Upload
outi
View
214
Download
0
Embed Size (px)
Citation preview
RESEARCH ARTICLE
Data quality in fMRI and simultaneous EEG–fMRI
Toni Ihalainen • Linda Kuusela • Sampsa Turunen •
Sami Heikkinen • Sauli Savolainen • Outi Sipila
Received: 30 April 2013 / Revised: 24 March 2014 / Accepted: 27 March 2014
� ESMRMB 2014
Abstract
Object To evaluate functional magnetic resonance
imaging (fMRI) and simultaneous electroencephalography
(EEG)–fMRI data quality in an organization using several
magnetic resonance imaging (MRI) systems.
Materials and methods Functional magnetic resonance
imaging measurements were carried out twice with a uni-
form gel phantom on five different MRI systems with field
strengths of 1.5 and 3.0 T. Several image quality parame-
ters were measured with automatic analysis software. For
simultaneous EEG–fMRI, data quality was evaluated on
3.0 T systems, and the phantom results were compared to
data on human volunteers.
Results The fMRI quality parameters measured with
different MRI systems were on an acceptable level. The
presence of the EEG equipment caused superficial artifacts
on the phantom image. The typical artifact depth was
15 mm, and no artifacts were observed in the brain area in
the images of volunteers. Average signal-to-noise ratio
(SNR) reduction in the phantom measurements was 15 %,
a reduction of SNR similar to that observed in the human
data. We also detected minor changes in the noise of the
EEG signal during the phantom measurement.
Conclusion The phantom proved valuable in the suc-
cessful evaluation of the data quality of fMRI and EEG–
fMRI. The results fell within acceptable limits. This study
demonstrated a repeatable method to measure and follow
up on the data quality of simultaneous EEG–fMRI.
Keywords Magnetic resonance imaging � Phantoms �Electroencephalography
Introduction
Functional magnetic resonance imaging (fMRI) measures
hemodynamic response in the brain during a predetermined
stimulus or task, or a resting state [1, 2]. Differences of
only a few percent in the blood oxygen level-dependent
(BOLD) signal are measured. From a technical standpoint,
an fMRI study consists of an echo-planar imaging (EPI)
sequence run over an extended period of time. EPI
sequences usually require the use of rapidly switching
gradients of maximum amplitude. Some system and
physiological noise are always present [3]. Even if some of
the noise can be reduced by data preprocessing, the quality
of the original data is essential. When the signal-to-noise
ratio (SNR) is low or the MRI signal stability is compro-
mised, the BOLD signal may be indiscernible or an
unexpected non-BOLD-related signal may correlate falsely
with the paradigm used [4]. Statistical averaging is limited
in the clinical setting, and sufficient SNR is crucial. EPI
sequences are also prone to several types of artifacts. When
fMRI is clinically used for presurgical evaluation, spatial
accuracy is essential [5, 6].
The demand for high stability together with maximum
instrumentation loading for a long period of time makes
fMRI one of the most demanding applications of MRI
T. Ihalainen (&) � L. Kuusela � S. Turunen � S. Savolainen �O. Sipila
HUS Medical Imaging Center, P.O. Box 340,
00029 HUS, Helsinki, Finland
e-mail: [email protected]
T. Ihalainen � L. Kuusela � S. Turunen � S. Savolainen
Department of Physics, University of Helsinki, P.O. Box 64,
00014 Helsinki, Finland
S. Heikkinen
Department of Chemistry, University of Helsinki, P.O. Box 55,
00014 Helsinki, Finland
123
Magn Reson Mater Phy
DOI 10.1007/s10334-014-0443-6
technology. Even more challenging is the simultaneous
acquisition of electroencephalography (EEG) and fMRI [7,
8], currently the subject of active research. Simultaneous
EEG–fMRI may help with localizing the foci of epileptic
seizures [9, 10]. Small signals are being measured with the
two techniques, which tend to interfere with each other’s
performance. The EEG equipment may introduce promi-
nent susceptibility-based image artifacts, as well as
reduction of image SNR [11, 12]. Krakow et al. [13] and
Mullinger et al. [14] have conducted phantom studies of
the effects of various EEG equipment on fMRI image
quality in simultaneous EEG–fMRI. They have reported
artifacts caused by EEG electrodes, as well as lower SNR
caused by increased coil loading. Most of the literature
discussing data quality issues of simultaneous EEG–fMRI
has focused on the quality of and artifacts in the EEG
signal (e.g., [15–17]).
Quality assurance tools are needed to ensure that MRI
systems meet the high performance demands of fMRI at all
times. MRI image quality can be assessed with test objects,
or phantoms, filled with a liquid or gel that is visible in
MRI. Several internationally acknowledged phantom sets
and protocols, as well as manufacturer-specific phantoms
are available for regular quality assurance [e.g., 18–20].
Friedman and Glover [21] have published a protocol for
fMRI quality assurance, focusing on signal stability
parameters and their automated analysis. This protocol has
served to ensure uniform measurements in the research
consortium Biomedical Informatics Research Network
(BIRN). Other studies of fMRI quality assurance protocols
that emphasize the role of automated data analysis have
also been published [22, 23].
In our organization, both clinical and research fMRI
studies were ongoing with five MRI systems of three dif-
ferent manufacturers; EEG–fMRI studies were ongoing
with two systems. The aims of this study were (1) to
evaluate fMRI data quality with a suitable phantom and
software, (2) to evaluate the data quality of EEG–fMRI as
much as possible using the same tools, and (3) to compare
the data quality measurements with EEG–fMRI data on
human volunteers.
Materials and methods
Phantom preparation
The desired properties of the phantom and analysis soft-
ware included (1) their suitability for several MRI systems
of different manufacturers and field strengths, (2) easy
measurement and data analysis, (3) an internationally val-
idated method, and (4) the possibility to include simulta-
neous EEG–fMRI. Because the method published by
Friedman and Glover seemed to fulfil these criteria, we
selected it for this study and constructed a uniform gel
phantom according to their instructions [21]. The gel was
prepared in a glass ball with an inside diameter of 18 cm.
We acquired high-resolution T1-weighted images of the
phantom to verify the absence of significant air bubbles.
The fMRI image quality measurements
We measured fMRI image quality using two 3.0 T systems
(from two different manufacturers) and three 1.5 T systems
(from three different manufacturers). The MRI systems
(A–E) used for the measurements as well as the number of
head coil elements in each system appear in Table 1. The
measurements were carried out with an axial gradient echo-
based single-shot EPI sequence: TR = 2,000 ms,
TE = 30/40 ms (3.0/1.5 T), field of view = 22 cm, num-
ber of slices = 30, slice thickness = 4 mm, slice
gap = 1 mm, flip angle = 77�, matrix size = 64 9 64,
receiver bandwidth = 2,298 Hz/pixel, and number of
acquired time frames = 200 [21]. If the number of slices
and the receiver bandwidth could not be set precisely on
some 1.5 T systems, we set the values as close to the ref-
erence values as possible. On MRI system B, we used
parallel imaging (Sensitivity Encoding, SENSE) with a
factor of 1.8.
We carried out two rounds of measurements with the
phantom approximately one year apart (later referred to as
measurements 1 and 2). The measurements using MRI
systems A and B were carried out with the image intensity
correction (IC) on and off, since use of the multi-element
head receiver coils permits normalization of the signal
intensity throughout the imaging volume with the help of
the sensitivity information of the coil elements. In addition,
control measurements were carried out using MRI system
B without parallel imaging to evaluate its effects on the
data.
For the phantom image analysis, we used the software
released by BIRN (BXH/XCEDE tools version 1.9.8.6,
downloaded at http://www.nitrc.org/projects/bxh_xcede_
tools/). The software automatically measured signal fluc-
tuation and drift, the SNR and the signal-to-fluctuation
Table 1 The MRI systems, their field strengths, and the number of
elements in the receiver head coils used in this study
MRI
system
Manufacturer and
model name
Field
strength (T)
Number of head coil
elements
A Siemens Verio 3.0 12
B Philips Achieva 3.0 8
C Siemens Avanto 1.5 12
D Philips Achieva 1.5 8
E GE Signa Hdxt 1.5 8
Magn Reson Mater Phy
123
noise ratio, the radius of decorrelation (the Weisskoff test),
the center of mass maximum displacement and drift in
three directions, as well as smoothness and ghosting. The
analysis software calculated two distinct values for drift;
we used the one that measured the value based on the
second-order polynomial fit of the signal curve. The radius
of decorrelation defined the size of the region of interest
(ROI) at which the voxels lost their statistical indepen-
dence [21]. The smoothness values reflected the smooth-
ness of the image noise expressed as the full width at half
maximum in three directions. In addition to measured
values, the software produced several parametric images of
the measurement series.
The EEG–fMRI data quality measurements
The measurements were carried out in the presence of the
EEG equipment using the 3.0 T MRI systems (A and B), as
these MRI systems were also being used in simultaneous
EEG–fMRI studies. A 64-channel EEG cap (BrainProducts,
Inc., Munich, Germany) was positioned on the phantom
surface as it would have been positioned on a human head
(Fig. 1). The electrodes were electrically connected with
ECI electrogel (Electro-Cap International, Inc., Eaton, OH,
USA) or EC2 Genuine Grass Electrode Cream (Grass
Technologies, Inc., Warwick, RI, USA) to prevent damage to
the EEG equipment. The electrocardiography (ECG) and
electrooculography (EOG) electrodes included in the cap
were also attached to the phantom. The EEG cap was con-
nected to two MRPlus amplifiers (BrainProducts, Inc.,
Munich, Germany) that were also in the magnet bore. The
EEG signal was also recorded with the same measurement
set-up on the 3.0 T systems. Recording began before the EPI
sequence was turned on and continued throughout the
acquisition. A separate recording was performed outside the
MRI room. We measured the effects of the helium pump as
well as in-bore lighting and ventilation on the EEG signal.
Because all the EEG channels were electrically connected
with gel, the analysis was limited to the evaluation of noise
and the detection of faulty channels. Two rounds of EEG–
fMRI measurements were carried out in connection with
fMRI measurements 1 and 2 described earlier, and the image
quality was also analyzed in a similar way. The EEG signal
was analyzed separately with BrainVision Analyzer 2.0
(BrainProducts, Inc., Munich, Germany). When necessary,
the recorded signal was corrected for artifacts related to
gradient switching using a well-known average artifact
subtraction method [24].
Based on the experience of the first measurements, we
conducted two further measurement sessions. Within each
session, we acquired four series with the EEG cap, and four
series without the EEG cap. Between each acquisition, the
phantom was removed from the magnet bore to introduce
some variation to the MRI system self-adjustments. The
position of the EEG cap was marked on the phantom surface
to make the positioning more repeatable. The ECG and EOG
electrodes suspected of having caused strong artifacts in the
previous measurements were now attached to the phantom
outside the field of view, as they would be in a study of a
human subject. The maximum penetration depth of the sur-
face artifacts caused by the electrodes was measured man-
ually slice by slice with a distance tool of ImageJ version
1.44p (Rasband, W., US National Institutes of Health, USA).
The SNR calculation [21] included the generation of a
signal image, which was a voxel-by-voxel average of the
images from the time series, and a difference image, which
was computed by summing all odd images and all even
images and then calculating a voxel-by-voxel difference of
these sum images. The SNR was then calculated as
SNR ¼ meanffiffiffiffiffiffiffiffi
r=np ;
where n was the number of time points, mean was the mean
voxel value of a signal image inside a selected ROI, and rwas the variance of voxel values from a difference image
inside the same ROI. To test whether the size of the ROI
affected the SNR measurement, we repeated the analysis
with ten different ROI sizes between 9 9 9 and 27 9 27
voxels in the four acquisitions of measurement 1 on MRI
system A. We calculated Pearson’s correlation coefficients
(r) for ROI size and SNR. This information served to assess
the validity of the SNR measurement in the presence of the
EEG equipment.
Comparison of the EEG–fMRI measurements to human
data
We retrospectively examined human fMRI data from three
imaging sessions (two volunteers, one participated in two
Fig. 1 fMRI phantom with the EEG cap
Magn Reson Mater Phy
123
sessions) conducted using MRI system B to investigate
whether the image quality findings from the phantom
measurements were comparable to the effects in studies of
human subjects. The ethics committee of the Hospital
District of Helsinki and Uusimaa approved the volunteer
studies, and the volunteers provided their written informed
consent. Each of these imaging sessions included four
identical gradient echo-based EPI sequences (TR =
3,000 ms, TE = 35 ms, number of signal acquisitions = 1,
voxel size = 2.8 9 2.8 9 4.0 mm, EPI factor = 79, and
number of time frames = 115) both with and without the
EEG equipment. No stimuli were delivered during these
acquisitions.
We used the ImageJ software to estimate the SNR of the
images of the volunteers. We first selected two slices, one
at the level of the centrum semiovale and the other, 24 mm
above it. We calculated the signal and difference images as
well as the SNR values in a manner similar to the one
previously described for the phantom case. Medial and
lateral square ROIs (30 9 30 mm) were placed on each
slice (Fig. 2). We used the same ROI coordinates for each
respective slice. We then calculated the SNR values for
each ROI. Because each voluntary session consisted of four
identical acquisitions both with and without the EEG
equipment, we calculated the average SNR values for these
four acquisitions and, subsequently, calculated for each
ROI the SNR difference caused by the EEG equipment.
Results
fMRI image quality
The results of the signal fluctuation (i.e., temporal noise)
and drift appear in Figs. 3 and 4, respectively. Apart from
four measurements, the percentage of the fluctuation value
was\0.2 % in all the measurements. In all cases, the signal
drift during the measurement of 200 time frames was
\1 %. The minimum and maximum SNR were 245 and
345 on 3.0 T systems, and 82 and 174 on 1.5 T systems,
respectively. The SNR results appear in Table 2. Pairwise
comparison of the SNR values of the respective measure-
ments with the signal intensity correction on and off yiel-
ded an average SNR reduction, with correction, of 8 % on
MRI system A and 0 % on MRI system B. The mean ghost
signal percentage was B2 % with MRI systems B and D,
and between 2 % and 5 % with other systems.
The radius of decorrelation, which describes the statis-
tical independence of the voxels during the fMRI time
series, varied substantially, with values ranging from 1.7 to
12.4 voxels; in general, the 1.5 T systems achieved higher
values than did the 3.0 T systems. The mean smoothness
measurements demonstrated up to two-fold greater differ-
ences despite the same nominal acquisition voxel size,
which was 3.4 9 3.4 9 4.0 mm in all measurements.
Fig. 2 Placement of the regions of interest in SNR determination of
the data on human volunteers. The 30 9 30 mm regions were placed
on two slices: at the level of centrum semiovale (left) and on the slice
24 mm above it
Fig. 3 The results of the signal
fluctuation (%) for two rounds
of measurements taken during
the fMRI time series on
different MRI systems. This
figure also presents the results
with EEG cap or intensity
correction (IC) on MRI systems
A and B, as well as the results
both with and without
Sensitivity Encoding (SENSE)
on MRI system B
Magn Reson Mater Phy
123
Maximum center of mass displacement values varied
from 0 to 0.4 mm (mean = 0.09 mm). On MRI systems B
and D, several abrupt but small (*0.2 mm) changes in the
phantom image center-of-mass were detected in the y
direction. We were also able to detect these changes in the
images when using careful window adjustment. Calculated
for the same MRI systems, the mean of all maximum
displacement measurements in the y direction was
0.20 mm, while it was 0.09 mm when calculated for the
other systems. The maximum absolute value for center-of-
mass drift was 0.36 mm (mean = 0.06 mm).
Except for a small temporal stability reduction in MRI
system D, we found no differences between measurements
1 and 2. Because the measurement of other MRI systems
produced equal results, the difference probably originated
from a small change in the stability of that system. When
control measurements were carried out without SENSE on
MRI system B, we observed no other effects than the 10 %
increase in SNR.
EEG–fMRI data quality
The image quality measurements carried out with the EEG
equipment resulted in superficial artifacts in the mean
image (Fig. 5). During the first two rounds of measure-
ments, the total signal dropout extended to 25 mm from the
Fig. 4 The results of the signal
drift (%) across the fMRI time
series for two rounds of
measurements on different MRI
systems. This figure also
presents the results with the
EEG cap or intensity correction
(IC) on MRI systems A and B,
as well as the results both with
and without Sensitivity
Encoding (SENSE) on MRI
system B
Table 2 Signal-to-noise ratios as measured in the two rounds of measurements
MRI system A MRI system B, with SENSE MRI system B, without SENSE
fMRI, 3.0 T
Measurement 1 263 294
Measurement 1, with IC 245 301
Measurement 2 270 288 345
Measurement 2, with IC 253 305 321
MRI system C MRI system D MRI system E
fMRI, 1.5 T
Measurement 1 82 98 150
Measurement 2 77 102 174
MRI system A MRI system B, with SENSE MRI system B, without SENSE
EEG–fMRI
Measurement 1 234 221
Measurement 1, with IC 208 189
Measurement 2 182 245 261
Measurement 2, with IC 170 272 284
IC intensity correction, SENSE sensitivity encoding
Magn Reson Mater Phy
123
phantom surface, and the effect on the signal level was
clearly visible as far as 50 mm from the surface. In the
additional measurements with improved electrode posi-
tioning, we measured the artifact depth slice by slice; the
key results appear in Table 3. The typical artifact depths
were 15 mm or less; the artifacts were deepest (as deep as
23 mm) in the two to three uppermost slices. The artifacts
showed no substantial differences in nature or strength
between the two 3.0 T systems. When the EEG equipment
was present in the first measurements, the SNR was
approximately 25 % lower (see Table 2). In the additional
measurements, the average SNR reduction was 15 %, with
an 11 % reduction on MRI system A and a 19 % reduction
on MRI system B (Table 3). The ROI size showed
no statistically significant correlation with SNR
(-0.52 B r B 0.31 in the four datasets tested). When the
EEG equipment was present, we found no effects on
temporal stability or other quality parameters.
The background noise in the EEG signal was approxi-
mately 2 lV peak-to-peak outside the MRI room and
10 lV inside the magnet. On MRI system B, we detected a
2-Hz interference signal with a maximal peak-to-peak of
50 lV that disappeared completely when the helium pump
was switched off; the general background noise level also
dropped to 5–8 lV peak-to-peak. In-bore lighting or ven-
tilation had no effect on the signal. The noise spectra of the
two amplifiers differed in some measurements even though
both amplifiers shared identical technical properties. The
amplifiers were placed on top of each other inside the
magnet, but the difference was independent of the order of
the amplifiers, and the interference did not depend on the
spatial distribution of the measurement electrodes on the
EEG cap. The interference did, however, correlate cycli-
cally with the channel number.
Comparison of the EEG–fMRI measurements
to the human data
Visual evaluation of three volunteer fMRI examinations
both with and without the EEG equipment revealed no
artifacts extending to the brain area. However, we did
observe a signal reduction in all three examinations with
the EEG cap, especially in the posterior brain. The average
SNR reduction in the presence of the EEG equipment was
24 % (18 % for medial and 30 % for lateral ROIs).
Discussion
In the first part of this study, we evaluated the image
quality of the fMRI protocol of several MRI systems. In
general, the results agreed well with previously reported
values and observations. Because the key parameters (i.e.,
signal drift and fluctuation values during an fMRI time
series) were in accordance with typical values reported by
Friedman and Glover [21], we could conclude that the
systems measured were operating at an acceptable stability
level. The differences in mean smoothness values for dif-
ferent MRI systems were notable despite the same nominal
acquisition and reconstruction voxel size. Taking into
account the differences in SNR and radius of decorrelation,
the probable explanation is the different filtering or other
manufacturer-dependent preprocessing of the fMRI data.
However, knowledge of the effects of data processing on
the resultant image quality for each MRI system would be
important in designing an fMRI study and interpreting the
data. The use of multi-element coils enables the use of
techniques such as parallel imaging or image intensity
correction. This may reduce SNR, which in some cases in
human studies may be critical. The use of parallel imaging
may, however, prove useful when optimizing the level of
SNR, artifacts, and speed [25].
The manufacturer did not confirm the reason for the
small changes in the phantom image center-of-mass on
MRI systems B and D. These changes likely have no effect
on the interpretation of data from an fMRI study of a
human subject, as the changes were small in comparison to
physiological movement or typical voxel size. The result,
Table 3 Signal-to-noise ratios and artifacts in EEG–fMRI
MRI
system
Session Series Relative
SNR (%)
Max artifact
depth in mm
(slice
number)
Average
artifact depth
(mm), slices
19–28
A 1 1 90 23 (29) 12
2 86 20 (29) 12
3 90 20 (29) 13
4 81 23 (30) 14
2 1 94 13 (24–30) 12
2 92 13 (25–30) 12
3 90 15 (26, 30) 12
4 89 14 (27–30) 12
B 1 1 83 18 (26) 14
2 75 19 (27) 15
3 86 19 (27) 15
4 90 18 (27, 28) 14
2 1 71 18 (27, 29) 13
2 80 19 (29) 14
3 84 18 (29) 14
4 76 20 (29) 14
The average artifact depth is the average value of the maximum
artifact depth of each slice 19–28
Relative SNR (%) is calculated by comparing the SNR of each series
to the average SNR of the four series in the respective session
acquired without the EEG cap
Magn Reson Mater Phy
123
however, indicated that the automated analysis is capable
of detecting occurrences that will likely not be observed if
the image data will undergo only visual inspection.
The superficial artifacts typically observed in the EEG–
fMRI measurement were comparable to those measured by
Bonmassar et al. [26] and Mullinger et al. [14]. The
obvious explanation for the signal dropout around the
electrodes was the B0 inhomogeneity. The strength of the
artifacts may depend on the type of electrodes and gel used.
During the first two rounds of measurements, we observed
effects extending as far as 50 mm from the phantom sur-
face. However, the strongest artifacts were absent from the
subsequent measurements when the ECG and EOG elec-
trodes were consistently positioned outside the field of
view. The largest observed artifacts in these measurements
were in the uppermost slices, where the number of elec-
trodes affecting the respective slice was greatest. The SNR
reduction probably resulted from higher coil impedance
when the EEG cap is inside the coil [11]. The effect of
EEG electrode-induced artifacts on the SNR measurement
can be ignored, because the SNR did not significantly
correlate with the analysis of ROI size. However, the SNR
reduction was somewhat smaller when the ECG and EOG
electrodes were positioned further from the imaging area,
which may explain the difference in observations.
In the EEG data, the intermittent noise differences
between the two amplifiers probably resulted from inter-
ference induced within leads or connectors between the
EEG cap and the amplifiers. The variation in the difference
is likely due to slight differences in the cable and amplifier
positions. Our previous experiences in human studies
support this supposition. The amplifier difference, as well
as other observed interferences were considered correct-
able with the help of conventional methods for correcting
EEG artifacts, as they generally perform well with regular
small-amplitude artifacts. The findings also supported the
recommendation to switch off the helium pump during
EEG–fMRI [27].
The artifacts detected in the phantom measurements were
not visible in the brain area in the images of the volunteers.
Thus, if the depth of the artifact in the phantom does not
exceed the depth in our measurements, the margin between
the electrode and the brain should be sufficient. In their study,
Mullinger et al. [14] showed that the artifacts that were
pronounced in the phantom image did not extend into the
brain due to the localized nature of the magnetic field inho-
mogeneities. Bonmassar et al. [26] reported a maximum
artifact depth of 15 mm from the surface, which should not
affect the clinical results. The SNR loss measured in the
images of volunteers, acquired with the EEG cap, was in line
with the phantom measurements. In a respective study,
however, Luo and Glover [28] demonstrated an SNR loss of
only 8 %. The difference may be attributed to the smaller
contribution of physiological noise (no stimuli delivered)
and the smaller voxel size in our scans. The lower SNR
resulting from the presence of the EEG equipment may have
some practical consequences, such as limiting the use of
diffusion tensor imaging or other methods where the SNR is
critical in the same imaging session. The type of EEG cap or
gel content may also affect the results [26, 29].
Because fMRI is one of the most demanding applica-
tions of MRI, an organization performing clinical and
scientific fMRI studies obviously requires a regular and
consistent quality assurance protocol. The measurements
and analyses carried out with the protocol suggested by
Friedman and Glover [21] proved straightforward. Suc-
cessful construction of the phantom required expertise in
chemistry, but phantom positioning, measurement and data
analysis were quick and simple procedures. The method
was also appropriate for the repeatable assessment of
Fig. 5 Superficial artifacts visible in the images representing mean pixel values across the fMRI time series. No EEG cap (left); with EEG cap
(right)
Magn Reson Mater Phy
123
image quality in the presence of EEG equipment. Com-
parison to the volunteer data supported the relevance of the
results of the phantom measurements. If simultaneous
EEG–fMRI is performed in a clinical or research envi-
ronment, such a test should be run with the EEG equip-
ment, at least whenever new pieces of equipment are
introduced, to ensure that artifacts and SNR remain within
acceptable limits. Recoding EEG data during the phantom
measurement can also reveal valuable information, such as
the EEG amplifier difference in our study. The results of
this study can serve in establishing a quality assurance
protocol and its reference values. Our results support the
typical values reported by Friedman and Glover for use as
reference values in assessing basic fMRI image quality. In
simultaneous EEG–fMRI, the acceptance limits for SNR
reduction and artifact depth could be set to 20 % and
20 mm, respectively. The ECG or EOG electrodes should
remain outside the imaging field of view. Because many
factors (e.g., cap and gel types) can influence the results,
baseline values may be site-specific.
An important limitation of many MRI quality assurance
protocols is that they fail to assess contrast parameters,
whereas the contrast is an essential property that sets MRI
above other modalities in many diagnostic areas. The fMRI
quality assurance method investigated in this study shares
this same limitation. Although some solutions have been
suggested [30, 31], the BOLD contrast is difficult to mimic
with phantoms. Possible future directions would incorpo-
rate the assessment of the contrast mechanism into the
quality assurance procedure discussed in this study.
Conclusion
We successfully evaluated the data quality of fMRI and
EEG–fMRI using a phantom in a multi-scanner clinical and
research environment. The phantom measurements showed
that the five evaluated MRI systems operated acceptably for
fMRI; nevertheless, we detected some differences in image
reconstruction and center-of-mass changes. We detected
superficial artifacts in the phantom data of EEG-fMRI, but
not in the brain area in the images of human volunteers. We
also observed an SNR reduction of comparable magnitude
in both the phantom and human data. This study demon-
strated a repeatable method for measuring and following up
on the data quality of simultaneous EEG–fMRI.
References
1. Ogawa S, Tank DW, Menon R, Ellermann JM, Kim SG, Merkle
H, Ugurbil K (1992) Intrinsic signal changes accompanying
sensory stimulation: functional braing mapping with magnetic
resonance imaging. Proc Natl Acad Sci USA 89:5951–5955
2. Biswal B, Zerrin Yetkin F, Haughton VM, Hyde JS (1995)
Functional connectivity in the motor cortex of resting human
brain using echo-planar MRI. Magn Reson Med 34:537–541
3. Jezzard P, Matthews PM, Smith SM (2001) Functional MRI: an
introduction to methods. Oxford University Press, New York
4. Purdon PL, Weisskoff RM (1998) Effect of temporal autocorre-
lation due to physiological noise and stimulus paradigm on voxel-
level false-positive rates in fMRI. Hum Brain Mapp 6:239–249
5. Zijlmans M, Huiskamp G, Hersevoort M, Seppenwoolde JH, van
Huffelen AC, Leijten FSS (2007) EEG–fMRI in the preoperative
work-up for epilepsy surgery. Brain 130:2343–2353
6. Tieleman A, Deblaere K, Van Roost D, Van Damme O, Achten E
(2009) Preoperative fMRI in tumour surgery. Eur Radiol
19:2523–2534
7. Goldman RI, Stern JM, Engel J Jr, Cohen MS (2000) Acquiring
simultaneous EEG and functional MRI. Clin Neurophysiol
111:1974–1980
8. Carmichael DW, Vulliemoz S, Rodionov R, Thornton JS, McE-
voy AW, Lemieux L (2012) Simultaneous intracranial EEG–
fMRI in humans: protocol considerations and data quality. Neu-
roImage 63:301–309
9. Gotman J, Kobayashi E, Bagshaw AP, Benar CG, Dubeau F
(2006) Combining EEG and fMRI: a multimodal tool for epilepsy
research. J Magn Reson Imaging 23:906–920
10. Lemieux L, Krakow K, Fish DR (2001) Comparison of spike-
triggered functional MRI BOLD activation and EEG dipole
model localization. NeuroImage 14:1097–1104
11. Carmichael D (2010) Image quality issues. In: Mulert C, Lemieux
L (eds) EEG–fMRI. Springer, Berlin, pp 173–199
12. Scarff CJ, Reynolds A, Goodyear BG, Ponton CW, Dort JC,
Eggermont JJ (2004) Simultaneous 3-T fMRI and high-density
recording of human auditory evoked potentials. NeuroImage
23:1129–1142
13. Krakow K, Allen PJ, Symms MR, Lemieux L, Josephs O, Fish
DR (2000) EEG recording during fMRI experiments: image
quality. Hum Brain Mapp 10:10–15
14. Mullinger K, Debener S, Coxon R, Bowtell R (2008) Effects of
simultaneous EEG recording on MRI data quality at 1.5, 3 and 7
tesla. Int J Psychophysiol 67:178–188
15. Ritter P, Becker R, Freyer F, Villringer A (2010) EEG quality:
the image acquisition artefact. In: Mulert C, Lemieux L (eds)
EEG–fMRI. Springer, Berlin, pp 153–171
16. Mullinger KJ, Castellone P, Bowtell R (2013) Best current
practice for obtaining high quality EEG data during simultaneous
fMRI. J Vis Exp 76:e50283
17. Ritter P, Becker R, Graefe C, Villringer A (2007) Evaluating
gradient artifact correction of EEG data acquired simultaneously
with fMRI. Magn Reson Imaging 25:923–932
18. Lerski RA, De Certaines JD (1993) Performance assessment and
quality control in MRI by Eurospin test objects and protocols.
Magn Reson Imaging 11:817–833
19. De Wilde J, Price D, Curran J, Williams J, Kitney R (2002)
Standardization of performance evaluation in MRI: 13 years’
experience of intersystem comparison. Concepts Magn Reson
15:111–116
20. Weinreb J, Wilcox PA, Hayden J, Lewis R, Froelich J (2005)
ACR MRI accreditation: yesterday, today, and tomorrow. J Am
Coll Radiol 2:494–503
21. Friedman L, Glover GH (2006) Report on a multicenter fMRI
quality assurance protocol. J Magn Reson Imaging 23:827–839
22. Simmons A, Moore E, Williams SCR (1999) Quality control for
functional magnetic resonance imaging using automated data
analysis and Shewhart charting. Magn Reson Med 41:1274–1278
Magn Reson Mater Phy
123
23. Stocker T, Schneider F, Klein M, Habel U, Kellermann T, Zilles
K, Shah NJ (2005) Automated quality assurance routines for
fMRI data applied to a multicenter study. Hum Brain Mapp
25:237–246
24. Allen PJ, Josephs O, Turner R (2000) A method for removing
imaging artifact from continuous EEG recorded during functional
MRI. NeuroImage 12:230–239
25. Preibisch C, Pilatus U, Bunke J, Hoogenraad F, Zanella F, Lan-
fermann H (2003) Functional MRI using sensitivity-encoded
echo planar imaging (SENSE-EPI). NeuroImage 19:412–421
26. Bonmassar G, Hadjikhani N, Ives JR, Hinton D, Belliveau JW
(2001) Influence of EEG electrodes on the BOLD fMRI signal.
Hum Brain Mapp 14:108–115
27. Nierhaus T, Gundlach C, Goltz D, Thiel SD, Pleger B, Villringer
A (2013) Internal ventilation system of MR scanners induces
specific EEG artifact during simultaneous EEG–fMRI. Neuro-
Image 74:70–76
28. Luo Q, Glover GH (2012) Influence of dense-array EEG cap on
fMRI signal. Magn Reson Med 68:807–815
29. Negishi M, Abildgaard M, Laufer I, Nixon T, Constable RT
(2008) An EEG (electroencephalogram) recording system with
carbon wire electrodes for simultaneous EEG–fMRI (functional
magnetic resonance imaging) recording. J Neurosci Methods
173:99–107
30. Renvall V, Joensuu R, Hari R (2006) Functional phantom for
fMRI: a feasibility study. Magn Reson Imaging 24:315–320
31. Olsrud J, Nilsson A, Mannfolk P, Waites A, Stahlberg F (2008) A
two-compartment gel phantom for optimization and quality
assurance in clinical BOLD fMRI. Magn Reson Imaging
26:279–286
Magn Reson Mater Phy
123