13
1 Development of a tomographic system for online dose measurements in BNCT Minsky D.M. 1,2 , Valda A.A. 1,2 , Burlon A.A. 1,2,3 , Kreiner A.J. 1,2,4 y Somacal H. 1,2 1. Departamento de Física (CAC, CNEA) 2. Escuela de Ciencia y Tecnología (UNSAM) 3. Fundación J.B. Sauberán 4. CONICET

Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

Embed Size (px)

Citation preview

Page 1: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

1

Development of a tomographic system for online dose

measurements in BNCTMinsky D.M.1,2, Valda A.A. 1,2, Burlon A.A. 1,2,3, Kreiner A.J. 1,2,4 y Somacal H. 1,2

1. Departamento de Física (CAC, CNEA) 2. Escuela de Ciencia y Tecnología (UNSAM)

3. Fundación J.B. Sauberán4. CONICET

Page 2: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

2

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Epithermalneutrons

Figure adapted from:http://web.mit.edu/nrl/www/bnct/info/description/description.html

[ ][ ] 1:4

ssueHealthy ti BTumor B

10

10≈

Page 3: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

3

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

10BThermalneutron

10B + n → 11B*→ 7Li* + 4He

7Li(0.84 MeV)

4He(1.47 MeV)

γ (478 keV)

5 µm

9 µm

↓ t ~ 10-14 s

Li + γ (478 keV)

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

γ-ray Telescope SystemNetherlands Energy Research Foundation, Petten, The NetherlandsEindhoven University of Technology, Eindhoven, The NetherlandsNuclear Research and Consultancy Group, Petten, The Netherlands

WFAR Verbakel, F Stecher-Rasmussen, Nucl. Instr. and Meth. in Phys. Res. A, 394, 163-172, 1997.WFAR Verbakel, W Sauerwein, K Hideghety, F Stecher-Rasmussen, Int. J. Radiation Oncology Biol. Phys. 55(3), 743–756, 2003.

Related works: γ-ray telescopeand PG-SPECT

PG-SPECT (Prompt Gamma-ray SPECT)Research Reactor Institute, Kyoto University, JapanResearch Institute for Radiation Biology and Medicine, Hiroshima Univ., Japan

T Kobayashi, Y Sakurai, M Ishikawa, Med. Phys. 27(9), 2124-2132, 2000.

Page 4: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

4

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Collimator design

Requirements:• Trans-septal penetration

less than 6%• Spatial resolution 1cm

at 20 cm from detector

Results:• t=2 mm• ∅=4.4 mm• e=80 mm

e

t∅

Collimator thickness

Geo

met

rical

effic

ienc

y[%

]

Collimator Thickness (e) [mm]

Geo

met

rical

effic

ienc

y[%

]

Spatial Resolution [mm]

Page 5: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

5

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

System numerical simulation

• System matrix (influence of each pixel on each detector for every projection).

MonteCarlo simulation with MCNP-4C of:

Some data of the simulations:

• Image consisting in 21*21 pixels• Pixels of 1cm*1cm size• 37 detectors in each projection• 20 projections regularly spaced

by 9O

Page 6: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

6

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

P1 P2 P3 P4

P5 P6 P7 P8

P9 P10 P11 P12

P13 P14 P15 P16

B6

B7

B8

B9

B10

B1 B2 B3 B4 B5

System matrix calculation

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

…………

…m34m32m31

…m23m22m21

…m13m12m11

p3

p2

p1

b3

b2

b1

=.

System matrix calculation

Page 7: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

7

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

………

…m34m32

…m23m22

…m13m12

0

0

1

b3-1

b2-1

b1-1

=.

m31

m21

m11

b3-1

b2-1

b1-1

System matrix calculation

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

……...

…m34b1-2

…m23b1-2

…m13b1-1

0

1

0

=.

m33

m23

m12

b3-2

b2-2

b1-2

b3-2

b2-2

b1-2

System matrix calculation

Page 8: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

8

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Phantoms numerical simulations

Page 9: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

9

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

WithoutBackground

→→→→

→→→→

↑↑↑↑Simulatedsource

↑↑↑↑Simulatedsinogram

↑↑↑↑Reconstructed

image

Numerical simulations of phantoms

WithBackground

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Influence of statistical fluctuations

Page 10: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

10

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Influence of tumor size

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Page 11: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

11

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

TANDARLaboratory

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Previous experimental measurements

Page 12: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

12

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Experimental setup under characterization

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Page 13: Development of a tomographic system for online dose ...scoccola/iguazu/NPA/04_NPA4-Minsky.pdf · 3 VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic

13

VI Latin American Symp. on Nuc. Phys. and Applications Development of a tomographic system for online dose measurements in BNCT

Conclusions and future work

• A tomographic SPECT system for online dose measurements in BNCT have been designed

• The whole system has been simulated by means of MonteCarlo calculations

• Sinograms from numerical phantoms have been simulated and used for image reconstruction

• Statistical fluctuations must be kept under 3-5%• The influence of the image reconstruction algorithm on

quantitative concentration measurements will be studied• The experimental setup is under characterization and

images will be acquired in the near future