2-6 July - Facoltà di Scienze - Pisa, Italy ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement V. Herrero *, R. Colom*, R. Gadea*, J. Cerdà*, A. Sebastià *, C.W. Lerche**, J. Benlloch** * DSD – Departamento Ingeniería Electrónica – Universidad Politécnica de Valencia ** IFIC – Instituto de Fisica Corpuscular – Consejo Superior de Investigaciones Científicas / Universidad de Valencia

ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

  • Upload
    others

  • View
    8

  • Download
    0

Embed Size (px)

Citation preview

Page 1: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

2-6 July - Facoltà di Scienze - Pisa, Italy

ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

V. Herrero *, R. Colom*, R. Gadea*, J. Cerdà*, A. Sebastià *, C.W. Lerche**, J. Benlloch**

* DSD – Departamento Ingeniería Electrónica – Universidad Politécnica de Valencia

** IFIC – Instituto de Fisica Corpuscular – Consejo Superior de Investigaciones Científicas / Universidad de Valencia

Page 2: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

2ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Research Project Collaboration

IFIC (Corpuscular Physics Research Institute)High Energy Physics research.Nuclear Medical Instrumentation

Portable Gamma Camera for medical applicationsDevelopment of PET systems.

PET mammography.Small Animal PET.

DSD (Digital Systems Design)Mixed Signal, front-end and read-out electronics for high energy phisics.High-Speed Data Acquisition and Digital Signal Processing Systems for PET Mammography.Mixed Signal ASIC development.

Page 3: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

3ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Detector Front-End

First Development of Detector Front-End and Read-Out

qContinuous

LSOScintillator

Flat Panel 8x8 MultianodePMT Hamamatsu H8500

…64 Anode Outputs

Resistive Current Division Network [1]

. . .

Ia Ib

IdIc

Ia

Ib

Ic

Id

I / V Amp-Shapers

Last Dynode Timing Signal

I / V Fast Shaper

Data Acquisition and Digital Signal

Processing [2]

[2] J. D. Martinez, J. M. Benlloch, J. Cerdà, Ch. W. Lerche, N. Pavon, A. Sebastia. “High-speed data acquisition and digital signal Processing system for PET imaging techniques applied to mammography” IEEE Trans. Nucl. Sci., Volume 51, Issue 3, Part 1, June 2004 Page(s):407 - 412

Channel 1

Channel 2

[1] S. Siegel, R. W. Silverman, Y. Shao, and S. R. Cherry, “Simple charge division readouts for imaging scintillator arrays using a multi-channel pmt,”IEEE Trans. Nucl. Sci., vol. 3, no. 43, pp. 1634–1641.

Page 4: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

4ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Detector Front-EndNew system requirements and extended features.

Depth of Interaction Measurement.Individual Anode Gain Adjustment.Possible use of different types of Photomultipliers (SiPMs).Reduce Front-End delay and Increase timing and spatial resolution.

qContinuousScintillator

Crystal

Ia, Ib, Ic, IdData

Adquisitionand Digital

Signal Processing

Multianode PMT / SiPM

4

.. Σ VChn/64

DOI measurement [3]

Preamplifier Stage

1/2 1/4 ...1

Current Amp

Digital Gain ControlI2C Interface

Serial In

Ch1 Ch8

Ch56 Ch64

[3] C. W. Lerche, J. M. Benlloch, F. Sanchez, N. Pavon, B. Escat, E. N. Gimenez, M. Fernandez, I. Torres, M. Gimenez, A. Sebastia, and J. Martinez, “Depth of gamma-ray interaction within continuous crystals from the width of its scintillation light-distribution” IEEE Trans. Nucl. Sci., vol. 52, no. 3, pp. 560–572, 2005.

Mixed Signal ASIC

Anode Voltages

Page 5: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

5ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Anode Preamplifier (I)

Rz Cm

(W/L)1

(W/L)1/K

Vbias1

Vbias2

(W/L)2 (W/L)2/K

Vbias3

ANODE_IN

M1

M2

M3

M4

Vbias4

Gain Adjustment Mirror Array

Iin IA

IB

Iin = IB - IA

IB / K

IA / K Iin / K

Iin / K

(Iin / K)*{0~2}

Current-Current Feedback Amp

Page 6: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

6ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Anode Preamplifier (II)~200MHz *

* Including Cout PMT ≈ 10pF

Low Input Impedance increases BandwidthGain Adjustment compensates Anode Gain dispersion.Isolates PM from Resistor Network.

Resistor values can be designed to optimize VCh(increases DOI precision)

Reduction of capacitances in Resistor Network input nodes

Cpar = 60fF **** AMS 0.35um - Extracted layout

Page 7: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

7ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Anode Preamplifier (III)

Conection to PMT and SiPM

PMT Scheme

Iin

Iout

Page 8: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

8ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Anode Preamplifier (IV)

SiPM Scheme. Biasing.

Idiode

Iout

Page 9: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

9ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Impact Position Calculation

Centre of Gravity of total energy of gamma event: dcba

dcba

IIIIIIIIY

++++−+

=)()(

dbca

dbca

IIIIIIIIX

++++−+

=)()(

Sources of error in Impact Position Calculation

XXIIIIIIII

dbca

dbca ∆+=++++−+ )()(

Thermal Noise from Resistor Network

PreAmplifier Noise

Frequency Response of Resistor Network ( ) ( ) ( )23

22

21 XXXX ∆+∆+∆=∆

1X∆

2X∆

3X∆

Page 10: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

10ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Impact Position Calculation

. . .

Ia Ib

IdIc

R/10

R

R/10

R/10 R/10RR

R

R RR/10 R/10R/10 R/10

Error in Impact Position Calculation

PreAmplifier Noise: Very Low (~40nA) and affects uniformly all network inputs.

Thermal Noise from Resistor Network: Noise

Model• Instantaneous noise generated by in2

V in Ia and Ic is anticorrelated(same as in Ib and Id).

• Noise from upper and lower horizontal resistor chain is uncorrelated.

2nHi

2n. nVi

2nHi

⎥⎦

⎤⎢⎣

⎡++

+∂∂

+∂∂

+++∂∂

+++∂∂

=∆

ndbncadbca

ndbdb

ncaca

IIIIII

XII

X

IIII

XIIII

XX

).()(.)(

.)(

.2

)(.)(

)(.)(

22

22

2

( )22

2 .8

dcba

nH

IIIIiX

+++=∆

dbca

dbca

IIIIIIIIX

++++−+

=)()(

( ) 22 .2 nHnca iII =+

( ) 22 .2 nHndb iII =+

Page 11: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

11ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Impact Position Calculation

Error in Impact Position CalculationThermal Noise from Resistor Network. Conclusions

( )22

2 .8

dcba

nH

IIIIiX

+++=∆

( )22

2 4.

dcba

nV

IIIIinY

+++=∆

10).1(

..42

Rn

fTKinH

+

∆=

RnfTKinV ).1(

..42

+∆

=X∆ ≠ Y∆

n=8Do not depend on impact position1.

SiegelHigher R values do not reduce error because: (VchMAX < Vdd ) (Ia+Ib+Ic+Id) must decrease

2.

Examples:Vchmax=2.6V

R=6k6

BW=150MHz

∆X =2.5E-4

(0.12% of Pitch)

Vchmax=2.6V

R=1k

BW=150MHz

∆X =1E-4

(0.04% of Pitch)

( )R

VIIII chMAXdcba .K=+++

K factor depends on impact depth [3]

[3] C. W. Lerche, J. M. Benlloch, F. Sanchez, N. Pavon, B. Escat, E. N. Gimenez, M. Fernandez, I. Torres, M. Gimenez, A. Sebastia, and J. Martinez, “Depth of gamma-ray interaction within continuous crystals from the width of its scintillation light-distribution”IEEE Trans. Nucl. Sci., vol. 52, no. 3, pp. 560–572, 2005.

Page 12: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

12ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Impact Position CalculationError in Impact Position Calculation

Frequency Response of Resistor Network [4]

Current Signal Amplitudes are affected by Frequency Response of resistor network (Capacitances in input nodes).

Frequency response depends on input node position.

Horizontal chain resistors 10 times lower Horizontal resistor chains effects not included.

Currents in different input points will not sum in phase // Higher amplitude signals are close to centre of impact.

[4] A. Pullia, W. F. J. Muller, C. Boiano, and R. Bassini, “Resistive or capacitive charge-division readout for position-sensitive detectors” IEEETrans. Nucl. Sci., vol. 49, no. 6, pp. 3269–3277, 2002.

Page 13: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

13ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Impact Position CalculationFrequency Response of R. Network Read-Out.

Discrete Imp. Without PreAmplifierR=6k6

Cap=60fFASIC with PreAmplifier

R=1k

Cap=10pFHigherBandwidthReduced Position Dependency of Response

Linear Phase!!Position Dependent Time Delay

Page 14: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

14ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Impact Position CalculationError in Impact Position Calculation

Evaluation of error due to Frequency response of R. Network Read-Out.Gaussian Pulse (10n rise / 50n decay, LSO-PMT systems).Measure the position error in every input point of a vertical resistor chain.

Discrete Imp. Without PreAmplifier

1 2 3 4 5 6 7 8-15

-10

-5

0

5

10

15

Input Node

dX in

% o

f pitc

h

Time Delay > 15ns

∆X ≈ ±13% of pitch

Page 15: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

15ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Impact Position CalculationError in Impact Position Calculation

Evaluation of error due to Frequency response of R. Network Read-Out.Gaussian Pulse (10n rise / 50n decay, LSO-PMT systems).Measure the position error in every input point of a vertical resistor chain.

ASIC with PreAmplifier

1 2 3 4 5 6 7 8-0.2

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

Input Node

dX

(% o

f pitc

h)

0.8ns of max Time Delay

∆X ≈ ±0.17% of pitch

Page 16: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

16ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Depth of Interaction Measurement

Depth of Gamma-Ray Interaction within a continuous scintillator crystal is related to the width of the generated light-distribution [3].

The light-distribution width can be calculated with the sum of all anode voltages.

[3] C. W. Lerche, J. M. Benlloch, F. Sanchez, N. Pavon, B. Escat, E. N. Gimenez, M. Fernandez, I. Torres, M. Gimenez, A. Sebastia, and J. Martinez, “Depth of gamma-ray interaction within continuous crystals from the width of its scintillation light-distribution”IEEE Trans. Nucl. Sci., vol. 52, no. 3, pp. 560–572, 2005.

ΣVchn

Transconductance Amp.

Page 17: ASIC front-end for Position Sensitive SiPM and Vacuum PMT ...iworid-8.df.unipi.it/presentations/bosch.pdf · ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment

17ASIC front-end for Position Sensitive SiPM and Vacuum PMT with gain adjustment and depth of interaction measurement

Conclusions

PreAmplifier stage isolates read-out from PM.Gain Adjustment can equalize dispersion of PM characteristics.Frequency Response improvement in ASIC implementation reduces overall front-end delay.Implementation of ASIC front-end increases timing and spatial resolution.