151
DESIGN OF A GAS HOLDUP SENSOR FOR FLOTATION DIAGNOSIS DEPARTMENT OF MIN~NG AND METALLURGICAL ENG~NEERING MCGILL UNIVERSITY MONTREAL, CANADA AUGUST, 1998 A TiiESiS SL~B~IITTEDTO THE FACULTY OF GRADUATE SI-UDIES AND RESEARCH IN PARTIAL FULFlLLhlENT OF Tt IE REQUiREMENTS OF THE DEGREE OF MASTER OF ENGINEERING

DESIGN A GAS HOLDUP SENSOR FLOTATION DIAGNOSIS

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

DESIGN OF A GAS HOLDUP SENSOR FOR FLOTATION DIAGNOSIS

DEPARTMENT OF MIN~NG AND METALLURGICAL ENG~NEERING

MCGILL UNIVERSITY

MONTREAL, CANADA

AUGUST, 1998

A TiiESiS SL~B~IITTEDTO THE FACULTY OF

GRADUATE SI-UDIES AND RESEARCH

IN PARTIAL FULFlLLhlENT OF Tt IE REQUiREMENTS OF THE DEGREE OF

MASTER OF ENGINEERING

I I

National Library l

BiMiothéque nationale du Canada

Acquisitions and Acquisitions et Bibliographie Services services bibliographiques

The author has granted a non- exclusive licence ailowing the National Library of Canada to reproduce, loan, distribute or seU copies of this thesis in microfom, paper or electronic formats.

The author retains ownership of the copyright in this thesis. Neither the thesis nor substantiai extracts fiom it may be printed or otherwise reproduced without the author's permission.

L'auteur a accordé une licence non exclusive permettant à la Bibliothèque nationale du Canada de reproduire, prêter, distribuer ou vendre des copies de cette thèse sous la fonne de microfiche/nlm, de reproduction sur papier ou sur format électronique.

L'auteur conserve la propriété du droit d'auteur qui protège cette thèse. Ni la thèse ni des extraits substantiels de celle-ci ne doivent être imprimés ou autrement reproduits sans son autorisation.

This fhesis is dediccred !O my wife Karen

And my son Javier who is leurning #O live

The mineral processing group of McGill University developed a novel gas

holdup probe that consists of two conductivity flow cells: an open ce11 used

to estimate dispersion (sluny + air) conductivity and a syphon ce11 used to

estimate sluny conductivity. These two values are entered into Maxwell's

mode1 to calculate gas holdup.

In this work some design criteria for conductivity flow celis and the syphon

ce11 are given. Effect of geometrical ceil constant and electrode width on

ce11 behaviour are analyzed. New dimensions for open and syphon cells are

proposed. Conditions to prevent bubbles from being entrained into the

syphon cell are established.

The new design of the gas holdup probe was tested successfully over a

prolonged period at the INCO Matte Separation Plant (Copper Cliff,

Ontario). Tests carried out in waste paper de-inking at BOWATER

(Gatineau, Quebec) showed ihc gas holdup va

with the values obtained froni prcssurc. The

holdup signal obtaincd by thc prohc could

operation.

lues given by the probe agreed

noise associated with the gas

be used to diagnose sparger

Gas holdup showed some degree of correlation with notation efficiency of

waste paper de-inking and tlotation rate constant. Chemistry and rheology of

the feed are other factors to be considered. Estimation of bubble surface area

flux should consider rheology properties (liquid viscosity and density).

Le groupe de procédés des minerais de l'université McGill a développé une

nouvelle sonde de fraction gazeuse qui consiste en deux cellules de

conductibilité du débit: une cellule ouverte, utilisée dans l'estimation de la

conductibilité de la dispersion (pulpe + air), et une cellule siphon, utilisée

dans l'estimation de la conductibilité de la pulpe. Ces deux valeurs entrent

dans le modèle de Maxwell utilisé, afin de calculer la fraction gazeuse.

Dans ce travail, quelques critères de design pour des cellules de

conductibilité du débit, et la cellule siphon, sont donnés. Les effets de la

constante géométrique de la cellule et de la largeur de l'électrode sur le

comporiement de la cellule sont analysés. De nouvelles dimensions pour des

cellules ouverte et siphon sont proposées. Des conditions de prévention

d'entraînement des bulles dans la cellule siphon sont établies.

Le nouveau design de la sonde de fraction gazeuse a été testé avec succès,

sur une période prolongée, à l'usine INCO Matte Separation Plant (Copper

Cliff. Ontario). Les tests effectués dans le désencrage de papier recyclé à

BOWATER (Gatineau, Québec) ont démontré que les valeurs de fraction

eazeuse obtenues par la sonde étaient en accord avec les valeurs obtenues b

des mesures de pression. L'interférence associée au signal de fraction

gazeuse obtenu par la sonde pourrait être utilisée dans le diagnostic

d'opération du barboteur.

La fraction gazeuse a démontré un certain degre de corrélation avec le

rendement de flottation du désencrage du papier recyclé et la constante du

taux de flottation. La chimie et la théologie de l'alimentation sont d'autres

facteurs à considérar. L'estimation du flux de surface des bulles devrait tenir

compte des propriktés rhéologiques (viscosité et densité du liquide).

1 would like to thank my supervisor, Professor J.A. Finch and Dr. C.O.

Gomez for their invaluable contributions and support of my work,

especially, for the opportunity of participating in the McGill mineral

processing group.

Also, 1 want to thank al1 members of the sensor/column flotation group at

McGilI University, and specialiy Mr. H. Hemindez, Dr. R. Escudero, Dr. F.

Tavera, Mr. G. Leichtle and Mr. C. Hardie for a most pleasant working

atmosphere.

1 want to express my gratitude to Mr. R. Agnew and Mr. C. Pollard for their

constant support during the plant tests and al1 the staff and crew of the

Matte Separation Plant, INCO ltd.

Finally, I want to thank the staff and crew at the waste paper de-inking plant,

BOWATER Pulp and Paper Inc.

1.1. UNDERSTANDING THE PROBLEM

1.2. OBJECTIVES AND ORGANIZXT~ON OF THE THESIS

CHAPTER 2: BACKGROUND

2.1. DEFINITION OF CONDCCTIVITY

2.3. ~IEASCRISG COSDCCTI VITI'

2.3. MASWELL'S MODEL

2.4. COSDL'CTIVITI- FLOW CELLS

2.5. Gxs HOLDCP PROBE: ~ ~ C G I L L PROTOTYPE

2.6. PRISCIPLES OF Tt IE Sj'Pt IO\ CELL

2.7. DESCRIPTIOS OF FL-OTXTIO\ C'OLLSIX

2.8. EFFECT OF OPERATISG AXD DESICX VARIABLES OS GAS HOLDUP

2.8.1. OPERATING VARIABLES

2.8.2. DESIGS VARIABLES

2.9. GAS HOLDUP ROLE 1s FLOTATIOS PERFORMANCE

2.1 0. BUBBLE SURFACE AREh FLUX ESTIMATION

CHAPTER 3: DESIGN CRITERIA FOR GAS HOLDUP SENSOR

3.1. INTRODUCTION

3.2. PLUGGING

3.3. DESIGN CRITERIA FOR THE CONDUCTIVITY CELL

3.4. DESIGN CRlTERlA FOR SYPHON CELL

3.5. EXAMPLE OF DESIGN

CHAPTER 4: EXPERIMENTAL

4.1. INTRODUCTION

4.2. DATA ACQUISITION SYSTEM

4.3. INCO: M A ~ E (CU-NI) SEPARATION PLANT.

4.4. BO WATER: DE-INKING FLOTATION COLUMN

CHAPTER 5: INCO RESULTS

5.1 . TEST OF RELIABILITY

5.2. RESPONSE TO A CHAKE OF THE A I R FLOW RATE.

5.3. GAS HOLDUP READISGS AT TH'O DIFFERENT DEPTHS.

CHAPTER 6: BOWATER R~sut:rs 6 1

6.1. GAS HOLOUP ~- IEASI~REVE~T~ FROU PRESSURE AND CONDUCTIVITY 61

vii

1.1. GAS HOLDUP ROLE IN FLOTATION

1.2. GAS HOLDUP ESTIMATION USlNG PRESSURE VALUES

1.3. GAS HOLDUP ESTIMATION USlNG MAXWELL'S XIODEL

1.4. SCHEMATIC REPRESENTATION OF GAS HOLDUP PROBE MODEL

2.1. SCHEMATIC DIAGRAM OF A CONDUCTIVITY FLOW CELL

2.2. THE CObIBINATION OF "OPEN'. ,AND "SYPHON.. CONDUCTIVITY CELLS

2.3. SCHEMATIC REPKESENTATION OF THE SYPHON CELL IN AN AIR-LIQUID

DISPERSION.

2.4. SCHEMATIC DIAGRAM OF A CONVENTIONAL FLOTATION COLUMN

2.5. GAS HOLDUP AS A FWCTIOX OF GAS RATE

2.6. GÂS HOLDUP VERSUS GAS RATE, EFFECT OF DOWNWARD LIQUID VELOCITY

2.7. GAS HOLDUP VERSCS G.4S RATE, EFFECT OF FROTHER DOSAGE

2.8. GAS HOLDUP VERSL'S G A S RATE .AS .-1 FUSCTION OF RS

2.9. GAS HOLDUP VERSL'S COSCESTR:\TE SOLIDS \1.4SS FLOW

2.10. COSIPARISON OF S,, TO c,,

3.1. S ~ i - i ~ h l t I ~ l ~ REPRESEST.4TIOS OF .4 COSDL'CTIVITY CELL WITH A THREE-

ELECTRODE ARRX\GE\IEST

3.2. CELL CONSTAST BE11 -\\'IOC'R I\ 'Ti{ O DIFFEREKT FLOW CELLS

~ .~ .DE\Y.~TIos FROXI GEO\~~:TRIC-:IL. (*I.I.l. <'O\Sî.-l\ l ' FOR 4 CELLS

\i'ITH 3-ELECTRODE :\HU \ \ C ; I . \ l l \ I

3.4. EFFECT OF THE RATIO :\, , , :\,, ( ) \ Ill \ I -\ 1 IO\ FROXf THE GEOXIETRICAL

CELL CONSTANT

3.3. EXPER~XIENTAL SETI'P t S13D l ' O < ' i l \ R . \ < ' l f KILE THE BOTTOM ORIFICE

Dl AhlETER OF SYPHOS CELI.

3.6. VA AS A FUNCTIOS OF l I FOU ix-II ORIFIC-t: DIAXIETER

VA AS A FbXCTIOS OF C,, FOR i<:\<'il ORIFICE DIAXIETER

DETERWISATION OF P:lR:\3lt~Tt.KS C',, : I \D Ci

MAXIXIUM ORIFICE DlASlETER AS r\ FI\CTION OF THE BUBBLE TERMINAL

VELOCITY AXD GAS HOLOUP

3.10. SCHEMATK REPRESENTATION OF THE GAS HOLDUP PROBE DESIGNED FOR AN

APPLICATION WlTH & = 1 O O ? AND DB = 1.0 MM.

4.1. SCHEMATIC REPRESENTATION OF THE DATA ACQUISITION SYSTEV

4.2. MA~N MENU OF THE DATA ACQUISITION SOFTWARE

5.1, RESULTS ON INCO COLUMN 2 FROM MARCH 13 TO MARCH 30

5.2. GAS HOLDUP RESPONSE TO CHANGES IN THE AIR FLOW RATE TO

M C 0 COLUMN 3 (3 1 -MARCH- 1998)

5.3. CONDUCTIVITY VALUES AND RATIO Y RESPONSES TO CHANGES IN THE

AIR FLOW RATE TO INCO COLUMN 3 (3 I-MARCH-19%)

5.4. ~ c O V E R Y RESPONSES TO CHANGES LN THE AIR FLOW RATE TO

M C 0 COLUMN 3 (3 1 -MARCH- 1 998)

5.5. FROTH DEPTH VARIATIONS IN COLUMN 3 (3 1 -MARCH- 1998)

5.6. GAS HOLDUP SIGNALS OF SENSOR 1 AND SENSOR 2 WORKING AT DIFFERENT

DEPTHS INSIDE INCO COLUhm 3(3-APRIL-I 998)

6.1. GAS HOLDUP ESTIMATES FROM CONDUCTIVITY AND PRESSURE

~IExSCRE~IENTS FOR BOWATER TESTS

6.2. G . ~ s HOLDUP EFFECT ON FLOTATION EFFICIENCY FOR TWO RESIDENCE TIMES

6.3. GAS HOLDUP EFFECT ON FLOTATION EFFICIENCY FOR TWO FROTH HEIGHTS

6.4. EFFEC~ OF SPXRGER SURFACE AREA ON GAS HOLDUP FOR TWO

ISlTlXL CONDITIOKS

6.5. EFFECT OF SPARGER SURFACE ARE.4 OS FLOTATION EFFICIENCY

6.6. FLOTATION EFFICIESCY VERSCS G A S HOLDUP

6.7. COLLECTIOX ZOSE RATE COSSTAST VERSCS GAS HOLDUP

6.8. BCBBLE SURFACE AREA FLLS \'ERSUS GAS HOLDUP

6.9. GXS HOLDLP VERSUS R.-ITIO JG/DB

6.1 O. BL'BBLE SURFACE AREA FLUX VERSLS COLLECTION ZONE RATE CONSTANT

C. 1. INCO M.ATTF. SEPARATION PLAST 94

3.1 - COLUMN CHARACTERISTICS, N C O MATE SEPARATION PLANT

4.2. CALIBRATION PARAMETERS N C O TESTS

4.3. CALIBRATION PARAMETERS FOR BOWATER TESTS

A. 1. CONDUCTIVITY AND CONDUCTANCE VALUES OBTAINED FOR

TWO 3-ELECTRODE CELLS

A.2- CELL CONSTANT VALUES (MEASURED AND GEOMETRICAL)

FOR 4 CELL DIAMETERS (3-ELECTRODE CELLS)

A.3. CELL CONSTANTS (MEASURED AND GEO~ETRICAL) FOR

THREE ELECTRODE WIDTHS (3-ELECTRODE CELLS)

A.4.A. LIQL'ID VELOCITY INSIDE THE SYPHON CELL AS FUNCTION OF

G . 6 HOLDUP AND H

Pl .4 .~ . CO;~TINWAT~ON FROhl A . ~ . A

C. 1 . CALIBRATION DATA. CONDUCTIVITY h1ETER FOR M C 0 TESTS

C.2. CALIBRATION DATA. OPEN CELL OF SENSOR 1 FOR INCO TESTS

C.3. CALIBRATION DATA. SSPHON CELL OF SESSOR 1 FOR INCO TESTS

C.4. CXL~BRATION DATA. OPEN CELL OF SENSOR 2 FOR INCO TESTS

C.5. CALISRATIOS DATA . SI'PHOS CELL OF SESSOR 2 FOR N C O TESTS

C.6. CALIBRATIOS DATA. COSDCCTI\'ITI' LIETER FOR BOWATER TESTS

C.7. CALIBRATIOS DATA. OPEN CELL FOR BOWATER TESTS

C.8. C.-\L~BRXTIOS DATA, sw ~ O V CELL. FOR BOWATER TESTS

D. 1 . INCO ~ ' ~ A T E SEFIARATIO\ PL.-I\T COl.1 \lx 2. G.As HOLDUP DATA

COLLECTED MARCH 13 ( 1998) L Sl\G SEXSOR 2 .

D.2. lNCO MATE SEPARATIOX P L A ? ~ r (-01.1 2 . GAS HOLDUP DATA

COLLECTED MARCH 14 ( 1998) cSIXG SEXSOR 2.

D.3. [ N C O MATE SEPXRATIOS PLAST C O L L ~ ~ N 2 . GAS HOLDUP DATA

COLLEcTED MARCH 15 ( 1998) L'SISG SENSOR 2.

DA. INCO MATE SEPARATIOS PLAST COLC'US 2. GAS HOLDUP DATA

COLLECTED MARCH 16 ( 1998) USISG SESSOR 2.

D.5. R\JCO MATE SEPARATION PLANT COLUIL& 2. GAS HOLDUP DATA

COLLECrED MARCH 17 (1 998) USWG SENSOR 2.

D.6. INCO MATE SEPARATION PLANT COLUMN 2. GAS HOLDUP DATA

COLLECTED MARCH 18 (1998) USING SENSOR 2.

D.7. iNCO MATE SEPARATION PLANT COLUMN 2. GAS HOLDUP DATA

COLLECTED MARCH 19 (1 998) USING SENSOR 2.

D.8. INCO MATE SEPARATION PLANT COLUMN 2. GAS HOLDUP DATA

COLLECTED MARCH 20 (1 998) USlNG SENSOR 2.

D.9. Iru'CO MATE SEPARATION PLANT COLUMN 2. GAS HOLDUP DATA

COLLEnED MARCH 2 1 (1 998) UStNG SENSOR 2.

D. 10. INCO MATE SEPARATION PLANT COLUMN 2. GAS HOLDUP DATA

COLLECTED MARCH 22 (1 998) USlNG SENSOR 2,

D. 1 1 . INCO MATE SEPARATION PLANT COLUMN 2. GAS HOLDUP DATA

COLLECTED MARCH 23 (1 998) USING SENSOR 2.

D. 12. INCO MATE SEPARATION PLANT COLUMN 2. GAS HOLDUP DATA

COLLECTED MARCH 24 ( 1 998) USING SENSOR 2.

D. 13. INCO MATE SEPARATION PLANT COLUX~N 2. GAS HOLDUP DATA

COLLECTED MARCH 25 (1 998) USlNG SENSOR 2.

D. 14. INCO MATE SEPARATION PLANT COLUMN 2. GAS HOLDUP DATA

COLLECTED MARCH 26 ( 1998) CSlh'G SENSOR 2.

D. 15. INCO MATE SEPARATION PLANT COLL'hlN 2. GAS HOLDUP DATA

COLLECTED MARCH 27 (1998) L'SING SENSOR 2.

D. 1 6 . INCO MATE SEPARATION PLAST COLUhlN 2. GAS HOLDUP DATA

COLLEC Ï F D MARCH 28 ( 1998) L'SiXG SENSOR 2.

D. 17. INCO MATE SEPARATION PLAST COLt'hlN 2. GAS HOLDUP DATA

COLLECTED MARCH 29 ( 1998) USING SENSOR 2.

D. 18.lNCO MATE SEPARATION PLANT COLUhlN 2. GAS HOLDUP DATA

COLLECTED MARCH 30 ( 1998) USING SENSOR 2.

D. 19.~. INCO MATE SEPARATION PLANT COLUMN 3. GAS HOLDUP DATA

COLLECTED MARCH 3 1 ( 1998) USING SENSOR 1.

D. 1 9 . ~ . CONTINUATION FROM D. 19.A.

D. 19 .~ . CONTINUATION FROM D-19 .~ .

D.20. MEASURED GRADES FOR WC0 TEST

D.2 1. GRADE VALUES USED TO ESTIMATE T H E WEIGHTS FOR

THE MASS BALANCE ADJUSTMENTS

D.22. ADJUSTED GRADES FOR WC0 TEST

D.23. RECOVERY AND TYPICAL DEVIATION VALUES FCR NCO TEST

D.23. DATA O N AIR FLOW RATE. LEVEL (FROTH DEPTH) AND SOLID PERCENT

FOR INCO TEST.

D.25.~. DATA FROM TWO SENSORS AT DIFFERENT DEPTHS INSIDE COLUMN 3

D . 2 5 . ~ . CO~TINUAT~ON FROM D.25.~.

D.23.~. C O N T ~ N U A T ~ O N FROM D.25.~.

E. 1. GAS HOLDUP FROM PRESSURE VERSUS GAS HOLDUP FROM CONDUCTIVITY

FOR BOWATER TESTS

E.2. TEST CONDITIONS FOR BOWATER TESTS

E.3. FLOTATIOX EFFICIENCY AND BRIGHTSESS GAIN FOR BOWATER TESTS

O E.4. TEST COSDITION FOR TEST 2B-%P HF = 100 CM: 5 SPARGERS: T = 6 MIN.

E.5. TEST COSDITIOS FOR TEST 2 I 3 - 4 ~ ~ HF = 100 CL1; 3 SPARGERS; f = 6 h1lN.

E.6. TEST COSDITIOS FOR TEST 2 B - ~ S P HF = 100 Chl: 3 SPARGERS; f = 6 hlIN.

E.7. TEST CONDITION FOR TEST ? A - ~ S P HF = 100 CX1: 5 SPARGERS: T = 3 hllN.

E.8. TEST COSDITIOV FOR TEST 2 A - k ~ Ht = 100 CXI: 4 SPARGERS: T = 3 h W .

E.9. TEST COSDITIOX FOR TEST 2.4-~SP Hf = 100 CM: 3 SPARGERS: T = 3 hllN.

E. 1 O. BLBULE SIZE. BCBBLE SC'RF--\CE :\RE.-\ FLCS AND COLLECTIOS ZONE

RXTE C O S S T A S T ESTIXIATED FOR BON'ATER TESTS.

parameters in equation 4.1

parameters in equation 4.2

parameters in equation 4.3

cross-sectional area of column, cm2

cross-sectional area in point A (Equation 2.1 1 )

cross-sectional area in point B (Equation 2.1 1)

cross-sectional area of conductivity ce11

sparger surface area

electrode exposed area

final ink concentration, ppm

initial ink concentration

geometrical cell constant. cm -

parameter in Equation 3.10

parameter in Equation 3.10

intemal diameter ot~conductivity cell, cm

syphon cd l intemal diameter

open cell intemal diamcter

bubble diametcr

column diamcter

orifice diameter at syphon çell bottom

friction coefticient

friction losses. cm2/s'

fraction volumrtric o f dispersed phase

de-inking notation efficiency

et-

Ef

f F.E.

gravitational acceleration, crn/s2

length of vessel, cm

height difference, cm (Figure 3.5)

collection zone height

froth zone height

current, A

current density, Akm'

superficial velocity, cm/s

superficial feed velocity

superficial gas velocity

superficial liquid velocity

superficial tailings velocity

AJAS

conductance, mS

collection zone rate constant. I /min

gap between electrodes. cm

open cell gap

syphon cell gap

total syphon cell Icngih. cm

parameter in Eqiiaiion 2.24

pressure in poini A. g cms-

pressure in poini B

pressure in poini A'

pressure in point B'

heat, cm%' (Equat ion 2.1 0)

Volumetric tlow rate. cm3/s

xiv

Resistance, R

Reynolds number of a single bubble (Equation 2.29)

Reynolds number of a bubble in a swarm (Equation 2.26)

collection zone recovery

fioth zone recovery

overall recovery

bubble surface area flux, 1 /s

slip velocity between gas and sluny, cm/s

bubble terminal velocity

liquid velocity in point A, cm/s (Equation 2.10)

liquid velocity in point B (Equation 2.10)

fluid work, cm%'

electrode width, cm

open ce11 electrode width

syphon ce11 electrode width

position in point A, cm

position in point B

bubble surface area moving through a vessel, cm%

parameter in equation 1.6

parameter in equation 2.10

parameter in equation 2.10

apparent relative conductivity

gas holdup

Psi

conductivity, mS/cm

dispersion conductivity

apparent conductivity (Equation 3.6)

liquid conductivity (Equation 2.6)

slurry conductivity

conductivity o f continuous phase (Equation 2.7)

conductivity of dispersed phase (equation t -7)

bubble density, g/cm3

slurry density

residence time

time taken a passing through a length h, s

time taken a passing through a length db

potential. volt

potential in electrode A. volt

potential in electrode B

slurry viscosity. @cm s

1.1 Understanding the Probtern

For many years a relationship between the performance of the flotation

process and some process variables has been sought. One of the latest

attempts suggests that the bubble superficial surface flux (or bubble surface

area rate) has a linear relationship with the flotation rate constant [ I l . The

bubble superficial surface flux, Sb is defined by the gas superficial velocity

and bubble diameter. The bubble superficial surface rate is difficult to

calculate because of the difficulty in obtaining a good estimation of the

bubble diameter in a 3-phase system. The gas holdup, an easier variable to

estimate, could be related to Sb.

When gas is introduced in the column, liquid (or slurry) is displaced. The

volumetric fraction displaced is called the gas holdup, E,. The role of gas

holdup in the flotation process has long been discussed but no relationship

has been found yet because of the lack of reliable industnal data. In theory,

gas holdup is defined by bubble size, liquid flow rate, solids content and

rnixing patterns [7-31. Gas holdup defines flotation kinetics and canying

capacities, which are associated with the performance 0f.a flotation device,

Figure 1.1.

Patterns )

Size Content

These facts have generated the following questions:

* How can we obtain a good rneasurement of gas holdup?

1s gas holdup related to flotation performance?

+ 1s there an optimum gas holdup?

The first question is raised because we do not have a reliable gas holdup

sensor for industrial applications. The common method to measure gas

holdup is based on pressure and estimations of the sluny density. A

description of this rnethod is presented in Figure 1.2. The main

disadvantage is the dificulty of obtaining the sluny density.

In an effort to solve the lack of si reliable sensor, the

group at McGill University devcloped a sensor based

estimate gas holdup. usine Maxwcl los model [4-51.

mineral processing - .

on conductivity to

Electrical conductivity is an intrnsive propcny. which is a fùnction of the

concentration of the difkrent phascs prcscnt in a dispersion. According to

Maxwell's model, if we have 3 disprsion ( s l uny plus air), we can determine

the volume concentration of the non-conducting phase (in this case air), if

we know the dispersion conductivity and sluny conductivity. In ~ i ~ u r e 1.3

we can see a conceptual description ol'Maxwell's model.

To fulfill the requirements of Maxwell's model, the gas holdup probe

applies the pnnciple of separation of phases. The probe consists of two flow

cells: an open ce11 and a syphon cell. A flow ce11 is defined as one that

allows a fluid or dispersion to flow through fieely while the electrical

conductivity is measured [4].

Dispersion

(Slurry + Air)

The open cell measures the slum.-air dispersion conductivity by ailowing a

relatively free flou of thc dispersion through the cell. The syphon ce11

measures the conductivity bt-tlic s lum. only. which requires the exclusion of

bubbles. This is achievrd by hm in- tlic upper end open and the lower end

constricted. With this desiyn. air bubbles are prevented fiom entering the

cell, which causes the drnsity of the crll contents to exceed that outside, thus

inducing a syphon effect. This means that fresh slurry is drawn in at the top,

replenishing the cell contents and. on its exit through the constriction

(orifice). further ensures the esclusion of bubbles [ 5 ] . Figure 1.4 shows a

schernatic representation of the gas holdup probe (open and syphon cell).

(dis ersion con B uctivity)

Syphon ce//

Figure 1.4. . Schematic representat ion of gas holdup probe.

1.2 Objectives and Organization of the Thesis

1.2.1 Objectives

The main objectives of this project can be summarized as follows:

Re-design of sensor to avoid plugging (syphon cell).

Establish some design criteria for the sensor.

Conduct long terrn testing of the gas holdup sensor at industrial level.

Study the relationship between gas holdup and flotation performance.

1.2.2. Organization

The thesis is organized into 7 chapters. Chapter 2 examines the background

to conductivity measurements and conductivity ceIl design, requirements of

sensor design and the relationship of flotation to some flotation variables.

Chapter 3 describes the design criteria for conductivity cells and critical

sensor parameters. An example of a complete design is given.

Chapter 4 examines the experïmental setup for two industnal applications:

mineral flotation (Copper Cliff, INCO) and de-inking (Gatineau.

BOWATER). Chapten 5 and 6 give the results obtained at these two sites.

Finally, Chapter 7 is devoted to discussion and conclusions with suggestions

for fùrther work.

This chapter defines conductivity and describes Maxwell's model, which

relates the volumetric concentration of air (gas holdup) in a dispersion to the

conductivity values of the dispersion and the slurry (dispersion without air).

Conductivity flow cells are described. The principles of the components of

the gas holdup sensor are descnbed: the open cell, used to measure the

dispersion conductivity, and the syphon cell, used to measure slurry

conductivity. The effects of operating and design variables on gas holdup

are also presented. Finally, the role of gas holdup in flotation performance is

discussed.

2.1. Definition of Conductivity

E lectrical conductivity, also called specific conductance [6-81, specific

conductivity [9-101, or conductivity [ l 1-1 71, is the ability of a substance to

conduct electric current. In this thesis, the term conductivity and the symbol

K will be used.

The conductivity is the proportionality constant in Ohm's law [18]:

where i is the current density (A/cm2), Av is the potential gradient (volt/crn),

and K is the conductivity (R-'/cm). In fact, the conductivity is the reciprocal

of the resistivity (Rcm) and the conductance is the reciprocal of the

resistance (R). The SI unit name for the conductance is the Siemens (S): 1 S

= 1 0 ' . Thus the SI units of conductivity are S/m. For convenience, the cgs

unit S/cm (or rnS/cm) will be used in this thesis.

The conductivity is an intensive property that may be thought of as the

conductance of a cube of 1 cm edge, assuming the current is perpendicular

to opposite faces of the cube [13, 181.

2.2. Measuring Conductivity

Al1 substances conduct electricity to some degree. Some materials are good

conductors (such as metals) and others are poor conductors (insulators). In

this work, we will discuss the conduction of electric energy in an aqueous

electrolyte solution.

The resistance of an electrolyte solution cannot be measured using direct

current, because it changes the concentration of the electrolyte. The

accumulation of electrolysis products at the electrodes also alters the

9

resistance of the solution. Instead of direct current, an alternating current is

used to overcome these effects [ 1 81.

In a conductivity ce11 with facing plate electrodes, it is assumed that the

current flux is at right angles to, and constrained by, the area of the plates;

under this assumption, the resistance of the electrolyte is given by:

drop of potenrial ( v , - v , ) R = - -

current I

where v, and are the potentials on the plates electrodes A and B,

respectively, and 1 is the current in the electnc circuit. In the case of a linear

conductor, the current density. i. on any equipotential surface is constant.

Therefore,

where A,,, is the cross sscrianril arc3 01' the cell. L is the length of the cell,

and a and b are the positions » t thc clectrodes A and B, respectively.

Substituting Equations 2.1. 2.3 and 2.4 into equation 2.2, yields:

where K is the conductance of the electrolyte. In Equation 2.5 the terni

A,,ll& is referred to as the ce11 constant, expressed in cm.

2.3. Maxwell's Model

The conductivity method is a technique to measure the gas holdup in a gas-

liquid system. Generally, the conductivity of a gas-liquid dispersion (a

continuous liquid phase with a dispersed gas phase) depends on the

conductivities of the two phases and their relative arnounts. However, the

relationship between the conductivity of the dispersion and the concentration

of the dispersed phase is not linear [19].

Maxwell considered a large sphere (continuous phase) containing many

small spheres (dispersed phase) of different conductivity. Assuming the

distance between small spheres is large enough so that their effect in

disturbing the path of the currrnt may be taken as negligible, the apparent

conductivity of this large sphere is given by [19-201:

where r , is the conduçtivitj. of' thc continuous phase, f is the volumetric

fraction of the dispersed phase oi'conductivity ~ 2 , and P is given by:

Recent research has found that this model c m be used to estimate the gas

holdup in three-phase flotation systems [3-51. Adapting Maxwell's model

by treating the sluny (solid plus water) as the continuous phase, then,

a, = ad, apparent conductivity of the mixture (dispersion) for any E,;

al = K , ~ , conductivity of the slurry (E, = 0);

a2 = O, conductivity of air;

f = E,, fractional gas holdup;

Thus, the gas holdup can be expressed in terms of apparent relative

conductivity y,

where

2.4. Conductivity Flow Cells

From experiments, it was shown that the design of a conduc tivity ce11 used

to measure the conductivity of a dispersion is important [3-5, 21-23].

Unifotm and parallel curent lines between two opposite electrodes in the

cell are essential. Once this condition is met, Equation 2.8 is satisfied.

A flow ce11 is defined as one that allows a fluid or dispersion to flow through

freely while the electrical conductivity is measured. One of the most

important features of a flow ce11 is the ce11 constant. The ce11 constant has

been defined as the ratio of the effective surface area used to transfer

electrical energy (normal to the 'flux of elecmc current) to the distance

between two points where the electrical energy is transferred [4]. Figure 2.1

shows a schematic of a conductivity flow cell.

Cross section arcri

Elcctrodes

Figure 2.1. Schematic diagram of a corrtlrrctivi@flow cell.

Experimental data on open cells indicate that the electric field formed inside

the ce11 extends beyond the edges of the electrodes [4]. This is because the

isopotential planes formed between the electrodes are not parallel but

concave to the cross section area of the cell. This extension of the electric

field increases with increasing conductivity of the aqueous media, as

reflected in a decrease of the ceIl constant. .

The ce11 constant of a flow ce11 is determined by calibration against

electrolyte solutions of known conductivity. The ce11 constant depends

mainly on ce11 dimensions; the experimental observations suggest that the

ce11 constant is independent of the type of electrolyte [4].

The effect of the addition of non conductive bodies in the flow ce11 has been

verified experimentally. It is concluded that the ce11 constant is not affected

by the presence of these bodies [4].

A three-ring arrangement was proposed to limit the electric field to the

volume in between the top and bottom rings: by maintaining these two

electrodes at the same potential and the central electrode at opposite polarity,

a current path outside this volume is avoided [3-41.

2.5. Cas Holdup Probe: McCill Prototype

An industrial prototype was built at McGill University based on the direct

measurement of the air-free dispersion conductivity. As described, the

simultaneous measurernent of the dispersion conductivity and the liquid

conductivity can be accomplished by using a combination of two

conductivity flow cells (open and syphon cells):

Open c d : a vertical cylinder open at both ends, with three intemal ring

- electrodes flush mounted to the wall, and used to measure the conductivity

of the dispersion (rd).

Syphon cell: a vertical cylinder with a conical bottom end, opened at the top .

and having û small opening (orifice) at the boaom. Again, three intemal

ring electrodes flush mounted to the wall were used. This ceil was used to

measure the liquid conductivity (K,,).

The prototype was built by combining these two cells in a single unit.

Figure 2.2 shows a schematic of the gas holdup probe.

Ring electrodes

Ring electrodes

Open cell Syphon cell

2.6. Principles of the Syphon cell

When a swarm of air bubbles passes through a column of water, the bubble-

water dispersion presents a lower (dispersion) density relative to that of

water with no air bubbles. This difference in density was exploited to

develop a technique to separate the continuous phase (water, in two-phase

systems; slurry in three-phase systems) of the dispersion and allow the

measure of its conductivity [4].

Because the syphon ce11 has a small hole at the bottom, it does not allow the

ascending air bubbles to enter the cell; therefore, the ce11 becomes filled with

liquid (sluny) without air, creating a hydrostatic pressure difference between

the orifice at the bottom of the cell and the sumoundings which causes the

liquid to flow out. This liquid flow increases until the pressure difference is

equal to zero. At steady state. a continuous replenishment of fkesh liquid

takes place fkom the top of the ce11 creating, in the end, a syphon effect.

Successfùl operation requires that the liquid velocity at the top of the ce11 be

lower than the terminal velocity of the rising bubbles outside the cell,

otherwise air bubbles could be drsiwn into the cell.

Figure 7.3 shows a schrmatic representation of the syphon ce11 in an air-

liquid (slurry) dispersion. The conscnation of energy between points A and

B is piven by:

where Q = O; w = 0; p = 0.5 for laminar tlow and P = 1 for turbulent. From

the mass conservation law Ive haw ihat:

As the fluid at both points A and B is in contact with the dispersion, we have

that:

P,, - P, = Plr - P r

and applying Bernoulli's principle between points A' and B',

where (ZT-ZB) = (Zr--ZB-) = LT. NOW substituting Equations 2.1 1 , 2.13, and

7.13 in Equation 2.10:

The friction loss, Er, can be defined as:

neglecting the term containing V,,;

Equation 2.1 7 relates the velocity of the fluid at the top of the ce11 with the

length of the syphon ce11 and the gas holdup of the system.

2.7. Description of Flotatioa Column

Column flotation consists of ta11 vertical reacton where gas bubbles are

eenerated with spargers at the bottom and wash water is added at the top. *

The main advantages of columns inciude: improved separation performance,

low capital and operational costs, low floor space requirements, and

adaptability to automated control [34]. Many different types of coiumn ce11

configuration exist. The conventional, or open flotation column, is

illustrated in Figure 2.4.

in column flotation. rising gas bubbles interact with the descending sluny

(or pulp) in the collection zone (H,). Gas bubbles are normally generated

with intemal spargers. such as porous stainless steel pipes or variable gap jet

spargers.

Wash water is added to stabilize the froth, by replacing the water that

naturally drains from the froth. The remainder of the wash water flows

through the froth and cleans the entrained particles. Therefore, the froth

zone is also called the cleaning zone (Hf). The flow of water moving

through the froth is called the bias water, a positive value corresponding to a

net flow downwards. For efficient cleaning of entrained particles bias rate

must be positive.

The overflow is rich in the floatable (hydrophobic) material, which often

forms the concentrate in minera1 systems. The unfloatable (hydrophilic)

material is collected as underflow from the bottom of the column. De-inking

is a reverse flotation process where the valuable (fibre) material (the accepts)

forms the underflow, the overflow consisting of the waste (ink) matenal (the

rej ects).

The flow rates of the different streams in a column are normally expressed

as superficial velocities (or rates). Superficial velocity is the volumetric

flow rate of a particular stearn divided by the column cross sectional area:

where i cm be gas (G), feed (F), wash water (W), bias (B). or

tai 1 ings/accepts (T). Superficial velocities are usefùl foi corn paring columns

of different diameters and are usually expressed in c d s .

0 When the transport in the collection zone is described by the plug flow

mode1 with residence time r, the overall recovery, when using the collection

zone rate constant k, (I/min). is given by [I l :

whcre R,- is the recovery in the froth zone. If we take R, as the collection

zone recovery, the overall flotation column recovery Rfc is [ 3 ] :

For de-inking, flotation efficiency (F.E.) is used to define cotation

performance. Flotation eficiency in this thesis is defined as:

where c is the concentration of ink and the subscripts i and f are for initial

and final, respectively. In order to estimate the flotation rate constant in

Equation 2.19, flotation efficiency can be taken as recovery.

2.8. Effect of Operating and Design Variables on Gas Holdup

2.8.1 Operating variables

Superficial Cas Rate (Jd

Volumetric gas flow varies with the static head. Unless moted otherwise, -

values of gas rate are referrrd t» the column overflow, Le., at atmospheric

pressure. Figure 2.5 shows the gencral rclationship between cg and J,. Gas

holdup increases approsimatrl? linccirl?. thcn deviates from linearity above a

cenain range of J,. The linear scçtion is characterized by a hornogeneous

distribution of bubbles of fairly unilimii s i x . rising at a fairly uniform rate.

This is called the bubbly tlow regime. Above the transition J,, gas holdup

becomes unstable and the flow is chancterized by large bubbkes rising

rapidly, displacing water and small bubbles downward. This is the chum-

turbulent flow regime. The desirable column operation regime is the bubbly

flow [3 ] .

Figure 2.5. Gus holJup US a fincrion of gas rule. general relarionship. From Finch and Dobbjl[.?$

Superfcia! Liquid Rate (Jr or JT)

Figure 1.6 shows the result of increasing JL countercurrently to the bubbles

(the usual column operation). As JI increases for a given J,, E, inrreases.

This is expected since the bubble rise velocity (relative to a stationary

observer) decreases. lncreasing J I decreases the maximum J, that can be

tolerated for operation to remain in the desired bubbly flow regime.

The addition of frothcr up 1 0 a certain concentration has a pronounced

impact on reducing bubble s i x . A reduced bubble size means reduced

bubble rise velocity and consrqurntly a gas holdup increase. The impact of

frother concentration is shown in Figure 2.7.

1 1 . i i i l A . . . l . . . os 1.0 1.5

Figure 2.6. Gus holdrcp versus gus rate. effer! of donnward liquid velociiy. From Finch

1 2 3 4 Sup.if icial gaa rata, Jl (cmls)

Figzo-e 2.7. Gus holJirp verslis gas rute. c'$&~t offiothrr dosuge. From Finch und Dobby Pl-

Solids Concentration

The role of solids is difficult to describe. A general consideration is that

soiid particles may promote or retard coalescence depending upon their

surface properties. Specific to flotation is that the solids distribute between

the liquid phase (i.e., remain in suspension) and the gas phase (i.e., attach to

the bubble).

Mineral solids in suspension generate a slurry density, psi, and a slurry

viscosity, psi, which have opposite effects on the bubble rise velocity. A

higher sluny density increases the bubble rise velocity while decreasing E,.

On the other hand, higher sluny viscosity decreases the bubble rise velocity

and increases E ~ . Finally, solids attached to the bubble yield a bubble-

particle aggregate density, pb. greater than zero, causing the bubble rise

velocity to decrease and, gas holdup to increase [ 2 ] .

A general observation in flotation columns is that increasing the percent

solids decreases gas holdup [35-261.

2.8.2. Design Variables

Sparger surface (internai spargers)

It is known that gas holdup decreases as the ratio R, (= Ac/As) increases (see

Figure M), i.e., the bubble diameter increases as Rs increases. As R,

increases there is also a tendency to produce a less unifom bubble size, and, e

E~ versus J,, starts to deviate fiom linearity. This is evidence of a transition

fiom bubbly flow to churn-turbulent fiow [27]..

30

O O 0.5 1-0 1.5 2 .O

Suparticial 0.8 velocity, J, (cm/*)

Figzlre 3.8. Gas holdicp i7rrsi<s gus rute us u function of R, . Frorn Finch and Dobby [q

2.9. Cas Holdup Role on Flotation Performance

The effect of gas holdup on flotation performance is not clear. Some work

suggests that gas holdup could strongly be related to flotation performance

and that this relationship could be used to control the flotation process.

Figure 2.9 shows a relationship between concentrate solids flow rate and gas

holdup, for individual columns with similar recoveries (>95%). This

relationship was developed in an atternpt to design a control strategy 128-

291.

Little or no correlation between flotation rate constant and the gas holdup

has been reported for mechanical cells. Recent results have however shown

a strong relationship between bubble surface area flux and the flotation rate

constant [ l , 301. Bubble surface area flux is defined by:

From Equation 2.27 we can see that the main dificulty in calculating Sb is

obtaining a reliable estimation of db. This is even more problematic if we

consider that the bubbles have a size distribution rather than a unique size.

Exploring the role of Sb in de-inking flotation columns, Leichtle has

suggested a linear relationship between E, and Sb. Figure 1.10 shows the

results obtained by Leichtle for laboratory and pilot columns [3 11.

Figure 2.10. Cornparison of Sb 10 g using data points j iom all tests (i. e. laboratory and pilot scale columm) [31].

Apart from its possible direct role on flotation performance, gas holdup also

has diagnostic applications. For example, a ruptured sparger is readily

detected as the increase in bubble size will cause the gas holdup to drop

suddenly. . -

2.10. Bubble Surface Area F lux Estimation

The bubble size can be estimatrd by using direct [ l , 30, 321 or indirect

rnethods. Drift flux anûlysis. an indirect method, has been applied to

estimate bubble size in both two phase and three phase systems [2, 33-34].

A brief review of drift flux analysis. taken from Finch and Dobby [2], is

included.

The slip (relative) velocity of gas and liquid in countercurrent systems with

uniform bubbles is defined by :

where upward flow is positive.

The slip velocity is related to the system variables. For bubbie sizes o f

interest here, db =< 2 mm (Re =< 500), a suitable expression is an adaptation

of the multi-species hi~dered settling equation of Masliyah [2 ] , wt-itten

below for the gas-slurry system:

where

and m is a function of the Reynolds number:

1 < Re, < 200

m = 4.45 ~ e 2 . I 200 < Re, < 500

and

where U, is the terminal nse velocity of a single bubble in an infinite pool

(this is calculated using Equation 2.74 and assuming E, = 0).

Using drift flux analysis bubble size can be estimated

E,. Jg and Ji. Following is the step-by-step procedure.

1 ) Set J i = Jr

3 estimate dh;

3 calculate U,, Equation 2.14 with E, =O;

4) calculate Reh. Equation 1-29;

5 ) calculate m. Equation 1.27 or 2.28;

6 ) calculate Reh,, equation 1.19;

7) calculate dh , Equation 1.15; iterate on dh; Go to 2

ffom measurements of

CHAPTER 3: DESIGN CRITERIA FOR GAS HOLDUP SENSOR

3.1 Introduction

- The original sensor designed to measure gas holdup [3-51 experienced a

serious problem in industrial application, with plugging of the syphon ce11

by minera1 particles.

The design of the syphon ce11 had to overcome this problem while respecting

such electrical criteria as cell constant. separation between electrodes, and

electrode width.

3.2 Plugging

The original syphon ceIl was composed of a vertical cylinder (3.8 cm

diameter) closed at its bottom end ( 4 5 O angle) with a lateral opening (about

10 mm diameter) [3]. The first step was to modiQ the bottom of the ce11 to a

conical shape ( I O 0 angle) with an orifice at the bottorn. The second

modification was to increase the ce11 diameter from 3.8 cm to 7.5 cm. With

a larger diameter c d , we are in a position to increase the size of the (sluny

discharge) orifice at the bottom. Giveri that the previous design successfully

prevented bubbles fiom entering at both the top (entrained by the slurry

entenng the cell) and bottom, a sirnilar slurry velocity inside the ce11 can

now be obtained by increasing the orifice diameter. The larger orifice will

reduce plugging by solid particles.

From experience the intemal diameters of both cells were established at 9.6

cm (Do: open) and 7.5 cm (D,: syphon), while the lengths (LT) were

established at 50 cm (in the syphon case, this length includes the conical

section).

3.3 Design Criteria for the Conductivity Cell

In the design of a gas holdup sensor, we have to know the values of the ce11

constants and how they relate to the electrode width and the separation

between electrodes.

Figure 3.1 shows a schematic of the electrode arrangement using both the

open and syphon cells. A thtee-ring electrode arrangement is seen: a cectral

electrode at one polarity, with the two outer electrodes at opposite polarity.

With this design, the current flux is constrained to flow only inside the cell,

allowing for a more stable ceIl constant [4]. In Figure 3.1 W is the electrode

width, D, the intemal ce11 diameter, and L, the separation between

electrodes.

F i g e 3 1 Schemaiic of u condrrcriiiry ceff rith a three-elecrrode arrangement.

In fact, the ce11 constant is constant throughout; the degree of deviation fiom

ideality depends on the ce11 parameten (D, W and L). Figure 3.2 shows a

typical conductivity cell behaviour.

O O 3 6

Specific Conductivity, (&/cm)

23

A simple relationship between the cell parameters and a geometrical ceIl

constant, C, was proposed: Equation 3.1.

where

Using data obtained by F. Tavera during the development of the original gas

holdup sensor (measured ceil constants for di fferent ce11 arrangements) [4],

we can evaluate the effect of the cell parameters on the ce11 constant. From

Figure 3.3 we can see that deviation from th-e geometrical value became less

signiîïcant when the geometricol cell constant decreased. Thus, to ensure

the closest possible estimation of the real ceIl constant, the smallest feasible.

geornetrical value has to be used. C

Another important panrnetcr in thc design of a conductivity ce11 is the

electrode width (W). WC will Jetinc the esposed electrode area by:

.-f,,. = nDCV

Figure 3.3. Deviat ion f iom g~~onrctricul c d consfarrf for 4 cells with 3-elecrrode arrangement. (see =Ippc.nJi\: -4. Tuhie .4.2). Daia O biained by F. Tavera [JI .

Figure 3.4 shows the deviation from geornetrical ce11 constant values

(Equation 3.1 ) for cells with different ratios of AceII/A,v. We can see that the

deviation decreases if the ratio ACcII/A, decreases. As a design criterion a

ratio ACcII/A, equal to I could be chosen. for which W is defined as:

Substituting Equation 3.3 and 3.5 in Equation 3.1 and rearranging terms, we

get that the separation between electrodes necessary to yield a selected ce11 C

constant is given by:

F i 3 Eflect of the rafio ACe,/A,, on deviafion fiom the geomefricai ce11 consfant &y rrsing Equa f ion 3.1 (see Appendk A. Table A. 3). Data obrained &y F. Tavera [4J

3.1. Design criteria for the Syphon Cell

Since the length and diameter of the syphon ce11 are already defined (see

Section 3 4 , the diameter of the orifice, dmifice, at the bottom is the

remaining critical parameter.

Figure 3.5 shows the experimental setup used to characterize different

orifice diameters: 14, 16. 18. 20 and 22 mm. For each diameter, different

water flow rates were passed through the cell until a steady value of H was

obtained.

Fiprc 3 .3 Erperimrntol serup uscd ro churacterize the bottom orifce diurneter of sj phon cell l~&,,/;~$.

A

C . 1

If we apply Bernoulli's pnnciple between point A (ce11 top) and B (ce11

bottom) the water velocity at the top, V,,, is a functinri of H and is given by:

where g is the gravity acceleration. P = 0.5 for laminar flow, and P = 1 for

turbulent flow. The syphon cell diameter is represented by Ds (in this case,

Ds = 7.5 cm). Figure 3.6 shows the experimental data, VA versus H,

obtained for each orifice diameter.

4

9

B

J

H

1

d :13, 16.18, 20 and 22 mm

1 w

Figirrr 3.6. as a funcrion of H for each orajke diorneter, dorflm (See Appendir A. Tables A.4.a and A.4.b).

From section 1.6, when a syphon ce11 is working in a dispersion of slurry

and gas, VA is related to the gas holdup, as follows:

where LT is the syphon ceIl length (in this case 50 cm). Taking equations

3.6 and 3.17 the gas holdup is related to H for a givenVA by:

0 An important parameter to be considered for a particular application, is the

bubble terminal velocity, Ut. A method to estimate this parameter is

presented in Equations 3.8 and 3.9 [2 ] .

where

Figure 3.7 shows the curves V , venus E, (using Equation 3.7), and the

0 bubble terminal velocity, for three different bubble diameters (0.5, 1.0 and

1.5 mm). The bubble terminal velocities were calculated for a water-air

system. For a given application (where E, and db are known), Figure 3.7

indicates that every orifice diameter that gives a V, value lower than U,

(defined by dh) would allow a bubble-free environment inside the cell.

As an example, for a bubble diameter of 1.5 mm, al1 five oritice diameters

could be used (for any E,) because al1 five curves are uiider the terminal

velocity line for a 1.5 mm bubble. If we have a system with E, = 30 % and

db = 0.5 mm, we can use only an orifice smaller than 14 mm.

Figzu-e 3.7. I.:, as a func I ion of G;. -fir rwch or~$ce diameter (LT=50 cm). Bubb1e rerminai \;rlocir ies were esrima~rd us-ing Eqicrrrioris 3.8 artd 3.9 (water-air system. 4 = 1.5. 1. O and 0.5 rnrn)(see Appendik A. IU~/L?.V .-l. 4. tr tmd -4.4. h).

Rearranging Equation 3.6 yields:

where Co and CI are constants u hiçh ;ire independent of the orifice diameter.

Figure 3.8 shows the detemination of Ci, and CI using the data obtained for

the orifices. Substituting these panmeters in Equation 2.17, we can estimate

the maximum orifice diameter as a function of the bubble terminal velocity

and gas holdup (see Figure 3.9).

1200 -- - - . - - ..

Parameters Equation 3.10 Ce = 1.20

Figure 3.8. Determination ofpuron~c.rers Co arrd Cl (Equation 3.10). Data obtained for 5 orifice diamerers (1 4. 16. 18. .?O. und 22 mm).

- - -

4 X I 2 16 20

Bu hbleTcrminal Velocity, cm/s

3.5. Example of Design

A gas holdup sensor was designed for a flotation column application

(mineral system): eg = 10% and db = 1 .O mm. From Section 3.2, Do = 9.6

cm, Ds = 7.5 cm and LT= 5 0 cm (for both cells).

a The two ce11 constants must be the same to use the same conductivity

meter.

O The lowest possible C in this case (LT= 50 cm) is 8 cm.

O Substituting D, and Do in Equation 3.5: W, = 1.9 cm and W, = 1.4 cm.

Substituting C, Ds and Do in Equation 3.6: L, = 7.3 cm and L, = 13.3 cm.

From Figure 3.7. al1 VA values at E, = 10% are lower than U, = 1 1 .I cm/s

(dh = 1.0 mm; water-air system). Considering that smaller bubbles (0.5

m m ) could be present. a orifice diameter, d,nfi,,, of 15 mm is chosen.

Figure 3.10 is a schematic, with al1 dimensions, of the gas holdup sensor

design for this application (E, = 10% and db = 1 .O mm).

Top View Open Cell

Section AB 9.6 cm . -

Syphon Cell

Figr o-e 3.1 O. S c h e s r r i t ic (!/ /lie ,~CJ.V k M / i p pro he J e s i p e ~ l ji>r un applicuiion with .cc 1 0'% (117d dh = 1. O ??l??l.

4.1 Introduction

Two industrial applications in flotation were considered in this work:

mineral concentration (MC0 at Copper Cliff, Ontario) and de-inking of

waste paper (BOWATER at Gatineau, Quebec). The same data acquisition

system is used for both applications: a gas holdup sensor and an electronic

interface that converts the rough signal fiom the sensor to a digital signal,

which can be processed in any cornputer.

This chapter concems the description of the electronic interface, data

acquisition system cal ibration and the two processes. Three gas holdup

sensor units were used. The three sensors have the sarne characteristics as

the ones described in Figure 3 . l 0 (with donecc = 1.5 cm).

1.2. Data Acquisition System.

Figure 4.1 shows a schematic of the electronic associated with the gas

holdup sensor. The sensor (open plus syphon cells) is placed inside the

column. A relay board (Keithley Metrabyte, model REL- 16) perinits each of

the conductivity cells to be measured independently by using a conductivity

meter (Bailey, model TB440). The output of the conductivity meter,

consisting of a signal between O and 10 volts, is converted to digital values

by using an AID converter (Keithley Metrabyte, model DAS-8). The relay

board and the A/D converter are mounted inside a 386.33- cornputer.

b + + Multiplexer

REL-16

1 Conductirneter 1 lllL4

Cornputer 386-33MHz

A A/D Converter " 1 DAS-8

Figirre 4. I . Schematic representation ofthe data acquisition sysrem.

When the digital voltage value (v) is obtained, it is converted to conductance

by using the calibration curve of the conductivity meter (quadratic equation)

relating voltage to conductance, see Equation 4.1.

This conductance value is transformed to conductivity by using the

calibration curve of the specific ce11 that is being read (open or syphon cell).

This ce11 calibration curve is also of the quadratic form. For the open cell,

the equation is as follows:

K,, = An + B,K +C,J2

and for the syphon cell:

K,, = A, + B , K +c,K'

The data acquisition process is driven by a program in Visual Basic 3.0

(Appendix B). Figure 4.2 shows the main menu of the data acquisition

'are.

In order to control the REL-16 and DAS-8 boards, Visual Basic needs

special drivers. To control the REL-16 board and to provide special graph

features the driver package VTX (Keithley Metrabyte) was used. The DAS-

8 board was controlled by using DRIVERLINX-DASSNB (Keithley

Metrabyte).

4.3. INCO: Matte (Cu-Ni) Separation Plant, Column circuit

This plant is located at Copper Cliff, Ontario. The matte separation plant

has five flotation columns (see Appendix C, Figure CA). The main

characteristics of the columns are presented in Table 4.1. Al1 five columns

use Minnovex variable gap jet spargers. Two gas holdup sensors were used

in these tests.

The following experiments were carried out:

a Experiment 1: Sensor 2 \ras placed inside column 2 at a Cmeter depth

for two weeks. The objective \ras to detemine if the syphon ce11 could

work continuousl\. nithout plugginç in an industriai environment.

a Experiment 2 : Sensor I \vas placcd inside column 3 at a 4-meter depth,

and changes in the air f l o ~ rate wcre introduced. Samples of the feed,

concentrate, and tailings. were obtained before and after every air flow

rate change. Grades of Cu. Ni, Co and Fe were determined. Mass

balances were carried out using a method described in the literature [35-

361. The objectives were to determine if the sensor was able to detect

changes in gas holdup. induced by changes of the air -flow rate, and

whether changes in the column performance could be related to gas

holdup.

Table 4.1

Characteristics of lNCO marte separution plant

I i Column I

Height (m) 11.6 11.6 11.6 11.6 Diameter (m) 1.83 1.83 2.13 t -07 Baffles - No Yes / No Y es No S ~ a r g e r TY pe Vanable gap

Air flow (std.m3/hr) 235 - I 233

- - I 340 I 40

<Jas holdup ( O h ) - LU - 25

Midd. Rghr. 1 Lnd L U LIN. 1 Ni C h . 1 Midd. Chu.

Experiment 3: both sensors were placed inside column 3, sensor 1 at a 4-

meter depth and sensor 2 at a 7-meter depth. The objective was to

determine if there were variations in gas holdup with depth.

Calihrdon: Conductivir). meter and Sensor

The conductivity meter calibration was carried out by replacing the gas

holdup sensor with a known resistor, R (R), and reading the voltage value

displayed by the data acquisition software. Using several resistors, a set of

conductance ( 1 / R) values versus voltage was obtained.

The calibration of both cell constants (open and syphon) was carried out by

using electrolyte solutions of known conductivity (without air); the

conductance in each ce11 was measured with the data acquisition software,

using the conductivity meter calibration curve. The conductivity of each

electrolyte solution was measured using a portable conductivity meter

(VWR Scientific, mode1 2052). Calibration parameters for the conductivity

meter, sensor 1 and sensor 2 used in Equations 4.1 to 3, are presented in

Table 4.2.

Table 4.2

Calibrarion parumeters for the conducrivity meter, sensorl und sensor2

(See Appendix C. Tables C. i to C.5)

Conductivity meter Sensor 1 .

Open cell Syphon cell

Sensor 2 Open cell Syphon cell

Parameter

4.4. BOWATER: De-inking Pilot Flotation Column

The equipment was a pilot flotation column constructed of PVC, with an

intemal diameter of 50 cm and a height of 5.1 m. This column is described

elsewhere [3 1,371. The installed instrumentation consisted of mass air flow

meters, magnetic liquid flow meters ( I ) , pressure transducers (3), and

centrifuga1 pumps (2) with control valves (2). The software used for data

collection and control was FIX DMACS (32-bit) by Intellution. The 110

interface between the software and al1 instruments was an OPTOl serial

board by Transduction. The following parameters were continuously

monitored and registered by FIX DMACS: feed and accepts (i.e., underfiow)

superficial flow rates, superfi~cial gas rate, gas holdup obtained fiom pressure

values, bottom pressure, middle pressure, and level.

One gas holdup probe was used in this application: sensor 3. Two kinds of

bubble generating devices were used in these experiments: five sintered

stainiess steel powder (porous rigid) spargers and one variable gap

(Minnovex) sparger.

The experiments carried out were:

O Experiment 1: the E, versus J, was obtained by using the five porous rigid

spargers for three di fferent conditions (Hf= 100 cm; r = 3 min, Hf = 100

cm: r = 6 min, and Ur = 50 cm; t = 6 min). The flotation effkiency and

brightness gain were measured.

Experiment 2: from an initial condition with 5 porous rigid spargen, the

system was evaluated after closing one sparger, and then two spargen.

Two initial conditions were evaluated (J, = 2.5 cm/s; Hf = 100 cm; t = 3

min, and J, = 2.5 cmk; Hf = 100 cm; t = 6 min). The objective was to

detemine the effect on gas holdup and column performance resulting

from a sparger surface area (A,) decrease.

a Experiment 3: fiom an initial condition with 5 porous rigid spargers, the

system was evaluated after closing one sparger, and increasing J, until the

previous gas holdup with five spargers was reached. Two initial

conditions were evaluated (J, = 2.0 crn/s; Hf = 50 cm; T = 6 min, and J, =

1.5 c d s ; Hf = 50 cm; r = 6 min). The objective was to determine if

performance was the same for the same gas holdup obtained in different

ways.

a Experiment 4: the E, versus J, was obtained by using a variable gap

sparger. For this experiment the air flow meter was set to the maximum

(J, = 3.5 cm/s) air flow rate for that gap. The objective of this experiment

was to compare the response o f the system to both sparger types.

O In order to limit errors, al1 the experiments were evaluated 15 minutes after

the steady state was achieved (stable signals). To obtain the flotation

efficiency and brightness gain, pads of feed and accept pulp were prepared

according to the CPPA C.4U method [%]. An average of 10 ERIC values

(ink concentration in the pad) were obtained for flotation efficiency

calculations with an average of 1 O brightness values for brightness gain.

Cnlibration: Conductiviîy meter and Sensor

The calibration procedure of the conductivity meter and the sensor for this

application is the same as the one described in Section 4.3. Calibration

parameters for the conductivity rneter and sensor 3, used in Equation 4.1 to

3, are presented in Table 4.3.

Table 4.3

Calibration parameters for the conductivity meter and sensor 3.

(See Appendix C. Tables C. 6 fo C.8)

1

Conductivity meter Sensor 3

Open cell Syphon cell

Parameters

-7.63239

-0.2756 -0.23 187

4.29777

O. 14052 0. i 2879

0.07425

0.000 15 0.00069

Chapter 5: INCO Results

This chapter describes the results obtained in columns 2 and 3 at the Matte

Separation Plant, INCO limited (Copper Cliff, Ontario). Two gas holdup

sensors were used in these tests. The main objective of these expenments was

to evaluate the re-designed gas holdup sensor under industrial conditions

(especially with regard to plugging of the syphon cell). One o f the sensors

0 was tested continuously for 17 days in column 2. A second test was run in

column 3, to determine if the sensor was able to detect changes in gas holdup

from changes in the air flow rate to the column. Finally, both senson were

installed in column 3. to detemine variations of the gas holdup value with

depth. The raw data are given in Appendix D.

5.1. Test of Reliabiliîy

Figure 5.1 shows the results obtained by gas holdup sensor 2 in column 2 from

March 1 3 to, March 30 ( 1 998 ). Figure 5. l .a shows the gas holdup values and

Figure 5.1 .b, the conductivity values of the dispersion and slurry.

Figwe 5- I . Results on colt~mn 2fiom kfurch 13 to March 30 (sensor 2): (a) gus holdup. (b) conc/irctirity values (dispersion and s lurry). See Appendix D. Tables D. I fo D. 18.

Al1 the values plotted in Figure 5.1 are the averages of fifieen minutes of

a operation. The actual rate of sampling is 1 value per minute. Point A in

Figure 5.1 shows a spike, probably the level in the column was below the

sensor position. The colurnn quickly recovered normal operation, both

conductivity values increasing (unsteady state situation). Between point B

and point C (about 1 day) the column was shut down, both conductivity values

were close to zero since the column was empty. At point C the column was

re-started. Between point C and point D the plant was in unsteady operation.

At point E the air flow was increased (gap between two conductivity values

increases) leading to a gas holdup increase. AAer point E the gas holdup was

stable around 8%, although the conductivity values continued to present

notable changes.

During the test period, the syphon ceIl did not experience any sign of

plugging, even though the column was shut down for almost one day. In the

previous probe design a column shut down caused the syphon to plug. Afier

the test, the sensor was checked with solutions of known conductivity and it

did not show significant deviation from the readings prior to the test, Le., the

calibration (K versus K ) had not changed.

5.2. Response to a- Change of the Air Flow Rate.

Figure 5.2 shows the gas holdup response to a change in the air flow rate

injected into column 3. All the values plotted in Figure 5.2 are the averages of

5 minutes of operation (sampling rate: 1 value per minute). Sensor 1 was

placed at a 4 meter depth.

Figure 5.2. Gas holdup response to changes in the air flow rate to cofumn 3 (31-March- / 998). See Appendix D. Table D. 19

At point A, a change in the superficial air rate was introduced: fiom 1.83 cm/s

to 1.03 crn/s. The gas holdup response was slow, reaching a steady gas

holdup after one hour. In contrast, when the air rate was decreased, fiom 2.03

c m k to 1.69 cm/s (point B), the system quickly reached a new gas holdup.

The same data is presented in Figure 5.3, but in terms of conductivity values

and ratio y (dispersion conductivity, open cell, over slurry conductivity,

syphon cell) responses to air flow rate changes. From this figure we can see

that at point B, the dispersion conductivity decreased (increasing the

difference between the two conductivities and giving a higher ratio y).

Ratio gamma I

i Syphon cell (slurry) - k

Figure 5.3. Conduc~iviy values and ratio responses to changes in the air j lo\r . rare ro coltrrnn 3 (3 1 -1March- 1 998). See rlppemiix D. Table D. 1 9.

Samples of the feed, concentrate and tailings were taken and the grades of Cu,

Ni, Co, and Fe were obtained (see Appendix D, Table D.30). Representative

samples of the three streams were analyzed 10 times for each of the 4

elements. This allowed an estimation of the relative standard deviations of

each stream for mass balance calculation purposes (weight calculations, see

Appendix D, Table D.3 1 ).

The adjusted grades, obtained from the mass balances, are presented in

Appendix D, Table D.22 The recoveries and typical deviations (95%

confidence interval) for each element are presented in Appendix D, Table

D.23. Unfortunately, typical deviations of Coppet, cobalt and iron are high

&d conclusions based on these elements are avoided. Fortunately, nickel

typical deviations are low, and can be used.

Figure 5.4 shows the recoveries of copper, nickel, cobalt and iron. Al1 of the

elements exhibit the same trend. Between A and B, we see a slight increase in

the recoveries initially, with a larger decrease at the end. From Figure 5.2, we

know that the gas holdup was increasing duhg this period.

This could be explained by the ohsenations in Figure 5.5. Figure 5.5 shows

the variation in level ( froth depth) during this experiment. From this figure

we can see that the level increased strongly at the end of the period A-B. This

could be related to the recovery decrease.

Figrcre 5.3. Fror h Jeph \*ariations of coltrmn 3 (3 I -March- 1 998). See Appendix D. Table D.24. Dura obtained 6). using a conJltc~irity probe ro d e r e . ~ level.

5.3. Gas Holdup Readings at TWO Different Depths.

Sensor 1 was placed at a 4-meter depth and sensor 2 at a 7-meter depth inside

column 3. Figure 5.6 shows similar results for both sensors. The average gas

holdup of sensor 1. at a +meter dcpth. was 4.7% + 0.1% (95% confidence)

and sensor 1. at a 7-mctcr depth. was 4.9% + 0.1% (95% confidence). These

results show that there was not a significant difference between the two

measurements. This is in contrast with some of the previous work, showing a

gas holdup decrease with increasing depth [ M I .

Figzrre 5.6. Gus holdup signals of sensor I and sensor 2 working at ciifferen! depths inside column 3: sensor 1 ai 4-meter @th: sensor 2 at 7-meter deprh (3-April-1998). See IIppendix D. Tables D.25.a to D.25.c.

CHAPTER 6: BOWATER RESULTS

This chapter describes the results obtained in the pilot column at the De-

inking of Waste Paper Plant, BOWATER Pulp and Paper Inc (Gatineau,

Quebec). Sensor 3 was used in these tests. One objective of the

experiments was to evaluate the re-designed gas holdup sensor on fibre

pulps, which are quite different from mineral slumes. Gas holdup values

obtained from pressure and conductivity were compared. The effect of gas

holdup on flotation efficiency was explored. The relationship between

bubble surface area flux (derived from drift flux analysis) and gas holdup

was tested. Changes in sparger surface area were induced to observe the gas

holdup response and its effect on flotation performance. Finally, a jetting

sparger (MINNOVEX) was tested. The raw data are given in Appendix E.

6.1. Gas Holdup Measurements from Pressure and Conductivity

Figure 6.1. summarizes the results including the ones for the jetting sparger.

From Figure 6.1, some scatter is evident. One explanation is that

conductivity gives a more local estimation than pressure, which is an

average between two points. Differences could be expected if gas holdup

would be distnbuted uniformly in the column

Figure 6.1. Gas holdup BOWATER tests. Every .-lppenJi-r E. Table E. 1.

estimaies fjom conductivity and pressure meusuremenrs for point is the average of 15 minutes of stable operation. See

a

To test it if the difference is signifkant, the t-student value for paired

samples rnust satisfy -1.03 < t (= 1.61) < 3.03, where 2.03 is obtained for

95% confidence and 36 degrees of freedom (t distribution). From this

analysis we c m conclude that the gas holdup obtained from conductivity

( 1 1.47% t 0.44%) is not different from the value given by pressure

( 1 1.77%). A slope analysis (m = 0.87 f 0.02) suggests that the gas holdup

obtained from conductivity, tends to underestirnate the one obtained fiom

pressure at low gas holdup, and to overestimate, at high values.

6.2. Effect of Cas Holdup on Flotation Efficiency

The relationship between flotation efficiency and gas holdup for two

residence times (3 and 6 min) is shown in Figure 6.2. It is clear that gas

holdup affects the flotation eff~ciency, but the latter is also affected by the

residence time, as we expected fiom the discussion in Section 2.7.

Another factor that can affect flotation efficiency is the fioth height. Figure

6.3 shows flotation efficiency venus gas holdup for two fioth heights (50

and 100 cm). For this particular case, Figure 6.3 shows fkoth height does not

seem to strongly affect the flotation performance. One might have expected

a higher recovery in the 50-cm froth case.

r = 6 min

O

T = 3 min

Cas holdup, %

Figwc 6.2. Gus hoidup elflecr on jlortrrion efficiency for IWO reside)tce rimes (3 und 6 r i Froih heighr (Hj is 100 cm. See Appendix E, Tables E. 2 and E. 3.

- - . - . - - - - - -

8 12

Cas holdup, ./.

Figure 6.3. Gas hoidup effecr on j h u r i o n eflciency for hvo fi.ofh heights (50 and 100 cm). r = 3 min- See Appendk E. Tc~hlt~s E.2 u r ~ d E.3.

The effect of sparger surface area. A,. on gas holdup, for two conditions, is

shown in Figure 6.4. When sparger surface area decreases, gas holdup

decreases for the same J,. as seen in Section 2-82. From this figure, we can-

see that the sensor reveals changes in the noise of the gas holdup signal. It

implies that the signal noise is rrlatcd to the bubble size, which increases as

the number of spargen decreasc.~. Thus. the gas holdup sensor could be a

useful diagnostic tool to monitor the spargw.

A test was perfomed to evaluate i F two similar values of gas holdup,

obtained with different sparger surface areas, give similar flotation

efficiency. Figure 6.5 shows the results for two initial conditions.

3 spargers

. - .

Timc

Figtire 6.4. Eflec~ of spurgw srlr$utu orrJtl on prs holdup for mfo conditions: (a) H/ = 100 cm: r = 6 min: 4 = 7 5 ends (hl I I , = i 00 cm r = 3 min: J, = 2.5 cm/s . See Appendix E. Tubles E.3 r~ : , E. 9.

.. - 0 60 . 1.33 cmh k 4 sprrgerr

2.93 cmh 4 spargem

Cas hoidup. O/,

Figure 6.5. Eflecr of sparger surface areo onfloiafion efficiency. Hf = 50 cm; r = 7 min. Ser -4ppendi-Y E. Tables E. 2 and E. 3.

Figure 6.5 indicates that gas holdup alone cannot de fine flotation efficiency.

From this figure we can see that a gas holdup even higher than the initial.

one. produces a lower flotation efficiency. We assume here that the system

chemistry has not changed during the test, which should be verified in future

work.

From Section 1.9, a relationship between gas holdup and flotation

performance could be expected. Figure 6.6 presents flotation efficiency

versus gas holdup for residence times close to 6 min. A relationship is

suggested. A mode1 fit is given with the parameters and a X' statistic

indicated.

FE. =7l.8(1 -~xp(.(&&6)~~)) xz = 3.3 c 16.9 (95% confidence. 9 df)

Gas holdup, %

Figure 6.6. Floiarion eflciencv versus gus holdup for r = 6 min. See Appendix E. Tables E. 3 and E. 3.

A relationship between gas holdup and flotation rate constant has been

reported in mechanical cells for particular cases [ I l . Figure 6.7 shows the

pas holdup versus the collection zone rate constant, k. The collection zone C

rate constant was estimated using the considerations outlined in Section 1.7

(plug flow), assuming Ri equal to 0.5 (same for froth height of 50 and 100

cm). Al1 the results are shown in Appendix E, Table E. 10.

I t is evident from Figure 6.7 that there is some correlation (R' = 0.72)

between the collection zone rate constant and the gas holdup. The scatter in

the data could be explained by changes in the chemistry or rheology of the

feed during the tests.

5 10 15

Cas holdup, ./.

Figure 6.7. Collection zone rate consfant versus gas holdup. Plugflow inside the colurnn M*US asslrmed to calculate rate constant. Froth recovery. RJr is eqtrol50%.. See Appendir E. Tables E. 2 and E- 10.

6.3. Relationship between Sb and E,

To estimate the bubble surface area flux, the bubble diameter was calculated

using drift flux analysis (described in Section 2.10). The properties of the

liquid were assumed to be those of water (psi = 0.01 gkms; psi = 1.0 &cm3)

and bubbles were considered without a load of particles (pb = O gkrn3).

Results giving bubble diameters bigger than 0.2 cm were discarded.

Figure 6.8 shows a linear relationship between gas the holdup and the bubble

surface area flux, as expected from Section 2.9. This relationship shows a

good fit (R' = 0.95).

8 12

Cas holdup, O h

Figure 6.8. Bubble surface area flux versus gus holdup al point is the average of 1.5 minutes of stable operurion. See

323 K and pressure P. Erery Appendix E. Table E. 10.

Following Figure 6.8, if we plot J,/dh (at a pressure and temperature of the

pas holdup measurement) versus gas holdup, we obtain another linear C

relationship (see Figure 6.9). This time however, a slight improvement in

the fit is observed (R' = 0.96). Such a relationship could be used to estirnate

the bubble diameter for a given application.

Bubble surface area flux has been reported having a linear relationship with

the flotation rate constant [ I l . From Figure 6.8 we could anticipate also a

linear relationship between gas holdup and the flotation rate constant. But,

as we saw in Figure 6.7, the gas holdup is not linearly related to the flotation

rate constant.

Figure 6.9. Cas hoIdup i*ersirs ruiio Jg/Jh (ut 323 K and pressure of measwement) . Eivery point is the average 4 / 5 rniniifcs 0 f stable operation See Appendix E. Tubles E. 2 un J E. 3.

Figure 6.1 0 indicates some degree of correlation between bubble surface

area flux and the collection zone rate constant. The intercept represents the

minimal Sh required for overflow of the column (Sb0 = 13 US) .

To explain the linear relationship hctween gas holdup and bubble surface

area evident in Figure 6.8. consider a hubble surface area moving through a

vessel, a (m2/s), and a section of thc vcssel of length h (m), with a gas

holdup E, (%). Then. the tirne t<iLcn h?. u to pass through h, (s), is given

by :

O 20 10 60 80 100 120

Bubblc surface area flux, 1 /s

Figure 6.10. Collecrion zone rate constuttt versus bubble surface areayux (ut 273 K and pressure P). Pfugfloar r w s ussunicd to calciriate the rate consfanf. See Appendix E. Tubles E. 2 and E. 1 0.

If we take h = db. then we have a T,," that is independent of db. Rearranging

Equation 6.1 :

The parameter ruo must be a tunciion of the superficial liquid velocity, the

rheology (viscosity and densit~ ot'thc liquid), and bubble load. From Figure

6.8 we can see that the paranleter r.," was independent of JT (the superficial

liquid rate) for these tests. If the system had any significant fluctuation in

rheology, a non-linear relationship between gas holdup and bubble surface

area flux would be expected. In this case, a linear relationship was obtained,

with constant raO. This suggests a constant rheology or may just be an

artifact of taking water as the liquid to estimate bubble size using the drifi

flux analysis. Ta have considered a constant rheology (water) may explain

the absence of a better correlation between the bubble surface area flux and

flotation rate constant.

Substituting Equation 2.1 9 in Equation 6.2 and rearranging:

This relationship is demonstrated in Figure 6.9. Taking the slope of the plot O in Figure 6.8' we have that T, is about I l ms. Therefore, the bubble

diameter c m now be estimated directly from Equation 6.3.

CHAPTER 7: CONCLUSIONS AND FUTURE WORK

7.1 Conclusions

- Some criteria to design the gas holdup probe were established:

The electrode width is a critical parameter in a conductivity

flow ce11 design. An electrode width equal to a quarter of

the cell diameter is recommended.

The lowest possible cell constant is recommended. Cells

with low ce11 constant behave closer to ideality (less

deviation from geometri cal ce1 1 constant).

A simple relationship for three electrode cells was verified,

which relates the geometrical cell constant to the gap

between clectrodes.

A new syphon cell design. having a conical bottom with a

small orifice, was presented.

The syphon ce11 diameter was increased (7.5 cm) from the

original design to allow a larger diameter orifice, which

decreases the chance of plugging.

0 An equation that relates the velocity inside the syphon ce11

(cylindrical part) to the gas holdup and the orifice diameter

was presented. The fluid velocity inside the ce11 must be

lower than the terminal velocity of the rising bubbles to

ensure a fiee-bubble environment inside the syphon cell.

The new design of the gas holdup probe was tested 17 days at INCO Matte

Separation Plant (Copper Cliff, Ontario). The syphon cell did not plug. The

gas holdup probe was able to detect changes in gas holdup upon changes in

air flow rate. Flotation performance (recovery) did not change with air flow

rate, probably due to uncontrolled changes in froth depth.

Tests camied out in waste paper de-inking at BOWATER (Gatineau,

Quebec) showed the gas holdup values obtained by the probe agreed with

those obtained from pressure measurements. This indicates that a bubble-

free environment was obtained inside the syphon cell.

Tests carried out using sparger surface areas showed that the noise

associated to the gas holdup signal could be related to the bubble diameter.

This noise may be used to diagnose sparger operation.

Gas holdup is an important variable in flotation, but alone does not define

flotation performance. The relationship between the gas holdup and the

collection zone rate constant showed some degree of correlation. Chernistry

and rheology of the feed are other factors to be considered. Bubble surface

area flux also showed little correlation with the collection zone rate constant.

The dispersion in the Sb data might be related to the assumption made to

calculate the bubble diameter using the drift flux analysis.

From a conceptual analysis, the bubble surface area flux would be a function

of the gas holdup and a parameter raO. This parameter is defined as the time

it takes a bubble surface area, moving through a vessel, to cover a distance

equal to the bubble diameter. This parameter seems to be constant for a

system with constant rheology. A Iinear relationship between gas holdup

and bubble surface area flux is predicted. This linear relationship was

verified in Chapter 6.

7.2 Future Work

The potential use of this sensor in the diagnostic of sparger operation may

justify further exploratory research in that direction. Trying to relate

changes in the noise of the gas holdup signal to changes in bubble diameter

is one approach.

An exhaustive analysis of the gas holdup relation to the bubble surface area

f l ~ i s is recommended. Minerd systems may have smaller changes in

rlicolorrv L d than those we suspect occurred For waste paper de-inking.

Gorain, B.K., "The Effect of Bubble Surface Area Flux on the Kinetics of

Flotation and its Relevance to Scale-Up". Ph.D. Thesis, University of

Queensland, Australia, 1997, pp. 59-1 44.

Finch, J.A. and Dobby, G.S., Column Flotation, Pergamon Press, 1990.

Gomez, CO., Tavera, F., Finch, J.A., Perez, R. and del Villar, R., 1995

(b). "Continuous Gas Holdup Measurements in Flotation Columns".

Proc. IV Meeting of the Southem Hemisphere on Mineral technology,

Castro S. and Alvarez J. Editors. Vol. 3, Concepcion, Chile, 1994; pp.

123- 136.

Tavera, F., "Flow Cells to Measurr Electrical Conductivity: Use in

Estimating Gas Holdup in Flotation Systems". Ph.D. Thesis, McGill

University. Canada. 1996.

Tavera. F., Gomez. C O . and Finch. LA.. "Novel Cas Holdup Probe and

Application in Flotation C'olut~ins". 'I'rans. Instn. Min: Metall. (Sect. C:

Mineral Process. Esir. MetriIl. ). 105. May-August 1996, pp. 99- 104.

Condon, E.U. (Ed.). Handbook of Physics. Chapter 9, McGraw Hill Book

CO., 1 967, pp. 146- i 69.

Andrews, D.H., I ntroductory Physical Chemistry, McGraw Hill Book

Co., 1970, pp. 689.

8. Barrow, G.M., Physical Chernistry, McGraw Hill Book Co., 1973, pp.

787.

9. Adamson, A.W., A Textbook of Physical Chemistry, Academic Press,

1979, pp. 953.

1 O.Kasper, C., "The Theory of the Potential and Technical Practice of

Electrodeposition", Trans. Electrochem. Soc., 1940, Vol. 77 pp. 353-384,

Vol. 78 pp. 131-161.

1 1 .Atkins, P. W., Physical Chemistry, W.H. Freeman and Company, 1981,

pp. 1095.

P.Levine, LN.. Physical Chemistry, McGraw Hi11 Book Co., 1988, pp. 920.

13.Braunstein. J. and Robbins, G.D., "Electrical Conductance Measurements

and Capacitive Balance". J. Chem. Ed.. 197 1, Vol. 48, No. 1, pp. 53-59.

14.Gilmont. R.. and Walton, R. F.. "New Design of an Electrolytic Ce11 for

the Study of Electroplating Phenomena", Jour. Electrochem. Soc., 1956,

Vol. 103, No. 1 O, pp. 549-552.

I5.Lord Rayleigh, "On the Influence of Obstacles Arranged in Rectangular

Order Upon the Propenies of a Medium", Phil. Mag., 1892, Vol. 34, pp.

48 1-502.

lG.Meredith, R.E and Tobias, C.W., "Resistance to Potential Flow through a .

Cubical Array of Spheres", J. Appl. Phys., 1960, Vol. 3 1, No. 7, pp.

17.Wagner, C., "The Scope of Electrochemical Engineering", in Advances

in Electrochernistry and Electrochemical engineering, Vol. 3. (ed.

Tobias, C. W.), Inerscience Publishers, 1962, pp. 1 - 14.

18.Shen, G., "Bubble Swarm Velocities in a Flotation Column". Ph.D.

Thesis, McGill University, Canada, 1 994.

19.Maxwel1, J-C., A Treatise of Electricity and magnetism, 3rd Edn., Oxford

university Press, 1 891. Vol. 1. part II, Chapter IX, pp. 435-449.

20.Tumer. J .C .R.. "Two-Phase Conductivi ty-The Electrical Conductance of

Liquid Fiuidized Beds of Spheres". Chem. Eng. Sci., 1976, Vol. 3 1 , pp.

487-492.

2 I .Y ianatos. J.B.. "Column Flotriti.on Froths", Ph.D. Thesis, McGill

University. Canada. 1987.

1 2 . X ~ . M.. "Radial Gûs Holdup Protilr and Mixing in the Collection Zone

of Flotation Columns". Ph.D. Thesis, McGill University, Canada, 1990.

23.Uribe-Salas, A., "Process Measurements in Flotation Columns Using

Electrical Conductivity". Ph.D. Thesis, McGill University, Canada, 1991.

24. Wills, B.A., Mineral Processing Technology, Pergamon Press, 1993.

25.Banisi, S., Finch, J.A., Laplante, A.R. and Weber, M.E., "Effect of Solid

Particles on Gas Holdup in Flotation Columns - 1. Measurements",

Chem. Eng. Sci., 1995, Vol. 50, No. 14, pp. 3319-2334.

26.Banisi, S., Finch, LA., Laplante, A.R. and Weber, M.E., "Effect of Solid

Particles on Gas Holdup in Flotation Columns - II. Investigation of

Mechanism of Gas Holdup Reduction in Presence of Solids", Chem. Eng.

Sci., 1995, Vol. 50, No. 14. pp. 335-2342.

27.Xu. M. and Finch. J.A.. "Eflrct of Sparger Type and Surface Area on

Bubble Size in a Flotation colurnn". Canadian Metallurgical Quarterly,

1989. Vol. 38 ( 1 ). pp. 1-6.

XDobby. G.S.. Amelunscn. K. and Finch. J.A., "Column Flotation: some

Plant Esperience and h l d c l I>cvelopment", Symp.. Automation for

Mineral Resourcc Dwcl.. .-Iiisir<iliri. 1985. pp. 559-264.

29.Arnelunsen. R.. "'Hi2 !tlcch.irii~-3 <if Opention of Column Flotation

Machines". Proc. 17"' :\nnci.il Mxt ing of the Canadian Mineral

Processors. C.1 .M.. Jmucir? I W 5 . pp. 1 - 7 4 1 .

30.Gorain, B.K., Manlapig. E X . and Fnnzidis, J.P., "The Effect of Gas

Dispersion Propefiies o n ihc Kinctics of Flotation", Proc. of the

international Symposium on C'olumn Flotation (Gomez. C.O. and Finch,

J.A. ed.), the Metallurgical Society of the Canadian Institute of Mining,

Metallurgy and Petroleum, Montreal, Canada, 1996, pp. 299-3 1 3.

3 1 .Leichtle, G.F., "Analysis of Bubble Generating Devices in a De-inking

Column", Master Thesis, McGill University, Canada, 1998.

3?.Yianatos, J.B. and Puelle, P.C., "Development and Calibration of a

Flotation Bubble Size Sensor", Proc. IV Meeting of the Southem

Hemisphere on Mineral technology, Castro S. and Alvarez J. Editors.

Vol. 4, Concepcion, Chile. 1 994, pp. 1 57- 164.

33. Wallis, G.B., One Dimensional Two-Phase Flow, McGraw Hill Book

Co., 1969, Chapter 9.

34.Banisi. S. and Finch. J A . . Tcchnicd Note: " Reconciliation of Bubble

Size Estimation blethods Using Drift Flux Analysis", Minerals Eng.,

1997. Vol. 1 O. p l i 367-380.

35. Simpson. D E . \'ol lcr. \'.K. and i%rrcti. M.G.. "An Efficient Algorithm

for Minera1 Procrssin< d;)t;i Adjiistriicnt". Int. J. Miner. Process., 199 1 ,

NO. 3 1 , pp. 73-96

36.Cortés-Lopez. F.. "liwluriç iim del Comportamiento de la Molibdenita en

el Circuito de Flotacion C'olcctiva de Division Andina de CODELCO-

Chile", B.Sc. Thesis. Uniwrsidrid Técnica Fedérico Santa Maria, Chile,

1994.

3 7. Watson, J.A., "De-inking Recycled Paper Using Co lumn Flotation".

Master Thesis, McGill university, 1996.

38.CPPA Usefbl Method C.4U, "Forming Pads for Reflectance Tests of

Pulp", CPPA Usefbl method, CPPA Standard Method.

APPENDIX A

Table A. 1. Conduciivity and conductame values obtained for nt.0 3-elecrrode c e f k Data obtained by Francisco Taveru [4]

Tub le A. 3. Cell constants (measured and geometrical) for three electrode width (3- electrode cells). D = 7.5 cm: L = 10 cm. Data obtained by Francisco Tmera [-/]

Tubir .J.-l.u. Liqrrid relocify inside the syphon cell as function of gus holdup and H (as

18 mm H Eg VA

(cm) ( O ) (cm/s)

3.6 7.2 4-20 4.0 8.0 4.47 6.1 12.4 5.67 6.7 13.4 5.94 8.7 17.4 6.70 8.9 17.8 6.82

I 1.8 23.6 7.96 12.2 24.4 8.1 1 14-5 29.0 8.85 15.1 30.2 9.10 17.3 34.6 9.75 18.1 36.2 10.00

. 14 mm

H

(cm) 7 6 -. 4.1 5.6 7.4

1 1 . 1 1 1 . 1 15.5 19.9

4 ( O h )

3 2 8.2

I 1.2 14.8 22.2 22.2 3 1.0 39.8

V A (cm/s)

7 70 - .- 2.87 3.30 3-84 4.71 4.75 5.59 6.30

Table A. 4. b. ContinuarionJion Table A. 4. a.

Module INITIAL

Variable definition

Option Explicit

Dirn dia$ Dirn hor$ Dirn fiIn$ Dim d As Long Dirn b As lnteger Dirn i As lnteger Dirn dummy As lnteger Dirn InputMode As lnteger ' best Al mode, moâified by Get-LOD-lnfo Dirn StopEventType As lnteger ' manual (single) or automatic (continuous) Dirn TriggerSource As String ' trigger source Dirn AI-Channel As lnteger ' AID channel Dirn vbdataarrayl () As Single ' VB array to store voltage values Dim SampleRate As Long ' sample rate in Hz

a Dirn SampleTime As Single ' calculated display time interval for graph Dirn numsamples As lnteger ' number of sarnples Dirn NumGraphPoints As lnteger ' number of points to graph Dim UpdatingGraph As lnteger ' flag to indicate data is being processed

Sub Form-Load ()

'in itialize program variables InitGlobalVars

'activating readings b = l

'setting open cell: relays 1 and 2 d = 3

'open file that contains parameters Open "c:\holdup\kparam mi" For Random As 81 Len = 6

'loading conductivity meter parameters (3 ranges) Get #1, 1, ar l Get #1, 2, br l Get #1, 3, cr l Get #1, 4, ar2 Get #1, 5, br2 Get #1, 6, cr2 Get # I l 7, ar3 Get #1, 8, br3

Get #1, 9, cr3 'defining active conductivity meter range

Get #1, 22, r 'loading open cell parameters

Get # I l 10, a01 Get #1, 11, bo l Get #1, 12, col

'loading syphon ceII parameters Get #1, 13, as1 Get #1, 14, bs1 Get #Il 15, csl Close #1

'setting relays related to the conductivity meter I f r = l T h e n a = O If r = 2 Then a = 40960 If r = 3 Then a = 61440

'setting control status editexesrEnabled = False 'start disable command2.Enabted = False 'sensor disable comrnand3.Enabled = False 'conductivity meter disable command5.Enabled = False 'end disable cornmandl .Enabled = True 'stop enable

'setting date and time dia$ = Date$ hors = Tirne$

'defining name of data file fiIn$ = "c:\data\" + MidS(dia$. 9. 2) + Left$(dia$, 2) + Mid$(dia$, 4, 2) + ".datWt

'displaying date and time on INITIAL from fecha.Caption = dias hora-Caption = Times

'closing relays rel-DOScalar = d + a rel.ArmState = 1 'running relay control

End Sub

Sub InitGlobalVars ()

'initialize variables 'Al-Channel = O 'SampleRate = 10000 'numsamples = 1500 StopEventType = DL-TCEVENT

End Sub

Sub reiJrocessDone (WamingCode As lnteger)

a . 'routine associated to relay control

reLArmState = 2 Call waiting(8) das8

End Sub

'stopping relay control 'calfing delay routine 'calling AIC converter routine

Sub waiting (s As Integer) 'delay routine

Dirn ax As lnteger Dim m$ Dirn tnew$ Dirn tne - Dirn told

ax = O told = O

30 ttt$ = Tirne$ tnew$ = Mid$(ttt$, 8, 1) tne = Asc(tnew$) If told = O Then told = tne If tne = told GoTo 30 ax = ax + 1 told = tne If ax < s GoTo 30

End Sub

Sub das8 () ' N C converter routine

ReDim vbdataarrayl(1500) Dim j As lnteger Dirn in As lnteger 'defining number of samples and sampling rate (Hz)

numsamples = 1500 SampleRate = 10000

'initiaking N C converter board init-device sr-ai

'defining A/C converter parameters sr-ai. Reqop = dldevice sr-ai.Req-device = O sr-ai. Req-subsystem = d l a i soi.Req-mode = dljolled

'starting AIC converter

sr-ai.Req-0~ = dl-start sr-ai. Refresh

' Call procedure ta start data acquisition GetAnalogBuffer sr-ai, InputMode, AI-Channel. numsamples. SampleRate. StopEventType 'Filling the output array

durnmy = VBArrayBufferConvert(sr-ai, O, 0. numsamples, vbdataarrayl (O), DL-tsingle, O, 0)

End Sub

Sub SR-AI-ServiceDone (task As Integer, device As Integer, subsystem As Integer, mode As Integer) 'routine associated to N C converter control

If (d = 3) Then 'identifying cell 'calculating the average voltage measured for the open cell

aux = O For i = 1 To 1500

aux = aux + vbdataarrayl (i) Next i vo = aux / 1500

'calculating open cell conductance aux = conductance(r, vo) vol = VO

'calculating open cell conductivity ko = a01 + bol ' aux + col ' aux * aux

'displaying open cell conductivity advafuel .Text = Format(Str$(ko), "#.##") advaluel Refresh advaluel .ArrnState = 1 'running open cell label control

Else 'calculating the average voltage measured for the syphon cell

aux = O - For i = 1 To 1500

aux = aux + vbdataarrayl(1) Next i vs = aux 1 1500

'calculating syphon cell conductance aux = conductance(r, vs) vs1 = VS

'calculating open ceIl conductivity ks = as1 + bs1 ' aux + csl ' aux ' aux

'dispiaying open cell conductivity

advalue2.Text = ~orrnat(Str$(ks); "#-WB) advalue2 .Refresh advalue2.ArmState = 1 'ninning syphon cell label control

End If

End Sub

Function conductance (r As Integer, aux As Single) As Single

If r = 1 Then conductance = arl + b r l ' aux + c r l ' aux 'aux If r = 2 Then conductance = ar2 + br2 " aux + cr2 ' aux ' aux If r = 3 Then conductance = ar3 + br3 * aux + cr3 *aux ' aux

End Function

Sub ADVALUEI-ProcessDone (WarningCode As Integer) 'routine associated to open cell label control

advaluel .ArmState = 2 d = 12 relay (a + d)

End Sub

Sub relay (m As Long) 'relay routine

closing relays forml .rel.DOScalar = m form 1 .reI.Refresh forml .rel.ArmState = 1

End Sub

'stopping label control ' setting syphon cell: relays 3 and 4 ' calling relay routine

'setting relays

'running relay control

Sub ADVALUE2-ProcessDone (WarningCode As Integer) 'routine associated to syphon cell label controt

advalue2.ArmState = 2 d = 3 graph 1 .ArmState = 1

'stopping label control 'setting open cell relays: 1 and 2 'running conductivity graph control

End Sub

Sub GraphlProcessDone (WarningCode As Integer) 'routine associated to conductivity graph control

Dim aux1 As Currency Dim aux2n As Single

Dim aux3n As Single graphl .Anstate = 2 'stopping conductivity graph control aux2n = Val(advalue1 .Text) If aux2n < .O1 Then aux2n = -001 aux3n = Val(advalue2.Text) If aux3n < -01 Then aux3n = -001

'calculating ratio y auxl = (aux2n) / (aux3n)

'calculating gas holdup aux1 = (1 - auxl) / (1 + -5 ' auxl) eg = auxl ' 100 holdup.Text = Str$(eg) holdup.Refresh holdup.AmState = 1

'displaying gas holdup value

'ninning gas holdup label control

End Sub

Sub holdup-ProcessDone (WarningCode As Integer) 'routine associated to gas holdup label control

holdup-ArrnState = 2 'stopping gas holdup label control GRAPH2.ArmState = 1 'running gas holdup graph control

End Sub

Sub graph2-ProcessDone (WarningCode As Integer) 'routine associated to gas holdup graph control

GRAPH2.ArmState = 2 'stopping gas holdup graph control 'open file to save data

Open fiIn$ For Append As #l Print #1. hor$, Format(vo1, "#.#Mt"). Format(vs1, "#.fi##'') Close #1

'determining if click on stop If b = O Then

'stopping data acquisition b = l d = 3 'setting open cell: relays 1 and 2 editexesr. Enabled = True 'start enable cornmand2.Enabled = True 'sensor enable command3.Enabled = True 'conductivity meter enable command5.Enabled = True 'end enable editexesr.SetFocus GoTo 100 'stop data acquisition

End If Start 'calling routine start

100 End Sub

Sub start () 'routine start

'defining date and time .

hor$ = Tirne$ dia$ = Date$

'defininig name of data file fiIn$ = "c:\data\" + Mid$(dia$. 9. 2) + Left%(dia$. 2) + Mid$(dia$, 4. 2) +

".datu 'displaying data and time

fecha-Caption = dia$ hora.Caption = Time$ relay (a + d) 'calling relay routine

End Sub

'if a control (button) is enable can induce an event (click)

Sub Commandl-Click () 'routine executed after click on stop button .

b = O command 1 .Enabled = False 'stop button disable

End Sub

Sub Command2-Click () 'routine executed after click on conductivity meter button

form 1 .Enableci = Faise form3.Show

End Sub

Sub Command3-Click () 'routine executed after click on sensor button

form 1. Enabled = False form2.Show

'initial form disable 'Go to conductivity rneter module

'initial form disable 'Go to sensor module

End Sub

Sub editexesr-click () 'routine executeâ after click on start button

editexesr. Enab ted = False comrnand2.Enabled = False cornmand3.Enabled = False command5Enabled = False commandl .Enabled = True commandl .Setfocus start

End Sub

Sub CornmandS-Click () 'routine executed after click on end button

End

End Sub

'start button disable 'sensor disable 'conductivity meter disable 'end button disable 'stop button enable

'calling start mutine

'end of program

Module General 'Copyright 1994 DriverLlNXNB Source Code Example Library

Option Explicit

Dim i As fnteger Dim dummy As lnteget Global aux As Single Dim Sample() As Single Dim NumSamples As Integer Dim numchannels As lnteger

Conductivity meter parameters Global r As lnteger

Range 0.1 Global arl As Single Global br1 As Single Global crl As Single

' Range 1 .O Global ar2 As Single Global br2 As Single Global cr2 As Single

' Range 10 Global ar3 As Single Global br3 As Single

Global cr3 As Single

' Siphon cell parameters Global as1 As Single Global bs1 As Single Global cs1 As Single

' Open cell parameters Global a01 As Single Global bol As Single Global w l As Single

Global ko As Single Global ks As Single Global eg As Single Global vo As Single Global vs As Single

Global a As Long

Global vol As Single Global vsl As Single

Sub GetAnalogBuffer (SR As Control. Mode As Integer. Channel As Integer. NumSamples As Integer. SampleRate As Long. StopEventType As Integer)

' Specify timing event SR.Evt_Tim-type = DLRATEEVENT SR.Evt-Tim-rateChannel = DCDEFAULlTIMER SR-Evt-Tim-rateMode = DL-RATEGEN SR. Evt-Tim-rateclock = DL-INTERNALA

'lkample rate in Hz SR. Evt-TimratePeriod = DLSecsZTics(SR. SR-Evt-Tim-ratechannet (1 1

SampleRate))

' Specify start event SR.Evt-Str-type = DL-COMMAND

SR-Evt-Str-type = DL-AIEVENT ' SR-Evt-Str-aiChannel = O ' SR-E-Str-aiGainCode = DLGainPCode(SR. -1) ' SR-Evt-Str-aiUpperThreshold = 1000 ' SR.Evt-Str-aiLowerThreshold = 900 ' SR-Evt-Str-aiSIope = True

' Specify stop event SR. Evt-Stp-type = StopEventType

' Specify number and size of buffers SR.Sel-buf_N = 1 SR-Sel-b-ize = DLSamplesZBytes(SR. Channel. NumSamples) SR-Sel-buf-notrfy = True

' Specify data format and channels to acquire SR-Sel-chan-format = DLtNATlVE SR-Sel-chan_N = 2 SR-Selchan-start = O SR-Selchan-startGainCode = DLGain2Code(SR, 1 ) ' unipolar, unity gain SR-Selchan-stop = Channel SR-Selchan-stopGainCode = DLGain2Code(SR, 1) ' unipolar, unity gain

' Initiate the Service Request SRSR-execute = 1

' Display a result message End Sub

Sub init-device (SR As Control)

'initialize device and service request control SR.Req-subsystem = DL-device SR.Req-mode = DL-other SR.Req-op = DL-INITIALIZE

End Sub

BESSEMER MA-€

Table C. 1. Calibration data. conduct ivity meter for M O tests. .

8 Average ' 068 -- 2.069 2.815 2.817 3.575 3.573 4.304 4.299 5.022 5.016 5.707 5.701 6.420 6.410 7.181 7.169 7.838 7.823 8.420 8.a3 8.918 8.897

Tuhle C. 2. Calibra fion data. open ceif qf sensor l for I'VCO tests.

S 10 . -

6.05 7.40 9.55 12.49 15.16 17.93 21.12 23 -85

Average 4.20 6.05 7.39 9.55 1 2.18 15.16 17.93 21-12 23.83

Table C.4. Ca/ibralion data. open ce// of sensor 2 for iNCO tests.

Average

0.62 0.80 1 .O8 1.18 1'86 2.26 2.74 3.12

TXdr C. S. Calibra~ion duru. s~pllon cdl of sensor 2 for INCO rem.

Average

0.62 0.80 1 .O8 1.48 1.86 2.26 2.71 3.1 t

Average

2.797 3.631 4.445 S. 236 5.998 6.789 7.639 8.36 1 8.999 9.490

Table C. 7. Calibrafion duru, open cetifor BOWATER tests.

Average 6.24 10.63 14-19 18.87 24.10 29.25 34.26 39.3 I

Tubk C. 8. Calibrurion duru. s~pitort ce// for BOIK4 TER resrs

APPENDIX D

Table D. 1. INCO Matfe Separafion Plant column 2. Gus holdup dura collected hfarch

99

Table D.2. INCO LMarte Separaiion Plant column 2. Gas holdup data collected i e c h

Lg (Y* ) 8.48

8.47 8.63 8: 50 8.19 8.25 8.52 8.48 8.36 7.9 1 8.27 7.75 7.62 7.02 7.50 7.79 7.83 7.40 7.57 7.x5 7.57 7.47 7 2.3 - - 7: ' 43 7 10 7 52 7 X? O - ' . 5 - o. , -8

?hl - -0 - -n - M i

- 1: - -- - t.1 - - 7

+

- t,G - C'

- 'JI t X? - 54 - -* 1 1 -

7 71 7

'hl 7 3

k sypnon (mS/cm)

7 - - 1.01 1-01 1 .O0 1 .O0 0.99 0.99 0.99 0.99 1.00 1.00 1.01 1 .O2 1 *O2 1 .O2 1 .O3 1 .O3 1 .O3 1-03 1 .O3

. 1 .O3 1 .O1 1.06 1-07 1 .O9 1-10 1.1 I 1.11 1.13 1.13 1.13 1.13 1.12 1.1 I 1.10 I .O4 1 .O8 1 .O7 1 .O7 1-06 I -0s 1 .O5 I .O4 1.03 1 .O2 1-01

Table 0.3. INCO Mutte Separation Plant c h m n 2. Gus holdup data col/ected March 15 (1 998) using sensor 2.

w open (mSfcm) 0.89

0.89 0.88 0.88 0.88 0.87 0.87 0.88 0.88 0.87 0.87 0.87 0.88 0.87 0.87 0.87 0.87 0.86 0.86 0.85 0.86 0.86 0.85 0.84 0.84 0.82 0.82 0.8 1 0.82 0.82 0.8 1 0.8 1 0.80

0.8 1 0.8 1 0.n 1 0.82 0.83 0.83 0.85 0.88 0.9 1 0.93 0.94 0.95 0.96 0.95 0.95 0.93

w sypnon 1 (mSkm) 1.01 1 .O0 0.98 0.97 0.97 0.96 0.95 0.94 0.93 0.93 0.93 0.93 0.93 0.91 0.95 0.96 0.99 1 .O I 1-03 1 .O6 1 .O8 1-10 1.11 1.12 1.12 1.1 1 1.10 1-10 1.09 1 .O8 I .O8 1 .O7 1.06 1 .O6 1 .O5 1 .O5 1 .O3 1-04 I .O4 1 .O3 1 .O2 1-02 1 .O2 1 .O2 1 .O2 1 .O2

r open (mS/cm)

0.9 3 0.9 1 0.90 0.89 0.88 0.87 0.87 0.8 1 0.8 I o. 80 0.80 0.79 0.8 1 0.8 1 0.82 0.83 0.86 O. 89 0.9 1 0.93 0.95 0.96 0.98 0.97 0.97 0.97 0.97 0.96 0.95 0.96 0.96 0.96 0.95 0.95 0.94 0.94 0.93 0.93 0.93 0.92 0.9 1 0.9 1 0.9 1 0.91 .

0.9 1 0.9 1

m

101

Table D.4. INCO Matte Separorion Plant column 2. Gus holdup data collecred March

101

Table D.5. INCO Matte Separation Plant colzrmn 2. Gas holdrrp data collecred Jhrcli

K sypnon (mS1cm) 7 m r

0.928 6.93 0.93 0.93 0.93 0.93 0.93 0.93 0.93 0.93 0.94 0.95 0.95 0.96 0.96 0.97 0.97 0.07 0.w O 97 o 97 11 YX

0 9%

0 crX O crn t 1 ' J X

O &JX

( t ' f i #

(1 V'I

O '4')

(1 ' F b

(1 'F4

1 tr l

[ i i !

l i a i

1 i n i

1 l m 1

I l ' C d

I I ' F I

t l '4s (1 ' 4 X

O un ( 1 (J-

( 1 '4s 0 ' 4X

( 1 '1-

Il 'lx ( I vx -

k sypnon (mS/cm)

0-99

1 .O0 1 .O1 1 .O2 1 .O2

-1 .O3 1 . 0 1 1 .O6 1.1 1 1-15 1.18 1.17 1-15 1. Id 1.13 1.12 1.12 1.12 1.12 1.13 1.13 1.13 1.13 1.14 1-15 1.17 1.18 1.20 1.21 1.22 1 . 3 1 -24 1 -24 1.25 I . 3 1.25 1.25 1-24 1.24 I .Z3 1.12 i 22 1.21 1-21 1.21 1-20

Table D.6. INCO Matte Separafion Plant column 2. Gus holdup data collected March 18 (1 998) using sensor 2.

#

0:22:40 0:37:42 0:52:46 1:07:49 1 :22:5 1 1 :37:57 1 :53:O2 2:08:06 2:X :O9 X 8 : 14 2:53: 17 3:08: 18 3 2 3 2 1 3:38:23 3 5 3 2 6 J:08:27 4 2 3 2 9 4:38:3 1 1:53:36 5:08:38 5:23:J 1 5:38:46 5 5 3 3 1 6:OX:F-l 6:23:57 h:39:00 6:SJ:02 7:09:02 7:24:07 7:;v: IO ?;M: 14 n.00. 1 t i X:2J:2 1 X:30:2j X:Sl.I4 '#:09:27 Y:24.2V 9: 39: j 5 9:5J:40 l OWi: J2 I0:2J:J7 1 O: j9: J9 10:jJ:jO 1 1 :09:53 1 1 :24:55 l 1:40:0l 1 I :55:04 12: IO: (KJ

k ' ' P B (mSlcm)

1.0/ 1 .O7 1 .O4 1.06 1 .O6 1 .O4 1 .Cu 1-04 1 .O3 1 .O3 l .O3 1-02 1 .O3 1 .O3 1.02 1-01 1 .O1 1.01 1 .OC)

1 .O()

1 .(Ml

I .OZ I .O I 1 .O2 1.02 1 .O'

1 .O3 1 .O<

1 0' 1 ( 1 1

1 H l

I (il ( t <FI

O 'W

O 'U,

O 9 5 0 ')J

i t (1;

O 'II

( 1 '):

i t ' ) Z 0 OZ

0.9;

0 . 0 5

0.95 0.95 O.'H* o. oh (i.9..

E7YJjm (mS/cm) l m r

1.190 1.19 1.18 1.18 1.17 1.17 1.16 1.15 1.15 1.14 1-14 1.1 J

1.14 1.11 1-13 1.13 1.12 1.12 1.12 1.12 1.13 I . I J I.1J 1 I J I 1' 1. 15 1.15 1 I J I 1; 1.1:

1 II I I i J f 1 in. I if(

I IJ 1 1,:

I il'

1 Il' I il: 1 o; 1 o'i 1 OP

1 0 5 l tut

1 (n, 1 in1 1 tu?

104

Table 0 . 7 . INCO Matte Sepuration Plant column 2. Gas holdup data collected :tfurch

I Z:J5:ZJ l3:ûO:X 13: 1 S:32 13:30:33 13:45:34 1 J:00:37 14: I5:38 l4:3O:39 14:45:13 1 5:00:46 15:15:51 15:30:54 15:45:59 l6:ûû:J9 16: 1559 l6:3 1 :O2 l6:-t6:O7 17:01:13 17:16:15 17:31:17 17:46:7 1 18:C) 122 ln: 1623 183 125 18:46:30 l9:O 1 5 6 19: 16:39 19:; 1:40 1 9:46:45 2o:o 1 145 20: lh:J7 2o:j 1:49 31~4Ci:Sj 2 l :O 155 ZI:17:01 2 I :32:06 21:47:[ 1 22:02: 1 1 22: i7:12 2252: 15 22:47: 19 23:O2:2 1 23:17:27 23:32:29 23:47:33

n sypnon (mSicm) T

1 .O3 1.04 1-04 1 .O4 1 .O5 1 .O5 1 .O5 1 .O6 1 .O5 1-05 1 .O5 1 .O5 1.04 1.04 1 -04 1 .O3 1 .O3 1 .O2 1 .O2 1.01 1.01 1.01 1 .O1 1 .O 1 1-01 1 .O0 1.00 0.99 0.99 0.99 0.99 0.98 0.98 0.98 0.98 0.99 0.99 0.98 0.98 0.98 0.97 0.97 0.97 0.96 0.97

IO5

Table D.8. INCO Marre Separaiion Plani column 2. Gus holdup data collected hfarch

lo6

Table D.9. INCO Matte Separation Plant column 2. Gas holdup data collected hfarch 2 I (1 998) using sensor 2.

1 O7

Table D. 1 0. INCO Matte Separarion Plant column 2. Gus holdup .dam collected Marck

(mS/cm)

1.220 1.2 1 1.21 1-20 1.19 1.18 1.17 1.16 1.15 1.14 1.13 1.12 1.12 1.1 1 1.10 1.10 1.09 1.09 1-04 1 -08 1 .O8 I .on I .O9

1.12 1.14 1.17 1.19 1.12 1 .23 l 24 1.3 1 1' I Zr) 1 1% 1 1 - I 15 11; 1 12 I Ili I trr I tif 1 .im I .oc I .(cl I .CU 1-03 I .o; 1.03

Table D. I I . IXCO Matte Separarion Phnr column 2. Gas hoZdup data collected Mrch

Table D. I 2 INCO Matte Separation P la~t column 2. Gar holdup data collected March 24 (1 998) using sensor 2.

110

Table D. 13. INCO Marie Separafion Plant coiumn 2. Gus hofdup dala collecred March

1 1 1

Table D. 14. INCO Matte Separation PIant column 2. Gas holdup data collected March 26 (1 998) using sensor 2.

w sy pnon (mS/cm)

1.14

I . I1 1.15 1-17 1-19 1-21 1.23 1.24 l -24 1 .23 1.23 1-22 1.21 1-20 1.20 1.19 1.19 1.18 1.18 1.18 1.18 1.18 1.18 1.18 1-20 1.22 1-24 1-25 1.27 1.28 1-29 1.29 1.30 1.29 1-29 1-28 1.27 1.26 1-24 1-24 1-23 1.24 1.26 1.27 1 .26 1-25

112

Table D. 15. INCO Mutte Separation Plant column 2. Gas holdup data colfected lMarch 27 (1 998) using sensor 2.

113

Table D. 16. INCO Marte Separaiion Plant column 2. Gus holdup.daro colleetrd Match ,

w =Y Pnom (mSlcm)

1.057 I .O5 1 .O5 1.04 1 .O4 1-03 1 .O3 1 .O2 1.02 1 .O2 1 .O2 1.02 1.01 1 .O1 1.01 1.01 1 .O 1 1.01 1.02 1 .O2 1.02 1 .O 1 I .O 1 1-01

1 . ( X I

1 .in) 1 .(MI 0.w 0.w

O . ' N

0 9%

O trx

O C r

O 9'

( t 'r ( 1 'Hl

0 <HI ( 1 Vf t I 'P) I ( H l

1 O'

I 0 3 1 .O3 1 .O4 1 .OM I .or I .oc 1.05

u sypnon (mYcm)

1.u5 1.05 1 .O5 1-04 1 .O3 1 .O2 1 .O2 1.01 1.01 1.01 1.01 1.01 1.01 1 .O t 1 .O2 1 .O3 1 .O3

- 1-04 1 .Cu 1-05 1 .O5 1 .OS 1 .O5 1 .O8 1. 1 1 1.11 1.17 1-20 1.22 1.24 1 . 3 1 -24 1.14 1 .2j 1.22 1-21 1-20 1.18 1.17 1.16 1.15 1-14 1.13 1.13 1.13 1.13

114

Table D. 1 7. M O Mutte Separarion Plant column 2. Gus holdup data collecrrd Jfarch 29 11 998) using sensor 2.

l q x z (mSJcm) l T T

1.120 1.1 1 1.10 1.09 1 .O8 1 .O8 I .O7 1 .O6 1 .O5 1 .O5 1 .O5 I .04 I .M 1 .O3 1 .O2 I .O 1 l .(K) 0 . w O w O V*

o vn O v- O 'J-

O km 0 Wb

( 1 '45

O V C

il W r

O '4-

Il us ( I 'CL1

1 I I ,

1 JI

1 11 l 11:

1 1 1 :

l f l :

1 1 1 :

I ( 1 I 1 ( i l

O 'P l

I I Y'

( J 'Ur

(1 'Wr

Il '45 O '14

II '1C

4 1 '15

Table D. 18. INCO Maffe Separaiion Plant column 2. Gus holdup data collecied March

OP="

(mS1cm) 1-01 1 .O8 1 .O8 1.09 1 .O9 1.1 1 1.12 1.17 1.21 1.29 1.35 l,39 r .43 I .J7 I .J9 1 .SZ 1.54 I .F6 1.57 I 57 1 .vl 1 5 5 I 5 ; 1 1'4 1 4 5 l J J

1 l ( 1

1 :9 I :K 1 :- 1 ;z

1 :! 1 :;

l :: I ; 1

i ; I 1 :4 i

1 I l -

II6

Table D.19.a. INCO Matte Separalion Plant column 3. Gas holdup data colfectrd

Table D. 19.6. Continuation fiom Table D. 19.a.

=Fi== (mS1cm) 1.23

1-22 1-22 1.21 1-21 I.20 1-20 1.20 1.19 1.18 1.18 1.17 1.17 1.16 1.16 1-15 1-15 1-15 1.15 1.15 1.16 1.15 1.16 1-16 1-10 1.16 I . l h 1.16 1.10 I . l a 1.16 1.16 1.16 I Ih

1 Ifr

1 Itt I I t r

1 It* 1 l t l

I 1t.

I t r * 1 lh

1 . I t l 1 Ir, 1.16 1.16 1.16 1.16 1.Ib

Table D. 19. c. Continuation fiom Table D. 19.6.

Table D. 20. Measured grades for INCO test.

Ti- m 10.56 11.10 13.82 13-97 13-93 12.2 1 10.88 10.15 -

Tuhfe D. .? I . Grude d u e s rrsv J to estikiu~re rhe weighrsfor the mass balunce u~trsrrnmrs

I I 1 1 I I I I I Average I 1 8.06 0.14 1 0.42 174.17 1 3.511 1 0.06 I 0.30 1 b4.33 1 I I

7262 1 0.22 1 0 . c I 1 I I I I 1 1 I 1 1 I

Table 0.23. Recovery and @pical deviation d u e s for NCO test.

Tuhfr D.24. Data on Airfloii. rate. Ia*el ( f i d depth) and solidpercent for /.!*CO lesr.

Table D.2j.a. Dafafiom two sensors ut di#ierent deplhs inside column 3. Sensor I al 4- rneïer deprh and sensor 2 at 7-meter depth.

Time

8:07:28 8: I N 9 8: 1729 8 2 2 3 0 8:27:30 8:32:30 8:37:30 8:42:30 8:47:30 8:52:31 8:57:3 1 9:O2:3 1 9:07:32 9: l2:32 9: 17:32 9:22:32 9:27:33 9 3 2 3 3 9:37:33 9:41:33 9:47:33 9 5 1 3 3 9:57:34 10:02:34 1 O : O i : j j

10: 1 3 5 I O : 1 7:3C 10:22:3t> 10.27;jh i ri:32:36 10.3?33 I o J : 3 7 lo:J7:37 10:52:37 10:57: 37 1 1 :O2127 1 1 :0?:37 I I:I2:3X 1 1: 17:3X

,

k syphon (mS1cm)

1 S U 1.50 1-50 1.50 1-50 1.50 1.50 1.50 1.50 1.50 1 .JO l .49 1-49 1.49 I .4U 1.49 1 .JX I Jn 1 4% 14% 1 J' 1 J' I J' 1 4- I Ji-

l Jti

1 4h 1 10

I 4fr

1 l f 1

I JI.

f 1'

I If.

l 1 8 %

- 1 0.

l W .

l I r .

1 U.

l a*.

I J -

Table 0.25. b. Continuation fiom Table D. 25. a.

k syphon (mSlcm)

1.48 1-49 1.50 1.51 I .5Z 1.53 1.54 1.55 1-56 1.57 1-58 1.59 1.59 1.60 1.61 1.6 1 1.62 1.62 1.63 1-64 1.63 1.63 1.62 1.61 1.61 1.63 I .hZ 1 .tl l I .HI 1 .CK 1.57 1 . i t l 1.35 1 3-1 1 ..il 1 . z ; I 3.; 1 2' 1 .SZ 1.51

Table E. l. Erperimental dam on gas holdup: gas holdup f iom pressure versus gas. hoId~ipJ;orn conduciivity. Every point is the average of 25 min of operation. (Zast 4 points correspond to jetting sparger tests). Temperature 323 K.

Table E. 2. Test conditions for BO WA TER tests

1 s t R

1 A- l 1 A-? I A-3 I A 3 1 A-5 10-1 1 8-2 1 B-3 1 B 3 1 B-5 1 C- l I C-2 I C-3 1 C-l 1 C-5

2.4-5sp 2 A-Ssp 2.4-3sp ZB-5sp ZB-lsp îB-jsp 5.4-Ssp 3.4-4sp

-;.A-4sp-b -; B-Ssp ;B-4sp

;B-lsp-b 4. l - I 4.A-2 4 .A - 3 4.A-4

(min-) 3.z 3.2 3 -2 3.1 3 -2 6.0 6.1 6.3 6.2 6.1 5.8 6.2 6.2 6.1 6.2 3.3 3.4 3 -5 6.0 6.2 6.6 6.9 7.2 6.9 7.3 7.4 7.2 6.0 6.1 6.1 6.7

Table E. 3. Flotat ion eficiency and brighrness gain for BO WA TER tests.

. 1 est e

1A-1 1 A-2 1 A-3 I A 4 I A-5 1 5 1 I B-2 I B-3 I B-l I f3-5 fC-l I C-2 I C-3 I C-l 1 C-5

ZA-SSP ZA-lsp ?A-3sp ZB-5sp ZB-lsp 2 B-3 sp 3 t l - 5 ~ ~ ;A-lsp

;A-4sp-b 3B-jsp 3 B-Jsp

3R-lsp-b

- - m teed

1075 975 934 915 12% 1225 1308 1226 1 146 1230 1187 1207 1319 1 290 101 1 98 1 1023 1255 1167 I 142 120 I 1095 1233 1100 94 5 IO43

n ) accepts - 3443

3 89 441 739 927 308 31 l 338 476 1 O7 1 32 1 348 3 72 460 1048 446 F 16 869 3 16 4;o 94 5 275 332 439 298 355 352

t tticrency ( 9 6 ) bl.7

63.8 54.7 20.9 0.0 74.8 74.6 74.2 6 1.2 6.5 73 -9 70.7 69.1 65.1 18.8 55.9 47.1 15.1 74.8 63.2 17.3 77.1 w 7

t 4 4 1.

-9 O h2.5 M. 2

t eed 45.95

36.06 47.43 48.02 47.96 42.95 42.7 1 42-48 43.60 43.45 43.50 43.35 42.94 43 -77 42.95 44.20 46.03 46.53 43-16 44.35 45.45 43.4 I 45.4 1 44.87 45.73 47.39 46.23

accepts 3 1-06 56.07 55.85 50.92 47.62 -56.2 1 56.5 1 55.9 1 52.97 46.25 56.03 55.90 55.35 54 -27 45.56 52.67 52.23 48.29 57.40 55.13 48.19 58.42 57.36 55.58 58.31 57.00 56.66

C a m - 11.1 I 10.0 1 8.42 2.90 0.00 13.26 13.80 13.43 9.37 2.80 12-53 12.55 12.4 1 10.50 2.6 1 8.47 6.20 1.76

IJ.24 10.79 2.74 15.0 1 1 1.95 1 O.? 1 12-58 9.6 1

10.4;

Table E. 4 Test condition for test dB-jsp Hf = 100 cm: 5 spargers: r = 6 min..

Table E.5 Test condition for test ZB-4sp Hf = 100 cm: 4 spargers: r = 6 rnin..

Table E.6. Test condition for test 223-3sp Hf = 100 cm: 3 spargers; r = 6 min..

tlP (*A)

S.P 9.2 9.9 9.7 8.4 9.2 9.5 8.7 9.4 9.9 9.4 8.7 9 2 9.6 8.3 8.8 9.7 9.5 9.0 9.4 8 -4 8.3 8.9 9.1 8.5 9.1 9.;

8.X n.7 n-3 'J. I X.7 Y . -: 9. I 8.7 n . 9 - - I I

x 'J

Y' 0 I

O

n 5 X X x 5 n 2

Table E. 7. Test condition for test 2A-jsp H/ = 100 cm: 5 spargers: r = 3 min..

Table E.8. Test condition for tesr ?A-4sp 4 = 100 cm: 4 spargers: -r = 3 min..

Table E. 9. Test condition for test ZA-3sp Hf = 100 cm: 3 spurgers: r = 3 min..

Ym (cmHZ01

Laa.Y 288.7 289.0 288.6 288.9 289.7 289.0 290.0 289.5 289.8 288.3 289.8 289.4 289.0 289.5 290.3 290.0 289.2 289.5 289.8 290.5 291.1 289.8 290.6 289.7 290.3 189.7 190.R 289.5 289.5 29 l .O

289.5 290.2 2nv.x ?XX h

2x9 4 2VO.h 2n9 - '90 x 'X0.5 29 1 .ll .xu K 291 O

190.' 'XV ' -

Table E. IO. Bubble size. bubble surface aareaflux and collection zone rate constant . esfirnafed for BOWATER tesrs. Bubble sizes over 0.2 cm are invalid

lcsl s

I A-T I A-2 I A-3 I A 4 I A-5 I B-I I B-2 1 5 3 1 B4 1 B-5 1C-1 1 C-2 IC-; ICA I C-5

2 A-Ssp A 4 s p Z A-3sp 2 6-5 sp 2BAsp 25-isp .;A-Ssp .:A-lsp

;A-!spb :R-?sp 7n4sp

:B4\pb J A - I 4A-2 JA-; 4 A-!

JI3 (JU h ( a n i s )

, i -.- 1.77 1 36 0.86 O66 2.79 '3.4 1.86 1 37 090 2 70 2.29 181 1 32 O 89 2 JI 2 35 2.32 2.34 2 22 2.31 1 76 I f 5 . 9.7

l -3' l -73

1 68 2 32 7 94 1 40 1 08