Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
Comparative Evaluation of Sport Port
(Tidal Port) vs. Vortex Port for
Therapeutic Plasma Exchange in Vivo
Rebecca Dill, RN, Jennifer Wintz, RN, Shelli Chernesky, RN, Lisa
Dalton, RN, Sean G. Yates, MD, Nicole De Simone, MD, Ravi Sarode,
MD, Karen Matevosyan, MD
University Of Texas Southwestern Medical Center at Dallas
• UT Southwestern Medical Center
• Over 600 beds
• Provide in- and outpatient apheresis services
• 3,300 therapeutic apheresis procedures per year
Background
• Angiodynamics Vortex Ports single and double lumen have been
routinely utilized at our UT Southwestern Medical Center for
vascular access for extracorporeal photopheresis, RBC exchange
and, to a limited extend, Therapeutic Plasma Exchange (TPE)
• Its use for TPE has resulted in lower flow rates due to limited flow
capacity of the port
– Single-lumen 60 mL/min, dual-lumen 50 mL/min
• Long term use of Vortex port requires frequent use of
tissue plasminogen activator (tPA) due to sludge and fibrin
sheath formation in the chambers of the port
– likely due to the cylindrical design of the Vortex port chamber
Background
• In 2014 Sport/Tidal Port (Norfolk Medical, Skokie, IL) was evaluated at our institution through in vitro studies
• The study supported company’s claims that the Sport/Tidal Port was capable of supporting continuous flow rates up to 100 mL/min, suitable for TPE- Reported at 2015 ASFA annual meeting
• Additional claim by Norfolk Medical• the spherical design of the Sport/Tidal Port should significantly reduce
sludging
• Resulting in decreased tPA requirement
Background
Port Chambers: Rounded vs Spherical
6
Rounded Spherical
• Sport/Tidal Port has been used since April 2015
• The Sport/Tidal port was used for inlet, while peripheral vein
used for return
• Bilateral Single-lumen Vortex Ports served as control
• The purpose of the study was to evaluate the performance of
Sport/Tidal Port in vivo and compare its performance to that of
bilateral single-lumen Vortex ports in patients undergoing TPE
Purpose
• 40 procedures were evaluated on 10 patients
• Sport/Tidal Port - 26 procedures (5 patients)
• Vortex Port - 14 procedures (5 patients)
• 1-1.2 plasma volume TPE • 5% albumin or normal saline/albumin replacement
• COBE Spectra and Spectra Optia cell separators
Study Design
Patients Demographics
• All patients required long-term TPE
• Age, mean (range) - 51.7 (36-62)• Male (1)• Female (9)
• Diagnoses• Multiple sclerosis (6)
• Neuromyelitis optica(1)
• Myasthenia gravis (2)
• Chronic inflammatory demyelinating polyneuropathy (1)
• All TPE's were performed at the highest allowable alarm-free inlet rates
• Mann-Whitney U-test was used for comparative analysis
Methods
• Procedural parameters• Inlet volumes
• Inlet flow rates
• Total Procedure Time
• Pure Procedure Run Time (instrument clock time)
• Inlet pressure alarms
• Operator interventions including port flushing
• Use of tPA
Data Collected
Parameters SportPort (N=5)
Vortex Port (N=5)
p value
Total procedures, n=40 26 14
Inlet volume (mL), mean (SD) 3929 (+777) 4066 (+643)
Inlet flow rate (mL/min), mean (SD) 67.4 (+11.0) 56.1 (+10.2) 0.002
Pure procedure run time (min), mean (SD) 86.2 (+35.7) 122.9 (+35.7) 0.001
Total procedure time (min), mean (SD) 99.8 (+32.0) 136.2 (+33.5) 0.002
Inlet pressure alarms per procedure, n 1.7 5.1
tPA requirement, n (%) 1 (4.3%) 3 (21.4%)
Results
• Both ports exhibited slower than optimal inlet flow rates for TPE
• Sport/Tidal Port allowed a slightly faster Inlet flow rate (p value
0.002)
• Sustained flow rates approaching 100 mL/min observed in the in
vitro study, were not achievable
• Vortex Port had 3 times greater Inlet flow/pressure-related
alarms compared to the Sport/Tidal Port
Conclusion
• Pure Procedure Run Time and Total Procedure Time were
shorter with Sport /Tidal Port (respective p values 0.001 and
0.002)
• Sport/Tidal Port generated 3 times fewer inlet flow/pressure
related alarms compared to Vortex port
• Sport/Tidal Port required 3x fewer tPA treatment compared to
Vortex port (4.3% vs. 21.4%)
Conclusion
• These initial results suggest that the Sport/Tidal Port is the
fastest continuous implantable port on market suitable for TPE
• However, flow rates of 100 mL/min initially observed in in
vitro evaluation could not be achieved in vivo
• The significant improvement in the procedure time as well as
decreased need for interventions suggest that the Sport/Tidal
Port may be a favorable option over Vortex Port in the clinical
setting for TPE’s
Conclusion
• Study Limitations
• Due to the small sample size of patients and procedures
evaluated further study is needed to verify the superior
performance of the Sport/Tidal Port for TPE
• Due to the limitations, further research is needed for the role
Sport/Tidal port will play on Apheresis procedures other than
TPE’s
Conclusion
ApheresisTeam!