Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
autoclart® plus User Manual Version 2, September 2016
1
USER MANUAL
autoclart® plus – User Manual
autoclart® plus User Manual Version 2, September 2016
2
autoclart® plus
CLART, CLART-Strip and SAICLART are brands registered by GENOMICA.
For greater detail in relation to the information set forth in this manual please access the
following website: www.genomica.com
GENOMICA, S.A.U.
Parque Empresarial Alvento, Edificio B
Calle Vía de los Poblados, 1 – 1ª planta
28033 Madrid, Spain
www.genomica.com
Version 2
September 2016
autoclart® plus User Manual Version 2, September 2016
3
1. TABLE OF CONTENTS
1. TABLE OF CONTENTS ................................................................................................... 3
2. PRINCIPLE OF THE ASSAY ............................................................................................. 6
3. INTENDED PURPOSE ................................................................................................... 8
4. USE8
5. CLASSIFICATION ........................................................................................................... 9
6. PRODUCT TRADE NAME AND REFERENCES ..................................................................... 9
7. DESCRIPTION ............................................................................................................. 10
8. COMPONENTS / CONSUMABLES .................................................................................. 11
8.1 COMPONENTS ................................................................................................... 11
8.2 CONSUMABLES. ................................................................................................. 12
9. ACCESSORIES. ............................................................................................................ 13
10. CHARACTERISTICS. ................................................................................................... 15
11. UNPACKING, STORAGE AND PACKING. ....................................................................... 16
11.1. UNPACKING. ................................................................................................... 16
11.2. STORING THE PACKAGING. ............................................................................... 17
11.3. STORING THE EQUIPMENT. ............................................................................... 18
11.4. TRANSFER / TRANSPORTATION. ........................................................................ 18
11.5. PACKING. ........................................................................................................ 18
12. ENVIRONMENTAL CONDITIONS .................................................................................. 19
12.1. VENTILATION. ................................................................................................. 19
12.2 - ENVIRONMENTAL CONDITIONS. ........................................................................ 19
autoclart® plus User Manual Version 2, September 2016
4
13. PRECAUTIONS OF USE ............................................................................................... 21
14. PREPARATION OF THE FACILITY PRIOR TO INSTALLATION .............................................. 23
15. INSTALLATION .......................................................................................................... 24
16. TURNING THE EQUIPMENT ON AND OFF ..................................................................... 26
16.1. ON ................................................................................................................. 26
16.2. OFF ................................................................................................................ 27
17. START OF THE ANALYSIS ............................................................................................ 28
17.1. ENTRY OF THE TYPE OF ASSAY ........................................................................... 30
17.2. ENTRY OF THE SAMPLE REFERENCES .................................................................. 31
17.3. WORK LIST ...................................................................................................... 33
17.4. ANALYSIS SETTINGS. ........................................................................................ 34
18. ANALYSIS ADDING THE SAMPLE AUTOMATICALLY ........................................................ 34
19. ANALYSIS ADDING THE SAMPLE MANUALLY ................................................................ 43
20. READING ONLY ANALYSIS .......................................................................................... 45
21. READING ................................................................................................................. 45
22. RESULTS .................................................................................................................. 46
23. REANALYSIS OF THE SAMPLE ..................................................................................... 49
24. MAIN MENU RECORDS ............................................................................................. 52
22.1. SEARCH BY RUN. .............................................................................................. 52
22.2. SEARCH BY SAMPLE. ........................................................................................ 54
25. MAIN MENU SETTINGS ............................................................................................. 54
22.3. INSTALLING / UPDATING A PROGRAM. ............................................................... 55
22.4. SYSTEM .......................................................................................................... 56
autoclart® plus User Manual Version 2, September 2016
5
22.5. BACK-UP / EXPORT .......................................................................................... 57
22.6. SERVICE .......................................................................................................... 58
DECONTAMINATION PROGRAM ......................................................................................... 59
CLEANING PROGRAM .......................................................................................................... 60
26. MAINTENANCE / CALIBRATION .................................................................................. 61
27. CLEANING ............................................................................................................... 61
28. CARING FOR THE EQUIPMENT AFTER USE ................................................................... 63
29. FAQ ........................................................................................................................ 64
29.1 UPDATING AND INSTALLATION OF SOFTWARE ...................................................... 64
30. TECHNICAL ASSISTANCE SERVICE ............................................................................... 65
APPENDIX A. TABLE OF FREQUENT ERRORS AND RESPECTIVE SOLUTIONS ........................... 66
APPENDIX B. DECLARATION OF CE CONFORMITY .............................................................. 69
Declaration of CE Conformity ...................................................................................... 69
APPENDIX C. ABBREVIATIONS USED ................................................................................ 70
APPENDIX D. WARNING SYMBOLS AND SIGNS. ................................................................. 71
APPENDIX E. DECONTAMINATION CERTIFICATE. ............................................................... 73
autoclart® plus User Manual Version 2, September 2016
6
2. PRINCIPLE OF THE ASSAY
autoclart® plus has been developed for the automation of both viewing and reading and
interpreting the results of the in vitro diagnostic kit manufactured by GENOMICA.
The technique is based on:
1. The viewing of a specific reaction through the precipitation of an insoluble
product in microarray areas in which the amplified products have undergone
hybridization.
2. The capacity to associate the tip pattern generated in the array with a specific
diagnostic. This is conducted using an image capturing and processing system and a
specific algorithm for each kit. The technique is based on a system designed to
capture and process the images obtained in the microarray, which will result in the
automatic generation of a single report for each sample analysed. The algorithms used
in each kit are comprehensively validated during the process involving the evaluation of
the performance of the product.
The results obtained by CAR® are automatically displayed on the touch screen and can
be printed out or exported to the LIMS system or a memory unit using the USB
connection.
The image capture device includes a LED light source, a diffusor (to homogenize the
light) and a CCD camera to capture the image.
autoclart® plus User Manual Version 2, September 2016
7
Furthermore, the device is equipped with an automatic positioning system to ensure the
plate is in the correct place.
The analysis of the microarray image is conducted using the SAICLART® software
developed by GENOMICA for the interpretation of microarrays. The analysis of the
image is fully automatic, preventing subjectivity and the interaction of the user, thereby
ensuring accurate results in a rapid and repeatable manner.
Compared to classic microarray systems, this system significantly simplifies the
hybridization and viewing processes, allowing for the simultaneous detection of multiple
molecular markers for diagnostic use, while providing the controls required to
guarantee the reliability of the results.
autoclart® plus User Manual Version 2, September 2016
8
3. INTENDED PURPOSE
The equipment will be used exclusively by personnel from accredited laboratories and
using GENOMICA S.A.U. in vitro diagnostic kits, illustrated on the website
www.genomica.com
Any use other than that specified may undermine the performance of the equipment.
4. USE
IN VITRO DIAGNOSTIC HEALTH PRODUCT
autoclart® plus is for exclusive use with the reagents supplied by
GENOMICA.
Follow the instructions carefully. We suggest you read these instructions
before putting them aside. They contain useful information on the correct
and efficient management of the equipment.
This apparatus should not be used by people without the appropriate
experience or knowledge, unless under the surveillance of someone
responsible for their safety or who has been given instructions on the use
of the apparatus.
The use of autoclart® plus in a manner other than that specified in this
manual will result in the loss of the guarantee provided.
autoclart® plus User Manual Version 2, September 2016
9
5. CLASSIFICATION
autoclart® plus complies with the provisions of Council Directive 98/79 / CE on in vitro
diagnostic health products, in accordance with Spanish law pursuant to Royal Decree
1662/2000 of 29 September 2000, and is classified under “other products” in Appendix
III of the Directive.
Moreover, GENOMICA S.A.U. declares that the autoclart® plus equipment has been
manufactured in accordance with the following safety and quality standards and
requirements in force (EN ISO 13485, EN 61010-2-101 and EN 61326-2-6).
6. PRODUCT TRADE NAME AND REFERENCES
autoclart® plus
Reference: ACP
autoclart® plus User Manual Version 2, September 2016
10
7. DESCRIPTION
Figure 1. autoclart® plus.
autoclart® plus is a fully automated system for processing plate format assays.
The tool is designed for processing the sample, the addition of reagents, the removal of
liquids, heating, cooling, agitating the incubation plate and scanning the plate at the end
of the assay.
As such, it is equipped with a Peltier system capable of maintaining precise control over
the temperature during the assay, an incubation agitation system and a system to
transfer the plate to the core of the equipment where the camera for capturing images
and analysis is located.
autoclart® plus features an X/Y/Z axis mechanism for adding and removing samples and
reagents. It also features a system for disposable tips, as well as an independent 4-
needle reagent suction system connected to a peristaltic pump.
The control of volumes during the addition of samples is conducted by piston pumps
driven by “micro-step” type motors controlled by software. autoclart® plus is equipped
with a computer which operates by means of a touch screen and an interface designed
autoclart® plus User Manual Version 2, September 2016
11
specifically for the equipment.
The user can use the touch screen to view the number of samples and the assay to be
executed, whereby it is possible to process a minimum of 4 and a maximum of 96
samples per assay.
The equipment will conduct the analysis, reading and the interpretation of the results
automatically.
8. COMPONENTS / CONSUMABLES
8.1 COMPONENTS
The different components of autoclart® plus are illustrated in the diagram below:
Figure 2. Components of autoclart® plus.
autoclart® plus User Manual Version 2, September 2016
12
The instructions for the assembly of the different components of the equipment are set
forth in the table below.
Remember that most of the components need to be placed in a specific position.
8.2 CONSUMABLES.
1. Distilled water bottle
holder
Place the holder on the markers in the position
marked 1.
Remember it will only fit in one position.
2. Distilled water bottle
Screw on the cap. Place the bottle in the holder in
the position marked 2. Connect the tube from the
piston pumps to the top of the bottle.
3. Tip rack holder Place the holder on the markers in the position
marked 3.
4. TL trough Place the TL trough on the markers in the position
marked 4.
5. SH trough Place the SH trough on the markers in the position
marked 5.
6. CJ and RE troughs
Place both troughs on the markers in the position
marked 6. The troughs are labeled and only fit in
one position.
7. Liquid waste trough
Place the trough on the markers in the position
marked 7. Connect the tube from the peristaltic
pump.
8. Tip waste trough
Place the trough on the markers in the position
marked 8.
autoclart® plus User Manual Version 2, September 2016
13
Disposable tips will need to be used in addition to the GENOMICA diagnostic kit.
The equipment has been designed to be used with a specific disposable tip model.
Please contact our local distributor for more information.
Do not attempt to use any other type of tip. This may cause the equipment to
malfunction or result in a breakdown which is not covered by the guarantee.
9. ACCESSORIES.
The equipment comes with the following accessories.
ACCESSORY 1: SOFTWARE.
The analysis of the images and the interpretation of the results are conducted by
specific software for each kit, designed and implemented by GENOMICA.
GENOMICA distributes this software to its clients in accordance with the diagnostic kits
it provides.
Take particular care to ensure the USB device used both for updating software
and exporting data is not infected with a virus. GENOMICA will not be held liable
for removing a virus which has infected the system as a result of improper use
by the user.
In the event the user selects the wrong software, an error message appears
in the results report. Please review the analysis you wish to conduct and
select the correct software.
autoclart® plus User Manual Version 2, September 2016
14
ACCESSORY 2: A/C POWER CABLE.
Used to connect the equipment to the power mains with a plug.
autoclart® plus has been designed for use with GENOMICA diagnostic kits, and ,
as such, no other type of tube, vial or material should be used. The use of
autoclart® plus in a manner other than that specified in this manual will result
in the loss of the guarantee provided.
Do NOT use any accessory which has not been supplied by GENOMICA.
GENOMICA will not be held liable for the use of materials not supplied
by the manufacturer.
All the accessories required for the correct use of the equipment will be
supplied by the manufacturer.
In the event any of these components is missing or damaged, please
contact GENOMICA or a sales representative immediately.
autoclart® plus User Manual Version 2, September 2016
15
10. CHARACTERISTICS.
Type of instrument Table equipment
Processing capacity 4 to 96 samples per assay
Processing time Normally less than 3 hours for 96 samples
Temperature control. PID control
Processing volume 3 L to 100 L
Dispensing and tolerance system
Piston pumps:
3-10 µL ±10%
100 µL +0% / -5%
+0% / -5%
Software Integrated software
Network voltage and frequency 100 to 240 VAC with 5A grounding
50/60 Hz
Fuses 5A UL Approved – Ø5x20 mm - T5AL250VAC
Dimensions 620(H) x 650(D) x 800(W) mm / (W = 1150 mm with the screen)
Weight 80 kg
Power 500 W (max)
Noise level (dB) ~ 60 dB
Electromagnetic compatibility Classes A 30 MHz to 1000 MHz
autoclart® plus User Manual Version 2, September 2016
16
Electromagnetic susceptibility
10 V/m 80 MHz to 1000 MHz
3V/m 1.4 GHz to 2 GHz
1V/m 2 GHz to 2.7 GHz
UV lamp T8 15W UVC G13 (GEMICIDA)
11. UNPACKING, STORAGE AND PACKING.
We suggest you read the entire manual before unpacking and installing the equipment
or using it. Pay attention to all the hazard alerts and warnings. Failure to do so may
result in serious injury to the user or damage to the equipment. In order to ensure the
protection this equipment provides is not damaged, do not use or install it in a manner
other than that specified in this manual.
11.1. UNPACKING.
BEWARE, HEAVY EQUIPMENT!
At least 2 persons are required to move the equipment. The packed equipment
weighs around 110 kg.
The use of gloves is recommended when handling the equipment as some of
the components of the container may have sharp edges.
On receiving the equipment check the packaging is not damaged; if it is, please notify
GENOMICA immediately.
autoclart® plus User Manual Version 2, September 2016
17
Follow the steps below to open the packaging:
1. Place the box in a vertical position to open it.
2. Open the box and remove the user manual; continue in accordance with the
instructions set forth in the manual.
3. Remove the flaps and the lid.
4. Pull off the outer wooden cover and remove the protective bubble paper.
5. Lift the equipment to one side and remove the side foam at the same time. Repeat
this procedure on both sides.Two persons are required to take the equipment out
of the container, one in front of it and one behind it.
7. Inspect the equipment to check for any clear signs of damage during
transportation. Report any damage to your local representative
immediately.Check the serial number and the number on the delivery note match.
9. Open the door to the equipment and remove the box of accessories from inside.
Para in order to prevent any kind of damage to autoclart® during delivery, the
internal mobile components are held in place using fastening systems
which need to be removed prior to use.
autoclart® plus features a safety screw to protect the optical system from
damage while being transported. This screw must be removed prior to
using autoclart® plus.
11.2. STORING THE PACKAGING.
Fold the wooden box in order to store the packaging correctly.
autoclart® plus User Manual Version 2, September 2016
18
Keep the original packaging materials!
Equipment returned for repair should be dispatched in the
original packaging material in order to protect it from damage
during transportation.
11.3. STORING THE EQUIPMENT.
If the equipment is temporarily stored on the ground or an uneven surface, the
appropriate measures should be taken to protect it.
The equipment features delicate components which operate in a vertical position, and
care should be taken not to twist the structure of the equipment while it is being stored.
This manual should be kept close at hand for everyday
reference!
In the event of transfer, ensure the manual is kept together with
the equipment to provide the new owner with the instructions
of use and the respective warnings.
11.4. TRANSFER / TRANSPORTATION.
In the event the equipment needs to be transported after use in the laboratory, it will
need to be cleaned and decontaminated. Moreover, a certificate of decontamination
should be issued.
In the event the equipment needs to be moved to another location, it should be lifted
and transported. It should never be pushed or dragged.
11.5. PACKING.
The safety screw and the three new fastening straps must be in place when transporting
the equipment. Follow the respective instructions below:
autoclart® plus User Manual Version 2, September 2016
19
1. Turn off the equipment completely. Unplug the equipment from all power
sources.Place the safety screw in its housing located on the lower right-hand side
in the rear of the working area and screw it in. Move the robot arm fully to the left
and put the safety straps in place.
12. ENVIRONMENTAL CONDITIONS
12.1. VENTILATION.
Take the following recommendations into account:
Do not place the equipment in areas in which there are currents of air.
Moreover, the equipment does not require specific ventilation.
Place the equipment at least 100 mm away from walls and equipment.
12.2 - ENVIRONMENTAL CONDITIONS. Take the following recommendations into
account:
The equipment is for exclusive use indoors and should not be exposed to major
variations in temperature and humidity. It has been designed for use in normal
laboratory conditions.
Keep the equipment away from solvents and acidic vapours.Avoid vibrations and
direct sunlight to ensure accurate results.
Altitude Up to 2000 m.
Temperature Ambient temperature: 18 °C – 24 °C.
Humidity Maximum relative humidity of 80% for temperatures of up to 31 °C
decreasing linearly to around 50% for 40 °C.
Network voltage 100 to 240 VAC – 5A 50/60 Hz with variations of up to ± 10 %.
autoclart® plus User Manual Version 2, September 2016
20
Transport and
storage conditions From 5 °C to 50 °C.
Once the equipment has been assembled wait for 1 hour before turning it
on to avoid potential problems caused by condensation. All waste liquid is
stored in the equipment until the end of the test, and, as such, it does not
need to be drained.
Do not use this equipment in the vicinity of powerful electromagnetic
radiation as this may affect its performance.
The user is responsible for guaranteeing an electromagnetic environment
which is compatible with the equipment to ensure the device operates
correctly.
When working in a magnetic environment the user should check the device
is performing correctly prior to routine use.
autoclart® plus User Manual Version 2, September 2016
21
13. PRECAUTIONS OF USE
Please read the following warnings carefully.
All the precautions contained on this page should be strictly complied
with in order to prevent harm to individuals and/or damage to the
equipment.
In the event of a malfunction, contact the technical support service and
do not attempt to deal with the problem using your own means.
Please acknowledge and comply with all the warning signs on the equipment and its
accessories.
CAUTION
Do not operate
instrument with
cover open -
Risk of injury
CAUTION
Risk of trapping
fingers during
instrument
operation
CAUTIONHot Surface
Both the lower and upper microplate heaters may reach temperatures of up
to 65 °C during the assay. Make sure you do not touch these surfaces when
the assay has been paused.
Avoid touching the equipment with wet hands. Do NOT touch the internal
components of the equipment. You could suffer an electric shock. Any
such harm or damage would not be covered under the guarantee.
autoclart® plus User Manual Version 2, September 2016
22
The equipment will turn itself off automatically when the door is opened.
Do not handle the equipment´s safety components. Never touch the
equipment during an assay.
The equipment may be used with infectious / or hazardous substances.
We recommend wearing a laboratory coat and protective gloves when
handling the equipment and reagents. If a reagent should come into
contact with the skin or the eyes, wash with plenty of water.
Avoiding looking into the UV light when the UV decontamination
procedure is underway.
Do not open the door or touch the door sensor during the UV
decontamination procedure.
Take special care with the electrical components. Comply with safety
measures at all times. Repairs to the electrical system may only be
conducted by qualified personnel.
Do not touch any of the equipment´s electronic components. In the event
of breakage or failure contact the GENOMICA technical support service.
Always unplug autoclart® plus when performing work involving the use of
liquids or when conducting maintenance on the electrical system or the
equipment´s internal parts.
In the event the power supply in the laboratory is cut off or surges, we
recommend the use of an uninterrupted power supply (UPS) source.
Do not attempt to open or force the outer protective housing of the
apparatus. Avoid vibration or any sudden movement of the equipment
during the reading and analysis process.
autoclart® plus User Manual Version 2, September 2016
23
Gloves should always be worn when handling potentially infectious
substances (such as human samples and reagents) and which may come
into contact with the skin.
The user should take care not to contaminate the inside or outside of the
equipment with biological samples.
14. PREPARATION OF THE FACILITY PRIOR TO INSTALLATION
Follow the detailed instructions below carefully for the correct preparation of the facility
prior to installation:
Take the following recommendations into account:
Always place the equipment on a firm, flat surface. The assembly surface should be
level in all directions to ensure the equipment functions correctly. Make sure
autoclart® plus is installed on a totally horizontal, stable surface.The minimum
surface area required is 1200 mm in length by 700 mm in depth. The equipment
measures 800 mm in length, but requires an additional 350 mm for the installation of
the touch screen.The equipment should not be placed more than one metre away
from the power socket.
The equipment opens vertically and it is 1000 mm in height when fully open. Make
sure you clear the top of the equipment before opening it.The autoclart® plus main
switch is located on the left hand side panel of the equipment. Place the equipment
where it can be easily accessed.
autoclart® plus User Manual Version 2, September 2016
24
Before connecting the equipment to the power mains, check that the
voltage is 110-230 V (AC) and 50/60 Hz and that the system is
earthed.
Do not install autoclart® plus near a source of light, intense heat or
external vibrations.
Never touch the switches with wet hands.
Turn the equipment off completely before unplugging it from the
mains.
15. INSTALLATION
Follow the detailed instructions below carefully for the correct installation of
the equipment.
autoclart® plus User Manual Version 2, September 2016
25
15.1. INSTALLATION OF THE TOUCH SCREEN
Follow the steps detailed below carefully:
Figure 3. M6 x 80 screw.
1. Check the device is on a level surface.
2. Fasten the arm of the screen by
tightening the rear clamp using the M6 x
80 screw.
3. Insert the M4 screw on the other side to
limit the inclination of the touch screen.
We recommend requesting the assistance of a second person from this point on.
Figure 4. The yellow circle highlights the M5
screws for the bracket. The red circle highlights
the screen cable.
4. Ensure the screen is in the correct position and
fasten it by tightening the two M5 screws in the
bracket.
5. Once it is in place, check the screen cannot be
moved.
6. Connect the screen cable to the equipment and the
USB cable to one of the equipment´s USB ports.
autoclart® plus User Manual Version 2, September 2016
26
16. TURNING THE EQUIPMENT ON AND OFF
16.1. ON
Follow the steps detailed below carefully:
Connect the power cable to the panel on the side of the equipment and plug it into the
mains (connecting the equipment to the mains automatically ensures the connection to
earth). Press the ON button.
Figure 5. Power cable and ON switch.
Once it has been turned on, autoclart® plus will take a few seconds to start up. During this
time the device will conduct a series of internal checks on both the hardware and software,
and both the axes and the Peltier plate will be placed in their original position.
If the equipment door is open the message “please close the door” will appear on the
screen. Please close the door to continue.
Whilst on the screen, the progress bar will expand.
autoclart® plus User Manual Version 2, September 2016
27
Figure 6. Starting up the equipment
The main screen will appear once the equipment start up has concluded.
16.2. OFF
1. Remove the microplate from the equipment if you have not done so already.Remove
the deionized water bottle from its bracket.
3. Remove the waste troughs (liquids and solids) and dispose of the waste in accordance
with local and state provisions.
4. Remove the reagents troughs and dispose of the waste in accordance with local and
state provisions.
5. Clean the reagent troughs with water (see the cleaning section) and put the
accessories back in the equipment.
6. Clean the waste troughs with bleach diluted to 4% (see the cleaning section) and
rinse. Put the accessories back in the equipment.Close the door to the equipment.
8. Return to the main screen and press the OFF button. When the screen darkens, turn
off the switch on the side of the equipment.
autoclart® plus User Manual Version 2, September 2016
28
Figure 7. The main menu. The OFF button is highlighted in red.
9. We recommend disconnecting the equipment completely for long periods of
inactivity.
17. START OF THE ANALYSIS
The reading and analysis of a sample is a simple sequential process consisting of a series of
menus from which to select the different options involved in the process.
Press the “new analysis” option on the main menu, which appears automatically when the
equipment is turned on.
Figure 8. “New Analysis”, highlighted in red
A new screen will appear featuring a diagram of a 96-well microplate.
autoclart® plus User Manual Version 2, September 2016
29
Figure 9. Diagram of a microplate on the “New Analysis” menu.
This is the application´s most important menu, as it enables the user to establish:
The number of samples to be analysed. The assay reference number.
The sample reference number.
autoclart® plus User Manual Version 2, September 2016
30
17.1. ENTRY OF THE TYPE OF ASSAY
Automatic method:
A bar code reader needs to be connected to autoclart® plus in order to select the assay
automatically.
Press the “auto assays” button located on the lower left-hand side of the screen and read
the assay code provided by GENOMICA.
All the CS will appear in light-grey in the microplate diagram, indicating that the assay has
not been selected.
Then, read the specific bar code for the test provided by GENOMICA and the CS in which
the assay has been defined will automatically change to the specific colour of the same.
Once the process has been concluded press the tick.
Figure 9. Selection of “auto assays”.
Manual method:
Select the CS number to be used in the assay. Press the X at the top of the specific position
and then press the wells. Unselecting all the wells will unselect the entire CS.
CLART® HPV4 will be assigned to all CS by default; to change the software press the edit
symbol above the selected CS.
A drop-down menu will appear enabling you to change only the CS selected or all the CS
selected as of this time appear with the software chosen on selecting “auto select”.
autoclart® plus User Manual Version 2, September 2016
31
Figure 11. Selection of the bands to be processed
Figure 12. Manual entry of the assay references.
N.B: Remember that if you intend to conduct the same assay in all the bands, press “AUTO SEL”
after selecting the assay. The symbol “A” will appear alongside the assay selected.
17.2. ENTRY OF THE SAMPLE REFERENCES
In the same manner as the previous sections, the sample references can be entered
automatically by installing a bar code reader, or manually.
autoclart® plus User Manual Version 2, September 2016
32
Automatic entry of sample references
A bar code reader needs to be connected to autoclart® plus in order to enter the sample
references automatically.
Select “samples” on the lower right-hand side of the screen and a sub-menu will appear
featuring a list of the positions.
Scan the codes of the samples to be analysed. The number of samples to be analysed will
depend on the number of references added, which will always occupy consecutive
positions.
Once all the references have been added press the tick to return to the previous menu.
Figure 13. Entry of samples.
Manual entry of sample references
Select “samples” on the lower right-hand side of the screen and a sub-menu will appear
featuring a list of the positions.
Type the sample references into the selected positions.
Once all the references have been added press the tick to return to the previous menu.
In the event no specific reference is assigned to the sample, the software
will assign each sample a consecutive number from 1 to 96, starting with
the sample located in position 1A.
autoclart® plus User Manual Version 2, September 2016
33
Figure 14. Entry of samples. Positions sub-menu.
Automatic entry of sample references from files
To enter references automatically from files, first contact the GENOMICA product specialist.
17.3. WORK LIST The work list will enable us to check all the data entered up to now is
correct.
If you should need to correct any data, select “Back” (left blue arrow) to edit the test or
reference, as explained in the previous sections.
If all the information is correct, press “continue” (right blue arrow) to start the analysis.
autoclart® plus User Manual Version 2, September 2016
34
Figure 14. Work list
17.4. ANALYSIS SETTINGS.
The equipment can be used in three different settings:
To add samples automatically.To add samples manually.For reading only.
Select the desired option on the touch screen.
Figure 15. Analysis settings
18. ANALYSIS ADDING THE SAMPLE AUTOMATICALLY
Select the “add samples automatically” option.
Check the following list of steps which appears on the touch screen.
Press the blue arrow (“next step”) to move on to the next step.
autoclart® plus User Manual Version 2, September 2016
35
STEP 1/11: LOAD COMPLETE NEW TIP RACKS
Figure 16. Loading tip racks.
Place the tip racks in their holder, taking
into account there is only one possible
position.
1 ml tip racks at the back of the
holder (red rack).
10 μl tip racks at the front of the
holder (white racks).
Ensure the racks are full. Remember that
the consumption of 1 ml tips is the same
for all assays regardless of the number of
samples analysed.
Press down on the four corners of the
racks to ensure they are properly in place
and close the fastening clamps (highlighted
in yellow circles),
Figure 17. Tip racks
autoclart® plus User Manual Version 2, September 2016
36
Step 2/11: LOADING THE MICROPLATE
Figure 18. Loading the Microplate
Raise the microplate holder cover and insert the
plate with as many CS as required for the analysis of
the samples, as illustrated in the image on the right.
Ensure position A1 is in the top left corner.
Figure 19. Position of the microplate
Ensure the CS adaptor is correctly inserted and press
it firmly downwards.
Close the cover and turn the lock.
Figure 20. Closing the cover, highlighted in
yellow.
autoclart® plus User Manual Version 2, September 2016
37
Step 3/11: CHECK THE USED TIP
TROUGH IS IN PLACE AND EMPTY.
Figure 21. Empty tip trough
Step 4/11: CHECK THE WASTE TROUGH
IS IN PLACE AND EMPTY.
Figure 22. Empty waste trough
Step 5/11: CONNECT THE TUBE TO THE
WASTE TROUGH.
Figure 23 a. Connecting the tube
autoclart® plus User Manual Version 2, September 2016
38
Place the waste trough in its
position.
Attach the connecting tube to the
waste trough and put on the cover
as illustrated in the image on the
right.
Figure 23 b. Connecting the tube to the waste trough, as highlighted in
yellow.
Step 6/11: LOAD THE WATER BOTTLE
WITH 250 ML OF DISTILLED WATER.
Figure 24 a. Bottle of distilled water
autoclart® plus User Manual Version 2, September 2016
39
Fill the bottle with 250 ml of distilled water and place
it in its holder.
Take into account that the consumption of water is
the same regardless of the number of samples
processed. The bottle should be refilled prior to each
assay.
Close the bottle with the red cap and insert the pump
tube through the cap as illustrated in the image on
the right.
Figure 24 b. Bottle of distilled water. Tube
inserted through the red cap.
REMEMBER THAT REAGENTS AND THE AMPLIFICATION PLATE SHOULD NOT BE ADDED UNTIL
THE TROUGHS HAVE COOLED DOWN.
Figure 25. autoclart® plus audio signal highlighted in red.
The equipment will emit an audio signal when the troughs have been prepared.
This audio signal indicates that the troughs have cooled down and the assay may continue.
Touch the audio symbol to turn it off.
The equipment automatically calculates the volume of reagents required for viewing in
accordance with the number of samples to be processed. Refer to the kit manual used at
the time of preparing the assay reagents.
autoclart® plus User Manual Version 2, September 2016
40
Take note that the viewing reagent troughs only fit in the correct position.
TL SH Troughs
CJ RE Troughs
Figure 26. Position of the viewing reagent troughs
Step 7/11. PLACE THE SPECIFIED
VOLUME OF WASH BUFFER IN THE
TANK.
Figure 27. Wash buffer volume.
autoclart® plus User Manual Version 2, September 2016
41
Step 8/11. PLACE THE SPECIFIED
VOLUME OF HYBRIDISATION SOLUTION
IN THE TANK.
Figure 28. Volume of hybridisation solution.
Step 9/11 – PLACE THE SPECIFIED
VOLUME OF DILUTED CONJUGATE IN THE
TANK.
Figure 29. Volume of conjugate
Step 10/11. PLACE THE SPECIFIED
VOLUME OF DEVELOPER IN THE TANK.
Figure 30. Volume of developer
autoclart® plus User Manual Version 2, September 2016
42
Step 11/11 – LOAD THE AMPLIFICATION
PLATE.
Figure 30 a. Load PCR plate.
Open the cover and place the PCR
plate in the holder.
Check the PCR plate is in the correct
position and position A1 is in the top
left corner and press it firmly
downwards.
Figure 30 b. Loading the PCR plate. Position in the holder.
Close the cover and turn the lock as
illustrated in the image on the right.
Figure 30 c. Loading the PCR plate. Locking the PCR plate.
autoclart® plus User Manual Version 2, September 2016
43
Once all these steps have been completed, close the cover to the equipment and press the
continue button (white arrow) to start the analysis. The screen will show the estimated
time of the end of the assay, the processing phase (which will change throughout the assay)
and the warning not to open the door to the equipment.
Figure 31. Countdown of the time remaining.
19. ANALYSIS ADDING THE SAMPLE MANUALLY
Select the “add samples manually” option.
autoclart® plus User Manual Version 2, September 2016
44
Check the following list of steps (as explained in item 2 of the previous section) on the
touch screen. Press the blue arrow (“next step”) to move on to the next step.
Figure 32. Manual addition of the samples.
Once all these steps have been completed, close the cover to the equipment and press the
continue button (white arrow) to start the analysis.
When the equipment reaches the sample addition step it will emit an audio signal and the
message “load samples” will appear on the screen.
Figure 33. Loading samples
Remove the autoclart® plus microplate and add the denatured PCR product to each of the
CS.
Then place the microplate back in autoclart® plus and turn the lock. Close the equipment
and press the white arrow to continue.
autoclart® plus User Manual Version 2, September 2016
45
20. READING ONLY ANALYSIS
Select the “reading” option.
Figure 34. Analysis reading settings.
Load the microplate with the desired number of CS to be read and turn the lock.
Close the equipment and press the white arrow to continue.
21. READING
When the test has been concluded the equipment will automatically read the microplate for
the purpose of analysis. A progress bar will appear when the analysis begins, indicating the
status of the process.
autoclart® plus User Manual Version 2, September 2016
46
Figure 35. Reading progress bar
Figure 36. End of the reading
When the progress bar reaches 100% the test has been completed and the results can be viewed,
exported and/or printed out.
22. RESULTS
22.1. VIEWING RESULTS
Press “report” to view the results of the complete run. The work list will reappear, which
will now include the information on the results.
autoclart® plus User Manual Version 2, September 2016
47
In order to view the results in an individual manner press the corresponding will in the
image of the microplate. This will access the specific report on the sample.
Figure 37. Individual results.
The following information is available on this sub-menu:
- Raw data: Provides access to the numeric, raw and standarised values obtained from
the analysis of the arrays.Image: Provides access to the image of the array taken by
autoclart® plus. This image may be exported or printed out individually.
FIgure 38. Image of a result.
- Export: Enables the user to export the information to an external server, LIMS or mass
data storage system established as an export route beforehand.
autoclart® plus User Manual Version 2, September 2016
48
22.2. OTHER OPTIONS.
The following options appear on the results menu:
Export:
Enables the user to export the complete assay, reports and results to an external server,
LIMS or mass data storage system established as an export route beforehand.
On selecting Export a list will appear of all the assay samples from which to select the
results you wish to export.
Select the samples you wish to export. This can be conducted individually or by selecting
the entire assay by pressing “select all”. If you wish to delete what you have selected press
“Delete selection”. If you wish to close the export menu click on “Close”.
Press “send” to export the results. A new window will then appear with the message
“Please wait, exporting”. Do not close the window until the process has ended.
Once the assays have been exported a new window will appear with the message “export
concluded”. Press “Ok”.
Figure 39. Selecting the test to export to LIMS
Figure 40. Export concluded.
Reports:
Once the analysis of the samples has ended autoclart® plus will generate two types of
report. A report with the results of the complete assay (work list) and individual reports for
each of the samples (see item 22.1.).
autoclart® plus User Manual Version 2, September 2016
49
The work list includes the following information:
- Date and time of the process
- Position of the sample in the microplate
- The code of the analysis conductedThe type of analysis
conductedThe sample referenceThe results
Figure 41. Report on the assay work list. The print option is highlighted in red.
The work list can be printed out. To do so connect a printer to autoclart® plus and press
the printer icon.
23. REANALYSIS OF THE SAMPLE
If, after having analysed the plate, a sample is colored red, this means the grid could not be
aligned automatically by the software. However, the user may reanalyse the sample, or in
other words, align the gird manually.
In order to reanalyse a sample press the same and the “Reanalysis” option menu will
appear on the screen automatically.
Only wells which CAR® was unable to analyse will be available for reanalysis.
autoclart® plus User Manual Version 2, September 2016
50
Figure 42. Initial menu of the reanalysis option. Change of assay reference.
First of all, check the assay selected is the correct one, and if not, change it. To do so press
the drop-down menu located in the middle of the screen.
Selecting “Start reanalysis” will access the application, in which the user can use a zoom
device to help align the grid correctly.
The correct alignment consists of placing the top left marker of the grid in the correct
position, and then doing the same with the bottom right marker. The latter marker enables
the user to increase or reduce the size of the grid. Once both markers are in position, select
“Reanalysis”.
Figure 43. Procedure for the manual alignment of the grid.
To put the markers in place, simply press on them on the screen and drag them into their
final position.
autoclart® plus User Manual Version 2, September 2016
51
Once the markers are in position the software provides a general view of how the grid
would fit on the array. If the grid is in the correct position, press “Accept” in order to
continue with the reanalysis.
Figure 44. Position of the grid on the array.
Aligning the grid manually implies the intervention of the user in the results, which will be
registered in the final report generated in the form of the following message.
The image has been reanalysed by the user.
Figure 45. Message on the manual alignment of the samples.
autoclart® plus User Manual Version 2, September 2016
52
24. MAIN MENU RECORDS
The “records” option on the main menu provides a results search tool.
Figure 46. The main menu. Access to assay records.
This results search tool can be used for an analysis or run or a sample.
22.1. SEARCH BY RUN.
Fig 47. Search by run.
This search should be conducted using:
autoclart® plus User Manual Version 2, September 2016
53
1. Two dates established by the user, one for the start and one for the end of the
search.
2. Search parameter: There are three different search parameters to choose from: run
ID, assay and plate.
Figure 48. Search parameters.
Once the parameters have been established, a list will appear with all the assays
which comply with the search parameters defined. Select the assay in question and
press “Next” to access the desired results.
Figure 49. List of analyses displayed.
autoclart® plus User Manual Version 2, September 2016
54
22.2. SEARCH BY SAMPLE.
Figure 50. Search by sample.
On selecting “Search by sample” a sub-menu will appear as in the “Search by run”
procedure, which will enable the user to set the parameters by date, run ID, assay ID,
plate ID, and particularly, sample ID.
Once the parameters have been established, a list will appear with all the samples which
comply with the search parameters defined. Select the sample in question and press
“Next” to access the desired results.
25. MAIN MENU SETTINGS
autoclart® plus User Manual Version 2, September 2016
55
Figure 51. Settings.
Press the “Settings” option to access a sub-menu with a number of options for
configuring autoclart® plus.
Figure 52. Settings sub-menu
22.3. INSTALLING / UPDATING A PROGRAM.
Select the “install / update” option to access a sub-menu containing a list of all the
programs currently installed in the equipment and version of the same.
Selecting a specific program and pressing “delete” will enable you to remove that program.
The “install” options provides access to a search menu from which to select the external
memory device connected to autoclart® plus containing the new software to be installed.
autoclart® plus User Manual Version 2, September 2016
56
Select the .Zip file provided by GENOMICA to install the program. The installation is
automatic.
Figure 53. Assay install / update menu.
22.4. SYSTEM
The “System” option provides the user with information on the software, hardware
and the serial number of the equipment.
Figure 54. System information screen.
This option also provides:
autoclart® plus User Manual Version 2, September 2016
57
- Export: Select the export route. Pressing the “…” symbol will access a browser tab
from which to select the external memory or LIMS network you wish to use to export
results.Back-up: Select the back-up route. Pressing the “…” symbol will access a
browser tab from which to select the external memory you wish to use for the back-
up of the assays stored in the equipment (see item 24.3). Language: change the
language used on the equipment.
- By default: select the program which appears in a new run by default.
22.5. BACK-UP / EXPORT
This option enables the user to copy or copy and delete only. The
latter option will also remove these assays from the equipment´s
memory.
The final destination should be specified prior to exporting or backing up data.
The information contained in autoclart® plus may be stored on any external memory
device with a USB connection or exported to the laboratory´s LIMS network, provided the
user has selected the export route (see item 24.2).
Select “back-up” on the “settings” menu and press continue (tick). Select the data you wish
to copy in accordance with the date, run ID and / or assay ID.
There are two back-up options once these parameters have been defined:
- Copy: The results are copied on the defined back-up route.
- Move: The results are copied on the defined back-up route and deleted from the
autoclart® plus memory. This ensures a result cannot be deleted unless it has been
copied first. Select the option in question and the process will be conducted
automatically, returning to the settings” menu once it has been completed.
autoclart® plus User Manual Version 2, September 2016
58
Figure 55. The back-up process.
22.6. SERVICE
Press “Service” in the settings menu where you will find three protocols for the
maintenance of the equipment.
- Decontamination: This program decontaminates the equipment using UV
light.Cleaning: This program cleans the equipment´s tubing and pump
system.Technical Assistance: This menu is for the exclusive use of specialised
technical assistance personnel. It is protected with a password and users have no
access.
Figure 56. Service menu.
autoclart® plus User Manual Version 2, September 2016
59
DECONTAMINATION PROGRAM
Select the “decontamination” program on the “services” menu.
Then complete the following 6 steps displayed on the touch screen:
1. Put the empty tip holders in place.
2. Load the empty used tip trough.
3. Put the empty sand dry waste trough in place.
4. Put the empty reagent troughs in place.
5. Load the empty hybridisation solution trough.
6. Ensure neither the PCR plate nor the CS microplate has been loaded.
7. Close the equipment door.
Once completed press continue. Up to this point the user may reverse the process by
pressing the cancel button (X), which will take you back to the “services” menu.
The decontamination process will begin and a countdown will appear on the screen.
Ensure the cover to the equipment remains closed throughout this process.
Avoiding looking into the UV light when the UV decontamination procedure is
underway.
Do not open the door during the UV decontamination process.
autoclart® plus User Manual Version 2, September 2016
60
Figure 57. The decontamination process.
An audio signal will indicate the end of the decontamination process.
CLEANING PROGRAM
Select the “cleaning” program on the “services” menu and follow the instructions which
appear on the touch screen.
1. Fill the bottle with disinfectant: Fill the extra bottle with bleach diluted at 4% or
another disinfectant whenever no foam is formed. Avoid all soapy solutions.Check
the waste trough is present and empty.
Once completed press continue. Up to this point the user may reverse the process by
pressing the cancel button (X), which will take you back to the “services” menu.
A countdown will appear to determine when the user needs to intervene. Once the process
has finished autoclart® plus will flash and a message will appear on the touch screen to
replace the clean bottle with the full bottle of distilled water.
Once the cleaning program has ended the equipment will emit an audio signal once again.
autoclart® plus User Manual Version 2, September 2016
61
26. MAINTENANCE /
CALIBRATION
Both the maintenance and the calibration of the equipment should be
conducted by qualified personnel. In the event of an incident contact
your GENOMICA technical assistance department or your retailer.
Only change the fuses when they blow. Following these steps to do so:
1. Remove the fuse box by pressing the upper and lower tabs.
2. Replace the blown fuses and put the fuse box back in place by pressing down firmly
on the middle of the same.
3. The type of fuse is indicated on the autoclart® plus label: T5A 250V (5 X 20 MM).
27. CLEANING
The equipment should be cleaned on a regular basis.
Only the top and outer surface of the equipment should be cleaned, using towels soaked
in alcohol or patented decontaminants*, followed by water and appropriate detergent*.
This exterior cleaning process should be accompanied by the execution of the program for
cleaning the inside of the equipment (see item 25.4).
Take the following recommendations into account:
- Do not attempt to remove the panels with the aim of cleaning the inside of the
equipment, as this involves the risk of injury and may damage the device.
- Always put on protective gloves before cleaning the equipment.
autoclart® plus User Manual Version 2, September 2016
62
- Use an appropriate detergent to clean the surrounding surfaces.
- Use a damp cloth and detergent to clean all the areas around the base and the reagent
area.
- Use the same technique to wash all the parts which might come into contact with an accidental spill.
- Clean all the reagent trays at the end of each run and ensure the waste trough has been
emptied.
- We recommend the use of UV goggles as a precautionary measure when conducting the
UV decontamination.
* The user is responsible for not using cleaning or decontamination products which might
represent a hazard as a result of the reaction with parts of the equipment or the
material comprising the same.
In the event of doubt in relation to the compatibility of the decontamination or cleaning
agents, please contact your local agent or the GENOMICA technical assistance
department.
autoclart® plus User Manual Version 2, September 2016
63
28. CARING FOR THE EQUIPMENT AFTER USE
The device
This equipment should not be disposed of using public waste
systems. In compliance with domestic and local European legal
regulations (EU Directive 2002/96/CE), the users of European
electrical equipment are required to return old equipment and
whose life cycle has expired to the manufacturer.
N.B: With regard to recycling, please contact GENOMICA or the
supplier of the equipment for instructions on how to return
equipment whose life cycle has ended for appropriate disposal.
Packaging
Please keep the packaging material throughout the service life of the
equipment.
The packaging components (wood, cardboard, polyethylene, etc.) are
products comparable with normal urban solid waste and can be
disposed of easily. However, we suggest disposing of these products
separately (selective collection) in accordance with the requirements
in force in relation to correct recycling. DO NOT THROW THE
PACKAGING PRODUCTS AWAY INTO THE ENVIRONMENT.
autoclart® plus User Manual Version 2, September 2016
64
29. FAQ
29.1 UPDATING AND INSTALLATION OF SOFTWARE
Figure 58. USB memory with installation files.
1. Connect the external memory to a free USB port in the equipment.
2. Access the “Settings” menu.
3. Press the “Install / Update” button.
4. You should see unit C:\ of autoclart® plus and its external memory on the browser,
which is normally identified with the letter D:\.
5. autoclart® plus may take a few minutes to recognise the memory. Return to the main
menu, wait for a moment and try again.Access the USB memory by pressing the box
Take particular care to ensure USB device used both for updating
software and exporting data is not infected with a virus.
GENOMICA will not be held liable for removing a virus which has
infected the system as a result of improper use by the user.
autoclart® plus User Manual Version 2, September 2016
65
with the icon. You will now see the compressed files required for the
installation.Select the kit you wish to install and press “Install”.
8. The installation ends with a window indicating that the process has been successfully
concluded.
Please close the window by pressing the tab in the top right corner.
If you wish to install more programs, repeat the operation from step 4. When you have
finished return to the main menu and reboot the equipment. In the event error messages
appear, please contact your technical assistance department. Remember to quote these
messages in your request for assistance.
30. TECHNICAL ASSISTANCE SERVICE
Remember that only GENOMICA SAU technical service personnel or associates may
conduct authorised repairs and services.
You may contact the GENOMICA, S.A.U. technical assistance department on:
Tel: + 34 91 674 89 90
Fax: + 34 91 674 89 91
E-mail: [email protected]
autoclart® plus User Manual Version 2, September 2016
66
APPENDIX A. TABLE OF FREQUENT ERRORS AND RESPECTIVE SOLUTIONS
N.B: the correct working order of this equipment should be checked after all maintenance,
servicing or repair operations.
Dealing with simple problems
Visual inspection
A comprehensive visual inspection of the equipment prior to using it will often reveal real or
potential problems. Conducting the following checks will help to determine the state of the device.
Check for any signs of evident damage. With the equipment turned off, open the lid and inspect the
base of the device for any spills or obstruction. Remove the obstruction and clean up the spill where
applicable.
Mechanics
Ensure the tip picking / removal system is in the initial position, move the X and Y axes by hand to
check for obstructions. In the event of obstructions, look for mechanical impediments such as loose
screws.
Check the suction probe needles are clean and in good condition. Gently clean each probe with
cotton wool soaked in alcohol. The tips should be straight – if for some reason they are bent, a
qualified technician should replace them accordingly.
Ensure nothing is blocking the path of the reagent plate.
Fluids
Check the tubes are not damaged and ensure there are no visible folds along the 'Y' and the "X" axis
which might affect the processing of the sample. Pay particular attention to the tip picking /
ejection mechanism connections.
Any problem detected during the visual inspection needs to be dealt with before turning the
equipment on, and failure to do so may cause even more damage to the device.
Start-up
autoclart® plus User Manual Version 2, September 2016
67
The specific functions of the equipment should be checked during the start-up process.
The table below provides details of the start-up process:
Step Action Comment Possible cause Consequence Solution
Pass
1
Master Controller
constant load, check
modules and the
compatibility of
firmware
Immediately if
functioning
correctly
Corrupt data
Incorrect firmware
Module disconnected
Error code
Check BUS cables
Check modules
Contact the technical assistance
department
2 Start-up of module C Home sensor not
detected Error code
Contact the technical assistance
department
3 Start-up of module AZ Home sensor not
detected Error code
Contact the technical assistance
department
4 Start-up of module Z Home sensor not
detected Error code
Contact the technical assistance
department
5 Start-up of module XY Home sensor not
detected Error code
Contact the technical assistance
department
6 Start-up of module T Home sensor not
detected Error code
Contact the technical assistance
department
7 Start-up of module RP Home sensor not
detected Error code
Contact the technical assistance
department
8 Ejection of tips Home sensor not
detected Error code
Contact the technical assistance
department
9 Check of pumps Home sensor not
detected Error code
Contact the technical assistance
department
Error codes
autoclart® plus User Manual Version 2, September 2016
68
The equipment features an error management mechanism with parameters which can be
monitored in real time.
In the event of an error the device will stop immediately and an error code will appear on the touch
screen.
The error codes are stored in a register. ALL CASES of error codes (S) should be reported to the
service provider immediately for analysis, stating in which phase of the assay the error occurred.
New measures
If the error persists or the solution fails to DEAL WITH the error please contact the Authorised
Technical Assistance Department.
autoclart® plus User Manual Version 2, September 2016
69
NO
MIC
A,
S.A
.U.
Mad
rid
Mer
can
tile
Reg
iste
r, V
olu
me
40
4,
Fo
lio
17
4,
Sh
eet
nº
M7
90
6,
En
try
1,
Dat
e 0
5-1
1-9
0-
C.I
.F.
A-7
9434
718
APPENDIX B. DECLARATION OF CE CONFORMITY
Declaration of CE Conformity
Manufacturer: GENOMICA, S.A.U.
Parque Empresarial Alvento
Edificio B.Vía de los Poblados, 1 -1ª planta.
28033 Madrid
Product: Autoclart® Plus
Classification: Other products
References: ACP
Evaluation procedure: Appendix III to Directive 98/79/CE (Declaration of
CE Conformity).
Classification Class A. Rule 5. Appendix VII.
Genomica, S.A.U. hereby declares that the product autoclart ®Plus complies with the
requirements of Directive 98/79/CE of 27 October on health products for in vitro
diagnostic, registered in Spanish law under Royal Decree 1662/2000 of 29 September
2000.
Quality standards applied EN ISO 13485. EN ISO 14971. EN 980, EN 13612. EN
ISO 18113-3, EN 62304. EN 61010-2-101. EN
61326-2-6
Entity Notified Not applicable
CE certificate nº Not applicable
Validity of the certificate: Not applicable
Cargo: Directora General
autoclart® plus User Manual Version 2, September 2016
70
APPENDIX C. ABBREVIATIONS USED
CS: CLART-strip®
Fig: Figure
PCR: Polymerase Chain Reaction
PDF: Portable Document Format generated by Adobe.
LIMS: Laboratory Information Management System
USB: Universal Serial Bus
autoclart® plus User Manual Version 2, September 2016
71
APPENDIX D. WARNING SYMBOLS AND SIGNS.
This symbol indicates the danger of harm being caused to both the operator
and the robot.
This symbol warns of the possible damage or harm the operator may suffer in
the event the normal robot function is interrupted.
This symbol warns of the danger of possible damage or harm caused by the
high voltage of the equipment.
This symbol refers to important information on autoclart® plus. Please read
the attached instructions carefully.
Refer to the instruction for use.
Manufacturer
Manufacture sheet
Indicates that the equipment is used as an in vitro diagnostic product
Expiration date
Serial number
Batch code
Catalogue number
autoclart® plus User Manual Version 2, September 2016
72
CE marking
Fuses 2 x T4H250 V (5 x 20 mm)
Electrical and electronic products, do not mix with standard waste
Symbol Colour Standard Description Location
Symbol and
Border:
Black
IEC 60417 –5032 Alternate Current
A/C
Identifying label on
the
equipment
Symbol and
Border:
White
IEC 60417 –5007 ON On switch with
filter
Symbol and
Border:
White
IEC 60417 –5008 OFF On switch with
filter
Rear: Yellow
Symbol and
Border: Black
IEC 60417 –5032 Danger: Hot
Surface On the heated area
Rear: Yellow
Symbol and
Border: Black
IEC 60417 –5007 Danger: risk of
damage
Junto a etiquetas de advertencia
Rear: Yellow
Symbol and
Border: Black
- Danger: UV light On the warning
label
autoclart® plus User Manual Version 2, September 2016
73
APPENDIX E. DECONTAMINATION CERTIFICATE.
Decontamination Certificate
This equipment:
Model: ____________
Serial number: ____________
Has not been in contact with biological samples of an infectious nature.
Date:
Signature: