower systems are
desirable
Pelvic and lumbar studies
Higher speed film/screen combo systems
are desirable due to large part thickness.
8/3/2019 Film Screen Combinations
35/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 35
Chest Radiography
High kVp chest radiography is achieved
more efficiently with calcium tungstate
screens.*
8/3/2019 Film Screen Combinations
36/36
Copyright 2006 by Thomson Delmar Learning. ALL RIGHTS RESERVED. 36
Pediatric Radiography
High speed film/screen combinations
Allow for reduction in technical factors
reducing patient dose.
Shorter exposure times
Will limit repeats due to patient motion