Assessment of a Ventilator Circuit Adapter to be Used in Conjunction with the Respimat Soft Mist Inhaler RATIONALE W ith the introduction of the Respimat Soft Mist Inhaler (SMI) there is a desire for an adapter to allow its use in ventilator circuits. An adapter (RespiConnect* Adapter) was assessed in terms of drug delivery and safety features. Patients on mechanical ventilation either involving prophylactic delivery of bronchodilators, or receiving such medication to treat a severe COPD exacerbation may need continuous therapy Excellence By Design J. Suggett, M. Nagel Trudell Medical International, London, Canada. MD-800A-0717. * trade-marks and registered trade-marks of Trudell Medical International. trade-marks and registered trade-marks of Boehringer Ingelheim. Copyright © Trudell Medical International 2017. ERS International Congress Milan, Italy September 9 – 13, 2017 METHODS Each adapter (n=10) was inserted in the inspiratory limb of a heated adult ventilator circuit • A 7.0 mm Endotracheal Tube (ET) was attached to a filter and the distal side of the filter coupled to an adult test lung to simulate a patient • A Servo ventilator (SV900C, Siemens, Sweden) delivered the required breathing pattern: T idal volume (V t ) = 600 mL • 10 breaths/minute • Inspiratory/expiratory (I/E) ratio = 1:3 • 5 cm H 2 O Positive End Expiratory Pressure (PEEP) RESULTS • During the experiment the circuit did not lose pressure and no alarms were triggered, indicating that the self sealing valve was effective • Mean values of 30.3% and 29.4% LC were determined for SAL and IPR respectively CONCLUSIONS • It is known that not having a good consistent seal can jeopardize the safety of the patient through circuit leakages and a consistent seal is also needed in order to achieve adequate drug delivery to the patient The presence of a self-sealing valve and safety cap for the RespiConnect* Adapter should ensure ventilator circuit integrity and therefore also help protect the safety of the patient • Approximately 30% LC of both drug components in the SMI was delivered to the distal end of ET tube via the adapter This study provides clinicians assurance regarding circuit integrity and likely drug delivery performance when using the RespiConnect* Adapter • A ventilator circuit humidifier (model MR50JHU, Fischer- Paykel, Auckland, NZ) was used to maintain conditions close to body temperature and saturation with water vapour • Five actuations of Combivent SMI were delivered, each actuation followed by 6 complete breathing cycles • Between actuations the SMI was removed from the adapter and primed according to manufacturer’s instructions • Salbutamol (SAL) and Ipratropium (IPR) were recovered from the filter and quantified by HPLC as % Label Claim (LC) STUDY PURPOSE The RespiConnect* Adapter was designed for safety and mechanical performance The self-sealing mechanical valve maintains the integrity of the ventilator circuit at all times; a redundant cap is incorporated for extra safety This study assessed mechanical performance through its ability to retain circuit integrity as well as aerosol drug delivery in a ventilator RespiConnect* Adapter Trudell Medical International London, Canada RespiConnect* Adapter with Respimat SMI Experimental Lab Set-up Flow pattern of RespiConnect* Adapter with RespiConnect SMI engaged • Aerosol delivery was optimized through CFD analysis during adapter design Servo ventilator Inhalation (humidified air) Respimat SMI RespConnect* Adapter Wye connector ET Filter Adult test lung Circuit humidifier Exhalation Drug In Air + Drug Out Air Out

Assessment of a Ventilator Circuit Adapter to be Used in … · 2020-04-06 · Assessment of a Ventilator Circuit Adapter to be Used in Conjunction with the Respimat † Soft Mist

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Assessment of a Ventilator Circuit Adapter to be Used in … · 2020-04-06 · Assessment of a Ventilator Circuit Adapter to be Used in Conjunction with the Respimat † Soft Mist

Assessment of a Ventilator Circuit Adapter to be Used in Conjunction with the Respimat† Soft Mist Inhaler

RAtIonAle•WiththeintroductionoftheRespimat†SoftMistInhaler(SMI)thereisadesireforanadaptertoallowitsuseinventilatorcircuits.Anadapter(RespiConnect*Adapter)wasassessedintermsofdrugdeliveryandsafetyfeatures.

•Patientsonmechanicalventilationeitherinvolvingprophylacticdeliveryofbronchodilators,orreceivingsuchmedicationtotreatasevereCOPDexacerbationmayneedcontinuoustherapy

Excellence By Design

J. Suggett, M. NagelTrudellMedicalInternational,London,Canada.

MD-800A-0717.*trade-marksandregisteredtrade-marksofTrudellMedicalInternational.†trade-marksandregisteredtrade-marksofBoehringerIngelheim.Copyright©TrudellMedicalInternational2017.

ERS International Congress Milan, Italy

September 9 – 13, 2017

MethodS•Eachadapter(n=10)wasinsertedintheinspiratorylimbofaheatedadultventilatorcircuit

•A7.0mmEndotrachealTube(ET)wasattachedtoafilterandthedistalsideofthefiltercoupledtoanadulttestlungtosimulateapatient

•AServoventilator(SV900C,Siemens,Sweden)deliveredtherequiredbreathingpattern:

•tidalvolume(Vt)=600mL

•10breaths/minute

•Inspiratory/expiratory(I/E)ratio=1:3

•5cmH2OPositiveEndExpiratoryPressure(PEEP)

ReSUltS•Duringtheexperimentthecircuitdidnotlosepressureandnoalarmsweretriggered,indicatingthattheselfsealingvalvewaseffective

•Meanvaluesof30.3%and29.4%LCweredeterminedforSALandIPRrespectively

ConClUSIonS•Itisknownthatnothavingagoodconsistentsealcanjeopardizethesafetyofthepatientthroughcircuitleakagesandaconsistentsealisalsoneededinordertoachieveadequatedrugdeliverytothepatient

•thepresenceofaself-sealingvalveandsafetycapfortheRespiConnect*Adaptershouldensureventilatorcircuitintegrityandthereforealsohelpprotectthesafetyofthepatient

•Approximately30%LCofbothdrugcomponentsintheSMIwasdeliveredtothedistalendofETtubeviatheadapter

•thisstudyprovidescliniciansassuranceregardingcircuitintegrityandlikelydrugdeliveryperformancewhenusingtheRespiConnect*Adapter

•Aventilatorcircuithumidifier(modelMR50JHU,Fischer-Paykel,Auckland,NZ)wasusedtomaintainconditionsclosetobodytemperatureandsaturationwithwatervapour

•FiveactuationsofCombivent†SMIweredelivered,eachactuationfollowedby6completebreathingcycles

•BetweenactuationstheSMIwasremovedfromtheadapterandprimedaccordingtomanufacturer’sinstructions

•Salbutamol(SAL)andIpratropium(IPR)wererecoveredfromthefilterandquantifiedbyHPLCas%LabelClaim(LC)

StUdY PURPoSe•theRespiConnect*Adapterwasdesignedforsafetyandmechanicalperformance

•theself-sealingmechanicalvalvemaintainstheintegrityoftheventilatorcircuitatalltimes;aredundantcapisincorporatedforextrasafety

•thisstudyassessedmechanicalperformancethroughitsabilitytoretaincircuitintegrityaswellasaerosoldrugdeliveryinaventilator

RespiConnect*AdapterTrudell Medical International London, Canada

RespiConnect*AdapterwithRespimat†SMI

Experimental Lab Set-up

Flow pattern of RespiConnect* Adapter with RespiConnect† SMI engaged

•AerosoldeliverywasoptimizedthroughCFDanalysisduringadapterdesign

Servo ventilator

Inhalation (humidified air)

Respimat† SMI

RespConnect* Adapter

Wye connector

ET Filter

Adult test lung

Circuit humidifierExhalation

Drug In

Air + Drug Out Air Out