788 Programmable Weight Indicating Instrument

Calibration Manual

MODEL 788SUPS-1 and 788SUPS-2

8545-M258-O1 Rev A PO BOX 151 � WEBB CITY, MO 64870 Printed in USA 10/05 PH (417) 673-4631 � FAX (417) 673-5001 www.cardet.com/ups

Page 2: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

2

Page 3: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

3

TABLE OF CONTENTS

INTRODUCTIONS Page 1 SPECIFICATIONS Page 2

Standard Features Page 2 INSTALLATION Page 3

Unpacking Page 3 Placement Page 3

SITE PREPARATION REQUIREMENTS Page 3 INTERCONNECTIONS Page 5

Load Cell Connection with Over 30 Feet of Cable Page 5 Load Cell Connection Page 5

SERIAL I/O CONNECTIONS Page 8 POWER LAYOUT Page 10 SETUP and CALIBRATION Page 11 SCALE TOTALIZER Page 16 CONTRAST CONTROL Page 19 RECEIVING AND SENDING PROGRAMS Page 20

Receive Page 20 Send Page 23

778C SERIAL CARD Page 24 778A ANALOG CARD Page 26 ERROR and STATUS MESSAGES Page 28 BEFORE YOU CALL SERVICE Page 29 PROBLEM RESOLUTION Page 30 PROCEDURE FOR RE-FORMATTING RAM DISK Page 35

LIST OF FIGURES

Figure No. 1 Post and Base Mounting Figure No. 2 Connector Wiring Diagram Figure No. 3 New vs. Present Layout Figure No. 4 Serial I/O Pin Location Figure No. 5 Serial I/O Connections Figure No. 6 Power Layout Diagram Figure No. 7 Calibration Screw Figure No. 8 Start Menu Figure No. 9 778C Serial Card Figure No. 10 778A Analog Card Figure No. 11 to 20 Ram Disk Clearing Procedure

Page 4: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

4

EUROPEAN DECLARATION OF CONFORMITY Manufacturer: Cardinal Scale Manufacturing Company PO Box 151 203 East Daugherty Street Webb City, Missouri 64870-0151 USA Telephone No. 417.673.4631 Fax No. 417.673.5001 Product: Non-automatic Weight Indicating Instrument Model Numbers 777, 778 and 788 Serial Number EXXXYY-ZZZZ Where XXX = day of year YY = last two digits of year ZZZZ = sequential number The undersigned hereby declares, on behalf of Cardinal Scale Manufacturing Company of Webb City, Missouri, that the above-referenced product, to which this declaration relates, is in conformity with the provisions of: European Standard EN45501 and the equivalent International Recommendation OIML R76, edition 1992 Report No. DANAK – 193750 as amended Council Directive 90/384/EEC Non-automatic Weighing Instruments Certificate of EU Type-Approval No. DK0199.22 and DK0199.22/1 Council Directive 89/336/EEC EMC Immunity Report No. DANAK – 193688 Council Directive 93/68/EEC (22 July, 1993) Low Voltage The Technical Construction File required by this Directive is maintained at the corporate headquarters of Cardinal Scale Manufacturing Company, 203 East Daugherty Street, Webb City, Missouri, USA. ______________________________ Link Yeager Director, Quality Assurance

Page 5: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

1

INTRODUCTION Thank you for purchasing the Cardinal 788 Programmable Weight Indicating Instrument. This instrument was built with quality and reliability at our factory in Webb City, Missouri. The purpose of this manual is to provide you with a guide through installation, calibration, and operation of your new weight indicating instrument. Please read it thoroughly before attempting to install or operate your 788 Indicator and keep it handy for future reference. The 788 incorporates the latest in digital technology and innovative features for the weighing industry. With programmable software, a user friendly program language and modular hardware design, the 788 has the versatility and high performance to be used in any type of standard or custom application. Configuration and upgrades to the application can easily be performed in the field, while still maintaining the rigid control the most demanding installations require. This flexibility insures the 788 will be able to meet your weight indicating needs for years to come.

FCC COMPLIANCE STATEMENT WARNING! This equipment generates uses and can radiate radio frequency and if not installed and used in accordance with the instruction manual, may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of FCC rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment. Operation of this equipment in a residential area may cause interference in which case the user will be responsible to take whatever measures necessary to correct the interference. You may find the booklet "How to Identify and Resolve Radio TV Interference Problems" prepared by the Federal Communications Commission helpful. It is available from the U.S. Government Printing Office, Washington, D.C. 20402, stock No. 001-000-00315-4.

All rights reserved. Reproduction or use, without expressed written permission, of editorial or pictorial content, in any manner, is prohibited. No patent liability is assumed with respect to the use of the information contained herein. While every precaution has been taken in the preparation of this manual, the Seller assumes no responsibility for errors or omissions. Neither is any liability assumed for damages resulting from use of the information contained herein. All instructions and diagrams have been checked for accuracy and ease of application; however, success and safety in working with tools depend to a great extent upon the individual accuracy, skill and caution. For this reason the Seller is not able to guarantee the result of any procedure contained herein. Nor can they assume responsibility for any damage to property or injury to persons occasioned from the procedures. Persons engaging the procedures do so entirely at their own risk.

PRECAUTIONS

Before using this instrument, read this manual and pay special attention to all "WARNING" symbols:

ELECTRICAL WARNING IMPORTANT

Page 6: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

2

UPS Models 788SUPS-1 SINGLE SCALE 788SUPS-2 DUAL SCALE SPECIFICATIONS

Displayed Resolution .….…… 10,000 divisions (commercial) 100,000 divisions (maximum) (non - commercial)

Internal Resolution …….……. 400,000 divisions

Division Value ………….……. 1 to 9 X 10, 1, 0.1, 0.01, 0.001 and 0.0001

Sample Rate ………….….….. selectable 1 to 60 samples per second

Auto Zero Range …….…..….. 0, 0.5 divisions or 1 through 9 divisions

Load-cells …………….…….... Up to twelve (12) 350 � load cells (with sense)

Excitation Voltage ….……...… 10.75 VDC

Sensitivity Range ……….…… 30 nV/V/ division maximum

Calibration …….………….….. Digital

Dimensions ………………..… 18” H x 19” W x 8” D

Weight …………………….….. 35 lb (wall mount version)

Construction ……………….… All stainless-steel housing, NEMA 4

Power Requirements ……….. 115 VAC @ 3A, optional 230 VAC @ 1.5A 50/60 Hz Optional 800 W Heater

Temperature Range ……..…. -10 to +40 °C / 14 to 104 °F

Standard Features

� 14 "Soft" Keys placed around the display that can be user defined for each application.

� Large, bright dot matrix LCD graphics display.

� Keyboard tare.

� Gross, tare, net conversion.

� One (1) Bi-directional RS232 port (scanner).

� One (1) Bi-directional combination (Ethernet device)

� One (1) 778C Card for Com3 and Com4 ports (Com3 for second scanner, Com4 for future use)

� Selectable and custom filtering.

� Integral diagnostics.

� One (1) scale input card for Model 788S-UPS1

� Second scale input card for Model 788S-UPS2

Page 7: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

3

INSTALLATION

Unpacking Carefully unpack and remove the indicator from the packing box. Inspect the unit for any signs of damage due to shipping, such as exterior dents and scratches. Retain the packing and shipping carton for return of instrument if necessary. It is the responsibility of the Purchaser to file all claims for any damages or loss in transit incurred, unless the responsibility has been accepted by the Seller in its proposal. Placement See the UPS Installation Manual for instructions. Refer to Figure No. 1 on the next page for dimensions and placement of system components.

SITE PREPARATION REQUIREMENTS The 788 Programmable Weight Indicator is a precision measuring instrument. As with any precision instrument, it requires an acceptable environment to operate at its peak performance and reliability. This section is provided to assist you in obtaining such an environment. Electrical Power The 788 is designed to operate from 90 to 260 VAC at 50/60 Hz. However, heaters and other parts are different in 115V or 230V AC Input models. Verify before installing that the unit is rated for available power source.

Caution!!! - To avoid electrical hazard and possible damage to the indicator, DO NOT, under any circumstance, cut, remove, alter, or in any way bypass the power cord grounding prong.

On models requiring 230 VAC power, it is the responsibility of the customer to have a qualified electrician install the proper power cord plug which conforms to national electrical codes and local codes and ordinances. The power outlet for the 788 should be on a separate circuit from the distribution panel. This circuit should be dedicated to the exclusive use of the indicator. The wiring should conform to national and local electrical codes and ordinances and should be approved by the local inspector to assure compliance. To prevent electrical noise interference, make certain all other wall outlets for use with air conditioning and heating equipment, lighting or other equipment with heavily inductive loads, such as welders, motors and solenoids are on circuits separate from the 788. Many of these disturbances originate within the building itself and can seriously affect the operation of the instrument. These sources of disturbances must be identified and steps must be taken to prevent possible adverse effects on the instrument. Examples of available alternatives include isolation transformers, power regulators, uninterruptible power supplies, or simple line filters.

Page 8: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

4

Figure No. 1 Post and Base Mounting

Page 9: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

5

INTERCONNECTIONS Load Cell Connection With Over 30 Feet Of Cable For installations with over 30 feet of cable between the indicator and the load cells, sense wires should be used. The sense wires must be connected between the +SENSE, -SENSE terminals on the indicator and the +EXCITATION, -EXCITATION wires of the load cells or the +SENSE, -SENSE terminals of the load cell trim board or the section seal trim board. For the indicator to use the sense wires, the SENSE jumpers J5 and J6 must be open on the Scale Input Board. Load Cell Connections 788SUPS-1 AND 788SUPS-2 These models are designed to be retrofitted to the existing platform scales. In doing so, it is the intention of UPS to keep the existing scale weight indicator as a back-up weight indicator. This can be done quickly using the weather-resistant military plugs, if properly installed. Since the 788SUPS-1 and 788SUPS-2 units have the military plug female version installed on the indicator bottom, the scale will require wiring the male mating connector into the scale load cell cable. Similarly, wiring a female connector into the cable coming out of the existing weight indicator will enable easy switchover of the scale from the 788SUPS-1 and 788SUPS-2 unit to the existing weight indicator. Both the existing indicator and the 788SUPS are calibrated to work with the scale base. The male connector from the scale, and the female connector from the existing weight indicator, should be installed using the pin wiring table shown in the Figure No. 2. Refer to Figure No. 3 to compare the present scale layout with the new layout. Note that with the new layout, the scale can be connected to the new unit or the existing indicator.

Page 10: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

6

Figure No. 2 Connector Wiring

Page 11: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

7

Figure No. 3 New vs. Present Layout

Page 12: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

8

SERIAL I/O CONNECTIONS The 788 comes standard with two bi-directional serial ports that are used as I/O. These ports are wired to a serial transition board. The external devices are then wired to the terminal strip either via a direct connection or through the use of a short pigtail cable with connector. The 788 units provided to UPS will have a model 778C serial card for COM3 (25pin) and COM4 (9pin). The COM4 will be used for the second bar code scanner on the two scale system. The COM3 is reserved for a possible future printer. The bar code scanner gun # 1 is connected to the short pigtail connector (9 pin) from the RS232 (COM1). The Lantronix Ethernet device is connected to the pigtail connector (9 pin) from the RS232 (COM2). The second scales bar code scanner gun # 2 is connected to the short pigtail connector (9 pin) from the RS232 (COM4). Figure No. 5 on the next page, illustrates the connections. Figure No. 4 illustrates the pin location of COM1 and COM 2 as viewed from the cable attachment side. Refer to the table below for the identity of the pins used.

Figure No. 4

COM 1 (RS-232 only) COM 2 (RS-232 and RS-485)

PIN NO. FUNCTION PIN NO. FUNCTION

1 NO CONNECTION 1 NO CONNECTION

2 DATA INPUT (RXD) 2 DATA INPUT (RXD)

3 DATA OUTPUT (TXD) 3 DATA OUTPUT (TXD)

4 DTR (+9 VOLTS OUTPUT ) 4 RS-485 (-) TRANSMIT

5 SIGNAL GROUND (GND) 5 SIGNAL GROUND (GND)

6 NO CONNECTION 6 RS-485 (-) RECEIVE

7 INTERNALLY CONNECTED TO 8 7 RS-485 (+) TRANSMIT

8 INTERNALLY CONNECTED TO 7 8 RS-485 (+) RECEIVE

9 + 5 VDC @ 200mA max (OUTPUT) 9 + 5 VDC @ 200mA max (OUTPUT)

5 4 3 2 1 9 8 7 6

Page 13: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

9

Figure No. 5 Serial I/O Connections

Page 14: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

10

POWER LAYOUT

Figure No. 6 illustrates the wiring inside the enclosure.

Figure No. 6 Power Connections

Page 15: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

11

Calibration Sealing Screw

SETUP AND CALIBRATION

Your 788 Programmable Weight Indicator has been thoroughly tested and calibrated before being shipped to you. If you receive the indicator attached to a scale, calibration is not necessary. If the 788 is being connected to a scale for the first time or recalibration is necessary for other reasons, proceed as indicated. Calibration of the 788 indicator is accomplished entirely by the keypad. With the power off, first remove the calibration sealing screw on the scale input board on the back of 788, (see Figure No.7). With the screw removed, turn the 788 on with the power switch. The screen will display CARDINAL 788 for a few seconds, and then change to the Start Menu (see Figure No. 8). With the Start Menu showing, press the [SETUP]�key. The 788 will display the FULL SETUP OR FINE ADJUST Menu. Press the [SETUP]�key again. The 788 is now ready for setup and calibration.

Figure No. 7

Figure No. 8

����

��

��

� ����

�����

��

��

��

! "

# $

%

&

'

(

)

*+

,

-

.

/

0�

1 �

�%�!� 2

788 VERSION 03.0

1024K ROM 512K FLASH 521K RAM

DISPLY SEND

REVIEW RECEIVE

SETUP DIAG STD APP�

Page 16: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

12

SETUP AND CALIBRATION, Cont.

During the setup and calibration process it will be necessary to enter operational parameters via the 788’s keypad. Pressing the [ENTER] key without entering a new value will retain the current setting and advance the 788 to the next screen prompt. To change a setting, enter a new value and press the [ENTER] key. This will save the new value and advance the 788 to the next screen prompt.

NOTE! With the exception of the INTERVAL and CAL WT 1 prompts, pressing the [ESC] key at any setup prompt will return the 788 screen to the previous setup step. Pressing the [BKSPC (ESC)] key at the INTERVAL prompt will return the 788 to the Start Menu. Pressing the [BKSPC (ESC)] key at the CAL WT 1 prompt will save all the setup parameters (up to the C1�prompt), then return to the START Menu.

1. After pressing the [SETUP] key, the 788 will prompt: INTERVAL, the interval value (division value) and display the current setting. To change the setting, use the keypad to select a new interval value (1 to 99), then press the [ENTER] key to save the setting and proceed to the next step.

2. The screen will prompt: DPP, the decimal point position and display the current setting.

To change the setting, use the keypad to select a new value for the decimal point position (0 to 5), then press the [ENTER] key to save the setting and proceed to the next step. Numbers zero (0) through five (5) are available and correspond to the following decimal point positions:

0 = XXXXXXX 1 = XXXXXX.X 2 = XXXXX.XX

3 = XXXX.XXX 4 = XXX.XXXX 5 = XX.XXXXX

3. The screen will prompt: CAPACITY, for the scale capacity and display the current setting. To change the setting, use the keypad to enter a new capacity (0 to 9,999,999), then press the [ENTER] key to save the setting and proceed to the next step.

4. The screen will prompt: UNITS, for the weighing units and display the current setting. To

change the setting, use the keypad to select a new value for the weighing units (1 to 3), then press the [ENTER] key to save the setting and proceed to the next step.

1 = Pounds only 2 = Kilograms only 3 = Pounds/Kilograms

5. The screen will prompt: Z TRACK, for the automatic zero tracking range and display the

current setting. This is the value in scale divisions that will be automatically zeroed off. That is, if the scale divisions are five (5) and the zero tracking is set to three (3) the 788 will automatically zero weights of 15 or less. To change the setting, use the keypad to select a new value for the zero tracking range (0, 0.5 or 1 to 9), then press the [ENTER] key to save the setting and proceed to the next step.

Page 17: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

13

SETUP AND CALIBRATION, Cont.

6. The screen will prompt: Z LIMIT, for the zero tracking range limit and display the current setting. If YES is displayed, the push button zero and zero tracking features are limited to a maximum of 4% of the scale capacity. If NO is displayed, the push button zero and zero tracking feature will operate up to the full capacity of the scale. To change the setting, use the keypad to select a YES or a NO, then press the [ENTER] key to save the setting and proceed to the next step.

7. The screen will prompt: FILTER, for the digital filtering and display the current setting. Four

(4) levels of filtering (0, 1, 2 and 3) are available. They are as follows:

0 = NO FILTERING 2 = MODERATE FILTERING 1 = MINIMAL FILTERING 3 = CUSTOM FILTERING

To change the setting, use the keypad to select a new value for the digital filtering, then press the [ENTER] key to save the setting and proceed to the next step.

NOTE! Custom filtering, three (3) is used when zero (0), one (1) or two (2) are inadequate. If you select custom filtering, the 788 will display two (2) more prompts.

If you selected Custom Filtering (3), the 788 will prompt: F MAX for the maximum filter level and display the current setting. The filter level is a number from 1 to 256 that corresponds to the level of filtering with 256 being the greatest filtering and 1 the least. To change the setting, use the keypad to select a new maximum filter level, then press the [ENTER] key to save the setting and proceed to the next step. Next, the 788 will prompt: F BREAK, for the filter break range and display the current setting. The break range is a number from 1 to 64 that corresponds to the number of division change to break out of filtering. To change the setting, use the keypad to select a new value for the filter break, then press the [ENTER] key to save the setting and proceed to the next step.

8. The screen will prompt: MOTION, for the range of motion detection and display the current

setting. Changes in weight exceeding the selected number of divisions will cause the - MOTION - message to display. To change the setting, use the keypad to select a new value from 0 to 9 divisions for the motion range, then press the [ENTER] key to save the setting and proceed to the next step.

9. The screen will prompt: S RATE, for the sample rate and display the current setting. The

value displayed is the sample rate in samples per second. To change the setting, use the keypad to select a new value from one (1) to sixty (60) samples per second for the sample rate, then press the [ENTER] key to save the setting and proceed to the next step.

Page 18: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

14

SETUP AND CALIBRATION, Cont.

10. The screen will prompt: C1, the current value of the C1 number. If the 788 was calibrated previously and you recorded the four (4) "C" (calibration) numbers, you may enter the value for C1. By entering the "C" numbers previously recorded, you can return to that calibration setting without having to use test weights.* If you wish to use test weights for calibration, leave the C1 value unchanged and press the [ENTER] key.

* If any components have been changed that affect calibration and your scale is used in a commercial application and must be "Legal for Trade" you can not use "C" numbers to re-calibrate.

11. The screen will prompt: CAL WT 1, the first of two calibration weights and display the

current setting. This setting could be ZERO (NO LOAD) or the TEST WEIGHTS (TEST LOAD). If the first calibration weight is to be ZERO or NO LOAD, use the keypad to input a 0, then press the [ENTER] key. If the first calibration weight is to be the TEST WEIGHTS or TEST LOAD, use the keypad to input the value of the calibrated test weights. Place these weights on the scale platform, then press the [ENTER] key.

12. After a few seconds, the screen will prompt: CAL WT 2, the second of two calibration

weights and display the current setting. If the first calibration weight was ZERO or NO LOAD, this weight should be equal to the test weight total. If, however, the first weight was the test weight total, then this weight should be zero.

If this second calibration weight is to be zero, make certain the scale platform is empty, use the keypad to input a 0, then press the [ENTER] key. If this second calibration weight is to be the test load, use the keypad to input the value of the calibrated test weights. Place the weights on the scale platform, then press the [ENTER] key.

13. After a few seconds, the screen will prompt: FINE SPAN ADJUSTMENT. Place a known

weight on the scale. Any small error will be displayed and can be corrected as follows:

Press the [UP] key to increase displayed weight 1 grad per press.

Press the [DOWN] key to decrease the weight 1 grad per press.

Press the [ZERO] key to zero the scale.

Press the [EXIT] key to save the settings and proceed to the next step.

14. After a few seconds, the screen will display: CAL DONE, SCALE X CALIBRATED and the “C” numbers. Note that X = the scale number. This indicates the setup and calibration process has been completed. The "C" numbers displayed should be recorded for future use. After recording the "C" numbers, press the [EXIT] key.

Page 19: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

15

SETUP AND CALIBRATION, Cont.

15. The screen will prompt: OIML (YES OR NO) to determine whether or not the indicator will be used in an application requiring a protective interface.

When YES is selected, serial port COM1 will be dedicated to outputting scale gross weight whenever the BASIC application is executing. The BASIC application will not be able to manipulate this output in any way. The serial port communications parameters will be 9600 baud, 8 data bits, no parity bits and 1 stop bit. The format of the data will be:

<lf><s><r><n><m><f><xxxxxx.xxx><uuu><cr>

where:

<lf> = line feed <s> = status flags:

Z for center of zero, O for overcapacity, otherwise, a space

<r> = 1 <n> = G <m> = motion status:

M if scale is in motion, space if not <f> = scale number:

1 for scale 1, 2 for scale 2, etc. up through 9 for scale 9. A is for scale 10, B for scale 11, and so on.

<xxxxxx.xxxx> = weight <uuu> = units (lb or kg ) <cr> = carriage return

The data will be sent out at a rate of from 1 to 2 times per second, depending upon the sample rate.

To change the setting, use the keypad to select a YES or a NO, then press the [ENTER] key to save the setting and proceed to the next step.

The instrument will reset, display the Start Menu and then advance to the Weight Screen. Replace the calibration sealing screw on the scale input board, and proceed with normal operations.

NOTE! Changes to the 788 setup parameters can be performed WITHOUT affecting the calibration. This is accomplished by entering the changes necessary up to the screen prompt C1 and pressing the [ENTER] and [ESC] keys. That is, follow the standard Setup and Calibration procedure up to the screen prompt, C1. Then instead of entering a value for C1, press the [ENTER] key, then at the CAL WT 1 prompt, press the [ESC] key. The instrument will reset, display the Start Menu and then load the application program or standard program. Replace the calibration sealing screw on the scale input board, and proceed with normal operations.

This allows inputting new Setup parameters, without the need to re-calibrate the scale.

Page 20: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

16

SCALE TOTALIZER

The scale totalizer will sum the weights of two or more scales and make the total available to the application program. The totalizer can be referenced as if it were another scale. The totalizer’s scale number will be one plus the actual number of scales attached to the indicator.

Setup To setup the totalizer, remove the calibration sealing screws from all scale cards and press the [SETUP] key on the indicator’s main menu. If the scale cards have not been setup, continue with the scale card setup. Otherwise, at the SCALE ID prompt, bypass the scale card setup screens by pressing the [ESC] key. The display will change to show:

TOTALIZE (YES OR NO) CURRENT=??

Selecting NO (and pressing the [ENTER] key) will disable the scale totalizer. Selecting YES (and pressing the [ENTER] key) will enable the scale totalizer. Pressing the [ESC] key will leave the selection as is and proceed to the next setup function or back to the main menu. If YES is selected (and the [ENTER] key pressed), the display will show:

INCLUDE SCALE 1

CURRENT=?? Selecting YES (and pressing the [ENTER] key) means that weight from scale 1 will be used by the scale totalizer. Selecting NO (and pressing the [ENTER] key) means that scale 1 will not be a part of the totalizer. Note, that all the scales to be included in the totalizer must have the same scale division value (interval and decimal point precision), sample rate, and weight units. If this is not true, the error message:

INVALID SETUP FOR TOTALIZE

will be displayed. Press the [ESC] key to return to the INCLUDE SCALE prompt and select NO, then press the [ENTER] key to proceed to the next setup function. In addition, the firmware for the 778A card must be version 00.4 or higher, and the firmware for the 778K card must be version 00.2 or higher. If this is not true, the error message:

INVALID REV FOR TOTALIZE

will be displayed. Press the [ESC] key to return to the INCLUDE SCALE prompt and select NO, then press the [ENTER] key to proceed to the next setup function.

Page 21: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

17

SCALE TOTALIZER, Cont.

After the appropriate selection for including scale 1 has been entered, the display will ask if scale 2 should be included in the totalizer. The question will be repeated for each scale attached to the indicator. The display will then change to show:

CAPACITY (1 TO 9,999,999) CURRENT=??

The value entered, in primary weight units, will be used in determining when the totalizer’s overcapacity status will become true. The totalizer’s overcapacity status will be true if any scale is overcapacity or if the total gross weight exceeds the totalizer capacity by more than 9 scale divisions. After the totalizer capacity has been entered, the display will show:

MOTION (0 TO 9)

CURRENT=??

The value entered, in scale divisions, will be used in determining when the totalizer’s motion status will become true. The totalizer’s motion status will be true if any of the totalized scales is in motion or if the total gross weight changes by more than totalizer’s motion range. This completes totalizer setup. Operation When operating in STD mode, the totalizer will be identified by a T. The [TARE] key will not be active when the display is showing the totalizer weight. If the [ZERO] key is pressed while displaying the totalizer weight, all of the scales included in the totalizer will be zeroed. If the [ZERO] key is pressed while displaying a scale weight, only that scale will be zeroed. If pounds-kilograms conversion is enabled, pressing the [UNITS] key, when displaying the totalizer or any of the scales included in the totalizer, will switch the weight units of the totalizer and the weight units of all the scales included in the totalizer. Program Commands The application program can use the standard scale commands and functions to access the totalizer. The totalizer may be accessed by using the totalizer’s scale number in the command or function, or by using the CURSCALE command to make the totalizer the default scale. The function CURSCALE will return the totalizer’s scale number if the totalizer is the default scale. The function CARDS(17) will return one if the totalizer is enabled, zero if it is not. The SCALES function will return the total number of actual scales, not including the totalizer. The GROSS, TARE, NET functions will return the totals of the gross, tare, and net weights, respectively, for all of the scales included in the totalizer. The CZERO function will return a nonzero value if and only if all of the scales included in the totalizer are at center of zero. The MOTION function will return a nonzero value if any of the scales included in the totalizer is in motion, or if the totalizer gross weight changes by more than the totalizer motion range.

Page 22: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

18

SCALE TOTALIZER, CONT.

The OVERCAP function will return a nonzero value if any of the scales included in the totalizer is overcapacity, or if the totalizer gross weight is overcapacity. An error condition for the totalizer or any of the scales included in the totalizer will also cause OVERCAP to return a nonzero value. The BLOZERO function will return a nonzero value if the gross weight of any of the scales included in the totalizer is below zero, or if the totalizer gross weight is below zero. The CAPACITY function will return the capacity of the totalizer. The INTERVAL function will return the scale division value for the totalizer. The totalizer and all of the scales included in the totalizer must have the same scale division value. The WTMODE function will return 1 if the totalizer displays weight in pounds only, 2 if the totalizer displays weight in kilograms only, and 3 if the totalizer displays weight in pounds and kilograms. The totalizer and all of the scales included in the totalizer must have the same weighing mode. The CURUNIT function will return 0 if the totalizer is weighing in pounds, 1 if it is weighing in kilograms. The UNIT$ function will return "LB" if the totalizer is weighing in pounds, "KG" if it is weighing in kilograms. The TARETYPE function will return 0 if no tare is stored, 1 if a tare was stored for at least one of the scales included in the totalizer using the DOTARE command, and 2 if a tare was stored for at least one of the scales included in the totalizer using the TARE command. All tare weights stored for the scales included in the totalizer must be of the same type. The SCLSTAT function will return information about the scale’s status. The information is encoded in the bits of the return value and can be determined by AND’ing the return value with the appropriate mask value: SCLSTAT AND 2 = 1 = One or more of the scales included in the totalizer had some sort of

failure or the total weight could not be calculated because the scales were incompatible.

SCLSTAT AND 8 = 1 = Analog failure of one or more of the scales included in the totalizer.

SCLSTAT AND 32 = 1 = This scale is included in the totalizer.

SCLSTAT AND 64 = 1 = This scale is the totalizer. The CURUNIT command can be used to set the weight units for totalizer and its associated scales if pounds-kilograms conversion is enabled. The totalizer and all of its associated scales will be set to the same units when the command is executed. The DOZERO command can be used to individually zero any of the scales associated with the totalizer. If the totalizer is specified as the target of the command, all of the scales included in the totalizer will be zeroed.

Page 23: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

19

SCALE TOTALIZER, CONT.

The DOTARE command can be used to individually tare any of the scales associated with the totalizer. If the totalizer is specified as the target of the command, it is as if the command were repeated once for each scale included in the totalizer. If a keypad tare has already been stored for one or more of the scales included in the totalizer, the DOTARE command can only be used to clear the tare (that is, the gross weight is 0). It cannot store a new tare since that could result in more than one type of tare being stored. The TARE command can be used to individually tare any of the scales associated with the totalizer. The command cannot be used with the totalizer. If a pushbutton tare has already been stored for one or more of the scales included in the totalizer, the TARE command can only be used to clear the tare (that is, set the tare to 0). It cannot store a new tare since that could result in more than one type of tare being stored.

Contrast Adjustment

The display contrast can be adjusted (without returning to the Start Menu and selecting Display) during normal operations with the following steps:

1. Press and hold the [SHIFT] key.

2. Press the [ENTER] key to increase the display contrast OR

3. Press the [SPACE] key to decrease the display contrast.

4. When the contrast is at the desired level, release the [SHIFT] key.

5. To save the contrast level setting, press and hold the [SHIFT] until a long beep is heard.

6. The indicator is ready for normal operations.

Page 24: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

20

RECEIVING AND SENDING PROGRAMS

The 788 Programmable Weight Indicator can receive and send application programs to and from another 788 indicator or with the optional SmartWeigh software development package to and from a computer. To transfer application programs between 788 indicators, one indicator must be in the Receive mode, the other in the Send mode and an 8545-B099-0A (778 to HOST PC) transfer cable must be connected from the COM port of one indicator to the COM port of the other indicator. To transfer application programs between the computer and the 788 indicator, the SmartWeigh program must be loaded and running on the computer, and an 8545-B099-0A (778 to HOST PC) transfer cable connected between the COM port on the computer and the COM port of the 788. The indicator and the SmartWeigh program must be in the appropriate modes (one in send, the other in receive) to transfer application programs. Refer to the SmartWeigh Programming and Simulator Manual (8545-M095-O1) for operation of SmartWeigh transfer procedures. Receive A Program

The steps to receive (load) an application program on the 788 are as follows:

1. With the power off, first remove the calibration sealing screw on the scale input board on the back of 788, see Figure No.13. With the screw removed, turn the 788 on with the power switch. The screen will display Cardinal 788 for a few seconds, then change to the Start Menu.

NOTE! The calibration sealing screw must be removed before the indicator can receive an application program. If the it screw is not removed, the indicator will display:

CAL SEAL

PRESS ENTER

Press the [ENTER] key to return to the Start Menu, then remove the screw before attempting to receive an application program again.

2. While the Start Menu is displayed, press the [RECEIVE] key.

3. The screen will display, SELECT COM 1 OR COM 2.

4. Press the [COM1] key or the [COM2] key depending on which port you are using to receive the program OR press the [ESC] key to QUIT the load procedure and return to the Start Menu.

Page 25: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

21

RECEIVING and SENDING PROGRAMS, Cont.

Receive A Program, Cont.

5. The screen will display, RECEIVE? and the current baud rate selected.

� If the baud rate displayed is correct, press the [YES] key to continue and proceed to step 6.

� If the baud rate displayed is incorrect or you wish to use a different rate, press the [BAUD] key. The screen will display a list of baud rates available. Select the desired baud rate, then press the [YES] key to continue and proceed to step 6.

� When the [NO] key is displayed, the load procedure may be canceled. Pressing the [NO] key will display, CANCELLED LOCALLY. Press the [OK] key to return to the Start Menu.

6. The screen will display the following warning message:

WARNING APPLICATION PROGRAM

WILL BE ERASED DO YOU WISH TO PROCEED?

YES NO

7. If the [NO] key is pressed, the operation will be cancelled and the current application

program will be unchanged.

8. If the [YES] key is pressed, the current application program will be erased in preparation for downloading a new application program.

9. The screen will display ERASING PROGRAM and erase the application currently stored in the 788 memory.

10. After erasing the program, the screen will display:�

SETUP REMOTE TO SEND PRESS -READY- TO CONTINUE

READY

11. Make sure the remote (computer or other 788 indicator) is ready to send, and press

the [READY] key.

Page 26: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

22

RECEIVING and SENDING PROGRAMS, Cont.

Receive A Program, Cont.

12. The screen will display

RECEIVING FROM COM x

Receiving record # S

where: x = com port selected in step 4 # = record number the indicator is receiving S = a status message

13. To stop or pause the load process, press the [ESC] key. The screen will display

QUIT?

YES NO

� To stop receiving (loading) the program, press the [YES] key. The screen will display, CANCELLED LOCALLY. Press the [OK] key to return to the Start Menu.

� Press the [NO] key to continue receiving (loading) the program from the point at which it was interrupted.

14. When the 788 is finished loading (receiving) the program, the screen will display DONE. Press the [OK] key to return to the Start Menu, then press the [APP] key to load and run the program.

NOTE! To load and run your program quicker, press the [OK] key twice. This will bypass the five (5) second delay at the Start Menu.

Page 27: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

23

RECEIVING and SENDING PROGRAMS, Cont.

Send A Program

The steps to send an application program to the 788 are as follows:

1. Apply power to the 788 indicator.

2. While the Start Menu is displayed, press the [SEND] key.

3. The screen will display, SELECT COM 1 OR COM 2.

4. Press the [COM1] key or the [COM2] key depending on which port you are using to send the program OR press the [ESC] key to ���� the send procedure and return to the Start Menu.

5. The screen will display, SEND? and the current baud rate selected.

� If the baud rate displayed is correct, press the [YES] key to continue and proceed to step 6.

� If the baud rate displayed is incorrect or you wish to use a different rate, press the [BAUD] key. The screen will display a list of baud rates available. Select the desired baud rate, then press the [YES] key to continue and proceed to step 6.

� When the [NO] key is displayed, the load procedure may be canceled. Pressing the [NO] key will display, CANCELLED LOCALLY. Press the [OK] key to return to the Start Menu.

6. Make sure the remote (the computer or another 788 indicator) is ready and press the YES key to start sending. The screen will display:

SENDING TO COM x

Sending record # S

where: x = com port selected in step 4 # = record number the indicator is receiving S = a status message

7. To stop or pause the send process, press the [ESC] key. The screen will display QUIT?.

� To stop sending the program, press the [YES] key. The screen will display, CANCELLED LOCALLY. Press the [OK] key to return to the Start Menu.

� Press the [NO] key to continue sending the program from the point at which it was interrupted.

8. When the 788 is finished sending the program, the screen will display DONE. Press the [OK] key to return to the Start Menu, then press the [APP] key to load and run the program.

NOTE! To load and run your program quicker, press the [OK] key twice. This will bypass the five (5) second delay at the Start Menu.

Page 28: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

24

778C SERIAL CARD The 778C Dual Serial Interface card has been designed to interface to a communications device that uses either RS-232 or 20mA current loop. The 778C when coupled with the 788 Programmable Weight Indicator provides two full duplex DTE connections that can operate at data rates up to 19.2K bps. The serial input and output functions can be programmed by the customer from the 788 main terminal without the need for complicated programming devices. The 778C card resides in the 788 main enclosure and has a 16 pin dual in-line Parallel Internal Bus (PIB) connector and a 4 pin single in-line power connector to provide an easy connection to the 788 internal hardware. The 778C card has two external connectors, a DB-25P (In the UPS application is used for the second barcode scanner in the two scale application) and a DE-9P for connection to serial devices. There is also an 8K buffer on the card to buffer the serial data while waiting on the serial device. This "spooling" feature improves the throughput of the 788 system.

Figure No. 9 Specifications

Temperature Range: 14 to 104 �F (-10 to +40 �C)

Dimensions: 5.75" L x 5.6" H x 0.8" D

Internal Connections: (1) PIB (16 pin DIL) and (1) power (4 pin SIL)

External Connections: (1) DB-25P and (1) DE-9P

(optional (2) SMA Fiber Optic)

Input/Output Ports: (2) Full Duplex RS232 or 20mA

(optional Fiber Optic)

Card Address Range: 0 to 15 (jumper selectable)

Buffer Size: 8192 bytes

Page 29: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

25

Setting Card Address The 788 can address up to 16 Dual Serial Interface option cards. In most cases, you do not have to change the factory default jumper settings. However, when using multiple serial interface cards, each card must have a unique address. Jumpers on the card allow for 16 different addresses. Additional jumpers, if needed are available by ordering Cardinal Part number 6610-2023. The table below shows the jumper settings for each card installed and the corresponding ID as displayed in the 788 self DIAGnostic procedure.

Jumper A0 Jumper A1 Jumper A2 Jumper A3 CARD # ID

OFF OFF OFF OFF 1 20

ON* OFF* OFF* OFF* 2* 21*

OFF ON OFF OFF 3 22

ON ON OFF OFF 4 23

OFF OFF ON OFF 5 24

ON OFF ON OFF 6 25

OFF ON ON OFF 7 26

ON ON ON OFF 8 27

OFF OFF OFF ON 9 28

ON OFF OFF ON 10 29

OFF ON OFF ON 11 2A

ON ON OFF ON 12 2B

OFF OFF ON ON 13 2C

ON OFF ON ON 14 2D

OFF ON ON ON 15 2E

ON ON ON ON 16 2F

A3 A2 A1 A0*

� � � �

� � � �

Address jumpers as viewed from component side of card.

* FACTORY DEFAULT SETTING

Page 30: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

26

778A ANALOG CARD The second analog card is used for the two scale applications. All 778A cards are identical and provided with a jumper to set for the scale identification. The 778A Scale Input Card has been designed to meet the requirements of simultaneous weight data display from multiple scales to a single 788. Analog and programmable digital filtering is used for stable weight readings. It can be used singularly or in combinations of up to sixteen (16) cards. Calibration is digital and stored in nonvolatile memory. The customer can easily configure the 778A from the 788 Programmable Weight Indicator main terminal without the need for complicated programming devices. The 778A card resides in the 788 main enclosure and has a 16 pin dual in-line Parallel Internal Bus (PIB) connector and a 4 pin single in-line power connector to provide an easy connection to the 788 internal hardware. The 778A card has one external 9 pin "D" type (DE-9S) connector for connecting to strain gage weigh transducers and provisions for wire sealing. Figure No. 10 Specifications

Temperature Range: 14 to 104 °F (-10 to +40 °C)

Displayed Resolution: 10,000 divisions (commercial)

100,000 divisions (maximum, non-commercial)

Internal Resolution: 400,000 divisions

Internal Connections: (1) PIB (16 pin DIL) and (1) power (4 pin SIL)

External Connections: one (1) DE-9S Load Cell connector

Sample Rate: selectable 1 to 60 samples per second

Auto Zero Range: 0.5 divisions or 1 through 9 divisions

Load Cells: Up to 12 350 � load cells (with sense)

Excitation Voltage: 10.75 VDC

Sensitivity Range: 30 nV/V/ division maximum

Division Value: Programmable

Weighing Units: Programmable

Card Address Range: 0 to 15 (jumper selectable)

Page 31: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

27

Setting Card Address The 788 can address up to 16 Scale Input cards. In most cases, you do not have to change the factory default jumper settings. However, when using multiple scale input cards, each card must have a unique address. Jumpers on the card allow for 16 different addresses. Additional jumpers, if needed are available by ordering Cardinal Part number 6610-2023. The table below shows the jumper settings for each card installed and the corresponding ID as displayed in the 788 self DIAGnostic procedure. See Figure No. 1 for board jumper location.

Jumper A0 Jumper A1 Jumper A2 Jumper A3 CARD # ID

OFF OFF OFF OFF 1 10

ON* OFF* OFF* OFF* 2* 11*

OFF ON OFF OFF 3 12

ON ON OFF OFF 4 13

OFF OFF ON OFF 5 14

ON OFF ON OFF 6 15

OFF ON ON OFF 7 16

ON ON ON OFF 8 17

OFF OFF OFF ON 9 18

ON OFF OFF ON 10 19

OFF ON OFF ON 11 1A

ON ON OFF ON 12 1B

OFF OFF ON ON 13 1C

ON OFF ON ON 14 1D

OFF ON ON ON 15 1E

ON ON ON ON 16 1F

A3 A2 A1 A0*

� � � �

� � � �

Address jumpers as viewed from component side of card.

* FACTORY DEFAULT SETTING

Page 32: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

28

ERROR AND STATUS MESSAGES The 788 is equipped with a diagnostic software program that tests various portions of the instrument’s circuitry and verifies proper operation. Should a problem be detected, an error or status message will be displayed alerting the operator to that condition. The following lists these errors and status messages and their meaning.

Message

POSSIBLE CAUSE CORRECTIVE ACTION

��������

This message indicates that motion is present when attempting to power up, print, zero or perform a push button tare function.

Insure scale is stable, level and vibration-free. Wait for a stable weight display.

������

The ZERO key was pressed when the gross weight is outside the scale zero weight range.

Check Setup parameter Z LIMIT. If YES, only 4% of scale capacity can be zeroed. Change as required.

����� � �����

This message indicates the weight on scale exceeds the scale capacity.

Remove the weight from the scale platform. Insure the scale is at zero. If problem persists, contact your scale serviceman.

�����������

Scale Input Board (SIB) data interface not functioning.

Both data and power cables must be plugged into SIB. If cables check okay, replace SIB.

� ��� �

Attempting to enter Setup and Calibration mode with the calibration sealing screw installed.

Remove the calibration sealing screw from the Scale Input Board (SIB) mounting bracket (refer to Figure No. 10).

�����������

This message is displayed to indicate that the weight is within +/- 1/4 division of the center of zero.

No corrective action is necessary. This is a status message indicating the automatic zero tracking is working properly.

There are many operating system error codes. These errors are rare.

The cause for an operating system error may be caused by bad RAM disk memory. This memory may be corrupted by several causes.

The corrective action is to follow the procedure for clearing of RAM memory. In doing this all RAM memory is cleared therefore clearing all RAM disk files.

Page 33: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

29

BEFORE YOU CALL FOR SERVICE The 788 has been designed to provide you with years of trouble-free operation. In spite of this, troubles sometimes happen. Before calling for service assistance you should make some initial checks to verify that a problem does exist. The following describes several types of symptoms along with suggested remedies.

Problem Possible Solutions Screen does not turn on Is AC power cord plugged into wall

receptacle? Check wall receptacle for proper AC power. Try another electrical appliance in the same receptacle, does it work? Check the circuit breaker. Has there been a power failure of any kind?

Incorrect weight displayed Has the instrument been calibrated? Insure that the scale platform isn’t touching an adjacent object. Check the load cell connector wiring. Have proper operation procedures been followed?

Indicator will not display weight Refer to Error and Status Messages section.

Page 34: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

30

PROBLEM RESOLUTION 788 Problem Resolution Overview From time-to-time, it may be necessary to perform some on-site troubleshooting

when the Automated Ramp Scale equipment or its associated systems are not responding as expected.

The following tables should always be referenced as a first step in determining the cause of the problem.

788 Programmable Weigh Indicator - Problem Resolution Table

Problem Possible Cause Solution No Display � No Power

� Disconnected display cable

� Turn on power to 788 � Reconnect display

ribbon cable (either at the inside, front panel, or at the Main Circuit Board inside)

Vertical lines on LED � Damaged display

ribbon cable � Partially disconnected

display ribbon cable

� Power off 788 and check cable for damage and replace if necessary

� Power off 788 and reseat display cable at LED and Main Circuit Board

“No Response from Operator” message is displayed

� Timeout occurred, operator failed to press an available option after scanning/keying ULD data.

� Re-process the ULD

788 is “Locked-up” or “Won’t Respond”

� Unit too cold � Power surge � If intermittent, ASMA

(Scale Host) isn’t running

� Power on and allow to warm up 10-15 minutes before use (especially in cold weather)

� Power off, wait 5-10 seconds and power back on

� Verify that ASMA (Scale Host) is running

Page 35: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

31

Problem Resolution, Cont. 788 (continued)

Problem Possible Cause Solution “Server Not Connected” message is displayed

� Lantronix device not receiving AC power

� Ethernet switch functioning improperly

� May also be caused by a faulty Lantronix Device Server

� Verify that AC adapter is plugged in

� Verify Ethernet switch functions

� Verify that the Lantronix Device Server is functioning properly.

PSC Scanner won’t scan It has a solid green LED

� System lock-up � Defective scanner

� Power the 788 off for 5-10 seconds and restart

� Replace the PSC Scanner

PSC Scanner won’t scan It appears to have a red LED when the trigger is pressed and the scanner emits a scan beam

� Dirty lens � Cracked/scratched lens � Condensation inside

scanner

� Clean lens � Replace scanner if lens

is cracked/scratched or if condensation exists

PSC Scanner won’t scan No lights are working

� Internal disconnect � Defective scanner

cable � Defective pigtail � Defective scanner

� Verify that scanner is connected to appropriate COM port

� Verify that COM port pigtail is properly terminated

� Try a replacement scanner cable

� Try a replacement pigtail cable

� Replace the scanner

Scanner works fine, but occasionally will not read all ULD barcodes

� Barcode damaged � Too faded for scanner

to decode

� Press the ULD button for the corresponding scale and type in the ULD number including the IATA code, the ID number, and the owner code (i.e., AAY8281UPS - where AAY is the IATA code, 8281 is the ID number, and UPS is the owner code)

Page 36: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

32

788 Problem Resolution, Cont. 788 (continued)

Problem Possible Cause Solution Scale doesn’t show zero (0) weight when nothing is on the scale.

� Environmental conditions (i.e., strong wind)

� Access the Zero Scale option to reset the scale to zero (make sure nothing is on the scale)

“Server App Active” message is displayed.

� This is not a problem, it is a normal system response to a periodic communication test

� When no activity takes place, the 788 will periodically send a “Heartbeat” message to the ASMA (Scale Host) to make sure it can still communicate to it

Strange number show up on the 788 display that are not valid weights

� Improperly terminated scale input cable

� Verify that the scale input cable is terminated on the 788 Indicator connector instead of the Existing indicator connector

Page 37: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

33

ASMA Problem Resolution Overview The Automated Scale Monitoring Application (ASMA) is responsible for gathering

and storing ULD data collected at the scales. In the event that data related problems occur, ASMA may be the cause. The following problem resolution table should be referenced if ASMA is believed to be the cause of a data related problem.

ASMA Automated Scale Monitoring Application (ASMA) Problem Resolution Table

Problem Possible Cause Solution ULD’s have been processed, but some are not automatically updating in DWB

� ULD’s were processed as “Load Not Found”

� Load Not Found ULD’s do not update on the DWB Server

� Scale operator failed to accept or reject the returned load information

� Scale operator will need to press “ACCEPT” after this message displays to save data locally on the ASMA host or “REJECT” to discard it and begin again

� Data can then be queried to obtain the capture

A specific ULD can not be found using the query function

� Improperly entered ULD number

� ULD was originally weighed more than seven days ago.

� Re-enter the ULD number (with or without the IATA and owner codes)

� Verify that the ULD was weighed less than seven days ago, otherwise it has most likely been purged from the system

Page 38: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

34

DWB Problem Resolution Overview Data collected through the Automated Ramp Scales are transmitted via the

ASMA Scale Host and used to populate the Distributed Weight and Balance (DWB) system with valid ULD weights. On occasion, there may be reason to question the data being populated into the DWB system. This problem resolution table offers a starting point for resolving such issues.

DWB Distributed Weight and Balances (DWB) Problem Resolution Table

Problem Possible Cause Solution The weights are not updating for the ULDs

� In connected mode, but screen hasn’t been refreshed.

� Not in connected mode.

� In connected mode (i.e., a green light appears on the bottom right corner of the screen), simply click the REFRESH button

� If not connected, you must query the ASMA manually and key the capture ULD weights

ULD’s are updating with zero weight

� Container is being scanned or keyed and saved before the container actually reaches the scale

� Wait until the ULD is positioned on the scale and allow the weight to stabilize before scanning/keying the ULD

Page 39: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

35

PROCEDURE FOR RE-FORMATTING RAM DISK ON MODEL 788 DIGITAL WEIGHT INDICATOR STEP 1: POWER UP INDICATOR. THE SCREEN BELOW WILL APPEAR AFTER THE CARDINAL LOGO NOTE: SCREEN TEXT MAY DIFFER SLIGHTLY FROM SITE TO SITE

Figure No. 11

VERSION 00.9 1024K ROM 2048K FLASH 2048K RAM REVIEW

SEND

RECIEVE

SETUP DIAG STD APP

PUSH “DIAG” BUTTON TO MOVE TO THE NEXT SCREEN

Page 40: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

36

STEP 2:

Figure No. 12

APPLICATION SOFTWARE

FILE: CF-REV-S.BAS DATE: 12/23/1998 REG NO: 21803544 CO NAME: CARDINAL SCALE CK SUM: OK

TRANSLATED

PREV NEXT

PUSH “NEXT” TO MOVE TO THE NEXT SCREEN

Page 41: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

37

STEP 3:

Figure No. 13

PIB CARD TEST SIB: 11 000 ERROR SIO: 21 000 ERROR

PREV NEXT

PUSH “NEXT” TO MOVE TO THE NEXT SCREEN

Page 42: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

38

STEP 4:

Figure No. 14

RAM DISK 159552 TOTAL BYTES XXXXX BYTES IN XXX USER FILES XXXXXXX BYTES FREE

PREV NEXT

THIS BUTTON IS NOT LABELED. PUSH THIS “HIDDEN” KEY TO START THE RE-FORMATTING PROCESS

NOTE: THIS SCREEN MAY COME UP WITH A STATEMENT: “XXX BAD CLUSTERS” INSTEAD OF THE SCREEN SHOWN ABOVE.

Page 43: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

39

STEP 5:

Figure No. 15

WARNING FORMATTING

WILL PERMANENTLY ERASE ALL DATA IN THE RAM DISK

PREV NEXT

PUSH “NEXT” TO MOVE TO THE NEXT SCREEN

Page 44: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

40

STEP 6:

Figure No. 16

FORMAT?

YES NO

PUSH “YES” TO FORMAT RAM DISK

Page 45: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

41

STEP 7:

DISPLAY WILL FLASH “FORMATTING RAM DISK” BEFORE GOING TO NEXT SCREEN

Figure No. 17

ROOT DIR (16 TO 256 ENTRIES)

CURRENT=256

?

1

2

3

4

YES\5 NO\0

9

8

7

6

ESC BKSPC . ENTER

PUSH “ENTER” TO MOVE TO NEXT SCREEN

Page 46: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

42

STEP 8:

Figure No. 18

RAM DISK 1559552 TOTAL BYTES 1559552 BYTES FREE

PREV NEXT

PUSH “NEXT” TO MOVE TO THE NEXT SCREEN

Page 47: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

43

STEP 9:

Figure No. 19 WHEN YOU PUSH “NEXT” AT THIS SCREEN, THE DISPLAY WILL GO TO THE MAIN SCREEN AND AFTER A FEW SECONDS THE PROGRAM WILL BEGIN TO RUN. YOU WILL NOW HAVE TO RE-ENTER THE OPERATOR ID’s AND THE FORLIFT ID’s AND THEIR ASSOCIATED TARE WEIGHTS

RAM DISK DIRECTORY VOLUME IN A: 788-009 DIR OF A:\

PREV NEXT

PUSH “NEXT” TO COMPLETE THE FORMATTING PROCESS

Page 48: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

44

NOTES

Page 49: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

45

Page 50: Calibration Manual - midwestern.decardinalscale.midwestern.de/.../8545-M258-O1_788UPSCalibrationMa… · 788 Programmable Weight Indicating Instrument Calibration Manual MODEL 788SUPS-1

UNAM&

EDSI

GUN

ED

FACTRED

I N U S A...

UNAM&

EDSI

GUN

ED

FACTRED

I N U S A...