CAV®-2100/2200Automatic Viscometer

Installation Guide

Page 2: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup
Page 3: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

i

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CONTENTS

1 CUSTOMER PRE-INSTALLATION CHECKLIST FOR THE CANNONAUTOMATIC VISCOMETER 1

Purpose of this guide .......................................................................................................... 1Required utilities for CAV installation ............................................................................... 1

1.___Electrical power ............................................................................................. 12.___Compressed air/nitrogen and vacuum ........................................................... 13.___Sample/waste disposal ................................................................................... 24.___Cooling water ................................................................................................ 25.___Ventilation...................................................................................................... 26.___Laboratory environment ................................................................................ 2

Additional installation requirements .................................................................................. 31.___Size ................................................................................................................ 32.___Clearance ....................................................................................................... 33.___Personnel and training ................................................................................... 34.___Tools .............................................................................................................. 45.___Solvent ........................................................................................................... 46.___Computer ....................................................................................................... 4

2 CAV INSTALLATION 5CAV setup options .............................................................................................................. 5CAV components ................................................................................................................ 6

Component descriptions ......................................................................................... 6Unpacking components .......................................................................................... 7Arranging components ........................................................................................... 8

Bath unit connections ......................................................................................................... 9Bath pedestal .......................................................................................................... 9Computer connection.............................................................................................. 9Power connection ................................................................................................. 10Inserting viscometer tubes .................................................................................... 10Filling the bath(s) ................................................................................................. 10Coolant connections ............................................................................................. 12Sample trays/drains .............................................................................................. 13Splash guards ........................................................................................................ 14Ventilation exhaust shroud ................................................................................... 14

Page 4: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

ii

2 CAV INSTALLATION (continued)Service Unit connections ..................................................................................................... 15

Air connections ..................................................................................................... 15Optional Nitrogen connection .............................................................................. 16CSU/SDU connection ........................................................................................... 16Vacuum connections ............................................................................................. 16Power connections ................................................................................................ 16MAINS~ ............................................................................................................... 16Service Unit operational settings .......................................................................... 16Other options for house air/nitrogen/vacuum ....................................................... 17

Solvent system connections .............................................................................................. 17Safety ground connections ................................................................................... 17Solvent connections (single and dual solvent systems) ........................................ 18Waste can connections .......................................................................................... 19Emptying the waste can ........................................................................................ 19

Adjusting pneumatic controls ........................................................................................... 20Installing VISCPRO® software ......................................................................................... 22

Computer requirements ........................................................................................ 22Windows® installation .............................................................................................. 22Installation actions ................................................................................................ 22

3 Software configuration 23Logging in ............................................................................................................ 23Checking Instrument Settings............................................................................... 24Tray Settings: Tube and Bath ............................................................................... 25Tray Settings: Wash .............................................................................................. 27Testing the configuration ...................................................................................... 28

A APPENDIX A—Multi-Unit configuration 29Kit components ................................................................................................................. 29Procedure .......................................................................................................................... 29

Connecting components ....................................................................................... 29Completing RS-485 connections .......................................................................... 29

B APPENDIX B—Tubing connections/photography 31

Page 5: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

1

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CHAPTER

1 CUSTOMER PRE-INSTALLATIONCHECKLIST FOR THE CANNON

AUTOMATIC VISCOMETER

Purpose of this guide

The CANNON® Automatic Viscometer Installation and Setup Guide isdesigned to assist you in completing initial installation and setup of theCAV-2100 viscometer. For additional information on CAV operation andmaintenance, consult your Instruction & Operation Manual.

Assembly diagram An assembly diagram with additional installation notes may also havebeen provided on a separate sheet. You may also refer to the photographyand diagrams throughout this Guide.

Using the checklist This chapter uses a checklist format to detail the required utilities,accessories, clearances, personnel and training issues to be consideredprior to, and during installation of the CAV-2100 instrument(s) in yourfacility. Please review the checklist carefully and address any outstandingissues to facilitate a successful installation.

Required utilities for CAV installation

The following utilities must be supplied by the CAV user:

1.___Electrical powerMAINS~ The CAV is designed to operate from several world AC power sources.

However, voltage and frequency requirements are specified by the CAVuser when the CAV is ordered. The exact power requirements for theCAV are indicated on the label on the rear panel of the Bath Unit. Con-sult CANNON® with any questions regarding power requirements.

CAUTION Proper electrical grounding is extremely important. The CAV is suppliedwith a power cord with an integral safety ground which must be con-nected to an AC mains receptacle with an earth ground.

2.___Compressed air/nitrogen and vacuumThe CAV requires a source of compressed air or nitrogen. The com-pressed air must be free of water, oil and particulates. Consumption ofthe compressed gas is dependent upon the operations being performed.

Page 6: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

2

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

The compressor must be capable of producing a constant air flow of 3cfm (5 m3/hr) at 40 psi (275 kPa) minimum. Line pressure should be setat 75-100 psi (517-690 kPa). If the CSU-200 Service Unit is not beingused, vacuum must be supplied by the user to the rear panel VACUUMconnection on the Bath Unit. Vacuum should be approximately -8 in Hgto -10 in Hg, and should be suitable for the solvent/samples being run.

3.___Sample/waste disposalThe customer must provide for the containment and disposal of all usedsolvents and waste oils. Collection, storage, and disposal of all wastefrom the CAV is the responsibility of the customer, and must be done inaccordance with all local safety and environmental regulations.

4.___Cooling waterInstruments maintaining bath temperatures lower than 40°C require asource of cooling water. Cooling water connections are located at the rearof the CAV unit. The cooling water should be maintained at 15-20°Cbelow the bath temperature setting.

NOTE Temperature baths operating at 40°C may require circulating coolingwater, particularly if the room temperature of the laboratory exceeds24°C (75°F).

5.___VentilationAll models of the CAV are provided with a 4” (10.16 cm) diameterventilation shroud in the rear panel of the unit. A ventilation systemhaving a continuous airflow adequate to ensure removal of solvent fumesand vapors should be connected to the CAV. CANNON® InstrumentCompany does not supply ventilation fans, hoods, or systems.

CAUTION The CAV must be operated with ventilation precautions appropriate foryour facility and the sample materials and solvents being used.CANNON® recommends that the CAV be connected to an externalventilation system. The 4” (10.16 cm) diameter opening on the backpanel of the CAV will accommodate a standard 4" dryer hose. If anysolvent vapors are detected around the CAV, verify the operation of theexhaust ventilation system.

6.___Laboratory environmentIt is important that the CAV be kept in a suitable laboratory environment.The room temperature of the laboratory must be maintained between15°C and 27°C (60-80°F). The laboratory must be free of dust and dirt,which could possibly taint the samples. The laboratory must be equippedwith all of the standard safety features normally associated with use ofsolvents, corrosives, etc.

Page 7: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

3

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Additional installation requirements

1.___SizeSINGLE BATH UNIT—Principal crate containing the Bath Unit:

Dimensions Crated 28 inches (71 cm) high48 inches (122 cm) wide61 inches (155 cm) long

Dimensions Unpacked (see enclosed diagrams)

Weight Crated 450 lbs. (205 kg)

DOUBLE BATH UNIT—Principal crate containing the Bath Units:Dimensions Crated (Bath Unit) 38 inches (97 cm) high

48 inches (122 cm) wide61 inches (155 cm) long

Dimensions Unpacked (see enclosed diagrams)(Bath Unit)

Weight Crated (Bath Unit) 650 lbs. (295 kg)

Entrance and handling equipment must be able to accommodate thesedimensions. The principal crate is shipped in a horizontal position, but isdesigned to facilitate tipping of the crate to a vertical position during theinstallation. Handling equipment, or the assistance of 3 persons inaddition to the Service Engineer, will be needed to upright the instrumentcrate and position the Bath Unit during installation. Size and number ofother cartons shipped will vary with each order. In order insure a timelyinstallation it is advisable that the crates be moved to the location of theactual installation before the arrival of the Service Engineer.

2.___ClearanceThe CAV requires a ceiling clearance of 68 inches (173 cm) above thepedestal base or benchtop for installation and service (insertion ofthermometers and viscometer tubes). The instrument pedestal base mayequipped with casters for convenient moving. In many cases drop-ceilingpanels can be removed to facilitate access.

3.___Personnel and trainingIt is advisable that the personnel who will be responsible for maintainingthe instrument be present during assembly and power up (approximately1-2 days, depending on installation configuration). Training normallybegins with calibration on the second day. During this time, the operatingtechnician’s presence will be necessary. Training and calibration willrequire 2-3 days. The customer should be prepared to supply typicalsamples for running tests during the training process.

Page 8: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

4

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

4.___Tools6 ft. step ladder • 7/16 (11mm) socket with ratchet • hammer •cat’s-paw type crowbar • #2 Phillips screwdriver(All other tools utilized for assembly are included with the instrument.)

5.___SolventCANNON® Instrument Company does not specify or recommend thetype of solvents to be used with the CAV. Any solvent used with the CAVmust effectively clean the viscometer tubes, dry by evaporation withcompressed air and be compatible with valves and materials in the CAV.Solvents and test materials are the responsibility of the customer andmust be handled in accordance with environmental regulations and thecustomer’s safety policies. 5 GAL US (20L) of solvent are needed duringthe installation and training period. If the CAV is equipped with the dualsolvent option then 5 GAL US of each solvent are needed. Acetone orMEK are specifically prohibited as they are not compatible with theinternal components of the instrument.

CAUTION All sample and solvent handling components are either glass, stainlesssteel, brass, or FEP. All samples/solvents used with the CAV must becompatible with these materials.

6.___ComputerIn the event that the controlling computer is not supplied by CANNON®,refer to the computer specification sheet or contact CANNON® forcurrent computer specifications.

Page 9: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

5

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CHAPTER

2 CAV INSTALLATION

CAV setup options

The CAV has been designed as a modular assembly for flexibility andexpandability. The VISCPRO™ software for your computer (PC) permitsindividual control of up to eight viscometers in four independent tem-perature Bath Units. The accessory CSU-200 Service Unit has also beendesigned to service multiple baths. The instructions in the body of thismanual assume single-bath installation. For information on multiple-unitinstallations, consult APPENDIX A and the enclosed diagram(s). The BathUnit for the CAV-2100 or the CAV-2200 may be placed on the benchtop or aspecially-designed pedestal base, available from CANNON®.

CAV-2000 single and dual-bath installation on pedestal base

Page 10: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

6

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CAV components

Component descriptionsThe standard CAV installation includes three primary components:

Bath unit Solvent Dispensing Unit (SDU) Service Unit (CSU)

Bath unit The Bath Unit houses the primary control and operating systems for theCAV. The lower section of the Bath Unit includes a pneumatics drawerwith valves and controls for moving the sample trays, vials, and wastereceivers. The middle section of the Bath Unit includes the sample table,sample trays, and the temperature bath with viscometer tubes. The upperchassis is located above the bath and includes the control panel andadditional control hardware and electronics for local and automatic modeoperation. The CAV-2100 contains two viscometers within a single bath.The CAV-2200 contains twin baths, each with one viscometer tube.

SDU The Solvent Dispensing Unit (SDU-200) channels solvent from theuser’s pressureless solvent container to the CAV instrument.

Service Unit The Service Unit (CSU-200) contains a vacuum pump and pressureregulators for air and vacuum. Each CSU-200 is capable of providingregulated gas/vacuum for up to four Bath Units.

NOTE A compressed air source is required for connection to the Service Unit.See Service Unit Connections section for more information.

Page 11: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

7

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Unpacking components

CAUTION The CAV Bath Unit weighs approximately 91 kilograms (200 pounds)when the bath is empty. Use appropriate safety procedures when liftingand/or moving the unit. At least two or three people should assist.

1. Remove all components from the shipping containers. To remove theCAV Bath Unit from its crate, consult the diagrams below.

2. Remove all packing materials from the components, including foamcollars around bath and stirrer motor (requires removal of top pan-els)..

3. Verify reception of shipped materials by comparing equipment itemswith packing/parts list(s). Report missing items to CANNON®

Instrument Company immediately.

4. Inspect each component for signs of damage. Report damages to theshipper and CANNON® Instrument Company immediately.

A B

C D E

A Bath Unit in crate in shipping (horizontal) configuration(Locate in proximity of Bath Unit resting position)

B Unit with crate lid removed(To detach, remove lid lag screws with a 7/16” socket wrench)

C Crate in upright (vertical) position(Carefully rock the packing crate up on end)

D Crate with Bath Unit removed(Relocate Bath Unit to pedestal or benchtop)

E Bath Unit in upright (installed) position(See notes following)

Page 12: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

8

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

NOTES Particular care should be taken when unpacking and setting up the CAVBath Unit. The glass bath(s) is/are shipped installed inside the Bath Unit,supported with a foam collar, as is the motor stirrer. The collar compo-nents should be removed when the Bath Unit has been positioned in itsresting location (see photo). After the packing materials have beenremoved, check to make certain that each bath drain valve is in itsclosed position (handle perpendicular to valve) and that the drain plug isin place. Then attach the front window panel to the bath frame with sixPhillips-head screws (provided).

If you have ordered the Pedestal Base, install the Base Plate Runners tothe bottom of the Bath Unit per the instructions included with the Base(Runners may already be factory-installed on your unit).

If you are ordering more than one bath unit, each pair of Bath Units willbe shipped in a double crate in “stacked” configuration.

CAUTION Do not operate the CAV bath without the front window panel installed.

Removing foam collar(s) Checking Drain Valve Installing front window panel

Damaged items Retain all packing materials until the instrument is connected and func-tioning properly. If any component(s) must be returned toCANNON® Instrument Company, the damaged item(s) should be pack-aged in the original shipping container. Contact your local CANNON®

agent for procedures on returning products to CANNON® .

Arranging componentsCAV components should be placed adjacent to each other during setupand as close to their final positions as possible. Note that access to thefront and rear of CAV components will be necessary during setup. If youare completing a multiple unit configuration, refer to the Appendix of thisGuide for further information on setup options.

Removing motor stirrer collar

Page 13: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

9

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Bath unit connections

Refer to the instructions below when installing the bath and completingBath Unit connections. Supplementary diagrams and photography areavailable in the appendices.

Bath unit upper rear panel

Bath pedestalIf you have purchased the pedestal base for the CAV, place the pedestal atits intended location. Level the pedestal. Then position the Bath Unit(s) ontop of the pedestal. Follow the instructions included with the pedestal forsecuring the Bath Unit to the pedestal base.

CAUTION If you are installing a single CAV unit, make certain that the unit iscentered on the pedestal base. Two Bath Units may be placed adjacenton a single pedestal. See APPENDIX A for more information on multi-unitconfiguration.

Computer connectionAn IBM® (or IBM®-compatible) computer using the Windows® operatingsystem (OS) is required to operate the VISCPRO® controlling software.Connect the computer to the CAV via the standard 9-pin RS-232 connectorlocated on the upper rear panel of the Bath Unit. For information on installingthe VISCPRO® software, refer to the final section of this chapter.

NOTE Multiple CAV instruments may be networked to a single computer viaRS-485 cable connections available with the Multi-Unit Interface Kit. SeeAPPENDIX A for more information.

Page 14: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

10

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Power connectionThe Bath Unit Mains power connection is located in the center of the rearpanel. Voltage and frequency requirements are indicated on the rear panellabel.

MAINS~ The ~MAINS symbol indicates instructions or connections for the ACpower supply. The AC Power input must match the electrical specifica-tions listed on the label on the rear panel of the instrument. The suppliedAC Mains power cord must be attached to the connector labelled~MAINS. This connection serves as a means of emergency disconnectand should be readily accessible.

NOTE AC Mains power is fused on both the high and low side of the line (twofuses).

CAUTION Do not turn on the Bath Unit power switch until bath fluid has beenadded to the appropriate fill level (see next section for instructions onfilling the temperature bath).

Inserting viscometer tubesThe CAV is ordinarily shipped with viscometer tubes in place in thebath(s). If this is not the case, tubes should be initially installed by anauthorized CANNON® service agent.

Filling the bath(s)

NOTE The instructions in this section are for filling an EMPTY CAV bath. Forinstructions on maintaining the bath level, consult the CAV 2000 SeriesInstruction & Operation Manual.

CAUTION Complete all Service Unit and Dispensing Unit connections in thischapter BEFORE filling the CAV bath(s).

Bath fluid grades Each temperature bath is ordinarily filled with Dow Corning 200 siliconefluid, supplied by CANNON® Instrument Company. The viscosity of thefluid depends upon the temperature of the bath during most operations:

htaBegnargnitarepo

dednemmoceRytisocsivdiulfhtab

C°001-02 tSc01

C°051-101 tSc02

Page 15: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

11

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Initial pre-fill procedure 1. Make certain that the drain valve at the bottom of the bath is closedand that the plug is in place (see photo). If it is not, you will need toremove the front window panel to access the drain valve. Loosen andremove the screws securing the top rear cover to the CAV Bath Unit.Then remove the cover.

2. (CAV-2100 instrument only) If the bath is empty, remove the plug/cap from the Swagelok® fitting at the front of the top bath flange.This fitting is for thethermometer holder.Removing it will preventair lock during the initialfill procedure.

3. Obtain the (supplied)funnel and locate the filltube (CAV-2100, seephoto) or fill hole (CAV-2200, see photo) at the topof the temperature bath.For the CAV-2100, openthe flip-cap on the expan-sion vessel. For the CAV-2200, loosen the knurledthmbscrew securing thealuminum hole cover to thetop bath flange and slidethe cover free of theopening to insert thefunnel.

NOTE If you have installed a cooling device in place of the fill tube, you may fillthe CAV-2100 bath through the expansion vessel (see photo). The CAV-2200 does not have an expansion vessel.

4. Select the desired bath fluid grade for your operating temperature(s).

5. Using the funnel, slowly introduce the appropriate grade of siliconebath fluid. For CAV-2200 baths, continue adding fluid until the cold-fill line is reached. For CAV-2100 instruments, add fluid until theliquid meniscus is 1/4” (3.8 mm) below the top bath flange at the topof the bath. Reduce the flow rate as the meniscus approaches the topof the bath to avoid overfilling!

6. (CAV-2200 only) Turn on the CAV instrument to engage the pumpsystem. Monitor the liquid level as fluid enters the pump system andadd additional fluid as necessary to refill to the cold fill line.

7. Remove the funnel. Return the flip cap or hole cover to its originalposition.

8. (CAV-2100 only) Insert the reference thermometer into the thermom-eter holder and slide the thermometer holder into the fitting at the

CAV-2000 Fill tube filloption

CAV-2000 top view (lift flip cap to fillthrough the expansion vessel)

flip-cap

thermometerholder

expa

nsion

vess

el

CAV-2200 fill hole cover

Drain valve in closedposition w/plug in place

Page 16: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

12

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

front of the top bath flange. The holder has two O-rings under thefitting nut. Carefully slide these O-rings down the thermometerholder and into the fitting-top. Then place the fitting nut over the O-rings. Tighten the fitting finger-tight, then secure with a wrench.

CAUTION Do not overtighten the Swagelok® fitting or you may damage the O-rings.

NOTE For information on draining the bath, consult the Instruction & OperationManual.

9. After installing the thermometer and holder, check to make certainthat the other top bath flange fittings are sealed tightly.

CAUTION During the initial CAV-2100 heating cycle,monitor the liquid level asit rises into the expansionvessel. Check the bathfittings for leaks to com-plete the fill procedure.Tighten fittings if neces-sary.

NOTES The expansion vessel hasbeen designed to holdexcess liquid when theCAV-2100 is operating athigh bath temperatures(see photo, right). Asindicated previously, theexpansion vessel will beempty at ambient tem-perature. Do not fill thebath above the levelindicated in step #5 whenthe bath temperature is ator near ambient.

CAUTION When filling the temperature bath, do not fill above the top bath flange(CAV-2100) or the cold-fill line (CAV-2200).

Coolant connectionsThe CAV offers two different cooling options when operating the instru-ment at temperatures lower than 40°C—a heat exchanger tube or theCBC-2K CANNON® Bath Cooler (older models provided a coolingcoil). Both devices are installed via the 20-mm Swagelok® fitting in thetop bath flange. When cooling devices are in use, the fitting secures aconvenient fill tube for bath fluid. Remove the fill tube before installingyour cooling device!

CAV-2000 only: Elevated liquid level inexpansion vessel during heating

meniscus

Page 17: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

13

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Exchanger connections The heat exchanger tube is installed via the 20-mm Swagelok® fitting inthe top bath flange. Slide the 20-mm nut and two O-rings onto the heatexchanger tube. Insert the tube into the bath as far as it will go. Thentighten the nut finger-tight and secure with an additional 1/4 turn. Checkfor leaks when filling the bath (see next sections). Ensure that the internalcoolant lines from the 3/8” Swagelok® fittings on the exchanger tube areconnected to the COOLANT IN/OUT tubing ports on the rear panel ofthe CAV Bath Unit. Connect tubing from the external coolant ports (seediagrams/photos in APPENDICES) to the appropriate coolant inlet/outleton your cooling unit. Activate your chiller when cooling is needed.

Cooler/chiller requirements

stnemeriuqeRrellihC/relooClavomeRtaeH muminimC°02@W002

erutarepmeTlortnoC

C°0.3±ytilibatS,C°52~5tnioPteS

pmuPrab0taetunimrep)strauq(sretil51-8

)isp(

yticapaC )strauq(sretil3-2

NOTE Coolant should be maintained at a temperature 10-15°C below thedesired bath temperature. Contact CANNON® for recommendations onchiller manufacturers and models for your application.

CBC-2K connections The CBC-2K CANNON® Bath Cooler is installed in the 20-mm connec-tion in the aperture in the top bath flange of the Bath Unit. It is securedwith two O-rings. Slide the nut and two O-rings onto the CBC-2K tube.Then slide the CBC-2K into the bath until the bottom of the finger isapproximately 2” from the bottom of the bath (or until the bend in thetube prevents further insertion). Then tighten the cap nut finger-tight andsecure with an additional 1/4 turn.

CAUTION The CBC-2K is designed for cooling between 20°C and 100°C. Do notuse the CBC-2K with the CAV high-temperature bath option.

Sample trays/drainsBoth heated and unheated sample trays will slide into place along thegrooved track in the sample table. The plated steel vial adaptors shouldbe placed in the receptacles in the tray. Glass sample vials slide into theadaptors. Place the sample tray in the track and slide the tray forwarduntil the detente mechanism engages. At position 1 (first sample directlyunderneath the viscometer tube), a heated sample tray should snap intoposition against the detent mechanism when the front of the tray extendsapproximately 2 centimeters (3/4”) over the edge of the sample table. Theforward edge of an unheated tray will be indented approximately 2.2centimeters (7/8”).

Page 18: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

14

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Electrical connections For (optional)heated sampletrays, connectthe keyed three-pin cable fromthe sample trayto the interiorwall connectorabove thesample table(see photo, nextpage), orientingthe cable in a“U” configura-tion aside thesample tray.Turn the knurledring clockwise to engage and secure the connector.

Tray/drain heat operations To provide power to the heated sample trays and/or heated drain lineswhich may be installed on your CAV unit, you will use the heater ON/OFF switches on the left side of the front panel (see diagram, left). Thesetwo switches provide power to left- and right-hand heating elements. Toadjust temperature, use the Adjust knobs on the upper front panel of theBath Unit. Turn the knobs clockwise to increase temperature and counter-clockwise to lower it. The knobs regulate the duty cycle of the heaters forthe sample trays and the drain lines.

The low-temperature sample heaters will maintain temperature stabilityas high as 80°C. The high-temperature sample heaters will maintaintemperature stability as high as 100°C.

NOTE During normal operation, the green lamp on the heated sample trayholder will illuminate and/or blink in tandem with the red LED on the frontpanel ON/OFF switch. If the green lamp is NOT flashing, check cableconnections and sample block temperature (block may be overheatedand may need to cool down).

Using a temperature probe A recessed thermometer well (3.63 mm; 0.143” ID) at the front of thesample tray permits insertion of a temperature probe to monitor traytemperature for heated sample trays.

Splash guardsTo reduce the risk of contaminating samples, and to protect the viscom-eter tube from damage, the exposed viscometer tube tips should beshielded with splash guards (provided). Use the captive thumb screw onthe top of the splash guard to secure it to the underside of the Bath Unitframe in front of the viscometer tube tip. The felt pad should face thetube tip (see photo, previous page).

heaterconnection

wastereceiver

heateddrain

Installed heated sample tray (splash guard removedon left viscometer tube to reveal tube tip)

Page 19: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

15

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Ventilation exhaust shroud

The CAV requires installation of the exhaust shroud on the rear panel ofthe unit. The shroud must be connected to a customer-supplied exhaustfan/ventilation system. The air flow rate and design of the exhaust fanmust be suitable to handle the vapors from the customer supplied solventand samples. A 4" (10 cm) solvent-safe ventilation hose should beconnected to the exhaust shroud on each Bath Unit.

Service Unit connections

The CSU-200 Service Unit houses an internal vacuum pump and regula-tor, as well as two air pressure regulators. Pressure/vacuum settings aremonitored via the front panel dials. Up to four CAV instruments can beconnected to a single Service Unit.

CSU-200 connections To complete CSU-200 connections, refer to interconnect diagrams,including the diagram on the last page of this manual, and these instruc-tions.

CAUTION IMPORTANT SAFETY NOTE: The SDU-200 Solvent Dispensing Unitmust be properly grounded prior to use. See below for details on properinstallation and grounding.

Air connectionsUsing 1/4” OD Poly-Flo® tubing, complete an attachment from the userair source (maximum pressure—100 psi (690 kPa); minimum 75 psi(517 kPa)) to the MAIN INPUT connector on the rear panel of theService Unit. The compressor must be capable of producing a constantflow of 3 cfm (5 m3/hr) at 40 psi (275 kPa). A water trap and air filter

Customer adapter for ventilation hoses frommultiple CAV-2000 instruments

Installing vent hose on exhaust shroud

Page 20: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

16

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

must be installed by the customer on the laboratory air supply line. A 60-micron filter is factory-installed inside the CSU-200 on the main air inlet(MAIN INPUT).

Tighten the Poly-Flo® fittings finger-tight and secure with an additionalfull turn using an adjustable or open-end wrench.

Locate the AIR and NVIS ports on the lower rear panel of the Bath Unit.Using two short pieces of the 1/4” OD tubing, connect these ports to aPolyflo “T” fitting. Connect the third leg of the “T” to the CSU-200 portlabeled TO CAV.

Connect a short length of 1/4” OD tubing to the TO CAV connector onthe Service Unit. “T” the open end of the tube; then attach both legs tothe AIR and NSOLV connectors on the lower rear panel of the Bath Unit.

Optional Nitrogen connectionThe CSU-200 must be internally replumbed prior to using the nitrogenport. Consult CANNON for details of replumbing.

CSU/SDU connectionConnect the 1/4” OD tubing from the SOLVENT FLOW CONTROLNSOLV IN connection on the Solvent Dispensing Unit to the NSOLVOUT connector on the Service Unit. Tighten the Poly-Flo® fittingsfinger-tight and secure with an additional full turn using an adjustable oropen end wrench.

Vacuum connectionsConnect the CANNON®-supplied vacuum trap to the VACUUM TOCAV connector on the CSU-200. Then connect a length of 1/4” OD Poly-Flo® tubing from the other connector on the vacuum trap to theVACUUM connector on the CAV Bath Unit upper rear panel.

Tighten the Poly-Flo® fittings finger-tight and secure by tightening anadditional full turn with an adjustable or open end wrench.

Attach the vacuum pump muffler to the EXHAUST connector on theCSU-200.

Power connections

MAINS~ The Service Unit Mains power connection is located in the upper leftcorner of the CSU rear panel. Voltage/frequency requirements are clearlyindicated on the rear panel label.

CAUTION Do not turn on the Service Unit power switch until all connections havebeen completed correctly and the SDU has been properly grounded.

Page 21: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

17

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Service Unit operational settingsAfter completing connections for all bath, Service Unit and DispensingUnit components, turn on the power switch on the CSU-200 front paneland check the pressure/vacuum readings on the front panel dials:

DIAL SETTING: ........ DESIRED VALUEMAIN AIR .................. 40-50 psi (275~345 kPa)NSLV ......................... 15-25 psi (100~150 kPa)VACUUM .................. -8 inches of mercury (-8 in Hg)

Adjust the settings if necessary, using the control regulator knobs oppo-site the dials.

NOTE Readjustment of the Service Unit regulators is seldom necessary unlessthe dial settings are outside the parameters outlined above.

Other options for house air/nitrogen/vacuumIn the event that the Service Unit is not used, it is the user’s responsibilityto provide regulated vacuum and air/nitrogen per CANNON® specifica-tions. Consider the following if/when configuring your system:

1. Single solvent systems should be plumbed to solvent port ‘A’ andport ‘B’ must be capped at each Bath Unit. Solvent port B plugs arealready installed on factory-configured single-solvent systems.

2. A water trap and air filter (60-micron screen) must be installed by thecustomer on the house air supply line.

3. House air must not exceed 100 psi (690 kPa).

4. House air may be substituted for the Nitrogen. Pressure must notexceed 75 psi (520 kPa) for tube drying. Pressure must not exceed 25psi (150 kPa) for the NITROGEN connection.

5. Vacuum supplied to the rear panel VACUUM connection on the CAVBath Unit should be approximately -8 in Hg to -10 in Hg, and besuitable for the solvent/samples being run.

Solvent system connections

Safety ground connectionsThe SDU is equipped with a grounding terminal (see photo image, nextpage).

Obtain a ground strap with a 1 Meg resistor. Then connect the strap fromthe grounding terminal to the CAV ground stud (if present) or to the CAVmetal frame at the rear of the Bath Unit using the existing screws.

CAUTION Do not operate this unit before verifying correct installation of the groundstrap.

Page 22: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

18

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

The CAV is equipped with a solvent system used to wash the viscometertubes after each sample test. The system consists of the Solvent Dispens-ing Unit (SDU), the internal and external drain lines, and the user-supplied solvent and waste cans. The SDU will interface with the CAN-NON Service Unit (CSU) and the CAV instrument. Use the instructionsin the following section to complete solvent connections.

CSU SDU

Solvent connections (single and dual solvent systems)Refer to the included interconnect diagram for information on single ordual solvent connections. Solvent connections are made at the connectorson the rearpanel of theSolventDispensingUnit (SDU)and ServiceUnit (CSU)using fittingsappropriatefor the 1/4”-OD Poly-Flo® (air/nitrogen/vacuum) and1/4”-ODFEP (solvent) tubing.

SOLVENT port The SDU SOLVENT IN port supplies replacement solvent from theuser’s supplied solvent container to the Solvent Dispensing Unit. Insertone end of the 1/4”-OD Poly-Flo® solvent tube into the user’s washing

SDU solvent and grounding terminal connections

Page 23: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

19

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

solvent vessel. The other end of the tube should be attached to the SDUSOLVENT IN ‘A’ port. If using an optional drying solvent, insert oneend of the 1/4”-OD Poly-Flo® solvent tube into the user’s washingsolvent vessel. The other end of the tube should be attached to the SDUSOLVENT IN ‘B’ port.

SOLVENT OUT port(s) The SOLVENT OUT ports dispense solvent from the SDU to the CAV-2100 head unit upper rear panel connections. Use 1/4”-OD FED tubing tocomplete these connection to the appropriate solvent intake on the upperrear panel of the CAV Bath Unit (see APPENDIX photography andassembly diagram).

NOTES Single-solvent systems should be plumbed to the CAV bath unit solventport A. The Port B Bath Unit connection must be capped at each BathUnit. For single-solvent systems, this is done at the factory. Use only1/4”-OD FED tubing for solvent connections.

Waste can connectionsThe customer-supplied waste can collects solvent and sample residuedrained from the viscometer tubes during the cleaning process. CAN-NON® Instrument Company does not supply a waste can. Using theprovided 1/2”-OD Nylo-Seal® tubing, connect the CAV drain ports foreach viscometer to a suitable safe container, conforming to local safetyregulations. The drain system operates by gravity flow.

CAUTION The drain line must be kept short and neat (no loops or knots) or an airlock could result.

Procedure 1. Attach the “T” connector to the Left and Right copper Drain portson the lower rear panel of each Bath Unit. Tighten the Nylo-Seal®

fittings finger tight and secure by turning an additional full turn withan adjustable wrench.

2. Connect the waste can to the connector with an appropriate length of½” Nylon® tubing (see previous CAUTION statement).

NOTE Make certain that the connection is oriented in a manner which permitsgravity flow of waste from all outlets directly to the waste can.

3. Ventilate the waste can to the user’s ventilation system.

WARNING Do not store the waste can in an area in which temperatures exceed40°C. Do not store near open flames. Do not store in an enclosed area,where fumes and vapors may build up.

Drip pan drain A separate bulkhead connector is available at the rear of the CAV BathUnit for drainage of the drip pan. Attach a length of drain hose to theconnector and route the hose to a safe solvent/sample container (con-forming to local safety regulations), making certain that gravity flow isunimpeded (no loops or knots).

Page 24: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

20

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Emptying the waste canCheck the volume of the waste can at the beginning of every work shift,to avoid solvent/sample spillage or overflow.

Procedure 1. Disconnect the waste can from the drain tube(s).

CAUTION Be careful when handling the drain tube, to avoid spilling or splashingwaste solvent collected in the tube. Take safety precautions whenworking with the solvent lines. This includes using safety glasses andprotective gloves.

2. Pour the collected solvent out of the waste can into a proper wastestorage container.

3. Replace the drain tube(s) in the waste can.

WARNING Waste solvents can present possible environmental and health hazards.Dispose of all waste solvents according to environmental safety regula-tions.

Checking the solvent level Check the solvent liquid level frequently. Refill the solvent container asnecessary.

Adjusting pneumatic controls

Control pneumatics are adjusted from the CAV pneumatic drawer at thebottom of the Bath Unit. A key is required for access to pneumaticcontrols through the hinged front panel. These controls are normally setduring installation and seldom require adjustment.

Regulation of tube drying The NVIS regulator (NVIS gauge) regulates the compressed air/nitrogensupply used to dry the viscometer tubes. The pressure should be setbetween 20-30 psi. Pull the control knob out, then rotate the knob toadjust the pressure setting. When done, push the knob in to lock at thedesired setting.

Other viscometer controls Pneumatics for both viscometer tubes have four corresponding controls(ADVANCE, AIR RETURN, LOAD and SHIFT) located on the innerpneumatic drawer panel of the CAV. These controls adjust the air pres-sure used to regulate the speed at which respective mechanisms move.

Page 25: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

21

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CAV pneumatics drawer controls

The following table provides an explanation of the function of each panelcontrol:

stnemtsujdarotautcarewardscitamuenpVACECNAVDA yartelpmasehthcihwtadeepsehtslortnoC

.noitisopelpmastxenehtotsecnavda

RIANRUTER

ECNAVDAehtrofdeepsyrevocer/nruterehtslortnoC.srednilycDAOLdna

DAOL mroftalpelpmasehthcihwtadeepsehtslortnoC.pitebutehtotlaivelpmasehtsesiar

TFIHS revieceretsawehthcihwtadeepsehtslortnoC.ebutretemocsivehtrednudrawrofsevom

The speed of these mechanisms should be set to allow them to functionsmoothly and efficiently without spilling or splashing the sample. Turnthe control knob clockwise to increase the pressure. Turn the controlknob counterclockwise to decrease the pressure.

Adjust the tray/table pneumatic mechanisms in the following manner:

Set the AIR RETURN regulators to the minimum setting to allow forreturn/retraction of the LOAD and ADVANCE cylinders. Set the corre-sponding LOAD and ADVANCE regulators slightly higher than the AIRRETURN regulators to ensure that the cylinders can extend fully.

After adjustments are complete, close the hinged access panel.

NOTE Closing the access panel will also lock it.

Page 26: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

22

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Installing VISCPRO® software

To install the VISCPRO® software, follow the instructions below in thesequence presented. Make certain that you complete the sections onchecking instrument settings and calibration data. If you encounterdifficulties at any stage in the installation process, call CANNON®

service at 814-353-8000.

Computer requirementsIBM-compatible computer with pentium processor and current Windows®

OS installed and at least one available serial port (COM 1 or COM 2).See the computer specification sheet or contact CANNON for currentcomputer specfiications.

Windows ® installation1. Turn on your computer. Wait for the Windows® software to load.

2. From the Windows® Start Bar click Settings/Control Panel. Insertthe first VISCPRO™ installation disc or CD-ROM into the appropri-ate drive.

3. Double-click the Add/Remove Programs icon and follow theWindows prompts to complete the installation procedure. Theexecutable file for VISCPRO™ software installation is SETUP.EXE.

NOTE Store the disc in a secure location after you have completed the installa-tion.

Installation actionsThe installation program will:

create a directory for your data files. The default directory isC:\Program Files\Cannon Instrument\VISCPRO).write SETUP information to the Windows® registry.copy the software executable file and other necessary files to thedirectory you specify.update other files in your Windows® directories to versions fullycompatible with the current VISCPRO™ software.place a shortcut icon for the VISCPRO™ executable file on yourWindows® desktop.

Page 27: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

23

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CHAPTER

3 Software configuration

Each on-line CAV instrument must be individually configured.

Original installation Your software installation package may have included pre-set instrumentconfiguration data on an accompanying CD-ROM. Follow the instructionsincluded with the disc to copy configuration information to theVISCPRO™ directory on your computer immediately following initialinstallation of the CAV.

CAUTION Copying existing disc information to your VISCPRO™ directory will overwriteexisting sample data. If you wish to restore the original configuration,make certain to archive your sample data before doing so (see Chapter 15for information on using the Database Manager software).

Follow the procedures in the next several sections of this chapter toverify/edit the instrument settings to ensure that they conform to theactual characteristics of your CANNON® instrument.

Configuration protection To check the configuration settings for your instrument(s), you must login to the security system as a manager. The software is installed with adefault Manager account. This account has no password, allowing anyoperator access to manager-level software functions as long as thepassword is not activated/changed. If you would like to engage the full-release security options, see Security Options in Chapter 4 of the CAVInstruction & Operation Manual.

Logging in1. Use your mouse to click Main from the VISCPRO™ menu bar.

2. Click Log In from the Main menu options.

3. Click on the (arrow) on the right side of the User Name: list boxto display the list of registered users.

Page 28: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

24

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

4. Click Manager. Do NOT enter a password!

5. Click OK. The Log In window will close automatically and you willbe logged in as management personnel.

Checking Instrument Settings1. Use your mouse to click (select) Configure from the VISCPRO™

menu bar.

2. Select your instrument from the list of available instruments (theremay be only one instrument in the list).

3. Select option 2, Instrument Settings, from the list of configurationoptions. The Instrument Settings window will appear.

You will use the Instrument Settings window (see below) to describeand control CAV instrument operational features. These settings affectthe instrument as a whole. Check the instrument settings for your instru-ment per the instructions below, and make any necessary changes:

The Instrument Settings window

Click on the Dual Solvent Option check box if the instrument has beenconfigured with dual solvent capability. If this option is checked, addi-tional options will be available in other windows for configuring CAVtrays and tests.

Use the ID field to input instrument identification information using up to30 alphanumeric characters.

Page 29: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

25

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Use the S/N: field to input the four-digit serial number from the label onthe bottom right corner of the CAV rear service panel.

If you have a dual-solvent unit, click on this check box to use bothsolvents (A and B) during CAV operation. Solvent wash characteristicsare set using the Configure/Tray Settings: Tube and Bath options.

Click on the LAS Shuttle Option check box if the instrument has beenconfigured with the Laboratory Automation System Shuttle. This devicereplaces the sample tray/advance mechanism with a robotic arm capableof being controlled by the CAV software.

Click on this check box to synchronize the timing of wash cycles forviscometers in the bath. (While slowing the throughput rate, this optionmay enhance bath temperature stability during CAV testing for someapplications).

Click on this check box to automatically abort CAV testing if air pressuredrops below the minimum threshold required for CAV operation.

When you have verified all settings, click OK.

Tray Settings: Tube and Bath

Use the Tray Settings: Tube and Bath window (see graphic, next page)to describe and control test parameters for each individual viscometertube. Alter the settings using the instructions below. When you haveachieved the desired configuration, save the current instrument settingsby clicking OK. To exit the configuration screen without saving yourchanges, click Cancel. To restore software defaults, click Defaults.

To complete the tray configuration, click Configuration/[your instru-ment ID]/Tray Settings: Tube and Bath. Then select the desired tubenumber by clicking on the desired tube tab # at the top of the window.Tubes are numbered from left to right per their placement in the CAVtemperature bath.

Use the S/N: field to input the identifying tube serial number (you maywish to use the serial number handwritten on the viscometer tube).

NOTE For reporting and calibration purposes, tube serial numbers should beunique for all instruments controlled by the VISCPRO® software.

Click on the Fast Run Tube check box if the viscometer tube is the fastrun (dual bulb) type.

Page 30: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

26

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

The Tray Settings: Tube and Bath window

NOTE Settings for soak times (see below) can be configured for each bathtemperature setting.

Selecting sample run determination settings

Determinations Select the maximum number of determinations which can be run for thetube samples (values from 1 to 20 are acceptable). Enter the maximumacceptable difference in calculated results when using multiple determi-nations (the default setting is 0.35 percent). Select the radio button which corresponds to the preferred test methodology.

NOTES The CAV determines the sample flow time by averaging the values of thetwo sample runs which correspond to the chosen methodology.

If the Always perform maximum determinations radio button is selected,the last two flow times will be used for the viscosity determination.

Page 31: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

27

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Soak time Input the desired sample soak times for each bulb of the viscometer tube.The soak time is the length of time in which the sample will remain in theviscometer tube before being dropped. The soak time is important toensure temperature equilibration of the sample. Values between 5 and300 seconds are acceptable (the default value is 120 seconds).

Constants The Tray Settings: Tube and Bath screen displays the calibrationconstants (C and E) for each bulb of the viscometer tube. These valuescannot be edited from the Tray Settings . . . screen. For information ondetermining bulb constants for viscometer tubes, see Calibration later inthis chapter.

NOTE When you have completed the configuration and saved the current traysettings, you may wish to permanently store the current instrumentsettings using the Save Instrument option from Main.

Tray Settings: Wash

Use the Tray Settings: Wash window to determine wash settings.Different wash settings may be selected for each CAV tube individually.

Automatic calculation . . . The Automatic calculation of wash rinses check box is used to restorethe software default value for rinse cycles. These settings are recom-mended by CANNON®.

Solvent rinse settings The solvent rinse settings for Solvent A and Solvent B are determinedusing the Tray Settings: Wash window. These settings determine thenumber of rinse cycles performed by the CAV instrument after each test.

Solvent fill time(s) The Solvent fill time(s) setting determines the amount of time thatsolvent will be introduced into the viscometer tube during each rinsecycle. Acceptable values are 1 to 30 seconds.

Page 32: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

28

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

The Tray Settings: Wash window

Solvent soak time(s) The Solvent soak time(s) setting determines the amount of time thatsolvent will remain in the viscometer tube prior to the solvent flushduring each rinse cycle. Acceptable values are 1 to 30 seconds.

Solvent flush time(s) The Solvent flush time(s) setting determines the amount of time pressur-ized air will flush the remaining solvent from the viscometer tube followingthe solvent soak. Select a value from 1 to 60 seconds for this option.

Final dry time(s) The Final dry time(s) setting determines the amount of time forced air ischanneled through the viscometer tube to dry it at the conclusion of therinse cycle. Select a value from 15 to 300 seconds for this option.

Testing the configurationWhen all CAV units have been configured, verify a successful configura-tion by running a sample in Remote mode for each instrument per manualinstructions.

Page 33: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

29

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CHAPTER

A APPENDIX A—Multi-Unitconfiguration

The CAV instrument is capable of being networked with the controllingcomputer via available RS-485 connections. As many as four CAVinstruments may be connected to a single computer. Follow the instruc-tions in this appendix to complete assembly of the multi-unit networkingkit, available from CANNON®.

Kit components

The multi-unit interface kit contains an RS-232-to-RS-485 converter andRJ-45 cables capable of networking up to four CAV Bath Units to asingle computer.

Procedure

Using the interface kit components, complete the multi-unit configurationas follows:

Connecting componentsRefer to the instructions in the second chapter of this manual to completethe assembly of CAV-2000 components.

NOTE The interface kit was designed for Bath Units placed in an adjacentconfiguration. Custom configuration with longer RJ-45 cables is possible.

Make certain to install the VISCPRO® software on the controllingcomputer. Complete all component assembly before proceeding with RS-485 networking. The solvent and pneumatic connections should beplumbed per the assembly diagram(s) and the diagram on the final pageof this manual.

Completing RS-485 connections

CAUTION 1. Make sure the power is removed from the computer and all instru-ments to be configured.

2. Locate the RS-485 converter and required RJ-45 connecting cable(s).

Page 34: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

30

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

3. Attach the RS-485 converter directly to the computer via the stan-dard 9-pin RS-232 connector. Then attach the longer RJ-45 cablefrom the converter to the IN connector on the serial communicationsinterface on the upper rear panel of the first CAV bath.

4. Attach a shorter RJ-45 cable to the OUT connector on the serialcommunications interface of the first CAV bath. Then attach the otherend of the cable to the IN connector on the adjacent Bath Unit.

5. Continue the procedure from the previous step for each bath consecu-tively until you have completed RS-485 connections between eachBath Unit and the previous unit in the sequence (see diagram).

------ ------ ------ ------ -----

6. Note the serial number of each Bath Unit, marked on the labeladjacent to the mains connection. You will use this information incompleting VISCPRO® software configuration for the multi-unitinterface.

7. Restore power to each CAV Bath Unit and press the MENU key toaccess configuration parameters for each bath. Then select Instru-ment Address from the list of options. Press the key corresponding tothe desired instrument address (0-9); then press ENTER to save theaddress.

NOTE Each instrument address must be unique. Remember the address givento each instrument. You will require this information when configuringeach bath for automatic testing using the VISCPRO® software.

When configuring the VISCPRO® software for automatic operation ofthe CAV instruments, make certain to enter the appropriate RS-485address for each corresponding Bath Unit, as well as the Bath Unit serialnumber.

Page 35: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

31

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CHAPTER

CSU-200CANNONService Unit andSDU-200SolventDispensing Unitconnections

CAV bath unit solvent connections

B APPENDIX B—Tubingconnections/photography

Page 36: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

32

CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup GuideVersion 1.3g— August, 2011; CANNON® Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

CAV-2000 upper rear panel solvent connections (RS-485 multi-bath communication cable alsovisible) NOTE: Copper tubing may be replaced with FEP

CAV-2000 lower rear panel with exhaust manifolds and power cable in place

Page 37: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup
Page 38: CAV®-2100/2200 Automatic Viscometer...CAV-2000 single and dual-bath installation on pedestal base 6 CANNON® Automatic Viscometer Models CAV-2100 and CAV-2200 Installation & Setup

2139 High Tech Road | State College, PA 16803 | USA800-676-6232 | 814-343-8000 | Fax: 814-353-8007

[email protected] | www.cannoninstrument.com