CIEDrelatedartifact
AnishBhuva
BHFClinicalResearchTrainingFellowCardiologyRegistrar
NHSInnovationsFellow
Outline
• Determinantsofartifact• Distributionofartifact
Strategiestoreduceartifactfor:• Cine• LGE• Mapping
RiskvsBenefitTotal 67
Indication N(%)Amyloid 1(2%)PreVTablationscarimaging 15(22%)Cardiomyopathy 26(39%)Cardiomyopathyfollowup 8(12%)EFassessment 1(2%)Ischemiaandviability 8(12%)Myocarditis 6(9%)Thrombus 2(3%)
RiskvsBenefit
RISKNoclinicalcomplications13%minorleadparameterchange
BENEFIT49% Changeinmanagement6% Unexpectednewdiagnoses
ConventionalLGE
BENEFIT88% Changeinmanagement14%Unexpectednewdiagnoses
WB-MOCOLGE
Devicerelatedartifact:devicetype
ILR PPM ICD CRT-D
ILRs
RevealLinq
3/2/18
2
LargerILRs
MedtronicRevealDX
Devicerelatedartifact:deviceproximity Devicerelatedartifact:deviceproximity
3/2/18
3
SSFP SSFP GREGRE
Devicerelatedartifact:sequenceselection
• SSFPsequences(cine,perfusion,T1scout)• Balancedgradientsrequirefieldhomogeneity
• Lateenhancement• Inversionpulserequiresfieldhomogeneitytonullmyocardialsignalintensity
• Darkzoneadjacenttogenerator• Moreremotehyperintesnity fromincompleteinversion
• Reduceinhomogeneity• Pushgeneratorup• Rightsided
• Usesequenceslessaffectedbyinhomogeneity• Spinechoinsteadofgradientecho(localisers)
• NonbalancedGREcine• ReducedTE(increasebandwidth)
• AdjustinversionpulseforLGE
GRESSFP GRESSFP
Butsometimes..
Slicethickness4mmSlicethickness8mm
3/2/18
4
FreebreathingInspiration
Perfusion
T1mappingindevicepatients
Shao J et al Magn Reson Med 2016;2:1–10 SSFP MOLLI look- locker
T1error≈60ms
Kellman P et al J Cardiovasc Magn Reson 2013;15(1):63.SCMR 2017, Moderated poster finalist
*
*
*
*
*
?Amyloidosiswithrevealdevice
T2mapping STIR
T2mappingindevicepatients
3/2/18
5
T2starmappingindevicepatients
*
**
-70
-50
-6
36
87
185
Hemochromatosiswithpermanentpacemaker
Scarimagingiskey• LGEimagingpronetodeviceartifacts
ConventionalLGE:
MOCOSSFP WBMOCOFLASHPSIRBHFLASH
LeadartifactcanmakesubtleLGEdifficulttotell
3/2/18
6
Overall%ofpatientswithartifact
ILR
FLASH
PPM ICD CRTD CompleteCohort
SCMR 2017
DeviceartifactsinLGEimaging
• Hyperintensity artifact• Fromoff-resonance• Signalisnotinvertedduetonarrow
bandwidthofinversionpulse
• Signalvoidartifact• Dephasing acrossvoxeldueto
susceptibilitygradient• Predominatelyinslicedirection
Signalvoid
Hyperintensity
ConventionalFLASHLGE
Combattingartifacts
phase=2*∏*∆f*TE
Minimizeoff-resonancevariationbyusingthinslices
Minimizeechotime(TE)byusingwidebandwidth
B1 (mT)
Off
reso
nanc
e (H
z)
0 0.005 0.01 0.015 0.02 0.025
-2500
-2000
-1500
-1000
-500
0
500
1000
1500
2000
2500
B1 (mT)
Off
reso
nanc
e (H
z)
0 0.005 0.01 0.015 0.02 0.025
-2500
-2000
-1500
-1000
-500
0
500
1000
1500
2000
2500
• Widebandinversion
• Thinnerslices
Acquisitionsequence• Wideband,4mmslices,singleshotFLASH,flipangle11°• Cost:LossofSNR
• TogainSNR:respiratoryMOCOaveraging• 24averages
• Benefitsinclude:• Freebreathing• PSIR
Acknowledgement: Peter Kellman
56 year old, LAD infarct, LV 30%
3/2/18
7
Notjustimplantablecardiacdevices..Conclusions
• CardiovascularMRimagingisdiagnosticinnon-defibrillatorpatients
• Artifactisrelatedtodevice,locationandsequence
• Artifactreductionmayrequire‘onthefly’sequenceadjustment