EXP NO:DATE:
CALIBRATION OF THE, RTD, THERMOCOUPLE, TEMPERATURE INDICATOR
AIM
To calibrate (i ) Temperature indicator, (ii) RTD, (iii) Thermocouple, EQUIPMENTS REQUIRED
a) RTD-PT100b) Resistance boxc) Oil bathd) Temperature indicatore) Thermocouple(J or K type)
SPECIFICATIONS
Sensor-PT100
Temperature indicator range= - 50℃ to 200℃PROCEDURE
CALIBRATION OF RTD:
a) RTD -PT100 is placed inside the oil bath.b) Oil bath is first set to a particular temperature using set temperature
control.c) Resistance of RTD at that temperature is noted from meter.d) Temperature is varied in steps and their corresponding resistance is
noted.e) Measured resistance is compared with standard resistance at particular
temperature and their errors are calculated.
WIRING DIAGRAM FOR RTD AND THERMOCOUPLE CALIBRATION
230v
rtd
t/c
CALIBRATION OF RTD
TEMPERATURE STD RESISTANCE (ohm)
MEASURED RESISTANCE(ohm)
%ERROR
% ACCURACY
CALIBRATION OF THERMOCOUPLE :
Power supply
Oil bath multimeter
a) Thermocouple is placed inside the oil bath.b) Oil bath is first set to a particular temperature using set temperature
control.c) Voltage of the thermocouple at that temperature is noted from meter.d) Temperature is varied in steps and their corresponding voltages are
noted.e) Measured voltages are compared with standard voltages at particular
temperature and their errors are calculated.
CALIBRATION OF TEMPERATURE INDICATOR:
a) A resistance box is simulated as a RTD and it is connected to temperature indicator.
b) From the table, the standard resistance for each temperature within the range.
c) The particular standard resistance is set in resistance box and corresponding temperature indicator is noted.
d) This procedure is repeated for various temperatures and indicator value is noted.
e) Thus the measured value is compared with the standard value and error is noted.
CALIBRATION OF THERMOCOUPLE
TEMPERATUREIN ℃ THERMOCOUPLE
OUTPUT VOLTAGE(mv)%ERROR % ACCURACY
STANDARD MEASURED
Model Calculation for
% Error = Standard value−Measured value
Standard value
% Accuracy = 100 -%ErrorFull scale reading
Before calibration
Reading No:
After calibration:
Reading No:
WIRING DIAGRAM FOR TEMPERATURE INDICATOR:
(230v)
p n 1
n l 2
CALIBRATION OF TEMPERATURE INDICATOR:
TEMPERATURE (0C)
RESISTANCE (ohm)
MEASURED TEMPERATURE(0C)
ERROR%
ACCURACY %
Power supply
Temperature indicator
RESULT