Comparative assessment of Comparative assessment of various automated processes various automated processes following prEN/ISO 15883-1 following prEN/ISO 15883-1 Klaus Roth; Jürgen Gauer; Anna Draghici; Winfried Michels Klaus Roth; Jürgen Gauer; Anna Draghici; Winfried Michels Zentr Steril 2005; 13: 34-44 Zentr Steril 2005; 13: 34-44

Comparative assessment of various automated … · Comparative assessment of various automated processes ... prEN/ISO 15883-1 Annex B ... Comparative assessment of various automated

Embed Size (px)

Citation preview

Comparative assessment of Comparative assessment of various automated processes various automated processes following prEN/ISO 15883-1following prEN/ISO 15883-1Klaus Roth; Jürgen Gauer; Anna Draghici; Winfried MichelsKlaus Roth; Jürgen Gauer; Anna Draghici; Winfried Michels

Zentr Steril 2005; 13: 34-44Zentr Steril 2005; 13: 34-44

prEN/ISO 15883-1 Annex BprEN/ISO 15883-1 Annex B

B.2.8.2 Surgical instrumentsLoad the WD with the test instruments on their tray. If the WD can take additional instrument trays, fill it with clean surgical instruments on their trays according to the manufacturers instructions. If the WD can take more than four but less than nine instrument trays simultaneously, two of these should be with test instruments according to B.2.7.2. If it can take more than eight trays simul-taneously, three of these should be with test instruments according to B.2.7.2.

B.2.9.2 AcceptanceFor the cleaning efficiency of the WD to be deemed acceptable at least 95% ofall test objects for each test object type under test should be clean.

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

B.2.6.2 Surgical instruments

One instrument tray with:

•20 Mayo scissors, curved, 14 cm, stainless steel, polished clean, not lubricated;

•20 forceps, Crile, curved, 14 cm, stainless steel, polished clean, not lubricated.

Investigations Following the Investigations Following the Standards prEN/ISO 15883-1Standards prEN/ISO 15883-1

BH 145R(lower sieve basket)

BH 425R (upper sieve basket)

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

Contamination of the artery clamps Contamination of the artery clamps with 100µl coagulable sheep bloodwith 100µl coagulable sheep blood

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

First InvestigationsFirst Investigations

The radionuclide method can only be used in the laboratory, but it the results of the tests are not influenced by the quality of the recovery procedure. Therfore the experiments in practice have been performed with the Biuret- and modified OPA-Method

Radionuclide method

Correlation of the Radionuclide Method with Correlation of the Radionuclide Method with the OPA-Methodthe OPA-Method

• Standardized in-vitro contamination• Quantification of residues• Detection of critical spots• Method for validation of cleaning behavior • The results of RNM and OPA- method correlate

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

Correlation between RNM- and OPA-MethodCorrelation between RNM- and OPA-Method

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

1/0,02 2/0,02 3/0,02 4/0,02 6/0,10 7/0,10 8/0,10 9/0,10 11/0,25 12/0,25 13/0,25 14/0,25 16/0,40 17/0,40 18/0,40 19/0,40

Instruments

coun

ts/s

ec

0

0,1

0,2

0,3

0,4

0,5

0,6

0,7

0,8

0,9

Extin

ctio

n

RNM OPA-Extinktion

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

Influence of the pre-cleaningInfluence of the pre-cleaning

Initial contamination of each instrument was 120 counts/secComparison pre-cleaning and main wash

5 63 3 5 4 2 4 5 6 4 3 5 3 5 2 3

84

10

12

21

35

13 1019

15 1116

26

1722 19

9

17

149

41

23

33

0

10

20

30

40

50

60

6900

S69

00Z

6900

L69

00F

6900

769

006

6900

269

014

6900

969

012

6900

U69

00G

6900

V69

00N

6900

569

00P

6900

Y69

008

6900

069

013

Instruments

coun

ts/s

ec

Main wash Pre-cleaning

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

Influence of the pre-cleaningInfluence of the pre-cleaning

2180

390

89

2034

177

51

0

500

1000

1500

2000

2500

contamination pre-cleaning main wash

cleaning step

coun

ts/s

ec

sieve basket BH425Rsieve basket BH145R

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

- 2 min cold pre-wash using tap water

- Drainage

- 5 min wash at 55°C using 0,5% alkaline detergent

- Drainage

- 5 min wash at 55°C using 0,5% alkaline detergent and an oxidative additive 0,35%

- 1 min neutralisation at 38°C using an acid additive (0,1 %) based on citric acid

- Drainage

- 1 min intermediate rinse

- Drainage

- 1 min intermediate rinseInvestigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

OxivarioOxivario ProgrammeProgramme

Alkaline detergent A (phosphate, disodium- and potassium-metasilicate, tenside-free, with and without an oxidative additive B (hydrogen peroxide solution somewhat less than 30%) in the second wash step

Also the oxidative additive was tested in different concentrations (0,175 % and 0,35 %)

Alkaline detergent C (sodium phosphate, sodium silicate, potassium hydroxide, tenside-free) with and without an oxidative additive D (hydrogen peroxide solution somewhat more than 30%) in the second wash step at a concentration of 0,35%

Mildly alkaline detergent E (enzymes, tenside, caustic potash solution) with and without oxidative additive B

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

Investigated detergentsInvestigated detergents

Detergent and concentration pH-value Alkaline detergent A 0,5% 11,2

Alkaline detergent A 0,5% & H2O2 0,175% (oxidative additive B) 10,9

Alkaline detergent A 0,5% & H2O2 0,35% (oxidative additive B) 10,4 Alkaline detergent B 0,5% & H2O2 0,35% (oxidative additive D) 11,1 Alkaline detergent E 0,5% 9,8 Alkaline detergent E 0,5% & H2O2 0,35% (oxidative additive B) 9,2

pH-values of the applied concentrations pH-values of the applied concentrations of detergentsof detergents

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

Reduced cycle investigations:Reduced cycle investigations:Reinigungsverhalten im verkürztem Zyklus

8665

55 48

375

116

5945

28 21

278

41

0

50

100

150

200

250

300

350

400

Oxivario Reiniger A Oxivario Reiniger A +Zusatz B (0,175%)

Oxivario Reiniger C +Zusatz D (0,35%)

Oxivario Reiniger A +Zusatz B (0,35%)

Oxivario Reiniger E Oxivario Reiniger E +Zusatz B (0,35%)

Reinigungsverfahren

coun

ts/s

ec

BH425R BH145R

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

Reduced cycle investigations:Reduced cycle investigations: Anzahl der Klemmen über dem Grenzwert

23

96

81

1313

7 6

62

1210

7

59

11

6 6

53

106 6

52

96

47

87

42

8

38

8

28

6

25

6

22

6

14

6

1212

0

10

20

30

40

50

60

70

80

90

Oxivario Reiniger A Oxivario Reiniger A +Zusatz B (0,175%)

Oxivario Reiniger A +Zusatz B (0,35%)

Oxivario Reiniger C +Zusatz D (0,35%)

Oxivario Reiniger E Oxivario Reiniger E +Zusatz B (0,35%)

Reinigungsverfahren

coun

ts/s

ec

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

The test method described in the standard prEN/ISO 15883, when combined with the RNM, proves to be a suitable way of comparing the different wash processes with one another.

The question is then whether the principle of using a partial or half cycle of wash steps to gain improved differentiation, should be used for WDs - as it has already been used to validate sterilisation.

At the moment wash tests are carried out without any safety net, which means that no performance reserves are taken into account to deal with fluctuations that may occur in the practical situation, for example by drying-on of soil.

Discussion (1)Discussion (1)

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

Because the clamps were contaminated in a defined way with a pipette, a standardised starting contamination could be assured. Thus it can be shown that the results of washing with the Oxivario programme without an oxidative effect are definitely inferior, and also that the statistical deviation increases.

Even at a low hydrogen peroxide concentration there was more constant washing with fewer differences in washing results. Another result of the Oxivario process with added hydrogen peroxide is the proven cleaning effectiveness, even in areas that are difficult to reach. This special deep cleaning makes the process superior in comparison to the alkaline process run without an oxidative additive.

Discussion (2)Discussion (2)

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

ConclusionConclusionThe radionuclide method was used to investigate comparatively the cleaning performance of automatic washing/disinfection processes on the critical joint areas of artery clips. The defined initial soil was radioactively marked coagulable blood.

Washing using the Miele Oxivario programme with the addition of hydrogen peroxide achieves an increased cleaning performance compared to that from the alkaline detergent alone, and also attains better results in critical areas, so that the cleaning performance is standardised at high level.

Additionally, following elution of the joint areas, the direct measurements of the radionuclide method were correlated with the indirect determination of residual contamination in the eluate using the OPA method.

Investigation of Cleaning Performance Following the Standard prEN/ISO 15883-1. Zentr Steril 2005; 13: 34-44

In order to implement the up and coming EN ISO 15883 in practice, we worked out the “Guidelines of the DGKH, DGSV and AKI for validating and routinely monitoring automatic washing and disinfecting processes for thermo-stable medical devices, and principles for selecting a washer-disinfector (WD)”Part 1**.

In an inter-hospital trial conducted with 18 hospitals, the practicability and meaningfulness of the test method was investigated.

5 of 18 WD involved in the trial were over threshold.

Although the sensitivity of the semi-quantitative Biuret / BCA method used in the tests is fairly low, it can be presumed that the evaluation framework is sufficient for the moment to identify WDs with poor cleaning quality, and to monitor the quality of the measures carried out here.

Further investigationFurther investigation

Inter-hospital trials to determine minimal cleaning performance according to the Guidelines from DGKH, DGSV and AKI; Zentr Steril 2005; 13