38
NCAR/TN-340+STR NCAR TECHNICAL NOTE October 1989 Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ALTITUDE OBSERVATORY NATIONAL CENTER FOR ATMOSPHERIC RESEARCH BOULDER, COLORADO I - .- - ;

Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

NCAR/TN-340+STRNCAR TECHNICAL NOTE

October 1989

Calibration of the HAO Standard Opal Filter Set

J.L. Streete

HIGH ALTITUDE OBSERVATORY

NATIONAL CENTER FOR ATMOSPHERIC RESEARCHBOULDER, COLORADO

I -

. - - ;

Page 2: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the
Page 3: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

iii

TABLE OF CONTENTS

Page

LIST OF FIGURES v

PREFACE vii

I. INTRODUCTION1

II. INSTRUMENTATION 3

Sources 3

Receiver 4

Data Collection 6

III. MEASUREMENT PROCEDURE 6

IV. RESULTS 7

V. SUMMARY AND SUGGESTED IMPROVEMENTS 10

Suggestions 11

VI. ACKNOWLEDGMENTS 12

VII. REFERENCES 14

Page 4: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the
Page 5: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

v

LIST OF FIGURES

Figure 1. Tranmsittance Curve-450 nm filter

Figure 2. Transmittance Curve-550 nm filter

Figure 3. Transmittance Curve-650 nm filter

Figure 4. Transmittance Curve-800 nm filter

Figure 5. Transmittance Curve-All Filters

Figure 6. Schematic Diagram of Opal Calibration System

Figure 7. Opal Transmittance at 550 nm

Figure 8. Opal Transmittance at 650 nm

Figure 9. Opal Transmittance at 450 nm

Figure 10. Opal Transmittance at 800 nm

Figure 11. Calculated Optical Densities of the Neutral Filters

Figure 12. Transmittance of 2B, 2D, 2E, 2F Opals at 450, 550, 650and 800 nm

Figure 13. Transmittance of 2G, 2H, 21, 2K Opals at 450, 550, 650and 800 nm

Figure 14. Transmittance of 2L, 20, 2P, 2Q Opals at 450, 550, 650and 800 nm

Page 6: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the
Page 7: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

vi i

PREFACE

New measurements of the transmittance of the HAO "fundamental

standard set" of opal filters were begun in September, 1988. The

instrumentation used in the measurements was provided by Rhodes

College, Memphis, Tennessee and the High Altitude Observatory.

Since nearly two decades had passed from the time of the last

calibration measurements, there was concern that the transmittance of

the filters might have changed. Another motivation for recalibrating the

opal filters was the desire to know the transmittance values in spectral

regions above and below those used in 1970. The spectral regions chosen

were 450 nm, 550 nm, 650 nm and 800 nm.

Comparisons of transmittance are made with the previous

values in Table 1, and new transmittance values are given Table 2. The

new values agree with the 1970 measurements to within about 10 per

cent (usually considerably better than this) except for a few filters. The

transmittance of filter 2F at 650 nm shows the greatest discrepancy with

the old results, with the new value being a factor of 1.6 higher.

It is felt that the use of the direct radiation from the sun,

thereby insuring the proper geometry for the measurements, and

Page 8: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

viii

employing a silicon photodiode detector,whose response is extremely

linear over a wide dynamic range, have significantly improved the

accuracy of the transmittance measurements.

Jack Streete*Scientific VisitorHigh Altitude Observatory

*On leave from Rhodes College, Memphis Tennessee.

Page 9: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

1

I. INTRODUCTION

Solar radiation attenuators called "opal filters" have been used for

several decades to calibrate photographic film in instruments used to

measure the brightness of the solar corona (see for example 1,2). The

primary component used in fabricating these filters is opal glass. This

glass contains a suspension of minute colloidal particles which produce

multiple scattering of the incident radiation. The diffusion thus produced

is very close to Lambertian, i.e.,the scattered intensity varies as the

cosine of the angle between the normal to the scattering surface and the

direction of measurement. As an example, opal glass illuminated with a

collimated light beam normal to its surface is seen to have, at 45 degrees

from normal, about 90 per cent the brightness predicted for a perfect

diffuser.3 The size of the suspended particles is such as to produce a

slight enhancement in the scattering of shorter visible wavelengths.

Transmittance measurements made with collimated laser beams,

however, show a variation of only about a factor of 1.02 in transmittance

from 400 nm to 800 nm.4

The "fundamental standard set" of opals produced and used at the

High Altitude Observatory contains twelve filters, beginning with a bare

Page 10: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

2

opal whose attenuation factor is about 105. The rest of the filters of the

set were constructed by joining an opal glass filter with a neutral density

filter of a particular density. The last filter of the set has an attenuation

factor of about 109. In the field the filters are used by placing them in

front of the objective lens of the telescope which is pointed at the sun so

as to introduce a known attenuation to the photospheric radiation striking

the film.

The "standard set" was last calibrated in 19705 ,and there is concern

that the attenuation might well have changed with time due to changes in

the neutral density filters or the cement between them and the opals.The

purpose of the present measurements was to recalibrate the set, using the

sun as the radiation source, if possible.

For a perfect diffuser, a Lambertian surface, the attenuation

produced, except for the reflectance, is approximately the solid angle

subtended by the source at the opal divided by 2x(pi). In other words, the

attenuation is directly proportional to the angular size of the source.

Therefore, since the opal glass provides such a Lambertian surface, it is

important, when calibrating the opal filters to use a source which

subtends the same solid angle at the opal as will the source used in the

field. For this reason, the sun (or a source which geometrically simulates

Page 11: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

3

the sun). must be used in measurements of transmittance.

Although the attenuation of the filters (opal glass and neutral

density filter) is not strongly dependent on wavelength, there is a slight

wavelength dependence, mainly due to the neutral density filters. For this

reason, the transmittance of the filters was measured in four spectral

regions centered at 450 nm, 550 nm, 650 nm and 800 nm. These spectral

regions were isolated with broadband interference filters each having a

bandpass of approximately 75 nm. Figure 1 through Figure 4 are

transmittance curves for these four filters, and Figure 5 shows the

combined transmittances for the filters.

Even though the calibration of the opal filters is relatively

straightforward, a great deal care must be taken to eliminate scattered

light when the opal is in the light beam. This is especially true when

measuring the most dense filters with attenuations of about 109.

II. INSTRUMENTATION

Sources.

Initially, two radiation sources were used in the calibration

measurements, a quartz halogen lamp with optics to simulate the sun

geometrically, and the sun itself. The final measurements were made

using only the sun as the source of radiation. An important advantage of

Page 12: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

4

using the sun is that the irradiance produced in the plane of the detector

is great enough to measure directly even the densest filters. This was not

the case in 1969 when it was necessary to move the tungsten source

closer to the detector in order to have sufficient signal for filters of

attenuation greater than 108. The High Altitude Observatory's heliostat

was used to direct a solar beam into the laboratory. The tracking accuracy

was such that only one or two times during a measurement session of

about five hours was it necessary to realign the heliostat.

Receiver.

As shown in Figure 6, the receiver end of the system was comprised

of the interference filter, F1, whose holder was located in front of a

Geneva motion driven, four position aperture wheel which held the opal

filter, F2. Also the aperture wheel held, in the other three positions, a

calibrated neutral density filter required in the open position when using

the sun, and two opaque metal inserts used to block the radiation and to

provide an optical zero. The neutral density filter was measured to have

transmittances of 0.111, 0.105, 0.104, 0.103 respectively at the spectral

positions 450 nm, 550 nm, 650 nm, and 800 nm.

A 676 mm focal length off-axis parabolic mirror, M4, imaged the

solar disk onto the detector. An aperture over the mirror was slightly

Page 13: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

5

smaller than the solar beam passed by the filter wheel aperture and was

used to insure that radiation was collected over the same solid angle for

the opal-in and opal-out positions.

The detector was a silicon photodiode, model UDT-111A made by

United Detector Technology, Inc. An important feature of this detector is

that it has a 1 percent linear response over the range of its sensitivity

which is from 10-2 to 10-11 watt. There was provision for changing the

horizontal position of the detector slightly by means of a micrometer

drive. An adjustable aperture was installed in front of the detector to

provide a field stop slightly smaller than the solar image. Inserted

between the filter aperture and mirror M4 and between mirror M4 and the

detector was a set of baffles, B1 and B2, used to eliminate scattered

light.

A light tight box was installed over the instrument, so that the

room would not need to be completely dark when carrying out the

measurements. This was necessary since other projects were sometimes

going on in the same room. However, these projects did not require that

the overhead lights be on. The top of the box was hinged so that the optical

alignment could be checked regularly.

Page 14: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

6

Data Collection.

The detector was interfaced with an Integrated Solutions computer

and software was written to provide output voltages, standard deviations

of the voltages and calculations of transmittances for each of the twelve

opal filters in the four spectral regions provided by the interference

filters. Options were available to designate the aperture port, gain of the

detector, and number of measurements to be made at each position. The

data point sampling rate was one per second. Copies of the data were

available from a laser printer.

III. MEASUREMENT PROCEDURE

The measurements were carried out only when the sky was clear

and the winds relatively calm. After setting up the heliostat and

directing the radiation into the laboratory, a beam-directing mirror, M3

in Figure 6, was adjusted to center the solar radiation on the aperture

port of the filter wheel and off-axis parabolic mirror. Adjustment of the

micrometer drive was then made to center the solar image on the detector

aperture.

The opal filter being measured was clamped in the filter wheel in

one of the four available positions and behind one of the interference

Page 15: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

7

filters. A particular opal was measured with each of the four interference

filters before beginning measurements on the next opal filter. The

measurements were done at least twice with each opal/interference

filter combination and the results checked before inserting the next opal.

Only when there was good agreement between two transmittance values

was the next spectral filter inserted. Before beginning measurements

with a new interference filter the optical alignment was checked and

adjusted if necessary.

IV. RESULTS

Table 1 shows the results of the new transmittance measurements

for each of the twelve opal filters in each of the four spectral regions.

Also listed are the 1970 values for the two spectral regions measured at

that time. Statistical analysis of the results shows that the random error

for each of the separate measurements produced a standard deviation of

0.01 to 0.02 for filters 2B through 2L and about 0.08 for filters 20, 2P and

2Q. The systematic error, an measure of which can be gotten from the

third column varies from a fraction on a per cent to about six per cent for

the first nine filters to as high as 12 per cent for the 20 filter at 800 nm.

This systematic error results primarily from from image motion and

probably from stray light in the instrument, which is more of a problem at

Page 16: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

8

extremely low transmittances.

For the two wavelength regions where comparison can be made

between the 1970 and 1989 results, namely 550 nm and 650 nm, the

agreement is seen to be mainly better than 10 per cent for most cases.

These results are given in the fifth column for the 550 nm and 650 nm

data. Table 2 is included to give at a glance the most recently measured

transmittances of the opal filters.

Figures 7 and 8 are the transmittance curves for the twelve filters

for 550 and 650 nm, respectively. Graphs of the present measurements

(NEW VALUES) and 1970 (OLD VALUES) are shown. The main significant

difference is seen to be the 2F filter in the 650 nm region (Figure 8),

where, as is seen in Table 1, there is a difference of about 38 per cent. In

reference 5 this value is in parenthesis, possibly indicating (although not

specifically stated) that the value is questionable. Also, Figure 8 shows a

discrepancy in the transmittance of filter 2Q. Again Table 1 indicated a 27

per cent difference in the two values. Figures 9 and 10 show the

transmittance values at 450 and 800 nm respectively.

Calculations were made to determine the densities of the

neutral density filter part of the opal filter that would be required to

produce the measured transmittances. Assuming that the overall

Page 17: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

9

transmittance, T, is given by

T - K x (Topal x TND), where

K is a constant taken for this purpose as 1 since for this

calculation we are interested only in relative values,

Topal is the measured transmitance of the bare opal, filter 2B,

and,

TND is the transmittance of the neutral filter part of

filters 2D through 2Q and is given by,

TND = T/Topal 10-D, or

D log(Topai/T), and

D is the optical density of the neutral filter.

Figure 11 shows the resulting calculated neutral filter

densities for filters 2D through 2Q. It can be seen that the density at 800

nm is always higher than for the shorter wavelength regions. Although the

original density curves for the neutral filters can not be located, with

typical neutral density filters the density does indeed increase with

wavelength. Also, for typical neutral density filters, the rate of change of

slope of the density curves usually increases with the density of the

filter. This effect is also seen in Figure 11.

Figures 12 through 14 are plots of transmittances of the

Page 18: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

10

separate opal filters at the four wavelengths used. It can be seen that,

typically, the transmittance is slightly higher at 550 and 650 nm and

lower at 450 and 800 nm. However, in Figure 14, the transmittance of

filters 2P and 2Q is seen to be highest at in the 450 nm region. Since the

signal-to-noise was least in the 450 nm region and for filters 2P and 2Q

it is likely that these two values are too high by possibly as much as 50

per cent.

V. SUMMARY AND SUGGESTED IMPROVEMENTS

Measurements of the transmittance of the HAO "fundamental

standard set" of opal filters were begun in the fall of 1988. The

motivation behind this effort was concern that in the almost twenty years

since the last calibration of these filters, their transmittance might have

changed. Also there was interest in knowing the transmittances of the

opal filters in spectral regions outside those used in the 1970

measurements. Thus measurements were extended to include the 450 nm

and 800 nm spectral regions.

Since the transmittance of these filters is directly

proportional to the angular size of the source being observed, and since

the sun is the source used in the field, the decision was made to try to use

the sun as the source of radiation for the measurements. Previous

Page 19: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

11

measurements used incandescent sources which geometrically simulated

the sun. The use of the sun also permitted the direct measurement of

transmittance to levels of about 10-9.

The dynamic range over which the measurements must be

made is approximately 109. Therefore a silicon photodiode detector was

chosen as the detector since its response is linear to 1 per cent over a

range of 109.

The new transmittance values agree with the 1970

measurements to within approximately 10 per cent except for the 2F and

2Q filters at 650 nm and the 2P filter at 550 nm. The discrepancy is

greatest for the 2F filter at 650 nm where the new value is 38 per cent

higher (a factor of 1.6 higher) than the 1970 value. It is recommended that

researchers employing this filter in the future use the new value for

transmittance. It might also be important to check the literature to

determine whether or not this filter has been used in the past in

calibrating systems used in measuring coronal brightness.

Suagestions

Probably the least satisfactory part of the experimental setup

was the lack of tracking accuracy of the heliostat. Although the heliostat

itself had to be realigned only a few times during several hours of

Page 20: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

12

measurement, the beam directing mirror had to be readjusted quite often

to keep the solar image centered on the detector aperture. Lack of

centering could severely affect the value of transmittance obtained.

Correction of this problem would greatly improve the ease of making the

measurements as well as the accuracy of the transmittance values.

VI. ACKNOWLEDGMENTS

The author gratefully acknowledges the considerable assistance of

several people at HAO who helped with various phases of this project.

Among other things, David Elmore had the detector interfaced with his

computer and wrote the software necessary to provide data collection and

analysis. Lee Lacey helped with suggestions and well as design and

construction of components needed to perform the measurements. Dick

Fisher's help, encouragement and suggestions as well as manual

assistance in setting up the heliostat made my work much easier than it

otherwise would have been. Paula Rubins and Rick Sheffer actually did the

wiring and electronic fabrication necessary for the interfacing.

Others who helped set up the heliostat were Terry Leach, Charlie

Miller and Steve Tomsczyk. Several others including Chris Abato and

Charlie Miller assisted with the alignment of the optical system. The

author is indeed grateful to Art Hundhausen for providing the funds

Page 21: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

13

necessary to carry out the measurements.

Page 22: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

14

VI. REFERENCES

1. Poland, A.; MacQueen, R.; Munro, R.; Gosling, J.; "Radiance Calibration of

the High Altitude Observatory White-Light Coronagraph on Skylab", Applied

Optics, 16, (1977).

2. Newkirk, G.,Jr.; Dupree,R.; Schmahl,E.;"Magnetic Fields and the Structure

of the Solar Corona, II." Solar Physics, 15, (1970).

3. Smith, W., "Modern Optical Engineering", McGraw-Hill, 1966.

4. Boivin, L.,"Diffusers in Silicon-Photodiode Radiometers", Applied

Optics, 21, No. 5, (1982).

5. Elmore, D. F.; Streete, J. L.; Eddy, J. A.; "Calibration of Opal Glass

Attenuators", Astro-Geophysical Memorandum No. 178, (Feb. 27, 1970).

Page 23: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

/U

Tr 60a

P n 50e sr m 40c ie t 30n tt a 2 020

nc

10e

0390 410 430 450 460 480 500 520 540 560 580 600

Wavelength (nm)

Figure 1. Transmittance Curve-450 nm filterci1

=I A

Page 24: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

mU

Tr 70

p a 60n

e s 50rm

c i 40e tn t 30ta

20nc 10e

0460 480 500 520 540 560 580 600 620 640 660 680 700

Wavelength (nm)

Figure 2. Transmittance Curve-550 nm filter0)

ft0

Page 25: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

80T

r70

ap a 60n

e s 50rmc i 40etn t 30t a

nc 10e

0580 600 620 640 660 680 700 720 740 760 780 800

Wavelength (nm)

Figure 3. Transmittance Curve-650 nm filter __z-A,,

Page 26: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

70Tr 60an 50

e sr m 40c i

e t 30n t

a 20nc 10e

0700 720 740 760 780 800 820 840 860 880 900

Wavelength (nm)

Figure 4. Transmittance Curve-800 nm filter

Page 27: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

iTiI 1111111 111 111111111111111111111111111111111111*- : :ua~u~uuuau~uhu *uuu0 u 0 u1*uu*a a a I A mmmla I I I i u

.a *IN * * m I m m -m 1 .ml !u * M u.n. u u * mm Ma I U I · I I Ia

+- I 0 m0 mm 0m I I U * i * i * * i u u u u u a 1 i ua i u I i mm 1 1i 1 i 1 1 u ! 1u

.roi · · · ..1 1! · · · i · m u f m h · m m m

400 _ 50400 450

~lpwrwgm~as$*jTIdIr F Ii i ii ULr 1 F'

500 550 600 650 700 750Wavelength (nm)

800 850 900

Figure 5. Transmittance Curves-All Filters--A

cD

80

percent

70

60

50

40

30

20

Transmittance

10

- 450 nm filter

- 550 nm filter

- 650 nm filter

-- 800 nm filter

a& m m Emmi a a r ·1 1 oil .I 0 am ·oil · a I A I I I11I IIls __I ME I IIII IIII I .

-4 I I 1 r 1 I I r r Il I I I I ml I I IF I I 0Ia I I I m 1 v w I I · · I AL I I I · a I I 4

Page 28: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

M1, I

M3: B

M4: C

F1- if

1

FECOERF2: 4-Posltlon Filer Wheel (Opal Placed at This Position)

B1, B2: Baffes

Ught Path

: Electrical Connections TERMINAL COMPUTER

o

Figure 6. Schematic Diagram of Opal Calibration System

Page 29: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

TABLE 1. TRANSMITTANCE OF OPAL FILTERS AND COMPARISON WITH PREVIOUS RESULTS

* Per cent difernces between successve measurements

550 nm 450 nm

Opel 1 99 % DIFFERENCE 1969 % DIFFERENCEFilter VALUES 1989 VALUES' VALUES NEW/OLD VALUES

29 9.42E-06 2.5% 1.03E-05 -9.3%2D 1.05E-06 1.1% 1.1OE-06 -4.3%2E 6.87E-07 1.2% 5.74E-07 2.2%2F 3.10E-07 0.1% 3.00E-07 3.2%20 1.81E-07 1.4% 1.72E-07 5.2%2H 1.25E-07 2.0% 1.17E-07 6.2%21 7.96E-08 0.1% 8.83E-08 -10.9%2K 3.63E-08 1.3% 3.92E-08 -8.0%2L 1.44E-08 0.1% 1.61E-08 -12.1%20 8.95E-09 3.1% 1.00E-08 -11.7%2P 2.66E-09 10.3% 3.19E-09 -20.1%20 1.81E-09 9.0% 1.71E-09 5.5%

Opal 19 89 % DIFFERENCEFilter VALUES 1989 VALUES*

2B 8.37E-06 0.8%20 9.48E-07 0.1%2E 6.17E-07 3.3%2F 2.92E-07 2.3%2G 1.65E-07 0.3%2H 1.18E-07 1.3%21 6.88E-08 3.7%2K 3.22E-08 4.1%2L 1.15E-08 0.1%20 8.48E-09 5.3%2P 3.17E-09 3.6%2Q 2.71 E-09 2.0%

i50 nm

Opel 19 % DIFFERENCE 199 % DIFFERENCEFilter VALUES 19 VALUES' VALUES NEWIOLD VALUES

2B 1.00E-06 1.1% 1.01E-05 -0.7%2D 1.08E-06 0.5% 9.95E-07 8.1%2E 5.74E-07 0.5% 5.26E-07 8.4%2F 2.95E-07 2.0% 1.83E-07 38.0%20 1.87E-07 0.4% 1.47E-07 12.0%2H 1.14E-07 0.5% 1.19E-07 -4.7%21 7.68E-08 0.2% 7.99E-08 -4.0%2K 3.49E-08 2.3% 3.49E-08 -0.1%2L 1.34E-08 2.0% 1.39E-08 -3.8%20 8.45E-09 9.4% 7.92E-09 6.3%2P 2.66E-09 6.3% 2.49E-09 6.4%20 1.70E-09 3.8% 1.25E-09 26.5%

2 800 nm

Opll 1 9 8 % DIFFERENCEFilter VALUES 1989 VALUES*

20 1.02E-05 1.7%2D 1.01E-06 0.3%2E 4.87E-07 0.6%2F 2.41 E-07 1.4%2G 1.45E-07 6.4%2H 1.03E-07 0.4%21 5.66E-08 0.4%2K 2.78E-08 4.0%2 L 1.05E-08 3.0%20 6.69E-09 12.5%2 P 2.00E-09 5.8%20 1.34E-09 2.5%

-A

Page 30: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

TABLE 2. TRANSMITTANCE OF OPAL FILTERS

T-R AN S MIT T A N C E

- - ------ -WAVELENGTH REGION (NM) - - ----- -

450 550 650 800

8.37E-06 9.42E-06 1.00E-05 1.02E-059.48E-07 1.05E-06 1.08E-06 1.01E-066.17E-07 5.87E-07 5.74E-07 4.87E-072.92E-07 3.10E-07 2.95E-07 2.41E-071.65E-07 1.81E-07 1.67E-07 1.45E-071.18E-07 1.25E-07 1.14E-07 1.03E-076.88E-08 7.96E-08 7.68E-08 5.66E-083.22E-08 3.63E-08 3.49E-08 2.78E-081.15E-08 1.44E-08 1.34E-08 1.05E-088.48E09 8.95E-09 8.45E-09 6.69E-093.17E-09 2.66E-09 2.66E-09 2.00E-092.71 E-09 1.81E-09 1.70E-09 1.34E-09

OpalFilter

2B2D2E2F2G2H212K2L202P20

Page 31: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

1TRA 1

NSs 1

ITTANCE

12B 2D 2E 2F 2G 2H 21 2K 2L 20 2P 2Q

OPAL FILTER DESIGNATIONS

Figure 7. Opal Transmittance at 550 nm

- OLD VALUES

-- NEW VALUES II

Page 32: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

1.00E-04

1.00E-05

1.00E-06

1.00E-07

1.00E-08

1.00E-09

N<

2B 2D 2 2E F 2G 2H 21 2K 2L 20 2P 20OPAL FILTER DESIGNATION

Figure 8. Opal Transmittance at 650 nrr

TRANSMITTANCE

-*- OLD VALUES

-- NEW VALUES

\)-Ph

a a II

I I

I I

I I

I I

I I

I

F& I

I

I

I lk -I

I lk I

a

I

I

I

I

I

I i. I i

I

I I

I-

- -

I-

I.

0

I

I

I

I

II

II

IIIII

I

I

IIIII

a a II

I I

I - I

I I

I I

I

III

I

III

I

I

I

I

I

I

I

k

IIII 1

1 1

1 1

II

III

s

I

I

I

I

I

I

I

I

* --..

.4k-,,%-

I --= .. ,

i ^2*41r- 4i ii .I I . I.I II

a

i ii i

I IIII I

%I1 2sh 11 1IIIII

i

14,

I

I-I-

I

I-

I-

I-

I-

I-

I--

I_

a

-77__

I I

III I ,,- F

i i ii I i t ii i m i Zr, mI a

II I

I

IJ

I

=tm

40i I I I 1 I I r .- I I- - - -- 7 - -T -- -

I

4

4

4

N

N*.-

Page 33: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

1.00E-05

1.00E-06

1.00E-07

1.00E-08

I nnl:nQI W.vv -V

2B 2D 2F 2G 2H 21 2KOPAL FILTER DESIGNATIONS

2L 20 2P 2Q

Figure 9. Opal Transmittance at 450 nmen

TRANSMITTANCE

Page 34: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

1 It-4 -

1.00E-05

1.00E-06

1.00E-07

1.00E-08

1.00E-092B 2D 2E 2F 2G 2H 21 2K 2L 20 2P 2Q

OPAL FILTER DESIGNATIONS

Figure 10. Opal Transmittance at 800 nm h)

A2

TRANSMITTANCE

I- I

I

I

I

I'

I

I

I

I

I

I

I I

I - I

I I

I I

I

L- - E�a I m

II iI-

a A

I

m A

I

I

L.-

- -- ff

i-

F---- ----- -II I

I

II - II- II - I& II- IV I

I - x I

I N, I

II II II II II - I. I

I II III II II II I

I III II II I

I II I

I

I

I

IIIIIL - IIIIIIIIIII-IIIIII

IIIIIII

I

i

I

I

I 111~

I

I

I

I

I

a

I

mob.

4

II

I

I I

9II

I

I

I

II-III1.i

1,

IIII

I

I

II

I

IIII

I

I

I

I

I

I

I

I

I

I

IIIIIIIIII . .

1.

II

I

IIIIIII

I II II - Ii I

II I

II II II II

I I

II

II ---'1%, II - :!"b.1 %;�

i -

I-

I-

I-I

I-

I -~I-

I lokfbI W. I

-I Irad - ---

_m .A

1.00

Page 35: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

Densities of Neutral Filters Requiredto Produce Measured Transmlttances

* 800 nm

* 650 nm

* 550 nm

* 450 nm

2D 2E 2F 2G 2H 21 2K 2LOpal Filter Designations

Figure 11. Calculated Optical Densities ofthe Neutral Filters

20 2P 2Q

(I)

4.00

3.50

Density

3.00

2.50-

2.00.

1.50-

1.00

0.50

0.00

--~ ~ ~ ~ ~ ~ _

| ~ _ --

I

i

ii

Ni

Page 36: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

1.00E-04

1.00E-05

1.00E-06

1.00E-07450 550 650

BANDPASS CENTER OF INTERFERENCEFILTERS

Figure 12. Transmittance of 2B, 2D, 2E, 2FOpals at 450, 550, 650 and 800 nm

TRANSMITTANCE

--. 2B

-0- 2D

-O- 2E

-- 2F

800

I\)00

IIa

II II I

I--- iII

I I---- - i II-

i iII I i~~~~~~~~II Ii

I I

IIII

I I

WI - Ii i0 - -41I II I---

I I

I

III

I

I

I

1.IIIi -1IIIs I

I

t�----·----·--� I --

I ----� , -

�---I

I lI .--

I

I

Page 37: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

1.00E-06

1.00E-07

i nnFt-nA

~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~-J_ ----_- b

__ I

_~~~~ 1

-- 2G

-- 2H

-a- 21

-o- 2K

50 NM 550 NM 650 NMBANDPASS CENTER OF INTERFERENCE

FILTERSFigure 13. Transmittance of 2G, 2H, 21, 2K

Opals at 450, 550, 650, and 800 nm

TRANSMITTANCE

800 NM

CD

--

I 0%FVp" Ijq

4,1

Page 38: Calibration of the HAO standard opal filter set.109/datastream/… · Calibration of the HAO Standard Opal Filter Set J.L. Streete HIGH ... opal filters was the desire to know the

1 .OOE-07

I .OOE-08

1 .OOE-09

-- 2L

-- 2P

-- 2Q

_i i i i i lii.

.... i i i i i i i i i i i i i i i i i i J i i i-ii

i- i iiii

v~~~~~~~~LIl I

7~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~ i i,

450 NM

Figure14.Opals

550 NM 650 NMBANDPASS CENTER OF INTERFERENCE

FILTERSTransmittance of 2L,_ 20, 2P, 2Qat 450, 550, 650, and 800 nm

800 NM

A)CO

TRANSMITTANCE