Image Reconstruction: Motion Correction
David Atkinson [email protected]
Centre for Medical Imaging,Division of Medicine,
University College London
Problem: Slow Phase Encoding
• Acquisition slower than physiological motion.– motion artefacts.
• Phase encode FOV just large enough to prevent wrap around.– minimises acquisition time,– Nyquist: k-space varies rapidly making interpolation
difficult.
Any Motion Can Corrupt Entire Image
k-space acquired in time
images(x)
Fourier Transform
ikx
k
ekSxs )()(
The sum over all k in the Fourier Transform means that motion at any time can affect every pixel.
K-Space Corrections for Affine Motion
Image Motion
Translation (rigid shift)
Rotation
Expansion
General affine
K-Space Effect
Phase ramp
Rotation (same angle)
Contraction
Affine transform
dAx)(
'xA
VΑV kAk' T1)(
)(')det(
)(2)( k'
A
dk'k S
iS e
Rotation Example
Time
Example rotation mid-way through scan.
Ghosting in phase encode direction.
Interpolation, Gridding and Missing Data
FFT requires regularly spaced samples.Rapid variations of k-space make interpolation difficult.
K-space missing in some regions.
Prospective Motion CorrectionMotion determined during scan & plane updated using
gradients.• Prevents pie-slice missing data.• Removes need for interpolation.• Prevents through-slice loss of data.• Can instigate re-acquisition.• Reduces reliance on post-processing.
• Introduces relative motion of coil sensitivities, distortions & field maps.
• Difficult to accurately measure tissue motion in 3D.• Gradient update can only compensate for affine motion.
Non-Rigid Motion
• Most physiological motion is non-rigid.• No direct correction in k-space or using gradients.• A flexible approach is to solve a matrix equation
based on the forward model of the acquisition and motion.
Forward Model and Matrix Solution
mEρ“Encoding” matrix with motion, coil sensitivities etc
Measured data
Artefact-free Image
2min mEρρ
Least squares solution:Conjugate gradient techniques such as LSQR.
The Forward Model as Image Operations
motion-freepatient
motioncoilsensitivity
sample shot
=Measuredk-space forshot
FFTk i
Image transformation at current shot
Multiplication of image by coil sensitivity map
Fast Fourier Transform to k-space
Selection of acquired k-space for current shot
Shots
spin echo1 readout = 1 shot
single-shot EPI multi-shot
Forward Model as Matrix-Vector Operations
motion-freepatient
motioncoilsensitivity
sample shot=
Measuredk-space forshot
FFTk i
ρEm
*
Converting Image Operations to Matrices
• The trial motion-free image is converted to a column vector.
nn
n2
ρmotion-freepatient image
Expressing Motion Transform as a Matrix
?
= motioncoilFFTsampleMeasuredimage k i
=
• Matrix acts on pixels, not coordinates.• One pixel rigid shift – shifted diagonal.• Half pixel rigid shift – diagonal band, width depends
on interpolation kernel.• Shuffling (non-rigid) motion - permutation matrix.
Converting Image Operations to Matrices• Pixel-wise image multiplication of coil sensitivities
becomes a diagonal matrix.• FFT can be performed by matrix multiplication.• Sampling is just selection from k-space vector.
patient
= motioncoilFFTsampleMeasuredimage k i
=
Stack Data From All Shots, Averages and Coils
ρEm *
Conjugate Gradient Solution
• Efficient: does not require E to be computed or stored.
• User must supply functions to return result of matrix-vector products
• We know the correspondence between matrix-vector multiplications and image operations, hence we can code the functions.
wEEv H and
2min mEρρ
The Complex Transpose EH
• Reverse the order of matrix operations and take Hermitian transpose.
• Sampling matrix is real and diagonal hence unchanged by complex transpose.
• FFT changes to iFFT.• Coil sensitivity matrix is diagonal, hence take complex
conjugate of elements.• Motion matrix ...
motion FFT
H H H H
coil sample
Complex transpose of motion matrix
Options:• Approximate by the inverse motion transform.• Approximate the inverse transform by negating
displacements.• Compute exactly by assembling the sparse matrix
(if not too large and sparse).• Perform explicitly using for-loops and
accumulating the results in an array.
Example Applications of Solving Matrix Eqnaveraged cine
‘sensors’ from central k-space lines input to coupled solver for motion model and artefact-free image.
multi-shot DWI example phase correction artefact free image
Summary: Forward Model Method
• Efficient Conjugate Gradient solution.• Incorporates physics of acquisition including parallel
imaging.• Copes with missing data or shot rejection.• Interpolates in the (more benign) image domain.• Can include other artefact causes e.g. phase errors in
multi-shot DWI, flow artefacts, coil motion, contrast uptake.• Can be combined with prospective acquisition.• Often regularised by terminating iterations.• Requires knowledge of motion.
Estimating Motion
• External measures.• Explicit navigator measures.• Self-navigated sequences.
• Coil consistency.• Iterative methods.• Motion models.
Estimating Motion
• External measures.• Explicit navigator measures.• Self-navigated sequences.
• Coil consistency.• Iterative methods.• Motion models.
ECG,respiratory bellows,optical tracking, ultrasound probesspirometersaccelerometersPower depositionField Probes
Estimating Motion
• External measures.• Explicit navigator measures.• Self-navigated sequences.
• Coil consistency.• Iterative methods.• Motion models.
pencil beam navigator,central k-space lines,orbital navigators, rapid, low resolution images,FID navigators.
Estimating Motion
• External measures.• Explicit navigator measures.• Self-navigated sequences.
• Coil consistency.• Iterative methods.• Motion models.
repeated acq near k-space centre,PROPELLER, radial & spiral acquisitions,spiral projection imaging,
Estimating Motion
• External measures.• Explicit navigator measures.• Self-navigated sequences.
• Coil consistency.• Iterative methods.• Motion models.
Predict and compare k-space lines.
Detect and minimise artefact source to make multiple coil images consistent.
Estimating Motion
• External measures.• Explicit navigator measures.• Self-navigated sequences.
• Coil consistency.• Iterative methods.• Motion models.
Find model parameters to minimise cost function e.g. image entropy, coil consistency.
Estimating Motion
• External measures.• Explicit navigator measures.• Self-navigated sequences.
• Coil consistency.• Iterative methods.• Motion models.
Link a model to scan-time signal.
Solve for motion model and image in a coupled system (GRICS).
Golden Angle Sampling For Motion Detection and High Resolution Imaging• Determine motion from registration of images
(each from wide temporal range).• Use motion in CG high resolution reconstruction.
[Hansen et al #749]
Outlook
• Prospective corrections limited to affine motion.
• Reconstruction times, 3D and memory still challenging.
• Expect intelligent use of prior knowledge: sparsity, motion models, atlases etc.
• Optimum solution target dependent. Power in combined acquisition and reconstruction methods.