Bardy Diagnostics ® Table of Contents Instructions For Use Carnation Ambulatory Monitor CAM Instructions For Use The Carnation Ambulatory Monitor is a continuously recording P-wave centric ambulatory ECG patch monitor that records for up to the prescribed wear time. IN THE BOX Indications for Use The Carnation Ambulatory Monitor (CAM) is designed to provide extended duration cardiac monitoring for people who are suspected of having cardiac arrhythmias. Patient Population The intended patient population includes both males and females not weighing less than 10 kg (22 lbs) who may have cardiac arrhythmias. TABLE OF CONTENTS Instructions for Use Prepare the Skin Prepare the CAM Apply the CAM Record Symptoms Link the CAM to Patient Information Operational Instructions General Cautions Processing Cautions EMC Guidelines Symbols Technical Specifications PREPARE THE SKIN Step 1 Remove all hair over sternal area by shaving close to the skin. Do not merely clip hair. The prepared area should extend 2 inches past where the CAM will be placed. Step 2 Use all prep pads provided in the box to clean area shown. SCRUB the skin with the prep pads until they appear clean after use. Skin should be scrubbed well enough to be slightly reddened. Allow the skin to dry for 1 minute prior to applying. PREPARE THE CAM Step 3 On a flat, hard surface insert the narrow end of the Recorder into the Battrode first, and then click the Recorder into place with the event button facing up. Page 2 3 4 4 5 6 6 9 10 11 13 Battrode (packaged inside CAM box) Recorder (packaged inside CAM box) Patient Diary and Quick Reference Instructions (inside CAM box) 1 2 Instructions For Use 5 Instructions For Use 4 Instructions For Use 3 Cautions 8 Cautions 7 Cautions 6 prep pads CAUTION: Proper skin prep required to achieve full length of prescribed monitoring duration. A green LED light will blink for 10 seconds to confirm activation. Following the LED activation sequence, push down on the event button once to ensure the Recorder is clicked firmly in place. If the LED does not blink, contact Customer Service. NOTE: The activated 10 second LED blinking will only occur while Recorder is connected to the Battrode. No additional blinking should occur while the CAM is being worn. Contact Customer Service if the LED continues to blink. Step 4 Record the date and time of CAM activation, which is a required field to complete patient registration. Step 5 Gently peel the liner from the CAM by grasping the tab at the top of the device and peeling downward, carefully avoiding contact with the adhesive. OPERATIONAL INSTRUCTIONS If the CAM becomes soiled, patient may gently wipe exterior of device, using a clean, dry cloth. If Battrode becomes loose or detached from skin, patient should press the adhesive portions of monitor back in place. If you do not see the confirmation blinking LED upon initial connection of the Battrode and Recorder, contact Bardy Diagnostics Customer Service. Remind the patient to return the monitor following the prescribed period of wear. SAFETY ALERT DESCRIPTIONS The symbol shown below identifies a potential hazard category. CAUTION This alert identifies hazards that may cause minor personal injury, product damage, or property damage. GENERAL CAUTIONS This section lists general cautions. Those pertaining to specific functions and procedures are included in the text where appropriate. CAUTION: Contains ECG Electrodes, which can damage skin if used improperly. The CAM should be used by or in consultation with a health care provider familiar with its proper placement and use. Apply electrodes to intact, clean skin only. Do not apply over an open wound, lesion, infected or inflamed skin. CAUTION: Damage to skin Instruct patients to remove electrodes carefully to avoid damaging their skin. CAUTION: Post-application activity Instruct patients to avoid showering, bathing, or exercise immediately following application, and thereafter avoid activities or environments that result in excessive perspiration, as this may result in a decreased period of monitoring. CAUTION: CAM adhesive swelling It is normal for the CAM adhesive material to swell in humid environments or when exposed to moisture. Instruct patients to allow adhesive to dry following activities such as showering or exercise. If desired, patient can gently pat with a dry towel, but should not attempt to reposition the CAM. CAUTION: Submersion Submersion (such as during swimming or bathing) is not advised. Instruct patients to keep showers brief and the CAM out of the direct stream of water. CAUTION: Poor skin contact Poor contact of the CAM with the skin can negatively affect monitoring performance. Instruct patients to secure the CAM back in place if it becomes loose or detached. CAUTION: Skin irritation Patients with sensitive skin or with known skin conditions should use the CAM with caution. If irritation such as redness, severe itching or allergic symptoms (i.e. hives) develop, instruct, patients to remove the CAM immediately and have them contact their physician. CAUTION: Allergic skin reaction Do not use the CAM on patients with known skin allergies or family history of skin allergies. APPLY THE CAM Step 6 Locate the bone at the bottom of the sternum. This is the xiphoid process. Apply the CAM to the patient’s sternum with the bottom part sitting over the xiphoid process. Rub firmly around the edges of the patch for 1 minute to ensure adhesion. Place two fingers under the event button and press down firmly to adhere the top of the CAM to the patient's chest. RECORD SYMPTOMS Step 7 Instruct patients to gently press the button only once each time they feel symptoms, and record the date/time in the Patient Diary (included). Do not press button repeatedly or forcefully. REVIEW Step 8 Give the patient the Patient Diary and CAM box and explain their use. Review all instructions and Cautions and advise patient to avoid showering, bathing, or exercise immediately following application, and thereafter avoid activities or environments that result in excessive perspiration, as this may result in a decreased period of monitoring. LINK THE CAM TO PATIENT INFORMATION Step 9 Barcode stickers with the CAM identification number are found inside the CAM box and must be placed on the patient’s chart or typed into the patient’s EMR. IMPORTANT: Fill in all information on the front page of the Patient Diary, including the patient name, date/time when the CAM is activated, physician name and hospital/clinic. Step 10 Register the patient as early as possible on the BDxCONNECT patient management portal. Bottom of CAM over xiphoid process xiphoid process CAUTION: Inability to diagnose if patient and physician data is missing. The Patient Diary card must be kept with the CAM to ensure proper diagnosis. XXXXX-XXXXX Lorem XXXXX-XXXXX Lorem XXXXX-XXXXX Lorem XXXXX-XXXXX Lorem XXXXX-XXXXX Lorem PATIENT HEALTH RECORD CAUTION: Choking Hazard This is a prescribed medical device. Keep device and packaging away from young children. CAUTION: Equipment damage Use only as directed. Tampering with the CAM may render it unusable. There are no user serviceable parts. CAUTION: May interfere with defibrillation therapy To avoid ineffective defibrillation therapy, remove the CAM before applying any external cardiac defibrillators. CAUTION: May interfere with MRI scanning The CAM is not intended for use with MRI equipment and could reduce effectiveness of diagnostic scanning. Remove before any MRI scan. CAUTION: Ineffective electronic imaging The CAM is not intended for use with imaging equipment and could reduce effectiveness of diagnostic imaging. Remove before using electronic imaging systems. CAUTION: Equipment damage Use only with approved Bardy Diagnostics BDxStation. CAUTION: Enclosure damage Do not use any parts that appear damaged. Check the CAM for damage before using and reject any parts that have been damaged in shipping. CAUTION: Electrical shock The CAM contains electrodes for monitoring ECG. Do not allow conductive parts of the electrodes on the CAM to come in contact with other conductive parts (including earth). Bardy Diagnostics CAMQuick Reference Instructions Carnation Ambulatory MonitorLED CAUTION: Touching the adhesive can reduce adhesive performance. Hold onto tabs at the end of the CAM. CONNECT / / : 2.If you feel symptoms that may be related to your heart, gently push the button and record date/time in this diary. Do notpush button repetitively or forcefully. CAUTION: If your CAM becomes dislodged, contact your physician for assistance. PATIENT NAME: PHYSICIAN NAME: HOSPITAL/CLINIC: DEVICE APPLIED (Required): Prescribed Wear Time: Days Date/Time: 1. Following your normal daily activities, wear the CAM for the amount of time prescribed by your physician. Wearing the Carnation Ambulatory Monitor (CAM): Place barcode label here Carnation Patient Diary / / : AM PM MM DD YY HH MM 2. If you feel symptoms that may be related to your heart, gently push the button and record date/time in this diary. Do not push button repetitively or forcefully. CAUTION: If your CAM becomes dislodged, contact your physician for assistance. PATIENT NAME: PHYSICIAN NAME: HOSPITAL/CLINIC: DEVICE APPLIED (Required): Prescribed Wear Time: Days Date/Time: 1. Following your normal daily activities, wear the CAM for the amount of time prescribed by your physician. Wearing the Carnation Ambulatory Monitor (CAM): Place barcode label here Carnation Ambulatory Monitor Patient Diary PUSH ONLY ONCE when you feel SYMPTOMS

CAM AME: tion AME: Bardy Diagnostics IED (Required):Days ... · CAM back in place if it becomes loose or detached. CAUTION: Skin irritation Patients with sensitive skin or with known

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: CAM AME: tion AME: Bardy Diagnostics IED (Required):Days ... · CAM back in place if it becomes loose or detached. CAUTION: Skin irritation Patients with sensitive skin or with known

Bardy Diagnostics® Table of Contents Instructions For Use

CarnationAmbulatory Monitor™

CAM™

InstructionsFor Use

The Carnation Ambulatory Monitor is a continuously recording P-wave centric™ ambulatory ECG patch monitor that records for up to the prescribed wear time.

IN THE BOX

Indications for UseThe Carnation Ambulatory Monitor (CAM) is designed to provide extended duration cardiac monitoring for people who are suspected of having cardiac arrhythmias.

Patient PopulationThe intended patient population includes both males and females not weighing less than 10 kg (22 lbs) who may have cardiac arrhythmias.

TABLE OF CONTENTSInstructions for Use

Prepare the SkinPrepare the CAMApply the CAMRecord SymptomsLink the CAM to Patient InformationOperational Instructions

General CautionsProcessing CautionsEMC GuidelinesSymbolsTechnical Specifications

PREPARE THE SKIN

Step 1Remove all hair over sternal area by shaving close to the skin. Do not merely clip hair. The prepared area should extend 2 inches past where the CAM will be placed.

Step 2Use all prep pads provided in the box toclean area shown. SCRUB the skin withthe prep pads until they appear clean after use. Skin should be scrubbed wellenough to be slightly reddened. Allow theskin to dry for 1 minute prior to applying.

PREPARE THE CAMStep 3On a flat, hard surface insert the narrow end of the Recorder into the Battrode first, and then click the Recorder into place with the event button facing up.

Page

23445669

101113

Battrode™

(packaged inside CAM box) Recorder

(packaged insideCAM box)

Patient Diary and QuickReference Instructions(inside CAM box)

1 2

Instructions For Use 5Instructions For Use 4Instructions For Use 3

Cautions 8Cautions 7Cautions 6

prep pads

CAUTION: Proper skin prep required to achievefull length of prescribed monitoring duration.

A green LEDlight will blinkfor 10 seconds toconfirm activation.Following the LEDactivation sequence,push down on the eventbutton once to ensure theRecorder is clicked firmly in place.If the LED does not blink, contact Customer Service.

NOTE: The activated 10 second LED blinking will only occur while Recorder is connected to the Battrode. No additional blinking should occur while the CAM is being worn. Contact Customer Service if the LED continues to blink.

Step 4

Record the date and time of CAM activation, which is a required field to complete patient registration.

Step 5Gently peel the liner from the CAM by grasping the tab at the top of the device and peeling downward, carefully avoiding contact with the adhesive.

OPERATIONAL INSTRUCTIONSIf the CAM becomes soiled, patient may gently wipe exterior of device, using a clean, dry cloth.

If Battrode becomes loose or detached from skin, patient should press the adhesive portions of monitor back in place.

If you do not see the confirmation blinking LED upon initial connection of the Battrode and Recorder, contact Bardy Diagnostics Customer Service.

Remind the patient to return the monitor following the prescribed period of wear.

SAFETY ALERT DESCRIPTIONSThe symbol shown below identifies a potential hazard category.

CAUTIONThis alert identifies hazards that may cause minor personal injury, product damage, or property damage.

GENERAL CAUTIONSThis section lists general cautions. Those pertaining to specific functions and procedures are included in the text where appropriate.

CAUTION: Contains ECG Electrodes, which can damage skin if used improperly. The CAM should be used by or in consultation with a health care provider familiar with its proper placement and use. Apply electrodes to intact, clean skin only. Do not apply over an open wound, lesion, infected or inflamed skin.

CAUTION: Damage to skinInstruct patients to remove electrodes carefully to avoid damaging their skin.

CAUTION: Post-application activityInstruct patients to avoid showering, bathing, or exercise immediately following application, and thereafter avoid activities or environments that result in excessive perspiration, as this may result in a decreased period of monitoring.

CAUTION: CAM adhesive swellingIt is normal for the CAM adhesive material to swell in humid environments or when exposed to moisture. Instruct patients to allow adhesive to dry following activities such as showering or exercise. If desired, patient can gently pat with a dry towel, but should not attempt to reposition the CAM.

CAUTION: SubmersionSubmersion (such as during swimming or bathing) is not advised. Instruct patients to keep showers brief and the CAM out of the direct stream of water.

CAUTION: Poor skin contactPoor contact of the CAM with the skin can negatively affect monitoring performance. Instruct patients to secure the CAM back in place if it becomes loose or detached.

CAUTION: Skin irritationPatients with sensitive skin or with known skin conditions should use the CAM with caution. If irritation such as redness, severe itching or allergic symptoms (i.e. hives) develop, instruct, patients to remove the CAM immediately and have them contact their physician.

CAUTION: Allergic skin reactionDo not use the CAM on patients with known skin allergies or family history of skin allergies.

APPLY THE CAMStep 6Locate the bone at the bottom of the sternum. This is the xiphoid process.

Apply the CAM to the patient’s sternum with the bottom part sitting over the xiphoid process. Rub firmly around the edges of the patch for 1 minute to ensure adhesion. Place two fingers under the event button and press down firmly to adhere the top of the CAM to the patient's chest.

RECORD SYMPTOMSStep 7Instruct patients to gentlypress the button only onceeach time they feel symptoms,and record the date/time inthe Patient Diary (included).Do not press button repeatedlyor forcefully.

REVIEWStep 8Give the patient the Patient Diary and CAMbox and explain their use. Review allinstructions and Cautions and advise patientto avoid showering, bathing, or exerciseimmediately following application, andthereafter avoid activities or environmentsthat result in excessive perspiration, as thismay result in a decreased period of monitoring.

LINK THE CAM TO PATIENT INFORMATION

Step 9Barcode stickers with theCAM identification number are found inside the CAMbox and must be placedon the patient’s chart ortyped into the patient’s EMR.

IMPORTANT: Fill in all information onthe front page of the Patient Diary,including the patient name, date/time when the CAM is activated, physician name and hospital/clinic.

Step 10Register the patient as early aspossible on the BDxCONNECT patient management portal.

Bottom of CAMover xiphoid process

xiphoidprocess CAUTION: Inability to diagnose if patient and physician

data is missing. The Patient Diary card must be kept withthe CAM to ensure proper diagnosis.

XXXXX-XXXXX

LoremXXXXX-XXXXX

LoremXXXXX-XXXXX

LoremXXXXX-XXXXX

LoremXXXXX-XXXXX

Lorem

PATIENT HEALTH RECORD

CAUTION: Choking HazardThis is a prescribed medical device. Keep device and packaging away from young children.

CAUTION: Equipment damageUse only as directed. Tampering with the CAM may render it unusable. There are no user serviceable parts.

CAUTION: May interfere with defibrillation therapyTo avoid ineffective defibrillation therapy, remove the CAM before applying any external cardiac defibrillators.

CAUTION: May interfere with MRI scanningThe CAM is not intended for use with MRI equipment and could reduce effectiveness of diagnostic scanning. Remove before any MRI scan.

CAUTION: Ineffective electronic imagingThe CAM is not intended for use with imaging equipment and could reduce effectiveness of diagnostic imaging. Remove before using electronic imaging systems.

CAUTION: Equipment damageUse only with approved Bardy Diagnostics BDxStation.

CAUTION: Enclosure damageDo not use any parts that appear damaged. Check the CAM for damage before using and reject any parts that have been damaged in shipping.

CAUTION: Electrical shockThe CAM contains electrodes for monitoring ECG. Do not allow conductive parts of the electrodes on the CAM to come in contact with other conductive parts (including earth).

Bardy Diagnostics

CAM™

Quick ReferenceInstructions

CarnationAmbulatory Monitor™

LED

CAUTION: Touching the adhesive canreduce adhesive performance. Hold ontotabs at the end of the CAM.

CONNECT

/ / : AM PMMM DD YY HH MM

2. If you feel symptoms that may berelated to your heart, gently push thebutton and record date/time in thisdiary. Do not push button repetitivelyor forcefully.

CAUTION: If your CAM becomesdislodged, contact your physicianfor assistance.

PATIENT NAME:

PHYSICIAN NAME:

HOSPITAL/CLINIC:

DEVICE APPLIED (Required):Prescribed Wear Time: Days

Date/Time:

1. Following your normal daily activities, wear the CAMfor the amount of time prescribed by your physician.

Wearing the Carnation Ambulatory Monitor (CAM):

Place barcode label here

CarnationAmbulatory Monitor™

Patient Diary

PUSH ONLY ONCEwhen you feelSYMPTOMS

/ / : AM PMMM DD YY HH MM

2. If you feel symptoms that may berelated to your heart, gently push thebutton and record date/time in thisdiary. Do not push button repetitivelyor forcefully.

CAUTION: If your CAM becomesdislodged, contact your physicianfor assistance.

PATIENT NAME:

PHYSICIAN NAME:

HOSPITAL/CLINIC:

DEVICE APPLIED (Required):Prescribed Wear Time: Days

Date/Time:

1. Following your normal daily activities, wear the CAMfor the amount of time prescribed by your physician.

Wearing the Carnation Ambulatory Monitor (CAM):

Place barcode label here

CarnationAmbulatory Monitor™

Patient Diary

PUSH ONLY ONCEwhen you feelSYMPTOMS

Page 2: CAM AME: tion AME: Bardy Diagnostics IED (Required):Days ... · CAM back in place if it becomes loose or detached. CAUTION: Skin irritation Patients with sensitive skin or with known

We’re close to your heart™

Symbols 11EMC Guidelines 10Cautions 9

Technical Specifications 14Technical Specifications 13Symbols 12

Technical Specifications 15

EMC GUIDELINESThe CAM requires special cautions regarding EMC and needs to be put into service according to provided information.

CAUTION: Electromagnetic interference (EMI)Portable and mobile wireless communications equipment such as wireless home network devices, mobile phones, cordless telephones and their base stations, and walkie-talkies can affect the CAM and should be kept at least a distance of 30 centimeters away from the equipment. Position the CAM away from other electrical or electronic equipment, if possible. The presence of strong EMI fields, or electronic, surgical, or diathermy instruments in close proximity to the ECG device may cause trace noise or input overload conditions.

The ECG device is compliant with IEC 60601-1-2 EMC immunity requirements.

SYMBOLSBardy Diagnostics products display one or more of these symbols and warning labels.

SYMBOL DESCRIPTION

Attention: Consult accompanying documents

Warning and caution symbol

TYPE BF APPLIED PART: F-TYPE APPLIED PART complying with the specified requirements of 60601-1 to provide a higher degree of protection against electronic shock than that provided by TYPE B APPLIED PARTS.

Note: TYPE BF APPLIED PARTS are not suitable for DIRECT CARDIAC APPLICATION.

Protected against intrusion from fingers and small objects, and from water falling as a spray at an angle of up to 60 from vertical. Do not immerse in bathtub or swimming.

Sold by prescription onlyCAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare practitioner.

Single use only

Manufacturer

Use-by date

DESCRIPTION

Do not expose to temperatures outside of these limits. For more information on environmental parameters refer to the Technical Specifications section.

Atmospheric pressure must be within these limits. For more information on environmental parameters refer to the Technical Specifications section.

Humidity must be within these limits. For more information on environmental parameters refer to the Technical Specifications section.

Date of manufacture

Contains electronic equipment. Dispose of properly in accordance with local regulations.

Serial number

Orderable part number

Batch code

MR Unsafe

CE Mark signifying conformity to EU regulations

EU authorized representative

IP23

131°F(55°C)

32°F(0°C)

SYMBOL

SN

10%

95%

500hPa

1060hPa

SPECIFICATIONITEM

TECHNICAL SPECIFICATIONS

Performance Characteristics

ECG channels

Recording capacity

Recording format

Service life

Shelf life

Electrical Characteristics

Frequency response

Differential range

A/D sampling rate

Power Requirements

Battery type

Lithium content

Heavy metal content

UN compliance

1 channel

Up to 2 or 7 days

Continuous

Up to 2 or 7 days

24 months

0.67 Hz to 25 Hz

4 mV

171 Hz

Lithium primary (coin cell)

Lithium content < 1 g

Within weight limits of 2006/66/EC

Complies with UN 3090

SPECIFICATIONITEM

Physical Characteristics

Approximate dimensions

Weight

Enclosure material

Flammability rating

Classification

Type of protection

Degree of protection

Protection againstobjects and wateringress

Electrode Characteristics

Number of electrodes

Type

Supplied as

Lead wire length

Materials

178mm x 38mm x 14mm

<25g

Medical grade thermoplastic polymer

UL-HB

Internally powered

Type BF applied part

IP23 (Protected against intrusion from fingers and small objects, and from water falling as a spray at an angle of up to 60° from vertical)

2

Electrode incorporating electrode gel and internal lead wire

Disposable, non-sterile

11.6 cm (no patient contact)

Electrode gel: Medical grade conductive synthetic

Adhesive: Medical grade skin adhesive

Environmental Specifications (ECG Device)

Operating temperature

Operating pressure

Operating humidity

Transport temperature

Storage temperature

Transport / Storagehumidity

Transport / Storage pressure

Standards compliance

50° F to 113° F(10° C to 45° C)

700 to 1060 hPa

10% to 95%(non-condensing)

14° F to 130° F(-10° C to 55° C)

59° F to 77° F(15° C to 25° C)

10% to 95%(non-condensing)

500 to 1060 hPA

Applicable sections of IEC60601-1, 60601-1-2,60601-1-11, 60601-2-47

CarnationAmbulatory Monitor

DWG000474B 09/2019www.bardydx.com

By prescription only

Read all instructionsbefore using this product

For assistance in setting up, using, and maintaining the device, or to report unexpected operations or events:

Bardy Diagnostics, Inc.® 316 Occidental Ave S, Ste B310Seattle, WA 98104USA

This device is provided non-sterile.

For additional instructions andFrequently Asked Questions

visit www.bardydx.com

Instructions For Use

Find us onsocial media

US Customer Service:(844) 422-7393

PROCESSING CAUTIONSWhen the CAM is returned to the processing center, these additional cautions apply.

CAUTION: Contaminated surfacesWhen the CAM is returned to the reading center, it will have been in contact with human skin. Follow the facility’s procedures for appropriate handling.

CAUTION: Battery may present environmental hazardThe CAM contains a battery. Properly dispose of batteries in accordance with local regulations.

CAUTION: Electronic waste may present environmental hazardThe CAM is considered medical electrical equipment. Properly dispose of electronic waste in accordance with local regulations.

CAUTION: Single use onlyThe Bardy Diagnostics CAM is not intended for reuse, as the monitor becomes non-functional after the first use.

Medical Product Service GmbHBorngasse 2035619 BraunfelsGermany

EU Customer Service:+31 (0)46 7630422

0459

Page 3: CAM AME: tion AME: Bardy Diagnostics IED (Required):Days ... · CAM back in place if it becomes loose or detached. CAUTION: Skin irritation Patients with sensitive skin or with known

Electromagnetic Emissions Declarations

Guidance and manufacturer’s declaration – electromagnetic emissions

The CAM is intended for use in the electromagnetic environment specified below. The customer or the user of the CAM should assure that itis used in such an environment.

Emissions test

RF emissionsCISPR 11

RF emissionsCISPR 11

Harmonic emissionsIEC 61000-3-2

Voltage fluctuations/flicker emissionsIEC 61000-3-3

Guidance and manufacturer’s declaration – electromagnetic immunity

The CAM is intended for use in the electromagnetic environment specified below. The customer or the user of the CAM should assure that itis used in such an environment.

IMMUNITY test

Electrostaticdischarge (ESD)IEC 61000-4-2

Power frequency(50/60 Hz) magnetic fieldIEC 61000-4-8

RadiatedRF IEC 61000-4-3

Radiated RFProximity FieldsIEC 61000-4-3(per IEC 60601-1-2 Ed. 4)

Compliance

Group 1

Group B

Not Applicable

Not Applicable

Electromagnetic environment - guidance

The CAM uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The CAM has No AC Mains, and is suitable for use in all establishments, including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

IEC 60601 test level

± 8 kV contact

± 15 kV air

30 A/m

10 V/m80 MHz to 2,7 GHz

9 V/m - 28 V/m perIEC 60601-1-2 Ed. 4,Table 9

Compliance level

± 8 kV contact

± 15 kV air

30 A/m

10 V/m80 MHz to 2,7 GHz

9 V/m - 28 V/m perIEC 60601-1-2 Ed. 4,Table 9

Electromagnetic environment - guidance

Floors should be wood, concrete or ceramic tile. If floorsare covered with synthetic material, the relative humidityshould be at least 30%.

Power frequency magnetic fields should be that of atypical commercial or hospital environment.

Portable and mobile RF communications equipment shouldbe used no closer than 30 cm to any part of the CAM.

Interference may occur in the vicinity ofequipment marked with the following symbol:

Page 4: CAM AME: tion AME: Bardy Diagnostics IED (Required):Days ... · CAM back in place if it becomes loose or detached. CAUTION: Skin irritation Patients with sensitive skin or with known

firmly on patch adhesive for 1 minute to ensure adhesion.

Peel the liner from the back of the Carnation Ambulatory

Use both hands to find the edge of the ribcage whereit meets the abdomen. Slide both hands down towardsthe center of the body until they meet in a little notch.

1

Carnation Ambulatory Monitor

Instructions For Use

Read all instructions before using this product

By prescription only

For assistance in setting up, using, and maintaining the device, or to report unexpected operations or events. Contact us:

2

Table of Contents Instructions for Use

Prepare the site Prepare the Carnation Ambulatory Monitor Apply the Carnation Ambulatory Monitor (CAM)Link the (CAM) to the patient's information Review all instructions in Patient Diary w/patient

General CautionsProcessing Cautions

Symbols Technical Specifications

Page

3

56

79

1113

Recorder (packaged inside Recorder pouch)

BattrodeTM

(packaged inside Battrode pouch)

Return Mailer for patient to return used device containing recording.

Patient Diary and Clinician Instructions for Use

0433541204335412

Carnation Ambulatory Monitor

Battrode Recorder

Carnation Ambulatory Monitor

Address Street, Suite #City, State ZIP-CODE

Table of Contents

EMC Guidelines 10

4

6

3

Instructions For Use

Prepare the Carnation Ambulatory Monitor

STEP 3

Remove the Recorder and Battrode from the packaging.

STEP 4

4

STEP 5

Monitor.

Instructions For Use

Apply the Carnation Ambulatory MonitorSTEP 6

5

Instructions For Use

Link the Carnation Ambulatory Monitor to the patient’s information

STEP 8

CAUTION: Inability to diagnose if patient and physician data is missing. The patient diary card must be kept with the Carnation Ambulatory Monitor to ensure proper diagnosis.

REVIEW ALL INSTRUCTIONS IN THE PATIENT DIARY WITH THE PATIENTGive the patient the Patient Diary and Return Mailer and explain their use. Review all Instructions and Cautions in case there is a change in performance of the Carnation Ambulatory Monitor, such as the device coming loose from the patient’s chest.

6

Instructions For Use

If you do not hear a confirmation buzz, place the monitor upto your ear and press the button to confirm. If no buzz, returnto manufacturer.

If Carnation Ambulatory Monitor becomes loose or detached from skin, patient should press the monitor back in place.

Operational InstructionsIf the Carnation Ambulatory Monitor becomes soiled, patients may gently wipe exterior of device, using a clean, dry cloth.

STEP 10

STEP 9

Barcode stickers with the device’s unique serial number are

RecorderBattrode

A

CLICK*

B*

CAUTION: Proper skin prep required to achieve 48 hours wear.

Prepare the SkinINSTRUCTIONS FOR USE

NOTE:

Tell patient to gently press the button when they feel symptoms, and record the time/date in the Patient Diary.

On a flat, hard surface, snap the Recorder into the

Fill in all information on the front page of the Patient Diary, including start time/date, patient initials, and physician name and phone number.

In the Box:

Be sure that the prepared area extends over 1 inch larger than where the patch will sit.

Battrode by inserting the narrow end first. Within 2-3 seconds you should hear a confirmation buzz. If there is no buzz see Operational Instructions on this page.

CAUTION: Touching the adhesive can reduce adhesive performance. Hold onto the tabs at the ends of the Carnation Ambulatory Monitor.

found inside the mailer box and can be placed on the patient chart, or be typed into the patient's EMR.

The Carnation Ambulatory Monitor is a single patient use, continuous recording ambulatory ECG monitor that records for up to 48 hours.

Indications for UseThe Carnation Ambulatory Monitor is designed to provide extended duration cardiac monitoring for people who are suspected of having cardiac arrhythmias.

Patient PopulationThe intended patient population includes both males and females not weighing less than 10 kg (22lbs) who may have cardiac arrhythmias.

USA

US Customer service: (844) 422-7393 EU Customer service: +31 (0) 46 7630422

STEP 1If hair is present, shave the hair as close to the skin as possible.

STEP 2Use the skin prep pads provided in the mailer box to remove oils from the patient’s chest. Scrub skin with wipe or wipes until wipes remain clean after use. IMPORTANT: Ensure skin is completely dry before monitor application.

Instructions to PatientAvoid activities or environments that result in severe perspiration, as this may result in the device failing to maintain adherence to the skin.

STEP 7 Apply the CAM to the patient’s chest with thebottom of the electrode sitting over the xiphoid. Press

Ask the patient to inhale and exhale – after exhaling you should feel a pointed bone. This isthe xiphoid process.

overxiphoid

prep pads

MPS Medical Product Service GmbHBorngasse 2035619 BraunfelsGermany

Bardy Diagnostics Inc

Bardy Diagnostics Inc316 Occidental Av. S, Ste B310Seattle, WA 98104

Carnation Ambulatory Monitor Instructions For Use DWG00 140H 04/2019

316 Occidental AveB310

Page 5: CAM AME: tion AME: Bardy Diagnostics IED (Required):Days ... · CAM back in place if it becomes loose or detached. CAUTION: Skin irritation Patients with sensitive skin or with known

7

General CautionsThis section lists general cautions. Those pertaining to specific functions and procedures are included in the text where appropriate.

CAUTION: Contains ECG Electrodes which can damage skin if used improperly.This device should be used by or in consultation with a health care provider familiar with their proper placement and use. Apply electrodes to intact, clean skin only. Do not apply over an open wound, lesion, infected or inflamed skin.

CAUTION: Skin irritationPatients with sensitive skin or with known skin conditions should use this device with caution. If irritation such as redness, severe itching or allergic symptoms (i.e. hives) develop, instruct patients to remove the device immediately and have them contact their physician.

Safety Alert DescriptionsThe symbol shown below identifies a potential hazard category.

Cautions

CAUTION: Allergic skin reactionDo not use this device on patients with known skin allergies or family history of skin allergies.

CAUTION: Damage to skinInstruct patients to remove electrodes carefully to avoid damaging their skin.

CAUTIONThis alert identifies hazards that may cause minor personal injury, product damage, or property damage.

CAUTION: Equipment damageUse only as directed. Tampering with the Carnation Ambulatory Monitor may render it unusable. There are no user serviceable parts.

CAUTION: Equipment damageUse only with approved Bardy Diagnostics down-load station.

CAUTION: May interfere with defibrillation therapyTo avoid ineffective defibrillation therapy, remove the Carnation Ambulatory Monitor before applying any external cardiac defibrillators.

CAUTION: May interfere with MRI scanningThis device is not intended for use with MRI equip-ment and could reduce effectiveness of diagnostic scanning. Remove before any MRI scan.

CAUTION: Ineffective electronic imagingThis device is not intended for use with imaging equipment and could reduce effectiveness of diagnostic imaging. Remove before using electronic imaging systems.

CAUTION: Choking hazardThis is a prescribed medical device. Keep device and packaging away from young children.

8 9

snoituaC snoituaC

Processing CautionsWhen the ECG device is returned to the processing center, these additional cautions apply.

CAUTION: Contaminated surfacesWhen this device is returned to the reading center it will have been in contact with human skin. Follow the facility’s procedures for appropriate handling.

CAUTION: Battery may present environmental hazardThis device contains a battery. Properly dispose of batteries in accordance with local regulations.

CAUTION: Electronic waste may present environmental hazardThis device is considered medical electrical equipment. Properly dispose of electronic waste in accordance with local regulations.

CAUTION: Single use onlyThe Bardy Diagnostics Carnation Ambulatory Monitoris not intended for reuse, as the monitor becomesnon-functional after the first use.

CAUTION: Electrical shockThis Carnation Ambulatory Monitor contains electrodes for monitoring ECG. Do not allow conductive parts of the electrodes on this device to come in contact with other conductive parts (including earth).

CAUTION: Enclosure damageDo not use any parts that appear damaged. Check the Carnation Ambulatory Monitor for damage before using and reject any parts that have been damaged in shipping.

10

EMC Guidelines

EMC GuidelinesThe Carnation Ambulatory Monitor needs special cautions regarding EMC, and needs to be put into service according to provided information.

CAUTION: Electromagnetic interference (EMI) Portable and mobile wireless communications equip-ment such as wireless home network devices, mobile phones, cordless telephones and their base stations, and walkie-talkies can affect the Carnation Ambulatory Monitor and should be kept at least a distance of 3.3 meters away from the equipment. Position the Carna-tion Ambulatory Monitor away from other electrical or electronic equipment, if possible. The presence of strong EMI fields, or electronic, surgical, or diathermy instruments in close proximity to the ECG device may cause trace noise or input overload conditions.

The ECG device is compliant with IEC 60601-1-2 EMC immunity requirements.

Symbols Bardy Diagnostics products display one or more of these symbols and warning labels.

SYMBOL DESCRIPTION

Attention: Consult accompanying documents

Warning and caution symbol

TYPE BF APPLIED PART: F-TYPE APPLIED PART complying with the specified requirements of 60601-1 to provide a higher degree of protec-tion against electric shock than that provided by TYPE B APPLIED PARTS.

Note: TYPE BF APPLIED PARTS are not suitable for DIRECT CARDIAC APPLICATION.

11

Symbols

Single use only

IP23Protected against intrusion from fingers and small objects, and from water falling as a spray at an angle of up to 60 from vertical. Do not immerse in bathtub or swimming.

Sold by prescription onlyCAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare practitioner.

Manufacturer

Use-by date.

Do not expose to temperatures outside of these limits. For more information on environmental parameters refer to the Technical Specifications section.

131°F(55°C)

32°F(0°C)

SYMBOL DESCRIPTION

SN

Humidity must be within these limits. For more information on environmental parameters refer to the Technical Specifications section.

Date of manufacture

Contains electronic equipment. Dispose of properly in accordance with local regulations.

Serial number

Orderable part number

Batch code

10%

95%

12

Symbols

Atmospheric pressure must be within these limits. For more information on environmental parameters refer to the Technical Specifications section.

500hPa

1060hPa

CE Mark signifying conformity to EU regulations

EU authorized representative

MR Unsafe

Page 6: CAM AME: tion AME: Bardy Diagnostics IED (Required):Days ... · CAM back in place if it becomes loose or detached. CAUTION: Skin irritation Patients with sensitive skin or with known

PHYSICAL CHARACTERISTICS

178mm x 38mm x 14mm

<25gMedical grade thermoplastic polymer

CLASSIFICATION

Type of protection Internally powered

Degree of protection Type BF applied part

Protection against IP23 (Protected against intrusion

objects and water from fingers and small objects,

ingress and from water falling as a spray

at an angle of up to 60° from

vertical)

13

Technical Specifications

Technical Specifications

ITEM SPECIFICATION

PERFORMANCE CHARACTERISTICS

1 channel48 hoursContinuous48 hours

ECG channels Recording capacity Recording format Service life Shelf life months

ELECTRICAL CHARACTERISTICS

Frequency response Differential range A/D sampling rate

0.67 Hz to 25 Hz4 mV171 Hz

POWER REQUIREMENTS

Battery type Lithium primary (coin cell)

ITEM SPECIFICATION ENVIRONMENTAL SPECIFICATIONS (ECG DEVICE)

Operating temperature 50 °F to 113°F

Operating pressure

Operating humidity

Transport temperature

Storage temperature

Transport/Storage humidity

Transport/Storage pressure

Standards compliance

(10 °C to 45 °C)

700 to 1060 hPa

10% to 95%

(non-condensing)

14°F to 130°F

(-10 °C to 55 °C)

59°F to 77°F

(15 °C to 25 °C)

10% to 95%

(non-condensing)

500 to 1060 hPa

Applicable sections of IEC

60601-1, 60601-1-2,

60601-1-11,

60601-2-47

14

Technical Specifications Carnation Ambulatory Monitor

Approximate Dimensions

Weight

Enclosure material

Flammability rating UL- HB

Lithium content Lithium content < 1 g Heavy metal content Within weight limits of 2006/66/EC UN compliance Complies with UN 3090

ELECTRODE CHARACTERISTICS

Number of electrodes

Type

2

Electrode incorporating electrode

Supplied as

Lead wire length

Materials:

gel and internal lead wire

Disposable, non-sterile

11.6 cm (no patient contact)

Electrode gel: Medical grade

conductive synthetic

Adhesive: Medical grade skin

adhesive

15

Guidance and manufacturer’s declaration – electromagnetic emissions

The CAM is intended for use in the electromagnetic environment specified below. The customer or the user of the CAM should assure that itis used in such an environment.

Emissions test

RF emissionsCISPR 11

RF emissionsCISPR 11

Harmonic emissionsIEC 61000-3-2

Voltage fluctuations/flicker emissionsIEC 61000-3-3

Guidance and manufacturer’s declaration – electromagnetic immunity

The CAM is intended for use in the electromagnetic environment specified below. The customer or the user of the CAM should assure that itis used in such an environment.

IMMUNITY test

Electrostaticdischarge (ESD)IEC 61000-4-2

Power frequency(50/60 Hz) magnetic fieldIEC 61000-4-8

RadiatedRF IEC 61000-4-3

Radiated RFProximity FieldsIEC 61000-4-3(per IEC 60601-1-2 Ed. 4)

Compliance

Group 1

Group B

Not Applicable

Not Applicable

Electromagnetic environment - guidanceThe CAM uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The CAM has No AC Mains, and is suitable for use in all establishments, including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

IEC 60601 test level

± 8 kV contact± 15 kV air

30 A/m

10 V/m80 MHz to 2,7 GHz

9 V/m - 28 V/m perIEC 60601-1-2 Ed. 4,Table 9

Compliance level

± 8 kV contact± 15 kV air

30 A/m

10 V/m80 MHz to 2,7 GHz

9 V/m - 28 V/m perIEC 60601-1-2 Ed. 4,Table 9

Electromagnetic environment - guidance

Floors should be wood, concrete or ceramic tile. If floorsare covered with synthetic material, the relative humidityshould be at least 30%.

Power frequency magnetic fields should be that of atypical commercial or hospital environment.

Portable and mobile RF communications equipment shouldbe used no closer than 30 cm to any part of the CAM.

Interference may occur in the vicinity ofequipment marked with the following symbol: