Basis® Headwall SystemDesign and Planning Guide

Page 2: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Table of Contents + At a Glance / 4

+ Configuring the Basis Headwall / 5

+ Basis Headwall System Architecture / 6

+ Basis Horizontal Units / 7

+ Basis Vertical Units / 8

+ Single and Double Configurations / 9

+ On-Wall Ceiling Interface / 10

+ Basis On-Wall Installation Guide / 11

+ Basis In-Wall Installation Guide / 12

+ Interface Styles that Support Two Rooms / 13

+ Nurse Call Placement Options / 14

+ Equipping the Basis Headwall / 15

+ Basis Wall Panel System / 20

+ Design Options – Laminates and Reveals / 24

+ Design Options – Horizontal On-Wall Wood Caps / 25

+ Hill-Rom’s Online Toolbox / 26

+ Codes and Standard Requirements / 28

Basis® Headwall System

2

Page 3: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

• The next generation value headwall system offering a wide range of clinical utility while providing an architectural style for today and the hospital of the future.

• Provides architects with an extensive range of design configurations, materials and finish options that will compliment many design themes.

• Provides nursing clinical functions ergonomically organized around the patient saving steps and motion.

• Supports the essential patient room clinical functions and acuity from ICU to Med-Surg with gas, power, data and the potential for integrated communications.

• Provides a simple, clean, linear look tailored to hospitality designs with commercially available laminates.

• Intended to simplify the time and complexity of designing and installing a patient room headwall by providing a factory tested pre-plumbed, pre-wired headwall ready to be installed by a general contractor.

Basis® Headwall System

3

Page 4: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

At a Glance

The Basis® headwall is the value hospital patient room headwall that provides hospital architects and designers a platform that offers design and styling options to create a less clinical-looking hospital experience.

The Basis headwall offers an extensive selection of finishes and decorative options including high pressure laminates and wood caps (On-Wall horizontal only). Panel reveals are available in black or silver.

The Basis headwall is a versatile panel and framework system. This provides nursing with a complete array of service options and loading that can support any care delivery from ICU to Med-Surg. Clinical services and nursing features are ergonomically designed to reduce reaching, bending and wasted steps. Accessory rails built in to the aluminum edge trim support a full range of clinical accessories.

• Superior aesthetics – Integration into hospital designs

• Clean linear design – Faceplate free

• Superior cleanability – Minimal gaps and crevices

• Installation – Quick and simple to install

• Price – Value priced for today’s cost conscious customers

Ceiling and floor indicators

Vertical Accessory

Rails

Horizontal Accessory

Rails Integrated Monitor Track

Gas Outlets

Nurse Call

Service Panels

Access Panels

Side Section Side Section

Center SectionIntegrated Bed Locator

Basis Horizontal unit configured for Med-Surg

Center SectionStandard Bed Locator*

Basis Vertical units configured for ICU

Side Section Side Section

* Bed locator sold separately. Back-to-back (3.625" wall) and On-Wall horizontal headwalls require a bed locator or other type of wall protector which should protrude out an amount equal to the depth of the headwall to prevent damage from beds.

Integrated Bed Locator

Stand-Alone Bed Locator

Basis Panel System

4

Page 5: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Configuring the Basis® Headwall– Gas, Electric, Data

Recommended configuration of a Basis headwall should be based on the loading of specified services and medical devices that will be connected to the headwall. Medical devices such as flowmeters and regulators come in a variety of sizes that if not spaced properly can restrict the use of available outlets. In addition, outlets, switches and devices should be placed with an ergonomic reach zone of 18" to 72" off the floor to avoid severe bending or reaching for staff.

• Vertical spacing for Rows of Outlets or Devices – 6" minimum

• Vertical spacing for in-line Gas Outlets – 18" minimum

• Horizontal spacing Gas Outlets – 4.5" minimum

• Horizontal spacing Electrical Devices – 1.812" minimum

Minimum spacebetween electrical

devices

Minimum spacebetween gas outlets

Minimumvertical spacingbetween rows

of devices

1.812"

4.5"

6"

5

Page 6: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Basis® Headwall System Architecture

Basis headwall installation options

The Basis headwall can be installed with three wall applications:

1. Mounted On-Wall – the headwall is surface mounted

to the finished wall serving that respective room. Surface

mounted depth is 5" off the finished wall.

2. Mounted In-Wall – the headwall is installed into a framed

stud wall opening and sits .75" off the finished wall surface.

The In-Wall configuration is available with 3.625" frame

depth only, but can function properly in both a 3.625" or 6"

stud wall.

3. Double-Sided – the headwall has two sides that serve

two adjacent rooms and protrudes either 0.75" or 2" off

the finished wall surface (depending on wall depth). This

configuration is available with a 6" frame and can be

configured for a 6" or 3.625" stud wall. Gas and electrical/

data services on each side can be configured differently from

each other.

1. MOUNTed ON-WALL

2. MOUNTed IN-WALL 3. dOUBLe-SIded IN 3.625" WALL (6" WALL SIMILAR TO #2)

6

Page 7: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Basis® Headwall Horizontal Units

72.00"

34.00"

60.00"18.00" 24.00" 18.00" 24.00" 24.00" 24.00"

16"24"32"

16"24"32"

108.00" 120.00"36.00" 18.00"36.00"18.00" 18.00" 24.00"18.00"18.00" 24.00"18.00"

16"24"32"

34.00"

16"24"32"

84.00" 96.00"24.00" 36.00" 24.00" 30.00" 36.00" 30.00"

16"24"32"

16"24"32"

34.00"

OxygenAir

Vacuum

OxygenAir

Vacuum

OxygenAir

Vacuum

Max222

Max886

Max886

Max222

Max886

Max886

Notes:

• Electrical/Data outlets have separate service panels from gas outlets.

• Gas outlets occupy separate rows from electrical/data outlets. They cannot be combined.

• The Basis headwall can support all electrical capacity needs for normal, emergency and low voltage. Consult your Hill-Rom architectural specialist for more information.

WARNING: MedICAL GAS OUTLeTS SHOULd NOT Be INSTALLed IN CeNTeR ACCeSS PANeL IN HORIZONTAL HeAdWALLS. CAUTION: GASeS ON HeAdWALL SHOULd NOT Be PLACed BeHINd HOSPITAL Bed.

all sizes are availaBle as on-wall, in-wall or douBle-sided units

7

Page 8: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Basis® Headwall Vertical Units

Most sizes are availaBle as on-wall, in-wall, or douBle-sided

96.0"102.0"

108.0"

63.0"57.0"

18.0" 16" 24" 32" 16" 24" 32" 16" 24" 32"

60.0"

8' – 6"9' 9' – 6"10'

8'

8' – 6"9' 9' – 6"10'

8'

72.0"84.0"

90.0"

12.0"

51.0"

63.0"57.0"

18.0" 16" 24" 32" 16" 24" 32" 16" 24" 32"16" 24" 32"

On-Wall not available

On-Wall not available

Max O2

AirVac

Max O2

AirVac

16"222

24"884

24"884

24"884

24"884

32"886

32"886

32"886

32"886

16"222

24"884

32"886

16"222

16"222

24"884

16"886

16"222

16"222

24"884

16"886

16"222

Note: In-Wall vertical headwalls do not extend to the ceiling. Leave at least 6" of drywall above unit.

8

Page 9: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Single or Double Configuration

In-Wall or Double-Sided

16"16" 16"58"*

24"24" 24"58"*

32"32" 32"

58"*

29"*

29"*

29"*

SingleUnit

DoubleUnits

On-Wall

Or

Basis®Headwall

WidthOptions

*Recommended Bariatric bed space is 66" between double units and 33" between bed centerline and single unit.

32"

24"

16"

Recommended Single and Double Unit Spacing

9

Page 10: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

On-Wall Ceiling Interface

Basis® Headwall Frame Channel (3.625" or 6")

Drop CeilingTile

Basis® Headwall Panel

Finished Ceiling Height (spec’dby architect)

Basis® Headwall Gap Filler

Basis® Headwall AluminumAccessory Extrusion/ Trim

Sheet Rock(by contractor)

Stud Wall Channel(3.625" or 6"by contractor)

Basis® Headwall Electrical and Gas Conduits

Basis® Headwall Wall Mount Bracket

on-wall Ceiling interfaCe

• On-Wall vertical headwall units terminate at the drop ceiling.

• Gas and electrical services extend through the drop ceiling.

• Drop ceiling requires pass through opening (reference your projects “Hill-Rom As Built Drawings”).

10

Page 11: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Basis® Headwall On-Wall Unit Installation Guide

1. Install Hanger brackets per Hill-Rom As-Built Drawings. Hanger

brackets should be bolted to backer plate behind drywall.

2. Hang headwall unit onto hanger brackets ensuring proper

alignment with rough-in openings. If required, bolt headwall

to hanger brackets.

3. Secure unit to bottom brackets through wall into backer plate.

4. Install manifolds. On 16" vertical units, install hoses to

manifold first.

5. Braze manifold risers. Make final electrical connections.

6. Complete installation by installing service panel and

surrounding gap trim.

Note: Reference Basis installation instructions for detailed information.

1

2

12

3

4

4

Note: On-Wall headwalls require a 16 gauge backer plate at wall attachment points.

11

Page 12: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

J-Box

J-Box

Cut out

Manifold

Manifold

1

12

2

3

4

4

3Stud Mount

Stud

Basis® Headwall In-Wall Unit Installation Guide

1. Construct headwall openings according to the Hill-Rom As-Built

Drawings. Cut out pass through opening in top header for gas pipe

and electrical conduit clearance.

2. Insert headwall unit into opening ensuring proper alignment

with studs.

3. Secure unit to studs at mounting points shown.

4. Install manifolds. On 16" vertical units, install hoses to manifold first.

5. Braze manifold risers. Make final electrical connections.

6. Complete installation by installing service panel and surrounding

gap trim.

Note: Basis In-Wall units must be installed BeFORe dRYWALL!

Note: double-Sided installation is similar to In-Wall installation, however the accessory trim is factory installed on one side. After installation, the additional accessory rail trim needs to be installed on the opposite side.

12

Page 13: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Interface Styles that Support Two Rooms

6" Stud 7.25"

.805".150"

Sheet Rock

6" StudChannel

1.587"

3.625"Stud

4.875"

1.956"

1.587"

Sheet Rock3.625" Stud

Channel

Note: double-Sided and In-Wall headwalls are designed to interface with 6" or 3.625" studs with typical 5/8" drywall. Other wall configurations may need to be evaluated for feasibility.

douBle-sided unit installed in a 6" stud wall

douBle-sided unit installed in a 3.625" stud wall

1.956"

1.587"

Sheet Rock

3.625" StudChannel

6"Stud 7.25"

*Acoustic drywall and caulk suppled by others

BaCK to BaCK unit installed in a 6" stud wall

13

Page 14: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Right of centerCenterline Not Available

NurseCall

No Nurse Call(Rough - in connection access panel)

Left of center

No Nurse Call(Rough - in connection access panel)

NurseCall

Nurse Call Placement on 16" Tall Headwalls

For 60" tall headwalls, Nurse Call is limited to positioning under J-Box. J-Box is typically

positioned on right side of headwall.

Nurse Call behind bed on center panel is limited to 3 Gang backbox with 2" of depth

*Not recommended for GUI based nurse calls.

Nurse Call Placement on 24" Tall Headwalls

24"

32"*Nurse Call Placement on 32" Tall Headwalls

32"

16"

63.0"(57" on 60" headwall)

Nurse Call Placement Options

horizontal headwalls

A Nurse Call system can be located in the center panel of a horizontal headwall, however, there are certain limitations related to location and backbox size. The diagrams below illustrate the placement options available:

vertiCal headwalls

Except for the 60" tall headwall, the Nurse Call system is generally placed at 63" or 32" AFF.

14

Page 15: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

30.00" 30.00"36.00"

24.00" 24.00"

24" UNIT

16" UNIT

32" UNIT

36.00"

18.00"36.00"

18.00" 18.00"

24.00"

18.00"18.00"

24.00"18.00"

36.00"

51.80"59.80"

67.80"

6.45"

87.60"

99.60"

111.60"

123.60"

32.20"

Equipping Basis® Headwall

The Basis headwall offers a selection of options that can provide enhanced functionality or flexibility to accommodate different building and installation conditions. When considering these options, consult with your Hill-Rom specialist for recommendations or further information.

An Integrated Bed Locator (IBL) provides rugged wall protection and bed locating for any type of hospital bed. It provides a smooth, flush panel that integrates with a horizontal Basis headwall while providing ancillary outlets on both sides of the bed.

Provisions for eight receptacles including:

• Normal power and emergency power

• Low voltage provisions

• Night light

• Bed nurse call connection

15

Page 16: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

7.12"

22.5"

50"

Equipping Basis® Headwall

stand alone Bed loCatorThe Basis headwall offers a selection of options that can provide enhanced functionality or flexibility to accommodate different building and installation conditions. When considering these options, consult with your Hill-Rom specialist for recommendations or further information.

A stand alone Bed Locator is also available and is compatible with all Basis headwall styles and types whether In-Wall, On-Wall, Double-Sided, vertical or horizontal. It can also be mounted onto the Basis Wall panel System. Made with durable metal and plastic construction, this option also offers provisions for eight receptacles including:

• Normal power and emergency power

• Low voltage provisions

• Night light

• Bed nurse call connection

22.5"

50"

54"(Min. Recommended)

7.12"

stand alone Bed loCator

16

Page 17: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Equipping Basis® Headwall

interConneCt Between Basis headwall and hill-roM® lights and Bed loCators The Basis headwall offers interconnects that provide a raceway for wiring from a Basis On-Wall vertical frame headwall to Hill-Rom patient lights or bed locators. This may be useful in renovations or other conditions where it is not desirable to provide power through in-wall rough-ins.

68 7/8"

51 9/16"

5 3/4"

9 3/8"17 7/8"

68 7/8"

11 1/16" 17 7/8"

6 1/2"

30 3/4"

40 7/16"

39 9/16"

40 7/16"

30 3/4"

68 7/8"

51 9/16"

5 3/4"

9 3/8"17 7/8"

68 7/8"

11 1/16" 17 7/8"

6 1/2"

30 3/4"

40 7/16"

39 9/16"

40 7/16"

30 3/4"

Interconnect for Light note:

Use of the interconnect with a bed locator reduces the number of available receptacles from eight to six.

When interconnect is used with bed locator, electrical devices are not available in bottom row of vertical chase.

interConneCt for 48" 696 il light interConneCt for 36" 696 il light

Wiring fed from Basis to light and bed locator

Interconnect for bed locator

interConneCt for 0052B Bed loCator interConneCt for 201606 Bed loCator** FOR ISOLATED POWER ONLY

17

Page 18: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Equipping Basis® Headwall Chase and transition sleeve option for horizontal headwalls For On-Wall applications, Basis offers vertical chase options as well as horizontal transition sleeve options in a variety of sizes to easily retrofit existing facilities and couple multiple headwalls together. This allows the installation of headwalls without opening up the existing facility wall. For environments such as PACU or semi private Med-Surg units the transition sleeve allows single point attachment of gas, electrical and data while serving multiple bed spaces (up to 3 beds).

7.00"

32.20"

Bed Centerline72.5" - 126.5" in 6" Increments

13.4"

Transition available in 12.5" & 18.5" sizes

31.9"

78.1"

Chase

Transition

Integris® Light(696 style)

Light Rail

Optional ChaseExtension

• Chase is available on the left, right, or centered between two Basis Headwalls.

• The Chase/Transition Sleeve option allows the use of the Integrated Bed Locator.

• An optional Chase extension creates an improved room aesthetic (see elevation below).

• Transitions are available in two widths: 12.5" and 18.5"

• Transition and Chase are compatible with all 16", 24", and 32" tall horizontal units up to 108" in length.

• Up to 3 headwall units can be connected together with the chase/transition sleeve.

• A Light Rail is available which allows multiple Patient Lights (696 Style) to be chased together (up to 3 lights).

• The Light Rail provides optional provisions for electrical and data outlets.

18

Page 19: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Equipping Basis® Headwall

The Basis® headwall offers a selection of optional accessories that can be attached and repositioned along a horizontal or vertical accessory rail. Below is a sample of the many accessories we offer. When planning accessory placement, consider the location of attached components and adjoining gas and electrical outlets to avoid blocking outlets or creating interferences with attached devices.

see Basis headwall aCCessory Catalog for a CoMplete list of availaBle CoMponents.

Note:

• For patient monitors, horizontal headwalls must use a separate monitor track that mounts to the integrated accessory tracks.

• “Over the top” mounted accessories (Integris style) are only compatible with the On-Wall horizontal headwalls.

19

Page 20: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Basis® Wall Panel System

wall panel systeM for horizontal headwallsThe Basis headwall offers a flexible system of decorative wall panels that seamlessly integrates to all the different horizontal sizes and styles available. Whether your preference is In-Wall, On-Wall, or Double-Sided, the Basis Wall Panel System provides methods of seamless integration.

typiCal horizontal wall panel Configurations

20

Page 21: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Basis® Wall Panel System

typiCal vertiCal wall panel Configurations

wall panel systeM for vertiCal headwalls

21

Page 22: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Basis® Wall Panel System wall panel aesthetiC and utility optionsThe Basis headwall offers a number of aesthetic and equipment options to enhance both the appearance and functionality of a Basis system with Wall Panels. These range from lighting options to equipment management options to bed protection and electrical options.

ALUMINUMReVeAL

BUTT JOINT

ACCeSSORY TRACK

Bed LOCATOR Bed BUMPeRS

INTeGRIS® LIGHT

1 - 4 GANG BACK BOXeS

• Accessory Track only available horizontally. Accommodates most Basis headwall accessories.

• Aluminum reveals can be horizontal, vertical or both.

• Butt joints can be horizontal, vertical or both.

• Integris® Light available in 36" and 48" versions.

• Bed Bumpers available in horizontal or vertical arrangements.

• Back boxes are available for high and low voltage applications only.

}

.25”

1.13”

ALUMINUM ReVeAL BUTT JOINT ACCeSSORY TRACK

22

Page 23: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Basis® Wall Panel System

wall panel installation proCess for headwalls with interConneCts Between Bed loCators and lights

When using the interconnects that provide a raceway for wiring from a Basis On-Wall vertical frame headwall to Hill-Rom patient lights and bed locators (see page 14), the installation process is different. Wall panels will need to be installed before the headwall and other hardware is installed.

INTeRCONNeCTS (Page 14)

Walls are closed and finished before wall panels

are applied

Wall Panels are then attached to the walls

Headwall, bed locator, lights and interconnects are then mounted

onto the wall panel system

Headwall Hanger Bracket

1

2

3

23

Page 24: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Design Options – Laminates and Reveals

unliMited laMinate ChoiCes

Basis® headwall panels are available with most commercially available HPL finishes. Lead time varies on pattern availability.

triM Color ChoiCes

QuiCK ship laMinates

A standard selection of laminates are available for short lead time orders that include the following:

Customized style options can be developed based on volume, lead time and material. Materials must comply with building codes and UL requirements. Consult a Hill-Rom sales associate for further information.

Colors for reference only.

Black

Silver

Light Oak (WilsonArt)

Wild Cherry(WilsonArt)

Honey Maple(Pionite)

Almond Leather(WilsonArt)

Antique White (WilsonArt)

Natural Maple(Formica)

Light Neutral(Nevamar)

Latte Walnut (Formica)

english Oak(WilsonArt)

Limber Maple(WilsonArt)

24

Page 25: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Customized style options can be developed based on volume, lead time and material. Materials must comply with building codes and UL requirements. Consult a Hill-Rom sales associate for further information. Wood caps are only available on horizontal On-Wall configurations.

Colors for reference only.

Design Options – On-Wall Wood Caps

The Basis® On-Wall headwall is available with solid wood caps as an alternative option to anodized aluminum top and bottom trim. The caps come with an integrated accessory rail attached to the front. The wood caps are available for On-Wall horizontal headwalls only.

wood Cap option for horizontal on-wall headwalls are availaBle in the following standard stains:

Light Oak Natural MapleWild Cherry Colonial CherryMedium Oak Honey MapleMontana Walnut dark Cherry

25

Page 26: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Hill-Rom reserves the right to make changes without notice in design, specifications and models. The only warranty Hill-Rom makes is the express written warranty extended on the sale or rental of its products.

©2013 Hill-Rom Services, Inc. ALL RIGHTS RESERVED.

171802 rev 2 E 02/04/2013 ENG – US

Hill-Rom’s Online ToolboxHill-Rom offers an online toolbox for its Architectural Products, Liko® Overhead Ceiling Lifts, NaviCare® Nurse Call and Bed product lines.• Drag and drop product drawings

• Multiple file formats

• CSI specification sheets

• Design & planning guides

• AIA-accredited CEU courses

Access these tools by visiting www.caddetails.com/hillrom/

Revit Models Available• Nurse call

• Boom systems

• Headwall systems

• Lift systems

• Beds

HILL-ROM DESIGN PORTAL

To access CAD drawings, BIM (Revit) Objects and Specifications, click on one of the links below.

Hill-Rom®Architectural Products

HITSHospital Information Technology

Liko®The Safe Lifting People

Hill-Rom®Hospital Beds

Start Here >> Start Here >> Start Here >> Start Here >>

Hill-Rom’s Online Toolbox

www.caddetails.com/

26

Page 27: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Hill-Rom’s On-Line Toolbox

27

Page 28: Basis® Headwall System - CADdetails · • Electrical/Data outlets have separate service panels from gas outlets. • Gas outlets occupy separate rows from electrical/data outlets

Codes and Standard Requirements

teChniCal data

American National Standards Institute (ANSI)/ Underwriters Laboratories, Inc. (UL)

• ANSI/UL 723 Standard Test for Surface Burning Characteristics of Building Materials

• ANSI/UL 514A Metallic Outlet Boxes

American Society of Mechanical Engineers (ASME)

• ASME B16.22 Wrought Copper and Copper Alloy Solder Joint Pressure Fittings

ASTM International

• ASTM B819 Standard Specification for Seamless Copper Tube for Medical Gas Systems

Compressed Gas Association (CGA)

• CGA-G-4.1 Cleaning Equipment for Oxygen Service

• CGA V-5 Diameter Index Safety System (Non-interchangeable Low Pressure Connections for Medical Gas Applications)

National Fire Protection Association (NFPA)

• NFPA 70 National Electrical Code (NEC)

• NFPA 99 Health Care Facilities

National Electrical Manufacturers Association (NEMA) Underwriters Laboratories Inc. (UL)

• UL 94 Standard for Safety of Flammability of Plastic Materials for Parts in Devices and Appliances testing

• UL 1047 Isolated Power Systems Equipment

Approvals

• UL listed by Underwriters Laboratory Inc.

Technical Properties

• Copper vacuum service tubes are manu factured to ASTM B819 standards

• Duplex and simplex sockets are UL-listed, NEMA style, hospital grade

• Gas service systems are assembled and tested in accordance with NFPA 99

• Gas terminal outlets are UL-listed (VXHT) quick-disconnect or DISS, conforming to CGA V-5

Fire Performance

• The Basis Headwall System meets UL fire performance requirements based on power type used and meets all applicable current NFPA 70 and NFPA 99 standards.