e1unbref

Embed Size (px)

Citation preview

  • 7/30/2019 e1unbref

    1/30

    Agilent E7572A E1 (Unbalanced)

    4-Link Interface Module

    Reference Manual

  • 7/30/2019 e1unbref

    2/30

    Agilent Technologies5070 Centennial Boulevard

    Colorado Springs, CO 80919-2497 USA

    Copyright 2002

    Agilent Technologies, Inc.

    All rights reserved.

    Reproduction, adaptation,

    or translation without prior

    written permission is

    prohibited, except as

    allowed under the

    copyright laws.

    Part No. 37907-90234

    First edition, January 2002

    Printed in U.K.

    Trademarks

    Microsoft is a U.S.

    registered trademark of

    Microsoft Corporation.

    Windows and

    MS Windows are U.S.

    registered trademarks of

    Microsoft Corporation.

    Product Support

    Call your local Agilent

    Technologies

    representative, or:

    Tel: 1-800-452-4844

    Fax: 303-754-4802

    or call your local Agilent

    Sales and Service Office.

    DocumentationWarranty

    The information contained

    in this document is subject

    to change without notice.

    Agilent Technologies

    makes no warranty of any

    kind with regard to this

    material, including, but

    not limited to, the implied

    warranties or

    merchantability and fitnessfor a particular purpose.

    Agilent Technologies shall

    not be liable for errors

    contained herein or for

    incidental or consequential

    damages in connection

    with the furnishing,

    performance, or use of thismaterial.

    SAFETY

    Warning

    Warning denotes a hazard.

    It calls attention to a

    procedure, which if not

    correctly performed or

    adhered to could result in

    injury or loss of life. Do

    not proceed beyond a

    warning note until the

    indicated conditions are

    fully understood and met.

    Caution

    Caution denotes a hazard.

    It calls attention to a

    procedure, which if not

    correctly performed or

    adhered to could result in

    damage to or destructionof the instrument. Do not

    proceed beyond a caution

    note until the indicated

    conditions are fully

    understood and met.

    For details of safety, see

    the Safety Symbolssection on page 22.

  • 7/30/2019 e1unbref

    3/30

    Agilent E7572AE1 (Unbalanced) 4-Link InterfaceModule

    Reference Manual

    Agilent E7572AE1 (Unbalanced) 4-Link InterfaceModule

  • 7/30/2019 e1unbref

    4/30

    2

    The interface module provides the physical connection between the Signaling

    Advisor and the network you are testing. The interface module extracts the signaling

    information from up to four duplex links and adds a time stamp to the captured data

    before routing it to the mainframe for storage and analysis.

    Sources of information about the Signaling Advisor

    The documentation supplied is listed below:

    Mainframe Reference Manual - This manual describes the Signaling Advisor

    hardware. It also covers the connection and configuration of accessories via PC

    Cards in the PCMCIA slots.

    Users Manual - This manual tells you how to use the Signaling Advisor to monitor

    and emulate signaling data, view and interpret the results.

    Software Recovery Guide - This manual provides detailed instructions for rein-

    stalling Windows and the Signaling Advisor System Software. (Note: Your instru-

    ment is shipped with all software installed.)

    Interface Module Reference Manuals - Each interface module is provided with a

    manual which tells you how to install and verify the operation of the module.

    Unpacking and Installation Card - Identifies the items you will typically receive

    with your instrument. It also shows you how to proceed when you first receive your

    instrument.

    Quick Start and Quick Reference - A summary of useful operating information.

    Registration and Initial Setup for the Signaling Advisor - Contains the prelimi-

    nary procedures you need to follow when you first receive your instrument.

    Signaling Advisor Undercradle Installation Manual (Optional) - This manualcontains installation instructions on how to connect/disconnect the undercradle

    to/from Signaling Advisor.

    On-Line Help - The help comprises three parts: on-line help, context sensitive help,

    and copies of the manuals supplied with the instrument, support information, con-

    figuration information and frequently asked questions. Microsoft Windows on-line

    help is also available.

    Web Site - http://onenetworks.comms.agilent.com/

    Introduction

  • 7/30/2019 e1unbref

    5/30

    3

    List of Contents

    List of Contents

    Installing an Interface Module........................................................................... 5

    Connecting Interface Modules to Bearers ......................................................... 7

    BNC Cabling Information............................................................................... 8

    Lemo Cabling Information ............................................................................. 9

    BNC Mating Connector .................................................................................. 10

    Lemo Mating Connector................................................................................. 10

    Connecting to a Protected Monitor Point........................................................ 11Connecting to an Unprotected Monitor Point................................................. 12

    Terminating a Link in Monitor Mode............................................................. 13

    Connecting for End Emulation ....................................................................... 14

    Verifying Correct Operation.............................................................................. 16

    Description...................................................................................................... 16

    Verification Prerequisites................................................................................ 16

    Procedure ........................................................................................................ 17

    Product Identification ..................................................................................... 19Build Status..................................................................................................... 19

    Declaration of Conformity................................................................................. 19

    Statement of Compliance................................................................................... 20

    Electromagnetic Compatibility (EMC) Information....................................... 20

    Safety Information .......................................................................................... 20

    Noise Declaration (German)........................................................................... 20

    Declaration of Conformity................................................................................. 21

    Safety Symbols .................................................................................................. 22ESD Precautions ................................................................................................ 23

    Cleaning the Product.......................................................................................... 23

    Calibration Cycle ............................................................................................... 24

    Operating Environment ..................................................................................... 24

    Storage and Shipment........................................................................................ 25

    Returning for Service...................................................................................... 25

    Weight and Dimensions.................................................................................. 26

    Web Site for Additional Product Information ................................................... 26

  • 7/30/2019 e1unbref

    6/30

    4

  • 7/30/2019 e1unbref

    7/30

    5

    Installation and General Information

    Installing an Interface Module

    Installing an Interface Module

    You can have up to two modules installed in your Signaling Advisor. However, the

    number of modules can be extended to four if you have an undercradle fitted. The

    modules operate in either slot. If only one module is fitted to your Signaling Advisor

    then a Filler Panel must be fitted to the empty slot.

    WAR NI NG If this product is not used as specified, the protection provided by the product

    could be impaired. This product must be used in a normal condition only (in

    which all means for protection are intact).

    CAUTION To prevent possible system errors, you must turn off the power to the Signaling

    Advisor before installing or removing a module.

    To install a module (or Filler Panel) into a Signaling Advisor:

    1 Ensure the Module Release Latch is in the unlocked position.

    2 Fit and slide the module (or Filler Panel) into the opening on the side of the

    instrument, see the following figure.

    3 Push the module firmly into place until it is seated (the Module Release Latch

    automatically moves to the locked position). If you are fitting a Filler Panel, you will

    need to manually move the Module Release Latch to its locked position.

    INTERFACE MODULE

    LATCH

  • 7/30/2019 e1unbref

    8/30

    6

    Installation and General Information

    Installing an Interface Module

    To remove an interface module (or Filler Panel) from a Signaling Advisor:

    1 Depress the left end of the Module Release Latch (labeled PUSH TO UNLOCK)

    then slide the latch to the left end (unlocked position).

    2 Pull the module (or Filler Panel) out of the module.

    To install an interface module into an undercradle:

    1 Switch off the power to the instrument.

    2 Fit and slide the module into the opening of the undercradle.

    3 Push it into place until it is seated, see the figure below.

    To remove an interface module from an undercradle:

    1 Slide the Module Release Latch towards the left.

    2 Pull the module out of the undercradle.

    For more information, refer to the Installation Manual that is shipped with theundercradle.

    Pull knob goes toward

    outside of Signaling Advisor

    UP

  • 7/30/2019 e1unbref

    9/30

    7

    Installation and General Information

    Connecting Interface Modules to Bearers

    Connecting Interface Modules to Bearers

    NOTE Beginning January 2002, all shipments of this interface module will be supplied with

    BNC connectors. Information about Lemo connectors and cabling are provided in

    this manual for reference only.

    Connections to the network are made with the eight Lemo connectors or eight BNC

    connectors on the interface module. Each pair of connectors provides a full duplex link.

    BNC Connectors

    Lemo Connectors

    CAUTION When connecting or disconnecting, ensure that you are grounded or, make contact

    with the metal surface of the mainframe with your free hand to bring you, the module,

    and the Signaling Advisor to the same static potential. Modules remain susceptible

    to ESD damage while the module is installed in the Signaling Advisor. Additional

    ESD precautions are required when servicing, see page 23.

    RXa Ports - Receive E1 unbalanced signals.

    RXb/TX Ports - Receive E1 unbalanced signals in Monitor mode, and transmit E1unbalanced signals in Emulate mode.

    Bearer 4Bearer 3Bearer 2Bearer 1

    Bearer 2 Bearer 3 Bearer 4Bearer 1

  • 7/30/2019 e1unbref

    10/30

    8

    Installation and General Information

    Connecting Interface Modules to Bearers

    BNC Cabling Information

    Description Length

    Agilent Part

    No.

    Loopback cable, BNC to BNC (male). 230 mm 8120-2682

    75 unbalanced male BNC to BNC (push fit). 3 m E4555A

    75 unbalanced male BNC to Siemens 1.6/5.6 (male) 3 m E4554A

  • 7/30/2019 e1unbref

    11/30

    9

    Installation and General Information

    Connecting Interface Modules to Bearers

    Lemo Cabling Information

    Description Length

    Agilent Part

    No.

    Loopback cable, Lemo to Lemo. 300 mm 15981-60001

    Note: This cable is for self test on 75 ohm and 120 ohm balanced interface modules.

    75 unbalanced male BNC to Lemo. 3 m 15982-60001

    75 unbalanced female BNC to Lemo. 300 mm 15986-60001

    75 unbalanced Siemens 1.6/5.6 (female) to Lemo 300 mm 15994-60001

    75 unbalanced Siemens 1.6/5.6 (male) to Lemo 3 m 15995A-60001

    75 unbalanced Lemo to Lemo. Can be used for loopback

    tests.

    6 m 15985-60001

    You can make two customized cables from this cable by cutting it in half and fitting suitable

    mating connectors.

    Lemo connector cable connection details:

    pin 1: not connected

    pin 2: core (signal)

    pin 3: not connected

    Sheath connected to Lemo body at both ends (ground)

  • 7/30/2019 e1unbref

    12/30

    10

    Installation and General Information

    Connecting Interface Modules to Bearers

    BNC Mating Connector

    The following connector mates with the interface module connectors:

    *For 75 coax cable - RG 59.

    Lemo Mating Connector

    The following connector mates with the interface module connectors:

    *For 75 coax cable - RG 179B/U.

    Port Type Tajimi Part No.

    RXa and RXb/TX Tajimi *TBNCS-706-00

    Port Type Lemo Part No.

    RXa and RXb/TX Lemo *FGG.0B.303.CLAD52 Z

    sleeve:GMA.0B.025.DN

  • 7/30/2019 e1unbref

    13/30

    11

    Installation and General Information

    Connecting Interface Modules to Bearers

    Connecting to a Protected Monitor Point

    Set the Signaling Advisor bearer operating mode to 20 30 dB Terminated Monitor

    when connecting to a monitor point that attenuates the line signal by 20 to 30 dB. The

    Signaling Advisor terminates the cable from the monitor point properly and causes

    minimal reflections. Maximum cable length depends on cable quality and is specified

    as 6 dB loss at 1024 kHz, typically this will be more than 100 m. For more information

    on configuring E1 bearer settings, refer to the instruments On-Line Help.

    BNC Protected Monitor Point

    Lemo Protected Monitor Point

  • 7/30/2019 e1unbref

    14/30

    12

    Installation and General Information

    Connecting Interface Modules to Bearers

    Connecting to an Unprotected Monitor Point

    Set the Signaling Advisor bearer operating mode to Bridged 2k4 ohms when

    connecting directly to an active line (the high input impedance allows you to make the

    connection without affecting the signal). The cable length from the Signaling Advisor

    to the line being monitored should not exceed 3 m (longer cables cause reflections

    which may produce errors and lead to alarms).

    For more information on configuring E1 bearer settings, refer to the instruments On-

    Line Help.

    BNC Unprotected Monitor Point

    Lemo Unprotected Monitor Point

  • 7/30/2019 e1unbref

    15/30

    13

    Installation and General Information

    Connecting Interface Modules to Bearers

    CAUTION Do not connect to an unprotected monitor point when the Signaling Advisor bearer

    operating mode is set to 0 dB or 20 30 dB Terminated Monitor. This will load the

    network.

    Terminating a Link in Monitor Mode

    If you need to use the Signaling Advisor to terminate a link before you can monitor

    data, then set the Signaling Advisor bearer operating mode to 0 dB Terminated

    Monitor. Maximum cable length depends on cable quality and will typically beseveral hundred meters.

    For more information on configuring the Signaling Advisors E1 bearer settings in

    Monitor mode, refer to the instruments On-Line Help.

  • 7/30/2019 e1unbref

    16/30

    14

    Connecting for End Emulation

    Set the Signaling Advisor bearer operating mode to End Emulate when you

    want to send/receive messages to/from Network Equipment. The Signaling

    Advisor terminates the Receive side of the Bearer and generates E1 signals on

    the Transmit side. Maximum cable length depends on cable quality and will

    typically be several hundred meters.

    For more information on configuring the Signaling Advisors E1 bearer settings

    in Emulate mode, refer to the instruments On-Line Help.

    BNC End Emulation

  • 7/30/2019 e1unbref

    17/30

    15

    Installation and General Information

    Connecting Interface Modules to Bearers

    Lemo End Emulation

  • 7/30/2019 e1unbref

    18/30

    16

    Installation and General Information

    Verifying Correct Operation

    Verifying Correct Operation

    Description

    You verify correct operation in Post Capture mode, by selecting All Tests from the

    Signaling Advisors built-in Self Test function. When you select All Tests, the

    program automatically detects which interface module(s) are housed in the Signaling

    Advisor, it then selects and runs the appropriate self tests. It takes approximately 15

    minutes to verify each interface module. The default settings are used which means

    tests will continue to completion (even if an error is detected), and all results (Pass or

    Fail) will be logged to disk.

    For a detailed description of Self Test, refer to the instruments On-Line Help.

    CAUTION To prevent network disruptions, ensure the Signaling Advisor is disconnected from

    your network before you run Self Tests.

    Verification Prerequisites

    During Self Test you will be prompted to fit external loopback cables.

    To minimize verification time ensure that you have following cables available before

    you start the Self Test:

    E1 loopback cable(s), part number Agilent 8120-2682 (BNC connectors) or Agilent

    15981-60001 (Lemo connectors).

    Clock loopback cable (connects to the Signaling Advisor CLOCK/TRIG IN and

    OUT ports), part number Agilent 8120-8820.

    Telephone handset (this is an optional connection), part number Agilent 15722A.

  • 7/30/2019 e1unbref

    19/30

    17

    Installation and General Information

    Verifying Correct Operation

    Procedure

    1 Double click the Signaling Advisor icon in the desktop.

    2 Click the Post Capture button.

    3 Click on the Tools menu, then select Self Tests.

  • 7/30/2019 e1unbref

    20/30

    18

    Installation and General Information

    Verifying Correct Operation

    4 Click the All Tests button, then clickRun Tests.

    5 Test progress and results can be viewed on the screen. A full set of results is available

    from the specified log file.

    If any of the Self Tests fail, see Returning for Service, Page 25.

    6 Click the Close button to exit Self Test and return to the Monitor application.

  • 7/30/2019 e1unbref

    21/30

    19

    Installation and General Information

    Product Identification

    Product Identification

    Attached to the product is a serial number plate. The serial number on this label is in

    the form GB00000101. The option number on this label indicates which hardware

    options are fitted.

    The contents of this manual apply to products with serial numbers GB00000101 and

    above.

    Build Status

    Also attached to the module side panel is a BUILD STATUS label which indicates the

    revision of hardware built into this product.

    If you contact Agilent for service on this product the Agilent engineer may request the

    Build Status number.

    Declaration of Conformity

  • 7/30/2019 e1unbref

    22/30

    20

    Installation and General Information

    Statement of Compliance

    Statement of Compliance

    Electromagnetic Compatibility (EMC) Information

    This product conforms with the protection requirements of European Council Directive

    89/336/EEC for Electromagnetic Compatibility (EMC).

    The conformity assessment requirements have been met using the technicalConstruction file route to compliance, using EMC test specifications EN 55011:1991

    (Group 1, Class A) and EN 50082-1:1992.

    In order to preserve the EMC performance of the product, any cable which becomes

    worn or damaged must be replaced with the same type and specification.

    See the Declaration of Conformity on the next page.

    Safety Information

    This product has been designed and tested in accordance with publication

    EN61010-1(1993) / IEC 61010-1(1990) +A1(1992) +A2(1995) Safety Requirements

    for Electrical Equipment for Measurement, Control and Laboratory Use, and has been

    supplied in a safe condition. The instruction documentation contains information and

    warnings which must be followed by the user to ensure safe operation and to maintain

    the instrument in a safe condition.

    Noise Declaration (German)

    LpA

  • 7/30/2019 e1unbref

    23/30

    21

    Installation and General Information

    Declaration of Conformity

    Declaration of Conformity

    DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN45014

    Manufacturers Name: Agilent Technologies UK Limited

    Telecomms Networks Test Division

    South Queensferry

    West Lothian, EH30 9TG

    Scotland, United Kingdom

    Manufacturers Address:

    Declares that the product

    Product Name: Unbalanced E1 Link Interface Module

    Model Number: E7572A

    Product Options: This declaration covers all options of the above product as detailed inTCF A-5951-9852-01.

    EMC:

    Conforms with the protection requirements of European Council Directive 89/336/EEC on the approximation of the

    laws of the member states relating to electromagnetic compatibility, against EMC test specifications EN 55011:1991

    (Group 1, Class A) and EN 50082-1:1992.

    As Detailed in: Electromagnetic Compatibility (EMC)

    Technical Construction File (TCF) No. A-5951-9852-01

    Assessed by: DTI Appointed Competent Body

    EMC Test Centre,GEC-Marconi Avionics Ltd.,

    Maxwell Building,

    Donibristle Industrial Park,

    Hillend,

    Dunfermline

    KY11 9LB

    Scotland, United Kingdom

    Technical Report Number:6893/2200/CBR, dated 21 August 1997

    Safety:

    The product conforms to the following safety standards:

    IEC 61010-1(1990) +A1(1992) +A2(1995) / EN 61010-1(1993)

    The product herewith complies with the requirements of the General Product Safety Directive 92/59/EEC.

    South Queensferry, Scotland. 01 June 2000

    Installation and General Information

  • 7/30/2019 e1unbref

    24/30

    22

    Installation and General Information

    Safety Symbols

    Safety Symbols

    Safety Symbols Description

    The Instruction Documentation Symbol. The product is marked with

    this symbol when it is necessary for the user to refer to the

    instructions in the supplied documentation.Indicates hazardous voltages.

    Indicates the field wiring terminal that must be connected to earth

    ground before operating the equipment - protects against electrical

    shock in case of fault.

    Frame or chassis ground terminal - typically connects to the

    equipments metal frame.

    WARNING Warning denotes a hazard. It calls attention to a procedure, which if

    not correctly performed or adhered to could result in injury or loss of

    life. Do not proceed beyond a warning note until the indicated

    conditions are fully understood and met.

    CAUTION Caution denotes a hazard. It calls attention to a procedure, which if

    not correctly performed or adhered to could result in damage to or

    destruction of the instrument. Do not proceed beyond a caution noteuntil the indicated conditions are fully understood and met.

    This symbol indicates that a device, or part of a device, may be

    susceptible to electrostatic discharges (ESD) which can result in

    damage to the product. Observe ESD precautions given on the

    product, or its user documentation, when handling equipment

    bearing this mark.

    The CE mark shows that the product complies with all relevant

    European Legal Directives.

    The C-Tick mark is a registered trademark of the Australian

    Communications Authority. This signifies compliance with the

    Australian EMC Framework Regulations under the terms of the

    Radio Communications Act of 1992.

    ISM 1-A This is a symbol of an Industrial, Scientific, and Medical Group 1

    Class A product.

    ICES/NMB-001 This ISM device complies with Canadian ICES-001.Cet appareil ISM est conforme la norme NMB-001 du Canada.

    Installation and General Information

  • 7/30/2019 e1unbref

    25/30

    23

    Installation and General Information

    ESD Precautions

    ESD Precautions

    CAUTION When making connections to the modules, review Connecting Interface Modules to

    Bearers, Page 7.

    The module contains components sensitive to electrostatic discharge.

    To prevent component damage, carefully follow the handling precautions presentedbelow.

    The smallest static voltage most people can feel is about 3500 volts. It takes less than

    one tenth of that (about 300 volts) to destroy or severely damage static sensitive

    circuits. Often, static damage does not immediately cause a malfunction but

    significantly reduces the components life. Adhering to the following precautions will

    reduce the risk of static discharge damage. Keep the module in its conductive storage box when not installed in the Signaling

    Advisor. Save the box for future storage of the module.

    Before handling the module, select a work area where potential static sources are

    minimized. Avoid working in carpeted areas and non-conductive chairs. Keep body

    movement to a minimum. Agilent recommends that you use a controlled static work-

    station.

    Handle the module by its front-panel. Avoid touching any components or edge con-

    nectors. When you install the module, keep one hand in contact with the protectivebag as you pick up the module with your other hand. Then, before installing the mod-

    ule, ensure that you are grounded or make contact with the metal surface of the Sig-

    naling Advisor with your free hand to bring you, the module, and the Signaling

    Advisor to the same static potential. This also applies whenever you connect/dis-

    connect cables on the front-panel.

    Cleaning the Product

    Use a soft, clean damp cloth to clean the outer surfaces.

  • 7/30/2019 e1unbref

    26/30

    24

    Calibration Cycle

    This product does not require periodic calibration. Operational verification can

    be confirmed by running the built-in Self Test, see page 16.

    Operating Environment

    This product is designed for indoor use only.

    DO NOT operate the product in an explosive atmosphere or in the presence of

    flammable gasses or fumes.

    This product may be operated in environments within the following limits:

    Temperature: +5o C to +40o C

    Altitude up to 4,600 m

    Humidity 20% to 80% relative humidity non condensing to 40oC.

    This product should be protected from temperature

    extremes which may cause condensation within the product

    enclosure.

    To ensure adequate cooling do not obstruct air vents in the product cabinet. Air

    vents are located at the front, on the underside and both sides of the product.

    CAUTION This product is designed for use in Installation Category II and Pollution

    Degree 2 per IEC 61010 and 60664 respectively.

    Installation and General Information

  • 7/30/2019 e1unbref

    27/30

    25

    Storage and Shipment

    Storage and Shipment

    This product may be stored or shipped in environments within the following limits:

    Temperature -25o C to +60o C

    Altitude Up to 15,250 m

    Humidity 10% to 90% relative humidity to 60oC

    This product should be protected from temperature extremes which could cause

    internal or external condensation.

    Returning for Service

    When returning this product for service, please include the Signaling Advisor and

    accessory cables.

    Tagging for Service. If the module is being returned to Agilent for service, please

    complete a repair tag and attach it to the module. Ensure the following information is

    sent with it:

    Description of the problem (include the self-test results file, error messages or

    specific performance data related to the problem).

    Whether the problem is constant or intermittent.

    Name and phone number of technical contact person.

    Return address.

    Model and option number of returned Signaling Advisor and module(s).

    Full serial numbers of returned Signaling Advisor and module(s).

    Mark shipping container FRAGILE to ensure careful handling.

    Original Packaging. Containers and materials identical to those used in factory

    packaging are available from Agilent offices.

    Other Packaging. The following general instructions should be followed when

    repackaging with commercially available materials:

    Wrap instrument in heavy paper or plastic.

    Use a strong shipping container. A double wall carton made of 350 pound test

    material is adequate.

    Installation and General Information

  • 7/30/2019 e1unbref

    28/30

    26

    Web Site for Additional Product Information

    Use a layer of shock absorbing material 70 to 100 mm (3 to 4 inch) thick, around all

    sides of the instrument to provide firm cushioning and prevent movement inside thecontainer. Do not rely on pellets for shock absorption, as during transit the

    instrument will sink to the bottom of the pellets.

    Seal shipping container securely.

    Weight and Dimensions

    Weight: 0.5 kg

    Dimensions: (height x width x depth)

    21 mm x 122 mm x 150 mm

    Web Site for Additional Product Information

    You can use the Signaling Advisor to access the Web (requires a suitable LAN or

    modem connection), or you can use any suitable PC with access to the Web.

    http://onenetworks.comms.agilent.com/

    S l d S i Offi

  • 7/30/2019 e1unbref

    29/30

    Sales and Service Offices

    An up-to-date list of Agilent Offices is available through

    the Agilent Website at URL:

    http://onenetworks.comms.agilent.com/

  • 7/30/2019 e1unbref

    30/30

    In this book

    You will be shown how to install

    an E7572A E1 (Unbalanced)

    4-Link Interface Module into a

    Signaling Advisor. Also provided

    is cabling information and a

    verification procedure.

    Printed in UK 01/02

    37907-90234