7
Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas Hinrichs 1 , Sven Gragert 1 , and Michael Klein 1 Abstract It is a well-known fact that personnel activities can have an adverse effect on the protective functions of biological safety cabinets. Increased air turbulence within the work area or within the surrounding room can disrupt the essential directional airflows of the cabinet, which generate the protective barrier between the environment and the products handled inside. This may result in an enhanced escape and/or invasion of airborne particles or microorganisms. Nevertheless, manufacturers tend to reduce downflow and inflow velocities up to the limits of the allowed values to improve energy efficiency or to reduce noise and vibration. Reliable data based on standardized test procedures to estimate the consequences arising from these measures are rare. In this study, the influence of different static and dynamic disturbing factors on the personnel protection performance of 2 class II biological safety cabinets was quantified. A microbiological test procedure given by the relevant European and US standards (EN 12649, NSF/ANSI 49) was used, but in contrast to the low requirements defined there, 4 more complex and realistic test scenarios were chosen to simulate working activities: static covering of the front sash opening caused by a sitting or standing person was simulated by a ‘‘dummy worker’’ and a ‘‘body plate,’’ respectively. Dynamic airflow perturbations were generated by an artificial ‘‘moving arm’’ swinging regularly inside the cabinet and by a flat plate running outside the cabinet parallel to its front opening to simulate a ‘‘walking man’’. It could be demonstrated that dynamic airflow disturbances caused by rapid body movements have a major impact on the cabinet’s protecting performance. Compared with an undisturbed working situation, personnel-protecting capabilities of both safety cabinets tested declined substan- tially when the worker’s movement next to the front opening was simulated. Therefore, downflow and inflow velocities should not be reduced to minimum values, to allow a sufficient margin of safety for actual in-use laboratory conditions (‘‘disturbed’’). Keywords BSC, airflow perturbation, personnel activities, protection, laminar airflow Class II biological safety cabinets (BSCs) are used as small contained areas for the safe and aseptic handling of sensitive and/or infectious substances, tissues, or organisms. 1-3 The functioning principle of this type of BSC is based on 2 direc- tional airflows. The downflow moves downward to the interior working surface before it is drawn through the front or rear grille. Room air is directed inward across the working opening and passes into the front grille before it reaches the working zone (inflow). Downflow and inflow are mixed outside the working zone and then filtered by a combination of supply and exhaust HEPA (high-efficiency particulate air) filters to remove airborne contaminants. For class II, type A and B1 BSCs, clean (particulate-free) air is recirculated back to the working area (approximately 70%) and, to a smaller extent (approximately 30%), to the environment (laboratory room or building exhaust system). Both airflows are driven by 1 inter- nal fans. Downflow and inflow characteristics ensure that no particles or microorganisms are released from the containment (personnel protection) or that products handled inside the BSC are contaminated by agents from outside (product protection) or by impurities released from other products (cross- contamination protection). To ensure a correct and safe functioning of the cabinet, a proper balance between downflow and inflow velocities is nec- essary. Required values are given by the European standard 4 (EN 12469), with a minimum average inflow velocity of 78.7 ft/min (0.4 m/s) and a downflow velocity of 49.2 to 98.4 ft/min (0.25-0.5 m/s), or by the American standard 5 (NSF/ANSI 49), with a minimum average inflow velocity of 75 ft/min (0.38 m/s; type A1) or 100 ft/min (0.51 m/s; types A2, B1, and B2). Within this framework, suitable settings are defined by the manufacturer having regard to an adequate level of protection as well as to other criteria, such as noise level, 1 Berner International GmbH, Elmshorn, Schleswig-Holstein, Germany Corresponding Personnel: Imogen Britton, Berner International GmbH, Mu ¨hlenkamp 6, 25337 Elmshorn, Schleswig-Holstein, Germany. Email: [email protected] Applied Biosafety: Journal of ABSA International 1-7 ª The Author(s) 2016 Reprints and permission: sagepub.com/journalsPermissions.nav DOI: 10.1177/1535676016635369 apb.sagepub.com

Biological Safety Cabinets: Simulation · 2016-04-20 · Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Biological Safety Cabinets: Simulation · 2016-04-20 · Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas

Article

Biological Safety Cabinets: Simulationand Quantifying of Airflow PerturbationCaused by Personnel Activities

Thomas Hinrichs1, Sven Gragert1, and Michael Klein1

AbstractIt is a well-known fact that personnel activities can have an adverse effect on the protective functions of biological safety cabinets.Increased air turbulence within the work area or within the surrounding room can disrupt the essential directional airflows of thecabinet, which generate the protective barrier between the environment and the products handled inside. This may result in anenhanced escape and/or invasion of airborne particles or microorganisms. Nevertheless, manufacturers tend to reduce downflow andinflow velocities up to the limits of the allowed values to improve energy efficiency or to reduce noise and vibration. Reliable data basedon standardized test procedures to estimate the consequences arising from these measures are rare. In this study, the influence ofdifferent static and dynamic disturbing factors on the personnel protection performance of 2 class II biological safety cabinets wasquantified. A microbiological test procedure given by the relevant European and US standards (EN 12649, NSF/ANSI 49) was used, butin contrast to the low requirements defined there, 4 more complex and realistic test scenarios were chosen to simulate workingactivities: static covering of the front sash opening caused by a sitting or standing person was simulated by a ‘‘dummy worker’’ and a‘‘body plate,’’ respectively. Dynamic airflow perturbations were generated by an artificial ‘‘moving arm’’ swinging regularly inside thecabinet and by a flat plate running outside the cabinet parallel to its front opening to simulate a ‘‘walking man’’. It could be demonstratedthat dynamic airflow disturbances caused by rapid body movements have a major impact on the cabinet’s protecting performance.Compared with an undisturbed working situation, personnel-protecting capabilities of both safety cabinets tested declined substan-tially when the worker’s movement next to the front opening was simulated. Therefore, downflow and inflow velocities should not bereduced to minimum values, to allow a sufficient margin of safety for actual in-use laboratory conditions (‘‘disturbed’’).

KeywordsBSC, airflow perturbation, personnel activities, protection, laminar airflow

Class II biological safety cabinets (BSCs) are used as small

contained areas for the safe and aseptic handling of sensitive

and/or infectious substances, tissues, or organisms.1-3 The

functioning principle of this type of BSC is based on 2 direc-

tional airflows. The downflow moves downward to the interior

working surface before it is drawn through the front or rear

grille. Room air is directed inward across the working opening

and passes into the front grille before it reaches the working

zone (inflow). Downflow and inflow are mixed outside the

working zone and then filtered by a combination of supply and

exhaust HEPA (high-efficiency particulate air) filters to

remove airborne contaminants. For class II, type A and B1

BSCs, clean (particulate-free) air is recirculated back to the

working area (approximately 70%) and, to a smaller extent

(approximately 30%), to the environment (laboratory room or

building exhaust system). Both airflows are driven by�1 inter-

nal fans. Downflow and inflow characteristics ensure that no

particles or microorganisms are released from the containment

(personnel protection) or that products handled inside the BSC

are contaminated by agents from outside (product protection)

or by impurities released from other products (cross-

contamination protection).

To ensure a correct and safe functioning of the cabinet, a

proper balance between downflow and inflow velocities is nec-

essary. Required values are given by the European standard4

(EN 12469), with a minimum average inflow velocity of

78.7 ft/min (0.4 m/s) and a downflow velocity of 49.2 to

98.4 ft/min (0.25-0.5 m/s), or by the American standard5

(NSF/ANSI 49), with a minimum average inflow velocity of

75 ft/min (0.38 m/s; type A1) or 100 ft/min (0.51 m/s; types A2,

B1, and B2). Within this framework, suitable settings are

defined by the manufacturer having regard to an adequate level

of protection as well as to other criteria, such as noise level,

1 Berner International GmbH, Elmshorn, Schleswig-Holstein, Germany

Corresponding Personnel:

Imogen Britton, Berner International GmbH, Muhlenkamp 6, 25337 Elmshorn,

Schleswig-Holstein, Germany.

Email: [email protected]

Applied Biosafety:Journal of ABSA International1-7ª The Author(s) 2016Reprints and permission:sagepub.com/journalsPermissions.navDOI: 10.1177/1535676016635369apb.sagepub.com

Page 2: Biological Safety Cabinets: Simulation · 2016-04-20 · Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas

vibration, and filter life. The operational airflow adjustment

recommended by the manufacturer is called the ‘‘nominal set

point.’’ Within the last few years, the nominal set point velo-

cities have been lowered successively, primarily to meet

energy-saving requirements. These efforts make sense both

economically and ecologically, but lowering of fundamental

safety parameters can result in a system with a reduced ‘‘safety

margin.’’ Especially dynamic flow disturbances caused by

rapid or inappropriate movements inside or outside the cabinet

may affect the safeguarding air curtain spanned across the front

opening by the 2 airflows. Although these negative impacts

have been well known for several decades,6-8 they are not yet

integrated into the prescribed test procedures. Until today, air-

flow perturbation caused by rapid body or arm movement is

widely ignored. Relevant guidelines for the proper use of BSCs

(eg, Occupational Safety and Health Administration9) mention

the adverse effects of personnel activities, but only limited

quantitative information is available.

In this study, typical working situations potentially compro-

mising the BSC air barrier were simulated to evaluate their

influence on personnel protection. By comparing the perfor-

mance limits (ie, the lowest airflow balance settings with which

the cabinet passes microbiological testing), the interfering

influence of different kinds of personnel activity should be

quantified. Normally, BSCs were tested in a standardized,

undisturbed scenario. In contrast, data from this study should

help to estimate the personnel-protecting capabilities of safety

cabinets when used under realistic working conditions.

Material and Methods

Approach

The influence of static and dynamic airflow disturbances on the

personnel-protecting performance of 2 biosafety cabinets (BSC

1 and BSC 2) was evaluated. Performance was quantified by

microbiological tests specified in the European and American

standards, EN 12469 and NSF/ANSI 49, respectively. Starting

from the nominal set points, the cabinet airflow velocities

(downflow, inflow) were reduced stepwise with a maximum

step size of 9.8 ft/min (0.05 m/s) until the cabinets did not

fulfill the requirements for personnel protection. The last pair

of velocity values that met the standard criteria was defined as

the limit of the safe range. To mimic blocked or disrupted BSC

airflows caused by laboratory staff, different types of perturbing

devices (dummy worker, body plate, moving arm, walking man)

were mounted in front of and/or inside the cabinets. The effect

upon protecting capabilities could be quantified by comparing the

performance limits determined under these conditions with those

evaluated under undisturbed conditions.

Safety Cabinets

The tests were carried out with 2 class II, type A1 biosafety

cabinets from different manufacturers. Both units are compa-

tible with EN 12469 and can be used for the handling of

(potentially) hazardous microorganisms. Access opening mea-

sures were 7.9 � 46.9 in (20 � 119 cm, BSC 1) and 7.1 �49.2 in (18 � 125 cm, BSC 2). Common special features

include a 3-HEPA filter system and a V-shaped grille in front

of the working area. Nominal set points for inflow and down-

flow velocities can be taken from Table 1. Lower values could

be realized by downregulating the fan speed and/or readjusting

the exhaust damper. Airflow velocities were determined

according to EN 12469. For downflow measurements inside

the cabinet, a thermal anemometer (model 454; Testo AG,

Lenzkirch, Germany) was used. The sensor was placed in

8 positions in a horizontal plane 2 in (5 cm) above the bottom

edge of the front window. Inflow velocity was measured volu-

metrically with an air-capture hood kit (model 8710 microman-

ometer attached to a model 8375 flow hood; TSI Inc,

Shoreview, Minnesota). Both instruments were calibrated

before use. The cabinets were tested individually in a 194-ft2

(18-m2) laboratory room. Before testing, the room ventilation

system was switched off.

Microbiological Testing

To evaluate the ability of the safety cabinets to keep airborne

particles from migrating out of the working area, protection

experiments were conducted according to the relevant stan-

dards. The test setups given in EN 12469 and NSF/ANSI 49

are nearly the same. Spores of Bacillus subtilis var niger

(ATCC 9372, 5 � 108 colony-forming units [CFUs] / mL;

obtained from Presque Isle Cultures, Erie, Pennsylvania) were

aerolized inside or outside the cabinet with a CN31I NSF

Collison Nebulizer (BGI, Inc, Waltham, Massachusetts).

Spores that penetrated the air curtain were collected with dif-

ferent kinds of air samplers (impinger samplers, slit samplers

with trypticase soy agar [TSA] culture plates, TSA settling

plates). After sampling, media were kept in an incubator up

to 48 hours. Colonies grown during this period were counted.

Exact information about the settings and positioning of all

devices used for microbiological testing can be taken from the

standard instructions. Test conditions exactly corresponding to

the requirements of the EN 12469 standard were designated as

‘‘undisturbed.’’

Personnel Protection Testing

The nebulizer containing a spore suspension of 5.0 � 108

CFUs/mL was mounted inside the safety cabinet with the

Table 1. Airflow Velocities at the Nominal Set Points of Both BSCsTested.

Airflow Velocity, ft/min

Downflow Inflow

BSC 1 68.9 86.6BSC 2 84.6 84.6

Abbreviation: BSC, biological safety cabinet.

2 Applied Biosafety: Journal of ABSA International

Page 3: Biological Safety Cabinets: Simulation · 2016-04-20 · Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas

opening directed toward the front window (Figure 1). To

simulate the influence of an (unmoving) arm, a circular

stainless-steel cylinder has to be placed at the center of the

working area conforming to the standards. The duration of a

complete test cycle was 30 minutes, including a 6.5-minute

aerosolization period. Samples were collected outside the

containment with 6 type AGI 30 impinger samplers (Ace

Glass Inc, Vineland, New Jersey) positioned around the

cylinder, as well as with 2 slit-type air samplers (Impaktor

FH5; Markus Klotz GmbH, Bad Liebenzell, Germany) posi-

tioned to the right and left of the cabinet front (Figure 1). A

settling plate was placed over the front grille beneath the

cylinder to serve as control plate. Sampling fluid from all

impinger samplers was filtered by means of a sterile mem-

brane filter unit (0.2 mm, MicroFunnel Filter Unit; Pall Cor-

poration, Ann Arbor, Michigan) and placed on a small TSA

plate under aseptic conditions. Slit sampler plates, the filter

membrane–containing plate and the control plate were incu-

bated at 37�C and checked for microbial growth after 24

and 48 hours. By standard definition, the cabinet passes the

personnel protection test if CFUs recovered from the impin-

ger samplers do not exceed 10 CFUs and total slit sampler

plate counts do not exceed 5 CFUs. Each test was per-

formed at least 3 times.

Airflow Perturbation

Performance tests were carried out according to the usual prac-

tice (undisturbed) and under 4 simulated working conditions

(disturbed). In contrast to the test conditions given in EN 12469

and NSF/ANSI 49, static and dynamic disturbances were

caused with a dummy worker by introducing its arms through

the front aperture, a static body-shaped plate positioned

directly in front of the opening (body plate), a robot arm swing-

ing inside the working area (moving arm), and a rail-based

plate (walking man) moving parallel to the cabinet opening.

Dummy Worker

To simulate the static airflow perturbation caused by a person

working in an upright sitting position, a dummy was placed on

a chair in front of the BSC (Figure 2). The distances from head

and chest to the front window were adapted to real working

conditions determined before the testing. The penetration depth

of the arms into the opening was defined as 9.8 in (25 cm).

Body Plate

Airflow disturbances caused by a person standing directly in

front of the BSC opening were simulated by a plate formed

Figure 1. Personnel protection test setup according to European andAmerican standards EN 12469 and NSF/ANSI 49, respectively.

Figure 2. Personnel protection test setup with the dummy worker.

Hinrichs et al 3

Page 4: Biological Safety Cabinets: Simulation · 2016-04-20 · Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas

similar to the shape of a human body, with the following

dimensions: height, 31.1 in (79 cm); upper edge, 17.7 in

(45 cm); and lower edge, 13.0 in (33 cm). The plate was posi-

tioned parallel to the front window touching the outside edge of

the cabinet (Figure 3). As defined in the standards, a cylinder

was placed inside the working area to disrupt the airflow. Due

to the small distance between the plate and the front aperture,

the impinger samplers and the cylinder had to be brought into

standard position via square cutouts in the plate.

Moving Arm

To investigate the effect of a dynamic airflow perturbation within

the front opening and inside the working area, a robot arm was

used as an external disruptive factor (Figure 4). The arm could be

moved electrically in a 2-dimensional way on a horizontal plane

(maximum velocity: 35�/s). Reproducible movement patterns

were triggered by an external control unit. The arm was placed

centrally in front of the BSC extending 21.6 in (55 cm) into the

working area. The use of personal protective equipment was

simulated by providing the arm with a glove and a sleeve cover.

During the test sequence, it swung from the central starting posi-

tion 55� to the left or right side. At the turning points, the arm

moved 4.3 in (11 cm) out of the front opening, to imitate the

shoulder motion of a working person. Airflow perturbation was

considered separately for each side of the cabinet.

Walking Man

A dynamic airflow perturbation induced by a moving person

was simulated via a plate that ran parallel to the front aper-

ture (Figure 5). The geometry and test instructions were

derived from the European standard for fume cupboards.10

The plate, with a width of 15.7 in (40 cm) and a height of

74.8 in (190 cm), was mounted on a moveable slide, which

ran on a linear guide. The distance between the track system

and the front aperture was 15.7 in (40 cm). Starting cen-

trally in front of the opening, the plate ran 31.5 in (80 cm)

farther than the outer edges of the BSC. At the end point of

its motion path, the plate stopped for 30 s, before it returned

to the starting position. The motion speed was fixed to

196.8 + 19.7 ft/min (1.0 + 0.1 m/s).

Results

As expected, lowering of downflow and inflow velocities

weakens the personnel protection performance provided by

both biosafety cabinets tested. Figure 6 shows the performance

Figure 3. Personnel protection test setup with the body plate.

4 Applied Biosafety: Journal of ABSA International

Page 5: Biological Safety Cabinets: Simulation · 2016-04-20 · Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas

limits for the BSCs when downflow and inflow velocities were

downregulated stepwise under undisturbed conditions. Starting

from the nominal set point for downflow and inflow velocities,

personnel protection could be maintained over a wide range of

reduced settings (Table 1). However, air velocities <39.4 ft/min

(0.2 m/s) resulted in a rising number of CFUs escaping from the

working area. The safety limit below which protection cannot

be said to be ‘‘effective’’ according to EN 12469 or NSF/ANSI

49 was reached when downflow velocities fell <21.7 ft/min

(0.11 m/s, BSC 1) and <35.5 ft/min (0.18 m/s, BSC 2) and

inflow velocities were <37.4 ft/min (0.19 m/s, BSC 1) and

<25.6 ft/min (0.13 m/s, BSC 2). This represents a maximum

acceptable negative deviation from the set point (reduction

potential) of 69% or 58% for downflow velocities and 57%or 70% for inflow velocities, respectively. The average value

was 64%.

The performance limits of both cabinets decreased when

they were operated under disturbed airflow conditions

(Table 2). Static and dynamic factors influenced the protection

capacities to different extents. In general, dynamic perturbation

factors had a greater impact than static ones. Differences

between BSC 1 and BSC 2, as well as between the perturbation

of downflow and inflow velocities, could be detected but were

mostly small and showed a similar trend. Without any airflow

perturbation (undisturbed working situation), downflow and

inflow velocities could be reduced on average by 63.5% before

microbiological release from the containment exceeded the

limits mentioned in EN 12469 and NSF/ANSI 49. Disturbances

caused by the dummy worker had no or negligible effects on

the personnel protection capacities of both BSCs, as seen from

a nearly unchanged average airflow reduction potential (63%).

However, with the second static disturbing construction (body

plate) mounted directly to the cabinet opening, the tolerable

performance limit reduction was clearly lower (on average

53%). A further decrease of the protective potential could be

observed when the cabinets were exposed to dynamic disturb-

ing factors. Interestingly, results measured with the moving

arm differ in regard to the BSC side tested: The maximum

tolerable airflow reduction on the right side was relatively high

(on average 51%). The airflow velocities on the left side, in

contrast, could be reduced by only 37% to fulfill the require-

ments for personnel protection. The reason remains unclear,

but the differences may be due to the fact that the artificial arm

was ‘‘protected’’ with a glove and a sleeve cover to mimic

aseptic working conditions (Figure 4). Covering could have

an impact on the shape of the artificial arm so that airflow

perturbation showed an asymmetrical pattern. The highest dis-

turbing effect was triggered by the walking man. With this kind

Figure 4. Personnel protection test setup with the moving arm.

Hinrichs et al 5

Page 6: Biological Safety Cabinets: Simulation · 2016-04-20 · Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas

of working simulation, downflow and inflow velocities could

be reduced by only 24% before CFUs sampled outside the

cabinets exceeded the standard threshold values. When the

distance between the walking man and the front aperture was

increased, airflow perturbation declined. From a distance of

31.5 in (80 cm), no influence on the protective function of the

BSCs could be detected (data not shown).

Discussion

The personnel protection performance of 2 BSCs was tested

under undisturbed as well as reproducible static and dynamic

working conditions. The results show that, in particular,

dynamic airflow perturbations caused by the movement of the

arms or the whole body present a major challenge to these

devices. While downflow and inflow velocities could be

reduced up to 63% under undisturbed conditions to meet the

minimum standard requirements, this safety margin shrank to

only 24% when rapid walking of a person was simulated. BSCs

are therefore more susceptible to the effect of operator move-

ment and turbulence than what could be expected from the

results of certification testings carried out in accordance with

current standards. Although these deficiencies have been

known for many years, they become acutely relevant because

manufacturers tend to reduce nominal set points for inflow and

downflow velocity close to the normative threshold limits. This

is done in response to customer wishes for higher energy

Figure 5. Personnel protection test setup with the walking man.

0

20

40

60

80

100

0 20 40 60 80 100

Inflo

w v

eloc

ity [f

t/min

]

Downflow velocity [ft/min]

nominalsetpoints

Figure 6. Performance limits (personnel protection test) of bothbiological safety cabinets (BSCs) tested in this study under undisturbedconditions. Filled symbols represent airflow combinations meeting thestandard test requirements. Open symbols indicate airflow settingsthat did not fulfill the criteria (triangles: BSC 1, circles: BSC 2).

6 Applied Biosafety: Journal of ABSA International

Page 7: Biological Safety Cabinets: Simulation · 2016-04-20 · Article Biological Safety Cabinets: Simulation and Quantifying of Airflow Perturbation Caused by Personnel Activities Thomas

efficiency and more comfort. However, any attempt to reduce

energy consumption at the expense of safety may prove coun-

terproductive. The nominal set point for inflow and downflow

velocity must be selected in such a way that even significant

airflow perturbations caused by a moving worker can be com-

pensated. A safety cabinet can normally not be used without

external disturbances. To ensure that the protection capabilities

are high enough to guarantee safe working under ‘‘everyday’’

conditions, standard requirements for the testing of BSCs

should be extended. Artificial models, such as those used in

this study, represent a first attempt to create a realistic but nev-

ertheless reproducible test scenario. Especially dynamic devices

such as the moving arm and the walking man could be helpful to

mimic the most relevant airflow-perturbing factors within a typ-

ical work process. Even though not all problematic situations can

be taken into consideration with this test setup, it would be a

significant progress toward increased safety for laboratory staff

as well as for products handled inside a BSC.

Declaration of Conflicting Interests

The author(s) declared no potential conflicts of interest with respect to

the research, authorship, and/or publication of this article.

Funding

The author(s) disclosed receipt of the following financial support for the

research, authorship, and/or publication of this article: This work was

funded by the German Federal Ministry of Economics and Technology

under the program PRO INNO II (PROgramme to foster the INNOvative

capacity of small and medium-sized enterprises). The investigations

were conducted in cooperation with the Institute of Energy and

Environmental Technology e. V. (IUTA), Duisburg, Germany.

References

1. Collins CH. Safety in microbiology: a review. Biotechnol Genet

Eng Rev. 1984;1:141-166.

2. Kruse RH, Puckett WH, Richardson JH. Biological safety

cabinetry. Clin Microbiol Rev. 1991;4(2):207-241.

3. US Department of Health and Human Services, Centers for Dis-

ease Control and Prevention, National Institutes of Health. Bio-

safety in microbiological and biomedical laboratories, 5th ed.

http://www.cdc.gov/biosafety/publications/bmbl5/BMBL.pdf.

Published 2009.

4. European Committee for Standardization. Biotechnology—

Performance Criteria for Microbiological Safety Cabinets

[German version]. Berlin, Germany: Beuth Verlag, European

Committee for Standardization; 2000. European Standard

DIN EN 12469:2000-09.

5. NSF International Joint Committee on Biosafety Cabinetry. Bio-

safety Cabinetry: Design, Construction, Performance, and Field

Certification. Ann Arbor, MI: NSF International Joint Committee

on Biosafety Cabinetry; 2012. NSF International Standard/Amer-

ican National Standard NSF/ANSI 49.

6. Barbeito MS, Taylor LA. Containment of microbial aerosols in a

microbiological safety cabinet. Appl Environ Microb. 1968;16(8):

1225-1229.

7. Huang RF, Chou CI, Hung CH. Improving aerodynamics and

operator protection performance of biological safety cabinet

subject to dynamic influences. Environ Eng Sci. 2009;26(7):

1217-1226.

8. Macher JM, Fist MW. Effects of airflow rates and operator activ-

ity on containment of bacterial aerosols in a class II safety cabinet.

Appl Environ Microb. 1984;48(3):481-485.

9. Occupational Safety and Health Administration. Laboratory

safety: biosafety cabinets (BSCs) fact sheet. http://www.osha.

gov/Publications/laboratory/OSHAfactsheet-laboratory-safety-

biosafety-cabinets.pdf. Published 2011.

10. European Committee for Standardization. Fume Cupboards—

Part 3: Type Test Methods [German version]. Berlin, Germany:

Beuth Verlag, European Committee for Standardization; 2000.

European Standard DIN EN 14175-3:2004-03.

Table 2. Airflow Reduction Potential as a Function of Airflow Disturbing Factors.a

Airflow Reduction Potential, %

Undisturbed Dummy Worker Body Plate Moving Arm: R Moving Arm: L Walking Man

DownflowBSC 1 69 64 54 54 43 25BSC 2 58 63 58 53 35 29Mean 63.5 63.5 56 53.5 39 27

InflowBSC 1 57 57 43 43 34 9BSC 2 70 67 58 53 35 33Mean 63.5 62 50.5 48 34.5 21

Abbreviation: BSC, biological safety cabinet; L, left side; R, right side.aAirflow reduction potential: maximum acceptable negative deviation from the set point. Airflow-disturbing factors: airflow velocities at the nominal set points ¼100%.

Hinrichs et al 7