Journal of Automatic Chemistry, Vol. 14, No. 5 (September-October 1992), pp. 163-167 Determination of nutrients in seawater by segmented-flow analysis with higher analysis rate and reduced interference on ammonia M. Jodo Tokyo Kyuei, 6906-10 Tsurugamaru Shiba Kawaguchi, Saitama-ken, Japan K. Kawamoto, M. Tochimoto Bran + Luebbe KK, Shouei Bldg, 22-8, 4-Chome, Sendagaya, Shibuya-ku, Tokyo 151, Japan’ and $. C. Coverly* Bran + Luebbe, PO Box 1360, W-2000 Norderstedt, German A microbore continuous-flow analyser was used to determine ammonia, nitrite, nitrate and phosphate in seawater. Ninety samples per hour were measured by optimizing the hydraulic conditions and using a cadmium coil for nitrate reduction. The analysis range was extended and the detection limit was reduced by using two analytical ranges with automatic range-changing. Interference from magnesium on ammonia was investigated and reduced by using a different complexing agent. The results correlated well with a Japanese reference method. Introduction Since the introduction of air-segmented continuous-flow analysis (CFA) in 1958 there have been continual improvements in system performance, mainly in analysis rate and the automation of activities subordinate to the analysis, such as automatic data handling, sample dilution and reagent change-over. The analysis rate in segmented-flow systems is inversely dependent on dispersion. Reductions in dispersion have centred around reductions in the internal diameter of the reaction tubing, from 2"4 mm in the original Technicon AutoAnalyzer I to 2 rnrn in the AutoAnalyzer II and mm in modern instruments. The latest systems have been designed with a proper understanding of the way in which dispersion, tubing diameter and flow rate are interrelated, following Snyder’s definitive description ]. As Snyder has also explained, a further reduction in dispersion in commercial systems is unlikely for practical reasons [2]. Segmented-flow systems run faster than flow- injection (FIA) systems operating at the highest possible sensitivity with slow reactions such as colorimetric ammonia and phosphate, due to the time needed to introduce a large injection volume into the manifold and the time required for the signal to return to near zero in a medium-dispersion FIA system. A major source of dispersion in the determination of nitrate in seawater is the cadmium reduction stage, which Author to whom correspondence should be addressed. commonly uses a column packed with cadmium granules. Patton [3] first described the use of a cadmium coil for analysing nitrate in fresh water. The coil allows the segmented stream to flow uninterrupted through the reduction stage. This more than doubled the analysis rate for seawater samples compared to methods using a packed column. Previous experience using the AutoAnalyzer ammonia method [4] showed negative values in surface samples far from the coast. The method uses citrate and tartrate buffers to reduce magnesium interference similar to the method described by Hansen and Grasshof [5]. Experi- ments showed that this was not due to impurities in the standards and base-line reference solution, but, rather, to a suppression of the ammonia colour reaction due to magnesium. The complexing agents tartarte, citrate and EDTA were found to be effective against calcium but did not completely eliminate interference from magnesium. As described below, 1,2 cyclohexane diamine tetraacetic acid (CyDTA) proved to be a more effective complexing agent for magnesium. To attain low detection limits in automated methods, it is necessary to use a high sample-to-reagent ratio, a flowcell with a long path-length and high signal amplification. This restricts the analytical range so that high samples have to be diluted. The range-expanding facility of the analyser is used to maintain a wide analytical range and achieve lower detection limits. After optimizing the analysis conditions for nitrite, nitrate, ammonia and phosphate, the results from the automatic analyser were compared to those from a standard manual method [6]. The correlation (R) was greater than 0"99 for each parameter. Materials and methods Analyser A four-channel TRAACS 800 with random-access sampler and automatic dilution facility was used (Bran + Luebbe, Norderstedt, Germany). Reagents All chemicals were of analytical grade. Deionized water (<1 bts/cm) was passed through an activated carcoal column and a 0"4 btm filter. The numbers below in square brackets refer to the flow diagram (figure 1). 163 0142-0453/92 $3.00 ( 1992 Taylor & Francis Ltd.

Determination seawater by segmented-flow analysis higher

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Determination seawater by segmented-flow analysis higher

Journal of Automatic Chemistry, Vol. 14, No. 5 (September-October 1992), pp. 163-167

Determination of nutrients in seawater bysegmented-flow analysis with higheranalysis rate and reduced interference onammonia

M. JodoTokyo Kyuei, 6906-10 Tsurugamaru Shiba Kawaguchi, Saitama-ken, Japan

K. Kawamoto, M. TochimotoBran + Luebbe KK, Shouei Bldg, 22-8, 4-Chome, Sendagaya, Shibuya-ku, Tokyo151, Japan’

and $. C. Coverly*Bran + Luebbe, PO Box 1360, W-2000 Norderstedt, GermanA microbore continuous-flow analyser was used to determineammonia, nitrite, nitrate and phosphate in seawater. Ninetysamples per hour were measured by optimizing the hydraulicconditions and using a cadmium coil for nitrate reduction. Theanalysis range was extended and the detection limit was reduced byusing two analytical ranges with automatic range-changing.Interference from magnesium on ammonia was investigated andreduced by using a different complexing agent. The resultscorrelated well with a Japanese reference method.

Introduction

Since the introduction of air-segmented continuous-flowanalysis (CFA) in 1958 there have been continualimprovements in system performance, mainly in analysisrate and the automation of activities subordinate to theanalysis, such as automatic data handling, sampledilution and reagent change-over.

The analysis rate in segmented-flow systems is inverselydependent on dispersion. Reductions in dispersion havecentred around reductions in the internal diameter of thereaction tubing, from 2"4 mm in the original TechniconAutoAnalyzer I to 2 rnrn in the AutoAnalyzer II andmm in modern instruments. The latest systems have

been designed with a proper understanding of the way inwhich dispersion, tubing diameter and flow rate areinterrelated, following Snyder’s definitive description ].As Snyder has also explained, a further reduction indispersion in commercial systems is unlikely for practicalreasons [2]. Segmented-flow systems run faster than flow-injection (FIA) systems operating at the highest possiblesensitivity with slow reactions such as colorimetricammonia and phosphate, due to the time needed tointroduce a large injection volume into the manifold andthe time required for the signal to return to near zero in amedium-dispersion FIA system.

A major source of dispersion in the determination ofnitrate in seawater is the cadmium reduction stage, which

Author to whom correspondence should be addressed.

commonly uses a column packed with cadmium granules.Patton [3] first described the use of a cadmium coil foranalysing nitrate in fresh water. The coil allows thesegmented stream to flow uninterrupted through thereduction stage. This more than doubled the analysis ratefor seawater samples compared to methods using a

packed column.

Previous experience using the AutoAnalyzer ammoniamethod [4] showed negative values in surface samples farfrom the coast. The method uses citrate and tartratebuffers to reduce magnesium interference similar to themethod described by Hansen and Grasshof [5]. Experi-ments showed that this was not due to impurities in thestandards and base-line reference solution, but, rather, toa suppression of the ammonia colour reaction due tomagnesium. The complexing agents tartarte, citrate andEDTA were found to be effective against calcium but didnot completely eliminate interference from magnesium.As described below, 1,2 cyclohexane diamine tetraaceticacid (CyDTA) proved to be a more effective complexingagent for magnesium.

To attain low detection limits in automated methods, it isnecessary to use a high sample-to-reagent ratio, a flowcellwith a long path-length and high signal amplification.This restricts the analytical range so that high sampleshave to be diluted. The range-expanding facility of theanalyser is used to maintain a wide analytical range andachieve lower detection limits.

After optimizing the analysis conditions for nitrite,nitrate, ammonia and phosphate, the results from theautomatic analyser were compared to those from astandard manual method [6]. The correlation (R) wasgreater than 0"99 for each parameter.

Materials and methods

AnalyserA four-channel TRAACS 800 with random-accesssampler and automatic dilution facility was used (Bran +Luebbe, Norderstedt, Germany).

ReagentsAll chemicals were of analytical grade. Deionized water(<1 bts/cm) was passed through an activated carcoalcolumn and a 0"4 btm filter. The numbers below in squarebrackets refer to the flow diagram (figure 1).

1630142-0453/92 $3.00 ( 1992 Taylor & Francis Ltd.

Page 2: Determination seawater by segmented-flow analysis higher

M. Jodo et al. Determination of nutrients in seawater by segmented-flow analysis

NH-N20T H-B 5T

45"C

WASTE

Flow cell

0.5ram ID x 30ram

630nm

PO -P

o d Do

Flow cell

0.5ram ID x 50ram

880nm

NO+I,IO -N

20T

0000WASTE

Flow cell

0.5mm ID x 30ram

550rim

N0-N

5T

000

WASTE

WASTE

RED/RED (323) Debubble

GRN/GRN 635 R-1Air

ORN/GRN (50) Dil Sampl

AirR-2

oOR/,WST(tZI, R-3

GRN/GRN (6351

Sam Soln

20T 5T

0000 000

WHT/WHT 258 DebubbleWASTE . . ,,,,

’RED/RED(3231 R-5I-",. ,Air

G/GRN(635)Sam le

OORN/YEL(79)Dil Sample

Air

00N/,ORN(192) ,R-4

WASTE

Flow cell

0.5ram ID x 50ram

550nm

Sample Was.h Soln

E’r./ET- (43 i) a-9

5T .. Air (o Ifree

000 ,--’I,,G/GI:%II(635’’) ;’ample

WASTE < OYEL/YEL (4311

WASTE

q’lAir, (01 free)

WASTE D/RED (323)

.O YELl .B.LU 484 R-6

d)O Air(O, free)

O-"RN/RN 192 R.esamp,le

(d aeductor Air

O ORN/EL (79) R-8O

To Air Valve

To Sampler

oRED/RED(323) Air(0 free)

GRN/.,GRN(6351 Wash Soln

Figure 1. Flow diagram of the system.

Ammonia: Stock CyDTA solutioncontains:5 g sodium hydroxide5 g sodium carbonate113 g 1,2 cyclohexane diamine tetraacetic acid(CyDTA).

Dissolve in the stated order, then adjust to pH 10.

Stock nitroprusside11 contains 0"6 g sodium nitroprusside (disodiumnitrosylpentacyanoferrate-I I dihydrate).

Working CyDTA JR1 on flow diagram]Mix 240 ml stock CyDTA, 60 ml stock nitroprussideand 0"5 ml Triton X-100 (40% solution in ethanol orpropan-2-ol).

164

Page 3: Determination seawater by segmented-flow analysis higher

M. Jodo et al. Determination of nutrients in seawater by segmented-flow analysis

Alkaline phenol reagent [R2]contains:47"5 g phenol95 ml 5M NaOH solution

Store in a plastic container in the refrigerator. Stablefor one week.

Sodium hypochlorite [R3]After determining the free chlorine content ofcommercially available hypochlorite solution byiodometric titration, prepare a working solutioncontaining 1-1"5% chlorine.Prepare fresh daily.

Phosphate:Stock molybdate solution

contains:8 g sodium molybdrate dihydrate0-17 g potassium antimony tartrate50 ml sulphuric acid (conc.).

Working molybdate reagent [R4]Add 0"8 g ascorbic acid and 2 ml 15% sodiumdodecyl sulphate (purest grade) to 100 ml stockmolybdate.Prepare fresh daily.

Phosphate dilution JR5]contains:30 g sodium chloride10 ml sulphuric acid5 ml sodium dodecyl sulphate (15% solution inwater).

Nitrate:

Imidazole [R6]contains:4 g imidazoleml Triton X-100 (40% solution in propan-2-ol).

Adjust to pH 7"2.

Sulfanilamide JR7]contains:10 g sulfanilamide100 ml hydrochloric acid.

NEDD [RS]contains g N-l-naphthylethylenediamine

dihydrochloride.Nitrate dilution water [Rg]

contains 3 ml Triton X-100 (40% inpropan-2-ol).

Standards

Primary standards with a concentration of 10 mmol/1were prepared from ammonium sulphate, sodium nitrite,potassium nitrate and potassium dihydrogen orthophos-phate. An intermediate standard containing ammonia(500 tmol/1, nitrite (100 btmol/1), nitrate (300 btmol/1)and phosphate (100btmol/1) was prepared in 10%sodium chloride solution. This intermediate standardwas diluted 20 to 200 times with water to produce thedaily working standards.

Analytical conditions

The system was run at a sampling rate of 90/h and asample wash ratio of4 1. Oxygen was removed from the

Table 1. Results for seawater samples in gmol ammonia/l.

Sample Sample SampleA B C

Tartrate method -0.41 -0"23 -0"35CyDTA method 0" 11 0"35 0"21

Table 2. Effect ofdifferent Mg levels on ammonia determination,with various NaOH concentrations in alkaline phenol reagent(phenol 95 g/1). Results in mol ammonia/l.

NaOH NaOH NaOH35 g/1 40 g/1 45 g/1

Tatrate method:Seawater, 55 mmol Mg/1Standard, 30 mmol Mg/1Standard, 60 mmol Mg/1

CyDTA method:Seawater, 55 mmol Mg/1Standard, 30 mmol Mg/1Standard, 60 mmol Mg/1

-0"36 -0"05 0"30-0"37 -0"24 0"01-0"57 -0"45 0"10

0"29 0"23 0"240" 13 0"02 0"030"23 0"14 0"12

Table 3. Between-run reproducibility (N 30).

Con-centration Mean SD CV

Determinand (zmol/1) (mol/1) (mol/1) %

Ammonia: 0 0"0002 0"041Sample 0"87 0"044 5’0

5 5"02 0"039 0"7710 9"98 0"062 0"6220 20"04 0"068 0"33

Phosphate: 0 0"0002 0"008Sample 0"22 0"007 3"2

1"25 1"23 0"005 0"42"5 2"47 0"006 0"245 5.00 0.008 0.16

Nitrate: 0 0"008 0"033Sample 0"95 0"031 3"3

5 5"02 0-027 0"5310 10"11 0"02 0"1920 20"26 0"04 0" 19

Nitrite: 0 -0"006 0"014Sample 0"3 0"013 5"7

1"25 1"25 0"009 0"722"5 2"51 0"007 0"275 5’01 0"01 0"19

air used for bubble segmentation on the nitrate channelby passing it through an oxygen absorber (activated iron,commercial hand-warmer available in Japan). An alter-native would be to use nitrogen. This is necessary to avoidpremature degradation of the reduction performance ofthe cadmium coil [3].

ManifoldsThe flow diagram for all four channels is shown in figure1. The cadmium reduction coil (Bran + Luebbe part No.165-0301-01 was prepared and regenerated according tothe manufacturer’s instructions. The analysis concen-

165

Page 4: Determination seawater by segmented-flow analysis higher

M. Jodo et al. Determination of nutrients in seawater by segmented-flow analysis

O,g at/O

30-

10-

Correlation of NO3-NNO3-Ny 1.009-0.085

0.999

0 10 20 30 O,g at/,’)Manual Method

10-

Correlation of PO4-P

Figure 2

PO4-Py 0.995--0.017=0.996

10

Manual Method

Figure 4

15 (g at//’)

(g at/t)40- Correlation of NH4-N

NH-4y 0.945 + 0.186 /

30-

10-

0 10 K) 3 4bg at/t

Manual Method

Figure 3

Figures 2-5. Correlation of automated methods with reference methods.

tration ranges in the low and high ranges were:

Ammonia 10 and 40 tmol/1Nitrate 10 and 40 tmol/1Nitrite 2 and 8 tmol/1Phosphate 4 and 16 mol/1

Each manifold has two sample pump tubes of differentflow rates. In routine operation, samples were firstanalysed using the large pump tubes. If any samples felloff scale, the following sample was automatically re-sampled at the end of the analysis, to eliminate any effectof carry-over from the off-scale sample. At the end of therun, the off-scale samples, and high standards, were re-analysed with the small sample pump tubes, giving a

higher dilution. This allowed samples falling into bothranges to be placed randomly into the sample traywithout risk of inaccuracy. By splitting the analyticalrange like this a lower detection limit on the low rangewas attained.

(g at/t)4.0-

3.0-

2.0-

1.0-

Correlation of NO2-N

NO2-N1.040 + 0.022Y

0 1.0 2.0 3.0 4.0 g at/t)Manual Method

Figure 5

Results and discussion

The effects of reagent composition on magnesium inter-ference in the ammonia determination were investigated.Table shows the results for three open-sea samples withlow ammonia content, which were measured using withthe complexing reagent from the AutoAnalyzer method(citrate/tartrate) and with CyDTA. The magnesiumcontent of the samples was 50-55 mmol/1. EDTA wasalso tested: the results, are not shown, but lay betweentartrate and CyDTA.The effect of magnesium on theammonia determination could be due to interaction withsodium phenolate, which would cause a change in theresonant structure of the phenolate ion and a decrease incolour.

Table 2 shows the influence ofmagnesium on the originaltartrate method and the CyDTA method at differentNaOH concentrations, These results show that the

166

Page 5: Determination seawater by segmented-flow analysis higher

M. Jodo et al. Determination of nutrients in seawater by segmented-flow analysis

CyDTA method is less affected by variations in theNaOH concentration. This is important, becausehypochlorite solutions, even from the same .supplier,contain variable ratios of hypochlorite to hydroxideaccording to their age, and the quantity of the stocksolution required to prepare the working reagent variesaccording to the free chlorine concentration.

The sensitivity of each method in the low-level analysisrange, measured by the absorbance produced by the topstandard, was as follows: ammonia 0"1 at 10 Bmol/1,nitrate 0"23 at 10Bmol/1, nitrite 0"23 at 2 tmol/1,phosphate 0" 19 at 4 tmol/1.

Reproducibility was measured by analysing five differentconcentrations ofeach analyte 10 times, in random order,on three separate days. There were thus 30 results at eachconcentration. The day-to-day relative standard devi-ation was less than 1% for all parameters at levels above5 Bmol/1. These results are summarized in table 3.

The detection limit, defined as three times the standarddeviation of the zero samples in the above test, was0"12 tmol/1 for ammonia, 0"024 tmol/1 for phosphate,0-1 tmol/1 for nitrate and 0.052 Bmol/1 for nitrite. Basedon past experience, determining the detection limit by theEPA procedure for Method Detection Limit, whichmeasures 10 or more consecutive low-level samples,would produce figures about half those above.

To validate the results from the automated methods,about 100 samples were collected from coastal areasaround Japan and analysed according to the manualmethods in the Manual for Oceanographic Observation [6],

and on the automated system. The results showed goodagreement, as shown in figures 2 to 5.

Conclusions

By applying a flow-through cadmium reductor to amicro-bore segmented-flow analyser, reliable and repro-ducible results were achieved for nutrient analysis at arate of 90 samples/h. CyDTA proved to be an effectivecomplexing agent for magnesium, and reduced thenegative effect on the ammonia determination. It alsoimproved the robustness of the method to varying NaOHconcentration..A disadvantage ofCyDTA is its high cost.

By operating the analyser in the dual-range mode it waspossible to reduce the detection limit and increase theanalytical range by reducing the effect of carry-over onvery low samples. This also improved the convenience ofdaily operation.

References

1. SNYD.R, L. R., Journal of Chromatography, 125 (1976),287-306.

2. SnYDF.R, L. R., Analytica Chimica Acta, 114 (1980), 3-18.3. PATTOn, C.J., Design, characterisation and applications of a

miniaturized continuous flow analysis system. Doctoralthesis (Michigan State University, USA, 1982).

4. Technicon AutoAnalyzer II Method No. 154-71W.5. GRASSHOFF, K., EHRHARDT, M. and KREMLING, K., Methods

of Seawater Analysis (Verlag Chemie, D-6940 Weinheirn,1983), 363-365.

6. Manualfor Oceanographic Observation (Oceanographic SocietyofJapan, 1985).

167

Page 6: Determination seawater by segmented-flow analysis higher

Submit your manuscripts athttp://www.hindawi.com

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Inorganic ChemistryInternational Journal of

Hindawi Publishing Corporation http://www.hindawi.com Volume 2014

International Journal ofPhotoenergy

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Carbohydrate Chemistry

International Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Journal of

Chemistry

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Advances in

Physical Chemistry

Hindawi Publishing Corporationhttp://www.hindawi.com

Analytical Methods in Chemistry

Journal of

Volume 2014

Bioinorganic Chemistry and ApplicationsHindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

SpectroscopyInternational Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

The Scientific World JournalHindawi Publishing Corporation http://www.hindawi.com Volume 2014

Medicinal ChemistryInternational Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Chromatography Research International

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Applied ChemistryJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Theoretical ChemistryJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Journal of

Spectroscopy

Analytical ChemistryInternational Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Quantum Chemistry

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Organic Chemistry International

ElectrochemistryInternational Journal of

Hindawi Publishing Corporation http://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

CatalystsJournal of