Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s guide for NMR data collection using Bruker’s IconNMR Automation software for the 400 MHz Avance-IIIHD Bruker NMR spectrometer located in Chem93 managed by the UC Davis NMR Facility. For detailed help with both routine and advanced NMR experiments, please consult Bruker’s User Guides, which can be found within the Resources tab on our website, nmr.ucdavis.edu. CONVENSIONS Some user control will be done in Bruker’s Topspin 3.5 interface. For Topspin interaction, keyboard input is shown as boldface type in this manual. Note that in TS the “enter” key must be used after the command is typed; this is assumed through-out this manual and “enter” key strokes are not given explicitly. Commands in TS are typed in on the TS command line near the bottom of the TS window; again this is assumed and will not generally be stated explicitly herein. LMB, MMB, and RMB are used to indicate actions of the left, middle, and right mouse button respectively. On a PC the mouse wheel acts as the MMB. Click, and double click refer to pressing the LMB. GENERAL PROCEDURE You will find that the general procedure for acquiring NMR data on all NMR spectrometers is essentially the same. IconNMR handles the majority of the tasks listed below. The general procedure is as follows: 1 Sample Preparation 2 Login and Startup 3 Setup Initial Parameters 4 Check/Edit Acquisition parameters 5 Insert Sample 6 Lock onto Solvent 7 Tune the Probe 8 Shim 9 Check Receiver Gain 10 Acquire 11 Process Data 12 Remove Sample 13 Logout, Sign Logbook

Brief IconNMR and Topspin 3.5 User Guide for Bruker … IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s

Embed Size (px)

Citation preview

Page 1: Brief IconNMR and Topspin 3.5 User Guide for Bruker … IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s

BriefIconNMRandTopspin3.5UserGuideforBrukerNMRSpectrometers

AvanceIIIHD400MHzNMRwithAutosamplerinChemistryroom93

DISCLAIMER Thisdocumentis intendedtobeabrief,bare-bonesuser’sguideforNMRdatacollectionusingBruker’s

IconNMRAutomationsoftwareforthe400MHzAvance-IIIHDBrukerNMRspectrometerlocatedinChem93managed by the UC Davis NMR Facility. For detailed help with both routine and advanced NMRexperiments,pleaseconsultBruker’sUserGuides,whichcanbe foundwithin theResources tabonourwebsite,nmr.ucdavis.edu.

CONVENSIONSSomeusercontrolwillbedoneinBruker’sTopspin3.5interface.ForTopspininteraction,keyboardinputisshownasboldfacetypeinthismanual.NotethatinTSthe“enter”keymustbeusedafterthecommandis typed; this is assumed through-out this manual and “enter” key strokes are not given explicitly.Commands inTSaretyped inontheTScommand linenear thebottomof theTSwindow;againthis isassumedandwillnotgenerallybestatedexplicitlyherein.LMB,MMB,andRMBareusedtoindicateactionsoftheleft,middle,andrightmousebuttonrespectively.OnaPCthemousewheelactsastheMMB.Click,anddoubleclickrefertopressingtheLMB.

GENERALPROCEDURE YouwillfindthatthegeneralprocedureforacquiringNMRdataonallNMRspectrometersisessentiallythe

same.IconNMRhandlesthemajorityofthetaskslistedbelow.Thegeneralprocedureisasfollows:1 SamplePreparation2 LoginandStartup3 SetupInitialParameters4 Check/EditAcquisitionparameters5 InsertSample6 LockontoSolvent7 TunetheProbe8 Shim9 CheckReceiverGain10 Acquire11 ProcessData12 RemoveSample13 Logout,SignLogbook

Page 2: Brief IconNMR and Topspin 3.5 User Guide for Bruker … IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s

Experiment: RoutineProtonNMRParameterSet: PROTONInstrument: ChemBruker400usingIconNMR

SamplePreparation1. DissolveyoursampleinanappropriatedeuteratedNMRsolvent.Makesurethere

isnoun-dissolvedmaterial.Ifthereis,youwillneedtoeithercentrifugeorfilteryoursampletoremovecrystals/debris.

2. Transferabout550-600uLofsolventintoacleanNMRtube.Werecommendhigh-qualityNMRtubes-rated600MHzorhigher,buteconomytubeswillbeOKforroutineworkatlowerfields(400MHzandbelow).Yoursampleheightshouldbeabout4-5cm.

LoginandStartup

1. Logintoyouriconnmruseraccount.Ifaprevioususeriscurrentlyloggedin,youwillneed

toselectChangeUser,thenlogintoyouraccount.

TIP:useasmallamountofKimWipeinaPasteurPipettetoquicklyfilteroutanyundissolvedmaterial

NOTE:IfyourIconNMRaccounthasnotyetbeencreated,pleasereachouttoBennettAddison:[email protected]

Page 3: Brief IconNMR and Topspin 3.5 User Guide for Bruker … IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s

SetInitialParameters1. IdentifyanemptyholderintheBACS120autosampler.Westronglyrecommendyouuse

theholdersintheinnerring:holders1-60.2. Placeyoursampleintoanavailablebluespinner,setdepthusingthedepthgauge,and

dropyoursampleoffintheappropriateholder.DoNOTdropoffyoursamplewhilethecarouselorsample-armismoving.

3. IdentifythematchingholderinIconNMRmenuanddouble-clicktoaddanexperiment.4. Enteryourdetails:DataDirectory,ExperimentName,ExperimentNumber,Solvent,

Experiment/ParameterSet.

Directory:YourdirectoryshouldbebasedonyourPI.Itissetautomaticallywhenyoulogin.

ExperimentName.Usersoftenincludeinitials,labnotebookpage,date,andanyadditionalinfoifdesired.Notedonotusespaces,specialcharacters,etc.

ExperimentNumber.Thedefaultisincrementsof10(10,20,30…)foreachsample/holder,orincrementsof1(10,11,12…)ifyouarecollectingmultipleexperimentsonthesamesample/holder.

Solvent:Selectyoursolventfromthedrop-downmenu

Experiment:Selectyourexperimentfromthedropdownlist.Ifthereisanexperimentyouwouldliketorunbutitisnotonthelist,contactBennett.

Useinnerring:Innerringisthelowernumbers,outerringhighernumbers

Page 4: Brief IconNMR and Topspin 3.5 User Guide for Bruker … IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s

Check/EditAcquisitionParameters1. Ifdesired,hittheequalsbuttontoeditstandardProtonacquisitionparameters

includingnumberofscans,spectralwidth,acquisitiontime,offsetfrequency,etc.

2. Editthetitleifdesired.Youcanaddanysample/experimentdescriptionin

thistextbox.Youmayfinditusefullaterforyourbookkeeping.

3. EdittheStartTimeifdesired.Theclockissetto24hrtime.Note,ifyousetastarttimebetweenthehoursof8pm–8am(night-timequeue),pleasemakeareservationontheonlinescheduler.

4. Ifyouwouldliketoalsocollectanotherexperimentonthissample(e,g,C13CPD),you

canhighlightthecurrentexperimentandthenhittheAddbutton.Anotherexperimentwillappearbelowwithexperimentnumber11andanemptyexperimenttype.Selecttheappropriateexperimentandeditacquisitionparametersifnecessary.Detailsareprovidedinlaterpages.

Note:Thedefaultacquisitionparametersaregood,butfeelfreetochangethemasdesired.Ifyouwouldliketochangeanyotherparametersnotaccessibleinthislist,youcanselectParameters–EditAll,whichbringsyoutoTopspin.Youmayneedtodothisforsome2Dexperiments.

11pmStartTime

Note:Thisisredonlybecauseaparameter(ns)hasbeenchanged

Page 5: Brief IconNMR and Topspin 3.5 User Guide for Bruker … IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s

SubmityourExperiment1. Whenyouaresatisfiedwithallexperimentsandarereadytocollectyourdata,highlight

theappropriateexperimentorholderdirectory,andhittheSubmiticon.

AUTOMATIONPROCESSES:AfterhittingSubmit,theautosamplerwilltakeover.Theautomationsoftwarewillhandlethefollowingprocesses:

• InsertSample,• TunetheProbe,• LockontoSolvent,• Shim,• CheckReceiverGain,• AcquireData,• ProcessData,• RemoveSample,• BackupDatatoGoogleDrive.

CurrentProcess

IfFlashing:inProgressIfGreen:CompletedIfClear:SkippedorNotCompleted

Page 6: Brief IconNMR and Topspin 3.5 User Guide for Bruker … IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s

DataProcessing1. Tolookatyourspectra,simplydouble-clickonyourcompletedexperimentinthe

PrecedingExperimentslist.Ifdataacquisitionandprocessingiscomplete,youwillseeacheckmarkunderAcqandProcbyyourexperiment.Whenyoudoubleclick,youwillbesenttoTopspinforspectralviewing.

SpectralProcessingCommandsandTips:

1. NavigatetotheProcesstab,andhitProc.Spectrumbutton.ThiswillperformFouriertransform(efp),autophaseyourdata(apk),andperformanautomaticbaselinecorrection(abs).

2. Clickanddragtozoomin3. Double-clickorhittheFullViewbuttontozoomout.4. Usemiddlemousescrollwheeltozoomupanddown5. ClickandholdtheHandicontoshiftthebaselineupordownorsidetoside6. ReturntoIconNMRwhendoneprocessing/viewingspectrum

Identifyyourcompletedexperimentanddouble-click.Thelistisorderedchronologically.

Page 7: Brief IconNMR and Topspin 3.5 User Guide for Bruker … IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER This document is intended to be a brief, bare-bones user’s

LogoutProcedure:1. Whentheautosamplerhasreturnedyoursampletotheholder,youcanremoveyour

sampleandspinner.Caution:doNOTremoveyoursamplewhilethecarouselismoving,anddoNOTreachacrossthecenterwherethesensorislocated.Thisisespeciallyproblematicwhenthereisasamplecurrentlyinthemagnet.Trippingthesensorwithyourhandmayresultinshutdownoftheautosampler.Ifthishappens,besuretocallBennettoranNMRTA.

2. Whenyouaredoneandnolongerusingyourchosenholder,returntoIconNMR,selectyourcompletedholder,andhitDelete.Theholdershouldbecomeavailable.

3. Signthelogbook,includingDate,Name/PI,Nucleus,Drop-OffTime,PickupTime,andcheckifyouhavedeletedyourcompletedruninIconNMR.