Compressed Sensing for Diffusion MRI of the Mouse Brain: A Trade-off of Spatial and Directional Information by Anthony Albert Salerno A thesis submitted in conformity with the requirements for the degree of Master of Science Department of Medical Biophysics University of Toronto c Copyright 2018 by Anthony Albert Salerno

Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Compressed Sensing for Diffusion MRI of the Mouse Brain:A Trade-off of Spatial and Directional Information

by

Anthony Albert Salerno

A thesis submitted in conformity with the requirementsfor the degree of Master of ScienceDepartment of Medical Biophysics

University of Toronto

c© Copyright 2018 by Anthony Albert Salerno

Page 2: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Abstract

Compressed Sensing for Diffusion MRI of the Mouse Brain: A Trade-off of Spatial and

Directional Information

Anthony Albert Salerno

Master of Science

Department of Medical Biophysics

University of Toronto

2018

Diffusion Magnetic Resonance Imaging (MRI) uses loss of signal coherence caused by

movement of water molecules to determine how water diffuses in tissue. These scans are

known to take a significant amount of time owing to the repeated acquisition of images,

each measuring diffusion for a particular direction and weighting. Emerging methods

based on collection of increasingly large numbers of diffusion directions and/or weight-

ings promise to provide a more detailed characterization of brain microstructure than

previously possible. However, because the time required for these methods can render

them impractical, I present a method of acquiring and reconstructing diffusion MRI data

based on Compressed Sensing (CS) that enables flexible trade-off between spatial sam-

pling and diffusion direction/weighting sampling. This method enables the collection of

more diffusion data in a given scan time at the cost of extended reconstruction time.

ii

Page 3: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

This work is dedicated to my late cousin Enzo Cardillo who spent many days and

nights at The Hospital for Sick Children during his short life. No matter what state he

was in, he did everything he could to make me smile. It didn’t matter that he barely

had the energy to go to school, whenever he saw me, he made me laugh and be able to

just enjoy being a kid. If it wasn’t for him, I doubt I would have had the passion I do

for my work, or the gratitude for the amazing research that occurs at The Hospital for

Sick Children. In the same breath, I want to thank the doctors, nurses, and staff who

worked tirelessly to give me a few extra years with him.

Acknowledgements

Many thanks go to my parents Mary and Frank Salerno, my siblings Stefania, Muriel,

and Cesare, as well as Elisa De Luca for their love and support during this process, I

know I wasn’t the easiest person to deal with. I also want to thank all of the staff and

students at the Mouse Imaging Centre for their continued support and chats at my desk

where many coffees were shared.

Thank you to my committee members Dr. John Sled and Dr. Chris Macgowan for their

words of encouragement, probing questions, and brilliant ideas during my committee

meetings.

And finally, a special thank you to my supervisor Dr. Brian Nieman, who put in long

hours to help me with my project and to keep the Mouse Imaging Centre running at its

full potential.

iii

Page 4: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Contents

Acknowledgements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii

Table of Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv

List of Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii

List of Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix

1 Introduction 1

1.1 Magnetic Resonance Imaging . . . . . . . . . . . . . . . . . . . . . . . . 1

1.1.1 The Signal of MRI . . . . . . . . . . . . . . . . . . . . . . . . . . 2

1.2 Diffusion-Weighted Magnetic Resonance Imaging . . . . . . . . . . . . . 4

1.2.1 The Diffusion Process . . . . . . . . . . . . . . . . . . . . . . . . 4

1.2.2 Diffusion MRI Sequence . . . . . . . . . . . . . . . . . . . . . . . 4

1.2.3 Diffusion Tensor Imaging . . . . . . . . . . . . . . . . . . . . . . . 8

1.2.4 Higher Order Models of Diffusion MRI . . . . . . . . . . . . . . . 10

Diffusion Kurtosis Imaging (DKI) . . . . . . . . . . . . . . . . . . 10

High Angular Resolution Diffusion Imaging (HARDI) . . . . . . . 11

Neurite Orientation Dispersion and Density Imaging (NODDI) . . 11

1.2.5 Clinical Uses of Diffusion MRI . . . . . . . . . . . . . . . . . . . . 12

1.3 The Mouse as a Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

1.3.1 MRI in the Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . 13

1.3.2 Diffusion MRI in the Mouse . . . . . . . . . . . . . . . . . . . . . 14

1.4 Compressed Sensing (CS) . . . . . . . . . . . . . . . . . . . . . . . . . . 15

1.4.1 Introduction to Compressed Sensing . . . . . . . . . . . . . . . . 15

1.4.2 Current Uses of Compressed Sensing in Human Research . . . . . 19

iv

Page 5: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

1.5 Aim . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

2 Methods 22

2.1 Adaptation of Compressed Sensing for Diffusion Data . . . . . . . . . . . 22

2.1.1 Total Variation in the Diffusion Dimension . . . . . . . . . . . . . 23

2.1.2 Mathematical Approximations and Definitions . . . . . . . . . . . 24

`1 Norm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

TV Kernel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

2.1.3 Phase in CS Reconstructions . . . . . . . . . . . . . . . . . . . . . 26

2.2 Sampling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

2.2.1 Choosing Directions for Diffusion Acquisition . . . . . . . . . . . 28

2.2.2 Spatial Undersampling Across Diffusion Directions . . . . . . . . 29

2.3 Implementation of ex vivo Diffusion RARE . . . . . . . . . . . . . . . . . 32

2.3.1 Acquisition Parameters . . . . . . . . . . . . . . . . . . . . . . . . 32

2.3.2 Phase Encode Partitioning Scheme . . . . . . . . . . . . . . . . . 33

2.3.3 Correction of Diffusion-Gradient Induced Phase . . . . . . . . . . 34

Interleaving ±~d Acquisitions: The “Checkerboard” . . . . . . . . 34

Diffusion-Gradient Induced Phase Corrections . . . . . . . . . . . 36

2.4 CS Reconstruction with ex vivo Diffusion RARE . . . . . . . . . . . . . . 38

2.4.1 Slice-to-Slice Correction . . . . . . . . . . . . . . . . . . . . . . . 38

2.5 Mice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

3 Results 43

3.1 CS Reconstruction – Proof of Principle . . . . . . . . . . . . . . . . . . . 43

3.1.1 Retrospectively Undersampled Shepp-Logan Phantom . . . . . . . 43

3.1.2 Retrospectively Undersampled Anatomical in vivo Data . . . . . . 43

3.2 Diffusion MRI Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

3.2.1 “Checkerboard” Correction of Diffusion-Gradient Induced Phase . 45

3.2.2 Compressed Sensing Reconstruction . . . . . . . . . . . . . . . . . 47

3.3 Potential Use for HARDI Analysis . . . . . . . . . . . . . . . . . . . . . . 49

v

Page 6: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

4 Discussion 55

4.1 Phase Corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

4.1.1 ~d and -~d “Checkerboard” Acquisition Scheme . . . . . . . . . . . 56

4.2 Limitations of the Study . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

4.2.1 Phase Measurements . . . . . . . . . . . . . . . . . . . . . . . . . 56

4.2.2 Trade-Offs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

4.2.3 Empirical Tuning of the Weighting Factors (λi) . . . . . . . . . . 57

4.3 Future Work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58

4.3.1 Further Applications in Diffusion . . . . . . . . . . . . . . . . . . 58

4.3.2 Potential Use in Clinic . . . . . . . . . . . . . . . . . . . . . . . . 59

5 Conclusion 60

Bibliography 62

vi

Page 7: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

List of Figures

1.1 RARE pulse sequence example with the use of a diffusion gradient . . . . 7

1.2 Sparsifying terms applied to a Shepp-Logan phantom . . . . . . . . . . . 17

2.1 Comparison of 1a

ln(cosh(ax)) and sgn(x) . . . . . . . . . . . . . . . . . . 26

2.2 Differences in the probability density functions for uniform and variable

density undersampling methods . . . . . . . . . . . . . . . . . . . . . . . 30

2.3 Uniform density undersampling method with directional biases included . 32

2.4 Checkerboarded sampling pattern with ±~d . . . . . . . . . . . . . . . . . 35

2.5 Flowchart of phase corrections . . . . . . . . . . . . . . . . . . . . . . . . 39

2.6 Simplified flowchart of the data collection and processing using a diffusion

specific CS acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

2.7 Flowchart of the CS algorithm for a diffusion acquisition . . . . . . . . . 41

2.8 Slice-to-slice intensity corrections . . . . . . . . . . . . . . . . . . . . . . 42

3.1 Shepp-Logan phantom CS reconstruction . . . . . . . . . . . . . . . . . . 44

3.2 Anatomical CS reconstruction . . . . . . . . . . . . . . . . . . . . . . . . 45

3.3 Representative phase corrections . . . . . . . . . . . . . . . . . . . . . . . 46

3.4 Comparison of fully-sampled and undersampled image reconstructions for

an individual diffusion direction. . . . . . . . . . . . . . . . . . . . . . . . 48

3.5 Comparison of fully-sampled and undersampled image acquisitions with

matched acquisition times. . . . . . . . . . . . . . . . . . . . . . . . . . . 50

3.6 Comparison of FA maps in 30-direction fully-sampled and 120-direction

undersampled diffusion acquisitions. . . . . . . . . . . . . . . . . . . . . . 51

3.7 A depiction of the complete 30-direction and 120-direction data sets. . . 52

vii

Page 8: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

3.8 HARDI analysis of 120-direction undersampled diffusion MRI data. . . . 54

viii

Page 9: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

List of Abbreviations

ADC Apparent Diffusion Coefficient.

CEST Chemical Exchange Saturation Transfer.

CS Compressed Sensing.

CSF cerebro-spinal fluid.

DKI Diffusion Kurtosis Imaging.

DSI Diffusion Spectrum Imaging.

DTI Diffusion Tensor Imaging.

EPI Echo Planar Imaging.

FA Fractional Anisotropy.

fMRI functional MRI.

FOV field of view.

GDiff diffusion gradient.

GPE1 phase encode gradient 1.

GPE2 phase encode gradient 2.

GRO readout gradient.

ix

Page 10: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

HARDI High Angular Resolution Diffusion Imaging.

MD Mean Diffusivity.

MRI Magnetic Resonance Imaging.

NODDI Neurite Orientation Dispersion and Density Imaging.

ODF orientation distribution function.

PDF probability density function.

QBI Q-Ball Imaging.

RARE Rapid Acquisition with Relaxation Enhancement.

RF radio frequency.

SENSE Sensitivity Encoding.

SNR signal-to-noise ratio.

TV Total Variation.

x

Page 11: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Chapter 1

Introduction

1.1 Magnetic Resonance Imaging

MRI plays an integral role in medical imaging in both the clinical and research worlds for

non-invasive anatomical, physiological, and functional measurements of the body. MRI

has the ability to acquire images with multiple different contrasts by changing the pa-

rameters of acquisition [1]. Having this flexibility is important for research and allows

the scientist the opportunity to optimize image acquisitions for the assessment of human

or preclinical models, in vivo or ex vivo samples, and for variability in tumour charac-

teristics, for instance [2, 3, 4, 5]. In biomedical research, MRI provides high anatomical

resolution with excellent soft tissue contrast; however, it is plagued by long scan times,

especially for methods that probe physiology or microstructure of the sample. The long

scan times for the latter methods are attributable to models of underlying microstructure

that are based on contrast changes in an image series, where the small intensity differ-

ences between each scan are used to help derive outcomes of interest. In the interest of

using MRI resources to obtain as much information as possible, it is important that im-

age acquisitions are tailored to balance spatial resolution and information for modelling

of microstructure within the time and resources available.

1

Page 12: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

2

1.1.1 The Signal of MRI

In order to generate signal in MRI, first a strong static field ~B0 must be applied uniformly

over the sample. The following is described in a manner similar to [1].

Conventionally, ~B0 ≡ B0k. This generates a bulk magnetization ~M0 in the sample,

aligned with this strong field. In order to obtain signal, the net magnetization must

be rotated to have a component perpendicular to ~B0, which is achieved by applying a

radio frequency (RF) pulse (via another magnetic field, ~B1) at the Larmor frequency,

ω0 = γ| ~B0|, where γ represents the gyromagnetic ratio. Depending on the frequency

profile and duration of ~B1, ~M can be rotated to arbitrary angles from the main magnetic

field ~B0.

Once the net magnetization ~M has a component perpendicular to ~B0, it precesses

around ~B0. The precession can be detected through current induced in a RF coil via

the Faraday-Lenz law of induction, which is the signal that is measured. The precession

is described by both a magnitude and phase, represented through real and imaginary

components in the signal. The precessing magnetization is transient, and returns to its

equilibrium state as described by two relaxation times, T1 and T2, which represent the

longitudinal and transverse relaxation times respectively. The former (T1) describes the

rate of signal recovery along ~B0, while the latter (T2) describes the rate of signal decay

perpendicular to ~B0. This can all be represented by the Bloch Equation:

d ~M(t)

dt= ~M(t)× γ ~B(t)− (Mx(t)i+My(t)j)

T2

− (Mz(t)−M0)k

T1

(1.1)

~M(t) =

Mx

My

Mz

=

e−tT2 0 0

0 e−tT2 0

0 0 e−tT1

~Rz(ω0t) ~M0 +

0

0

M0(1− e−tT1 )

(1.2)

Where ~B is the total magnetic field and ~Rz is the rotation matrix for rotation about the

z-axis. It is also convenient to define Mxy such that the transverse component can be

Page 13: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

3

written more simply.

Mxy = Mx + iMy (1.3)

dMxy

dt=dMx

dt+ i

dMy

dt(1.4)

dMxy

dt= −

(1

T2

+ iω0

)Mxy (1.5)

The solution to Equation (1.5) describes the evolution of transverse magnetization with

explicit relaxing and precessing parts:

Mxy = M⊥e−t/T2e−iω0t (1.6)

Where M⊥ is the magnitude of the transverse magnetization.

For imaging purposes, signal across space is distinguished by changing the magnetic

field strength linearly as a function of position. Detection of the associated change in

frequency (“frequency encoding”) or equivalently the change in phase over a defined unit

of time (“phase encoding”) thus maps signal to spatial coordinates. The imposed linear

change in magnetic field is generally referred to as a magnetic field gradient (henceforth

referred to simply as “gradient”). For each of the three dimensions, the gradients are

applied such that:

~B = ~B0 + ~G · ~r (1.7)

Where ~r represents the location relative to the isocentre of the magnet where the strength

of ~B0 is unperturbed. Incorporating the spatial variation induced by the gradients into

Equation (1.6), we may write:

Mxy(~r, t) = M⊥(~r)e−t/T2(~r)e−iω0te−iγ∫ t0~G(τ)·~rdτ (1.8)

Finally, to obtain signal off of this magnetization, we would integrate over all three

dimensions:

s(t) =

∫x

∫y

∫z

Mxy(~r, t) dx dy dz (1.9)

Page 14: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

4

1.2 Diffusion-Weighted Magnetic Resonance Imag-

ing

1.2.1 The Diffusion Process

The process of molecular diffusion is defined by Brownian motion of particles as a result

of the thermal energy they hold [6]. In free space, the molecular displacement distribution

follows a uniform Gaussian in three dimensions; this is statistically described as a diffusion

coefficient, D. The diffusion coefficient is in SI units of m2s−1 (or more conveniently,

mm2s−1), relating to a mean squared displacement of each individual particle; for free

water at 37◦C this value is 3×10−3 mm2s−1 [7]. Within the brain, however, the molecular

displacements differ from an isotropic Gaussian model due to the presence of tissue

components such as cell membranes and macromolecules [7]. Within very short times in

these samples (short enough such that there will be minimal interaction with barriers),

the diffusion will look Gaussian. However as the diffusion time increases, the obstacles

become important and affect the diffusion observed.

1.2.2 Diffusion MRI Sequence

The microstructure of the brain can be probed by measuring how water diffuses in dif-

ferent directions or areas of the brain. Deviation from isotropic Gaussian diffusion is

particularly evident in white matter, as myelinated axons act as tracts, inhibiting the

diffusion of water perpendicular to the orientation of the axon. Diffusion MRI exploits

this phenomenon to elucidate the microstructure of the brain by probing the diffusion

rate of water within tissue at a multitude of different directions, ~d, with the expectation

that certain directions will restrict the diffusion of water more than others based on the

orientation of these barriers [8, 9, 10, 11, 12]. This indicates, for instance, the degree of

alignment of axon bundles.

Diffusion MRI is similar in its acquisition to anatomical MRI except for its one defining

factor – the introduction of strong diffusion sensitizing gradients. These gradients are

identical to the imaging gradients described, except applied with larger magnitude and

Page 15: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

5

usually separated in time. They are applied within the pulse sequence in order to probe

diffusion of water, a single vector direction at a time, mapping the result over all space

in the image volume [13, 14, 15, 16, 17].

The detection of diffusion through MRI originates from protons that move along the

diffusion gradient. The notion is that where diffusion exists, signal is attenuated due to

a loss of phase coherence between the protons within a voxel. The method is as follows:

A diffusion gradient is applied, and generates a change in the phase of each proton as a

function of position along the sample in the direction of the gradient. After this gradient

is turned off, a 180◦ pulse is then applied (thus reversing the accumulated phase), and

the same gradient is applied again. For a stationary signal the amount of phase gained

from the first application of the diffusion gradient will be exactly cancelled by the second

gradient, effectively leaving zero net phase. However, if the proton has moved along the

probed axis, phase induced by the second gradient does not match the first, and a net

residual phase is present. This will lead to reduced phase coherence within the voxel and

a loss of signal. Increased levels of diffusion, through pathology, temperature or other

changes, lead to greater signal loss due to decreased phase coherence within each voxel.

Larger diffusion gradients and timings also produce larger phase deviations, and hence

greater decreases in signal intensity.

The impact of diffusion can be calculated beginning from Equation (1.6), but collect-

ing the T2 relaxation term and the precession term into a single value, S0, for simplicity

[18], the magnetization can be written as:

Mxy(~r, t) = S0e−iγ( ~G(t)·~r)t (1.10)

If diffusion within the sample follows Fick’s law, the magnetization of the system can be

described by a modified differential equation accounting for diffusion:

dMxy(~r, t)

dt=[−iγ(~G(t) · ~r) + D∇2

]Mxy(~r, t) (1.11)

Where ∇2 represents the Laplacian and D represents the diffusivity of the sample, which

here has been assumed to be isotropic. The solution to this differential equation can

Page 16: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

6

be isolated by assuming a time-varying amplitude (referenced to S0) multiplied by a

precession term. It is convenient in this solution to define a parameter known as the

b-value – in units of s/mm2 – as follows:

Mxy(~r, t) = S0e−bDe−iγ( ~G(t)·~r)t (1.12)

b = γ2

∫ τ

0

[∫ t

0

G(t′)dt′]2

dt (1.13)

All diffusion imaging requires a comparison of data across two b-values, but the first

is generally set to b = 0, which is equivalent to a purely anatomical scan and called

the b0 image. For diffusion weighting based on a pair of matched idealized rectangular

gradients, as was solved in [18], the b-value can be simplified as follows:

b = γ2G2δ2

(∆− δ

3

)(1.14)

Where γ is the gyromagnetic ratio in MHz/T, G is the strength of the diffusion gradient

pulse in units of T/mm, δ is the duration of the pulse in units of ms, and ∆ is the time,

also in units of ms, between the beginning of the diffusion gradients. An example of

diffusion gradients (GDiff) and the respective timings are pictured as part of an imaging

sequence in Figure 1.2.

In order to probe the directional dependence of diffusion, an image is collected for

many different diffusion gradient directions covering the 3-dimensional Euclidian sphere,

which for a given b value will maintain a constant |GDiff|, but different combinations of

GRO, GPE1, and GPE2 to probe direction. The diffusion in any given region can be either

isotropic (not directionally dependent) or anisotropic (directionally dependent). If there

is isotropic diffusion at a given voxel, for all diffusion gradient directions there will be

consistent intensity differences from the b0 scan [19]. However, for anisotropic diffusion,

some directions will have higher intensities than others due to the restriction of water

diffusion. An example of this would be in the aforementioned white matter tracts where

myelin sheaths and axonal membranes act as barriers to motion perpendicular to their

principle axes, but have little effect on travel along their axis [19, 20, 21, 22].

Page 17: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

7

Figure 1.1: One TR of a Rapid Acquisition with Relaxation Enhancement (RARE) sequence witha diffusion sensitivity gradient applied. The diffusion gradient (GDiff) is applied across the readoutgradient (GRO), phase encode gradient 1 (GPE1), and phase encode gradient 2 (GPE2) but is shownseparately for readability. The values of G, δ, and ∆ can be changed to obtain the desired b-value.RDG represents the read dephase gradient (in black), SG represents the spoiler gradient (in gray), RGrepresents the read gradient (in green), and PEG represents the phase encode gradient (unshaded).

Since an entire image must be acquired for each diffusion direction assessed, diffusion

MRI scans are known to take significantly longer than standard anatomical scans. Con-

sequently, other means of speeding image acquisition are usually employed including the

use of different pulse sequences, such as Echo Planar Imaging (EPI) [23, 24, 25].

The ability to infer microstructural changes from diffusion scans is dependent upon the

model employed to describe diffusion-induced intensity changes, which in turn determines

the number of directions and b-values acquired. More sophisticated, flexible models

require a large number of directions and/or b-values to probe finer angular dependence of

diffusion or diffusion restrictions. The simplest and most widely employed model is called

Diffusion Tensor Imaging (DTI) and will be discussed below along with a description of

more advanced models.

Page 18: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

8

1.2.3 Diffusion Tensor Imaging

As one of the first models to explain diffusion MRI, DTI [21, 22, 26] attempts to probe

the anisotropy of water diffusion by generating an ellipsoidal probability distribution

[18, 27, 28, 29]. The model is applied across all voxels in a set of images and does not

use any prior information about anatomical location.

Through DTI, the intensity of images as a function of direction is fit according to the

diffusion tensor D.

Sk = S0 e−b gTk D gk (1.15)

D =

Dxx Dxy Dxz

Dyx Dyy Dyz

Dzx Dzy Dzz

(1.16)

In this representation, S0 represents the signal intensity of the b0 image, Sk is defined

as the signal intensity of the image at the kth diffusion direction, gk is a 3 × 1 unit length

column vector defining the kth diffusion gradient and D represents the diffusion tensor as

a 3×3 matrix [8, 21, 22]. By definition, the tensor is a symmetric matrix (i.e. Dij = Dji)

and as such there are only six unique elements that must be calculated – namely the

upper triangle. From the diffusion tensor, a number of diffusion properties can be inferred,

mainly based on the eigenvalues and eigenvectors of the diffusion tensor matrix. Here, it is

convenient to let λi represent the eigenvalues, and ~vi represent the individual eigenvectors.

Generally, the eigenvalues are presented such that λ1 ≥ λ2 ≥ λ3 ≥ 0. λ1 represents the

largest of the eigenvalues, and is often called the axial or parallel diffusivity – it can also

be represented by λ‖. Along with its eigenvector, it represents the degree and direction of

the greatest diffusion in the voxel. The other two eigenvalues can be averaged in order to

create the radial or perpendicular diffusivity, λ⊥, representing the diffusion in the plane

perpendicular to the axial direction. Finally, the Mean Diffusivity (MD) is the average of

all three eigenvalues – λ – which is also called the Apparent Diffusion Coefficient (ADC).

It is important to note that this measure is not sensitive to anisotropy, but will indicate

how much diffusion is occurring (an important indicator in many clinical applications,

Page 19: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

9

such as detecting edema). In white matter, anisotropy occurs predominantly due to an

inhibition of diffusion in the direction of eigenvectors ~v2 and ~v3, with limited or no change

in the direction of ~v1. As such, the ADC is generally larger in voxels where λ1 ≈ λ2 ≈ λ3

where diffusion becomes less restricted, which can be an indication of pathological change

in white matter [21, 22, 26, 30, 31, 32, 33].

λ‖ = λ1 (1.17)

λ⊥ =λ2 + λ3

2(1.18)

λ = ADC =λ1 + λ2 + λ3

3(1.19)

In healthy white matter, or other areas of restricted diffusion, one can use the eigenval-

ues to define the degree of anisotropy within each voxel, and thus within different regions

in the tissue. The primary method of calculating anisotropy within the tensor model

is by a metric called Fractional Anisotropy (FA). The FA is a value in the range from

zero to one, where an FA of zero represents a perfectly isotropic voxel (λ1 = λ2 = λ3),

and an FA of one represents a voxel where diffusion can only occur in one direction

(λ1 > 0, λ2 = λ3 = 0). FA is calculated with the following equation:

FA =

√1

2

(λ1 − λ2)2 + (λ1 − λ3)2 + (λ2 − λ3)2

(λ21 + λ2

2 + λ23)

(1.20)

In an idealized voxel of white matter, diffusion will occur preferentially in one direction

so that λ1 � λ2 ≈ λ3 – describing an ellipsoid with one long and two equivalent shorter

axes with FA . 1. Another important case is when λ1 ≈ λ2 ≈ λ3, which is found

generally in grey matter and CSF. In some cases, the isotropic diffusion may be indicative

of pathology such as a tumour [32]. The eigenvalue information thus provides a useful

tool to indicate the degree of directional-specific diffusion occurring through FA maps,

widely used both in research and the clinic.

Through the use of eigenvectors, these maps can be made more informative by defining

the principle direction of diffusion at each voxel – generally by assigning a colour to each

principle direction and mixing the colours at each voxel based on the principle direction of

Page 20: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

10

diffusion determined from ~v1. Combined, the directional information is given by colour,

and the intensity set by FA represents the preference for diffusion in that direction. This

representation can be used quite effectively to find brain tumours via delineation of white

matter surrounding them, and may be useful in surgical planning to determine how white

matter tracts have been distorted around tumour tissue [34, 35, 36, 37].

1.2.4 Higher Order Models of Diffusion MRI

Despite significant strengths of DTI, many regions of tissue are poorly described by el-

lipsoidal or isotropic diffusion. A classic problem case is two crossing fibres, such as

where white matter tracts cross one another. Here, diffusion in a single voxel contains

a mixture of unrestricted diffusion in two directions, and restricted diffusion in one, and

thus cannot be accurately described by a simple 3× 3 tensor. This leads to a relatively

small FA, even though there is significant anisotropy present at the microstructural level.

Moreover, the DTI framework does not model restrictions to diffusion per se, but ac-

counts for them empirically by allowing reduced diffusion in particular directions. This

motivated the development of alternative models that can describe more complicated

patterns of diffusion, with increased dependence on direction or degree of diffusion. This

in turn, depending on the model, requires acquisition of images for a larger set of b-values

and/or number of directions. Of course, in order to obtain these extra image contrasts,

additional scan time is necessary.

Diffusion Kurtosis Imaging (DKI)

Diffusion Kurtosis Imaging (DKI) advances the simple DTI model by characterizing the

deviation of diffusion from standard Gaussian diffusion, which is to be expected in regions

where membranes or other structures present barriers to diffusion. The kurtosis model

adds an extra term to the DTI model that is second order relative to the apparent

diffusion coefficient as shown in Equation (1.21).

Sk = S0 e−bk ADC + 16Kapp,k b2k (ADC)2 (1.21)

Page 21: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

11

Where Kapp represents the kurtosis factor and determines the “peakedness” of the Gaus-

sian bell-shape [38]. If the kurtosis is zero, then diffusion is Gaussian and can be described

completely by the simple DTI model. However, if the kurtosis value is greater than zero,

then the diffusion profile has a higher peak, with smaller tails where diffusion is re-

stricted. In order to obtain this information, more b-values are required, as there are 21

independent parameters in the kurtosis model in comparison to only 6 in the DTI model,

implying longer scan time.

High Angular Resolution Diffusion Imaging (HARDI)

High Angular Resolution Diffusion Imaging (HARDI), as the name implies, requires a

very large number of unique diffusion directions – at least 60 directions to have a deter-

mined set (compared to six for DTI). In practice, HARDI scans, like most other diffusion

MRI, uses 120 or more directions at one b-value to generate an overdetermined set (com-

pared to typically 30 directions and one b-value in DTI) [39, 40]. HARDI attempts to

overcome the basic assumption of ellipsoidal Gaussian diffusion within DTI, at the cost of

a large number of diffusion directions. However, through the use of a high angular resolu-

tion on the diffusion direction sphere, HARDI is able to create a more flexible depiction

of diffusion over all directions, and is thus able to handle cases such as crossing fibres [7].

Many different methods have been proposed to visualize and identify the distribution of

fibres from HARDI data. A popular approach is to fit the diffusion distribution with

spherical harmonic functions, but other related model free approaches also exist such as

Q-Ball Imaging (QBI) [39, 41]. From HARDI datasets, it is common to compute the ori-

entation distribution function (ODF) – which describes the orientation of white matter

tracts and diffusion within the voxel.

Neurite Orientation Dispersion and Density Imaging (NODDI)

Neurite Orientation Dispersion and Density Imaging (NODDI) is another higher order

diffusion MRI model that incorporates more parameters than DTI, and in some respects

combines aspects of DKI and HARDI, requiring both multiple b-values and a large num-

ber of diffusion directions. Namely, NODDI requires a minimum of two HARDI “shells”,

Page 22: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

12

where a “shell” is a set of diffusion directions at a specific b-value. The resulting image

intensities are modelled assuming three potential environments: intracellular (neurite)

space, extracellular space, and cerebro-spinal fluid (CSF) – with the underlying assump-

tion that each environment will inherently give a different diffusion MRI signal. The

signal behaviour from each environment is assumed to follow a particular model, and the

resulting image data is fit assuming a weighted summation of the models. The NODDI

model attempts to quantify the density of neurites within each voxel of the brain as well

as their orientation more accurately than DTI or HARDI are able to.

1.2.5 Clinical Uses of Diffusion MRI

Diffusion MRI is a tool that is integral to the diagnosis and staging of many diseases. Its

primary use clinically is for the diagnosis of stroke, but it can also be used to probe the

functional architecture of the brain – specifically white matter, as the highly myelinated

axons act as the “wiring” of the brain. Prime examples of diffusion MRI being useful

in the clinic would be cases of acute brain ischaemia [42], as the diffusion coefficient

decreases significantly within minutes of the ischaemic episode. Thus, one would see

a decrease in the MD as well in the FA of the brain region where the ischemia took

place in comparison to normal, while a standard T2-weighted scan is not diagnostic for

several hours [17, 43]. Similar effects are found in patients that have experienced strokes

[31, 44, 45, 46]. Diffusion-weighted MRI represents standard-of-care for diagnosis in these

cases.

Diffusion MRI has many additional clinical uses, such as measuring neuroanatomy

[47], stratification of brain tumours [48], prediction of cancer responses to chemoradia-

tion [49], and for visualizing effects of multiple sclerosis [50]. This is nowhere near an

exhaustive list, but shows the large range of pathologies that may be assessed based on

changes to water diffusion. The widespread use of diffusion MRI already promotes the

ease of implementation of new diffusion approaches, and motivates use of diffusion MRI

in research, both in human and small animal models.

It is common practice now within the clinic to utilize EPI as the imaging readout

for diffusion-weighted imaging as it is a very fast imaging technique. However, EPI is

Page 23: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

13

very sensitive to susceptibility and spatial distortion artifacts, which motivate the use

of alternative readouts in some applications. Other strategies to speed imaging include

parallel receive strategies of various methods [51].

1.3 The Mouse as a Model

The mouse is a convenient mammalian model with high homology to humans both struc-

turally and genetically [52]. In addition, the mouse has an accelerated lifespan, and there

are a large number of genetically-modified (and characterized) mouse strains. While us-

ing the mouse as a genetic model is not within the scope of this thesis, it motivates

the technical developments in this research. Moreover, it has been shown that 90% of

mouse mutants with altered behavioural phenotypes have accompanying changes in neu-

roanatomy [53], making MRI an efficient and quantitative surrogate for evaluating the

mouse brain. Although the mouse also has a number of limitations – only a 10% fraction

of white matter volume compared to 50% in humans and a complete lack of cortical

folding [54] –it remains one of the most widely used models for neuroscience.

1.3.1 MRI in the Mouse

Due to the significantly smaller size of the mouse, there are some differences between

mouse and human MRI. In order to obtain similar “anatomical” resolution, voxel dimen-

sions must be scaled approximately 15-fold [55] to linear sizes of 100µm or smaller, which

translates to obtaining higher spatial-frequency information in k-space. An important

consequence to this is a loss in signal-to-noise ratio (SNR) [1], which is often recovered

by higher magnetic fields, custom radiofrequency coils, and longer scan times – for ex-

ample through acquisition of multiple averages. This issue is exacerbated for diffusion

imaging where a whole scan series must be acquired. One approach employed at the

Mouse Imaging Centre to increase efficiency is to image multiple samples simultaneously

so that, although each scan may be long, many images are acquired. This approach is

referred to as multiple-mouse MRI [56], and is especially attractive for imaging of ex vivo

brain samples, a strategy largely unique to preclinical research.

Page 24: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

14

1.3.2 Diffusion MRI in the Mouse

Diffusion imaging for mouse data is often collected ex vivo due to the long scan times

required - it is not uncommon to have diffusion MRI scans taking 12 [57, 58] to 28

hours [59, 60]. Since the data is taken post mortem, there will be a general decrease

in the motion of water due to the loss of active transport of water within the brain

[61], but is still an excellent measure of relative diffusion within the sample. However,

even with the extended scan times and the post mortem acquisition, diffusion MRI in

the mouse provides an important contrast in ex vivo mouse research. It complements

laborious histological approaches – almost always restricted to select regions of interest

– by providing a whole-brain assessment to characterize white matter [62, 63] in models

of traumatic brain injury [64], stroke [65], or multiple sclerosis [66], as well as mapping

brain development and developmental pathologies through mouse models [67, 68, 69, 70].

It also, of course, provides a convenient link to a widely-used clinical tool, connecting

experimental work in the lab with observations available in patients.

The extended scan time of ex vivo mouse MRI, combined with the need for high-

resolution acquisitions of many different images, hampers the potential application of

methods such as DKI, HARDI, or NODDI in preclinical research. Finding a way to in-

crease the efficiency of these experiments would enable broad application for the scientific

community. This represents a classic trade-off in MRI, where spatial information must

be balanced against the number of images acquired to characterize another dimension

(such as contrast, time, frequency, or diffusion directions). With standard fully-sampled

image acquisitions, this trade-off is often unacceptable, requiring too low a resolution in

order to acquire the number of images needed in a given scan time. As an alternative,

I propose in this thesis an approach that improves the researcher’s ability to make this

trade-off by application of compressed sensing (CS), enabling a more flexible tuning of

the number of spatial samples and the number of diffusion directions.

Page 25: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

15

1.4 Compressed Sensing (CS)

The amount of data that is required for higher order diffusion imaging methods is

significantly greater than the simple DTI case. In mice, where scan times are generally

increased already, the demand for scan time quickly becomes prohibitive. For example,

in high resolution DTI scans of ex vivo brains at our centre, 30-direction scans last for

12 hours. Running 120 directions thus requires a continuous 48 hours! This is sufficiently

impractical that it cannot be routinely performed.

In the various diffusion approaches described here, there are up to five dimensions

that need to be sampled: 3 spatial dimensions, 1 b-value dimension, and the diffusion

directions dimension (which in itself, of course, includes three directions). In standard

diffusion scans, the spatial dimensions are fully sampled, while the b-value and diffu-

sion dimensions are sampled at a very low resolution – often only one b-value and 30

directions over the Euclidean half-sphere. Such an acquisition is greatly undersampled

in the diffusion dimensions for higher order models. To remedy this, one approach would

be to instead undersample in the spatial dimensions – decreasing the amount of time

required for the acquisition of each specific b-value and direction combination – which

can then be leveraged to sample the required number of b-value and/or direction images.

While this would normally imply that the spatial dimensions will have lower resolution,

CS is an approach to image reconstruction that employs constraints to compensate for

undersampled data.

1.4.1 Introduction to Compressed Sensing

The theory of CS was first introduced by Donoho in 2006 as a method to reconstruct

complete data from sparsely-sampled sets [71]. Suppose there exists some unknown vector

x that can be measured in Rm. In a fully sampled case, at least m measurements would

be required to estimate x. However, for CS, if x can be assumed to largely consist of

zeros (i.e. is sparse) in some space, it is possible to make only n � m measurements,

and to obtain an accurate rendition of x by enforcing sparsity in the result.

The mode of undersampling is important in CS, so it is important that sampling

Page 26: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

16

be planned to ensure certain conditions. Ideally, a method that acquires only from the

“important information” of the signal would be most efficient, but in practice a strategy

must be selected that works broadly for a given application. This requires establishing a

sampling space, the desired output space, and some sparse space (which can but need not

be identical to the output space). For MRI, all of the sampling occurs in k-space (also

denoted as “Fourier space”), and the desired output is in image space. To implement

the method, only a subset n of our m samples are collected in k-space such that n� m.

[71, 72, 73]. The CS algorithm has three major requirements [71, 72]:

a) The image is compressible (sparse) in some transform space;

b) Any artifacts due to the undersampling scheme are “noise-like” and do not

present as coherent aliasing artifact; and

c) The reconstruction method promotes sparsity in the selected space(s) and fi-

delity to acquired data.

The notion of having the data exist in a compressible space is readily met for the

MRI data that I am working with. Advances in compressing image data have been

occurring for years, specifically in the field of computing with respect to different ways of

compressing image files, such as the work done with JPEG-2000, which uses the wavelet

transform [74]. In wavelet space, most of the image can be described by a relatively

small number of high-intensity points, and the remaining points can largely be ignored.

There are a number of other transforms that define MRI data as sparse, and each of these

can be used to constrain the CS algorithm to exploit specific redundancies. A common

example is the Total Variation (TV). The TV is defined as an image of finite differences,

calculating a one-pixel gradient in image space [75]. The TV has significant intensities

only at tissue boundaries, and is close to zero elsewhere.

For MRI, the sampling in k-space is generally determined a priori, and for CS, must

be selected to ensure “noise-like” behaviour in image space. As mentioned earlier, n� m

samples will be acquired. Naively, one might undersample evenly across k-space; however,

in image space, this would cause an effective field of view (FOV) decrease, and cause

coherent aliasing due to sampling under the Shannon-Nyquist limit. The second naive

idea for undersampling, might be to take a low resolution image by only sampling the

Page 27: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

17

Figure 1.2: A representation of the sparsity generated through the use of the total variation and wavelettransforms on a Shepp-Logan phantom. It is important to note that most of the values in each of thetwo sparsifying transforms are zeros. The total variation is excellent at determining the locations ofedges within the sample, and the wavelet coefficients are often used for compression of data.

centre region of k-space covered by n� m points. This would avoid the aliasing problem

of the first approach, but is likely to result in too poor a spatial resolution. In order

for CS to enhance the image, farther regions of k-space need to be sampled to obtain

information about the edges in the image. Instead, the current standard choice for

undersampling utilizes a (pseudo)-random sampling pattern in k-space. Namely, first a

small fully-sampled centre of k-space is acquired. On one hand, this ensures that there is

a good (low-resolution) basis to initialize the CS algorithm and on the other, it provides

information necessary to estimate the phase of the MR image (which is inherent in the

data acquisition but not handled as part of the CS reconstruction). Subsequently, the rest

of the sampled points (n − ncentre) are distributed through the remainder of k-space on

the same grid as the fully sampled k-space. This approach samples the high frequency

regions that give edge information to the image without generating coherent aliasing

artifacts and results in image-space artifacts that appear as noise.

The sampling scheme in this work will be constrained to a Cartesian grid, however

this need not be the case as many different sampling trajectories are used in the MR field

such as radial imaging [76, 77, 78], or spiral imaging [79, 80].

The simplest form of the CS reconstruction itself for MRI data is based on a con-

Page 28: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

18

strained minimization described as follows [71, 72]:

minimize ||Ψx||1

s.t. ||Fux− y||2 < ε(1.22)

In this set of equations, x is our current guess for the reconstructed image (note that

later in Figure 2.7, our guess is in Wavelet space and is represented by ΨG); Ψ is a spar-

sifying transform, in this case the Wavelet Transform; Fu represents the “undersampled”

Fourier Transform (effectively applying a Fourier transform and only taking those points

acquired in undersampled k-space); y is the k-space data that was collected from the

scanner; and ε represents the tolerance for data consistency term. Finally, the ||x||n is

the `n norm of x, defined here as:

||x||p =

(n∑i=1

|xi|p) 1

p

for {p ∈ Z | p ≥ 1} (1.23)

||x||0 =n∑i=1

|xi|0 where 00 ≡ 0 (1.24)

Although the `0 “norm” in Equation (1.24) is in many ways the most intuitive mea-

sure of sparsity (i.e., counting the number of non-zero elements), it is computationally

impractical to work with. Instead, it is common to use the `1 norm, which has been

proven to be an acceptable solution to promote sparsity [81, 82, 83].

Equation (1.22) is a simple form of the CS equations and it is common to add addi-

tional terms to the minimization function, such as the TV term [72, 84].

minimize ||Ψx||1 + λ TV(x)

s.t. ||Fux− y||2 < ε(1.25)

The λ variable is used to tune the weight of the TV term relative to the wavelet sparsity

term. The equation is presented as such for the optimization:

argminx ||Fux− y||2 + λ1||Ψx||1 + λ2TV(x) (1.26)

Page 29: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

19

Where now, λ1 and λ2 are used to tune the weights of the wavelet and TV terms relative to

the data consistency term, respectively. By utilizing other terms within the optimization,

just like the TV term has been added, this algorithm can be adapted for different types

of redundancies in specific data sets. Some examples could be redundancies in the time-

dimension, longitudinal anatomical data, or diffusion directions, which is the focus of

this thesis.

1.4.2 Current Uses of Compressed Sensing in Human Research

CS works best when SNR is high, and therefore it is more easily applied in human scans

than in mouse scans. As such, there is already extensive use of CS in human research.

For example, 3D contrast-enhanced angiography, in which the image is naturally sparse

and temporal frames evolve smoothly, is a perfect candidate [72, 85], and represents one

of the areas where CS was arguably first applied.

Although there is some interesting investigations of CS in combination with multi-

channel parallelization, CS is not used frequently for strictly anatomical work. CS is

optimal in cases where additional dimensions of redundancy exist, such as in dynamic

imaging, even on a slice-by-slice basis when imaging the brain [86]. Research has also

investigated the possibility to obtain higher resolution images through undersampled

datasets for functional MRI (fMRI) using spiral acquisition, where the redundancy from

the temporal dimension is exploited [87]. Using CS in conjunction with Sensitivity Encod-

ing (SENSE) for Chemical Exchange Saturation Transfer (CEST), allows for the overall

time required to be decreased by a factor of four (R = 4) [88]. The Human Connectome

Project has also been looking into using CS for diffusion MRI within the human brain in

order to obtain high resolution scans and an accurate mapping of the connections within

the brain [89], specifically with respect to Diffusion Spectrum Imaging (DSI), another

diffusion imaging approach requiring both a large number of b-values and directions. In

addition to this, there has been a recent interest in the use of dictionary-based sparsity

constraints exploiting previously acquired data to improve reconstruction even for factors

of R ≥ 4 [90].

The expansion of CS research within the human sphere has been rapid, now cover-

Page 30: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

20

ing an immense number of potential applications. Specifically, in diffusion MRI, a CS

framework has been developed to estimate the degree of crossing fibres in the intra-voxel

environment within 30 direction DTI datasets [91]. In this experiment, CS is applied such

that mixing fractions are calculated based on the exploitation of the tensor directions

determined using DTI. This approach treats the diffusion direction dimensions as under-

sampled, and aims to discern information typically only available through a higher order

model like HARDI. Another experiment along these lines has attempted to use spherical

ridgelets to interpolate the data between diffusion directions to increase the resolution

of sampling in the diffusion directions dimension [92]. Each direction is fully sampled

and missing directions are inferred based on the expectation that each HARDI signal can

be expanded as a linear combination of spherical ridgelets. This model is effective, but

requires extremely high SNR to ensure the fits accurately reflect tissue properties, and

not image noise.

In diffusion MRI there are three fully sampled spatial dimensions, along with two

possible diffusion dimensions: diffusion directions and b-values, which are often sampled

at a much lower rate. However, in the current environment, where there is a significant

appetite for dense sampling in the diffusion direction or b-value dimensions, collecting a

large number of images with increasingly subtle intensity differences is required. An at-

tractive approach is to begin undersampling in the spatial dimensions in order to enable

increased sampling in the diffusion dimensions. For acquisitions based on Cartesian sam-

pling, it is convenient to undersample in only two dimensions because there is no benefit

gained from undersampling in the frequency encoding direction, which costs compara-

tively little time to acquire. Such an approach, in combination with a CS reconstruction,

could enable trade-offs in the amount of spatial versus diffusion information acquired, so

that more advanced diffusion models can be employed without significantly longer scan

time.

Page 31: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

21

1.5 Aim

The purpose of this study is to exploit redundancy found in diffusion MRI data in con-

junction with CS to enable undersampling of k-space and a flexible trade-off of spatial

and diffusion data during image acquisition. This strategy was tested through several

stages. First, the CS algorithm was implemented and tested with anatomical mouse MRI

at a sampling rate of 25% to confirm function of the algorithm and to evaluate its per-

formance in isolated anatomical scans without an extra dimension of redundancy. Next,

significant modifications of a pulse sequence used for ex vivo diffusion MRI were made

to allow prospective acquisition of data for CS reconstruction. The modifications were

critical to addressing artifacts affecting the CS algorithm. Finally, the algorithm was

tested prospectively by acquiring a four-fold undersampled scan with 120 diffusion direc-

tions scanned (four times more than the baseline 30 directions) evaluating the benefit of

trading off spatial for diffusion direction sampling.

Page 32: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Chapter 2

Methods

2.1 Adaptation of Compressed Sensing for Diffusion

Data

The optimization scheme used was based originally off of a series of Matlab scripts

written by Dr. Michael Lustig to accompany his 2007 paper – however, the custom scripts

have been significantly adapted and ported into Python (https://python.org) [93] using

the packages NumPy [94], SciPy [95], and PyWavelets [96]. Matplotlib [97] was used for

visualization. The code was written so that the user can choose any of the several

available optimization routines and is parallelized over many cores for each index in the

readout direction (or other selected dimension). All optimization routines available for

the ported code are found in the SciPy package, with a conjugate gradient being the

default, and used for all CS reconstructions in this thesis.

The goal of my work is to collect high-angular resolution diffusion data in ex vivo

mouse brain samples to further elucidate diffusion microstructure within the brain for

mouse models of human disease. Using the baseline 30-direction diffusion scan at a

resolution of 78 µm, that runs in 12 hours in conjunction with four-fold undersampling

in the spatial dimensions, the high-angular information can be obtained through a 25%

spatially sampled set of 120 directions. However, in this implementation, the b0 scans will

remain fully-sampled due to their importance in subsequent computation of the diffusion

22

Page 33: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

23

attenuation.

2.1.1 Total Variation in the Diffusion Dimension

The strategy I chose for enabling undersampling in the spatial dimensions was to enforce

similarity in nearby diffusion images as part of the CS constraint. There are multiple

approaches that could achieve this. However, as this strategy effectively seeks to mini-

mize variation in the diffusion direction dimension, it can be considered in many ways

analagous to the standard definition of TV in the spatial dimensions. Consequently, I

chose to expand the standard definition of TV to enable addition of the diffusion direc-

tion.

The design of a TV term for diffusion direction is complicated by the fact that each

diffusion direction is likely to have a large number of nearest neighbours. To account for

this, I assumed that over a small solid angle the intensity of the diffusion-weighted images

could be approximated as a linear function of vector direction. The gradient coefficients

of such an approximation represent a form of intensity gradient analagous to the spatial

TV terms. For any given direction, the directions considered “nearby” include those

that are nearly colinear, or those that are nearly 180◦ away. Using the image intensity

at each voxel and the difference in the diffusion vectors (i.e. creating a system where the

direction being analyzed is at the origin), a least squares fit can be used to determine

the intensity gradients in the diffusion direction space. Defined mathematically, for a

given direction r there are n closest directions [~dr,1, ~dr,2, ..., ~dr,n], where ~d represents the

diffusion direction for index n with respect to r. The local intensity is fit through a least

squares approach:

∆I =

Ir − Ir,1

...

Ir − Ir,n

(2.1)

Page 34: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

24

A =

1, ~dr − ~d′r,1

...

1, ~dr − ~d′r,n

(2.2)

Aβ = ∆I (2.3)

β = (ATA)−1AT∆I (2.4)

= [β0 βx βy βz]T (2.5)

TVdiff =√β2x + β2

y + β2z (2.6)

~d′r,n is used to select ±~dr,n that generates the positive value for ~dr · ~dr,n. In

this representation ∆I is an n × 1 column vector, A is an n × 4 matrix to account for

a first order fit in the three spatial axes as well as a constant term, which leads to our

fit, β, a 4 × 1 matrix that represents the equation of the plane that fits the intensity

variation across diffusion directions at each voxel. It is expected that there should be

only nominal intensity offset (β0 = 0) since the selected directions are close and the

intensity distribution is assumed not to have any sharp discontinuities, however, it allows

an offset at the origin so that, for instance, the extremum of an ellipsoid is not penalized.

For the purposes of the objective function and the gradients, the defined TVdiff is

combined with the standard definition of TV. I added an additional tunable scaling

factor λdiff that can be changed to scale the weighting of TVdiff relative to the spatial

TV. For all of the work in this thesis, λdiff was set to unity.

TV = TVspatial + λdiffTVdiff (2.7)

2.1.2 Mathematical Approximations and Definitions

`1 Norm

The most intuitive constraint to enforce sparsity is to count the number of non-zero

entries. This is referred to as an `0 norm. However the `0 norm is difficult to use due to

Page 35: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

25

its discrete nature, and difficulty in handling noise in data. Thus, it is convenient to use

the `1 norm [82]. As can be seen in Equation (1.23), at x = 0, the first derivative of the

`1 norm is discontinuous – following the curve of y = |x|. It is necessary to approximate

the `1 norm derivative because, often, the reconstruction is working in the realm where

x ≈ 0, at which point the derivative is ill-defined. In other applications of CS, the

approximation has been presented as [72]:

|x| ≈√x2 + µ (2.8)

d|x|dx≈ x√

x2 + µ(2.9)

Where µ is empirically chosen to limit the derivative near zero while providing a reason-

able approximation for abs(x). I chose an alternative approximation that easily tuned

and smooth through the origin:

|x| ≈ 1

aln [cosh(ax)] (2.10)

d|x|dx≈ tanh(ax) (2.11)

Where a is a user-tuned variable to effect the sharpness of the curve, as can be seen in

Figure 2.1. When a larger value of a is used, the curve is nearly identical to |x|. This

method allows for some tunability in the effects of very small values (where the curve is

smooth), and retains the benefit that it is continuous everywhere. For all calculations, a

value of a = 25 was used.

TV Kernel

The spatial TV terms were implemented as a convolution of a gradient kernel and the

image. This allows flexibility in how this term is applied in the algorithm, by allowing

for different weighting to surrounding pixels based on the kernel. This method can be

readily implemented for the standard TV method through the following kernels:

Page 36: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

26

Figure 2.1: (a) Shows a comparison between different mappings of 1a log(cosh(ax)) for increasing values

of a when compared to |x|. Note that for larger values of a the curve approximates |x|. (b) Shows acomparison between the derivatives of the above functions, represented by tanh(ax) and sgn(x).

TVx =

0 0 0

0 −1 1

0 0 0

TVy =

0 0 0

0 −1 0

0 1 0

(2.12)

In future implementations, it may be useful to apply a sparsity term with the third

dimension as well [75], which would require equivalent definition of TVz, or exploration of

alternative gradient kernels for edge detection – such as the Gabor filter – in the sparsity

constraint [98].

2.1.3 Phase in CS Reconstructions

Note that the CS reconstruction is defined using a real image, whereas MRI data is

complex with both magnitude and phase. One must therefore include a step in the CS

algorithm to account for phase prior to computations of sparsity in image space, but

reapply it for comparison with data collected in k-space. Indeed, the most important

initialization step for CS reconstruction is arguably the phase estimation. When accurate

Page 37: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

27

phase is used, reconstructions can be accurate with only 8.3% of the data in an idealized

phantom [72]. However, to have a perfect phase estimation, one requires a complete

image, which of course is contradictory to the notion of undersampling in the first place.

The primary method to obtain a phase estimation is thus to use a blurred, low-resolution

image attained by fully sampling a small region at the centre of k-space. While this

provides a good phase estimate in many cases, edge information is inevitably lost and

cannot be recovered in the reconstruction.

In the work presented here, I elected to map the phase based in large part on the fully

sampled b0 images. This approach, in principle, gives the highest resolution phase map

available. However, significant efforts were required to enable this strategy, owing to the

sensitivity of RARE to anomalous phase and the tendency of the diffusion gradients to

introduce anomalous phase. This is elaborated on in Section 2.3.3. One of the primary

areas for future research should be optimization of the phase estimation of undersampled

data, which would broaden the scope of CS applications.

2.2 Sampling

For all of the sampling schemes discussed within this paper, a cylindrical acquisition

scheme was used [55]. In this approach the corners outside a fixed radius in the plane

defined by the two phase-encode directions are not collected – thus only π/4 of the

full Cartesian grid is sampled – or a reduction of ∼22%. All references to percentage

sampling within this work are in reference to this baseline (i.e., 25% sampling indicates

0.25 × π/4 sampled points of the fully-sampled Cartesian grid). Only the two phase-

encode directions will be undersampled, as there is no benefit to undersampling in the

frequency-encode direction.

Generally, the sampling scheme in undersampled regions of k-space can be uniform or

variable density. If uniform undersampling is used, a probability density function (PDF)

is applied such that the probability of a point in k-space being sampled is equal for all

points (with the exception of the fully sampled centre). This sampling method represents

low- and high-frequency image information equally, however, it is also prone to artifacts

Page 38: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

28

from omission of low-frequency data and from a discontinuous PDF [72]. Variable density

undersampling on the other hand, focuses in on the fact that most of the signal can be

found closer to the centre of k-space, and attempts to acquire more of it at the expense

of information closer to the edge. As such, it applies a PDF that drops proportionally to

r−p from the centre of k-space, where p ≥ 1, generally bounded between 1 and 6. This

method of sampling has been shown to converge faster in the CS algorithm for cases

where no extra redundancy is present, such as in anatomical scans [72].

For the purposes of this thesis, all sampling schemes used will be at 25% sampling of a

fully sampled cylindrical acquisition. In addition, an alternative method of sampling will

be presented that was selected a priori in order to exploit redundancy across multiple

diffusion directions in the sampling in order to optimally sample k-space across diffusion

directions.

2.2.1 Choosing Directions for Diffusion Acquisition

Diffusion directions were selected based on a simple algorithm that treated each direction

that I wanted to probe as a pair of point charges constrained to the unit sphere – that

is to say, in spherical coordinates, points were placed at (ρ = 1, φ, θ) and (ρ = −1, φ, θ).

The use of two “charges” is motivated by the notion that in diffusion MRI, a positive

and negative diffusion gradient probe equivalent directions.

The points were initially placed on the sphere following a method that chose latitudes

at regular intervals, and then placed a varying number of points based on the size of the

cross sectional circle at that point of the sphere [99]. Using this starting point, the points

were moved on the sphere in order to minimize the following energy:

E =n−1∑i=1

n∑j=i+1

max(| ~ri ± ~rj |−1) (2.13)

A conjugate gradient system was used to solve this process with numerical calculation

of the gradient. Using this method, a set of equispaced directions on a unit sphere are

generated that can be used for uniform coverage of the unit sphere.

Page 39: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

29

2.2.2 Spatial Undersampling Across Diffusion Directions

It is convenient to have a good initial guess for the CS acquisition based on acquired

data. Since diffusion-induced intensity loss varies slowly with diffusion direction, k-space

from nearby directions can be used in place of unsampled data to generate an estimate

of the fully-sampled image. In order to utilize information between different diffusion

directions, I designed an undersampling scheme spread uniformly across directions, such

that a collection of nearest neighbour directions together had a nearly fully-sampled

image. To satisfy this requirement, a hybrid of variable density undersampling and

uniform density undersampling was used.

The sampling selection is performed once for an individual protocol with each diffusion

direction assigned its own unique sampling. The centre of k-space was fully sampled, and

then uniform density sampling was used around the periphery of k-space. In order to

ensure that there were no structured discontinuities in k-space at the border between

these regions, a “taper” region was included at the boundary. This can be seen in Figure

2.4(b) where three major regions exists:

1. Fully sampled centre

2. Variable density (“taper”) region

3. Spatially uniform undersampling constrained in distribution across the diffusion

directions

In order for neighbouring directions to have, when combined, a uniform sampling of

k-space, the sampling of neighbouring directions should be non-overlapping. Hence, a

second optimization method using simple electrostatics was implemented to minimize the

potential energy of groups of charges placed on the unit sphere at locations described by

the diffusion directions. One assumption in diffusion MRI is that diffusive motion runs

along both the positive and negative directions on a given axis, so that equal and opposite

diffusion directions are equivalent. For optimization, I therefore considered these points

equivalent. Consequently, the objective function was defined as follows:

Page 40: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

30

Figure 2.2: Probability density functions for CS undersampling: (a) Variable density undersampling, asis commonly used in CS reconstructions, shown both in 2D and a 1D line that is radially symmetric overthe sampling pattern. This method only has the fully sampled centre and the variable density region (b)The novel hybrid pattern with a central region which has both the fully sampled region and the variabledensity region exactly as before, as well as the external region where directionally biased sampling isused which is equivalent to uniform density sampling.

Page 41: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

31

obj =

q∑g=1

m−1∑i

m∑j=i+1

max(| ± ~rgi − ~rgj|−2) (2.14)

In this equation, g represents a group from the set of q total groups, with a total of

m members per group. The max() function accounts for the symmetry of diffusion along

the reference axis. For even a modest number of diffusion directions, this function has a

large number of local minima. A brute force method could be employed to find the global

minimum, however the number of different groups quickly becomes intractable because of

the binomial nature of combinations. Moreover, it is not essential that a global minimum

be identified for our purpose. As such, an alternative method was created and dubbed

“Team Picking”. The “Team Picking” algorithm is similar to children choosing teams

on the playground where the most favourable of the as yet unselected choices is made.

In this algorithm, directions are selected to fill q groups, where q is determined by

the number of directions (n) and the percentage of k-space to be sampled outside the

variable density sampled region, (p). Each group in k will have m or m+ 1 directions.

q = bp−1c (2.15)

m = bnpc (2.16)

Where b·c represents the floor function. The algorithm proceeded as follows: first start

with choosing a single direction – this becomes the first direction in group one. Then

the q − 1 directions closest to that direction are found. These will be the “captains”,

populating the first direction in groups 2 through q. With the remaining directions,

unique directions are added to each group in order to minimize the objective function

with each addition. Once one minimizing direction is found per group, they are added

to the corresponding “team” and used to calculate the objective function for all future

iterations. This process is then repeated until the teams are populated. By using this

algorithm, it is ensured that directions close to one another generally do not collect the

same data – and that over some small solid angle there exists a fully sampled k-space,

as can be seen in Figure 2.3.

Page 42: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

32

Figure 2.3: (a) Different “teams” of directions created using the team building method with p = 0.2and n = 30. Each direction is shown as a vector going from one end of the sphere to the other throughthe origin. Note that in this case, as this is only one example beginning with one direction, some ofthe members on a specific team (i.e. blue) are close together. (b) Representation of how k-space issampled, where white is the fully sampled region, grey represents variable density sampling and thecolours represent the data that will be collected by each team based on colour. Over a small solid angle,the entire region of k-space is sampled.

2.3 Implementation of ex vivo Diffusion RARE

2.3.1 Acquisition Parameters

To test the initial implementation of the CS algorithm independent of the added diffusion

component, high-resolution anatomical images were retrospectively undersampled and

reconstructed via the CS algorithm. This data was acquired using a 7.0 T Bruker MRI

(Bruker Corporation, Billerica, USA) with a 30 cm bore. A cylindrical k-space acquisition

[55] in a standard gradient-echo sequence was used with the following parameters: TR

26 ms, TE 8 ms, flip angle of 37◦, matrix size 334 × 294 × 294, with 2 averages, and 1

hour acquisition time. Twenty-four hours prior to imaging, the mice were injected with

50mg/kg MnCl2 to enhance contrast for T1-weighted imaging.

After confirming the algorithm appeared to be working, I established a sequence and

protocol for imaging ex vivo brain specimens. For this work, a multi-channel, 7.0 T MRI

magnet with a 40 cm diameter bore (Magnex Scientific, Oxford, UK) with an Agilent

Direct Drive console (Varian Inc., Paolo Alto, CA, USA) and Resonance Research Inc

(Billerica, MA) gradient system was used. Samples were placed in a custom 16-coil (4×4)

Page 43: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

33

solenoid array. In this configuration, the corner coils are not used for diffusion imaging

due to significant differences in gradient strength at the edge of the field [56, 100].

A diffusion imaging sequence was implemented for prospective acquisition of under-

sampled data. The diffusion sequence was based on a 3-dimensional RARE sequence,

which improves efficiency by collecting several k-space lines after a single excitation, us-

ing a train of refocusing pulses to repeatedly recover otherwise dephasing signal. For our

diffusion acquisitions, a 6 echo train was used with the centre of k-space acquired on the

first echo. A schematic of the pulse sequence used can be seen in Figure 1.2. The scan

parameters were TR 300 ms, TE 15 ms, TEeff 30 ms, 90 µm isotropic voxels, FOV of

1.62 × 1.62 × 2.91 cm, and a matrix size of 180 × 180 × 324. In the fully sampled 30

direction case, this scan takes approximately 12 hours to acquire, which implies that a

120 direction acquisition would take 48 hours. With the use of undersampling and CS,

the undersampled 120 direction case also takes approximately 12 hours to acquire.

As described, RARE provides improved efficiency by repeated refocusing pulses for

recovery of signal and acquisition of additional k-space lines. This has a few implications

that became integral to this work. First, each phase encode grid point in k-space must be

assigned to a particular repetition time and echo in the acquisition. Second, the phase of

the signal down the echo train must be carefully controlled and consistent. Echo-to-echo

inconsistencies in phase – either globally or linearly varying across the sample – lead to

prominent artifacts. Owing to the refocusing pulses in a RARE sequence, the erroneous

phase often separates between even and odd echoes. These are discussed further in the

next two sections.

2.3.2 Phase Encode Partitioning Scheme

Having selected a k-space acquisition scheme that omits the corners outside a given

radius, the remaining grid points in the PE1-PE2 plane must be divided across the (six)

echos of the echo train. The circular geometry lends itself to a partitioning based on

radius, with each echo separated into an annulus as can be seen in Figure 2.4(a). The

undersampling pattern described above must be combined in the context of this RARE

acquisition, resulting in a modified version of the PE1-PE2 sampling as a function of

Page 44: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

34

echo as shown in Figure 2.4(b).

Maintaining strict control of phase down the echo train is critical in the RARE pulse

sequence. One source of error is residual phase from phase-encode gradient pulses. These

may take the form of either a constant phase over the whole sample, or residual phase

gradients across the sample. Given good quality refocusing pulses, the former can be

corrected simply by measurement of the residual phase and correction in post-processing

prior to reconstruction. This is conducted as a standard part of our ex vivo RARE

acquisition [55], and was employed here as well. Residual phase gradient terms after

standard imaging pulses are sufficiently small that this effect can be neglected.

Unfortunately, the addition of diffusion gradients complicates things considerably.

The diffusion gradients are much larger than the imaging gradients, and can occur in any

orientation (determined by diffusion directions). Consequently, the potential for residual

phase and phase gradients is greatly increased. Indeed, our first implementation showed

significant artifact resulting from echo-to-echo inconsistencies in k-space acquisition. In

principle, induced phases that are uniform can be handled with the same retrospective

approach described for the phases induced by imaging gradients. Conversely, residual

gradient terms after the diffusion gradient are much more problematic because they have

the effect of shifting the effective position in k-space of the acquired data, so that it no

longer lies on a simple Cartesian grid. The most obvious way to handle this problem

might be to empirically tune gradient waveforms to nullify any residual gradient terms.

However, I found that this approach was not feasible over the whole gradient bore,

because some positions required different tuning than others. Thus, I instead designed

an approach to accommodate this problem as detailed in the next section.

2.3.3 Correction of Diffusion-Gradient Induced Phase

Interleaving ±~d Acquisitions: The “Checkerboard”

Imperfect behaviour of the magnetic field gradients is often attributed to eddy currents,

but can also be related to gradient hysteresis, concomitant fields or other unanticipated

behaviours. The resultant phase or phase gradients affect the subsequent acquisition,

Page 45: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

35

Figure 2.4: (a) Radially segmented splitting of echos shown in the fully sampled case to show how theannuli are split (b) A representative undersampled table using radially segmented k-space while stillcollecting the centre of k-space on the first echo. (c) Fully sampled example of how a radially segmentedphase encode table would be collected if the checkerboard pattern was used. The zoomed in region showshow each point alternates as to whether it is collected on ~d or -~d for each echo. (d) The undersampledversion. Note here that though the points are not all collected, the same checkerboard pattern is followed,where ~d and -~d are collected on the same mapping as in the previous figure.

Page 46: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

36

and is particularly problematic with a RARE readout, and became a major hurdle for

my implementation with ex vivo diffusion imaging.

In order to accommodate the problem of diffusion-gradient induced phase, I employed

both positive and negative diffusion gradients for each direction, in a pattern that alter-

nated in a “checkerboard” across k-space (see Figure 2.4(c) and (d)). Ideally, diffusion

in each direction should be identical, so this pattern should make no difference to the

resulting image. If, however, there are induced phase or gradient errors, these produce

a similar “checkerboard” pattern in k-space, which manifest in image space as ghosts in

the corners. This has two potential benefits. First, comparison of separate reconstruc-

tion of the “dark” and “light” checkerboards with the b0 reference allows a first-order

estimate and correction of these terms. Second, residual phase errors are more likely

to produce corner ghosts instead of characteristic ripple at edges in the image, a more

palatable ghosting pattern. In this way, the checkerboard sampling pattern allows for the

acquisition of data to correct for first-order problems induced by the diffusion gradients.

Diffusion-Gradient Induced Phase Corrections

Estimation and correction of diffusion-gradient induced phase was achieved by compar-

ison to the fully-sampled b0 reference scans. Correction of phase is applied in several

steps.

First, RARE phase corrections as described above are applied to both the b0 and b

data. The b0 data, which included 5 images acquired uniformly through the acquisition,

was averaged to create a high SNR reference, called b0. From b0, a mask is created

encompassing the brain tissue (by a simple threshold operation). This mask defines the

region of interest over which phase match will be evaluated.

Next, the TRs are split into the light and dark checkerboards (which correspond to

the positive and negative amplitude diffusion gradients) for both the b0 and the b data.

Since the b0 data is fully sampled, when it is split into the checkerboard pattern, the

image has perfect ghosts in the corner that can be joined to generate an exact replicate

of the brain in the centre. Within the defined mask, the phase of the b images is compared

to the respective b0 reference. The desired phase fit Φ is split into four parts:

Page 47: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

37

1. Constant Phase (φ0)

2. Frequency Encoding (Readout) Phase Gradient Term (φ′RO)

3. PE1 Phase Gradient Term (φ′PE1)

4. PE2 Phase Gradient Term (φ′PE2)

In this notation, φ′ is used to indicate a gradient. The constant phase is fit by

measuring the difference in overall phase for the regions of the images that fall within

the centre mask:

φ0 = ∠(∑

ej(∠Ib=0−∠Ib6=0))

(2.17)

The three phase ramps are all calculated in a similar manner to each other, where

the two datasets are projected onto the axis in question by summing across the other

two dimensions. The data is then zero padded to and Fourier transformed to determine

the distance between peak amplitudes, which maps directly to a phase ramp term. This

approach provided satisfactory results and was faster than other alternatives tested.

Subsequently, the derived coefficients are used either as a preprocessing correction (for

the constant and readout terms) or during the CS reconstruction (for the phase encode

gradient terms). In both cases, the corrections take into account the applied diffusion

gradient direction (i.e., light vs. dark checkerboard) and even vs odd echo.

The latter is necessary because each subsequent refocusing pulse reverses the phase,

and the applied correction must follow suit. Thus, each point in k-space is corrected

based on its (L)ight/(D)ark (i.e. ~d/-~d) and (E)ven/(O)dd echo assignment as follows:

1. Φ(L,O) = ΦL

2. Φ(L,E) = −ΦL

3. Φ(D,O) = ΦD

4. Φ(D,E) = −ΦD

Where Φx = [φ0 φ′RO φ′PE1 φ′PE2]TL/D. φ0 and φRO can be applied prior to CS because

the constant phase is applied element-by-element and the RO direction is fully-sampled.

Page 48: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

38

The two phase encode directions are undersampled, however, and the phase gradient

terms need to be included as part of the CS algorithm. It is therefore convenient to

define Φx = [φ′PE1 φ′PE2]TL/D, which will be used in Figures 2.5 and 2.7.

2.4 CS Reconstruction with ex vivo Diffusion RARE

The algorithm that the CS reconstruction follows is shown schematically in Figure 2.7.

The first step is to apply a FFT down kRO which is fully sampled. For convenient parallel

processing, the reconstruction is then performed on a slice-by-slice basis (taking “slice”

to indicate an index in the read direction), reducing the overall time required to complete

the reconstruction, and spreading the memory requirements. The latter of these points

becomes important as the number of directions increases.

2.4.1 Slice-to-Slice Correction

Introducing the slice-by-slice method of reconstruction is convenient for parallelization,

but can introduce slice-to-slice variability in overall intensity. To correct this, the final

volume of each direction undergoes intensity scaling in the frequency-encode direction

based on the intensity profile of the undersampled data from the scanner (i.e. FFT

reconstruction without the use of CS), as is shown in Figure 2.8.

2.5 Mice

Wild type CD1 mice at age P14 from the Centre for Phenogenomics (TCP) in-house

colonies were used for the in vivo anatomical CS experiments. For the diffusion biased

CS experiments, wild type C57/Blk6 mice were perfusion-fixed at age P14. These mice

were from a study of the Syngap1 gene (as provided by Dr. Gavin Rumbaugh). For all

animal experiments, approval was obtained from the Centre for Phenogenomics Animal

Care Committee.

Page 49: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

39

Figure 2.5: A flowchart representing the ±~d phase corrections required to correct diffusion-gradientinduced phase. This figures expands 1© in Figure 2.6. In this method, the constant phase and thegradient along the read direction (RO) are applied to generate modified k-space data (kaq), while thephase-encode phase gradient terms are fed to the CS algorithm.

Page 50: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

40

Figure 2.6: A simplified flowchart of how to acquire and correct the data for CS reconstruction. 1© canbe found in Figure 2.5, and 2© can be found in Figure 2.7.

Page 51: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

41

Figure 2.7: A simplified flowchart demonstrating the CS algorithm. This figures expands 2© in Figure 2.6.The diffusion specific addition for this project is captured within the left arm of the Total Variation. ΨG

represents the current guess (represented in wavelet space) that is being tested through the algorithm.

For each of the boxes, both the objective function f and its gradient ~∇f are calculated. For theData Consistency box, the ‘mask’ is the sampling mask used in acquisition, and determines which phaseΦ(B/W,E/O) is to be applied. The wavelet box is simple because the current image guess is representedin wavelet space. Note here that when ΨG is updated, a scalar c is applied to the gradient that representsthe line search term.

Page 52: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

42

Figure 2.8: Slice-to-slice intensity corrections. In this image, the readout direction is in the left-to-rightdirection. (a) One representative slice to show post-CS slice-to-slice differences. (b) Intensity profiledifferences used to scale each slice. Red is the intensity profile (summed over the two phase encodedirections) of the post-CS reconstruction and blue represents the pre-CS intensity profile. (c) The sameslice after intensity correction as per the profile correction.

Page 53: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Chapter 3

Results

3.1 CS Reconstruction – Proof of Principle

3.1.1 Retrospectively Undersampled Shepp-Logan Phantom

As a proof of principle, the CS implementation was tested using a digital phantom and

simulation of an undersampled (anatomical) acquisition. The digital phantom was gen-

erated from the Shepp-Logan phantom in Matlab (The MathWorks, Natick, USA) with

the addition of smoothly varying phase and complex noise. The data was retrospectively

undersampled from the complex dataset in k-space and run through the CS algorithm

(using the phase from the fully sampled case), as in [72]. The CS reconstruction reduced

the incoherent noise present from the zero-filled dataset, reconstructing even the three

small light-grey regions close to the bottom of the phantom as shown in the blue box in

Figure 3.1.

3.1.2 Retrospectively Undersampled Anatomical in vivo Data

A retrospective undersampling and CS reconstruction was used to assess the algorithm’s

performance in mouse data using in vivo data collected from the Bruker MRI. The data

was retrospectively undersampled using the variable density undersampling strategy at

25% sampling. If the CS-based acquisition had been acquired prospectively, it would

have taken approximately 7.5 minutes (as compared to 30 minutes for the fully-sampled

43

Page 54: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

44

Figure 3.1: Example of a Shepp-Logan phantom with 10% complex Gaussian noise added run throughthe CS algorithm at 25% sampling (R = 4). Parameters used: λ1 = λ2 = 0.05.

dataset). For the CS reconstruction shown below, one average of the fully-sampled data

was used to generate the phase map for the CS reconstruction. Without the use of this

phase, the reconstruction does not perform nearly as well. In this particular image, this

can be largely attributed to a large phase inhomogeneity at the upper left side of the

image, which appears darkened in the zero-filled data, and even shows minor artifact in

the fully-sampled image.

In the reconstruction, we can see that the post-CS data appears qualitatively better

than the zero-filled case owing to the removal of most of the incoherent sampling noise.

While the CS reconstruction on anatomical data alone does not recapitulate the anatom-

ical image, it provides an excellent quality with much of the detail in a fraction of the

time.

3.2 Diffusion MRI Data

After demonstrating that the algorithm was functioning in anatomical experiments, I

wanted to determine if CS could trade-off spatial sampling for an increased number of

diffusion directions. The scan that was used as the starting (and comparison) point was

is a 30 direction diffusion MRI scan with a b-value of 2000 s/mm2 and five b0 scans. The

total acquisition time is 12 hours for the fully sampled cylindrical acquisition of k-space.

Page 55: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

45

Figure 3.2: A test of CS reconstruction in an in vivo anatomical image of the mouse brain. At left, afully-sampled in vivo brain image is shown (inset at bottom). Undersampling (25% of the total data)results in the image in the middle column, where missing data is replaced with zeros. At right, the CSreconstruction is used to recover detail. Not the partial recovery of the subdural fat (blue arrow, toprow) and a blood vessel in the cortex (blue arrow, bottom row) through CS. Parameters used: a = 25,λ1 = λ2 = 0.05.

3.2.1 “Checkerboard” Correction of Diffusion-Gradient Induced

Phase

The diffusion gradients induce substantial residual phase. To test the “checkerboard”

approach – which uses interleaved acquisitions of positive and negative diffusion gradients

– for measuring and eliminating this phase, I first examined fully-sampled diffusion-

weighted images acquired in this fashion. An example is provided in Figure 3.3(a).

In this image, the phase inconsistencies between the ~di and −~di directions manifest as

ghosts in the corners of the image. In order to correct the phase differences between

the ~di and −~di acquisitions in the checkerboard data, the phase of each is mapped to

the corresponding b0 data (splitting of the b0 data by the same pattern). Application

of the obtained phase corrections then enables generation of a clean image where the

corner ghosts are minimized. Successful application of this method provides an approach

for minimizing the impact of diffusion-gradient induced phase errors and, where present,

manifests them as corner ghosts as opposed to edge ripple artifacts.

Page 56: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

46

Figure 3.3: Diffusion specific phase corrections from ±~d checkerboard data. (a) A diffusion (b > 0)dataset after RARE corrections applied. RARE corrections are not sufficient to remove the ghostingin the corner caused by the ±~d differences. (b) The top and bottom row are the ±~d data split fromeach other, respectively. The column on the left is the b data, and the column on the right is the b0data. The phase from the b data is fit to the respective b0 phase on the right. (c) After phase correctionas described in Section 2.3.3 the corner ghosts are substantially reduced, and the phase of the b datamatches the b0 phase.

Page 57: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

47

3.2.2 Compressed Sensing Reconstruction

With the CS algorithm demonstrated and a new approach for phase correction imple-

mented, I prospectively acquired a 120-direction diffusion-weighted data set with four-fold

undersampling. The sampling scheme is described in Section 2.2.2. In the same acqui-

sition time as the fully-sampled standard, the CS image acquisition and reconstruction

is able to obtain 120 diffusion directions acquired at the same b value in addition to the

five fully-sampled b0 scans. Thus by implementing this method, there is four times the

amount of diffusion direction data to work with, within the same acquisition time.

Figures 3.4 and 3.5 are representative examples providing comparison of individual

directions between fully-sampled images and the undersampled image acquisitions. The

latter are show both pre- and post-CS-reconstruction. For each CS reconstruction, the

tuning factors λ1 and λ2 were set to a value of 0.05 relative to the data consistency term.

Post CS-reconstruction, it can be seen that on an individual diffusion direction level,

at 25% sampling, there is some reduction in image quality compared to a fully-sampled

acquisition. For the duration of this work, I will be using the correlation as a measure of

image quality when comparing the fully-sampled, zero-filled, and post-CS datasets. As

a point of reference, the correlation between b0 data was 98.6% ± 0.1% across the five

b0 scans, where the error represents the range of data about the mean. The correlations

were measured over a mask covering only the brain in each image.

When applying undersampling and CS, the reduced quality is perhaps most evident

in the cerebellum, where fine details of foliation remain somewhat blurred in the CS

reconstruction. However, the CS-reconstruction provides significant improvement over

a zero-filled reconstruction. This may be most obvious in the corpus collosum. The

correlation across volumes for the pre- and post-CS data were 82.1% ± 5.1% and 87.4%

± 3.9% respectively.

The comparison in Figure 3.4 is biased, in the sense that the fully-sampled image in (b)

took four times longer to acquire than the CS reconstruction in (d). A comparison based

on four nearby directions from the undersampled acquisition to the fully-sampled result

was therefore also performed. This comparison is provided in Figure 3.5, which shows

Page 58: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

48

Figure 3.4: Comparison of fully-sampled and undersampled image reconstructions for an individualdiffusion direction. (a) A schematic display an outline of the brain in three orthogonal orientations withslices indicated the location of planes in (b) through (d). Double-headed arrows indicate the diffusiondirection probed in that plane. (b) A fully-sampled image (from a 30-direction DTI acquisition) shows

the baseline image result (note that this is without the use of the ±~d checkerboard acquisition scheme).

(c) Undersampled data with ±~d checkerboard acquisition scheme before undergoing CS reconstructionand in (d) after CS reconstruction. cer. represents the cerebellum and cc represents the corpus callosum

Page 59: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

49

the four individual undersampled images (a-d) and their average (e). When evaluated

in this way, the undersampled CS reconstruction performs very similarly to the fully-

sampled acquisition. Although some loss of detail remains (again, see the cerebellum),

the quality is very comparable, yielding a correlation of 94.2% ± 2.5%. The benefit of

the undersampling data begins to be evident in this figure, where the additional data

from multiple undersampled images (a-d) provides a basis for signal intensity analyses

that cannot be performed with one image. This increased diffusion direction resolution

is the basis for evaluating more subtle differences in intensity as a function of diffusion

direction.

In order to compare the 120-direction images with the 30-direction data for DTI

analysis, I computed FA maps across the entire brain for both cases. In Figure 3.6,

the FA maps are shown. Here it can be seen that the FA between the two datasets

exhibits anatomically very similar features. Interestingly, there is a general decrease in

the FA in grey matter in the 120-direction case, with a reduction from values around

0.4 using the original method, to 0.25 in the new one. In fact, the reduced value is

more consistent with expectations for grey matter. The decrease in FA is further shown

in a histogram comparing the FA between the fully sampled data and the CS data in

Figure 3.6(c). I suspect that this drop is largely due to the phase corrections enabled by

the checkerboard acquisition, which corrects diffusion-gradient induced phase errors and

provides a substantial reduction in artifact levels in specific directions.

3.3 Potential Use for HARDI Analysis

The real benefit of the CS undersampled acquisitions is the much increased diffusion

direction information. This is exemplified in Figure 3.7, where an individual slice from

all directions is shown, adjacent to a schematic showing all directions sampled for both the

fully-sampled (a) and undersampled (b) image series. The substantial increase in number

of directions enables more complicated mapping of the directional dependence of diffusion

than with DTI. As a proof-of-principle, a simple HARDI analysis using a Q-ball method

has been performed using the same data presented in Figure 3.7. The initial results of

Page 60: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

50

Figure 3.5: Comparison of fully-sampled and undersampled image acquisitions with matched acquisitiontimes. For each of these images, the arrows on top of each column represent the direction of the diffusiongradient in that plane. (a)-(d) Representative CS-reconstructed images acquired with 25% sampling areshown for the four closest directions to the fully-sampled direction in (f). (e) The average of images(a)-(d) provides a representative comparison to the fully sampled case. cer. represents the cerebellumand cc represents the corpus callosum

Page 61: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

51

Figure 3.6: Comparison of FA maps in (a) 30-direction fully-sampled and (b) 120-direction undersampledimage acquisitions are shown. Anatomical features are well-preserved between the two images, and mayeven be enhanced in some features of (b). An overall decrease in the FA of gray matter in the 120-direction case is attributed to improved handling of artifacts by the “checkerboard” phase correctionscheme and by the larger number of directions involved in the fit. (c) Comparison of FA values in 30-direction fully-sampled (FS) and 120-direction CS acquisitions are shown. A stark decrease can be seenin the FA values for the 120-direction case, which is likely due to spurious acquisition in the 30-directiondata which is corrected through both the higher angular resolution and the checkerboard acquisitionscheme.

Page 62: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

52

Figure 3.7: In the left column, the diffusion directions are plotted as vectors going through the originand touching the unit sphere on either side, representing the resolution in the dimension of diffusiondirections. In the right column, a representative slice is taken for each direction to visualize the sizes ofthe two datasets between (a) the original 30-direction method and (b) the novel method that incorporatesspatial undersampling and CS.

Page 63: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

53

this analysis are shown in Figure 3.8, which highlights the corpus callosum (a) and the

anterior commisure (b). With the anterior commissure and corpus callosum the overlaid

representations of the diffusion exhibit a much more elongated profile, produced primarily

by reduction in the perpendicular directions. Interestingly, many voxels (particularly at

boundaries of white matter) exhibit patterns that cannot be well described by a DTI

model. The underlying cause of these changes would interesting to study further. It has

been shown previously that approximately 90% of white matter has some level of crossing

fibers [101], and and this approach provides an opportunity to investigate such features.

While I have focused on a HARDI analysis at first, implementation for other models of

diffusion is possible as well.

Page 64: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

54

Figure 3.8: HARDI analysis of 120-direction undersampled diffusion MRI data. Anatomical imagesshowing the relative location of the enlarged ROIs are presented in the left column, with the rightcolumn showing zoomed regions with an HARDI analysis through an overlay of the ODF as calculatedby QBI (using DiPy [102]) represented as a deformed sphere based on the spherical harmonics on eachvoxel. Likelihood of diffusion in a direction is presented both by sphere deformation and colour, wherered implies strong diffusivity and blue represent low diffusivity. The specific regions that are probed are(a) corpus collosum (cc) and the anterior commissure (ac), and (b) anterior commissure (ac). Such ananalysis cannot be performed using the 30-direction fully-sampled acquisition.

Page 65: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Chapter 4

Discussion

In this thesis, a method was developed to permit a trade-off of spatial and diffusion

direction resolution in diffusion-based MRI analyses. The approach was further designed

to be acquired using a self-calibrating sequence by accounting for gradient imperfections

that induce spurious phase. The image data was then reconstructed using a model

that emphasized data similarity between diffusion directions close together on the unit

sphere. This enabled a large number of directions to be acquired, providing the potential

for accelerated acquisition of data sets appropriate for HARDI analysis. In this chapter,

I discuss the features and limitations of the proposed approach.

4.1 Phase Corrections

RARE acquisitions are notoriously artifact prone due to their extreme sensitivity to phase

errors through the repeated use of refocusing pulses down the echo train. As a result, the

use of this pulse sequence adds a layer of difficulty for the CS implementation, especially

since a high-quality phase estimation is critical for accurate CS reconstruction. I handled

this phase problem with a “checkerboard” acquisition strategy, in which the symmetry of

diffusion and properties of the Fourier transform were exploited to move artifact to the

corners of the image. Comparison to the b0 phase maps allowed for mapping of all images

to a common phase, which proved essential for the reconstruction. In future, extension

of the proposed work to other types of pulse sequences is likely to be less phase sensitive.

55

Page 66: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

56

4.1.1 ~d and -~d “Checkerboard” Acquisition Scheme

The utilization of both ~d and -~d diffusion directions for imaging has interesting appli-

cations for future use. Currently, most diffusion imaging protocols will only use ~d, and

assume that ~d and -~d are identical – as they should be based on the underlying physical

explanation if diffusion imaging. However, this assumption may fail if scan imperfections

cause other intensity alterations. By acquiring both directions, in the best case scenario,

no differences will be found and the image will be identical to the original method. On

the other hand, the mixing of both provides the opportunity to recognize and correct

erroneous phase, producing a cleaner data set for diffusion modelling. In this study this

method of sampling proved critical.

While this is largely attributable to our use of a RARE readout, the checkerboard

acquisition scheme is not limited to working with RARE acquisitions, nor is it limited

to sampling Cartesian k-space. In theory, all this method requires is a similar number

of ~d and -~d samples such that the phase of each can be defined. Here, the checkerboard

pattern was used because inconsistencies between the datasets were easily identified in

the corners of the image. However, one could imagine an experiment where both ~d and

-~d are probed, but with a spiral trajectory through k-space. On each TR, the sign of

~d could change, but the aliased inconsistencies would spread in a less coherent fashion

around the periphery of the image. A more likely application would be in multi-shot

EPI, with each shot could alternating sign on ~d, and therefore give a mixed dataset that

could help elucidate the differences present between ~d and -~d that should be removed

before the use of any models.

4.2 Limitations of the Study

4.2.1 Phase Measurements

In pursuing my thesis research, it became clear to me that the most important part of

the CS reconstruction for MRI is the phase that is used to move from real to complex

space (and back again). This represents the critical link between image space and the

Page 67: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

57

acquired data within the reconstruction. For many proof-of-principle demonstrations

of CS, fully-sampled data is retrospectively undersampled and the phase is taken as a

known quantity [72]. However obtaining an accurate phase for the CS algorithm is more

difficult given undersampled data. For this reason, CS may work best when samples

can be meticulously shimmed and employed with sequences relatively robust to phase

errors. Unfortunately, the RARE sequence employed here is particularly sensitive to

phase-induced artifact. As a means of compensation here, the redundancy of diffusion

MRI, fully sampled b0 images, and the checkerboard pattern of the diffusion directions

were employed to adapt or accommodate phases in each image.

4.2.2 Trade-Offs

Adequate resolution and sampling in all four (or five) of the dimensions in a diffusion MRI

experiment are critical to accurate diffusion modelling and delineation of the structures

of interest. Too great an undersampling in any one of the dimensions will lead to unsatis-

factory image results. In the mouse however, given the extensive length of time required

to image ex vivo brain samples at high resolution, this trade off of spatial information

through undersampling for an increased number of diffusion directions enables advanced

modelling techniques available such as HARDI. This trade-off will inevitably result in

poorer apparent spatial resolution, to the benefit of a much richer mapping of diffusion

direction space. The use of CS in this research made this trade-off feasible. Additional

research will be required in a cohort of mouse models with representative pathologies

in order to further optimize the balance between the two. The selection of 25% spatial

sampling in this research was empirically selected based on the appearance of structure

in wildtype mice only, and did not take into account the appearance of pathology.

4.2.3 Empirical Tuning of the Weighting Factors (λi)

Tuning of the different λ weighting values for any optimization is important, especially in

nonlinear optimization, as a small change in input can lead to a large change in output.

Since λs are empirically chosen, they may need to be retuned depending on the contrast,

Page 68: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

58

undersampling, or potentially the SNR of the dataset in use. Hence, the search space

has the potential to be quite large and may require a significant amount of retuning after

adjustment of the application, imaging protocol, or type of input data.

4.3 Future Work

4.3.1 Further Applications in Diffusion

In this body of work, the CS implementation focused on increasing the number of direc-

tions to utilize a more advanced model of diffusion. However, it would be particularly

interesting to consider alternative models of diffusion where the use of only one b-value

is not sufficient. In principle, a very similar approach could be used with a few b-values

with fewer diffusion directions for each. For example, if the same undersampling method

was used, in theory there could be four shells with 30 directions each. However, as the

directions are further apart in this case, the diffusion TV term proposed in this work

is forced to consider a broader solid angle (or be reduced to a simple anatomical CS

reconstruction). The implemented CS algorithm and TVdiff constraint are able to handle

this trade-off as well, though it might be of interest to explore the potential of using

constraints across not only diffusion directions but across different b-values as well.

The current method of increasing the number of diffusion directions, while decreasing

the acquired spatial data for each direction could be taken even further. On one side, one

could imagine trying to acquire fewer directions with greater spatial sampling in each,

for example, 90 directions at 33% sampling each. This would allow each independent

reconstruction to be more reliant on the data collected for that specific direction and less

dependent on sharing information between directions. By constraining the comparison

with a user-defined solid angle, there is the potential to adapt this constraint so that the

CS algorithm is more or less dependent on the comparison across diffusion directions. At

the other extreme, one could imagine taking a continuum of directions with very little

spatial data collected for each one. This would require modifications to the CS algorithm

used here since each individual direction would have too little data to be used in any way

Page 69: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

59

on its own. However, such an approach might have the advantage of being user tunable

retrospectively, letting directions be combined or divided according to the analysis needs.

While appealing, this strategy would likely be extremely dependent on accurate handling

of phase, especially since accurate representations of phase estimates require a minimum

of data with a fixed diffusion weighting. Thus in order for this to work, an algorithm

would need to be put in place to estimate the phase correction(s) required for each

direction based on the directions around it. In my experience, the sensitivity of CS to

phase means that this algorithm would need to be exceptionally robust. Advantageously,

the phase corrections required generally vary smoothly with gradient amplitude, so such

an approach is not out of the question. The range of possibilities are even broader if one

adds a wide range of b-values to the continuum of sampled directions

4.3.2 Potential Use in Clinic

Within the clinic, 2D imaging techniques are generally used with trajectories such as

EPI for an image readout, which is very different from 3D imaging with RARE used in

the ex vivo mouse MRI experiments presented here. Even though the faster 2D imaging

methods are in use, HARDI acquisitions still require long scan times, and thus are not in

broad use [103]. However, if the acquisition and reconstruction scheme discussed in this

thesis could be adapted to use a different trajectory, there is potential for it to accelerate

other scans as well. If this avenue is explored, optimization of the sampling scheme, and

the phase estimation are likely to remain key factors in success. Lustig emphasized as

much when he said “randomness is too important to be left to chance” [72].

Page 70: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Chapter 5

Conclusion

This study focused on the potential of using CS for diffusion MRI in the mouse by making

a trade-off between spatial sampling and the number of diffusion directions acquired. This

will enable the use of more sophisticated models of diffusion to characterize mouse models

of human disease. The implementation was based on the idea that nearby diffusion

directions have very similar intensities and may thereby serve as a constraint in the

CS-based reconstruction.

Numerous challenges were encountered in implementation, exclusively related to var-

ious kinds of induced image phase. A novel method where both ±~d were probed for a

given diffusion direction was implemented to accommodate and correct induced phase

errors. While this method may require further testing for use in other applications, the

CS reconstruction after phase correction was found to produce satisfactory image quality

in the spatial dimensions, and a significantly accelerated acquisition of diffusion direction

information. I demonstrated the approach by collection of 120-direction diffusion data

with 4-fold undersampling, for comparison to fully-sampled 30-direction data. The new

approach largely preserved anatomical detail while enabling a 4-fold increase in sampling

of diffusion directions. The new method further enabled the use of higher order models,

impossible for the data with fewer directions. Expansion of this approach will enable use

NODDI, or DKI framework where more than one b-value is also used. As the use of MRI

in preclinical applications continues to expand, the demand for quantitative measures

and an evaluation of their potential to reveal microstructural changes motivates an ex-

60

Page 71: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

61

panded evaluation of how spatial, diffusion, and indeed other contrasts may be combined

to maximize information content per unit scan time. CS and model-based analyses have

the potential to play and important role in the forward progress of this work.

Page 72: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

Bibliography

[1] Dwight G. Nishimura. Principles of Magnetic Resonance Imaging. Lulu.com, 2010.

[2] Robia G Pautler and Alan P Koretsky. Tracing odor-induced activation in the

olfactory bulbs of mice using manganese-enhanced magnetic resonance imaging.

Neuroimage, 16(2):441–448, 2002.

[3] Frank Wiesmann, Michael Szimtenings, Alex Frydrychowicz, Ralf Illinger, Andreas

Hunecke, Eberhard Rommel, Stefan Neubauer, and Axel Haase. High-resolution

MRI with cardiac and respiratory gating allows for accurate in vivo atherosclerotic

plaque visualization in the murine aortic arch. Magnetic Resonance in Medicine,

50(1):69–74, 2003.

[4] Jiangyang Zhang, Linda J Richards, Paul Yarowsky, Hao Huang, Peter CM van Zijl,

and Susumu Mori. Three-dimensional anatomical characterization of the developing

mouse brain by diffusion tensor microimaging. Neuroimage, 20(3):1639–1648, 2003.

[5] Rex A Moats, Sendhil Velan-Mullan, Russell Jacobs, Ignacio Gonzalez-Gomez,

David J Dubowitz, Takashi Taga, Vazgen Khankaldyyan, Linda Schultz, Scott

Fraser, Marvin D Nelson, et al. Micro-MRI at 11.7 T of a murine brain tumor model

using delayed contrast enhancement. Molecular imaging, 2(3):15353500200303112,

2003.

[6] Albert Einstein. Investigations on the theory of brownian motion. reprint of the

1st english edition (1926), 1956.

62

Page 73: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

63

[7] Denis Le Bihan, Cyril Poupon, Alexis Amadon, and Franck Lethimonnier. Artifacts

and pitfalls in diffusion MRI. Journal of magnetic resonance imaging, 24(3):478–

488, 2006.

[8] Jose Soares, Paulo Marques, Victor Alves, and Nuno Sousa. A hitchhiker’s guide

to diffusion tensor imaging. Frontiers in neuroscience, 7:31, 2013.

[9] Denis Le Bihan and E Breton. Imagerie de diffusion in-vivo par resonance

magnetique nucleaire. Comptes-Rendus de l’Academie des Sciences, 93(5):27–34,

1985.

[10] Klaus-Dietmar Merboldt, Wolfgang Hanicke, and Jens Frahm. Self-diffusion

NMR imaging using stimulated echoes. Journal of Magnetic Resonance (1969),

64(3):479–486, 1985.

[11] Denis Le Bihan, Eric Breton, Denis Lallemand, Philippe Grenier, Emmanuel Caba-

nis, and Maurice Laval-Jeantet. MR imaging of intravoxel incoherent motions: ap-

plication to diffusion and perfusion in neurologic disorders. Radiology, 161(2):401–

407, 1986.

[12] DG Taylor and MC Bushell. The spatial mapping of translational diffusion coeffi-

cients by the NMR imaging technique. Physics in medicine and biology, 30(4):345,

1985.

[13] Susumu Mori, Barbara J Crain, VP Chacko, and Peter Van Zijl. Three-dimensional

tracking of axonal projections in the brain by magnetic resonance imaging. Annals

of neurology, 45(2):265–269, 1999.

[14] Mark A Horsfield and Derek K Jones. Applications of diffusion-weighted and dif-

fusion tensor MRI to white matter diseases–a review. NMR in Biomedicine, 15(7-

8):570–577, 2002.

[15] NF Lori, E Akbudak, JS Shimony, TS Cull, AZ Snyder, RK Guillory, and TE Con-

turo. Diffusion tensor fiber tracking of human brain connectivity: aquisition meth-

Page 74: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

64

ods, reliability analysis and biological results. NMR in Biomedicine, 15(7-8):494–

515, 2002.

[16] Susumu Mori and Peter van Zijl. Fiber tracking: principles and strategies–a tech-

nical review. NMR in Biomedicine, 15(7-8):468–480, 2002.

[17] Denis Le Bihan. Looking into the functional architecture of the brain with diffusion

MRI. Nature Reviews Neuroscience, 4(6):469–480, 2003.

[18] EO Stejskal. Use of spin echoes in a pulsed magnetic-field gradient to study

anisotropic, restricted diffusion and flow. The Journal of Chemical Physics,

43(10):3597–3603, 1965.

[19] Christian Beaulieu. The basis of anisotropic water diffusion in the nervous system–a

technical review. NMR in Biomedicine, 15(7-8):435–455, 2002.

[20] Joseph V Hajnal, Mark Doran, Alasdair S Hall, Alan G Collins, Angela Oa-

tridge, Jacqueline M Pennock, Ian R Young, and Graeme M Bydder. MR imaging

of anisotropically restricted diffusion of water in the nervous system: technical,

anatomic, and pathologic considerations. Journal of computer assisted tomogra-

phy, 15(1):1–18, 1991.

[21] Peter J Basser, James Mattiello, and Denis LeBihan. Estimation of the effective

self-diffusion tensor from the NMR spin echo. Journal of Magnetic Resonance,

Series B, 103(3):247–254, 1994.

[22] Peter J Basser, James Mattiello, and Denis LeBihan. MR diffusion tensor spec-

troscopy and imaging. Biophysical journal, 66(1):259–267, 1994.

[23] Adolf Pfefferbaum, Edith V Sullivan, Maj Hedehus, Kelvin O Lim, Elfar Adalsteins-

son, and Michael Moseley. Age-related decline in brain white matter anisotropy

measured with spatially corrected echo-planar diffusion tensor imaging. Magnetic

resonance in medicine, 44(2):259–268, 2000.

Page 75: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

65

[24] Roland Bammer, Martin Auer, Stephen L Keeling, Michael Augustin, Lara A

Stables, Rupert W Prokesch, Rudolf Stollberger, Michael E Moseley, and Franz

Fazekas. Diffusion tensor imaging using single-shot SENSE-EPI. Magnetic Reso-

nance in Medicine, 48(1):128–136, 2002.

[25] Andrew L Alexander, Jee Eun Lee, Mariana Lazar, and Aaron S Field. Diffusion

tensor imaging of the brain. Neurotherapeutics, 4(3):316–329, 2007.

[26] Carlo Pierpaoli and Peter J Basser. Toward a quantitative assessment of diffusion

anisotropy. Magnetic resonance in Medicine, 36(6):893–906, 1996.

[27] Thomas L Chenevert, James A Brunberg, and James G Pipe. Anisotropic diffusion

in human white matter: demonstration with MR techniques in vivo. Radiology,

177(2):401–405, 1990.

[28] Michael E Moseley, Yoram Cohen, J Kucharczyk, J Mintorovitch, HS Asgari,

MF Wendland, J Tsuruda, and D Norman. Diffusion-weighted MR imaging of

anisotropic water diffusion in cat central nervous system. Radiology, 176(2):439–

445, 1990.

[29] Robert Turner, Denis Le Bihan, Joseph Maier, Robert Vavrek, L Kyle Hedges,

and James Pekar. Echo-planar imaging of intravoxel incoherent motion. Radiology,

177(2):407–414, 1990.

[30] J Peter Basser. Microstructural and physiological features of tissues elucidated by

quantitative-diffusion-tensor MRI. J Magn Reson B, 111(3):209–219, 1996.

[31] Peter J Basser and Derek K Jones. Diffusion-tensor MRI: theory, experimental

design and data analysis–a technical review. NMR in Biomedicine, 15(7-8):456–

467, 2002.

[32] Yaniv Assaf and Ofer Pasternak. Diffusion tensor imaging (DTI)-based white mat-

ter mapping in brain research: A review. Journal of Molecular Neuroscience,

34(1):51–61, Jan 2008.

Page 76: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

66

[33] Peter J Basser and Carlo Pierpaoli. Microstructural and physiological features

of tissues elucidated by quantitative-diffusion-tensor MRI. Journal of magnetic

resonance, 213(2):560–570, 2011.

[34] Saurabh Sinha, Mark E Bastin, Ian R Whittle, and Joanna M Wardlaw. Diffusion

tensor MR imaging of high-grade cerebral gliomas. American Journal of Neurora-

diology, 23(4):520–527, 2002.

[35] Brian P Witwer, Roham Moftakhar, Khader M Hasan, Praveen Deshmukh, Vic-

tor Haughton, Aaron Field, Konstantinos Arfanakis, Jane Noyes, Chad H Moritz,

M Elizabeth Meyerand, et al. Diffusion-tensor imaging of white matter tracts in

patients with cerebral neoplasm. Journal of neurosurgery, 97(3):568–575, 2002.

[36] Yaniv Assaf, Pazit Pianka, Pia Rotshtein, Michal Sigal, and Talma Hendler. Devi-

ation of fiber tracts in the vicinity of brain lesions: evaluation by diffusion tensor

imaging. Israel journal of chemistry, 43(1-2):155–163, 2003.

[37] Chris A Clark, Thomas R Barrick, Mary M Murphy, and B Anthony Bell. White

matter fiber tracking in patients with space-occupying lesions of the brain: a new

technique for neurosurgical planning? Neuroimage, 20(3):1601–1608, 2003.

[38] Andrew J Steven, Jiachen Zhuo, and Elias R Melhem. Diffusion kurtosis imaging:

an emerging technique for evaluating the microstructural environment of the brain.

American journal of roentgenology, 202(1):W26–W33, 2014.

[39] Christopher P Hess, Pratik Mukherjee, Eric T Han, Duan Xu, and Daniel B Vi-

gneron. Q-ball reconstruction of multimodal fiber orientations using the spherical

harmonic basis. Magnetic Resonance in Medicine, 56(1):104–117, 2006.

[40] Lawrence R Frank. Characterization of anisotropy in high angular resolution

diffusion-weighted MRI. Magnetic Resonance in Medicine, 47(6):1083–1099, 2002.

[41] David S Tuch. Q-ball imaging. Magnetic resonance in medicine, 52(6):1358–1372,

2004.

Page 77: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

67

[42] Alison E Baird and Steven Warach. Magnetic resonance imaging of acute stroke.

Journal of Cerebral Blood Flow & Metabolism, 18(6):583–609, 1998.

[43] Christopher H Sotak. The role of diffusion tensor imaging in the evaluation of

ischemic brain injury–a review. NMR in Biomedicine, 15(7-8):561–569, 2002.

[44] S Warach, D Chien, W Li, M Ronthal, and RR Edelman. Fast magnetic resonance

diffusion-weighted imaging of acute human stroke. Neurology, 42(9):1717–1717,

1992.

[45] A Gregory Sorensen, Ferdinando S Buonanno, R Gilberto Gonzalez, Lee H

Schwamm, Michael H Lev, Frank R Huang-Hellinger, Timothy G Reese, Robert M

Weisskoff, Timothy L Davis, Nijasri Suwanwela, et al. Hyperacute stroke: evalua-

tion with combined multisection diffusion-weighted and hemodynamically weighted

echo-planar MR imaging. Radiology, 199(2):391–401, 1996.

[46] A Gregory Sorensen, William A Copen, Leif Østergaard, Ferdinando S Buonanno,

R Gilberto Gonzalez, Guy Rordorf, Bruce R Rosen, Lee H Schwamm, Robert M

Weisskoff, and Walter J Koroshetz. Hyperacute stroke: simultaneous measurement

of relative cerebral blood volume, relative cerebral blood flow, and mean tissue

transit time. Radiology, 210(2):519–527, 1999.

[47] Heidi Johansen-Berg and Timothy EJ Behrens. Diffusion MRI: from quantitative

measurement to in vivo neuroanatomy. Academic Press, 2013.

[48] Bradford A Moffat, Thomas L Chenevert, Theodore S Lawrence, Charles R Meyer,

Timothy D Johnson, Qian Dong, Christina Tsien, Suresh Mukherji, Douglas J

Quint, Stephen S Gebarski, et al. Functional diffusion map: a noninvasive MRI

biomarker for early stratification of clinical brain tumor response. Proceedings of the

National Academy of Sciences of the United States of America, 102(15):5524–5529,

2005.

Page 78: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

68

[49] Andrzej Dzik-Jurasz, Claudia Domenig, Mark George, Jan Wolber, Anwar Pad-

hani, Gina Brown, and Simon Doran. Diffusion MRI for prediction of response of

rectal cancer to chemoradiation. The Lancet, 360(9329):307–308, 2002.

[50] M Rovaris, A Gass, R Bammer, SJ Hickman, O Ciccarelli, DH Miller, and M Filippi.

Diffusion MRI in multiple sclerosis. Neurology, 65(10):1526–1532, 2005.

[51] Anagha Deshmane, Vikas Gulani, Mark A Griswold, and Nicole Seiberlich. Parallel

mr imaging. Journal of Magnetic Resonance Imaging, 36(1):55–72, 2012.

[52] Nadia Rosenthal and Steve Brown. The mouse ascending: perspectives for human-

disease models. Nature cell biology, 9(9):993–999, 2007.

[53] Brian J Nieman, Ann M Flenniken, S Lee Adamson, R Mark Henkelman, and

John G Sled. Anatomical phenotyping in the brain and skull of a mutant mouse by

magnetic resonance imaging and computed tomography. Physiological Genomics,

24(2):154–162, 2006.

[54] Kechen Zhang and Terrence J Sejnowski. A universal scaling law between gray

matter and white matter of cerebral cortex. Proceedings of the National Academy

of Sciences, 97(10):5621–5626, 2000.

[55] Spencer Noakes, T Leigh, R Mark Henkelman, and Brian J Nieman. Partitioning

k-space for cylindrical three-dimensional rapid acquisition with relaxation enhance-

ment imaging in the mouse brain. NMR in Biomedicine, 30(11), 2017.

[56] Nicholas A Bock, Norman B Konyer, and R Mark Henkelman. Multiple-mouse

MRI. Magnetic resonance in medicine, 49(1):158–167, 2003.

[57] R Mark Henkelman, Leila Baghdadi, and John G Sled. Presentation of 3D isotropic

imaging data for optimal viewing. Magnetic resonance in medicine, 56(6):1371–

1374, 2006.

[58] Jacob Ellegood, Laura K Pacey, David R Hampson, Jason P Lerch, and R Mark

Henkelman. Anatomical phenotyping in a mouse model of fragile X syndrome with

magnetic resonance imaging. Neuroimage, 53(3):1023–1029, 2010.

Page 79: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

69

[59] Yi Jiang and G Allan Johnson. Microscopic diffusion tensor imaging of the mouse

brain. Neuroimage, 50(2):465–471, 2010.

[60] Yi Jiang and G Allan Johnson. Microscopic diffusion tensor atlas of the mouse

brain. Neuroimage, 56(3):1235–1243, 2011.

[61] Klaus Schmierer, Claudia AM Wheeler-Kingshott, Phil A Boulby, Francesco Scar-

avilli, Daniel R Altmann, Gareth J Barker, Paul S Tofts, and David H Miller.

Diffusion tensor imaging of post mortem multiple sclerosis brain. Neuroimage,

35(2):467–477, 2007.

[62] Susumu Mori and Jiangyang Zhang. Principles of diffusion tensor imaging and its

applications to basic neuroscience research. Neuron, 51(5):527–539, 2006.

[63] Dan Wu, Jiadi Xu, Michael T McMahon, Peter CM van Zijl, Susumu Mori,

Frances J Northington, and Jiangyang Zhang. In vivo high-resolution diffusion

tensor imaging of the mouse brain. Neuroimage, 83:18–26, 2013.

[64] Christine L Mac Donald, Krikor Dikranian, Philip Bayly, David Holtzman, and

David Brody. Diffusion tensor imaging reliably detects experimental traumatic

axonal injury and indicates approximate time of injury. Journal of Neuroscience,

27(44):11869–11876, 2007.

[65] S Wang, EX Wu, K Cai, H-F Lau, P-T Cheung, and P-L Khong. Mild hypoxic-

ischemic injury in the neonatal rat brain: longitudinal evaluation of white mat-

ter using diffusion tensor MR imaging. American journal of neuroradiology,

30(10):1907–1913, 2009.

[66] Matthew D Budde, Joong Hee Kim, Hsiao-Fang Liang, Robert E Schmidt, John H

Russell, Anne H Cross, and Sheng-Kwei Song. Toward accurate diagnosis of white

matter pathology using diffusion tensor imaging. Magnetic resonance in medicine,

57(4):688–695, 2007.

[67] Halima Chahboune, Laura R Ment, William B Stewart, Xiaoxian Ma, Douglas L

Rothman, and Fahmeed Hyder. Neurodevelopment of C57B/L6 mouse brain as-

Page 80: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

70

sessed by in vivo diffusion tensor imaging. NMR in biomedicine, 20(3):375–382,

2007.

[68] Tianbo Ren, Jiangyang Zhang, Celine Plachez, Susumu Mori, and Linda J

Richards. Diffusion tensor magnetic resonance imaging and tract-tracing analy-

sis of probst bundle structure in Netrin1- and DCC-deficient mice. Journal of

Neuroscience, 27(39):10345–10349, 2007.

[69] Sajjad Baloch, Ragini Verma, Hao Huang, Parmeshwar Khurd, Sarah Clark, Paul

Yarowsky, Ted Abel, Susumu Mori, and Christos Davatzikos. Quantification of

brain maturation and growth patterns in C57BL/6J mice via computational neu-

roanatomy of diffusion tensor images. Cerebral cortex, 19(3):675–687, 2009.

[70] Jacob Ellegood, Jason P Lerch, and R Mark Henkelman. Brain abnormalities in a

Neuroligin3 R451C knockin mouse model associated with autism. Autism Research,

4(5):368–376, 2011.

[71] D. L. Donoho. Compressed sensing. IEEE Transactions on Information Theory,

52(4):1289–1306, April 2006.

[72] Michael Lustig, David Donoho, and John M. Pauly. Sparse MRI: The application

of compressed sensing for rapid MR imaging. Magnetic Resonance in Medicine,

58(6):1182–1195, 2007.

[73] Emmanuel J Candes, Justin K Romberg, and Terence Tao. Stable signal recov-

ery from incomplete and inaccurate measurements. Communications on pure and

applied mathematics, 59(8):1207–1223, 2006.

[74] Michael W Marcellin, Michael J Gormish, Ali Bilgin, and Martin P Boliek. An

overview of JPEG-2000. In Data Compression Conference, 2000. Proceedings. DCC

2000, pages 523–541. IEEE, 2000.

[75] Leonid Iakov Rudin. Images, numerical analysis of singularities and shock filters.

1987.

Page 81: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

71

[76] Ti-Chiun Chang, Lin He, and Tong Fang. MR image reconstruction from sparse

radial samples using Bregman iteration. In Proceedings of the 13th Annual Meeting

of ISMRM, Seattle, volume 696, page 482, 2006.

[77] Jong Chul Ye, Sungho Tak, Yeji Han, and Hyun Wook Park. Projection reconstruc-

tion MR imaging using FOCUSS. Magnetic Resonance in Medicine, 57(4):764–775,

2007.

[78] Kai Tobias Block, Martin Uecker, and Jens Frahm. Undersampled radial MRI

with multiple coils. iterative image reconstruction using a total variation constraint.

Magnetic resonance in medicine, 57(6):1086–1098, 2007.

[79] Michael Lustig, Jin Hyung Lee, David L Donoho, and John M Pauly. In Proceedings

of the 13th annual meeting of ISMRM, Miami Beach, page 685. Citeseer, 2005.

[80] Juan M Santos, Charles H Cunningham, Michael Lustig, Brian A Hargreaves, Bob S

Hu, Dwight G Nishimura, and John M Pauly. Single breath-hold whole-heart MRA

using variable-density spirals at 3T. Magnetic resonance in medicine, 55(2):371–

379, 2006.

[81] Scott Shaobing Chen, David L Donoho, and Michael A Saunders. Atomic decom-

position by basis pursuit. SIAM review, 43(1):129–159, 2001.

[82] David L Donoho. For most large underdetermined systems of linear equations the

minimal `1-norm solution is also the sparsest solution. Communications on pure

and applied mathematics, 59(6):797–829, 2006.

[83] GM Fung and OL Mangasarian. Equivalence of minimal 0-and p-norm solutions

of linear equalities, inequalities and linear programs for sufficiently small p. Journal

of optimization theory and applications, 151(1):1–10, 2011.

[84] Leonid I Rudin, Stanley Osher, and Emad Fatemi. Nonlinear total variation based

noise removal algorithms. Physica D: Nonlinear Phenomena, 60(1-4):259–268, 1992.

Page 82: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

72

[85] Ricardo Otazo, Daniel Kim, Leon Axel, and Daniel K Sodickson. Combination of

compressed sensing and parallel imaging for highly accelerated first-pass cardiac

perfusion MRI. Magnetic Resonance in Medicine, 64(3):767–776, 2010.

[86] Urs Gamper, Peter Boesiger, and Sebastian Kozerke. Compressed sensing in dy-

namic MRI. Magnetic resonance in medicine, 59(2):365–373, 2008.

[87] Zhongnan Fang, Nguyen Van Le, ManKin Choy, and Jin Hyung Lee. High spatial

resolution compressed sensing (HSPARSE) functional MRI. Magnetic resonance in

medicine, 76(2):440–455, 2016.

[88] Hye-Young Heo, Yi Zhang, Dong-Hoon Lee, Shanshan Jiang, Xuna Zhao, and

Jinyuan Zhou. Accelerating chemical exchange saturation transfer (CEST) MRI

by combining compressed sensing and sensitivity encoding techniques. Magnetic

resonance in medicine, 77(2):779–786, 2017.

[89] Kawin Setsompop, R Kimmlingen, E Eberlein, Thomas Witzel, Julien Cohen-

Adad, Jennifer A McNab, Boris Keil, M Dylan Tisdall, P Hoecht, P Dietz, et al.

Pushing the limits of in vivo diffusion MRI for the human connectome project.

Neuroimage, 80:220–233, 2013.

[90] Berkin Bilgic, Kawin Setsompop, Julien Cohen-Adad, Anastasia Yendiki,

Lawrence L Wald, and Elfar Adalsteinsson. Accelerated diffusion spectrum imag-

ing with compressed sensing using adaptive dictionaries. Magnetic Resonance in

Medicine, 68(6):1747–1754, 2012.

[91] Bennett A Landman, John A Bogovic, Hanlin Wan, Fatma El Zahraa ElShahaby,

Pierre-Louis Bazin, and Jerry L Prince. Resolution of crossing fibers with con-

strained compressed sensing using diffusion tensor MRI. NeuroImage, 59(3):2175–

2186, 2012.

[92] Oleg Michailovich, Yogesh Rathi, and Sudipto Dolui. Spatially regularized com-

pressed sensing for high angular resolution diffusion imaging. IEEE transactions

on medical imaging, 30(5):1100–1115, 2011.

Page 83: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

73

[93] Travis E Oliphant. Python for scientific computing. Computing in Science &

Engineering, 9(3), 2007.

[94] Stefan van der Walt, S Chris Colbert, and Gael Varoquaux. The NumPy array: a

structure for efficient numerical computation. Computing in Science & Engineering,

13(2):22–30, 2011.

[95] Eric Jones, Travis Oliphant, Pearu Peterson, et al. SciPy: Open source scientific

tools for Python, 2001–. [Online; accessed Aug 25, 2017].

[96] Gommers R Wohlfahrt K OLeary A Nahrstaedt H Lee G, Wasilewski F and Con-

tributors. PyWavelets - wavelet transforms in python. https://github.com/

PyWavelets/pywt, 2006.

[97] John D Hunter. Matplotlib: A 2D graphics environment. Computing In Science &

Engineering, 9(3):90–95, 2007.

[98] Jamie Ludwig. Image convolution. Portland State University, 2013.

[99] Markus Deserno. How to generate equidistributed points on the surface of a sphere.

P.-If Polymerforshung (Ed.), 2004.

[100] Jun Dazai, Shoshana Spring, Lindsay S Cahill, and R Mark Henkelman. Multiple-

mouse neuroanatomical magnetic resonance imaging. JoVE (Journal of Visualized

Experiments), (48):e2497–e2497, 2011.

[101] B Jeurissen, A Leemans, JD Tournier, DK Jones, and J Sijbers. Estimating the

number of fiber orientations in diffusion MRI voxels: a constrained spherical de-

convolution study. In International Society for Magnetic Resonance in Medicine

(ISMRM), page 573, 2010.

[102] Eleftherios Garyfallidis, Matthew Brett, Bagrat Amirbekian, Ariel Rokem, Stefan

Van Der Walt, Maxime Descoteaux, and Ian Nimmo-Smith. Dipy, a library for the

analysis of diffusion MRI data. Frontiers in neuroinformatics, 8:8, 2014.

Page 84: Compressed Sensing for Diffusion MRI of the Mouse Brain: A … · 2018-07-18 · Abstract Compressed Sensing for Di usion MRI of the Mouse Brain: A Trade-o of Spatial and Directional

74

[103] Daniela Kuhnt, Miriam HA Bauer, Jan Egger, Mirco Richter, Tina Kapur, Jens

Sommer, Dorit Merhof, and Christopher Nimsky. Fiber tractography based on

diffusion tensor imaging compared with high-angular-resolution diffusion imaging

with compressed sensing: initial experience. Neurosurgery, 72(0 1):165, 2013.