Upload
yan-hua
View
217
Download
0
Embed Size (px)
Citation preview
Bone quality assessment based on conebeam computed tomography imaging
Yan HuaOlivia NackaertsJoke DuyckFrederik MaesReinhilde Jacobs
Authors’ affiliations:Yan Hua, Olivia Nackaerts, Reinhilde Jacobs,Oral Imaging Center, School of Dentistry, OralPathology and Maxillofacial Surgery, Faculty ofMedicine, Katholieke Universiteit Leuven, Leuven,BelgiumJoke Duyck, Biomat Laboratory, School ofDentistry, Oral Pathology and MaxillofacialSurgery, Faculty of Medicine, KatholiekeUniversiteit Leuven, Leuven, BelgiumFrederik Maes, ESAT-PSI, Medical Imaging Center,Faculty of Medicine and Faculty of Engineering,Katholieke Universiteit Leuven, Leuven, Belgium
Correspondence to:Reinhilde JacobsOral Imaging Center,School of DentistryOral Pathology and Maxillofacial SurgeryFaculty of MedicineKatholiekeUniversiteit LeuvenKapucijnenvoer7, 3000 LeuvenBelgiumTel/Fax: þ32 16 332410e-mail: [email protected]
Key words: cone beam computed tomography, fractal analysis, fractal dimension,
morphometry
Abstract
Objectives: The aim of this in vitro study was to investigate the accuracy of fractal analysis
and morphometry for bone quality assessment as measured with dual energy X-ray
absorptiometry (DXA).
Material and methods: Nineteen mandibular bone samples were used for the creation of
artificial bone lesions (n¼5) or decalcification (n¼12) to simulate osteoporosis; two
samples were used as controls. Cone beam computed tomography (CBCT) and DXA scans
were made before and after processing the samples. The image data obtained from the
CBCT scans were used to calculate the mean fractal dimension (FD), bone area and density
(morphometric analysis) of the samples. Bone mineral density (BMD) was obtained from the
DXA scans and set as a reference value for bone quality. The correlation between BMD and
FD and between BMD and morphometric results were calculated.
Results: A significant correlation between FD and BMD (r¼ þ0.71 to þ0.75; Po0.05) was
observed. Bone area and BMD of the specimens (r¼ þ0.69 to þ0.85; Po0.05) were also
significantly related, in contrast to the density analysis, for which no significant correlation
to BMD was found.
Conclusions: The results of this study suggest that fractal analysis and bone area
measurement have potential to evaluate bone quality on CBCT images, while density
measurement does not seem to be valid.
The definition of osteoporosis refers to
both low bone mass and micro-architec-
tural deterioration of the bone scaffold. The
presence of one or both factors results in
increased bone fragility and therefore in-
creased susceptibility to fractures (White
2002). Esposito et al. (1998) identified low
bone quality as one of the factors associated
with oral implant failures due to biological
causes. Such low bone quality or a quality
reduction can be caused by osteoporosis.
Therefore, preoperative bone quality as-
sessment is important for oral implant
surgery planning.
Bone mineral density (BMD) assessment
as well as micro-architectural analysis of
the bone, both have a nonnegligible role in
the diagnosis of osteoporosis (Apostol et al.
2006).
BMD, as a gold standard for the diagnosis
of osteoporosis, can be measured in differ-
ent ways. Dual-energy X-ray absorptiome-
try (DXA) can be used for osteoporosis
diagnosis at the spine and/or hip, and
also, although until now it has been used
solely in experimental settings, for jaw
bone evaluation (von Wowern 2001; Nack-
aerts et al. 2007).
Date:Accepted 22 October 2008
To cite this article:Hua Y, Nackaerts O, Duyck J, Maes F, Jacobs R. Bonequality assessment based on cone beam computedtomography imaging.Clin. Oral Impl. Res. 20, 2009; 767–771.doi: 10.1111/j.1600-0501.2008.01677.x
c� 2009 The Authors. Journal compilation c� 2009 John Wiley & Sons A/S 767
For studies on bone micro-architecture,
texture analysis has been applied earlier
(Apostol et al. 2006). Fractal analysis is
one of the approaches for texture analysis,
which received particular attention for ap-
plication in 2D radiographs (Apostol et al.
2006). In fractal analysis, there are different
methods to calculate fractal dimension
(FD), which can be used to compare normal
bone with osteoporotic bone (Geraets &
van der Stelt 2000). For example, Southard
et al. (2000) found that the average FD
decreased from 1.26 to 1.1 in radiographs
of decalcified human alveolar bone. In an
in vivo study (Southard et al. 2001) they
confirmed the clinical relevance of this
finding. FD and density of the alveolar
process bone were highly correlated, mean-
ing the FD could represent an objective
way to quantify bone quality, irrespective
of variations in exposure settings.
In the 3D domain, bone quality assess-
ment methods have also been studied.
Texture analysis has been applied in mi-
cro-CT (Apostol et al. 2006), while Houns-
field units (HU) have been used in spiral
CT as a measure, related to jaw BMD
(Stoppie et al. 2006).
Cone beam computed tomography
(CBCT) is a more recent development
than spiral CT. Its clinical application in
the field of dentomaxillofacial radiology is
gaining importance and spreading widely
(Guerrero et al. 2006; Loubele et al. 2006;
Scarfe et al. 2006). While the clinical re-
levance for presurgical assessment of jaw
bone density has been clearly demonstrated
(for a review see Esposito et al. 1998), a
reliable and easily applicable clinical tool
for this assessment does not yet exist. The
available research on CBCT-based bone
quality assessment is scarce and hampered
by the inherent technical constraints of
CBCT image data sets.
Indeed CBCT data have a larger amount
of scattered X-rays than conventional spiral
CT. This may enhance the noise in recon-
structed images, and thus affect the low-
contrast detectability (Endo et al. 2001).
Because of scatter and artifacts, HU values
in CBCT are not valid, and therefore the
method of correlating BMD to HU values
from CBCT is not ideal. Moreover, the
scatter and artefacts in CBCT get worse
around inhomogenous tissues with reduced
HU values up to 200 HU (Yoo & Yin
2006), which confirms that the HU in
CBCT is not a valid method for bone
quality assessment. Beam hardening is a
phenomenon resulting from the increase of
mean energy of the X-ray beam when it
passes through an object. Because of beam
hardening, the HU of certain structures
such as soft tissue and bone alters. The
single detector CBCT has a larger beam
width than the conventional multi-detec-
tor row CT. This causes a non-uniform
angular distribution of the X-ray beam
intensity known as the heel effect, which
also leads to HU that have no uniformity
either. This confirms that up till now
CBCT-based bone quality assessment is
neither accurate nor reliable, and thus,
there is an urgent need to find methods to
circumvent the shortcomings of this parti-
cular development, so as to have a reliable
way to assess bone quality.
Given the fact that HU values are not
valid in CBCT, there is a need for methods,
other than density measurements, for bone
quality assessment. Texture analysis may
thus come into play, which is strengthened
by the fact that bone quality may be
expressed by its micro-architectural com-
position.
The aim of this in vitro study was to
investigate the accuracy of fractal analysis
and morphometry for bone quality assess-
ment as measured with dual energy X-ray
absorptiometry (DXA). Fractal and mor-
phometric analyzes were based on CBCT
images.
Material and methods
Material
Human mandibular dry bone specimens
were obtained from the Department of
Anatomy (K. U. Leuven), after approval by
the ethical committee of the Catholic Uni-
versity of Leuven. A total of 19 specimens
were randomly allocated to two separate
groups: artificial bone lesion (group 1;
n¼ 5) and decalcification (group 2;
n¼ 12). Two samples were used as controls.
Methods
Preparation of bone specimens to simulatebone loss
Samples in group 1 were used to simulate
destruction of the trabecular bone struc-
ture. This was done by creating small bone
defects (diameter between 1 and 1.5 mm)
using a spoon excavator. Samples in group
2 were used to simulate bone loss by
decalcification. The samples were decalci-
fied with an HCl solution (Decals
, Serva,
Heidelberg, Germany) twice during
10 min. At each interval, the samples
were first rinsed with distilled water for a
few seconds in order to suppress decalcifi-
cation and then dried.
Methods for bone quality assessment
Imaging procedure. All samples were
weighed on a scientific scale, AB304-Ss
(N. V. Mettler-Toledo S. A., Zaventem,
Belgium; precision¼ 0.1 mg), before and
after the procedures for bone removal.
All samples were scanned with CBCT
and DXA before and after the procedures
for bone removal. For the CBCT images,
the samples were put into a polystyrene
container with water for soft tissue simula-
tion and scanned with an i-CATs
(Imaging
Sciences International Inc., Hatfield, PA,
USA) using the predefined protocol ‘Max-
illa 6 cm, 20 s’.
DXA scans were carried out with a
Hologic QDR-4500as
(Hologic Inc., Bed-
ford, MA, USA) to establish a reference
standard on bone density. The DXA scan-
ner was calibrated daily in accordance with
the manufacturer’s recommendations. The
regional high-resolution mode of the small
animal scan protocol [scan field 5
(width) � 7.4 (height) cm2, line spacing
and point resolution 0.0311 cm] was used.
The specimens were positioned on a plexi
support (thickness¼ 2 cm).
The CBCT image data were analyzed
using several bone quality assessment
methods, potentially suitable for CBCT
image analysis: 2D and 3D fractal analysis
and morphometric analysis. The methods
are described below.
2D fractal analysis. FD, which is the result
of fractal analysis, is a quantification of
surface roughness: the rougher the surface,
the larger the computed magnitude of FD
(Southard et al. 2001). FD was based on the
box counting method, which is only one of
several existing methods for fractal analysis.
MevisLabs
(MeVis Research GmbH,
Bremen, Germany) was used for this ana-
lysis. CBCT slices were consecutively pro-
cessed using this programming environ-
ment. The bone region on the slices was
segmented (‘live wire’ and ‘region grow-
Hua et al . Bone quality on CBCT images
768 | Clin. Oral Impl. Res. 20, 2009 / 767–771 c� 2009 The Authors. Journal compilation c� 2009 John Wiley & Sons A/S
ing’) and thresholding (window/leveling)
was applied to obtain a binary image (Fig.
1). Either the whole bone (cortical and tra-
becular bone) or only the trabecular bone
were set as the region of interest (ROI) and
fractal analysis was applied in Matlabs
(MathWorks Inc, Natick, MA, USA), using
the open source project FracLab. The mean
value of FD for each sample was calculated.
3D fractal analysis. Within the program-
ming environment of MevisLabs
, a 3D
cube counting method was developed for
fractal analysis as an extension to the 2D
box counting method. For this purpose, all
slices were processed in the volume mode.
Either the whole bone (cortical and trabe-
cular bone) or only the trabecular bone was
set as the volume of interest and fractal
analysis was applied. FD was determined
by the relationship between the number of
voxels covering bone and the voxel size
(cube size). The slope of the logarithm-
fitted curve of inverse voxel size vs. the
number of cubes needed to cover the bony
structure represented the FD (Zhang et al.
2006). An illustration of such curve is
shown in Fig. 2.
Morphometry. Morphometry was done
with the commercially available software
AxioVisions
(Carl Zeiss MicroImaging
GmbH, Koln, Germany). Each slice was
saved as a jpg image, on which morpho-
metry was performed. In this study, the
mean density value based on gray values
and the area of bony structure in each slice
were measured.
Data analysis
As data were not normally distributed,
Spearman’s r was calculated between FD
and the corresponding BMD and between
morphometric results and the correspond-
ing BMD. To assess the change in FD and
morphometric results, only descriptive sta-
tistics were used because of the small
sample size.
Results
Weight
The mean weight of the samples is shown
in Table 1. The weight of all samples
except for the control samples, decreased
after bone removal procedures.
2D fractal analysis
The mean FD of all CBCT slices was
calculated for each sample. Table 2 shows
the mean FD of the processed samples for
both methods. Although we were not able
to perform statistical analysis due to the
size of the experimental groups in this
exploratory study, it became evident that
the FD decreased after the modifications.
This was true for the analysis including the
cortical bone and the one including only
trabecular bone. Analysis of the control
samples showed a stable FD.
3D fractal analysis
The number of voxels containing bone for
the same cube size decreased after the
modifications to the bone samples (struc-
ture and Ca content). The FD decreased
when analyzing the entire bone sample.
However, FD increased when analyzing
only the trabecular bone.
Morphometry
The mean density, based on gray values,
and bone area of each sample are listed in
Table 3.
Fig. 1. (a) Cone beam computed tomography
(CBCT) slice in binary format, including cortical
bone. (b) CBCT slice in binary format, excluding
cortical bone.
1/Voxel size
N v
oxe
ls c
ove
rin
g b
on
e vo
lum
e o
f in
tere
st
Fig. 2. Illustration of the fractal analysis of one sample in 3D. The slope, a measure for fractal dimension (FD),
decreases after decalcification.
Table 1. Changes in bone specimen weight based on various procedures
Weight beforemodification (g)
Weight aftermodification (g)
Change % change
Artificial bone lesion 2.88 2.72 � 0.16 � 5.69Decalcification 2.97 2.31 � 0.66 � 22.11Control 1.89 1.88 � 0.01 � 0.27
Table 2. Results of 2D fractal analysis: (a) method 1 (both cortical and trabecular bone),(b) method 2 (trabecular bone only)
Mean FD (SD) beforemodification
Mean FD (SD) aftermodification
FD change % change
(a)Control samples 1.00 (0.07) 0.99 (0.08) � 0.01 � 1.0Modified samples 1.05 (0.04) 0.98 (0.04) � 0.07 � 6.7
(b)Control samples 0.71 (0.05) 0.71 (0.05) 0.00 0.0Modified samples 0.88 (0.14) 0.82 (0.13) � 0.06 � 6.8
Hua et al . Bone quality on CBCT images
c� 2009 The Authors. Journal compilation c� 2009 John Wiley & Sons A/S 769 | Clin. Oral Impl. Res. 20, 2009 / 767–771
The area decreased after the modifica-
tions to the samples. The density measured
in gray values increased.
BMD
Each sample was scanned twice with the
DXA device. The correlation between
BMD and FD, and between BMD and
area were calculated for both scans. A
significant correlation of þ 0.71 to þ 0.75
(Po0.05) was found between FD and
BMD. Likewise, area and BMD correlated
significantly (r¼ þ 0.69 to þ 0.85;
Po0.05). In contrast, density measured in
gray values did not significantly correlate
with BMD (r¼ þ0.13 to þ 0.21; P40.5).
Discussion
Considering the importance of bone quality
in jaw bone implant surgery or related
therapy (Bryant 1998), the need for an
accurate and reliable clinical tool for quan-
tifying this preoperatively is evident. Un-
fortunately such tool is not yet available.
At the same time, in many cases, low dose
CBCT can be advised for implant place-
ment, considering the possibility to gather
clinically relevant 3D data at a low dose
(Guerrero et al. 2006). However, CBCT
does not allow reliable and accurate bone
quality assessment when focusing on the
inherent radiographic density information
that is otherwise expressed by HU (Yama-
shina et al. 2008). This has led to the
present study from which the outcome
illustrates that there may be a potential
for structural analysis to become the
method for standard clinical jaw bone qual-
ity assessment. FD and bone area represent
objective measurements that avoid the dif-
ficulties associated with bone density eva-
luation in CBCT.
For the present study, we opted for an in
vitro approach to allow a reliable validation
and accuracy assessment of the new tech-
niques. Trabecular bone destruction and
decalcification of the bone specimens,
representing micro-architectural bone
deterioration and decreased bone mass,
were an attempt to characterize osteoporo-
tic bone.
The main finding of this study was that
2D FD of CBCT images was significantly
related to the actual jaw BMD. This was in
accordance with a study of Southard et al.
(2001), who found that the average FD
decreased from 1.26 to 1.1 in radiographs
of decalcified human alveolar bone.
Besides, the present study showed that
3D fractal analysis of the CBCT image data
facilitated quantification of the induced
decrease in bone mass by either artificial
bone lesion creation or decalcification.
When performing 3D fractal analysis based
on the trabecular bone alone, the calculated
FD increased after osteoporosis simulation.
Because this was a preliminary study on
development of an innovative method, this
study could not be related nor compared
with previous studies regarding this parti-
cular method in CBCT image data assess-
ment. In the 2D domain, there has been
previous evidence that the FD increased
after osteoporosis simulation. For example,
Berry et al. (1996) decalcified human ver-
tebral bone. Digitized radiographs were
made every 30 s and the FD increased
from 2 to 3. These findings suggest that
FD does not consistently decrease after
osteoporosis simulation. Therefore, it
could be more meaningful to observe both
bone mass and FD, or more in general
structural properties, rather than only con-
sidering FD (Jiang et al. 1999). We will
continue in vitro studies with larger sample
size and validate this approach in clinical
prospective studies.
The morphometric measurements
showed a decreased bone area, correspond-
ing well with the bone mineral reduction
for all samples after the modifications.
BMD does obviously not only depend on
the calcium content of bone alone, but also
on the structural characteristics. The ab-
sence of a clear correlation between density
based on the gray values and BMD based on
DXA measurements illustrated that den-
sity measurements based on CBCT images
were not useful as such, because of inten-
sity inhomogeneity.
As far as the researchers know, the pre-
sent study was the first one proposing an
objective bone quality assessment study
based on CBCT technology for dentomax-
illofacial applications. It might also be
stressed that this was the primary study
reporting the use of radiographic CBCT
slices in jaw bone morphometry.
Conclusions
From the present preliminary results, frac-
tal analysis and bone area measurement
seem to have some potential for bone
quality assessment on CBCT images,
while density measures do not seem valid.
More elaborate studies are necessary to
verify these results and test their clinical
applicability.
References
Apostol, L., Boudousq, V., Basset, O., Odet, C., Yot,
S., Tabary, J., Dienten, J.M., Boller, E., Kotzki,
P.O. & Peyrin, F. (2006) Relevance of 2D radio-
graphic texture analysis for the assessment of 3D
bone micro-architecture. Medical Physics 33:
3546–3556.
Berry, J.L., Towers, J.D., Webber, R.L., Pope, T.L.,
Davidai, G. & Zimmerman, M. (1996) Change
in trabecular architecture as measured by fractal
Table 3. Results for morphometry: (a) in goup 1 (artificial bone lesion), (b) in group 2(decalcification), and (c) in control group
Sample Before lesion After lesion Change % change
(a)Density (gray value) 240.47 261.25 20.78 8.64Area (pixel) 65,197 60,815 � 4382 � 6.70
Sample Beforedecalcification
Afterdecalcification
Change % change
(b)Density (gray value) 226.26 238.02 11.76 5.20Area (pixel) 40,814 32,846 � 8328 � 20.40
Sample Before rinsing After rinsing Change % change
(c)Density (gray value) 212.28 218.37 6.09 2.87Area (pixel) 34,216 33,466 750 2.20
Hua et al . Bone quality on CBCT images
770 | Clin. Oral Impl. Res. 20, 2009 / 767–771 c� 2009 The Authors. Journal compilation c� 2009 John Wiley & Sons A/S
dimension. Journal of Biomechanics 29: 819–
822.
Bryant, S.R. (1998) The effects of age, jaw site, and
bone condition on oral implant outcomes. Inter-
national Journal of Prosthodontics 11: 470–490.
Endo, M., Tsunoo, T., Nakamori, N. & Yoshida, K.
(2001) Effect of scattered radiation on image noise
in cone beam CT. Medical Physics 28: 469–474.
Esposito, M., Hirsch, J.M., Lekholm, U. & Thom-
sen, P. (1998) Biological factors contributing to
failures of osseointegrated oral implants. Eur-
opean Journal of Oral Sciences 106: 721–764.
Geraets, W.G.M. & van der Stelt, P.F. (2000) Fractal
properties of bone. Dentomaxillofacial Radiology
29: 144–153.
Guerrero, M.E., Jacobs, R., Loubele, M., Schutyser,
F., Suetens, P. & van Steenberghe, D. (2006)
State-of-the-art on cone beam CT imaging for
preoperative planning of implant placement. Clin-
ical Oral Investigations 10: 1–7.
Jiang, C., Giger, M.L., Chinander, M.R., Martell,
J.M., Kwak, S. & Favus, M.J. (1999) Character-
ization of bone quality using computer-extracted
radiographic features. Medical Physics 26: 872–
879.
Loubele, M., Maes, F., Schutyser, F., Marchal, G.,
Jacobs, R. & Suetens, P. (2006) Assessment
of bone segmentation quality of cone-beam CT
versus multislice spiral CT: a pilot study. Oral
Surgery, Oral Medicine, Oral Pathology, Oral
Radiology and Endodontics 102: 225–234.
Nackaerts, O., Jacobs, R., Horner, K., Zhao, F.,
Lindh, C., Karayianni, K., van der Stelt, P., Pavitt,
S. & Devlin, H. (2007) Bone density measure-
ments in intra-oral radiographs. Clinical Oral
Investigations 11: 225–229.
Scarfe, W.C., Farman, A.G. & Sukovic, P. (2006)
Clinical applications of cone-beam computed to-
mography in dental practice. Journal of Canadian
Dental Association 72: 75–80.
Shapurian, T., Damoulis, P.D., Reiser, G.M., Grif-
fin, T.J. & Rand, W.M. (2006) Quantitative
evaluation of bone density using the Hounsfield
index. The International Journal of Oral & Max-
illofacial Implants 21: 290–297.
Southard, T.E., Southard, K.A., Jakobsen, J.R., Hil-
lis, S.L. & Najim C, A. (2000) Fractal dimension
in radiographic analysis of alveolar process
bone. Oral Surgery, Oral Medicine, Oral Pathol-
ogy, Oral Radiology and Endodontics 89: 569–
576.
Southard, T.E., Southard, K.A. & Lee, A. (2001)
Alveolar process fractal dimension and postcranial
bone density. Oral Surgery, Oral Medicine, Oral
Pathology, Oral Radiology and Endodontics 91:
486–491.
Stoppie, N., Pattijn, V., Van Cleynenbreugel, T.,
Wevers, M., Vander Sloten, J. & Naert, I. (2006)
Structural and radiological parameters for the
characterization of jaw bone. Clinical Oral
Implants Research 17: 124–133.
von Wowern, N. (2001) General and oral aspects of
osteoporosis: a review. Clinical Oral Investiga-
tions 5: 71–82.
White, S.C. (2002) Oral radiographic predictors of
osteoporosis. Dentomaxillofacial Radiology 31:
84–92.
Yamashina, A., Tanimoto, K., Sutthiprapaporn, P. &
Hayakawa, Y. (2008) The reliability of computed
tomography (CT) values and dimensional
measurements of the oropharyngeal region using
cone beam CT: comparison with multidetector
CT. Dentomaxillofacial Radiology 37: 245–251.
Yoo, S. & Yin, F.F. (2006) Dosimetric feasibility of
cone-beam CT-based treatment planning com-
pared to CT-based treatment planning. Inter-
national Journal of Radiation Oncology, Biology
and Physics 66: 1553–1561.
Zhang, L., Liu, J.Z., Dean, D., Sahgal, V. & Yue,
G.H. (2006) A three-dimensional fractal analysis
method for qualifying white matter structure in
human brain. Journal of Neuroscience Methods
150: 242–253.
Hua et al . Bone quality on CBCT images
c� 2009 The Authors. Journal compilation c� 2009 John Wiley & Sons A/S 771 | Clin. Oral Impl. Res. 20, 2009 / 767–771