I/

Installation and Operating Instructions

Sartorius CC30002, CC50002Electronic Mass Comparator

98648-011-03

Page 2: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

3 General View of the Mass Comparator

4 Intended Use4 Warranty4 Equipment Supplied4 Storage and Shipping Conditions5 Warning and Safety Information

5 Important Note to Users

5 Installation Instructions5 Ambient Conditions6 C Marking

7 Getting Started7 Setting Up the Mass Comparator7 Leveling the Weighing Platform Using

the Level Indicator

7 Connecting the Mass Comparator to AC Power

8 Safety Precautions8 Selecting the Voltage

9 Operating the Mass Comparator9 Warmup Time9 Turning the Display On and Off

(Standby Mode)9 Self-test

10 Mass Comparator Display10 Displaying Model and Serial No. in the

Weight Display11 Setting ID No./Date/Time11 ID for GLP/GMP-compliant Record

12 Adjustment/Calibration12 Blocking the Adjustment/Calibration

Functions

13 General Directions for Working with the Mass Comparator

13 Working with the Mass Comparator13 General Directions14 Weighing Methods14 A-B-B-A Method14 A-B-A Method

15 Data Interface15 Connecting Electronic Peripheral Devices15 Connecting Devices to the

Mass Comparator

16 Care and Maintenance16 Service16 Cleaning16 Safety Inspection16 Instructions for Recycling the Packaging

17 Troubleshooting

Contents

18 Mass Comparator Operating Menu18 Changing Menu Code Settings19 Accessing the Operating Menu19 Reset: Undoing All Menu Code Changes

20 Mass Comparator Operating Parameters20 Adapting the Mass Comparator

to Ambient Conditions20 Standard Weighing – Manual Filling20 Stability Range20 Stability Symbol Delay20 Tare Parameter21 Auto Zero Function21 Calibration/Adjustment and Linearization

Functions

21 Interface Parameter Settings21 Baud Rate21 Parity21 Number of Stop Bits21 Handshake Mode22 Weight Units22 Display Resolution23 Configuring the Data Record23 Data Output Parameter23 Automatic Data Output24 Data Output at Defined Intervals 24 Automatic Tare Function24 Data ID Codes24 Individual Data ID Codes

25 Additional Functions25 Menu Access Function25 Mass Comparator Operating Menu25 Acoustic Signal25 Blocking the Keys25 Universal Remote Switch26 Blocking the Numeric Keys26 Power-On Mode26 Display Backlighting

27 ISO/GLP-compliant Printout or Record28 Printout/Record for Calibration/Adjustment

Functions28 Reset: Undoing All Menu Code Changes

29 Accessories

30 Specifications

2

Page 3: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

3

.

I/

0 1 2 3 4 5 6 7 8 9 .CF

F1

F2

TARE

123

5

6

7

8

9

10

11

4

25

24

23

22

21

20

19

18

17

16

15

14

13

12

No. Designation No. Designation

1 Weighing platform 14 w Toggle key2 Leveling foot 15 i Info key3 Level indicator 16 e On/off key4 Connector for system cable 17 c Clear Function key

(weighing platform to display and control unit) 18 Range indicator (bar graph)5 Application display 19 Connector for system cable6 Numeric keys (display and control unit to weighing platform)7 f Function key 20 Interface port8 F Function key 21 Menu access switch9 p Print key (data output) 22 Power socket10 Function display for f and F keys 23 Manufacturer’s ID label on the back of the housing11 t Tare key 24 Display and control unit12 Application program symbol 25 Weighing pan13 Weight display

General View of the Mass Comparator

Page 4: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

The models CC30002 and CC50002 are mass comparators from our full-range series,designed for high-precision mass determination based on the principle of differentialmass comparison. Practical applications for the mass comparator include, for example:

– testing high-precision work pieces,– mass determination of expensive materials,– detection of slight changes in the mass of objects, and – calibration of weights and mass standards.

If you wish to use your mass comparator for other purposes, please contact your Sartorius office or dealer.Sartorius accepts no liability for any damage that occurs as a result of use of the masscomparator for purposes other than those for which it was originally intended.Please read these operating instructions carefully before putting your new mass compara-tor into operation for the first time.

WarrantyDo not miss out on the benefits of our full warranty.Please contact your local Sartorius office or dealer for further information. If available,complete the warranty registration card, indicating the date of installation, and return thecard to your Sartorius office or dealer.

Equipment SuppliedCarefully unpack all components of the mass comparator.The equipment supplied includes the following components:

– Weighing platform (1)

– Weighing pan (25)

– Display and control unit (24)

– System cable

– AC adapter

Storage and Shipping Conditions– Allowable storage temperature: + 5°C to + 40°C (+ 41°F to + 104°F)– Do not expose the mass comparator unnecessarily to extreme temperatures, moisture,

shocks, blows or vibrations.

After unpacking the mass comparator, check it immediately for any visible damage as aresult of rough handling during shipment.If you detect any damage, proceed as directed in the chapter entitled “Care and Mainte-nance”, under “Safety Inspection”.In case you need to ship your mass comparator in future:

– Save the box and all parts of the packaging, to prevent damage in shipment.– Unplug all connected cables before packing your mass comparator.

To prevent damage, remove the weighing pan before shipping.

4

Intended Use

.

I/

Page 5: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

5

Warning and Safety Information– The mass comparator may be opened only by authorized service technicians who have

been trained by Sartorius and who follow Sartorius’ standard operating procedures formaintenance and repair work.

– Unplug the power cord from the wall socket (mains) before performing any work on themass comparator.

– Do not use this equipment in hazardous areas/locations.

– The mass comparator must not be exposed during operation to potentially explosive mixtures of gases and/or dust.

– The seals affixed to this equipment indicate that only authorized service technicians areallowed to open the equipment and perform maintenance work so that safe and trouble-free operation of the equipment is ensured and the warranty remains in effect.If the seal is broken, this will result in forfeiture of all claims under the manufacturer’swarranty.

Important Note to UsersMake sure to carefully read and follow sections marked with this symbol. They contain important safety instructions.

Before you begin to weigh, please read the following directions carefully. In this way, you will avoid making any operating errors which could result in damage to the weighingplatform.

– Working with the mass comparator requires a smooth, uninterrupted technique.

– After initially connecting your mass comparator to AC power (or after a relatively long power outage), allow at least 12 hours warmup time beforeoperating the equipment.

– Before connecting or disconnecting peripheral devices (such as a printer) to or from the interface port on the display and control unit, make sure to disconnect the mass comparator from power.

Installation Instructions

Ambient ConditionsYour Sartorius mass comparator is designed to provide reliable weighing results undernormal ambient conditions.Set up the mass comparator in a conditioned room kept at a constant temperature (maximum allowable fluctuation: DT < ±0.5 K/h), free of drafts and aggressive agents.The humidity in the room should be between 50% and 55% (maximum allowable fluctuation: <10%). During a weighing operation, avoid moving around in the roomwhere the mass comparator is set up.

Page 6: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

Select a suitable place of installation for the mass comparator:

– Set up the mass comparator on a stable, even surface (e.g. a Sartorius balance table)– Do not expose the mass comparator to heat radiation generated, for example, by direct

sunlight, light fixtures or heaters.– Protect the mass comparator from drafts that come from open doors or windows or from

air conditioning units.– Avoid exposing the mass comparator to any vibration during weighing.– Protect the mass comparator from aggressive chemical vapors.– Do not expose the mass comparator to extreme moisture over long periods.

Moisture in the air can condense on the surfaces of a cold mass comparator whenever it isbrought to a substantially warmer place. If you transfer the mass comparator to a warmerarea, make sure to condition it for about 2 hours at room temperature, leaving itunplugged from AC power.

– Avoid static electricity at the place of installationUse suitable materials for the floor covering, upholstery, operators’ clothing and tools.You can adapt the mass comparator to ambient conditions by setting the respective codesin the operating menu (for details, see “Operating the Mass Comparator”).

C MarkingThe C marking affixed to the equipment indicates that the equipment complies with thefollowing Directive: 89/336/EEC “Electromagnetic compatibility (EMC)”.

1. Electromagnetic compatibility:EN 61326-1 Electrical equipment for measurement,

control and laboratory use EMC requirements

Part 1: General requirementsDefined immunity to interference: Industrial areas, continuous, un-monitored operation

Limitation of emissions: Residential areas, Class B

Important Note:The operator shall be responsible for any modifications to Sartorius equipment and forany connections of cables or equipment not supplied by Sartorius and must check and, if necessary, correct these modifications and connections. On request, Sartorius will provide information on the minimum operating specifications (in accordance with theStandards listed above for defined immunity to interference). The manufacturer declaresthe conformity of the equipment.

Council Directive 73/23/EEC “Electrical equipment designed for use within certain voltage limits”.Applicable European Standards:

EN 60950 Safety of information technology equipment including electrical business equipment

EN 61010 Safety requirements for electrical equipment for mea-surement, control and laboratory equipmentPart 1:General requirements

If you use electrical equipment in installations and under ambient conditions requiringhigher safety standards, you must comply with the provisions as specified in the applicable regulations for installation in your country.

6

°C

Page 7: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

Setting Up the Mass Comparator§ Place the display and control unit (24) to the left or the right of the weighing platform,

depending on your requirements. Make sure you have a clear view of the application display (5).

§ Plug one end of the supplied system cable into the socket at the back of the weighingplatform (1) and the other end at the back of the display and control unit (24). Tighten the screws on the connectors.

§ Place the weighing pan (25) on the weighing platform.

Leveling the Weighing Platform Using the Level Indicator

§ At the place of installation, level the weighing platform using the leveling feet (2) so thatthe air bubble is centered within the circle of the level indicator.

§ To level the weighing platform, proceed as follows:– To raise the weighing platform, extend the leveling feet (turn clockwise)– To lower the weighing platform, retract the leveling feet (turn counterclockwise)

Connecting the Mass Comparator to AC PowerThe mass comparator is powered by the AC adapter supplied. Make sure that the voltagerating printed on this unit is identical to your local line voltage.If the voltage specified on the label or the plug design of the AC adapter does not matchthe rating or standard you use, please contact your Sartorius office or dealer.

§ Plug the AC adapter into the socket (22) at the back of the display and control unit (24).

§ Then plug the AC adapter into the wall outlet.

7

Getting Started

Page 8: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

Note:Use only original Sartorius AC adapters. Use of AC adapters from other manufacturers,even if these units have a registered approval rating from a national testing laboratory,requires the consent of an authorized Sartorius service technician.

Safety PrecautionsThe AC adapter rated to Class 2 can be plugged into any wall outlet without taking anyadditional safety precautions. The pole of the output voltage is connected to the housingof the display and control unit, which can be grounded for operation.

The data interface is also electrically connected to the housing of the weighing platform(ground).

Selecting the VoltageWhen you use the portable AC adapter model 69 71172 with a standardized Europeanplug, you can select the desired voltage (see “Accessories”).

Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules. These limits are designed to providereasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio fre-quency energy and, if not installed and used in accordance with the instruction manual,may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Changes or modifications notexpressly approved by Sartorius AG could void the user’s authority to operate the equipment.

8

Page 9: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

Warmup Time§ Once you set up the mass comparator at its place of use, allow

at least 12 hours acclimation time. After initially connecting your mass comparator to AC power (or after a relatively long power outage), the system will need to warm up for atleast 2 hours before you can operate it.

Turning the Display On and Off (Standby Mode)§ To turn the mass comparator display on or off, press the e key (16).

Self-test> After the mass comparator has been turned on, an automatic self-test of the mass

comparator’s electronic circuitry is performed.At the end of the self-test, either 0.00 g or 0.000 is displayed, depending on the model.

You can now begin weighing.Calibrate/adjust the weighing platform before performing the first weighing operation(see “Calibration/Adjustment” below).

9

Operating the Mass Comparator

.

I/

0 1 2 3 4 5 6 7 8 9 .CF

F1

F2

TARE

Page 10: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

10

Mass Comparator DisplayThe display shows the following special codes for your information:

<O> means standby> The display was turned off using the e key (16) and is currently in standby mode.

The mass comparator is ready to operate and does not need to warm up.

<b> means busy> Once you switch on the mass comparator, the “b” symbol is displayed until you press

a key.

> During operation, the “b” symbol indicates that the mass comparator processor is stillbusy processing a function and cannot accept another command to perform any otherfunctions at this time.

CAL> Press and hold the t key (11) and press the f key (7) to initiate the external

calibration/adjustment function (see also the chapter entitled “Calibration/Adjustment”).

Displaying Model and Serial No. in the Weight Display– Turn off the mass comparator– Turn the mass comparator back on– While all segments are displayed, press the p key (9) briefly> The model designation is displayed– Press the c key (17) to display the serial number– Press the c key (17) to quit the function

0 1 2 3 4 5 6 7 8 9 .CF

F1

F2

TARE

TARE

Page 11: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

11

Setting ID No./Date/Time

§ Turn off the mass comparator

§ Turn the mass comparator back on

§ While all segments are displayed, press the w key (14) briefly

§ Press the f key (7) to toggle between the ID no., “Id”,date “dAt” and time “tIM”

“ID for GLP/GMP-compliant Printout/Data Record”The ID number can be up to 8 characters long, and can contain the numbers 0 through 9as well as the dash (“-”) character. A zero at the beginning of the number is not output.

§ Enter the ID number; e.g., 4 . 3 2 . 1

– Press STO - F to confirm (8)

“Date”§ Enter a decimal point (“.”) between day, month and year by pressing the p key (9)

e.g., 1 0 . 0 4 . 0 2

– Press STO - F to confirm (8)

“Time”§ Enter the hours, minutes and seconds,

e.g., 1 9 . 0 5 . 3 0

§ The internal clock begins at the set time when you press the STO - F key (8)

§ If you enter a number between 0 and 12 for the hour, select “A” for a.m. or “P” for p.m.by pressing f (7) to select and

> then STO - F (8) to store,> or simply press STO F (8) to accept the default 24-hour time display

§ Press c (17) to end this function.

0 1 2 3 4 5 6 7 8 9 .CF

F1

F2

TARE

Page 12: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

12

Calibration/Adjustment

PurposeCalibration technically means to determine the difference between the weighing instrument readout and the actual weight on the platform to determine the accuracy.Calibration does not entail any modification of the calibration weight value stored in themass comparator.Adjustment is the correction of the difference between the measured value displayed andthe true weight (mass) of a sample, or the reduction of the difference to an allowable levelwithin maximum permissible error limits.During calibration, the mass comparator is adjusted to changes in the ambient conditions.Calibrate/adjust your mass comparator at the place of installation:

– after each warmup period– before the first weighing operation– following a change in ambient conditions– each time the mass comparator is set up in a different area

Use only calibration weights of an accuracy equal to or better than the particular accuracyclass of your mass comparator. See the section entitled “Accessories” for an overview ofthe calibration weight sets available. You will need calibration weights of the followingaccuracy classes:

Model CC30002 CC50002Weight (g) 1x10000 1x20000Accuracy class E2 E2

– Tare the mass comparator when the display has stabilized.

– Press and hold the t key (11) approx. 2 seconds

> The Application display shows “C.E.” (“calibration: external”), and the function displayshows “CAL”.

– Press f (7) to initiate the calibration routine; the calibration weight (10000 or 20000 g)is displayed.

– Center the calibration weight on the weighing pan.

> The mass comparator is calibrated automatically. At the end of calibration, the calibra-tion weight value is displayed with the weight unit and an acoustic signal is emitted.

Note:You can also enter the conventional weight of your calibration weight via the numerickeys to adjust the sensitivity of the mass comparator more precisely. To do this, press f(7) (CAL) to display the currently valid calibration weight. Now you can enter the desiredweight value using the numeric keys. Press f (7) again (STO) to store the weight value;the previous value is overwritten. Center the prompted weight on the weighing pan. The value you entered is stored automatically. If the error code “Err 02” is displayed, this indicates an error or disturbance in starting the calibration procedure.

– In this case, press t briefly and then, when the display shows a zero readout, press thef key again.

Note:Press and hold the f (7) key to store the values in non-volatile memory.

Blocking the Calibration/Adjustment FunctionsYou can block these functions by setting code 1 9 2.

Page 13: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

13

Working with the Mass ComparatorWorking with the mass comparator requires a steady hand and a smooth, uninterruptedtechnique.

– Perform a few trial weighing procedures before you begin the actual mass comparisonsequence, to get a feel for the handling method required.

> The weights and the mass comparator should remain next to one another in the weighingroom for an extended period to condition them to the same temperature.

– Place the weight gently on the weighing pan.Moisture or warmth from your hands can cause measuring errors.For best results, wear gloves or use weight handles or forceps while working with themass comparator (see “Accessories”).

– Always place the weight in the middle of the pan. If the weight is not centered, this mayresult in systematic deviations in the weighing results which can influence the accuracy ofyour measurement unnecessarily, resulting in longer response times.

General Directions– Press the e key (16) to turn on the mass comparator

– Center your reference weight on the weighing pan

– Calibrate/adjust the mass comparator if necessary

> After a brief period, a readout appears on the display. Press the t key (11) to set it to 0.00 g.If the displayed value changes after you have set it to zero, tare the mass comparatoragain. If the display continues to change, make sure you are observing the instructionsgiven in the section entitled “Working with the Mass Comparator.”

– Remove the reference weight from the weighing pan.

– Place your test weight on the weighing pan.

> After a brief period, the display shows the difference between the reference weight andthe test weight with the respective plus or minus sign.

– Read off the weighing result after the mass comparator has stabilized.

Note:Weighing with a mass comparator is a method of measuring by comparison, known as comparative weighing. The known mass of a reference weight is compared to the massof the test weight, in a two-part weighing procedure. The difference in mass is shown inthe weight display (13) along with the respective plus or minus sign.With full range mass comparators you can perform comparison weighing in any range, up to the maximum capacity.

General Directions for Working with the Mass Comparator

.

I/

Page 14: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

14

Weighing MethodsFor accurate results, we recommend using the “A-B-B-A” or “A-B-A” method, wherebythe difference in mass between test weight (B) and reference weight (A) is determinedrepeatedly and the average of the individual weighing results is calculated from the meanvalue of the measured differences.

A-B-B-A MethodPlace the reference weight (A) and the test weight (B) alternately on the weighing pan in the sequence shown below, maintaining a smooth, steady pace:

A– –B– –B– –A– –A– –B– –B– –A– –A– –B– –B– –A– etc.

1st cycle 2nd cycle 3rd cycle etc.

A-B-A MethodPlace the reference weight (A) and the test weight (B) alternately on the weighing pan in the sequence shown below, maintaining a smooth, steady pace:A– –B– –A– – –B– –A– –B– –A– –B– –A– etc.

1st cycle 2nd cycle 3rd cycle etc.

Please note when using either of these methods:The time intervals between A-B and B-A or between B-B and A-A should always be the same.This produces a smooth, steady working rhythm.

Note:Always lift the weight completely off of the weighing pan.

Page 15: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

15

Connecting Electronic Peripheral DevicesBefore you connect a peripheral device (such as a printer or PC) to the interface port:

§ Make absolutely sure to unplug the mass comparator from AC power - also before disconnecting a peripheral device from the mass comparator data interface (20).

> To document measurements with a Sartorius printer, connect the printer to the interfaceport (20).You do not need to adjust any settings.

§ Re-connect AC power.

§ To print or output data, press the p key (9).

For details on the data interface (such as the data output or input formats, pin assign-ment, etc.), see the “Interface Description” included with your mass comparator.

Connecting Devices to the Mass ComparatorNote:

– The data interface is electrically connected to the mass comparator housing.

– The accessory cabling supplied is shielded and electrically connected on both ends to thecases of the connectors. This electrical connection may result in interference caused byground loops or by transient currents if you have grounded the housing or connected theprotective grounding conductor for AC power. If necessary, connect an equipotentialbonding conductor to the mass comparator.If you have any problems with your mass comparator, please contact your nearest Sartorius Service Center.

Data Interface

Page 16: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

16

Care and Maintenance

ServiceRegular servicing by a Sartorius technician will extend the service life of your mass comparator and ensure its continued weighing accuracy. Sartorius can offer you servicecontracts, with your choice of regular maintenance intervals ranging from 6 months to 2 years.

CleaningBefore cleaning the mass comparator, unplug the AC adapter from the wall outlet (mains supply).Please do not use any aggressive cleaning agents (solvents or similar agents). Instead, use a cloth which has been wet with a mild detergent (soap).Make sure that no liquid enters the mass comparator housing. After cleaning, wipe downthe mass comparator with a soft, dry cloth.

Note:If you clean the mass comparator with water that is too hot or too cold, the difference in temperature between the water and the equipment can cause condensation to formwithin the equipment.This condensation can lead to malfunctions in the equipment.

Safety InspectionIf there is any indication that safe operation of the mass comparator with the AC adapteris no longer warranted, turn off the power and disconnect the equipment from AC powerimmediately. Lock the equipment in a secure place to ensure that it cannot be used forthe time being.

Safe operation of the mass comparator with the AC adapter is no longer ensured when

– there is visible damage to the AC adapter

– the AC adapter no longer functions properly

– the equipment has been stored for a relatively long period under unfavorable conditions (such as extreme humidity)

In this case, notify your nearest Sartorius Service Center or the International TechnicalSupport Unit based in Goettingen, Germany. Only service technicians who are authorizedby Sartorius and who have access to the required maintenance manuals are allowed toperform maintenance and repair work on the equipment.

Instructions for Recycling the PackagingTo ensure safe shipment, your mass comparator has been packaged using environmentallyfriendly materials. After successful installation of the mass comparator, you should returnthis packaging for recycling.For information on recycling old weighing equipment, consult your municipal waste disposal center or local recycling depot.

.

I/

Page 17: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

17

Problem Cause Solution

– No segments appear on the weight – No AC power is available – Check the AC power supplydisplay (13) – The AC adapter is not plugged in – Plug in the AC adapter

– The weight display shows “H” – The load exceeds the – Unload the mass comparatorcapacity of the mass comparator

– The weight display shows “L” – The weighing pan (25) – Position the pan on theis not in place mass comparator

– The weight display – The t key (11) was not – Press t when a zero readout briefly shows “Err 02” pressed before starting calibration is shown, then press and hold

the t key (11);then press f until thecalibration weight is displayed

– The special code “b” – None of the keys has been pressed since – Press a keyremains displayed the mass comparator was turned on

– The weight readout changes constantly – Unstable ambient conditions – Set up the mass comparatorin another area

– Too much vibration, or the – Access the menu to select the correct mass comparator is exposed to a draft code for the ambient conditions

– The mass comparator – A foreign object is caught – Remove the foreign objectreadout does not stabilize between the weighing pan

and the housing of themass comparator

– The weight readout – The mass comparator has been not – Calibrate the mass comparatoris obviously wrong calibrated

– The mass comparator was – Tare before weighingnot tared before weighing

– The air bubble of the level indicator – Level the mass comparatoris not within the circle (see page 7)

Troubleshooting

Page 18: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

18

In the operating menu, you can define how your mass comparator adapts to ambientconditions and also how it works to meet your special requirements.

The factory-set menu codes are identified by an asterisk (“*”) in this manual. You can select a function not marked with “*” by selecting the corresponding menu code.These codes are listed in the section entitled “Mass Comparator Operating Parameters”. If you need to change any of the factory settings, we recommend that you note the newsetting here, along with the date the change was made.

Changing Menu Code SettingsTo select specific functions, you will need to set the respective menu codes. There are three steps to changing a code:

– Accessing the menu

– Setting the code

– Confirming and storing the code

The keys have special functions when setting menu codes. The four keys used to set thecodes appear as arrows in the illustration on the left:

< w and > p = Move to the right and left

f and v F = Increase/decrease a digit value by one

t= Confirm a code setting

c = Store code setting and exit the menu

Example:Now try changing a code yourself. Follow the example below to change the power-onmode to “Automatic power-on”.

The menu code for “Automatic power-on” is 8 5 4.

<

Mass Comparator Operating Menu

Page 19: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

19

Accessing the Operating Menu§ Press the e key (16) to turn the mass comparator off.

§ Press e to turn on the mass comparator. While all segments are displayed, briefly hold down t (11).

§ Release the t key when “I” is displayed.

§ If “-L-” is shown in the Application display, remove the protective cap from the menuaccess switch (21) at the back of the display and control unit and move the switch in thedirection of the arrow (see the illustration below).

§ Press F (8) repeatedly until “8” is displayed.

§ Press p (9) until the second number of the code appears.

§ Press f until “5” is displayed.

§ Press p until the third number of the code appears (when you move to this number, the previously set menu code appears).

§ Press the f key to select “4”.

§ Press t (11) to confirm the code you have just set (this is indicated by the “o” after the code).

§ To store the new menu code setting, press and hold the c key (17) for more than 2 seconds.

Important Note:Changes to code settings are not stored if you turn off the mass comparator by pressing e (16) while selecting code numbers or before pressing t to save a setting. The current code setting in the mass comparator operating menu is identified by a small,superscript “o” after the last number. If you wish to change several menu code settings,you do not have to exit the mass comparator operating menu (c key) after each change.You can store individual settings as you go along, without exiting the menu. When youaccess the operating menu, the previously set code is displayed when you select the right-hand number, which means the entire menu code setting is displayed. This makes it easyfor you to check the previously set menu codes.Please do not forget to relock the operating menu (return the menu access switch (21) tothe “closed” position) to avoid inadvertent or unauthorized changes to the settings. The “-L-” symbol on the display indicates that the menu is locked.You can lock the menu with the menu access switch (21) anytime you wish once you haveaccessed it. However, it is best if you wait until you have changed the last code setting.

Reset: Undoing All Menu Code ChangesThe reset function lets you undo all menu code changes, which means the original factorydefaults (marked by an asterisk (“*”) in this manual) are restored. To use this function,select code 9 – 1°.

See the previous page for information on confirming and storing a menu code setting.

TARE

F2F2

0 1 2 3 4 5 6 7 8 9 .CF

F1

F2

TARE

TARE

Page 20: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

20

Mass Comparator Operating Parameters

Adapting the Mass Comparator to Ambient ConditionsThe mass comparator can be adapted to the prevailing ambient conditions at the place ofinstallation.

Menu code Factory setting Changed

Very stable conditions 1 1 1

Stable conditions 1 1 2

Unstable conditions 1 1 3 *

Very unstable conditions 1 1 4

Standard Weighing - Manual FillingYou can optimally adapt your mass comparator to meet either of these requirements. In the manual filling mode, the display compensates for fluctuations of the load on the masscomparator, giving you especially fast and stable readouts.

Menu code Factory setting Changed

Standard weighing mode 1 2 1 *

Manual filling mode 1 2 2

Stability RangeThe stability symbol is displayed steadily once the weight variation is within ±:

Menu code Factory setting Changed

0.25 digits 1 3 1

0.5 digits 1 3 2

1 digit 1 3 3

2 digits 1 3 4

4 digits 1 3 5 *

8 digits 1 3 6

Stability Symbol DelayThis setting enables your mass comparator to compensate for individual interfering factorswhich slowly subside. As a rule, you will not need to change this code.

Menu code Factory setting Changed

No delay 1 4 1

Short delay 1 4 2 *

Long delay 1 4 3

Extremely long delay 1 4 4

Tare ParameterYou can define when the mass comparator will perform the taring operation.

Menu code Factory setting Changed

At any time 1 5 1

Not until the readout is stable 1 5 2 *

Page 21: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

21

Auto Zero FunctionWhen this zero tracking function is activated, any changes off the zero readout that areequal to a defined fraction of digits per second are automatically tared. In other words, it ensures a stable readout.

Menu code Factory setting Changed

Auto Zero on 1 6 1

Auto Zero off 1 6 2 *

Calibration/Adjustment and Linearization FunctionsExternal calibration/adjustment and/or linearization by pressing the t key (11) (hold keylonger than 2 seconds) and the f key (7). The weights to be placed on the mass compara-tor for linearization are displayed one after the other in increasing order.

Calibration/Adjustment Menu code Factory setting Changed

External calibration 1 9 1 *

blocked 1 9 2

External linearization 1 12 1 *

blocked 1 12 2

Interface Parameter Settings(see “Interface Description”)

Baud RateMenu code Factory setting Changed

150 baud 5 1 1

300 baud 5 1 2

600 baud 5 1 3

1200 baud 5 1 4 *

2400 baud 5 1 5

4800 baud 5 1 6

9600 baud 5 1 7

19,200 baud 5 1 8

ParityMenu code Factory setting Changed

Mark 5 2 1

Space 5 2 2

Odd 5 2 3 *

Even 5 2 4

Number of Stop BitsMenu code Factory setting Changed

1 stop bit 5 3 1 *

2 stop bits 5 3 2

Handshake ModeMenu code Factory setting Changed

Software handshake 5 4 1

Hardware handshake with 2 characters after CTS 5 4 2 *

Hardware handshake with1 character after CTS 5 4 3

Page 22: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

22

Weight Units

Symbol in Symbol output Conversion factor Menu code Factory Changeddisplay to interface port** setting

1 g =

User-defined unit 1. 1 7 1

Grams g 1. 1 7 2 *

Kilograms kg 0.001 1 7 3

Carats ct 5. 1 7 4

Pounds lb 0.0022046226 1 7 5

Ounces oz 0.035273962 1 7 6

Troy ounces ozt 0.032150747 1 7 7

Hong Kong taels tl tlh 0.02671725 1 7 8

Singapore taels tl tls 0.02646063 1 7 9

Taiwanese taels tl tlt 0.02666666 1 7 10

Grains GN 15.43235835 1 7 11

Pennyweights dwt 0.643014931 1 7 12

Milligrams mg 0.0001 1 7 13

Parts per pound o /lb 1.1287667712 1 7 14

Chinese taels tl tlc 0.02645547175 1 7 15

Mommes m mom 0.2667 1 7 16

Austrian carats k k 5. 1 7 17

Tola t tol 0.0857333381 1 7 18

Baht b bat 0.06578947436 1 7 19

Mesghal m MS 0.217 1 7 20

** Some of the unit symbols printed on hardcopy or output to a computer will differ from the way they are shown on the mass comparatordisplay.

Display Resolution

With the mass comparator, the last digit displayed can be rounded off.

Display mode Menu code Factory setting Changed

Highest possible accuracy 1 8 1 *

Last numeral blankedwhen load changes 1 8 2

Rounding factor 2 1 8 3

Rounding factor 5 1 8 4

Rounding factor 10 1 8 5

Page 23: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

23

Configuring the Data Record

Sartorius mass comparators come with an interface as standard equipment.You can plug a Sartorius printer or a computer into this interface port. In addition, you canchoose to have data transmitted to or from your mass comparator to the connected deviceeither automatically or by pressing a key.The mass comparator operating menu lets you define the various parameters for data output.For information on the data formats and for interfacing a computer or a different peripheraldevice, please refer to the chapter entitled “Data Interface”.

Data Output Parameter

This parameter is coupled with the stability parameter; stability = stable readout or nomotion is detected.Print on request = data is output when the output key is pressed or a software command isreceivedAuto print = continuous, automatic data output

Menu code Factory setting Changed

Print on request regardless of stability 6 1 1

Print on request after stability, with storage of the function 6 1 2 *

Print on request at stability without storage of the function 6 1 3

Auto print regardless of stability 6 1 4

Auto print at stability 6 1 5

Automatic Data OutputYou can stop and start automatic data output (auto print function) by pressing the p key(9). To avoid operating errors or to ensure that the data is output continuously in the auto-matic mode, you can block this function.

Menu code Factory setting Changed

Start/stop auto print using the p key (9) 6 2 1

Auto print not stoppable 6 2 2 *

Page 24: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

24

Data Output at Defined Intervals

You can reduce the volume of data in the “auto print” mode by defining the interval atwhich data will be output automatically. This auto print interval is based on the number of display updates.

Auto print interval Menu code Factory setting Changed

1 display update 6 3 1 *

2 display updates 6 3 2

5 display updates 6 3 3

10 display updates 6 3 4

20 display updates 6 3 5

50 display updates 6 3 6

100 display updates 6 3 7

Automatic Tare Function

You can choose to have the mass comparator zero/tare automatically after the output key ispressed.

Menu code Factory setting Changed

No automatic taring 6 4 1 *

Automatic taring after the p key (9) is pressed 6 4 2

Data ID Codes

The ID code for weighed values consist of a code letter printed or displayed in front of thesevalues.If you select “without data code,” only the weighed value is output. This setting must be changed to generate ISO/GLP-compliant data records. The ID code increases the data output format for each weight readout from 16 to 22 characters.

Individual Data ID Codes

Menu code Factory setting Changed

Without 7 2 1 *

With 7 2 2

Page 25: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

25

Menu Access FunctionYou can override the physical menu access switch (21) by setting the code for the mass comparator operating menu to “accessible.” The mass comparator operating menu is thenaccessible at all times, which means you can change the menu codes at any time regardlessof the menu access switch position.

Mass Comparator Operating Menu

Menu code Factory setting Changed

Accessible: “-C-” is displayed 8 1 1

Depends on menu access switch setting: “-L-” is displayed 8 1 2 *

Acoustic Signal

If you wish, you can turn off the acoustic signal (beep tone).Acoustic signal

Menu code Factory setting Changed

On 8 2 1 *

Off 8 2 2

Blocking the Keys

You can block all keys on the mass comparator (except the e key (16)).

Key function Menu code Factory setting Changed

Accessible 8 3 1 *

Blocked 8 3 2

Universal Remote Switch

You can connect an external remote switch to the interface port (20) on the mass comparator.The external switch can be used to trigger one of the following key functions:

Key function Menu code Factory setting Changed

p key (9) 8 4 1 *

t key (11) 8 4 2

f key (7) 8 4 3

F key (8) 8 4 4

Additional Functions

Page 26: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

26

Blocking the Numeric Keys

You can block the numeric keys by setting the corresponding menu code.

Number keys: 0 – 9 and . Menu code Factory setting Changed

Blocked 2 5 1

Accessible 2 5 2 *

Power-On Mode

Depending on your requirements, you can change the power-on mode of your mass com-parator.

The factory setting is: (power) –> off <–- > on <–> standby <.

If you select “Toggle between on and standby,” the mass comparator automatically switcheson again after it has been temporarily disconnected from AC power or a power failure hasoccurred (menu code 8 5 3).

In the “Automatic power-on” mode, the mass comparator will automatically turn on againafter it has been temporarily disconnected from AC power, after a power failure has occurred,or when you press e (16). With this setting, pressing the e key does not switch the masscomparator off.

Power-on mode Menu code Factory setting Changed

(Power) off –> on <–> < 8 5 1 *

> on <–> standby < 8 5 3

Automatic power-on 8 5 4

Display Backlighting

You can turn the display backlighting on or off to suit your individual workplace require-ments.

Display backlighting Menu code Factory setting Changed

On 8 6 1 *

Off 8 6 2

Page 27: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

27

ISO/GLP-compliant Printout or Record

Application:

Use of the mass comparator as a test and measuring instrument in quality assurance systemsin compliance with the requirements of ISO, GLP, GMP and EN (European Standards) inwhich proof of the mass comparator’s accurate performance is required.

The mass comparator can record all completed calibration or adjustment operations andprint out data in compliance with the requirements of Good Laboratory Practice (GLP). The mass comparator, interfaced with a data printer or a computer, generates documentationthat includes the date, time, serial number and model designation, which means data can beunambiguously traced to the mass comparator that generated it and the time at which it was generated.

Select the ISO/GLP-compliant printout or record mode by setting the corresponding code inthe operating menu.

ISO/GLP-compliant printout/record modeMenu code Factory setting Changed

Off 8 10 1 *

Only for calibration/adjustment and linearization functions 8 10 2

Always on 8 10 3

The following menu code setting must be selected in order to obtain an ISO/GLP-compliantprintout/record:

With data ID codes 7 2 2

Important Note:ISO/GLP-compliant printouts/records are not generated when the factory setting, menu code 7 2 1, is active.With the “Auto print” data output parameter selected (code 6 1 4 or 6 1 5), only calibration/adjustment and linearization data is output.

Page 28: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

28

Printout/Record for Calibration/Adjustment Functions

A printout or record is generated at the end of the following functions:– all calibration and adjustment procedures– all linearization procedures

The printout can have the following lines:

—————————————MC1 - Sartorius : ManufacturerModel CC50002 : Mass comparator modelS/N 14203101 : Serial number of the mass comparatorId 4-32-1 : ID number (e.g., for entering a workstation or operator ID)—————————————Date : 24 May 02 : Current dateStart: 10:05:30 : Time at which the application startedCal. : Test : Calibration mode (in this case, “sensitivity test”)Diff.: -0.05 g : Weighed value for sensitivity testCal. : External : Type of adjustment function

(in this case, “external calibration”)Stat.: Complete : Status of adjustment/calibration or linearization functionEnd : 10:05:45 : Time application endedName : : Field for signature of the operator responsible—————————————

Set. :20000.00 g : Calibration weight value (only for “external calibration”)

Reset: Undoing All Menu Code Changes

The reset function lets you undo all menu code changes, which means the original factorydefaults (marked by an asterisk (“*”) in this manual) are restored.This can be useful if, for example, you do not know which menu codes have beenchanged.

Reset functionMenu code Factory setting Changed

Reset all codes 9 — — 1Off 9 — — 2 *

Page 29: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

29

Order no.

Data printer YDP03-0CEwith date/time and statistics functions, approved for use in legal metrologyPrint speed: approx. 1.5 lines per second Printer housing dimensions (W x D x H in mm): 150 x 138 x 43

Remote display(for connection to the interface port)– reflective YRD 12 Z– transmissive (for overhead projectors) YRD 13 Z

Plug-in AC adapter with selectable frequency 6971172

Foot switch with T-connector/ Universal remote control YFS 01switch for triggering the output, tare, F1 or F2 function(select function in the mass comparator operating menu)

T-connector YTC01

Weighing table YWT03

Draft shield YDS03C

Centermatic centering device YWP03C

External calibration weight (10000 g E2) YCW712 - 00(20000 g E2) YCW722 - 00

Crane without gripper (information on request) YLD 01 C

Gripper for stirrup weights (information on request) YLD 02 C

Hoist for 5 kg knob weight YAW 50Hoist for 10 kg knob weight YAW 51Hoist for 20 kg knob weight YAW 52Hoist for 50 kg knob weight YAW 52

Gloves (cotton) YAW 21Gloves (leather) YAW 22

Accessories

Page 30: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

30

Model CC30002 CC50002 CC50002with Centermatic

Accuracy class kg 0 – 41 0 – 61 0 – 51F1 kg 20 20 to 50 20 to 50F2 kg 10 to 20 10 to 50 10 to 50M1 kg 2 to 20 2 to 50 2 to 50

Maximum capacity Max kg 41 61 51

Electronic weighing range g 41 61 51

Readability mg 5 10 10

Tare range (subtractive) kg 41 61 51

Repeatability when load is0 - 30 kg mg <100 - 50 kg mg < 20 <15

External calibration weight kg 10 (Class E2) 20 (Class E2) 20 (Class E2)

Response time (average) s <10 <10 <10

Adaptation to ambient conditions and application requirements by selection of 1 of 4 optimized filter levels

Display update (depends on filter level selected) s 4 4 4

Operating temperature range °C +15°C to +30°C (+59°F to +86°F)

Power requirements, AC power source V~ via AC adapter TNG6; 230 or 115, + 15%...-20%

Frequency Hz 50–60

Power consumption (average) VA Max.: 18.7; average: 15

Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, /lb, tlc, mom, K, tol, bat, MS

Working space required mm2 700 x 400

Display and control unit net weight, approx. kg 3.1

Weighing platform (W + D + H) mm 400 + 3000 + 100

Display and control unit (W + D + H) mm 213 + 307 + 107

Centermatic centering device ( ∅ + H) mm 261 + 83

Weight, net kg 23 23 30

Built-in interface RS-232 C-S/V24-V28, RS-423/V10; 7-bit; even, odd, mark or space parity; transmission rates: 150 to 19,200 baud; 1 or 2 stop bits

Handshake Software/hardware

Specifications

Page 31: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

31

Page 32: Electronic Mass Comparator - scale service · Electronic Mass Comparator ... – The mass comparator must not be exposed during operation to potentially ... EN 61326-1 Electrical

Printed in Germany. Gedruckt auf chlorfrei gebleichtem PapierW1A000 · KTPublication No.: WCC6022-e02112

Sartorius AGWeender Landstrasse 94–10837075 Goettingen, Germany

Phone +49.551.308.0Fax +49.551.308.3289www.sartorius.com

Copyright by Sartorius AG, Goettingen, Germany. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior writtenpermission of Sartorius AG.The status of the information,specifications and illustrations in this manual is indicated by the date given below. Sartorius AG reserves the right to make changes to thetechnology, features,specifications and design of theequipment without notice.

Status: November 2002, Sartorius AG, Goettingen, Germany