The International Practical Temperature Scale (IPTS) defines six primary fixed points for reference temperatures in terms of: the triple point of equilibrium hydrogen 259.34°C the boiling point of oxygen 182.962°C the boiling point of water 100.0°C the freezing point of zinc 419.58°C the freezing point of silver 961.93°C the freezing point of gold 1064.43°C (all at standard atmospheric pressure) Temperature can be measured via a diverse array of sensors. All of them infer temperature by sensing some change in a physical characteristic.

Course of mesurement

Embed Size (px)

Citation preview

Page 1: Course of mesurement

The International Practical Temperature Scale (IPTS) defines six primary fixed points for reference temperatures in terms of:

the triple point of equilibrium hydrogen 259.34°C the boiling point of oxygen 182.962°C the boiling point of water 100.0°C the freezing point of zinc 419.58°C the freezing point of silver 961.93°C the freezing point of gold 1064.43°C(all at standard atmospheric pressure)

Temperature can be measured via a diverse array of sensors. All of them infer temperature by sensing some change in a physical characteristic.

Page 2: Course of mesurement

The various temperature measuring devices can be classified as:

Mechanical Methods

1 .Expansion of solids

) a (Solid rod thermometer

) b (Bimetallic thermometer2 .Expansion of liquids

) a (Liquid-in- glass thermometer

(b) Liquid-in- metal thermometer3. Expansion of gases (a) Gas thermometers

Page 3: Course of mesurement

Electrical Methods

1. Electrical resistance thermometers

2. Thermistors

3. Thermocouples and thermopiles

Based on Radiation Properties

4. Total radiation pyrometer

5. Optical pyrometer

6. Infra-red pyrometer

Page 4: Course of mesurement

Mechanical Methods

1 .Expansion of solids)a (Solid rod thermometer

The change ( Δ l) of the original length l of a solid due to change in temperature Δ t is given by ,

Δ l = l α Δ t Where, α is the coefficient of linear expansion usually taken as a constant over a particular temperature range for a metal.

)b (Bimetal thermometer

Bimetal thermometers are employed usually in the – 30oC to 550 oC temperature range

Page 5: Course of mesurement

Two dissimilar metals are bonded together into what is called a bimetallic strip.

Suppose metal A has a smaller coefficient of thermal expansion than does metal B. As

temperature increases, metal B expands more than does metal A, causing the bimetallic

strip to curl upwards as sketched.

One common application of bimetallic strips is in home thermostats, where a bimetallic

strip is used as the arm of a switch between electrical contacts. As the room temperature

changes, the bimetallic strip bends as discussed above. When the bimetallic strip bends

far enough, it makes contact with electrical leads that turn the heat or air conditioning on

or off.

Another application is in circuit breakers (the bimetallic strip is labeled “5” in the photo

to the right). High temperature indicates over-current, which shuts off the circuit [from

Wikipedia].

Another common application is for use as oven, wood burner, or gas grill thermometers.

These thermometers consist of a bimetallic strip wound up in a spiral, attached to a dial

that is calibrated into a temperature scale.

Page 6: Course of mesurement

2 .Expansion of liquids

) a (Liquid-in- glass thermometer

(b) Liquid-in- metal thermometer

(a )Liquid-in- glass thermometer

The most common and well-known thermometer is the liquid-in-glass

thermometer.

As the temperature rises, the liquid expands, moving up the tube. The scale is

calibrated to read temperature directly.

Usually, mercury or some kind of alcohol is used for the liquid.

Page 7: Course of mesurement

Pressure thermometer

A pressure thermometer operates by the expansion of a gas instead of a liquid or solid.

(Note: There are also pressure thermometers that use a liquid instead of a gas.)

Suppose the gas inside the bulb and tube can be considered an ideal gas. The ideal gas

law is PV = mRT, where P is the pressure, V is the volume, and m is the mass of the gas.

R is the gas constant for the specific gas (not the universal gas constant), and T is the

absolute temperature of the gas.

o Specific gas constant R is a constant. The bulb and tube are of constant volume, so V

is a constant. Also, the mass m of gas in the sealed bulb and tube must be constant

(conservation of mass). Hence, the ideal gas equation reduces to P= constant T.

A pressure thermometer therefore measures temperature indirectly by measuring

pressure.

Page 8: Course of mesurement

The gage is a pressure gage, but is typically calibrated in units of temperature instead.

A common application of this type of thermometer is measurement of outside

temperature from the inside of a building. The bulb is placed outside, with the tube

running through the wall into the inside.

The gage is on the inside. As T increases outside, the bulb temperature causes a corresponding increase in pressure, which is read as a temperature increase on the gage.

Page 9: Course of mesurement

Electrical Methods

1. Electrical resistance thermometers

2. Thermistors

3. Thermocouples and thermopiles

Page 10: Course of mesurement

Electrical Methods

Electrical Methods

Thermoresistive or Variable Resistance

Transducers

ThermistorsResistance

temperature detector )RTD(

Thermo-electric or e.m.f. generating

transducers

Page 11: Course of mesurement

The change of resistance with temperature in case of some materials forms the basis of temperature measurement.

These materials fall into two categories : one using conductors (metals) and other semi-conductors.

The resistance of highly conducting materials (generally metals) increases with temperature. Examples are nickel, copper, platinum, tungsten and silver.

A temperature measuring device using an element of this type is resistance thermometers or resistance temperature detector (RTD). Whereas, the resistance of semiconductors generally decreases with increase of temperature and are called thermistors.

Page 12: Course of mesurement

Resistance temperature detectors

A resistance temperature detector (abbreviated RTD) is basically either a long, small diameter metal wire (usually platinum) wound in a coil or an etched grid on a substrate,

The resistance of an RTD increases with increasing temperature.

In general, the resistance –temperature relationship for most metals over a wide range :of temperature is given by the following empirical relation

R = R0( 1 + AT + BT2)

Where, R = resistance at absolute temperature T R0 = resistance at the reference temperature

A, B = constants for the material

Page 13: Course of mesurement

Over a limited temperature range around 0oC (273 K) the following relationship can be used :

Rt = Ro (1 + α t)

Where, α = the temperature coefficient of resistance of material in oC-1

Ro = resistance at 0oC

t = temperature relative to

If a change in temperature from t1 to t2 is considered, then ,

R1 = Ro (1 + α t1)

R2 = Ro (1 + α t2) Rearrangement gives:

𝑡 2=𝑡1+(𝑅2−𝑅2 )𝛼𝑅𝑜

Page 14: Course of mesurement
Page 15: Course of mesurement

A thermistor is similar to an RTD, but a semiconductor material is used instead of a metal. A thermistor is a solid state device.

Thermistors

A thermistor has larger sensitivity than does an RTD, but the resistance change with temperature is nonlinear, and therefore temperature must be calibrated with respect to resistance.

Thermistors are labeled by their resistance at 25oC. For example, two popular thermistors are type 2252 (2252 Ω at 25oC) and type 5000 (5000 Ω at 25oC)

The upper temperature limit of thermistors is typically lower than that of RTDs. In fact, the maximum temperature of operation is sometimes only 100oC to 200oC for a typical commercial thermistor. However, thermistors have greater sensitivity and are typically more accurate than RTDs or thermocouples.

Thermistors usually have negative temperature coefficient, their resistance decreases as the temperature increases.

Page 16: Course of mesurement

Thermoelectric effect sensors )thermocouples(

Thermoelectric effect sensors rely on the physical principle that, when any two different metals are connected together, an e.m.f., which is a function of the temperature, is generated at the junction between the metals. The general form of this relationship is:

e = a1T + a2T2 + a3T3 + ……..+anTn

where e is the e.m.f. generated and T is the absolute temperature.

This is clearly non-linear, which is inconvenient for measurement applications. Fortunately, for certain pairs of materials, the terms involving squared and higher powers of T (a2T2 + a3T3 etc.) are approximately zero and the e.m.f.–temperature relationship is approximately linear according to:

e ≈ a1T

Wires of such pairs of materials are connected together at one end, and in this form are known as thermocouples.

Page 17: Course of mesurement

Thermocouples are manufactured from various combinations of the base metals copper and iron, the base-metal alloys of alumel (Ni/Mn/Al/Si), chromel (Ni/Cr), constantan (Cu/Ni), nicrosil (Ni/Cr/Si) and nisil (Ni/Si/Mn), the noble metals platinum and tungsten, and the noble-metal alloys of platinum/rhodium and tungsten/rhenium.

Only certain combinations of these are used as thermocouples and each standard combination is known by an internationally recognized type letter, for instance type K is chromel–alumel.

The e.m.f.–temperature characteristics for some of these standard thermocouples are shown in Figure 14.1: these show reasonable linearity over at least part of their temperature-measuring ranges.

Page 18: Course of mesurement

A typical thermocouple, made from one chromel wire and one constantan(a copper–nickel alloy) wire, is shown in Figure(a). For analysis purposes, it is useful to represent the thermocouple by its equivalent electrical circuit, shown in Figure (b). The e.m.f. generated at the point where the different wires are connected together is represented by a voltage source, E1, and the point is known as the hot junction.

The e.m.f. generated at the hot junction is measured at the open ends of the thermocouple, which is known as the reference junction.

Page 19: Course of mesurement

To save money of using wire to many centimeters, extension leads up to several metres long are normally connected between the thermocouple and the measuring instrument. This modifies the equivalent circuit to that shown in Figure (a).

There are now three junctions in the system and consequently three voltage sources, E1, E2 and E3, with the point of measurement of the e.m.f. (still called the reference junction) being moved to the open ends of the extension leads.

Page 20: Course of mesurement

The measuring system is completed by connecting the extension leads to the voltage measuring instrument. As the connection leads will normally be of different materials to those of the thermocouple extension leads, this introduces two further e.m.f.-generating junctions E4 and E5 into the system as shown in Figure (b). The net output e.m.f. measured (Em) is then given by:

Em = E1 + E2 + E3 + E4 + E5

and this can be re-expressed in terms of E1 as:E1 = Em - E2 - E3- E4 - E5

It is usual to choose materials for the extension lead wires such that the magnitudes of E2 and E3 are approximately zero, irrespective of the junction temperature.

A zero junction e.m.f. is most easily achieved by choosing the extension leads to be of the same basic materials as the thermocouple, but where their cost per unit length is greatly reduced by manufacturing them to a lower specification.

However, such a solution is still prohibitively expensive in the case of noble metal thermocouples, and it is necessary in this case to search for base-metal extension leads that have a similar thermoelectric behaviour to the noble-metal thermocouple.

In this form, the extension leads are usually known as compensating leads.

Page 21: Course of mesurement

A typical example of this is the use of nickel/copper–copper extension leads connected to a platinum/rhodium–platinum thermocouple. Copper compensating leads are also sometimes used with some types of base metal thermocouples and, in such cases, the law of intermediate metals can be applied to compensate for the e.m.f. at the junction between the thermocouple and compensating leads.

Example 1If the e.m.f. output measured from a chromel–constantan thermocouple is 13.419mVwith the reference junction at 0°C, the appropriate column in the tables shows that thiscorresponds to a hot junction temperature of 200°C.

Example 2If the measured output e.m.f. for a chromel–constantan thermocouple (reference junction at 0°C) was 10.65 mV, it is necessary to carry out linear interpolation between thetemperature of 160°C corresponding to an e.m.f. of 10.501mV shown in the tables and the temperature of 170°C corresponding to an e.m.f. of 11.222 mV. This interpolationprocedure gives an indicated hot junction temperature of 162°C.

Page 22: Course of mesurement

Thermocouple types The five standard base-metal thermocouples are:

chromel–constantan (type E),

iron–constantan (type J),

chromel–alumel (type K),

nicrosil–nisil (type N)

copper–constantan (type T). Chromel–constantan devices give the highest measurement sensitivity of 80 μV/°C,

with an inaccuracy of š0.5% and a useful measuring range of 200°C up to 900°C.

Iron–constantan thermocouples have a sensitivity of 60 μV/°C and are the preferredtype for general-purpose measurements in the temperature range 150°C to °C1000°C,where the typical measurement inaccuracy is ±0.75%. Thermocouples made from platinum and a platinum–rhodium alloy (type R and type

S) have a low inaccuracy of only ± 0.5% and can measure temperatures up to 1500°C, but their measurement sensitivity is only 10 μV/°C.

Alternative devices made from tungsten and a tungsten/rhenium alloy have a better sensitivity of 20 μV/°C and can measure temperatures up to 2300°C.

Page 23: Course of mesurement

Thermocouple manufacturers have standardized the use of certain pairs of metals for thermocouples.

Each standard thermocouple has been assigned both a letter and a color. For example, type J and type T thermocouples are used in our ME 345 lab.

Standard thermocouples

A type J thermocouple has the color black, and uses iron and constantan as its component metals. (Constantan is an alloy of ≈ 45% nickel and ≈ 55% copper.)

A type T thermocouple has the color blue, and uses copper and constantan as its component metals.

A type K thermocouple has the color yellow, and uses chromel and alumel as its component metals. (Chromel is an alloy of ≈ 10% chromium and ≈ 90% nickel.) Type K thermocouples are the most popular variety in use today. In fact, many digital multimeters (DMMs) can measure temperature by plugging in a type K thermocouple with standard connections.

By convention, the reference temperature for thermocouple tables is 0oC. Hence, all thermocouple voltages are given relative to 0oC – the voltage at 0oC is listed as zero volts.

If you wire the thermocouple according to one of the above diagrams with TR = 0oC, the thermocouple voltage at any sensing temperature should match with that listed in the tables.

Page 24: Course of mesurement

Quartz crystal thermometer

It is based on the principle that the resonant frequency of a quartz crystal is a function of temperature

The normal operating range of a quartz crystal thermometer is – 40oC to 230oC with a resolution of 0.1oC.

Page 25: Course of mesurement

Thermocouple Laws

First some notation: Let T1 be the temperature of bath 1, and T2 be the temperature of bath 2. Let V1-Rbe defined as the voltage produced by a thermocouple at temperature T1

when a proper reference junction at temperature TR is used (TR = reference temperature = 0oC). Note that V1-R is the voltage listed

in the thermocouple tables at temperature T1. Let V1-2be defined as the difference in voltage between V1-Rand V2-R, namely, V1-2= V1-R - V2-R.

Page 26: Course of mesurement

Law of intermediate metals

Page 27: Course of mesurement

What is a total radiation pyrometer? What are its advantages and disadvantages?

A pyrometer is used to measure the temperature of an object from a distance, without making contact.

The method used for making these non-contacting temperature measurements is known as radiation pyrometry. Non-contact temperature sensors use the concept of infrared radiant energy to measure the temperature of objects from a distance.

After determining the wavelength of the energy being emitted by an object, the sensor can use integrated equations that take into account the body's material and surface qualities to determine its temperature. You can read

The benefit of a radiation pyrometer is there is no need for direct contact with what you are measuring the temperature of. These pyrometers come in handy when the object is moving, as on an assembly line; when it is very hot, as inside a furnace; or when it is inaccessible, as inside a high-vacuum chamber.

Radiation Methods )Pyrometry(

Page 28: Course of mesurement

Under the following situations , the radiation method should be used :

1. The body is too hot and a contact may damage or melt the device or thermometer.

2. The body is not accessible for actual physical contact as it may be moving or is too far off.

3. When temperature distribution over the surface of the body to be determined.

Page 29: Course of mesurement

All bodies above absolute zero temperature radiate energy.

Heat radiations emitted from the body above temperatures 650oC are of sufficient intensity and are used for temperature measurements.

The instruments based on sampling and measuring radiant energy from the hot body re called pyrometrys.

There are three types of pyrometrys :

1. Total radiation pyrometrys.

2. Optical pyrometrys.

3. Infrared pyrometrys.

Page 30: Course of mesurement

o Radiative properties of an object change with temperature.o So, radiative properties are measured to infer the temperature of the object.

Principle of operation:

o The advantages of radiative pyrometry are: There is no physical contact with the object whose temperature is being

measured. Very high temperatures can be measured.

Page 31: Course of mesurement
Page 32: Course of mesurement

A pyrometer is any temperature-measuring device that includes a sensorand a readout. However, in this section we will discuss only radiationtypepyrometers. A radiation pyrometer is a noncontact temperaturesensor that infers the temperature of an object by detecting its naturallyemitted thermal radiation. An optical system collects the visible and infraredenergy from an object and focuses it on a detector, as shown inFigure 7-21. The detector converts the collected energy into an electricalsignal to drive a temperature display or control unit.

Total Radiation Pyrometry :

Page 33: Course of mesurement

Objects radiate electromagnetic energy. The higher the temperature of an object the more electromagnetic radiation it emits. This radiation occurs within the infrared portion of the electromagnetic spectrum (see Figure 3-11).

Infrared

Infrared temperature sensors work by measuring the amount of infrared energy emitted by an object, allowing us to measure the temperature of an object without being in contact.

Page 34: Course of mesurement

Ideal black body is the basis of infrared temperature measurement and gives rise to the emittance setting.

The emittance of a real surface is the ratio of the thermal radiation emitted by the surface at a given temperature to that of the ideal black body at the same temperature.

Emittance is a decimal number that ranges between 0 and 1 or a percentage between 0% and 100%.

o An infrared pyrometer infers the temperature of a hot surface by measuring the temperature of a detector inside a detector chamber as sketched to the right.

o The detector itself is usually a thermopile. It measures Tdet, the temperature of the detector inside the chamber.o Tind is the indicated temperature, which is calculated from Tdet, from the known geometry and the radiation equations. Tind is calibrated as a function of TH for a body of some assumed emissivity εassumed, typically 0.95 or so.

Page 35: Course of mesurement
Page 36: Course of mesurement

o The instrument is set up such that Tind is a function of the voltage output, and the display typically indicates temperature Tind rather than voltage Vdet.

o Tind can be thought of as an uncorrected estimate of TH, since the emissivity of the object may not be the same as that assumed by the infrared pyrometer. In other words, if the actual emissivity of the object is not the same as the assumed emissivity, Tind is incorrect.

Page 37: Course of mesurement

o Optical pyrometer

o An optical pyrometer is useful for measuring very high temperatures of things that glow (even flames).

o It consists of a telescope with an eye piece and a built-in internal wire through which electrical current is passed until it gets so hot that it glows (much like a lightbulb filament).

o The temperature of the glowing wire is calibrated as a function of the supplied current.

o An optical pyrometer works by comparing the glowing wire of known temperature to the glow (optical radiation) from a hot object onto which the telescope is focused.

o When the internal wire and the glow of the object are the same color, the temperatures are approx. equal.

o Since the temperature of the internal wire is controlled and known, the temperature of the object is inferred

The working of an optical pyrometr is based on the principle that at a given wavelength the radiant intensity (brightness) of a black-body varies with temperature

Page 38: Course of mesurement
Page 39: Course of mesurement

Temperature measurement in high speed flow

Page 40: Course of mesurement

Temperature Measurement Using Radiation

A radiation thermometer is an instrument which collects radiation from a target and produces an output signal, usually electrical, related to the radiance, which is used to infer

the temperature of the target.

Page 41: Course of mesurement

Instruments to measure temperature can be divided into separate classes according to the physical principle on which they operate. The main principles used are:

The thermoelectric effect Resistance change Sensitivity of semiconductor device Radiative heat emission Thermography Thermal expansion Resonant frequency change Sensitivity of fibre optic devices Acoustic thermometry Colour change Change of state of material.