EXP NO: DATE: CALIBRATION OF THE, RTD, THERMOCOUPLE, TEMPERATURE INDICATOR AIM To calibrate (i ) Temperature indicator, (ii) RTD, (iii) Thermocouple, EQUIPMENTS REQUIRED a)RTD-PT100 b) Resistance box c)Oil bath d) Temperature indicator e) Thermocouple(J or K type) SPECIFICATIONS Sensor-PT100 Temperature indicator range= - 50℃ to 200℃ PROCEDURE CALIBRATION OF RTD: a) RTD -PT100 is placed inside the oil bath. b) Oil bath is first set to a particular temperature using set temperature control. c) Resistance of RTD at that temperature is noted from meter.

Calibration of the Temperature Indicator

Embed Size (px)

DESCRIPTION

LAB MANUAL

Citation preview

Page 1: Calibration of the Temperature Indicator

EXP NO:DATE:

CALIBRATION OF THE, RTD, THERMOCOUPLE, TEMPERATURE INDICATOR

AIM

To calibrate (i ) Temperature indicator, (ii) RTD, (iii) Thermocouple, EQUIPMENTS REQUIRED

a) RTD-PT100b) Resistance boxc) Oil bathd) Temperature indicatore) Thermocouple(J or K type)

SPECIFICATIONS

Sensor-PT100

Temperature indicator range= - 50℃ to 200℃PROCEDURE

CALIBRATION OF RTD:

a) RTD -PT100 is placed inside the oil bath.b) Oil bath is first set to a particular temperature using set temperature

control.c) Resistance of RTD at that temperature is noted from meter.d) Temperature is varied in steps and their corresponding resistance is

noted.e) Measured resistance is compared with standard resistance at particular

temperature and their errors are calculated.

Page 2: Calibration of the Temperature Indicator

WIRING DIAGRAM FOR RTD AND THERMOCOUPLE CALIBRATION

230v

rtd

t/c

CALIBRATION OF RTD

TEMPERATURE STD RESISTANCE (ohm)

MEASURED RESISTANCE(ohm)

%ERROR

% ACCURACY

CALIBRATION OF THERMOCOUPLE :

Power supply

Oil bath multimeter

Page 3: Calibration of the Temperature Indicator

a) Thermocouple is placed inside the oil bath.b) Oil bath is first set to a particular temperature using set temperature

control.c) Voltage of the thermocouple at that temperature is noted from meter.d) Temperature is varied in steps and their corresponding voltages are

noted.e) Measured voltages are compared with standard voltages at particular

temperature and their errors are calculated.

CALIBRATION OF TEMPERATURE INDICATOR:

a) A resistance box is simulated as a RTD and it is connected to temperature indicator.

b) From the table, the standard resistance for each temperature within the range.

c) The particular standard resistance is set in resistance box and corresponding temperature indicator is noted.

d) This procedure is repeated for various temperatures and indicator value is noted.

e) Thus the measured value is compared with the standard value and error is noted.

Page 4: Calibration of the Temperature Indicator

CALIBRATION OF THERMOCOUPLE

TEMPERATUREIN ℃ THERMOCOUPLE

OUTPUT VOLTAGE(mv)%ERROR % ACCURACY

STANDARD MEASURED

Model Calculation for

Page 5: Calibration of the Temperature Indicator

% Error = Standard value−Measured value

Standard value

% Accuracy = 100 -%ErrorFull scale reading

Before calibration

Reading No:

After calibration:

Reading No:

WIRING DIAGRAM FOR TEMPERATURE INDICATOR:

Page 6: Calibration of the Temperature Indicator

(230v)

p n 1

n l 2

CALIBRATION OF TEMPERATURE INDICATOR:

TEMPERATURE (0C)

RESISTANCE (ohm)

MEASURED TEMPERATURE(0C)

ERROR%

ACCURACY %

Power supply

Temperature indicator

Page 7: Calibration of the Temperature Indicator

RESULT