Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
EVALUATING THE EFFICIENCY OF CYCLIC VOLTAMMETRY USING MODIFIED AND UNMODIFIED GLASSY CARBON ELECTRODES FOR THE ANALYSIS OF
SULFUR CONTAINING COMPOUNDS IN WATER
ASHLEY LYNN RYAN
A THESIS SUBMITTED IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE
DEGREE OF MASTER OF ENVIRONMENTAL SCIENCE
SCHOOL OF GRADUATE STUDIES NIPISSING UNIVERSITY NORTH BAY, ONTARIO
JULY, 2017
©ASHLEY LYNN RYAN, 2017
ii
Signature Page
iii
Abstract
When low oxidation state sulfur compounds such as sulfides dissolve in water, they create an
undesirable rotten egg smell. A prolonged exposure to concentrations of about 0.01 mg L1- of
H2S has been known to have negative effects on human gastro-intestinal, and nervous
systems. Concentrations of dissolved low oxidation state sulfur-based compounds in the
range of 0.01 to 0.5 mg L1- in aquatic systems have been shown to have chronic and acute
toxic effects in aquatic organisms. The ability to qualitatively and quantitatively determine
sulfur-based compounds in water, and further be able to speciate these compounds in water
would be of great significance. Sulfur-based compounds tend to disproportionate in aqueous
solutions to form compounds, the identity of which is dependent on the pH, temperature, and
ionic strength of the solution. The work that was investigated in this study involved the
detection of simple sulfides (H2S, HS- and S2-) in water through use of cyclic voltammetry
(CV), iodometric, and colorimetric techniques. In addition, the analysis of polysulfide and
polythionate compounds using these techniques were also investigated in this study. The
sodium salts of disulfide (Na2S2), trisulfide (Na2S3), and tetrasulfide (Na2S4) were used in
this study as representatives of polysulfides. These compounds were not available
commercially and were therefore synthesized. With regards to the CV technique, the bare
glassy carbon electrode (BGCE) and the vanadium oxide modified glassy carbon electrode
(MGCE) were investigated for their suitability to analyze the polysulfides and thiosulfate.
The results obtained in this work demonstrated a potential for use of the CV technique for the
qualitative analysis of the above class of compounds. It was however found necessary to first
reduce the above type of sulfur compounds to simple sulfide (S2-) before they could
successfully be analyzed using the iodometric and colorimetric techniques. Even though the
latter two techniques were successfully used for the quantitative analysis of sulfide, the CV
technique did not show much success. This was primarily as a result of the unpredictable
magnitudes of the various peak currents that resulted. In addition, the background currents
especially when the MGCE was used were found to interfere with the peaks associated with
sulfide. The colorimetric method and iodometric methods were found to have a great
agreement to each other (F (1,10) = 0.418, p = 0.532) for the determination for reduced sulfur-
based compounds.
iv
Keywords
Electrochemical Analysis, Sulfide, Polysulfide, Glassy Carbon Electrode, Vanadium Oxide,
Electrode Modification
v
Acknowledgments
I would like to thank my supervisor Dr. Stephen Kariuki for offering me the opportunity to
take on this project, and for his continued guidance and support throughout. All of his advice
and encouragement was greatly appreciated. I would also like to thank my committee
members, Dr. Mukund Jha and Dr. Samuel Mugo, for reviewing the enclosed research, for
their advice and Dr. Jeffery Dech for chairing my defense.
Exclusive thank you goes to Dr. Lesley Lovett-Doust and Dr. Jeffery Dech for their endless
candor, knowledge and guidance through the editing process. Thank you goes out to anyone
else who was brave enough to read this document. It was very much appreciated. I would
also like to thank Dr. Krys Chutko for collecting one of the water samples that was used to
test the methodologies outlined in this work.
Recognition also goes out to the Instrumental Laboratory in Lakehead University (LUIL) for
the work on the surface characterizations using SEM-EDX technology, and ICP analyses for
the water sample. To iCAMP though Canadore College for the preliminary exploration of
the electrode surfaces through SEM. Also, thanks to Willam Zhe from Laurentian University
for the SEM and EDX work done there as well.
I am thankful for my family, friends, and colleagues for their advice, guidance and continued
fortitude throughout this endeavor. I am especially thankful to my spouse, Tommi Jo, for
being an ever patient and supportive partner while I took the time to complete this work.
During the course of this project I spent a sleepless night watching a re-run of Back to the
Future. I am reminded of the antics of Dr. Emmett Brown (Christopher Lloyd) and can relate
a few quotes from the movies to this work. It was challenging to pick just one.
I end with this:
“You're not thinking fourth dimensionally!” – Dr. Emmet Brown
Cheers
vi
Table of Contents
Signature Page .................................................................................................................... ii
Abstract .............................................................................................................................. iii
Acknowledgments............................................................................................................... v
Table of Contents ............................................................................................................... vi
List of Tables ..................................................................................................................... ix
List of Figures .................................................................................................................... xi
List of Abbreviations and Symbols................................................................................... xv
Overview and Research Objectives .................................................................................... 1
Chapter 1 ........................................................................................................................... 14
Sulfide (HS-, H2S and S2-) ................................................................................................. 14
1.1 INTRODUCTION ................................................................................................ 14
1.2 MATERIALS AND METHODS .......................................................................... 15
1.2.1 Reagents and Solutions ............................................................................. 15
1.2.2 Standardization of 25 mM Thiosulfate and 25 mM Iodine Solutions ...... 17
1.2.3 Standardization of Sulfide Using Iodometric Titration ............................ 18
1.2.4 Electrochemical Analysis of Sulfide ......................................................... 18
1.3 RESULTS ............................................................................................................. 21
1.3.1 Standardization of Iodine and Thiosulfate Solutions ................................ 21
1.3.2 Standardization of Sulfide Solutions using Iodometric Titration ............. 21
1.3.3 Electrochemical Analysis of Sulfide ......................................................... 22
1.4 CONCLUSIONS................................................................................................... 49
Chapter 2 ........................................................................................................................... 54
Analysis of Polysulfides and Polythionates ...................................................................... 54
2.1 INTRODUCTION ................................................................................................ 54
vii
2.2 MATERIALS AND METHODS .......................................................................... 56
2.2.1 Making Na-polysulfides............................................................................ 56
2.2.2 N,N-Diethyl-p-phenylenediamine (DEPD) Analysis of Sulfide ............... 59
2.2.3 Reduction of Na-polysulfides and polythionates Using Chromium as the
Reducing Agent ........................................................................................ 59
2.3 RESULTS ............................................................................................................. 61
2.3.1 Making Na-polysulfides............................................................................ 61
2.3.2 Reduction and Analysis of Na-polysulfides and Polythionates ................ 63
2.4 CONCLUSIONS................................................................................................... 66
Chapter 3 ........................................................................................................................... 68
Electrochemical Analysis of Na-polysulfides and Polythionates ..................................... 68
3.1 INTRODUCTION ................................................................................................ 68
3.2 MATERIALS AND METHODS .......................................................................... 70
3.2.1 Equipment ................................................................................................. 70
3.2.2 Dissolution of Na-polysulfides and Thiosulfate Ions ............................... 70
3.3 RESULTS ............................................................................................................. 71
3.3.1 Electrochemical Analysis of Na-polysulfides and Polythionates at BGCE
and MGCE ................................................................................................ 71
3.4 CONCLUSIONS................................................................................................... 80
Chapter 4 ........................................................................................................................... 82
Analysis of Low Oxidation-State Sulfur Species in Water .............................................. 82
4.1 INTRODUCTION ................................................................................................ 82
4.2 MATERIALS AND METHODS .......................................................................... 83
4.2.1 Water Source and Collection Parameters ................................................. 83
4.2.2 Reduction and DEPD Analysis of Lake Water Sample ............................ 84
4.2.3 Electrochemical Analysis of Lake Water Samples ................................... 84
4.3 RESULTS ............................................................................................................. 85
viii
4.3.1 Reduction and DEPD Analysis of Lake Water Sample ............................ 85
4.3.2 Electrochemical Analysis of Lake Water ................................................. 90
4.4 CONCLUSIONS................................................................................................... 91
General Conclusions ......................................................................................................... 93
Appendix A ....................................................................................................................... 96
Appendix B ....................................................................................................................... 99
References ....................................................................................................................... 102
Curriculum Vitae: Ashley Ryan (nee Marcellus) ........................................................... 133
ix
List of Tables
Table 1. Standardized values (mM) for thiosulfate (S2O32-) and iodine (I2) used in iodometric
determination of sulfide. Values are the mean of n = 5. ......................................................... 21
Table 2. Mean potential positions and peak current (µA) values in the anodic segment for a 1
mM solution of HS- through the 2nd to the 10th CV cycle. ...................................................... 27
Table 3. Peak currents at the cathodic peak at -1208 mV for CV cycles 1 through 10 for a 1
mM HS- solution. .................................................................................................................... 29
Table 4. Mean of six replicate runs and ± SE of current responses (Ipc) for CV analysis of
sulfide obtained at the BGCE at the Epc of -1263 mV. ........................................................... 31
Table 5. Mean CV anodic currents (Ipa) for the sulfide analysis at the BGCE at potentials of
0, +565, and +1142 mV. ......................................................................................................... 31
Table 6. Peak current, potential and scan number for 0.36 mM HS- run over MGCE at a scan
rate of 100 mV s1-. .................................................................................................................. 46
Table 7. Mean ± SE for anodic current response Ipa (µA) of various concentrations of HS- at
MGCE; at a mean anodic potential Epa of +200 mV, (n = 6; R2 = 0.937). ............................. 48
Table 8. Mean ± SE for cathodic current response Ipc (µA) of various concentrations of HS- at
MGCE at a mean cathodic potential Epc of -570 mV, (n = 6; R2 = 0.907). ............................ 49
Table 9. Temperature profile for the preparation of Na-polysulfides (Rosen and Tegman,
1971). ...................................................................................................................................... 58
Table 10. Comparison mean % yields ± SE of reduced Na-polysulfides from two separate
batch products, summer 2015 and winter 2016. ..................................................................... 63
Table 11. Mean and SE ± for various concentrations of sulfide used to develop a standard
curve for DEPD analyses (n=11) ............................................................................................ 64
Table 12. Mean percent yields ± SE of the reduced polysulfides after applying heat at 60 °C
and reducing the solution for 4 hours (n = 5). ........................................................................ 66
x
Table 13. Weights (g) and concentrations (mM) of Na-Polysulfides used for CV analyses .. 71
Table 14. Anodic segment peak potential (mV) and peak current (µA) for CV analysis of Sn2-
ions (1 mM) at BGCE (n = 3). ................................................................................................ 73
Table 15. Cathodic segment peak potential (mV) and peak current (µA) for CV analysis of
Sn2- ions (1 mM) at BGCE (n = 3). ......................................................................................... 73
Table 16. Inorganic metal concentrations (ppm) from site IG9 for water sample collected in
October of 2015. ..................................................................................................................... 87
Table 17. Sediment concentrations (ppm), of the metals (and sulfur) associated with smelting
of nickel ores. Mean values (mg/kg ± standard error) (Chase et al. 2016, unpublished results).
................................................................................................................................................. 87
Table 18. Comparison of percent recoveries and recovered concentrations (mM) of sulfide
spiked water samples for filtered and unfiltered samples. ...................................................... 88
xi
List of Figures
Figure 1: Electrochemical cell setup for chemical analysis. (A) From left to right: Reference
electrode (Ag/AgCl) shaded grey, glassy carbon electrode (GCE) shaded black and platinum
auxiliary electrode (Pt) represented as light grey coil. (B) The BASi setup for
electrochemical analysis, following the same electrode order from left to right as is the case
in A.......................................................................................................................................... 19
Figure 2. Fraction dissociation plot for sulfide as a function of pH. Where indications of
sulfide species in solution are H2S(aq) (♦), HS- (■) and S2- (▲). ............................................. 23
Figure 3. Overlapped voltammograms showing five CV cycles of HS- (1 mM), with the
supporting electrolyte at BGCE with a scan rate of 100 mV s-1. ............................................ 25
Figure 4. Overlapped voltammograms showing 10 – 15 CV cycles of 1 mM HS- solution at
the BGCE with supporting electrolyte solution (dashed line), at scan rate of 100 mV s-1. .... 25
Figure 5. Epa positions (mV) at -6 (♦), +434 (▲), and +1158 (■) with peak current values
(Ipa) µA at CV cycles 2 and 5-10 for 1 mM HS-. .................................................................... 26
Figure 6. Epc position (mV) at -1208 with peak current values (Ipc) µA at CV cycles 2 and 5-
10 for 1 mM HS-. .................................................................................................................... 28
Figure 7. CV analysis for the various HS- concentrations at a mean Epc = -1263 mV at BGCE.
Error bars represent ± S.E. of six replicate runs (R2 = 0.988). ............................................... 30
Figure 8. CV anodic currents (Ipa) for various HS- concentrations at Epa = 0 mV (♦) R2 =
0.628, +565 mV (■) R2 = 0.926, and +1142 mV (▲) R2 = 0.897 on BGCE. Error bars
represent ± S.E. of six replicate runs. ..................................................................................... 32
Figure 9. BGCE SEM images A) is polished and electrochemically pretreated in 0.1 M
phosphate buffer, B) BGCE after 10 CV cycles of 1 mM HS-, in 0.1 M potassium phosphate
buffer, chevrons indicate some product build-up from sulfide electrochemical oxidation. ... 33
Figure 10. EDX spectra of BGCE pretreated by running CV scans of the BGCE in phosphate
buffer from -200 to +800 mV at 100 mV s1-; where C = carbon. ........................................... 34
xii
Figure 11. EDX spectra of BGCE after 10 CV cycles (-1600 to +1600 mV) in 1 mM HS-
solution made in 0.1 M phosphate buffer. Where C = carbon and O = oxygen. ................... 34
Figure 12. SEM of BGCE obtained after scanning 10 CV cycles (-1600 to +1600 mV) in a 1
mM HS- solution prepared in 0.1 M phosphate buffer. Chevron indicates a deposition onto
electrode surface. .................................................................................................................... 35
Figure 13. SEM images of BGCE (A) and MGCE (B), showing the surface characterizations
between the BGCE and MGCE. ............................................................................................. 36
Figure 14. The fifth cycle of a CV output of 0.36 mM concentration of HS- (solid line) at
MGCE, and the supporting electrolyte solution (dashed line) at a scan rate of 100 mV s1-. .. 38
Figure 15. SEM image of MGCE showing foreign deposits on the film modification after 10
CV cycles of 1 mM HS- at 100 mV s1- (A) and MGCE showing 10 CV cycles in the
supporting electrolyte solution at 100 mV s1- (B). .................................................................. 38
Figure 16: EDX spectra of MGCE after 10 CV cycles (-1600 to +1600 mV) of 1 mM HS-
solution made in 0.1 M phosphate buffer. Where C, O, S and Cu represent carbon, oxygen,
sulfur and copper respectively. ............................................................................................... 39
Figure 17: EDX spectra of MGCE after 10 CV (-1600 to +1600 mV) cycles in supporting
electrolyte solution (0.1 M phosphate buffer), where C = carbon. ......................................... 40
Figure 18. SEM image of MGCE (left), showing sites of interest for EDX along with
supporting EDX data expressed as weight % (right) for V2O5 modifier. ............................... 40
Figure 19. SEM image of MGCE after 10 CV cycles in 1 mM HS- solution, showing sites of
deposits on surface (left) and EDX data expressed as weight % (right) for MGCE after 10 CV
cycles in 1 mM HS- solution. .................................................................................................. 41
Figure 20. CV analysis of supporting electrolyte solution on MGCE from cycles 1 to 10, scan
rate 100 mV s1-; the numbers indicate the scan number of the CV cycle. .............................. 43
Figure 21. CV graph at MGCE of HS- solution (0.44 mM); scan rate of 100 mV s1- showing
cycles 2, 5, and 10 as well as the blank (dashed line). ............................................................ 44
xiii
Figure 22. Peak current values (Ipa1) µA at CV cycles 1 to 10 for 0.36 mM HS- (♦), compared
to CV cycle numbers 1 to 10, and compared to the supporting electrolyte solution (▲) on
MGCE. .................................................................................................................................... 45
Figure 23. Peak current values (Ipa2) µA at CV cycles 1 to 10 for 0.36 mM HS- (♦), compared
to CV cycle numbers 1 to 10, and compared to the supporting electrolyte solution (▲) on
MGCE. .................................................................................................................................... 45
Figure 24. Peak current values (Ipc) µA at CV cycles 1 to 10 for 0.36 mM HS- (♦), compared
to CV cycle numbers 1 to 10, and compared to the supporting electrolyte solution (▲) on
MGCE. .................................................................................................................................... 46
Figure 25. CV analysis increasing towards more positive potentials Ipa (µA) for various,
concentrations of HS- at Epa = +200 mV on MGCE after blank corrections. Error bars
represent ± S.E. of six replicate runs (R2 = 0.937). ................................................................ 48
Figure 26. CV analysis increasing towards more negative potentials Ipc (µA) for various
concentrations of HS- at Epc = -570 mV on MGCE after blank corrections. Error bars
represent ± S.E. of six replicate runs (R2 = 0.907). ................................................................ 49
Figure 27. Pictures of the synthesized Na-polysulfides: (A) Na2S2; (B) Na2S3; (C) Na2S4 .... 59
Figure 28. Purge and trap system set-up for the reduction of aqueous Na-polysulfides (Sn2-)
and thiosulfate (S2O32-): flask (A) (Sn
2- or S2O32-); flasks B and C (traps for reduced sulfide).
................................................................................................................................................. 61
Figure 29. Calibration curve of sulfide obtained using the DEPD method (R2 = 0.999). ...... 64
Figure 30. CV graph at a potential scan rate of 100 mV s-1 for S22- [1.39 mM], S3
2- [1.31 mM]
and S42- [1.22 mM] at BGCE. With supporting electrolyte solution as the blank (dashed
line). ........................................................................................................................................ 72
Figure 31. CV graph of the comparison of simple sulfide ion with di-, tri- and tetrasulfide
ions at a potential scan rate of 100 mV s-1 for HS- [1 mM], S22- [1.39 mM], S3
2- [1.31 mM]
and S42- [1.22 mM] at BGCE. With the supporting electrolyte solution as the blank (dashed
line). ........................................................................................................................................ 74
xiv
Figure 32. CV graph of the comparison of quantitative amounts of S2O32- at a potential scan
rate of 100 mV s-1 for S2O32- [1.06 mM], and 2 x the amount of S2O3
2- [2.12 mM] at BGCE.
With the supporting electrolyte solution as the blank (dashed line). ...................................... 75
Figure 33: CV graph of the comparison of polysulfide ions with thiosulfate ion at a scan rate
of 100 mV s-1 for S2O32- [1.06 mM] (dotted line), S2
2- [1.39 mM], S3
2- [1.31 mM] and S42-
[1.22 mM] at the BGCE. With supporting electrolyte solution as the blank (dashed line). .. 76
Figure 34. CV graph of the comparison of thiosulfate ion, sulfide ion and a combination of
thiosulfate ion and sulfide ion.at a potential scan rate of 100 mV s-1 for S2O32- [1.06 mM],
sulfide [1 mM], and mixture of HS- + S2O32- at BGCE. With the supporting electrolyte
solution as the blank (dashed line). ......................................................................................... 77
Figure 35. CV graph of the comparison of thiosulfate ion, trisulfide ion and a combination of
thiosulfate ion and trisulfide ion at a potential scan rate of 100 mV s-1 for S32- [1.31 mM] and
a mixture of S32-
[1.31 mM] and S2O32- [1.06 mM], and S2O3
2- [1.06 mM] at the BGCE. With
supporting electrolyte solution as the blank (dashed line). ..................................................... 78
Figure 36. CV graph at potential scan rate at 100 mV s1- at the MGCE for S22-, S3
2-, S42-,
S2O32-, and HS- (dotted line). With the blank as the supporting electrolyte solution (dashed
line). ........................................................................................................................................ 80
xv
List of Abbreviations and Symbols
Aux Auxillary Electrode
BASi Bioanalytical Systems Inc.
BGCE Bare Glassy Carbon Electrode
cm Centimeter
CTFE Chlorotrifluoroethylene
CDR Colour Developing Reagent
Cu Copper
DEPD N,N-Diethyl-p-phenylenediamine
EC Electrochemical Cell
GCE Glassy Carbon Electrode
g Grams
HDPE High-density polyethylene
Fe Iron
FeS Iron Sulfide
PbS Lead Sulfide
HgS Mercury Sulfide
μ Micro
μA Micro Amps
μM Micromolar
mM Millimolar
mL Milliters
mV Millivolts
MOE Ministry of Environment
MnS Manganese Sulfide
MGCE Modified Glassy Carbon Electrode
M Molar
N2 Nitrogen Gas
N Normal
I Peak Current
Ipa Peak Current Anodic (oxidation) μA
Ipc Peak Current Cathodic (reduction) μA
Epa Peak Potential Anodic (oxidation) mV
Epc Peak Potential Cathodic (reduction) mV
s1- Per second
Pt Platinum
MΩcm Resistivity measure unit MegOhms
centimeter
ν Scan rate (mV s1-)
Ag/AgCl Silver/Silver chloride
Ag2S Silver Sulfide
S Sulfur
V Volts
Zn Zinc
ZnS Zinc Sulfide
1
Overview and Research Objectives
Sulfur in the Environment
Interactions between the abiotic factors of the environment and the living organisms of
the biosphere are accompanied by a continuous cycling of matter in nature. Different
species of living organisms assimilate substances needed for their growth and to support
life. Living and non-living organisms emit by-products of metabolism, complex
minerals, and organic compounds of chemical elements in the form of non-assimilated
food or dead biomasses into the environment. This evolution to the biosphere formed a
stable connection of global biogeochemical cycles. With well-known and most important
biogeochemical cycles being: carbon, nitrogen, oxygen, phosphorous, sulfur and water.
Land and water ecosystems play an import role in the dynamics of biogeochemical
cycles, which have been agitated due to anthropogenic processes. The impact from these
anthropogenic processes has caused unpredicted changes to climate, increased
greenhouse gas production, decreases in biodiversity, progressive desertification, as well
as other factors (Krapivin and Varatsos, 2008; Ciais et al., 2014). Biogeochemical cycles
naturally aim at employing an equilibrium state in order to balance the cycling of the
elements between land and surface compartments for example carbon, nitrogen,
phosphorous, and sulfur. Impact of human activity to these biogeochemical cycles sees
most of the resources taken from nature. Those resources that were once taken from
nature are returned as waste products, which is more often than not poisonous or
unsustainable. These anthropogenic practices, thus, tip the equilibrium scales that should
exist between both the biosphere and humankind.
The nitrogen, sulfur and carbon biogeochemical cycles play important roles in the
regulation of many biological, chemical and geochemical processes. Carbon is an
essential element to life forms so too is nitrogen, phosphorus and sulfur. For example,
three of the essential Amino Acids (for humans) contain sulfur –these are cysteine,
cystine, and methionine (Canfield & Raiswell, 1999; Sievert et al., 2007). Amino acids
are proteins essential for functional metabolisms in the body. Cystine is formed by the
2
oxidation of two cysteine molecules. Cystine is found in the skeleton, hair, nails,
connective tissues, and is required to form glutathione. Other well-known sulfur
containing compounds in biological systems are the tripeptide glutathione, and many
important (protein) enzymes, coenzymes, vitamins, and hormones. Microorganisms can
use inorganic forms of sulfur compounds, such as sulfate, to process energy referred to as
assimilation (Andreae, 1990; Canfield & Raiswell, 1999; Sievert et al., 2007; Tang et al.,
2009; Knöller & Schubert, 2010). Due to an ever increasing impact from anthropogenic
activities such as the production and combustion of fossil fuels, this includes peat, coal,
oil and natural gas. These activities have increased the availability of sulfur compounds
in natural and managed ecosystems.
Numerous chemical and biological processes contribute to various transformations of
sulfur from one form to another through oxidation-reduction reactions. Sulfur based
compounds can occur in a variety of oxidation states, with -2 (sulfide), 0 (elemental
sulfur) and +6 (sulfate) being the most common (Kuhn et al., 1983; Luther III, 1985;
Andreae, 1990; Kariuki et al., 2001; Keller-Lehmann et al., 2006; Sievert et al., 2007;
Tang et al., 2009). Of the sulfur-based compounds, the more stable form sulfate can
function as an electron acceptor in metabolic pathways to be utilized by microorganisms,
and can be converted to sulfide (Sievert et al., 2007; Tang et al., 2009; Fazzini et al.,
2013). On the other hand, reduced sulfur compounds, like sulfides, can serve as electron
donors converting those compounds to elemental sulfur and sulfate (Friedrich et al.,
2001; Rohwerder and Sand, 2007; Tang et al., 2009). Some microorganisms that utilize
this type of conversion are sulfur reducing bacteria. Sulfur reducing bacteria represents a
diverse group of anaerobes and aerobes thriving in both oxic and anoxic environments
The biochemical reactions involved in the oxidation and reduction of sulfate to sulfide
and vice versa are elaborate. Different groups of sulfur reducing bacteria utilize these
pathways in different ways, depending on their environmental conditions, to suit their
needs. According to authors Friedrich et al. (2001), Friedrich et al. (2005), Rohwerder
and Sand (2007), Mohapatra et al. (2008), Ghosh and Dam (2009), and Tang et al.,
(2009), the metabolic processes of sulfur reducing bacteria have been extensively
reviewed. In brief, sulfate reduction occurs though two pathways, assimilatory and
3
dissimilatory. Both pathways require activation of sulfate by adenosine triphosphate
(ATP). The attachment of sulfate to ATP results in the formation of adenosine
phosphosulfate (APS) then is catalyzed by the enzyme ATP sulphurylase. The sulfate
portion of the APS is further reduced to sulfite by the enzyme APS reductase. In the
assimilatory pathway sulfate is used to generate reduced sulfur compounds, using organic
compounds, for the synthesis of amino acids and proteins. The assimilatory pathway
does not excrete sulfide directly. Rather, the developed sulfide is incorporated into
organic sulfur compounds, such as dimethyl sulfide and dimethyl sulfoxide. In the
dissimilatory pathway the reduction of sulfate or sulfur is converted to inorganic sulfide,
such as hydrogen sulfide, by available anaerobic sulfate and the resulting sulfide
produced is thus excreted into the surrounding environment.
Some sulfur-based compounds which are of particular interest include: hydrogen sulfide
(H2S), hydrosulfide (HS-), sulfide (S2-), elemental sulfur (S0), thiols (R-SH), sulfate
(SO42-), sulfite (SO3
2-), polythionates (S2On2-) and polysulfides (Sn
2-) (Ciglenečki &
Ćosovíc, 1997). In anoxic environments, there may be an accumulation of reduced sulfur
species such as, hydrogen sulfide, metal sulfides, polysulfides, thiosulfates and elemental
sulfur (Brouwer and Murphy, 1995). Sulfur species formed depend largely on
temperature, pH, ionic strength, the source of the sulfur species and oxidizing agent
present (Boulegue et al., 1982; Koh, 1990; Petre and Larachi, 2006; Kaasalainen and
Stefánsson, 2011; Pan et al., 2013). As outlined by Holmer and Storkholm (2001) and
Mohapatra et al. (2009), the oxidation metabolism of inorganic sulfur compounds starts
with sulfide oxidase which catalyzes the oxidation of hydrogen sulfide to elemental
sulfur. Sulfur oxidizing enzymes oxidize elemental sulfur to sulfite. Further oxidation of
sulfite to sulfate is catalyzed by sulfite oxidase. During the oxidation of sulfide to the
various sulfur compounds, sulfite and elemental sulfur can react chemically together to
form thiosulfate. Inorganic forms of sulfur-based compounds can be found in oxic,
anoxic, fresh and marine aquatic systems (Al-Farawati & van den Berg, 1999; Dutta et
al., 2010).
Sulfur based compounds are released naturally from living organisms, some examples
include metabolic excretions from bacteria, decomposition of organic plant matter under
4
anaerobic conditions, and discharge from agriculture livestock. Several anthropogenic
processes such as wastewaters, petroleum refineries, burning of fossil fuels, and paper
mills also produce sulfur-based compounds. Releases of high concentrations of reduced
sulfur compounds, like sulfide, can seriously disrupt ecosystems by producing elevated
levels of sulfur-based contamination downstream (Witter and Jones, 1997; Ateya et al.,
2007). Brouwer and Murphy (1995), reported that hydrogen sulfide in concentrations
down to 0.015 mg L1- cause chronic toxicity to aquatic organisms, with levels of 0.5 mg
L1- exhibiting acute toxicity to rainbow trout, and salmon in a Spanish fish farm (Ortiz et
al., 1993). Hydrogen sulfide may be produced from the decomposition of organic matter
underground, such as decaying plant material, or by chemical reduction of sulfate by
sulfate-reducing bacteria. Hydrogen sulfide can be found in deep or shallow wells. It is
often present in areas underlain by shale or sandstone, near coal or peat deposits, and near
oil fields. Hydrogen sulfide can also occur naturally in groundwater systems. Some of
the sources of hydrogen sulfide in ground water are located near farm areas and swamps
in Ontario, as well as areas in Nova Scotia where there are sulfide bearing mineral
deposits and out west through Saskatchewan in the gas fields (Edwards et al., 2011). The
level of contamination from the hydrogen sulfide, in the ground water, depends on the
level of the water table. Dissolved sulfide in sewage and paper mill waste water effluent
were reported to be released at concentrations of 47 mg L1- and 51 mg L1- respectively
(Dutta et al., 2008 and 2010). Another example of discharge into the environment is the
effluent from the un-hairing process that occurs during tanning of animal hides.
Discharges from tanning process effluents have reported sulfide concentrations in the
range of 700 – 2000 mg L1- (Font et al., 1996). The resulting untreated waste water, from
tanneries, has also been reported to have dissolved sulfide concentrations of 20 mg L1-
(Font et al., 1996).
Development of reliable methods for the detection of various sulfur-based compounds
has become an important goal for analytical and environmental chemists (Lawrence et
al., 2000; O’Reilly et al., 2001; Khudaish & Al-Hinai 2006; Dutta et al., 2009; Titova et
al., 2009; Paim and Straditto 2010). The importance of the detection of sulfur-based
compounds has increased due to the increased amount of occupational exposure to
hydrogen sulfide. Sulfur content in crude oils is 0.3 to 0.8 wt. % and the hydrogen
5
sulfide content in natural gas ranges from 0.01 to 30 wt. % (Lawrence et al., 2000(a)).
Further increases to the extraction of oil and natural gas products to meet future supply
and demands will cause further increases to hydrogen sulfide concentrations in both oil
and natural gas. Clinical cases involving sulfide poisoning can range from 0.03 to 1 mg
L1-; with lethal doses ranging from 0.3 to 3 mg L1- (Lawrence et al., 2000(a); Ardelean et
al., 2014). As a weak acid, hydrogen sulfide has corrosive properties that can lead to
damage to infrastructure in municipal and city sewer systems. Zhang et al. (2008)
reported that at levels of 0.1 – 0.5 mg L1- of sulfide content in waste water will have
minor effects to infrastructure. However, at sulfide concentrations of > 2.0 mg L1-, severe
corrosion problems are prominent. Sulfide present in waste waters from sewer systems,
combined with the enclosed environment related to sewer systems, the exposure to levels
of sulfide that induce corrosion to infrastructure pose a risk of toxicity to sewer system
workers. Several authors such as Lawrence et al. (2000a), Lawrence et al., (2000b),
Lawrence et al. (2004), Ateya et al. (2007), Lawrence et al. (2007), Titova et al. (2009),
Edwards et al. (2011), Pikaar et al. (2011), and Hu & Mutus (2013), have reported some
clinical cases that have arisen from the exposure to sulfide. These authors have also
reported toxic and chronic effects with continued exposure. Exposures to sulfide, even at
concentrations as low as 0.01 mg L1-, have been reported to have neurotoxic effects as
well as causing chronic side effects to the skin, eyes, circulatory system, digestive and
respiratory tracts (Lawrence et al., 2000(a); Lawrence et al., 2000(b); Khudaish & Al-
Hinai 2006; Hughes et al., 2009; Titova et al., 2009; Hu & Mutus, 2013). The risks
associated with sulfide, not only to environments that are close to wastewater effluents
but also to surrounding occupational hazards mean that, continued investigation is
essential to the understanding of the environmental processes undergone by sulfur-based
compounds.
Current Detection Methods for Sulfur-Based Compounds
Sulfide analyses are well represented across a variety of analytical methods. The sulfide
anion can be rather versatile, in that, the anion can form other intermediate species of
sulfur based compounds to which they can be determined by an assortment of analytical
methods. A few of those intermediate sulfur based compounds that can form from the
6
sulfide anion are thiosulfate, sulfite, sulfate, and acid volatile sulfides. Some major
analytical methods for the detection of the sulfide anion, and other sulfur-species include
titration (Pawlak & Pawlak, 1999; Ciesielski & Zakrzewski, 2006), spectroscopy (Silva et
al., 2001; Kariuki et al., 2008), liquid chromatography (O’Reilly et al., 2001; Kamyshny
et al., 2006), ion chromatography (Miura et al., 2005; Stefansson et al., 2007), gas
chromatography (Wardencki, 1998; Kristiana et al., 2010) and electrochemistry (García-
Calzada et al., 1999; Rozan et al., 2000b; Giovanelli et al., 2003; Cheng et al., 2005;
Manova et al., 2007; Titova et al., 2009; Piam & Stradiotto, 2010; Manan et al., 2011;
Huang et al., 2012). For the detection of simple sulfide, the methodologies vary over a
broad instrumental platform, with each method posing its own set of benefits and
limitations.
In a review published by Lawrence et al. (2000a), the authors have presented methods for
analysis of the sulfur-species and their respective limitations. Iodometric titration is
reported to be reliable for the standardization of pure, laboratory-generated sulfide
samples for calibration since the overall concentration of dissolved total sulfide is ≥ 1.0
mg L1-. The authors suggest that titrimetric methods have significant limitations in terms
of sensitivity and selectivity when used to analyze environmental samples. In the
presence of interfering ions like thiosulfate, sulfite, and organic compounds, both
dissolved and solid iodometric titration methods can be hindered (Clesceri et al., 1998).
These interferences cause a reduction in sensitivity towards the overall detection of the
sulfide ion in environmental samples, due to the formation of side reactions whose
products have detection limits that range from 0.2 to 4.0 mg L1- (Lawrence et al., 2000a).
Spectroscopic Determinations of Sulfur Based Compounds
There are a variety of spectroscopic detection methods known for analyzing sulfides.
The most common analysis quantifies the concentration of sulfide with the Methylene
Blue complex, detected via the UV/visible spectrum. This particular reaction involves
aqueous samples of sulfide to react with N,N-dimethyl-p-phenylenediamine, through
oxidative coupling in the presence of ferric ions under acidic conditions, which forms a
pentacyclic phenothiazine dye, blue coloration referred to as methylene blue that has an
7
absorbance maximum at 670 nm (see Scheme 1) (Cline, 1969; Spaziani et al., 1997;
Lawrence et al., 2000a; Lawrence et al., 2000c; Silva et al., 2001; Kariuki et al., 2008).
The methylene blue method is reported to be consistently sensitive, selective and simple.
For example, wastewater samples analyzed for sulfide using methylene blue have upper
and lower detection limits of 3.2 x 10-2 and1.0 x 10-4 mg L1- respectively (Lawrence et
al., 2000a). Although, the Methylene Blue approach has been popular, the sample
turbidity and the formation of H2S when S2- is in contact with oxygen tend to limit the
accurate evaluation of sulfide in aqueous media (Lawrence et al., 2000c). Also, reducing
agents cause interference with the methylene blue reaction, for example, thiosulfate in
concentrations of 10 mg L1- may prevent or impair the formation of the blue coloration
(Clesceri et al., 1998). When analyzing natural water samples with a complex matrix,
combined with low sulfide concentrations, interferences from dissolved organic material
can occur. The absorbance shoulder of dissolved organic material can spread over the
wavelength of 668 nm, could falsely increase the methylene blue signal (Tang &
Santschi, 2000). Other chromogens that have been mentioned to be analogous to
methylene blue, towards the detection of sulfide, are: resazurin, aszurea, thionine and
toluidine blue (Lawrence et al., 2000a). Other spectroscopic methods, use the
fluorimetric protocols with 2,7-dichlorofluorescien. This method suffers from problems
of poor selectivity in the presence of thiol (RSH) compounds, since the latter reacts in a
similar way as the sulfide anion does (Lawrence et al., 2000a).
Scheme1
Chromatographic Determinations of Sulfur Based Compounds
Chromatographic separations, like those of capillary electrophoresis (Font et al., 1996;
Petre & Larachi, 2006), gas (Wardencki, 1998; Kristiana et al., 2010), liquid (Tang &
Santschi, 2000; Pan et al., 2011), and ion chromatography (Casella et al., 2000; Ohira &
Toda, 2006) hold advantages for complex media. Chromatographic methods can offer
8
better resolution and separation from other sulfur based compounds present in the
sample. For capillary electrophoresis, earlier work by Font et al. (1996), observed good
linearity for sulfide in concentration ranges from 0.5 to 10 mg L1- at pH of 10.5, using
internal and external calibration standards. The authors reported a limit of detection for
sulfide of 0.2 mg L1-. They also observed that thiosulfate, nitrate, and nitrite could be
resolved using this technique and these compounds did not cause any interferences with
the resolution of the sulfide anion. However, in work performed by Petre and Larachi
(2006), the authors used capillary electrophoresis to resolve inorganic sulfur bearing
species. These species included sulfate, sulfite, polythionates, polysulfides, and sulfide
anions. Over a pH range of 8.2 to 12.2 that the separations were tested, a pH of 9.5 was
found suitable for the separation of thiosulfate, sulfate, sulfide, tetrathionate and sulfite at
a sulfide concentration of 6 mg L1- and the other sulfur species at concentrations of 10 mg
L1-. They also observed that polysulfide distribution depended largely on pH. In addition,
they noted that the polysulfide ions were unstable over the pH range of 8.2 to 12.2. A
complete quantitative determination of these ions was not possible with this method.
Separations and quantification of sulfide is possible with capillary electrophoresis, for
examining environmental effluents from leather, paper processing and mining water with
detection ranges from 5.0 x 10-4 mg L1- to 20 mg L-1 (Lawrence et al., 2000a). However,
some sulfur containing compounds like polysulfides appear more challenging to separate
and quantify due to their rapid dissociation with changes in pH of the solution being
analyzed. With liquid and gas chromatography separation analyses, these techniques are
susceptible to common interferences such as hydroxyl ions and metal ions that may be
present in the sample. Also, many liquid and gas chromatography separations employ
much of the emphasis on the organic forms of sulfide, such as dimethyl sulfide, rather
than the inorganic forms of sulfide (Wardencki, 1998; Kristiana et al., 2010). However,
there has been success in separating sulfide, sulfite, sulfate and thiosulfate anions using
ion-pair chromatography. This technique requires the conversion of sulfide to
thiocyanate and sulfite to sulfate in order to be analyzed. In concentrations of up to 10 -
mM, the common anions listed above, did not interfere with the determination of sulfur
anions. The recoveries of these anions from hot-spring waters were reported to be 99.5 to
101.3 % respectively (Miura & Kawaoi, 2000; Miura et al., 2005). Pre-column
9
derivatizations based on methylene blue and florescent labeled monobromoimane have
been utilized towards the detection of sulfide anion through liquid chromatography. Ion
chromatography, using gel sorbents or solid lead (II) based chromate columns suffer
impairment from hydroxyl ions as well as metal ions that could be found in complex
sample matrices, such as, calcium, magnesium and iron (Lawrence et al., 2000a). The
technique also lacks selectivity with regards to other redox species of sulfur anions.
Electrochemical Determinations of Sulfur Based Compounds
Electrochemical detection methods have become popular in the analysis of sulfur
compounds (Ciglenečki & Ćosović, 1997; García-Calzada et al., 1999; Lawrence et al.,
2000b; Chadwell et al., 2001; Lawrence et al., 2002; Diligin et al., 2012; Huang et al.,
2012; Hu & Mutus, 2013). The responses of these methods are based on the oxidation or
reduction of an electroactive species, like those of sulfur-based species. Unlike some
other analytical methods, the electrochemical based methods can provide information
about reaction kinetics, chemical behavior of the electroactive species in solution, and
information regarding possible adsorbed products to the surface of the working electrode
(Kissinger and Heineman, 1996; Scholz, 2010; Zanello et al., 2012). The
electrochemical cell contains a reference (saturated calomel (SCE) or silver/silver
chloride (Ag/AgCl)), an auxiliary, and a working electrode. Working electrodes like
those of inert metals such as, gold, silver, and platinum offer favorable electron transfer
kinetics and a wide potential range (Kissinger and Heineman, 1996; Scholz, 2010;
Zanello et al., 2012; Li and Miao, 2013). However, the cathodic potential is restricted
due to the low hydrogen overvoltage, which forms surface oxides and hydrogen layers
leading to high background currents affecting the kinetics of the electroactive species at
the working electrode. Mercury electrodes offer high hydrogen overvoltage and an
extended cathodic potential window (Kissinger and Heineman, 1996). The most
commonly used mercury electrode is the hanging mercury drop electrode, which possess
a highly reproducible, renewable, smooth surface (Kissinger and Heineman, 1996;
Umiker et al., 2002; Scholz, 2010; Li and Miao, 2013). The hanging mercury drop
electrode does not require cleaning before an experiment since it has a self-renewing
surface, but the limited anodic range and toxicity around handling and disposal has
10
researchers looking towards alternative working electrode materials that have comparable
chemical function to that of mercury. (Li and Miao, 2013). Carbon based electrodes have
soft surface properties, in other words, the surface can be easily polished compared to
metal electrodes like silver and platinum. Carbon surfaces can be easily renewed for
electron exchange during polishing and cleaning procedures (Kissinger and Heineman,
1996; Scholz, 2010; Zanello et al., 2012). Carbon electrodes have a complex underlying
microstructure, broad potential window, low background current, which allows for the
formation of a wider variety of surface bonds and functional groups (McCreery, 2008;
Scholz, 2010; Zanello et al., 2012). The cost of carbon based electrodes is quite low
compared to that of traditional metal electrodes, which makes them a favorable material
for electrochemical analyses (Li and Miao, 2013). Common carbon based electrodes
include pyrolytic graphite, glassy carbon, carbon paste, carbon-fiber, nanotubes, and
carbon composite electrodes (Kissinger and Heineman, 1996; Scholz, 2010; Zanello et
al., 2012; Li and Miao, 2013).
Electrochemical techniques include polarography (Umiker et al., 2002), cathodic
stripping voltammetry (Ciglenečki & Ćosović, 1997) anodic stripping voltammetry
(Huang et al., 2012), and cyclic voltammetry (Lawrence et al., 2004; Lawrence et al.,
2007). Electrochemical methods can provide fast, sensitive detection while probing
reaction mechanisms of electroactive species. Differential pulse polarography has been
used for many years as a direct method for sulfur compound speciation (Rozan et al.,
2000). When compared to direct current and normal pulse polarography, differential
pulse polarography better resolves multi component systems and exhibits closely spaced
half-wave potentials especially when both oxidized and reduced species of the redox
couple are present (Umiker et al., 2002). When utilizing the hanging mercury drop
electrode for sulfide analysis, the cathodic reaction at the mercury electrode (scheme 2)
shows that sulfide oxidizes the mercury through the formation of mercury sulfide.
Similar sulfur based species like those of sulfite, thiols, polythionates, and polysulfides
have a similar reaction with mercury (Rozan et al., 2000; Umiker et al., 2002). Umiker
et al (2002) conducted differential pulse polarography to quantify sulfur species in soil
samples. The authors reported that the concentrations for sulfite, thiosulfate, cysteine and
sulfide were 0.580, 3.16, 1.58, and 0.296 µM, respectively.
11
𝐻𝑆− + 𝐻𝑔 ↔ 𝐻𝑔𝑆 + 𝐻+ + 2𝑒−
Scheme 2
Cathodic striping voltammetry can also be used to determine sulfur species in the
presence of sulfide. In a study done by Ciglenečki & Ćosović (1997), the authors
measured sulfur based compounds in anoxic sea water using the mechanism outlined in
scheme 2, but utilizing the reverse reaction when the potentials are moved to more
negative regions and the resulting current of mercury (II) to elemental mercury is
measured. The authors report that sulfide and elemental sulfur concentrations were 42
µM, with thiosulfate concentrations being 8 µM in sea water samples. They also reported
that sulfide and thiosulfate had good linearity in the range of 0.01 to 1 µM and 1 to 100
µM respectively. Huang et al (2012) investigated the use of bismuth film modified
electrodes towards the detection of sulfide in water samples using anodic stripping
voltammetry. The use of anodic stripping voltammetry is based on the selective reaction
between cadmium (Cd2+) and sulfide to form cadmium sulfide precipitate and then
stripping the Cd2+ from the formed sulfide deposit. The process leads to a current which
is proportional to the Cd2+ concertation in the sample. Bismuth, as an electrode modifier,
has been reported towards electrochemical detection of metal ions (Wang, 2005). Huang
and others go on to report that bismuth films have the same advantages of mercury, but
with less toxicity. In their study they compared the anodic stripping voltammetric
method to the classic methylene blue reaction. The initial analysis of the water samples
yielded no detectable limits for either method, but upon spiking the water samples with
sulfide, recoveries of 101 ± 0.583 µM L1- for the anodic stripping method resulted. The
methylene blue reaction showed no detectable levels of sulfide, even after spiking except
in two of the ten water samples they analyzed. Good linearity and reproducibility was
also reported towards the detection of sulfide, as well as little to no matrix interferences.
Lawrence et al. (2007) used cyclic voltammetry and various carbon based substrates,
more specifically edge-plane pyrolytic graphite, to examine the direct oxidation of sulfide
in a river water sample. The authors compared the techniques of cyclic and square wave
voltammetry, reporting a linear range of sulfide being detected 5 – 6 µM and 10 – 60 µM
12
respectively. Lawrence et al. (2007) chose to use cyclic voltammetry due to its lower
linear range for detectable sulfides. These authors further report recoveries of sulfide
spiked river water of 104 %. The authors also noted that the approach they used would
suffer when determining sulfide in matrices that contain oxidizable electroactive species.
Another group from Lawrence et al. (2004) used carbon nanotubes to modify glassy
carbon electrodes for the detection of sulfide. Although these authors did not apply their
modifications to environmental samples, they had good linear range for sulfide using bare
glassy carbon and modified glassy carbon with carbon nanotube electrodes, in the range
of 12.5 – 50 µM and 1.25 – 112.5 µM respectively. Many of the techniques summarized
here show adequate selectivity with some sensitivity, and vice versa towards the
determination of sulfide in various matrices, with sample handling and clean up measures
being required in most cases. To increase the comprehensive assessment of detecting
sulfide and other sulfur based compounds it appears that combining the various analytical
techniques, could perhaps enhance the selectivity and sensitivity of detection of sulfur
based compounds.
This work will attempt to examine if electrochemical analyses, such as cyclic
voltammetry, with the use of bare and modified glassy carbon electrodes will be efficient
towards possible quantification of lab-generated sulfide samples. Chemically modified
electrodes function to immobilize molecules with specific functions on the electrode
surface by either physical or chemical means (Kissinger and Heineman, 1996; Li &
Miao, 2013). The use of glassy carbon electrodes is favored more over most traditional
metal electrodes like gold, silver, platinum (Kapusta et al., 1983; Mohtadi et al., 2005) or
mercury (Umiker et al., 2002). This study will apply cyclic voltammetry using
chemically modified and unmodified glassy carbon electrodes against lab-generated
sulfide, along with polysulfides, namely, disulfide, trisulfide and tetrasulfide. Also, a
representative of the polythionates, such as thiosulfate, will be examined to explore
whether these compounds can be distinguished individually and as part of a mixture with
simple sulfide. The use of cyclic voltammetry will also be compared to some of the
classical methodological approaches, for example, iodometric titrations and ethylene
blue, analogous to methylene blue, for its efficiency in the determination of sulfides,
polysulfides and polythionates. Lastly, this work will be applied to a natural water
13
sample in order to test cyclic voltammetry, iodometric titration and ethylene blue
reactions against environmentally relevant concentrations of sulfur-based compounds
such as the ones mentioned above.
14
Chapter 1
Sulfide (HS-, H2S and S2-)
1.1 INTRODUCTION
Electrochemical techniques measure the oxidation and the reduction signals of the sulfur
compounds at the surface of a working electrode. Modifying the working electrode
surface with a metal oxide, such as vanadium oxide (V2O5), has been shown to enhance
the detection of inorganic sulfide species like those of HS-, S2-, and H2S (Park et al.,
1998; Park et al., 2002; D’Elia et al., 2004; Khudaish and Al-Hinai, 2006). Khudaish
and Al-Hinai (2006) have outlined the electrochemical deposition of vanadium oxide
films on glassy carbon electrodes according to a procedure described by D’Elia et al.
(2004). Vanadium compounds are not soluble in water and lead to a slow
electrochemical response. Once they are deposited on the electrode surface the catalytic
activity, towards the complexation with the sulfide ion in solution, are increased and can
be used in aqueous solutions (Li et al., 1996; D’Elia et al., 2004; Khudauish and Al-
Haini, 2006; Salimi et al., 2006). The V2O5 film deposition was realized through the
electrochemical oxidation of the vanadyl species, VO2+, that was obtained by scanning
potentials of a solution containing VO2+ from 0 to + 2000 mV versus Ag/AgCl. The
suggested reaction which occurs during this process is shown in Scheme 3 (Khudaish and
Al-Hinai, 2006). Other metal oxide catalysts for the oxidation of hydrogen sulfide have
been explored, such as, bismuth-molybdenum oxide (Li and Cheng, 1966), iron-
antimonate and iron-tin (Li et al., 1997); aluminum oxide, titanium dioxide, and iron (III)
oxide (Park et al., 1998). The authors that indicated that these metal oxides can oxidize
hydrogen sulfide did not propose any mechanisms details for the oxidation process.
However, Li and Cheng (1997) have described a two-step mechanism towards the
oxidation of hydrogen sulfide to elemental sulfur using mixed oxides of bismuth and
molybdate. These authors describe that, first, hydrogen sulfide reacts with the oxygen of
the metal oxide catalyst to form elemental sulfur, and the metal oxide is then partially
reduced. Subsequently, the partially reduced metal oxide is re-oxidized by the oxygen
present in the reaction mixture (Scheme 3).
15
2𝑉𝑂2+ + 3𝐻2𝑂 → 𝑉2𝑂5 + 6𝐻+ + 2𝑒−
Scheme 3
Voltammetric measurements can provide insight into chemical mechanisms in regards to
reaction mechanisms, kinetics, electron transfer, reversibility or irreversibility of a
reaction, and the behavior of a species in solution (Batchelor-McAuley et al., 2015). The
cyclic voltammetric technique allows for fast and easy accumulation of data. This
technique employs the use of various substrates for a working electrode, with the most
common being metal based electrodes (gold, silver, platinum or mercury). With the
enhanced awareness of environmental protection and health concerns towards human
exposure, the applications of mercury electrodes ought to be reduced due to their high
toxicities (Huang et al., 2012). Carbon based electrodes such as glassy carbon, carbon
paste, carbon fiber, or carbon composite electrodes are more advantageous over
conventional metallic based electrodes. The working surfaces of glassy carbon electrodes
are readily recharged by mechanical polishing, and can be modified by applying
electrochemical pretreatments (Wang & Hutchins, 1985; Pocard et al., 1992; McCreery,
2008). This chapter outlines the electrochemical analysis of sulfide using cyclic
voltammetry, with the bare glassy carbon and also with the modified glassy carbon
working electrodes.
1.2 MATERIALS AND METHODS
1.2.1 Reagents and Solutions
A 0.1 M potassium phosphate buffer solution (pH = 10.3) was prepared and used to make
the sulfide stock solutions. This buffer solution also served as a supporting electrolyte
for all the electrochemical experiments that were carried out in this work. Once the
sulfide was dissolved in the buffer solution, the sulfide stock solution (1.2.1.1) was
standardized against 25 mM standardized iodine and thiosulfate solutions. The
standardization of the sulfide stock solution was required since the sulfide solution can
easily be oxidized when exposed to air. Therefore, to ensure the concentrations of sulfide
16
used during the electrochemical and colorimetric experiments were as accurate as
possible, iodometric titrations were conducted. The procedures about how the solutions
used for the iodometric titration and for the phosphate buffer preparation are summarized
in Appendix A. These include the preparation of 6 M hydrochloric acid, 25 mM sodium
thiosulfate, 2 mM potassium bi-iodate, 25 mM iodine and 2 % (w/v) starch solution.
1.2.1.1 Stock Sulfide Solution
Working in a Polymer Series 100, Baxter Glove Box, equipped with a dual purge
nitrogen flow control, as well as a humidity and temperature control capability (Terra
Universal Inc.), a sulfide stock solution of 106 ± 1 mM (n = 9) was prepared by weighing
and dissolving 0.158 g of anhydrous sodium sulfide with a purity of ≥ 90.0 % (Acros
Organics, Fisher Scientific) into 20-mL of pre-purged, 0.1 M phosphate buffer with a pH
of 10.3, contained in a 20 mL, 28 x 61 mm borosilicate glass scintillation vial with
polyethylene caps (Fisherbrand, Fisher Scientific). The solution was well sealed and
stored at a temperature of 4 ºC, until it was required for analysis.
1.2.1.2 Vanadium Tetroxide (V2O4) Solution
A 0.01 M of V2O4 was prepared in a 250-mL volumetric flask by dissolving 0.415 g of
V2O4 (99.9 %, SIGMA) in 1.0 M phosphoric acid (H3PO4) (85 %, Fisher Scientific). The
V2O4 solution is reported to have been used for the modification of a glassy carbon
electrode surface (D’Elia et al., 2004; Khudaish and Al-Hinai, 2006). These authors
report that they used the modified glassy carbon electrode for the electrochemical
detection of sulfide. The vanadyl species present in solution. According to formal
reduction potentials (Harris, 2010), is VO2+.
In the present study, the VO2+ solution was to be used for the modification of the glassy
carbon electrode surface for the electrochemical analysis of not only simple sulfide, but
also of other sulfur species. The process used for the electrochemical deposition of the
V2O5 film is described in section 1.2.4.4.
17
1.2.2 Standardization of 25 mM Thiosulfate and 25 mM Iodine
Solutions
1.2.2.1 Standardization of Thiosulfate Solution against Bi-Iodate Solution
Sodium thiosulfate (Na2S2O3) is a common secondary standard which therefore, requires
standardization through the use of a primary standard. A 25 mM S2O32-
solution,
prepared as described in Appendix A, was standardized with a solution of potassium bi-
iodate (KIO3), also prepared as outlined in Appendix A. In the standardization of S2O32-,
a weighed amount of KIO3 was reacted with excess potassium iodide (KI) under acidic
conditions. As shown in reaction R – 1.1 (Appendix B), the reaction releases iodine (I2)
which then reacts with the S2O32-
solution being standardized (Appendix B, R – 1.2). The
summary of how the standardization was carried out is as follows. A 20-mL aliquot of 2-
mM KIO3 solution, prepared as shown in Appendix A, was pipetted into a 150-mL
solution containing 2 g KI. The resulting solution was acidified with 2 drops of
concentrated sulfuric acid (ACS Grade, Caledon Labs) and then diluted with ultra-pure
water to a volume of 200-mL. The 200-mL solution was then titrated with 25 mM S2O32-
prepared as specified in Appendix A. A few drops of 2 % (w/v) starch solution prepared
as described in Appendix A, was used as the indicator of the end point for the titration.
The reactions used to calculate the concentration of the thiosulfate ion are outlined in
Appendix B.
1.2.2.2 Standardization of Iodine Solution against Standardized Thiosulfate Solution:
A 20-mL aliquot of the standardized S2O32- (25 mM) was added to a 250-mL Erlenmeyer
flask. Two drops of 2 % (w/v) starch solution was added to the S2O32-solution. The
resulting solution was titrated with iodine solution (Appendix A) to a blue-starch
complex. The balanced reaction between thiosulfate and iodine is shown in Appendix B,
reaction R – 1.3. The molarity of the iodine solution was calculated using equation Eq.
1.7 (Appendix B). This process was repeated for a total of three replicates.
18
1.2.3 Standardization of Sulfide Using Iodometric Titration
A 25-mL portion of the standardized iodine solution (Appendix A) was transferred into a
250-mL Erlenmeyer flask. The iodine solution was acidified by pipetting 2-mL of 6 M
hydrochloric acid (Appendix A), and was swirled to mix. A 0.5-mL aliquot of the sulfide
stock solution (section 1.2.1.1) was then pipetted into the iodine/acid mixture. To limit
exposure to air, the sulfide solution was discharged below the surface of the iodine/acid
mixture. As the reaction R – 1.4 Appendix B shows, the added sulfide solution reacted
with I2 which had been added in excess. The unreacted I2 was then back-titrated with
standardized S2O32- solution (Reaction R – 1.5, Appendix B). For this titration, a
standardized S2O32- solution (Appendix A) was titrated into the iodine solution mixture
until a pale straw colour was achieved, after which a few drops of 2 % (w/v) starch
solution (appendix A) was added. The addition of the starch solution led to the formation
of a blue starch complex coloration which indicated the presence of some unreacted I2.
Further titration with the S2O32- led to the disappearance of the blue coloration. This
marked the end point of the titration. The reactions R – 1.4 and R – 1.5 (Appendix B)
were used in the computation of the sulfide in solution.
1.2.4 Electrochemical Analysis of Sulfide
1.2.4.1 Equipment
All electrochemical measurements were conducted using the BASi work station with
Epsilon USB software (Bioanaytical System Inc.). The three-electrode cell consists of a
C3 glass cell vial with a dimension of 50 mm x 59 mm, that houses the supporting
electrolyte solutions, and electrodes used for analysis. A glassy carbon electrode with a
solvent resistant coating of chlorotrifluoroethylene with a working surface diameter of
3.0 mm (area = 0.017 cm2); was used as the working electrode. A coiled Pt wire, 23-cm
in length was used as the auxiliary electrode. An Ag/AgCl, filled with 3.0 M NaCl, was
used as the reference electrode. All electrodes were supplied through BASi, and were
assembled as shown in Figure 1.
19
A)
B)
Figure 1: Electrochemical cell setup for chemical analysis. (A) From left to right:
Reference electrode (Ag/AgCl) shaded grey, glassy carbon electrode (GCE) shaded black
and platinum auxiliary electrode (Pt) represented as light grey coil. (B) The BASi setup
for electrochemical analysis, following the same electrode order from left to right as is
the case in A.
1.2.4.2 Cleaning the Electrodes
Between experiments, the glassy carbon electrodes were cleaned using various grades of
diamond and alumina polish over their respective polishing pads. Surfaces were rinsed
well with ultra-pure water, and sonicated (FS20, 3.0 qt; Fisher Scientific) for 5 minutes to
ensure polishing particulates were removed. The Ag/AgCl reference electrode and the Pt
auxiliary wire were rinsed with ultra-pure water between experiments.
1.2.4.3 Electrochemical Pretreatment of Glassy Carbon Electrode
Using the electrochemical cell, described in section 1.2.4.1, all glassy carbon electrodes
were pretreated electrochemically using the 0.1 M phosphate buffer solution (Appendix
A) as the supporting electrolyte solution. To do this, CV was used to cycle the glassy
carbon electrode potential from -200 mV to +800 mV at 100 mV s-1 for 25 cycles. This
procedure, according to a number of authors, Nagaoka and Yoshino, 1986; Kamau, 1988;
Yang and Lin, 1994; McCreery and Cline, 1996 (Ch. 10); Dekanski et al., 2001; Kiema et
al., 2003; Zhao et al., 2008 and McCreery, 2008, helps to remove any residue on the
surface of an electrode that may have remained during the electrode polishing process.
20
1.2.4.4 Electrochemical Deposition of V2O5 Film for Modified Glassy Carbon
Electrode (MGCE):
The electrochemical deposition of V2O5 onto glassy carbon electrodes has been
recommended by Khudaish & Al-Hinai (2006) and D’Elia et al. (2004). The VO2+
solution (section 1.2.1.2) was added to the electrochemical cell so that there was 2 cm of
head space at the top of the cell. The pretreated working glassy carbon, reference and
auxiliary wire electrodes were then placed in the cell. Using CV, the electrochemical
deposition of the V2O5 film was coated onto the surface of the glassy carbon electrode by
cycling the electrode potential from 0 to +2000 mV, at a scan rate of 50 mV s-1, for 20
cycles. The MGCE was removed and rinsed gently with ultrapure water, and stored in
the supporting electrolyte solution (Appendix A) until needed. The MGCE’s were used
for the electrochemical experiments the same day that they were modified.
1.2.4.5 Electrochemical Analysis Using Cyclic Voltammetry (CV)
Electrochemical measurements of the sulfide solution (section 1.2.1.1) using CV were
conducted by cycling the electrode potential from -1600 mV to +1600 mV, at a scan rate
of 100 mV s-1 for 10 cycles. The peak potential (mV), height (µA), and area (µC) were
recorded for all 10 cycles. Six replicate runs were carried out for each experiment over
bare glassy carbon electrodes (BGCE) and MGCE. All the potential measurements were
carried out using the set-up described in section 1.2.4.1, and the electrochemical
parameters outlined above were also used for the experiments involving the polysulfides
and polythionates; which are outlined in chapter 3.
1.2.4.6 Characterization of Bare and Modified Glassy Carbon Electrode Surfaces
Working electrode surface characterizations on the BGCE and vanadium oxide MGCE
were conducted through an external laboratory using a Hitachi SU-70 Schottky Field
Emission Scanning Electron Microscope coupled with Energy Dispersive Spectrometer
(Oxford Aztec 80 mm/124 eV EDX).
21
1.3 RESULTS
1.3.1 Standardization of Iodine and Thiosulfate Solutions
As shown in Table 1, the standardized concentrations for the iodine and thiosulfate
solutions as prepared in Appendix A were 25 ± 0.2 mM and 25 ± 0.1 mM respectively.
The small standard error points to the fact that iodine and thiosulfate were of very high
purity. As indicated in reaction R – 1.3 (Appendix B), iodometric titration involves the
reduction of iodine with the thiosulfate ion. Prior to using the sulfide solution for
electrochemical analyses, the sulfide solution needed to be standardized using iodometric
titration.
Table 1. Standardized values (mM) for thiosulfate (S2O32-) and iodine (I2) used in
iodometric determination of sulfide. Values are the mean of n = 5.
Solution Expected Conc. (mM) Mean Conc. (mM)
S2O32- 25 25 ± 0.1
I2 25 25 ± 0.2
1.3.2 Standardization of Sulfide Solutions using Iodometric
Titration
As described above, the standardization of sulfide requires the addition of the iodine
solution to be in excess. Since the iodine is present in excess, the iodine oxidizes the
entire sulfide that is present in solution (Appendix B, R – 1.4). The excess iodine is
determined by back titrating with standardized thiosulfate (Appendix B, R – 1.3). The
concentration of the sulfide in the stock solution (section 1.2.1.1) was found to be 106 ± 1
mM (n = 9). The difference between the calculated, based on the original weight of Na2S
prior to dissolution, and the experimental values determined iodometrically was 4 ± 1
mM (n = 9). It should be noted that there was some precipitate formed in solution during
the titration process. This was most likely as a result of the formation of elemental sulfur
during the oxidation of sulfide by iodine (Appendix B, R – 1.4). Also, according to
Pawlak & Pawlak (1999), some of the concentration differences between the calculated
22
and the titrated values could be attributed to the oxidation of sulfide to sulfate during
transfer of the sulfide from the basic medium (pH ≥ 10) to the acidified iodine solution.
Also, the preparation of the sulfide stock solution (section 1.2.1.1) which involved the
dissolution of a small weight of the anhydrous Na2S into a small volume of the buffer
(20-mL) could also have generated the significant difference between the calculated and
experimental values as observed above.
1.3.3 Electrochemical Analysis of Sulfide
1.3.3.1 Electrochemical Analysis of Sulfide using BGCE
According to the sulfur fractional dissociation of sulfide plot (Figure 2), drawn as a
function of pH, the main species present in the sulfide containing solution having a pH of
10.3 is primarily HS-. A small amount of S2- may also have been present. The acid base
equilibria for the dissociation of sulfide in solution are outlined in reactions R – 1.5 and R
– 1.6, in Appendix B. The Ka1 and Ka2 values used in the plotting of Figure 2 were 9.1 x
10-8 and 1.1 x 10-12 (Harris, 2010).
As the HS- becomes oxidized it is possible that more than one oxidation product is
formed at the electrode surface (Kuhn et al., 1983). As shown in Appendix B, the
oxidation products of sulfide may include S0, S2O32-, S4O6
2-, SO32-, and SO4
2-. It has been
reported that at the pH ranges of 7.5 – 11, sulfide oxidation may result in the formation of
SO32-, S2O3
2-, and SO42- (Zhang and Millero, 1993).
23
Figure 2. Fraction dissociation plot for sulfide as a function of pH. Where indications of
sulfide species in solution are H2S(aq) (♦), HS- (■) and S2- (▲).
Figure 3 shows six overlapped voltammograms, obtained at the BGCE. Five of the
voltammograms are for 1 mM HS- solution and the other one for the supporting
electrolyte solution. As indicated in Figure 3, the voltammograms were obtained by
scanning potential from -1600 to +1600 mV, the potential at which it was switched back
to -1600 mV. The results in Figure 3 show that more than one oxidation products were
obtained while anodically scanning the potential. On the other hand, one cathodic
product appears to have formed while scanning the potential in reverse. There was no
presence of electroactive species in the blank under these analysis conditions.
The purpose for conducting several CV cycles for each solution was to monitor any
changes that may have occurred at the surface of the electrode. Specifically, the potential
was switched back and forth ten times between -1600 mV and +1600 mV for each
solution to provide information on probable adsorption onto the surface of the electrode.
In order to produce a more distributed spread between the CV cycles, only the even
cycles are represented in Figure 3. The Epa appears at three distinct potentials, with their
0
10
20
30
40
50
60
70
80
90
100
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14
% C
om
po
siti
on
of
sulf
ide
sp
eci
es
pH Units
H2S ♦HS- ■ S2- ▲
24
mean values being -6, +434, and +1158 mV vs. Ag/AgCl respectively. There is also a
peak present in the Epc at a mean value of -1208 mV vs. Ag/AgCl. Peak current values,
at the indicated potentials in the Epa and Epc segments, exhibit increases from cycles 1 to
10. However, when the peak currents alone are examined, as shown in Figures 4 and 5,
versus the cycle number, there is more or less a steady current response from cycles 5 to
10 in most cases. More cycles, were examined, up to 15, but it became clear that there
were no significant increases in terms of the CV outputs after the 10th cycle (Figure 4).
Since this was the case, the 10th cycle was selected to represent the CV analysis for each
of the proceeding experiments using BGCE.
25
Figure 3. Overlapped voltammograms showing five CV cycles of HS- (1 mM), with the
supporting electrolyte at BGCE with a scan rate of 100 mV s-1.
Figure 4. Overlapped voltammograms showing 10 – 15 CV cycles of 1 mM HS- solution
at the BGCE with supporting electrolyte solution (dashed line), at scan rate of
100 mV s-1.
-160
-100
-40
20
80
140
200
260
320
380
-1600 -1200 -800 -400 0 400 800 1200 1600
I /
µA
E / mV vs. Ag/AgCl
Blank
2
4
6
810
-200
-125
-50
25
100
175
250
325
400
-1600 -1200 -800 -400 0 400 800 1200 1600
I / µA
E mV vs. Ag/AgCl
Blank
10-15
26
As shown in Figure 5, peak current responses (Ipa), moving towards more positive
potentials, at the mean potential positions listed above from Figure 3 exhibit an increase
in their values from cycle 2 to cycle 5. However, from cycles 5 to 10 the peak current
plateaus at potential positions -6 and +434 mV (Figure 3), with a mean Ipa = 15 ± 0.45 µA
and 57 ± 0.55 µA, respectively. The peak current at the potential position +1158 mV
displays a gradual increase from cycle 5 to cycle 9 with the peak current dropping off by
11 µA in cycle 10 (Table 2). The peak(s) through this potential region get masked by the
electrochemical response to the presence of water molecules in solution as the CV cycles
increase. Khudaish and Al-Hiani (2006) have suggested that this region is close to the
electrolysis of water, and one should use caution when interpreting data in this region.
Figure 5. Epa positions (mV) at -6 (♦), +434 (▲), and +1158 (■) with peak current values
(Ipa) µA at CV cycles 2 and 5-10 for 1 mM HS-.
0
10
20
30
40
50
60
70
80
90
100
2 5 6 7 8 9 10
I pa
/ µ
A
CV cycle number
27
Table 2. Mean potential positions and peak current (µA) values in the anodic segment for
a 1 mM solution of HS- through the 2nd to the 10th CV cycle.
Potential Positions Epa (mV)
-6 +434 +1158
CV Cycle No. Peak Current Ipa (µA)
2 0 45 46 3 8 51 68
4 12 51 74
5 13 56 79
6 15 56 82
7 15 56 84
8 15 57 89
9 16 57 92
10 16 57 81
Outlined in Figure 6 are the peak current responses (Ipc) at the potential, Epc, of -1208
mV. The results indicate that from CV cycle 2 to cycle 5, there is an increase in peak
current from 52 µA to 97 µA in the 2nd and 5th cycles respectively. The peak current then
becomes steady at a value of 106 µA from the 6th to the 10th cycle (Table 3).
As stated above, the results shown in Figure 3 and summarized in Table 2 and Table 3
indicate that when 1 mM HS- is scanned anodically, three peaks result whose peak
currents appear to become steady from the 6th CV cycle. The anodic peaks appear at – 6,
+ 434, and at +1158 mV. The cathodic peak appears at -1208 mV. A number of authors
such as Kissinger and Heineman (1996), Scholz (2010) and Zanello et al. (2012) have
pointed out that for a reversible redox reaction, the difference between the anodic peak
and the corresponding cathodic peak potentials (∆Ep) should be 59
𝑛 mV, where n
represents the number of electrons involved in the redox reaction. The minimum
separation between the anodic peak potential and that of the only cathodic peak potential
(Figure 3) is 1214 mV. The redox reactions that occurred during the CV of sulfide
cannot therefore be considered reversible. According to Kissinger and Heineman (1996),
electrochemical irreversibility is caused by slow electron exchange of the electroactive
species at the working electrode. In some cases, irreversibility may be, according to the
authors, be a result of diffusing away the product formed at the working electrode into the
bulk of the solution before the reverse scan, at the particular potential of interest, is
28
reached. These authors also state that the irreversibility of a redox system may be
influenced by smaller rate constants. They also go on to state that irreversibility
influences the peak current ratio, in that smaller peak current values may be observed in
the reverse scan than in the forward scan. Based on these facts, there was no evidence
from the data that was obtained (Figure 6 and Table 3), of an indication of reversibility
through this oxidation-reduction couple. Also, the anodic peak current values were 16,
57 and 92 µA after 10 CV cycles (Table 2), the anodic peak currents showed no increase
in value from CV cycles 10-15 (Figure 4). This phenomenon could suggest a quasi-
reversible redox system (Zanello et al., 2012).
Figure 6. Epc position (mV) at -1208 with peak current values (Ipc) µA at CV cycles 2 and
5-10 for 1 mM HS-.
0
20
40
60
80
100
120
2 5 6 7 8 9 10
I pc
/ µ
A
CV cycle number
29
Table 3. Cathodic peak curretns at -1208 mV for CV cycles 1 through 10 for a 1 mM HS-
solution.
Potential Position Epc (mV)
-1208
CV Cycle No. Peak Current Ipc (µA)
1 52 2 52
3 76
4 89
5 97
6 104
7 106
8 106
9 106
10 106
Depicted below in Figure 7 and Table 4, are the results of the CV analysis of the various
concentrations of sulfide at -1263 mV on the BGCE. There is a good positive linear
correlation towards increasing concentrations of sulfide (R2 = 0.914). However, the
precision of the current signal decreases as the concentration of sulfide increases. Table
4 outlines the mean peak current response as a function of the sulfide concentration
solution. At lower sulfide concentrations (0.6 mM), the standard error is decreased over
the upper concentrations examined (2.2 mM), where the standard error is 4 times greater
than that of the lower end of the concentration range examined. Three to six replicates
were examined in order to narrow the observed error over the concentration range of
sulfide, and to provide adequate information to assess the reproducibility of the
experiment. However, even with the increase to the number of replicates, the error
decreased only slightly. Table 5 and Figure 8 illustrate the CV analysis and the peak
current response of the various concentrations of sulfide in solution with potentials
moving to more positive regions. Of the three oxidative peaks, the one at + 565 mV has
the best linear correlation between the sulfide concentration and the current response.
The R2 value at this potential is 0.926 while as at the other two peaks, 0 mV and +1142
mV, the R2 values are 0.628 and 0.897, respectively. There are 3 oxidative products that
correspond to the potentials of 0, +565 and +1142 mV. Based on the half-reaction and
standard formal potentials outlined by Bouroushain (2010), the possible oxidative
products formed could be elemental sulfur, sulfate, sulfite and thiosulfate. As reported by
30
authors Lawrence et al. (2000b) and Dutta et al. (2009), the error that is outlined in
Tables 4 and 5 could be attributed to the mechanical polishing that is required between
experiments in order to remove any adhered oxidative products, such as elemental sulfur,
that may be on the surface of the electrode after analysis. Polishing is known to re-
activate the surface of an electrode (Kamau, 1988; Chen and McCreery, 1996; Kiema et
al., 2003; McCreery, 2008). Even though every effort was made to polish the electrodes
that were used during the CV analyses of the HS- solutions as consistently as possible,
there was no guarantee that every residue on the surface of any given electrode was
removed. Further, since it was not possible to consistently use only one electrode
through the CV experiments, it is likely that imperfections in the various electrodes could
have contributed to some of the standard error values that were observed.
Figure 7. CV analysis for the various HS- concentrations at a mean Epc = -1263 mV at
BGCE. Error bars represent ± S.E. of six replicate runs (R2 = 0.988).
0
50
100
150
200
250
300
0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25 2.5
Ipc
/ µ
A
Bulk Sulfide Concentration (mM)
31
Table 4. Mean of six replicate runs and ± SE of current responses (Ipc) for CV analysis of
sulfide obtained at the BGCE at the Epc of -1263 mV.
Bulk Conc. Sulfide (mM) Mean Ipc (µA)
0.6 33 ± 8
1.1 69 ± 16
1.4 95 ± 22
1.6 124 ± 26
1.9 218 ± 20
2.2 243 ± 36
Table 5. Mean CV anodic currents (Ipa) for the sulfide analysis at the BGCE at potentials
of 0, +565, and +1142 mV. Mean (µA) ± SE
Conc. Sulfide (mM) Ipa(1) (0 mV) Ipa(2) (565 mV) Ipa(3) (1142 mV)
0.6 5 ± 1 27 ± 5 39 ± 8
1.1 12 ± 3 46 ± 8 50 ± 8
1.4 21 ± 3 63 ± 7 77 ± 7
1.6 27 ± 8 73 ± 12 80 ± 13
1.9 68 ± 14 71 ± 14 ―
2.2 70 ± 13 106 ± 22 ―
2.5 41 ± 8 98 ± 13 ―
32
Figure 8. CV anodic currents (Ipa) for various HS- concentrations at Epa = 0 mV (♦) R2 =
0.628, +565 mV (■) R2 = 0.926, and +1142 mV (▲) R2 = 0.897 on BGCE. Error bars
represent ± S.E. of six replicate runs.
Figure 9 shows scanning electron microscopy (SEM) images of the BGCE surfaces for a
polished and electrochemically pretreated electrode (image A) and after running 10 CV
cycles (-1600 to +1600 mV) of a 1 mM HS- solution on the polished and
electrochemically pretreated electrode (image B). As indicated in section 1.2.4.3 the
pretreatment of the electrode was accomplished by running CV scans by varying the
potential of the working electrode from –200 to +800 mV at 100 mV s1- for 25 cycles.
The chevrons, shown in Figure 9 (B), indicate small deposits present on the surface of the
BGCE. Energy-dispersive X-ray (EDX) was conducted in conjunction to SEM work
(Figure 9). However, the only spectral band present in both samples was carbon for the
results attained though EDX (Figure 10 and Figure 11). Based on the EDX shown in
Figure 9, there appears to have been insufficient sulfur build-up on the electrode surface
for detection. The SEM image in Figure 9 (B) shows that there was some deposition that
had built up on the surface of the electrode. The SEM and EDX work performed was
conducted at an external laboratory. The time it took for the sample preparation to be
done and the impending analysis to be completed was substantial and could have affected
0
20
40
60
80
100
120
140
0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25 2.5
I pa
/µA
Bulk concentration of sulfide (mM)
33
the chemistry of any adsorbed material on the surface of the electrode. Figure 12 shows
another SEM image of a BGCE after 10 CV cycles (-1600 to +1600 mV) of a 1 mM HS-
solution made up in 0.1 M phosphate buffer. The SEM image, taken using Joel JCM-
6000 (Hoskin Scientific), was obtained on the same day that the CV cycles of 1 mM HS-
solution were analyzed on BGCE. Although it was observed that there was more
deposition on the surface of the electrode, as opposed to the image observed in Figure 9
(B), the composition of the surface could not be determined due to the limitation of the
SEM utilized for the image in Figure 12. The SEM utilized for the acquisition of the
image in Figure 12 was limited due to the lack of an alternative detection source (EDX)
for appropriate characterization of the deposits observed. The images attained, using that
particular SEM (Joel JCM-6000), were for exploratory purposes. Due to this limitation
the presence of sulfur or any other product that may have been deposited on the BGCE
could not be confirmed.
Figure 9. BGCE SEM images A) is polished and electrochemically pretreated in 0.1 M
phosphate buffer, B) BGCE after 10 CV cycles of 1 mM HS-, in 0.1 M potassium
phosphate buffer, chevrons indicate some product build-up from sulfide electrochemical
oxidation.
34
Figure 10. EDX spectra of BGCE pretreated by running CV cycles of the BGCE in 0.1 M
phosphate buffer from -200 to +800 mV at 100 mV s1-; where C = carbon.
Figure 11. EDX spectra of BGCE after 10 CV cycles (-1600 to +1600 mV) in 1 mM HS-
solution made in 0.1 M phosphate buffer. Where C = carbon and O = oxygen.
0
250
500
750
1000
1250
1500
1750
2000
2250
2500
0 2.5 5 7.5 10 12.5
Inte
nsi
ty (
Co
un
ts)
keV
C
0
1200
2400
3600
4800
6000
7200
8400
9600
10800
12000
0 2.5 5 7.5 10 12.5
Inte
nsi
ty (
Co
un
ts)
keV
C
O
35
Figure 12. SEM of BGCE obtained after scanning 10 CV cycles (-1600 to +1600 mV) in
a 1 mM HS- solution prepared in 0.1 M phosphate buffer. Chevron indicates deposition
onto electrode surface.
1.3.3.2 Cyclic Voltammetric Analysis of Sulfide Solutions at MGCE
As explained in section 1.2.4.4, the modified electrode surface film was prepared by the
electrochemical deposition of an oxide of vanadium on the GCE surface. This was
intended to enhance the detection of sulfide at lower concentrations. With repetitive
potential cycling between 0 to +2000 mV, for 20 cycles at a scan rate of 50 mV s1-, the
current density and peak potential in the anodic segment decrease. This decay is
attributed to the consumption of VO2+ to form V2O5 at the electrode surface (Weckjuysen
and Keller, 2003; D’Elia et al., 2004; Khudauish and Al-Haini, 2006). Weckjuysen and
Keller (2003) and Harris (2010) have indicated that the reaction responsible for the
formation of the V2O5 film from VO2+ is given by the Reaction R – 1.7, shown below.
2𝑉𝑂2+ + 3𝐻2𝑂 → 𝑉2𝑂5 + 6𝐻+ + 2𝑒− (𝑅 − 1.7)
The deposition of the film can be observed below in Figure 13 (B) where the MGCE and
the BGCE Figure 13 (A) are compared. As shown below in Figure 13 (A) the BGCE has
pores on the surface which according to McCreery (2008) are formed during fabrication.
The MGCE SEM image shown in Figure 13 (B) shows a uniform coverage over the
36
glassy carbon surface which is consistent with literature that presents characterization of
the electrodeposition of V2O5 films onto GCE (D’Elia et al., 2004).
Figure 13. SEM images of BGCE (A) and MGCE (B), showing the surface
characterizations between the BGCE and MGCE.
Figure 14 summarizes the comparison of the blank (supporting electrolyte solution) and a
0.36 mM sulfide solution at MGCE. There are peak current responses in both the Ipa/Ipc
segments of the supporting electrolyte solution, but the current density shows a marked
increase in the presence of sulfide solution. This overlap in current was corrected for by,
taking a blank measurement before each set of experiments with MGCE, taking the
current responses (Ipa/Ipc), from those blank measurements, and then subtracting the blank
current responses from the current responses of the sulfide solutions along the potential
regions of interest (Epa/Epc). In the case of Figure 14, the blank current response was 37 ±
4 µA at 132 ± 18 mV, and the current response measured for 0.36 mM sulfide was 70 ± 3
µA at 145 ±18 mV. The actual current response for 0.36 mM of sulfide taken at 145 ± 18
mV was 33 ± 3 µA with background correction applied. In contrast, the supporting
electrolyte response in the case of the BGCE (Figure 3) does not interfere with the
current responses from the sulfide present in solution as it does in the case of the MGCE.
EDX and SEM analyses of the surface of the MGCE after 10 CV cycles in just the
supporting electrolyte solution did not show any accumulation of foreign products on the
film. Also, contrary to the observation of D’Elia et al. (2004), the analyses did not show
the presence of vanadium bands. However, when the sulfide was added and 10 CV
37
cycles were conducted between -1600 and +1600 mV, there was a presence of deposits
on the MGCE surface (Figure 15 and Figure 18). EDX analyses of these electrode
surfaces revealed the presence of copper, sulfur, carbon and oxygen bands (Figure 16).
As noted earlier, even though the CV of the supporting electrolyte has significant residual
currents, the bands found on MGCE did not show any interfering products (Figure 17 and
Figure 19) that could explain the signal observed in Figure 14. Since the EDX and the
SEM with just the supporting electrolyte solution on MGCE did not show any traces of
deposition to the surface related to the supporting electrolyte solution, that would be
indicate that they interaction between the deposited V2O5 film and sulfide is responsible
for the deposits on the MGCE shown in Figure 15. Khudaish and Al-Hinai (2006)
proposed the following electrochemical mechanism for the oxidation V2O4 to soluble V5+
as shown in reactions R – 1.8 and 1.9. The proposed mechanisms (Khudaish and Al-
Hinai, 2006) for the reactions happening during CV cycles at MGCE suggest that the
vanadium oxide modifier is a self-regenerating film, which could be the cause of the
segment peaks observed in the anodic and cathodic segments (Figure 14) at Epa/Epc =
+303, +1125 and -641 mV respectively. When the MGCE is in the presence of sulfide,
the catalytic response from the MGCE increases the peak current in the overlapped
potential regions where the blank is present.
𝑉2𝑂4 + 2𝑂𝐻− → 2𝑉𝑂3− + 2𝐻+ + 2𝑒− (𝑅 1.8)
2𝑉𝑂3− + 𝐻2𝑂 → 𝑉2𝑂5 + 2𝑂𝐻− (𝑅 1.9)
38
Figure 14. The fifth cycle of a CV output of 0.36 mM concentration of HS- (solid line) at
MGCE, and the supporting electrolyte solution (dashed line) at a scan rate of 100 mV s1-.
Figure 15. SEM image of MGCE showing foreign deposits on the film modification after
10 CV cycles of 1 mM HS- at 100 mV s1- (A) and MGCE showing 10 CV cycles in the
supporting electrolyte solution at 100 mV s1- (B).
-80
-60
-40
-20
0
20
40
60
80
100
120
-1600 -1200 -800 -400 0 400 800 1200 1600
I /
µA
E/ mV vs. Ag/AgCl
39
Figure 16: EDX spectra of MGCE after 10 CV cycles (-1600 to +1600 mV) of 1 mM HS-
solution made in 0.1 M phosphate buffer. Where C, O, S and Cu represent carbon,
oxygen, sulfur and copper respectively.
0
250
500
750
1000
1250
1500
1750
2000
2250
2500
0 1 2 3 4 5 6 7 8 9 10 11 12
Inte
nsi
ty (
Co
un
ts)
keV
Cu
Cu
Cu
SO
C
40
Figure 17: EDX spectra of MGCE after 10 CV (-1600 to +1600 mV) cycles in supporting
electrolyte solution (0.1 M phosphate buffer), where C = carbon.
Figure 18. SEM image of MGCE (left), showing sites of interest for EDX along with
supporting EDX data expressed as weight % (right) for V2O5 modifier.
0
250
500
750
1000
1250
1500
1750
2000
2250
2500
0 1 2 3 4 5 6 7 8 9 10 11 12
Inte
nsi
ty (
Co
un
ts)
keV
C
41
Figure 19. SEM image of MGCE after 10 CV cycles in 1 mM HS- solution, showing sites
of deposits on surface (left) and EDX data expressed as weight % (right) for MGCE after
10 CV cycles in 1 mM HS- solution.
Figure 20 shows 10 overlapped CV cycles of the supporting electrolyte obtained using
MGCE. A progressive decay in the anodic current density from cycle 1 to 10 at an Epa of
+1200 mV is observed. According to authors Barrado et al. (1997), Weckjuysen &
Keller (2003), D’Elia et al. (2004), and Khudaish & Al-Hinai (2006), the likely cause of
this decay is the electroxidation of V2O5 to V2O4. This occurs when the V2O5 undergoes
an electron transfer in the presence of water molecules. This current decay at +1200 mV
can also be observed in Figure 21 which contains an overlap of 3 CV cycles of 0.44 mM
sulfide solution at the MGCE. This process, for the response of the V2O5 film towards
the electrochemical response to sulfide, is outlined below as reaction R – 1.10 as
proposed by McCleverty & Meyer (2004) and Khudaish & Al-Hinai (2006). According
to D’Elia et al., (2004), the presence of this decay in current density has been attributed
to the presence of dissolved oxygen. These authors pointed out that at potentials greater
than +1230 mV, the oxidation of water occurs. This in turn could enhance the
electrooxidation of V2O5 as depicted in reaction R – 1.10. Figure 20 and Figure 21 show
that between 0 and +200 mV, there is an anodic peak that develops at the MGCE for both
the supporting electrolyte alone and the 0.44 mM sulfide solution. However, the anodic
peak current in 0 to +200 mV region is more enhanced in the presence of the sulfide ion.
42
According to McCleverty & Meyer (2004) and Khudaish & Al-Hinai (2006), the
variation in current density in the Epa region 0 to +200 mV, in the presence of sulfide, is
attributed to the oxidation of V2O5·HS- which creates elemental sulfur (Reaction R –
1.10). As later CV cycles occur, further accumulation of elemental sulfur on the surface
may cause the active regions of the vanadyl film to become fouled (Figure 15 A).
𝑉2𝑂5 ∙ 𝐻𝑆− → 𝑉2𝑂4 + 𝑆 + 𝑂𝐻− (𝑅 − 1.10)
It is possible that the build-up of elemental sulfur on the MGCE surface could reduce the
availability of the reactive sites of the electrode surface (Figure 16) for continuous HS-
oxidation over a series of CV cycles. The presence of elemental sulfur was supported by
the SEM and EDX data provided from Figure 15 (A) and Figure 16 for the analysis of
sulfide at MGCE. In addition, the Epc shifted to a more negative region with each
successive CV cycle. The Epc shown in Figure 17 has an average Epc of -540 mV. When
compared to the Epc obtained with the BGCE (-1208 mV), the catalytic activity of the
V2O5 used in the MGCE may be considered responsible for a potential shift of
approximately 668 mV. These changes in current densities over the MGCE when in the
presence of the sulfide ion are also illustrated in Figure 22 to Figure 24 as well as in
Table 6. It should be noted that in Table 6 for both the Epa1 and Epc there is significant
peak shift, where the shift is 456 mV towards more positive potentials from CV cycles 1
to 10, and a 247 mV shift towards more negative potentials from CV cycles 1 to 10,
respectively. This peak shift was also observed in the MGCE CV cycles with just the
supporting electrolyte solution, where the peaks shifted 466 mV towards more positive
potentials and 325 mV towards more negative potentials for Epa1 and Epc, respectively.
As shown in Figure 22 to Figure 24, there is little separation in the current response
between the supporting electrolyte solution at the MGCE compared to the current signal
when in the presence of sulfide. However, as shown in Figure 22, the best separation in
the current response between the sulfide signal and just the supporting electrolyte signal
at the MGCE occurs at the 4th CV cycle. As a result of this, the 4th CV cycle was used
for the sulfide solution analysis at the MGCE Figure 25 and Table 7.
43
Figure 20. CV analysis of supporting electrolyte solution on MGCE from cycles 1 to 10,
scan rate 100 mV s1-; the numbers indicate the scan number of the CV cycle.
-200
-140
-80
-20
40
100
160
220
280
340
400
-1600 -1200 -800 -400 0 400 800 1200 1600
I /
µA
E / mV vs. Ag/AgCl
1
2
3
4
}5-10
44
Figure 21. CV graph at MGCE of HS- solution (0.44 mM); scan rate of 100 mV s1-
showing cycles 2, 5, and 10 as well as the blank (dashed line).
-200
-150
-100
-50
0
50
100
150
200
250
300
350
-1600 -1200 -800 -400 0 400 800 1200 1600
I / μ
A
E / mV vs. Ag/AgCl
2
10
5 & blank
45
Figure 22. Peak current values (Ipa1) µA at CV cycles 1 to 10 for 0.36 mM HS- (♦),
compared to CV cycle numbers 1 to 10, and compared to the supporting electrolyte
solution (▲) on MGCE.
Figure 23. Peak current values (Ipa2) µA at CV cycles 1 to 10 for 0.36 mM HS- (♦),
compared to CV cycle numbers 1 to 10, and compared to the supporting electrolyte
solution (▲) on MGCE.
0
20
40
60
80
100
0 1 2 3 4 5 6 7 8 9 10
I pa
1/
µA
CV Cycle No.
0
50
100
150
200
250
300
0 1 2 3 4 5 6 7 8 9 10
I pa
2/
µA
CV Cycle No.
46
Figure 24. Peak current values (Ipc) µA at CV cycles 1 to 10 for 0.36 mM HS- (♦),
compared to CV cycle numbers 1 to 10, and compared to the supporting electrolyte
solution (▲) on MGCE.
Table 6. Peak current, potential and scan number for 0.36 mM HS- run over MGCE at a
scan rate of 100 mV s1-.
Scan No. Ipa1 (µA) Epa1 (mV) Ipa2 (µA) Epa2 (mV) Ipc (µA) Epc (mV)
1 57 -250 274 1235 156 -423
2 45 -125 217 1234 128 -448
3 62 -45 135 1184 95 -492
4 77 64 76 1107 73 -543
5 71 141 41 1107 58 -594
6 57 185 25 1120 48 -620
7 50 211 18 1120 43 -664
8 43 224 14 1117 39 -643
9 38 233 14 1142 35 -668
10 37 206 12 1090 34 -670
Six concentration levels were examined in an attempt to quantify various sulfide
concentrations at the MGCE. The sulfide concentrations examined at the MGCE were:
0.67, 0.44, 0.36, 0.31, 0.22 and 0.13 mM. The sulfide concentrations used for the
calibration curve in the case of MGCE were lower than when BGCE was used (Figure 7
and Figure 8). According to Khudaish and Al-Hiani (2006), there was a significant
change in the concentrations of the sulfide solutions used in the BGCE analyses when
0
40
80
120
160
200
0 1 2 3 4 5 6 7 8 9 10
I pc
/ µA
CV Cycle No.
47
compared to that of the MGCE analyses. This change in the sulfide concentrations for
the MGCE analyses was in the part from the enhanced catalytic activity provided by the
V2O5 modifier film on the surface of the electrode. Figure 25 shows that there is a good
linear correlation between the peak current at +200 mV and the various concentrations of
sulfide in the anodic segment. These results indicate that the oxidative products of HS-
slightly increase linearly with increasing concentration of sulfide (Table 7). The results
in Figure 26 and Table 8 show that in the cathodic segment, at a potential of -570 mV, is
a linear increase in the peak current as the sulfide concentration increases during the
reduction of HS- at the MGCE. However, the standard error results (Table 8) for the HS-
concentrations 0.13, and 0.31 mM raises some concerns. For those two concentrations
(0.13 and 0.31 mM), after background corrections were conducted, had values of 0.7 ± 3
and 1 ± 4 µA. The standard error is larger than the measurement attained, which would
suggest that since these two measurements are less than their standard errors, it would be
difficult to differentiate them from background noise and are therefore unreliable. The
lower peak current values observed in the cathodic segment provide evidence that the
redox system being analyzed at MGCE is irreversible, since electrochemical
irreversibility influences the peak current ratio and that the more irreversible a couple the
smaller the peak current values will be on the reverse segment (Tables 7 and 8)
(Kissinger and Heineman, 1996; Zanello et al., 2012). Although, direct proportionality
between the peak current at -570 mV and the sulfide concentrations is weak, the signal
for detection for the reduced HS- concentrations in the cathodic segment are detected at
values ≥ 0.44 mM. Nonetheless, there is a catalytic increase in current response at the
MGCE of solutions containing sulfide when compared to the supporting electrolyte
solution without sulfide (Figure 14).
48
Figure 25. CV analysis increasing towards more positive potentials Ipa (µA) for various,
concentrations of HS- at Epa = +200 mV on MGCE after blank corrections. Error bars
represent ± S.E. of six replicate runs (R2 = 0.937).
Table 7. Mean ± SE for anodic current response Ipa (µA) of various concentrations of HS-
at MGCE; at a mean anodic potential Epa of +200 mV, (n = 6; R2 = 0.937).
Bulk Conc. Sulfide (mM) Mean Ipa (µA)
0.13 35 ± 3
0.22 36 ± 4
0.31 40 ± 6
0.36 52 ± 6
0.67 54 ± 4
25
30
35
40
45
50
55
60
0.00 0.13 0.26 0.39 0.52 0.65 0.78
I pa
/ µ
A
Bulk Sulfide Concentration (mM)
49
Figure 26. CV analysis increasing towards more negative potentials Ipc (µA) for various
concentrations of HS- at Epc = -570 mV on MGCE after blank corrections. Error bars
represent ± S.E. of six replicate runs (R2 = 0.907).
Table 8. Mean ± SE for cathodic current response Ipc (µA) of various concentrations of
HS- at MGCE at a mean cathodic potential Epc of -570 mV, (n = 6; R2 = 0.907).
Bulk Conc. Sulfide (mM) Mean (Ipc)
0.13 0.7 ± 3
0.31 1 ± 4
0.36 9 ± 4
0.44 13 ± 3
0.67 23 ± 3
1.4 CONCLUSIONS
As mentioned above fifteen CV cycles were initially examined in order to optimize an
appropriate number of cycles for the determination of sulfide at BGCE and MGCE. Only
ten CV cycles were used for each of the six replicates examined at each concentration
level listed for BGCE and MGCE. The reason for choosing 10 CV cycles was the
observation that after the 10th CV cycle there were no increases to the CV outputs (Figure
4). As illustrated in the SEM images in Figure 12 and Figure 15 (A), it is believed that
surface fouling on both the BGCE and MGCE may be attributed to the buildup of
-5
0
5
10
15
20
25
30
0.00 0.13 0.26 0.39 0.52 0.65 0.78
I pc
/ µ
A
Bulk Sulfide Concentration (mM)
50
elemental sulfur on the working surface of the electrode. Such surface fouling leads to a
progressively reduced active working electrode surface as a series of the CV scans
proceed. This effect could, in turn, cause a progressive decrease in the output current.
As the CV output presented in Figure 3 shows, there are three oxidative products that
appear at three distinct potentials, namely at 0, 565, and 1142 mV. The possible
oxidative products that may have formed during the CV analysis of sulfide at these
potentials are elemental sulfur, sulfate and thiosulfate. Kinetic studies conducted on the
oxidation of sulfide in aqueous solutions by several authors (Avrahami & Goulding,
1968; O’Brien & Birkner, 1977; Millero, 1986; Kotronarou & Hoffmann, 1991) have
proposed mechanisms for the formation of sulfur intermediates based on the presence of
oxygen in sulfide solutions being analyzed. In a study done by Avrahami and Goulding
(1968) that examined the oxidation of sulfide in water in the pH range of 11-13, the
disappearance of sulfide was found to follow first order kinetics, with sulfate and
thiosulfate being the products of oxidation. The authors further suggest that the oxidation
of sulfide forms sulfite in the presence of oxygen. The sulfite is then further oxidized to
sulfate. These authors have presented possible mechanisms for the intermediate
oxidative products of sulfur, formed from sulfide. O’Brien and Birkner (1977) also
examined the oxidation of sulfide in water, through the pH range of 7.5-11, and they
suggested that oxygen as well as the pH of the sulfide solution influence the formation of
sulfur species during oxidation. Based on their kinetic study, the authors proposed that
with low sulfide to oxygen ratios the sulfur products formed would be sulfite, thiosulfate
and sulfate. However, when the ratio of sulfide to oxygen is high the formed products are
elemental sulfur and polysulfides. In the work presented in this chapter the potentials
were cycled from -1600 to +1600 mV for 10 cycles. As reported by Harris (2010), the
electrolysis of water occurs at potentials of -1230 mV. Since the CV analysis conducted
in this work included potentials where water molecules are oxidized to oxygen, this could
potentially lead to the formation of sulfur intermediates during the anodic sweeping. The
speciation of the specific sulfur intermediates that form may be realized through further
investigations of the chemistry occurring at the surface of the electrode.
As observed from the results obtained in this chapter, there was a fair amount of
variability among the experiments performed. Increasing the number of replicates and
51
performing Dixon’s Q test to recognize and remove outliers did not seem to help reduce
the variability of sequential replicate values observed here. The results in this chapter
further indicate that the variability was reduced when CV analysis of HS- was conducted
over MGCE (Table 8 & Table 9), compared to the variability that resulted from the
analysis of sulfide at BGCE (Table 4 & Table 5).
It is clear that the higher the concentration of HS- being measured, the greater the
variability of replicate measurements (S.E. increases with increasing concentration of
sulfide). It is possible that once the sulfide is in solution, any exposure to air could cause
rapid oxidation of the HS-. Even though precautions were taken to ensure that the
exposure to oxygen was limited, i.e. significant purging of solutions with nitrogen was
maintained, HS- solution was discharged below the surface of the electrochemical cell
solution, and all procedures were carried out under a blanket of nitrogen, the variable
readings from the same sample suggest that there could have been slight exposure to air.
It has been reported that the presence of oxygen can cause the oxidation of HS-, which
leads to for formation of intermediate sulfur-based compounds such as S0, SO32-, SO4
2-
and S2O32- (Avrahami and Golding, 1968; O’Brien and Birkner, 1977; Millero, 1986;
Kotronarou and Hoffmann, 1991). Secondly, the variation of current responses could
have occurred when more than one GCE was inevitably used. Even if it were possible to
use just only one electrode, slight differences in the polishing/cleaning procedures
between analyses could also lead to current output variations.
The results obtained in this chapter indicate that the MGCE has an enhanced detectability
of HS- ion at lower concentrations than those over the BGCE. Using a MGCE, as more
elemental sulfur gets deposited onto the active vanadium film, there is decline in the
current observed over the series of ten CV cycles (Figure 21). The shift in potential and
the reduced current response after four of the ten cycles examined suggests that useable
HS- analysis could be carried out with fewer cycles than is the case for the analyses
carried out using BGCE. A major disadvantage realized when using MGCE is that the
current responses for the blank sample occurs at the same potential as do those of the HS-
solutions. This could limit the use of this approach to measure low detectable levels of
sulfide. Even though, the lowest sulfide concentration detected using MGCE was 0.13
52
mM, the observed standard error results indicated at concentrations of 0.13 and 0.31 mM
(Table 8), for the cathodic segment, these values would not be considered quantifiable.
To achieve the electrodeposition of V2O5 on the GCE, twenty CV cycles were required to
achieve the film coverage over the GCE surface. There may have been microabrasions
on the surface, like those observed in Figure 9 (A), which could have prevented uniform
deposition of the V2O5 film on the GCE surface, a phenomenon that could reduce the
precision of the results. Although the V2O5 film has previously been described as a stable
product (Barrado et al., 1997; Park et al., 2002, Weckhuysen & Keller, 2003; D’Elia et
al., 2004) the present study shows that there is a significant decay in current density in
the supporting electrolyte solution as well as in the HS- solutions. A possible explanation
for this is that the V2O5 may be undergoing secondary reactions over the sequence of CV
cycles. When the results using the BGCE and the MGCE are compared in terms of their
ability to detect sulfide, the MGCE allows lower levels of sulfide to be detected.
However, during the CV analysis at the MGCE there appears to be current output
stability after 10 cycles. When the HS- solution was added to the electrochemical cell
and CV cycles were resumed, the MGCE surface seemed to show inactivity towards HS-
oxidation. The proposed theory by Khudaish and Al-Hiani (2006) does not seem to
support the regeneration of this modification film. It has been reported elsewhere
(Barrado et al., 1997) that the V2O5 may undergo side reactions with the supporting
electrolyte as well as with any present oxygen during the CV cycles when the potential is
swept from -1600 to +1600 mV. Perhaps as a result of deposition of elemental sulfur on
the GCE surface, the accumulation of these adsorptive products can eventually deactivate
the surface of the working electrode requiring renewal (polishing) (Kissinger and
Heineman, 1996). Zanello et al. (2012), suggested that the MGCE acts like a membrane
which regenerates under different ionic environment thereby causing a shift in the
potential around a particular peak potential, at the working electrode surface. This might
help explain the potential shifts observed in Table 6.
53
Overall, the results obtained from the BGCE and the MGCE experiments casts doubts on
the reliability of these methods for the routine quantitative analysis of sulfide. The
methods nevertheless seem to provide useful qualitative information about the presence
of sulfide in a sample, and it may be possible to modify the apparatus to address the
sources of drift in measured current density and surface changes in the electrodes
themselves as further measurements are made.
54
Chapter 2
Analysis of Polysulfides and Polythionates
2.1 INTRODUCTION
Polysulfides and polythionates are a unique class of sulfur compounds, known to
participate in many environmentally significant processes due to their reactivity as strong
nucleophiles, and reducing agents (Chadwell et al., 2001; Kamyshny et al., 2004;
Kamyshny et al., 2006; Kamyshny et al., 2008 and Kristiana et al., 2010). For example,
polysulfides are important geochemically, and have a high affinity for transition metal
ions. They form sulfide minerals like pyrite (iron sulfide) (Chadwell et al., 2001 and
Kristiana et al., 2010). Polysulfides also participate in the sulfurization of organic matter
in aquatic systems, and are precursors to the formation of volatile organic sulfur
compounds such as dimethyl disulfide (Gun et al., 2000; Kamyshny Jr. et al., 2008 and
Kristiana et al., 2010). Previous reports have stated that polysulfides are present in oxic
and anoxic environments (Kariuki et al., 2001 and Kristina et al., 2010). Sulfur chain
lengths of up to S182- have been reported (Chadwell et al., 2001) being present in
biofilms, and in drinking water distribution systems (Kristiana et al., 2010). The concern
with the presence of sulfides and polysulfides in drinking water systems is that these
species consume disinfectants, dissolved oxygen and react with metal ions. These
processes can produce insoluble metal sulfides and can cause taste and odor problems
(Kristiana et al., 2010). However, they are challenging to analyze, due to their low
concentrations in natural environments (10 – 20 µM), thermal instability, and
susceptibility to oxidation and transformation (Chadwell et al., 2001; Kariuki et al., 2001;
Kamyshy Jr. et al., 2004; Kamyshny Jr. et al., 2006; Kamyshny Jr. et al., 2008 and
Kristiana et al., 2010). The dissociation of polysulfides to polythionates, sulfur and
sulfide depends on the pH, temperature and ionic strength of an aqueous solution.
Along with sulfide and polysulfides, polythionates, such as the thiosulfate ions are also
present in natural environments. These compounds have a similar chemical structure and
their metabolism appears closely related. Polythionates are important intermediate
55
species in the redox transformations of sulfur compounds in many environments and in
the metabolism of sulfur-oxidizing and sulfur-reducing microorganisms (Koh, 1990;
Druschel et al., 2003; Mohapatra et al., 2008). The versatility of thiosulfate is that it can
be oxidized to sulfate or tetrathionate, reduced back to sulfide, or disproportionate into
both sulfide and sulfate, through the metabolism of sulfate reducing bacteria (Ciglenečki
and Ćosović, 1997). Controlling and understanding the distribution of sulfur-base
compounds could aid in the reduction of hydrogen sulfide production from these redox
processes, which in turn could be useful in the management of odors released in anoxic
environments and water treatment processes as well as managing outbreaks of taste and
odour problems in drinking water supplies.
The availability of polysulfide compounds commercially is rare and when available, the
compounds are largely costly. These compounds were therefore synthesized as needed in
the present study. The methods used to synthesize the sodium salts of disulfide,
trisulfide, and tetrasulfide from sodium sulfide and the elemental sulfur are described in
this chapter. The analysis of the synthesized Na-polysulfide compounds using the N,N-
diethyl-p-phenylenediamine (DEPD) colorimetric method, and also the iodometric
titration method was carried out as described in section 1.3.2. Using these two methods
to directly analyze polysulfides and polythionates would be the preferred option. If the
polysulfide ions cannot be directly analyzed using DEPD, they would be reduced to
simple sulfides (S2-) using chromium. The reduced Sn2- ions will be analyzed using
DEPD and iodometric titrations. The DEPD and iodometric titrimetric methods, also
used to analyze thiosulfate ions, are also described in this chapter. The thiosulfate ion
was used as a representative of the class of compounds known as polythionates. An
attempt on how these two methods (DEPD and iodometric) were used to quantify the
polysulfides and thiosulfate directly without transforming them first is also described in
the chapter. Ultimately, it became necessary to reduce these compounds to simple sulfide
before being quantified with these two techniques.
56
2.2 MATERIALS AND METHODS
2.2.1 Making Na-polysulfides
2.2.1.1 Equipment and Preparation
Commercially available polysulfide compounds are only obtainable in the tetrasulfide
form. Even though available, the product is pricy. A 5-g quantity of sodium tetrasulfide
currently costs $500 and even then, its purity is not guaranteed to that of analytical grade.
A preparative method outlined by Rozen and Tegman (1971) was used to make the
polysulfide compounds of interest for this study. Due to the high reactivity of sodium
sulfide (Na2S) with air, the weighing, mixing and transfer of the reactants used in the
making of the polysulfides into the reaction vials was carried out in a Polymer Series 100,
Baxter Glove Box, equipped with a dual purge nitrogen flow control, for humidity, and
temperature monitoring (Terra Universal Inc.). Stoichiometric amounts of elemental
sulfur (S0) and anhydrous Na2S were weighed out on an analytical balance (Ohaus
Explorer Pro; model no. EP114C), in a nitrogen purged environment. The equations 2.1
(a & b), 2.2 (c & d), and 2.3 (e & f) (Appendix B) illustrate the proportions of Na2S and S
that were used to synthesize sodium disulfide (Na2S2), sodium trisulfide (Na2S3), and
sodium tetrasulfide (Na2S4). For example, to make a polysulfide containing an n-sulfur
chain (Sn2-), the molar ratio of Na2S to S0 would be 1:(n – 1).
The relative humidity of the glove box in which the anhydrous Na2S and elemental sulfur
were prepped was set at a range of 20-30%. Ambient temperature in the glove box was
maintained. Mortars and pestles, scoopulas, glass tubes, and glass enclosure vials, used
in this procedure were dried in an oven at 150°C for 12 hours, removed, wrapped in
aluminum foil, and cooled to ambient temperature before using. The glass tubes, and
glass enclosure vials were purged with nitrogen gas, and their openings were sealed with
parafilm before introducing them in the glove box. The glove box was initiated and the
main chamber was evacuated. Manipulations of weighing or mixing the Na2S or S did not
occur until the atmosphere inside the main chamber reached 30% relative humidity or
lower.
57
2.2.1.2 Polysulfide Mixing Process
Stoichiometric amounts of Na2S were weighed out and ground using a mortar and pestle.
A stoichiometric amount of S0 was weighed and added to the same mortar containing the
Na2S. The two reactants were mixed, using the pestle, until completely homogenized.
The homogeneity of the Na2S and S reactants was required to ensure that there would be
uniform melting, and re-crystallization of the final product. Stoichiometric portions of
the individually identified polysulfide mixtures were placed into 18 x 150 mm
borosilicate glass tubes (Baxter Scientific Products, T1290-9A), and sealed with parafilm,
before removing them from the glove box via the air lock port. This was to ensure that
any exposure to oxygen prior to evacuating the nitrogen from the tube during the sealing
process was minimized as much as practically possible.
2.2.1.3 Sealing the Glass Tubes for Final Preparation
Using a water aspirator to create a vacuum, the parafilm-sealed tube containing the
reaction mixture was quickly un-wrapped and attached to the vacuum tubing. Making
sure the sealing from the vacuum tube to the sample tube was air tight. The reactant tube
was evacuated for 60 minutes. Using a hand torch cylinder (MG9 - 14.1 oz. MAP-Pro,
Worthington Cylinders U.S.A.), the tube was sealed by gently softening the upper portion
of the glass tube. This was done by carefully rotating the glass tube around the flame
until the tube walls collapsed, creating a sealed environment for the reaction mixture.
2.2.1.4 Preparation and Retrieval of Na-polysulfides
Outlined in Table 9 is the temperature profile (Rosen and Tegman, 1971) that was used to
form the Na-based polysulfides. The Na-polysulfide reactant mixture was conducted in
an Isotemp Muffle Furnace (Fisher Scientific) using the temperature profile outlined
below in Table 9. Once the reaction was complete, each of the polysulfide products was
removed from its respective glass tube and placed in the mortar for crushing and grinding
to a fine powder. Before being used, the mortars, pestles, and the glass storage vials had
been baked in an oven at 150 °C for 12 - 24 hours and cooled to room temperature. The
58
glass vials into which the synthesized polysulfides were placed had been purged with
nitrogen prior to transferring them into the glove box.
Table 9. Temperature profile for the preparation of Na-polysulfides (Rosen and Tegman,
1971).
Temperature
Segment
Reaction
Temperature (°C)
Reaction
Time
(hours)
Comments
Segment 1 230 10 – 12
Solid state conversion, 80-90 % of
reaction occurs at this segment
Segment 2 300 – 490 ½
Liquid state reaction, complete
conversion
Segment 3 205 1 - 10
Tempering period, recrystallization
occurs
Retrieval of the Na-polysulfides from the glass tubes in which they were synthesized was
conducted in the Baxter Glove Box, under the same conditions as those used for the
initial mixing of the reactants as outlined above. To detach the synthesized polysulfides
from the walls of the glass tubes in which the reactions were carried out, the glass tubes
were carefully tapped with an object of moderate weight. Each glass tube was then
gradually broken apart, starting from the top of the seal and moving downward towards
the polysulfide product. Once the Na-polysulfide was removed from the glass tube it was
placed in a mortar, and a pestle was used to crush and grind it to a powder. The latter
was then transferred into a 12 x 75 mm glass storage vial, with threaded screw cap
(Figure 27). The glass vials containing the Na-polysulfides were packed under nitrogen,
in zipper-lock bags, and placed in the fridge until needed for further analyses.
The final Na-polysulfides were sent to an external laboratory for X-Ray Diffraction
(XRD) (Pananalyical Expert Pro, Pixcel Detector) analysis for purity. The system was
operated at 45 kV, 40 mA, in which the incidence angle spanned from 6° to 140°2θ,
under argon. A copper anode was used in this analysis.
59
Figure 27. Pictures of the synthesized Na-polysulfides: (A) Na2S2; (B) Na2S3; (C) Na2S4
2.2.2 N,N-Diethyl-p-phenylenediamine (DEPD) Analysis of Sulfide
The spectrophotometric method for the analysis of sulfide using DEPD was adopted from
another study by Kariuki et al. (2008). Into a 25-mL volumetric flask, a small amount of
the alkaline ultra-pure water, which had been previously purged with nitrogen, was added
to a 25-mL volumetric flask. A 250-μL aliquot of the 106 mM sulfide stock solution was
then added to the flask, along with 2-mL of the color developing reagent, prepared as
described in Appendix A. The alkaline ultra-pure water was used to top up the flask to
the 25-mL total volume. The 1 mM sulfide solution was left to react for a 30-minute
period. The resulting solution was used to create 25, 15, 10, 5 and 2.5 µM of sulfide. The
DEPD spectrophotometric method is analogous to the Methylene Blue method whose
main reaction is shown below. The diluted solutions were immediately analyzed using a
UV-VIS spectrophotometer (GENESYS, Thermo Scientific) at a fixed wavelength of
670 nm, using a rectangular quartz glass cuvette with a path length of 1 cm (Agilent
Technologies). A blank was prepared by using nitrogen-purged ultra-pure water.
2.2.3 Reduction of Na-polysulfides and polythionates Using
Chromium as the Reducing Agent
The set-up for the reduction of the polysulfides and the thiosulfate ion is shown in Figure
28. The reduction for these compounds was adopted from a procedure reported by
Kariuki et al. (2008). In brief, a purge and trap set-up consisting three 125-mL
60
Erlenmeyer flasks were connected in series. A 50 mL aliquot solution of the sulfur-
containing compound being reduced was placed in the flask labeled A. If thiosulfate was
being reduced, the solution was made up in water. The polysulfides were made up in
0.1 M NaOH solution. To minimize air oxidation of the polysulfides, their solutions were
made up in the Baxter glove box under a nitrogen rich environment. Before removal
from the glove box, the solutions were sealed with a rubber stopper and parafilm prior to
removal from glove box for reduction.
The 0.1 M NaOH solution was pre-purged with nitrogen gas prior to using with
polysulfides. This was to limit any prior exposure to oxygen in order to control the rapid
oxidation of the polysulfides. The thiosulfate solution was made up in pre-purged ultra-
pure water. The thiosulfate is a very stable product and the need for high pH solutions, as
is the case for simple sulfides and polysulfides, was not required for this compound. In
each of flasks (B and C) 60 mL 0.1 M NaOH was placed. Three, two gas-tube stoppers
were connected with Tygon tubing and placed on the top of each of the flasks (Figure
28). The third flask (C) in the sequence (see Figure 28) was intended to function as an
extra ‘catch flask’. If recovery of polysulfides from flask (B) were to be <100 %,
following reduction, some of that reduced solution might escape into flask (C) and be
trapped there. Nitrogen gas was connected to the inlet source tube for reaction flask (A),
and a gentle stream of gas flow was initiated. Flasks B and C were connected first,
wrapped in parafilm, and quickly, the third stopper was connected to flask A. All three
flasks were purged for 30 minutes. With the nitrogen gas still flowing, approximately
1.50 g sample of chromium metal was weighed, added to reaction flask (A), and then 7
mL of concentrated hydrochloric acid was also added to the same flask. The gas-tube
stopper was quickly placed on the top of the flask, immediately sealed with parafilm, and
continuously purged for 4 hours, at 60 °C while the reaction completed. After the
reaction was complete, working under a blanket of nitrogen gas, flasks B and C were
removed and the reduced solutions were poured, individually, into a 60 mL Teflon
storage bottle (Nalagene®, Fisher Scientific). The DEPD reaction, as outlined in section
2.2.3, was conducted on reduced solutions captured in flask B. These reduced solutions
were also used for electrochemical experiments, which will be described in Chapter 3.
61
Figure 28. Purge and trap system set-up for the reduction of aqueous Na-polysulfides
(Sn2-) and thiosulfate (S2O3
2-): flask (A) (Sn2- or S2O3
2-); flasks B and C (traps for reduced
sulfide).
2.3 RESULTS
2.3.1 Making Na-polysulfides
As described in the procedure section, 2.2.1 the synthesis the Na-polysulfides involved
several stages. One of the stages involved the sealing of the reaction mixture in
borosilicate glass tubes, before placing the tubes in a muffle furnace. Extra care was
required to ensure that the sealing of the glass tubes was perfect. This appeared to have
been a delicate step because only about 50% of sealing attempts ended up being
successfully done. Also, due to the low moisture environment inside the glove box, the
glass tubes became highly charged with static electricity. When the Na2S and S
homogenized mixtures were transferred from the mortar to the glass tube, some of the
mixture tended to adhere to the side of the tube rather than settling at the bottom of the
tube. An attempt was made to circumvent this problem by having the powdered mixture
delivered through a rolled 15.24 x 15.24 cm weigh paper (Fisherbrand, Fisher Scientific).
This helped deliver most of the mixture to the bottom of the tube, rather than sticking to
the tube walls. Any of the mixture retained on the walls would not mix with the rest of
the reactant mixture that did make it to the bottom of the tube. That could have reduced
the yields below 100 % of the final Na-polysulfide product. The colors of the Na2S2,
62
Na2S3, and Na2S4 synthesized in our lab were yellow, orange or yellow-orange, and olive
green, respectively (Figure 27). The colors for Na2S2 and Na2S4 agreed with those
reported by Rosen and Tegan (1971). When the polysulfides were ground into a powder-
like form some of the physical characteristics between the different polysulfides were
quite evident. This was the case when the trisulfides were extracted from their glass; they
did not grind into a dry-like powder, but more crystalline and seemed to have more
moisture. This observation, for the trisulfide, did not seem to be affected by the humidity
levels in the lab environment or by the relative humidity levels inside the glovebox. The
disulfide and tetrasulfide compounds ground easily into a dry, powder-like form.
The standardization of the polysulfides through iodometric titration (Section 1.3.2) and
the DEPD method (Section 2.2.2), without first reducing them to S2-, was unsuccessful. It
is believed that the white cloudy precipitate that formed during the use of the two
standardization methods impaired the ability of the apparatus to accurately measure and
standardize the aqueous Na-polysulfides. Specifically, in the DEPD experiment, the
cloudy precipitate blocked the incident light in the spectrophotometer from passing
through the cuvette. To circumvent the problem of precipitate formation, the aqueous
Na-polysulfides needed to be reduced to the simple sulfide ion (HS- or S2-), as described
in section 2.2.3, prior to analysis using DEPD. Such a reduction of the polysulfides to
sulfide also allowed the use of the iodometric titration for the standardization of the
polysulfides.
A comparison of two separate batches of Na-polysulfides following reduction is outlined
below in Table 10. As the results indicate, the purity of the disulfide, trisulfide, and
tetrasulfide prepared in January 2016 was 94%, 87%, and 89%, respectively. This purity
level of the polysulfides was better than of the ones prepared in August 2015 by 9%, 7%,
and 5% for disulfide, trisulfide, and tetrasulfide, respectively. There is a possibility that
the relative humidity at the time of synthesizing the polysulfides could have impacted the
purity of the synthesized polysulfides. There is usually higher humidity in the lab in the
month of August than in January.
63
Table 10. Comparison mean % yields ± SE of reduced Na-polysulfides from two separate
batch products, summer 2015 and winter 2016.
Polysulfide August 2015 January 2016
Na2S2 85 ± 7 94 ± 3
Na2S3 80 ± 3 87 ± 7
Na2S4 84 ± 19 89 ± 5
A typical relative humidity range in August is anywhere between 45 - 55 %; compared to
January relative humidity levels which tend to be between 18 - 20%. An increased level
of relative humidity in the lab environment may lead to an increased level of moisture in
the air. Sodium sulfide and the Na-polysulfides are very hygroscopic. As reported by
Petri and Larachi (2006), excess moisture could lead to loss of these types of compounds
due to oxidation to hydrogen sulfide as a gas. Steps were taken to ensure humidity levels
were reduced as far as possible by running a de-humidifier and split air-conditioning unit.
However, the two external units were only able to reduce the humidity to about 45 % in
the lab space. As a result, even working inside the glove box and manipulating the
samples during the month of August occurred in an environment where the relative
humidity could only be reduced to 30 - 35 %. This is much higher than measurements
made in January when the relative humidity was between 20 - 25 %.
The Na-polysulfides were analyzed using XRD, under conditions as outlined above.
Unfortunately, due to the low signal to noise ratio, the final products could not be
quantified. There was also the presence of some impurities such as disodium sulfite
found within these samples. Some of these impurities did not coincide with the oxidative
products described in appendix B or listed in Table 12. This may suggest the existence of
side reactions during the synthesis of the polysulfides at high temperatures. The
quantification of the side products was not determinable using XRD.
2.3.2 Reduction and Analysis of Na-polysulfides and Polythionates
According to a number of authors (Cline, 1969; Lawrence et al., 2000; Kariuki et al.,
2008; Reese et al., 2011), sulfide, the product of reduction of sulfur containing
compounds with chromium, gets protonated in an acidic medium to give off H2S gas
(Reaction R-2.4). The H2S gas generated, when collected in an alkaline solution, reverts
64
back to S2− which in turn reacts with DEPD to give an intense blue solution. The
intensity of the blue colour, from the DEPD reaction, is proportional to the concentration
of S2− in the solution being analysed.
𝑆2−(𝑎𝑞) + 2𝐻+(𝑎𝑞) → 𝐻2𝑆(𝑔) (𝑅 − 2.4)
Figure 29. Calibration curve of sulfide obtained using the DEPD method (R2 = 0.999).
Table 11. Mean and SE ± for various concentrations of sulfide used to develop a standard
curve for DEPD analyses (n=11)
Concentration (µM) Mean Absorbance ± SE
25 1.5 ± 0.07
15 0.91 ± 0.04
10 0.63 ± 0.03
5.0 0.32 ± 0.02
2.5 0.17 ± 0.01
Figure 29 shows the calibration curve obtained by plotting the concentrations of sulfide
in the DEPD reaction versus absorbance at a wavelength of 670 nm. A correlation with
an R2 value of 0.999 was obtained for the sulfide analysis through the DEPD method. To
account for possible matrix affects, a method blank was prepared using ultra-pure water
and 0.1 M NaOH. Both were separately put through the reduction process and run
through DEPD and iodometric titration. Both the ultra-pure water and 0.1 M NaOH
0
0.2
0.4
0.6
0.8
1
1.2
1.4
1.6
0 3 6 9 12 15 18 21 24 27
Ab
sorb
an
ce a
t 6
70
nm
Concentration of Sulfide (µM)
65
showed no presence of sulfide after analysis through the DEPD and iodometric titration
methods.
Earlier attempts to reduce the aqueous Na-polysulfides, specifically the tetrasulfide
compound, were not very promising. When the reduction was run at room temperature
for 1.5 hours, the yield was ≤ 80 % conversion of the aqueous disulfide and trisulfide
polysulfide compounds to their reduced forms of Sn2-. There was ≤ 60 % conversion of
the tetrasulfide compound under these conditions. After the reduction time had ended,
and the solutions from flask B and C were recovered, an odor of H2S was apparent from
reaction flask A. The presence of this characteristic “rotten egg” odor indicated that not
all of the aqueous Na-polysulfide had been captured in flask B and converted to their
simple sulfide form. This was confirmed after completing the DEPD analysis (section
2.2.3) when the recovery values of Sn2- were shown to have been ≤ 80 % for the disulfide
and trisulfide compounds, and ≤ 60 % for the tetrasulfide. As noted by Kariuki et al.
(2008), the increase in chain length of the aqueous Na-polysulfide makes the reduction of
these compounds more difficult under these conditions. The method was modified in
terms of applying heat to 60 °C, and the reduction of the polysulfides was set up to occur
over a 4-hour period rather than the original 1.5 hours. The heat source was turned off
after 3.5 hours and the N2 gas flow was increased slightly as the reaction solution cooled
to room temperature. This was done to expel any remaining reduced Sn2- from reaction
flask A before retrieval of the solutions for further analyses. The additional heat
treatment and reaction time caused the % yields of recovered Sn2- to increase by 19, 12
and 40% for disulfide, trisulfide and tetrasulfide, respectively. Table 12 summarizes the
% yields of the reduced Sn2- products after applying heat at 60 °C and reducing the
solution for 4 hours. With this modification of the procedure there was no smell of H2S
from reaction flask A after applying the reduction treatment. As indicated below in Table
12, there are some remaining challenges in terms of getting closer to the desirable level of
100 % conversion of reduced Sn2- of the trisulfide and tetrasulfide.
66
Table 12. Mean percent yields ± SE of the reduced polysulfides after applying heat at 60
°C and reducing the solution for 4 hours (n = 5).
Na-Polysulfide Mean ± SE
Na2S2 [S22-] 93 ± 5
Na2S3 [S32-] 85 ± 4
Na2S4 [S42-] 83 ± 5
The percentage purities of the synthesized polysulfides using the DEPD method were in
agreement with those obtained through the iodometric titrations. In order to directly
compare the two sets of results, an Analysis of Variance (ANOVA) was carried out to
compare the effect of each method on reduced sulfide compounds. There was no
significant difference between DEPD and iodometric titration in terms of the
determination of reduced sulfide compounds (F (1,10) = 0.418, p = 0.532). This supports
the conclusion that the DEPD and iodometric titration methods are in agreement, and that
the DEPD method provides a reliable means for determining the concentrations of
reduced sulfur compounds.
2.4 CONCLUSIONS
The several steps required to make the Na-polysulfides can be a source of significant
errors, which in turn may lead to incomplete product formation. Through several batch
adjustments attempts were made to optimize successful product formation. The Na-
polysulfides made in January 2016 have proved the best products to date. The early
round of testing of the final Na-polysulfides in August 2015 was challenging; for
example, when the Na-polysulfides were added to the supporting electrolyte solution
(Appendix A) at a pH of 10.3, a fine black to grey-ish precipitate formed instantly. It has
been suggested that the polysulfide molar distribution is pH dependent, with tetrasulfide
being dominant at pH ≈ 8.0 with trace amounts of trisulfide; and at a pH ≥12.0 trisulfide
disulfide becoming the predominant species (Petre and Larachi, 2006). Also, when
testing the aqueous Na-polysulfides in alkaline media (pH > 12.0) against DEPD method,
with no reduction analysis done, the polysulfide ions in the acidic environment
precipitated a white cloudy substance that only allowed for a slight blue coloration to
form. Sonne and Dasgupta (1991), showed that where H2Sx has a value of x > 1, the
67
compounds are unstable in acidic solutions, thereby producing elemental sulfur and
hydrogen sulfide. It is likely that this was the case when the Na-polysulfide prior to the
reduction process were subjected to the DEPD and iodometric titration methods.
An improvement that can be made to the study of reduced aqueous Na-polysulfides is to
vary the pH of the trap solution (Flask B) or vary the pH during the initial mixing of the
Na-polysulfides prior to conducting the reduction process. Heating the reaction flask (A)
increased the rate of conversion (Table 12). Another consideration could be the use of
other reducing agents besides chromium that would have the potential to increase the
conversion of Na-polysulfides to simple sulfide (HS- or S2-).
The exact purity of the Na-polysulfides could not be determined though XRD analysis.
Some of the deviation from the high purity of the synthesized polysulfides as determined
through the XRD analysis could have been as a result of a significant time delay from
when the samples were synthesized to the time when they were analyzed through XRD. It
is likely that side reactions could have been occurring during storage of the samples. In
future, the ideal situation would be the XRD analysis of the polysulfides shortly after
their synthesis.
A drawback was recognized when using the DEPD (section 2.2.3) and iodometric
titrimetric (Section 1.2.3) methods for the reduced Sn2- or SnO2
2- solutions. The two
methods do not allow differentiation between the different reduced Sn2- or SnO2
2- species
if they are all present in solution at the same time. A further investigation of the methods
of analysis for the polysulfides and polythionates might lead to the speciation of these
compounds. In the following chapter, electrochemical analyses will be used to
investigate whether these aqueous Na-polysulfides and polythionates can be readily
differentiated.
68
Chapter 3
Electrochemical Analysis of Na-polysulfides and Polythionates
3.1 INTRODUCTION
Direct determination of polysulfides and polythionates still remains a challenge. This is
partially due to their high redox reactivity in the natural systems 10 - 20 µM (Kamyshny,
2006; Chadwell et al., 2001). Polysulfide ions are subject to autoxidation when present
in solution. This autoxidation reaction tends to be rapid, when in solution, but when the
polysulfides are in a solid state the reaction is slower (Kamyshny, 2006; Steudel, 2003;
O’Reilly et al., 2001; Sonne and Dasgupta, 1991). Most of the previous attempts to
distinguish and separate polysulfide and polythionates in environmental samples use
capillary electrophoresis (Petre and Larachi, 2006), ion chromatography (Miura et al.,
2005), titrimetric methods (Kamyshny et al., 2004), and electrochemical analyses (Manan
et al., 2011; Kariuki et al., 2001; Rozan et al., 2000b). Ion chromatography and capillary
electrophoresis often require pH changes, which could alter the speciation of sulfur based
compounds present in samples. Derivatization using substances like methyl iodide are
sometimes required for analysis of samples using the above techniques. The uses of
modifiers, such as a chromate electrolyte in a solution of hexamethonium bromide, have
also been used towards the determination of polysulfides and polythionates in samples.
These changes to pH, derivatizations, and added modifiers can alter the relative
abundance of the sulfur-based species originally present in a given environmental sample.
For example, S2O32- under acidic conditions will decompose to elemental sulfur and
sulfite (SO32-) (Ciglenečki and Ćosović, 1997). This distribution change seems to alter
the sulfur-based compound speciation in the original sample. Kamyshny et al. (2004),
reviewed the various published attempts of the polysulfide speciation, concluding that the
speciation and distribution of polysulfides and polythionates seem to vary widely from
lab to lab. These authors also pointed out that not one lab used the same technique or set
of techniques to determine sulfur based compounds in a consistent manner. However,
these authors also identified the thermodynamic constants (pKn) for polysulfide
disproportionation which could be useful in understanding polysulfide and polythionate
69
kinetics in solution. They also developed a new approach to determining the polysulfide
speciation in aqueous media. Their results show a narrower range of disproportionation
to polysulfide speciation by accounting for variations of the pH of the solutions that the
polysulfide was dissolved in. Their results support the findings of Petre and Larachi
(2006), where they classify the abundance of the polysulfide species present in solution
over a pH range from 8.0 to 12.0.
Electrochemical analyses for the detection of polysulfides using mercury drop electrodes
(Chadwell et al. 2001; Rozan et al. 2001) showed that there is an interaction between
polysulfides and bisulfides with divalent cations such as iron, nickel, cobalt, copper and
zinc. Both groups of authors concluded that a metal complex was formed through the
decomposition of the polysulfide to bisulfide in solution, rather than through direct
complexation with the polysulfide ion. This might suggest that in characterizing
environmental samples, one should be aware that the matrix of the solution affects the
determination of polysulfide concentrations. Also, other authors who have used the
polarographic electroanalytical techniques for the analysis of sulfur compounds have
observed a significant interference from higher order polythionates (Steudel 2003;
Kariuki et al. 2001; Rozan et al. 2000).
This chapter reports on the electrochemical analyses of aqueous solutions of the Na-
polysulfides (Na2S2, Na2S3 and Na2S4). In addition, the analysis of thiosulfate as a
representative of the class of polythionate compounds is also presented. Cyclic
voltammetry using both the modified and the unmodified glassy carbon electrodes used
for the characterization of the polysulfides and polythionate in alkaline media is also
discussed. The electrochemical analyses for the polysulfides and polythionate presented
here will demonstrate that the mixture is best analyzed when the sulfur compounds get
converted to simple sulfide.
70
3.2 MATERIALS AND METHODS
3.2.1 Equipment
Electrochemical measurements were conducted using the BASi work station with Epsilon
USB software (Bioanaytical System Inc.). The three-electrode cell consists of a C3 glass
cell vial 50 mm x 59 mm housing the supporting electrolyte solutions, and electrodes for
analysis. The working electrode was a glassy carbon electrode with a working surface
diameter of 3.0 mm (area = 0.017 cm2). The auxiliary electrode used was a 23-cm long
coiled platinum wire. An Ag/AgCl probe was used as the reference electrode. All
electrodes were supplied by BASi. In each case the electrochemical measurements
(section 1.2.4.5) of the Na-polysulfides, the polythionate, and reduced polysulfides and
the polythionate solutions were conducted using cyclic voltammetry. Cyclic
voltammetric cycles were operated by scanning the electrode potential from -1600 mV to
+1600 mV, at a scan rate of 100 mVs-1 for 10 cycles. The peak potential (mV), height
(µA), and area (µC) were recorded for all 10 cycles. Three sets at 10 CV cycles per
replicate were analyzed for each experiment. MGCE experiments were also carried out
on the above solutions using the procedures outlined in section 1.2.4.4.
3.2.2 Dissolution of Na-polysulfides and Thiosulfate Ions
3.2.2.1 Na-Polysulfides
The final products selected for electrochemical analysis of the Na-polysulfides that were
synthesized as outlined in section 2.2.2.5 were those that generated the highest percent
yield following the chromium-reduction process described in section 2.2.4. The weights
of the various Na-polysulfides that were used are shown in Table 13. Each compound
was dissolved in 20 mL 0.1 M NaOH as described in section 2.2.3.1, stored in a 30 mL
Teflon storage bottle (Nalgene®, Fisher Scientific), and placed in the fridge at 4 °C until
required. Due to the instability of the Na-polysulfides, once they were dissolved, they
had to be used within 48 hours. All procedures described here were carried out inside the
Baxter Glovebox as described in section 1.2.1.2.
71
Table 13. Weights (g) and concentrations (mM) of Na-Polysulfides used for CV analyses
Na-Polysulfide Weight (g) Concentration (mM) in 20 mL 0.1 M
NaOH Na2S2 0.149 67.8 Na2S3 0.175 61.4 Na2S4 0.175 50.1
3.2.2.2 Polythionates (S2O32-)
The thiosulfate ion was used to represent the polythionates group of sulfur compounds.
For the thiosulfate ion, 0.539 g of sodium thiosulfate was weighed and dissolved in 20
mL of ultra-pure water, placed in a 20 mL glass scintillation vial (Fisherbrand, Fisher
Scientific), and stored in the fridge at 4 oC until required later for analyses. The
thiosulfate concentration in this solution was 109 mM. The thiosulfate ion is not readily
oxidized by exposure to air, so the preparation (weighing and dissolution process) was
carried out on the bench top rather than in the glove box.
3.3 RESULTS
3.3.1 Electrochemical Analysis of Na-polysulfides and Polythionates at
BGCE and MGCE
Cyclic voltammetry was performed using each of the BGCE and MGCE for the aqueous
Na-polysulfides (Na2S2, Na2S3, Na2S4), the polythionate (thiosulfate), as well as for the
reduced polysulfide solutions (section 2.2.4). For the BGCE experiments, the three Na-
polysulfides analyzed are shown in Figure 27. The Na2S2 showed no peak current
response through the potential scan from -1600 to +1600 mV. Indeed even when the
Na2S2 concentration was doubled relative to that shown in Figure 30 (1.39 mM), no
current response was observed.
72
Figure 30. CV graph at a potential scan rate of 100 mV s-1 for S22- [1.39 mM], S3
2- [1.31
mM] and S42- [1.22 mM] at BGCE. With supporting electrolyte solution as the blank
(dashed line).
It is unclear why there is no CV response for the S22- ion. Some researchers have
suggested that at extremely high pH values (≥ 12.0) the disulfide ion becomes the
predominant species (Petre et al., 2006 and Steudel, 2003). However, in this
electrochemical analysis the pH of the supporting electrolyte solution was 10.3, which is
significantly below the pH of 12.0 that they were considering. Neither our own work, nor
other published reports clearly address whether the disulfide ion breaks down to other
sulfur species when the solution is at a pH < 12.0. It was, however, observed that after
the analysis on the BASi equipment, the electrochemical cell containing the disulfide ion
released the characteristic odor of H2S. This may indicate that a reduced volatile sulfide
was present, but perhaps not in quantities that could be detected by this particular
analytical method. It is worth noting that humans are particularly sensitive to the odor of
H2S, and some people can detect the gas at levels as low as 0.47 ppb. There were,
however, clear current responses for both S32- and S4
2- ions (Figure 30), occurring in the
-100
-75
-50
-25
0
25
50
75
100
125
150
175
-1600 -1200 -800 -400 0 400 800 1200 1600
I /
µA
E / mV vs. Ag/AgCl
blank
S22-
S42-
S32-
73
anodic and cathodic segments. The values for their potential position(s) and current
response(s) are summarized in Table 14 and Table 15 with the values for HS- ion added
for comparison.
Table 14. Anodic segment peak potential (mV) and peak current (µA) for CV analysis of
Sn2- ions (1 mM) at BGCE (n = 3).
Na-Polysulfide Epa (mV) Ipa (µA)
S22- ― ―
S32- -8, +1050 17, 30
S42- -15, +1055, +1339 21, 28, 18
HS- -10, +412, +1114 18, 69, 87
Table 15. Cathodic segment peak potential (mV) and peak current (µA) for CV analysis
of Sn2- ions (1 mM) at BGCE (n = 3).
Na-Polysulfide Epc (mV) Ipc (µA)
S22- ― ―
S32- -1135 43
S42- -1100 71
HS- -1200 115
In Appendix A, the table for oxidation states of various sulfur compounds, provides some
insight as to which species could be produced through the oxidation of S32- and S4
2- ions.
Some of the potential positions for both the anodic and cathodic segments shown in
Figure 32 match the CV analyses conducted on sulfide in Figure 3. However, there is a
less marked rise in the current baseline in the tri- and tetrasulfide CV graphs at the Epa
positions at ≈ +1060 and +1340 mV, compared to the response observed for the sulfide
CV (Figure 31). The current responses in the tri- and tetrasulfide CV graphs are easier to
integrate at potentials of ≈ +1060 to +1340 mV due to the less marked rise in baseline
current in these regions of high potential. The current responses of HS- ion at potentials
of +1600 mV are twice as high as the current response of the tri- and tetrasulfides at a
potential of +1600 mV (Figure 31). All of the concentrations examined for the ions of
Sn2- and HS- comparison were close to 1 mM.
74
Figure 31. CV graph of the comparison of simple sulfide ion with di-, tri- and tetrasulfide
ions at a potential scan rate of 100 mV s-1 for HS- [1 mM], S22- [1.39 mM], S3
2- [1.31
mM] and S42- [1.22 mM] at BGCE. With the supporting electrolyte solution as the blank
(dashed line).
When thiosulfate was analyzed at the BGCE there were only two peak current responses
in the anodic segment of the CV trace. There were no peak(s) present in the reductive
segment (Figure 32). A 1.1 mM solution of S2O32- in the electrochemical cell gave
potential positions Epa = +1037 and +1415 mV with a current response of Ipa = 42 and 59
µA, respectively. A check was done to see if the S2O32-
could, on its own, be
quantitatively determined on a BGCE (Figure 33). This was, indeed, found to be the
case. However, when thiosulfate was combined with sulfide, the potentials at which
oxidation of the two species appeared were indistinguishable (Figure 34). Overall the
results shown in Figure 32 and Figure 33 suggest that thiosulfate can be detected on its
own, but when it is mixed with sulfide its response cannot be readily distinguished from
that of the HS- ion. In contrast with that result, Figure 32 and Figure 35 show that when
-150
-100
-50
0
50
100
150
200
250
300
350
-1600 -1200 -800 -400 0 400 800 1200 1600
I /
µA
E / mV vs. Ag/AgCl
HS-
blank
S22-
S32-
S42-
75
S2O32- is mixed in with a Sn
2- such as S32-, the S2O3
2- can still be individually
distinguished from the baseline current.
Figure 32. CV graph of the comparison of quantitative amounts of S2O32- at a potential
scan rate of 100 mV s-1 for S2O32- [1.06 mM], and 2 x the amount of S2O3
2- [2.12 mM] at
BGCE. With the supporting electrolyte solution as the blank (dashed line).
-50
-25
0
25
50
75
100
125
150
175
200
225
250
275
-1600 -1200 -800 -400 0 400 800 1200 1600
I /
µA
E / mV vs. Ag/AgCl
blank
S2O32-
(2X) S2O32-
76
Figure 33: CV graph of the comparison of polysulfide ions with thiosulfate ion at a scan
rate of 100 mV s-1 for S2O32- [1.06 mM] (dotted line), S2
2- [1.39 mM], S3
2- [1.31 mM] and
S42- [1.22 mM] at the BGCE. With supporting electrolyte solution as the blank (dashed
line).
-100
-75
-50
-25
0
25
50
75
100
125
-1600 -1200 -800 -400 0 400 800 1200 1600
I /
µA
E / mV vs. Ag/AgCl
blank
S2O32-
S22-
S42-
S32-
77
Figure 34. CV graph of the comparison of thiosulfate ion, sulfide ion and a combination
of thiosulfate ion and sulfide ion. At a potential scan rate of 100 mV s-1 for S2O32- [1.06
mM], sulfide [1 mM], and mixture of HS- + S2O32- at BGCE. With the supporting
electrolyte solution as the blank (dashed line).
According to Steudel (2003) polysulfide anions are subject to autoxidation in the
presence of molecular oxygen. Reaction R – 3.1 shows the S42- example. In the presence
of molecular oxygen, polysulfides autoxidize to form thiosulfate and sulfur (R – 3.1).
Even though measures had been taken in this experiment to exclude oxygen, the
potentials that are used in this work can help further explain the source of oxygen that
may readily react with polysulfides, as shown in reaction R – 3.1. Anodic oxidation of
sulfide can yield products like elemental sulfur, polysulfides, thiosulfates and sulfates
(Al-Kharafi et al. 2010). This may explain the two peaks observed at potentials of +1050
and +1339 mV respectively (Figure 30 and Table 14).
𝑆4 2− +
3
2 𝑂2 → 𝑆2𝑂3
2− + 1
4 𝑆8 (𝑅 − 3.1)
This proposed mechanism and explanation is further supported by the results of Lessner
et al. (1993), where spectroscopic and chemical investigations revealed that in alkaline
-150
-100
-50
0
50
100
150
200
250
300
350
400
450
-1600 -1200 -800 -400 0 400 800 1200 1600
I / µ
A
E mV Vs. Ag/AgCl
(HS- + S2O32-)
HS-
S2O32-
blank
78
pH ranges, the polysulfide and bisulfide species predominate. This can be represented by
the following reaction (R – 3.2):
4𝑆42− + 8𝑂𝐻− + 𝐻2𝑂 ↔ 3𝑆2𝑂3
2− + 10𝐻𝑆− (𝑅 − 3.2)
According to the reaction R – 3.1, as the CV cycles are continuously run on the BGCE, in
the presence of sulfur-based compounds, elemental sulfur builds up on the working
surface of the electrode, thereby reducing the area of activity. It should be noted that
thiosulfate on its own does not exhibit a reductive segment. This may suggest that in the
case of thiosulfate there is no reduction to form HS- and that the thiosulfate ion is
oxidized to tetrathionate (S4O62-). According to O’Reilly et al. (2001) and Koh (1990),
there is no production of elemental sulfur in the electroxidation process that could impair
the working surface of the glassy carbon electrode.
Figure 35. CV graph of the comparison of thiosulfate ion, trisulfide ion and a
combination of thiosulfate ion and trisulfide ion at a potential scan rate of 100 mV s-1 for
S32- [1.31 mM] and a mixture of S3
2- [1.31 mM] plus S2O3
2- [1.06 mM], and S2O32- [1.06
mM] at the BGCE. With supporting electrolyte solution as the blank (dashed line).
-60
-20
20
60
100
140
180
220
-1600 -1200 -800 -400 0 400 800 1200 1600
I /
µA
E / mV vs. Ag/AgCl
blank
S2O32-
S32-
S32- + S2O3
2-
79
As illustrated in Figure 36, the analysis of the Sn2- ions at MGCE’s exhibit similar
behavior to that of HS-. After background corrections were adjusted at Epa ≈ +83 mV
(LOD = 28 µA at MGCE), the HS-, S22-, S3
2-, S42- ions have a current response of 50 µA,
22 µA, 35 µA, and 43 µA respectively. It should be noted that there is a shift in potential
to the left of the Sn2- and S2O3
2- ions relative to the blank reference at an Epa of +249 mV.
However, there is no apparent shift in potential for the current responses located at a
mean Epa of +1138 mV (n = 5) relative to the blank at a Epa of +1107 mV. Based on the
proposed formation of a complex of V2O5 ·HS- at MGCE (Khudaish and Al-Hinai 2006)
(Equation R – 3.3), the interactions of the Sn2- with the MGCE surface may be explained.
𝑉2𝑂5 ∙ 𝐻𝑆− → 𝑉2𝑂4 + 𝑆 + 𝑂𝐻− (𝑅 − 3.3)
With regard to the activity of the Sn2- ions at the MGCE, the present study may suggest
that the reaction that may have occurred can be described by the reaction R – 3.4. This
hypothesis is based on the mechanism proposed by Khudaish and Al-Hinai (2006), and
shown as R – 3.3.
𝐻2𝑂 + 𝑉2𝑂5 ∙ 𝑆𝑛2− → 𝑉2𝑂4 + 𝑛𝑆 + 𝑂𝐻− (𝑅 − 3.4)
Figure 36 clearly shows that it would be very difficult to attribute the peak responses
from a series of CV scans on an MGCE, separately for HS-, S2O32-, and Sn
2- ions present
in a mixture. This observed overlap with all these sulfur anions poses a problem for the
application of this modification agent towards environmental and industrial samples.
Samples that come from environment and industrial waste basins would be more likely to
contain a mixture of these substances rather than pure solutions that were individually
analyzed in Figure 36.
80
Figure 36. CV graph at potential scan rate at 100 mV s1- at the MGCE for S22-, S3
2-, S42-,
S2O32-, and HS- (dotted line). With the blank as the supporting electrolyte solution
(dashed line).
3.4 CONCLUSIONS
At the BGCE, S32- and S4
2- ions can be detected, but they do not show unique, distinctive,
or characteristic potential positions during CV cycles. According to Lessner et al.
(1993), thiosulfate can be a side product formed from the dissociation of polysulfide ions
in solution. Thiosulfate, when mixed with a polysulfide, appears to show the same
potential positions, between +1000 to +1300 mV (Figure 32). The thiosulfate ion can be
distinguished when it is in a pure solution rather than being part of a mixture, and it is
unique in that it has no reductive segment. The CV results of the thiosulfate ion at the
BGCE suggests that under the CV conditions described above, only the oxidation peak,
and not the reduction peak, is observed. This observation also agrees with that of several
other authors (Zanello et al., 2012; Scholz, 2010; Kissinger and Heineman, 1996). As
observed in this chapter, there are several potential positions that overlap in the CV
-100
-75
-50
-25
0
25
50
75
100
125
150
175
200
-1600 -1200 -800 -400 0 400 800 1200 1600
I / µA
E / mV vs. Ag/AgCl
S22-
S32-
S42-
S2O32-
HS-
blank
81
graphs for the S32- and S4
2- ions. There are several potential positions that overlap for the
S32- and S4
2- ion CV graphs. This overlap in potential positions makes these two
compounds hard to distinguish, which would pose a problem when analyzing
environmental samples since these samples would likely contain a mixture of these
compounds. The problem is particularly apparent when Sn2- ions are combined with HS-
(Figure 31 and Figure 34). There is a catalytic response to Sn2- and S2O3
2- ions at
concentrations of about 1 mM, but their potential positions did not vary enough during
CV analyses to allow for distinction between these two classes of sulfur species. The
interference of the supporting electrolyte solution during the experiment at MGCE would
suggest that at concentrations lower than 1 mM for the Sn2- and S2O3
2-, the current
response of these ions would be unquantifiable. However, this phenomenon was not
observed in reactions carried out using BGCE. As was the case with the BGCE, the
potential positions and current responses of the Sn2-, HS-, and S2O3
2- ions were very
similar and therefore, indistinguishable under the CV conditions used in this work. The
unexpected fluxes in terms of the increases and decreases in the current recorded over a
series of ten CV cycles at the MGCE (Figures 22 to 24) may need further investigation.
Earlier studies have successfully determined quantifiable polysulfides in solution
mixtures in combination with thiosulfate and sulfide ions (Kariuki et al. 2001; Rozan et
al. 2000). However, these analyses were carried out using a mercury drop electrode,
which provides a fresh working surface with each sequential drop. With the mercury
drop electrode, there was no need for manual manipulation or electrode cleaning as was
the case with the glassy carbon electrodes used in this study, where the surface had to be
polished between analyses.
Although the polysulfides could not be distinguished from each other using the BGCE
and MGCE, valuable qualitative data was gathered regarding the individual ions and how
they might interact during CV analyses.
82
Chapter 4
Analysis of Low Oxidation-State Sulfur Species in Water
4.1 INTRODUCTION
Sulfur compounds are prevalent in environmental and industrial processes. Substances
such as sulfide, thiosulfate, polysulfides, sulfate and sulfite can be generated by natural
processes (biological or geochemical) or by anthropogenic activities such as petroleum-
based or mining industries (Casella et al. 2002). Identification and quantification of these
sulfur compounds is important due to the health risks associated with exposure to
hydrogen sulfide and the effects these compounds have in aquatic systems (Kariuki et al.
2008; Lawrence et al. 2007; Lawrence et al. 2002; Witter and Jones 1998; Zhang and
Millero 1993). As previously stated, the levels of dissolved H2S at 0.01 mg L1- have been
reported to have a chronic toxicity to aquatic organisms (Brouwer and Murphy, 1995).
Levels of dissolved H2S approaching 0.5 mg L1- exhibit acute toxicity to rainbow trout
and salmon (Ortiz et al., 1993). The sulfide anion can be rather versatile in that it forms
intermediate species of sulfur based compounds. These intermediate sulfur based
compounds include but are not limited to: thiosulfate, sulfite, sulfate, and acid volatile
sulfides. According to Gun et al. (2000), polysulfides are an integral part of the sulfur
cycle and they are required for the formation of volatile sulfur compounds, which are a
group of organic volatile sulfur species such as dimethyl sulfide. Also, the polysulfide
ions have the ability to disassociate to form reduced inorganic sulfur species like those of
HS- and H2S during the organic decomposition of R-SH groups prompted by sulfur
reducing bacteria (Gosh and Dam, 2009; Kamyshny et al., 2008; Kamyshny et al., 2006;
Friedrich et al., 2001). The standard detection methods in use were outlined in an
extensive review by Lawrence et al. (2000) and were briefly summarized in the
background information. Among those detection methods, electrochemical analyses have
become more popular for the detection of environmental contaminants such as sulfide
species. Examples of electrochemical methods used for sulfide determination are
potentiometry (using an ion-specific electrode), voltammetry and amperometry (Ardelean
et al., 2014).
83
In this chapter, a water sample collected from Callander Bay, Ontario was examined
using the methods described in the earlier chapters. The electrochemical, iodometric
titration and ethylene blue analyses employed earlier have set the frame work for testing
the efficiency of these methodologies towards environmental samples to determine if
sulfide, polysulfides and polythionates can be detected at environmentally relevant
concentrations.
4.2 MATERIALS AND METHODS
4.2.1 Water Source and Collection Parameters
Water was collected from a site in the middle of Callander Bay of Lake Nipissing,
Ontario. This site has been identified by the Ontario Ministry of the Environment
(OMOE) as site IG9 (Easting: 624587 Northing: 5119449) where the water is 9 – 10 m
deep (OMOE Oct 6, 2015). The temperature of the water, dissolved oxygen
concentration (DO) and the pH of the water were 13.8 °C, 75.5% (7.88 mg/L), and 6.5,
respectively. The total watershed area for Lake Nipissing, including Callander Bay, is
296.12 km2 of which 83% is the Wasi watershed (Karst-Riddoch 2011). Samples for two
one-litre high-density polyethylene (HDPE) bottles pre-purged with N2 gas and sealed,
were collected using a Van Dorn water sampler to a depth of 10 m. The water sample
was filled to the very top of the HDPE bottle to ensure no headspace was left. The two
1-L HDPE sample bottles were returned to the lab and stored in the fridge (4 ºC) for no
longer than 48 hours before being repackaged and frozen until needed for further
analyses. The contents of the bottles were then split up into 15 individual 125-mL HDPE
bottles. In the Baxter Glove Box, under an atmosphere of nitrogen, the individual bottles
were sealed and frozen for subsequent analyses. Three 50-mL sub-samples of the water
were portioned into 125-mL flasks to be used for reduction and DEPD analyses. Those
samples were carefully sealed and stored in a fridge (4 ºC) until the analyses could be
completed. The three 50-mL samples were also prepared in the Baxter Glove box as
described above. A subsample of the lake water, from site IG9, was filtered (0.45µm)
and sent out for Ion-coupled Plasma Atomic Emission Spectroscopy (ICP-AES) analysis
at an external lab for trace inorganic metal analysis.
84
4.2.2 Reduction and DEPD Analysis of Lake Water Sample
Using the procedures for the reduction of sulfur compounds with chromium as described
in section 2.2.4, three replicates of the unfiltered water samples were first reduced
without any spiking of a known concentration of sulfide. Also, using the same
procedure, another triplicate set of untreated water samples was then spiked with a
known concentration of the sulfide stock solution and was reduced with chromium
(section 1.2.1.2). Similarly, another set of water samples was acidified without the
addition of chromium and analyzed using the same procedure as the above two sets of
water samples. The last two sets of water samples were being tested in order to detect the
presence of any acid-labile sulfides in the water sample. The procedure above was
repeated on filtered lake water samples. The filtration of the water samples was carried
out through a 0.45-µm membrane. All the three sets of replicates for the filtered and
unfiltered water samples were tested using DEPD, following the procedure outlined in
section 2.2.3, to detect the presence of sulfide in the lake water. A control group, using
ultra-pure water, with combinations of chromium and sulfide was analyzed using this
procedure to determine if there were any matrix interferences.
4.2.3 Electrochemical Analysis of Lake Water Samples
Electrochemical measurements were conducted using the BASi work station with Epsilon
USB software (Bioanaytical System Inc.). The three-electrode cell consists of a C3 glass
cell vial 50 mm x 59 mm, housing the supporting electrolyte solutions, and electrodes for
analysis. A glassy carbon, with a working surface diameter of 3.0 mm (area = 0.017
cm2); was used as the working electrode. The auxiliary electrode, Pt wire, in the form of
a 23-cm long coil was used and the reference electrode was a Ag/AgCl. All electrodes
were supplied by BASi. The electrochemical set up for the analysis using BASi is shown
in Figure 1. Electrochemical measurements (section 1.2.4.5), conducted using CV, where
the electrode potential was cycled from -1600 mV to +1600 mV at a scan rate of 100 mV
s-1 for 10 cycles. The peak potential (mV), height (µA), and area (µC) were recorded for
all 10 cycles. Three sets at 10 CV cycles per replicate were performed for each
experiment for BGCE and MGCE. The unfiltered lake water sample was examined using
85
the procedures outlined above; also the sulfide spiked lake water sample was examined.
The reduced lake water sample described in section 4.2.1.2 was also assessed.
4.3 RESULTS
4.3.1 Reduction and DEPD Analysis of Lake Water Sample
The reduction of the lake water samples with chromium and their analysis using the
DEPD methodology was carried out on a couple of water samples, chosen to represent a
freshwater sample that was collected from Lake Nipissing. The present study was
designed to test current methodologies outlined in the earlier chapters in order to
determine whether these methods would be able to detect sulfur compounds at
environmentally relevant concentrations. It was also an opportunity to see if individual
sulfur compounds could be distinguished and to see if natural samples showed any form
of matrix interferences.
The first subsample of lake water was reduced without being spiked with simple sulfide.
Two reduction procedures were carried out, the first without the addition of chromium,
and the second with chromium being added to the water sample. Also, filtered and
unfiltered samples were analyzed. When these reductions were completed, there were no
detectable traces of sulfide observed using the DEPD method (section 2.2.2). It is worth
noting that when the lake water samples were reduced without the addition of chromium,
a white precipitate formed at the bottom of the reaction flask (A), shown in Figure 26.
From the results indicated in Table 16 and Table 17 for water and sediments collected
from site IG9, there are very few of the divalent cations that are found in the water
sample (Table 16). However, as indicated in Table 17 provided by Chase et al. (2016),
there are several of the key divalent cations in the sediment samples from site IG9.
According to Kaasalainen and Stefánsson, (2011); Kariuki et al., (2008); Rozan et al.,
(2000) and Janssen et al., (1999), upon acidification of samples with metallic sulfides
such as, FeS2, PbS, Ag2S, MnS and HgS, there is a production of a white precipitate
forming during the reduction processes. Acidification of a sulfur containing sample
promotes the dissociation of any acid-volatile sulfide complexes present in the water
86
sample (Brouwer and Murphy, 1995). Such metal sulfide complexes have been shown to
not only increase the rate of oxidation of sulfide but also to change the distribution
products formed during sulfide-oxidation (Zhang and Millero, 1993). It has been
reported that metal sulfide complexes tend to occur for the bisulfide (HS-) of the divalent
cations like iron, cobalt, nickel, copper, and zinc (Chadwell et al., 2001; Al-Farawati and
van den Berg, 1999; Zhang and Millero, 1994). Since the pH range of most aquatic
systems falls in the slightly acidic to slightly basic range, and according to the fractional
dissociation for sulfide as a function of pH (Figure 2), the HS- ion is the most dominant
and readily available species. In Table 17 the presence of inorganic metals found in fresh
water sediments have been shown to form metal sulfide complexes, and are responsible
for controlling the solubility of trace metals in anoxic waters, and are also thought to be
important for the stabilization of H2S in oxic surface waters (Al-Farawati and van den
Berg, 1999; Zhang and Millero, 1994). Further examination of Table 17 indicates that
the values for iron and sulfur are above the limits set out in the guidelines by
Environment Canada for sediment samples. It has been suggested by Rozan et al., (2000)
that in natural water samples, iron sulfides are the most abundant, and that iron sulfides
diffuse from the anoxic sediments into the water column. Although the iron content in
the water sample, from IG9, was below the limit of detection along with the other metals
(Table 16), there was a likelihood of presence of sulfur in the form of sulfide complexed
with the metals. The presence of sulfur, found in the water sample at IG9, is a possible
cause for the white precipitate that formed at the bottom of reaction flask (A) (Figure 26)
after carrying out the chromium-oxidation reaction A possible reaction that may have
occurred to generate the precipitate is shown Reaction R – 4.1 (Bouroushain, 2010).
𝐻2𝑆(𝑔) → 𝑆(𝑠) + 2𝐻+ + 2𝑒− (𝑅 − 4.1)
87
Table 16. Inorganic metal concentrations (ppm) from site IG9 for water sample collected
in October of 2015.
Site Aluminum Barium Copper Iron Lead Magnesium Nickel Zinc Sulfur
IG9 ― 0.012 0.005 ― ― 1.99 ― ― 1.59
*CCME 0.005 ― 0.002 0.300 0.007 ― 0.025 0.030 ―
*CCME: Canadian Council of Ministers of the Environment, water quality data table for
the protection of aquatic life. (―) indicates no data or below limit of detection.
Table 17. Sediment concentrations (ppm), of the metals (and sulfur) associated with
smelting of nickel ores. Mean values (mg/kg ± standard error) (Chase et al. 2016,
unpublished results).
*ECG indicates Environment Canada Guidelines for sediment for each element (also
expressed in mg/kg)
According to several authors Kaasalainen and Stefánsson (2011); Kariuki et al. (2008);
Rozan et al. (2000) and Janssen et al. (1999), the addition of chromium to reaction flask
(A), in combination with HCl, promotes the dissociation of the metal sulfide complexes
(PbS, MnS, HgS, etc.). This in turn increases the recovery of simple sulfides in the
sample being tested (Rozan et al. 2000). However, Cainfield et al. (1986) and Luther III
et al. (1985) had earlier suggested that the addition of chromium would not promote the
dissociation of metal sulfide complexes of PbS, MnS, HgS, etc. In the present study,
after the chromium was added to the water sample and reduced again, there was still no
detectable sulfide present. Initially, aliquots of 1, 5 and 10 mL of the water samples
treated with chromium were analyzed through the DEPD measurement. However, since
there was no detectable sulfide in either of these sample volumes, a 40-mL aliquot of a
water sample treated with chromium was analyzed for the presence of sulfide through the
DEPD method. Despite the increased sample volume for both with and without added
chromium, the DEPD analyses still did not detect any sulfides in the water sample.
However, when the water sample was spiked with a known concentration of sulfide and
analyzed for sulfide recovery both with and without chromium, there were positive
Site Iron Nickel Copper Cobalt Arsenic Lead Zinc Sulfur
IG9 32,987
(629) 47
(0.5) 40
(0.3) 15
(0.2) 6
(0.4) 48
(0.7) 166 (1)
1038 (18)
*ECG 20,000 16 16 22 6 31 120 1000
88
results for the detection of sulfides using the DEPD and the iodometric titration methods,
as outlined below in Table 18.
Table 18. Comparison of percent recoveries and recovered concentrations (mM) of
sulfide spiked water samples for filtered and unfiltered samples. Experiment
Sample and Treatment DEPD (mM) % Recovery Iodometry (mM) % Recovery
Duchesnay Creek (UF) + S2- +Cr 3 85 3 78
IG9 (UF) + S2- + Cr 0.2 4 BLD ―
Duchesnay Creek (UF) + S2- 3 67 2 47
IG9 (UF) + S2- 0.1 4 BLD ―
Duchesnay Creek (F) + S2- + Cr 4 108 4 105
IG9 (F) + S2- + Cr 5 113 3 82
Duchesnay Creek (F) + S2- 3 72 2 53
IG9 (F) + S2- 4 98 3 74
Type I water + S2- + Cr 4 91 3 90
Type I water + S2- 2 56 2 41
UF = Unfiltered; F = Filtered (0.45 µm); BLD = Below Limit of Detection
The results in Table 18 indicate that the addition of chromium to the reaction flask A
(Figure 28) enhances the overall recovery of sulfides in the natural water samples
collected from the two different sites. This observation holds both for the filtered and
unfiltered natural water samples. The results in Table 18 also indicate that there are
matrix effects encumbered in natural water samples. According to the results, the matrix
effects were more enhanced in the unfiltered water samples since the sulfide recovery,
when compared to that of the filtered water samples, was lower. As stated above, even
though the addition of chromium to unfiltered water samples enhanced the recovery of
sulfides, the recovery of the sulfides in natural water samples is enhanced when the water
sample is filtered. This increase in sulfide recovery is in agreement with the results of
Mylon et al. (2002), Kariuki et al. (2008) and Rozan et al. (2000) who showed that the
addition of chromium to acidified water samples containing sulfides could cause the
transition metal sulfide complexes, if contained in the water samples, to release the
chromium-labile sulfides.
89
The collection of the IG9 water sample (Table 18) was done in early October of 2015,
and initial recoveries of unfiltered water samples yielded very low percent recoveries of
sulfide. Even after the sample’s treatment with chromium, low percent recoveries were
still observed. Based on the chemistry of typical field samples, and as described in
section 4.2.1, collection was after the fall overturn event, when stratification of water at
different temperatures and densities breaks down and water from the hypolimnion
(deeper zone) mixes freely with water in the epilimnion (Dodson, 2005). As a result, at
the depth of 10 m where the sample was collected, the water sample would have been
relatively well-oxygenated, despite the fact that earlier, during summer stratification, the
deep water region can become severely anoxic. Since the initial IG9 water sample was
unfiltered, and the fall-overturn had occurred, there is a good possibility that the
sediment-water interface has mixed. This mixing alters the chemistry at the sediment-
water interface, which would stir up fine colloid suspensions, chemotrophs, and bacteria;
thus could impair the conversion of simple sulfides in a natural water sample (Table 16
and Table 17). However, the conversion of sulfides was greatly increased when the water
sample was filtered prior to reduction in both the treatments applied.
The results in Table 18 indicate that there were elevated percentage yield recoveries with
some water samples that were treated with chromium over those that did not get the
chromium treatment. This could suggest that some sulfur-based intermediates such as
polysulfides, thiosulfate or tetrathionate may have been present. Chadwell et al., 2001
suggested that polysulfides can be formed by the oxidation of H2S by O2, iron III, and by
manganese III and IV oxides in the absence of O2 or by micro biotic processes. Some of
this micro biota can be removed if the water sample is filtered prior to treatment with
chromium, thus limiting the activity of their use of sulfur intermediates for assimilation
and dissimilation. These sulfur-based intermediates would normally not be detected
using DEPD, unless they first get reduced with chromium (Kariuki et al. 2008). If there
were any aqueous simple sulfides (H2S, S2- or HS-) in the field samples, those compounds
would have become oxidized by the fall turnover in the lake by air. Therefore, these
oxidized sulfur species would not have been detected as simple sulfides (Sorokin 2011;
Kuwabara et al. 1999) except in the case where the samples were first treated with
chromium (Morse and Luther III, 1999). For example, if the polysulfides were present in
90
the water samples, they could have been oxidized to thiosulfate, and without the
chromium reduction of the sample, these compounds would not have been detected.
Also, since the condition of the lake was oxic at this time of year, it is possible that the
metal-sulfide complexes may have formed, in the presence of oxygen (Al-Farawati and
van den Berg 1999; Morse and Luther 1999; Zhang and Millero 1994). These metal-
sulfide complexes would also go undetected if one were simply using the DEPD method
to analyze for just acid labile sulfides. At the time of sampling, the reported pH value for
the water sample was 6.5. According to Rozan et al. (1999), metal sulfide complexes are
likely to be present in natural water samples at this pH. In order to preserve the water
samples prior to analysis they were separated and frozen. O’Reilly et al. (2001) have
reported that freezing the samples failed to prevent the oxidation of sulfide. Since the
water samples were slowly frozen at -20 ºC and not flash frozen using liquid nitrogen,
there is a possibility that oxidation could still have been occurring during the freezing
process. The presence of naturally occurring thiosulfate in the water samples analyzed
through treatment of the samples with chromium could cause these compounds to
become detectable using DEPD, and this in turn could have inflated the observed percent
recovery of sulfide.
4.3.2 Electrochemical Analysis of Lake Water
Even though there was no detectable sulfide found in the lake water samples prior to their
being reacted with chromium, the lake water was still compared against the
electrochemical parameters described above. There was no detectable current response in
terms of the potential scan window of -1600 to +1600 mV on BGCE and MGCE. The
same results held for the chromium-reduced lake water samples in combination with the
3.63 mM spiked sulfide to the water sample prior to addition of chromium. As well,
there was no detectable current signal for the water samples spiked with 3.63 mM sulfide,
and with no chromium added to it. Using a larger amount of the chromium-reduced water
sample in the EC analysis still provided no current response. More CV cycles were run,
going up to 20 cycles, using the BGCE only, since the MGCE showed physical
breakdown of the film after 10 CV cycles. Nevertheless, no current response was seen.
At this point it is not clear why this was the case.
91
4.4 CONCLUSIONS
For the original, unfiltered and filtered water sample, after going through the reduction
process without being spiked with sulfide, the sulfide was not readily detected through
DEPD or CV analyses. This was also the case with the water sample being reduced with
and without chromium. When the unfiltered water sample was spiked with sulfide and
reduced, with and without chromium there was sulfide detected (4 mM) through DEPD
reaction. However, when the water sample was filtered, spiked with sulfide, and reduced
with and without chromium the amount of sulfide detected was close to and above 100 %
recovery (Table 18) for DEPD reaction at site IG9. When using the filtered, sulfide
spiked lake water samples, and applying the electrochemical conditions there was no
detectable sulfides present over BGCE and MGCE. This should have been the case since
the aliquot added to test this procedure exceeded the lowest detectable amount of sulfide
at BGCE (≥ 0.57 mM). The exceedingly high percent recovery values for these
chromium-reduced water samples (Table 18), when compared to the results from the
original water sample, prior to spiking with sulfide, from the reductions done raises some
concerns. As indicated in Tables 16 and 17, there is a presence of transition metal ions in
both the water and in the sediments at site IG9. Metal-sulfide complexes may have
formed, in the presence of oxygen during the change in conditions of the lake from
anoxic to oxic (Al-Farawati and van den Berg 1999; Morse and Luther 1999; Zhang and
Millero 1994). According to Morse and Luther III (1999), these metal-sulfide ion
complexes are disassociated when in the presence of chromium during the reduction
procedure. The interactions between the metal ions in the water and the sediment, at the
pH measurement taken during sample collection (pH = 6.5) could be the cause for the
higher than expected recoveries noted in Table 18. Also, it should be noted that the lower
recoveries observed in Table 18 for the unfiltered water sample could suggest that
bacterial activity is promoting the dissociation of the sulfide ion (Gosh and Dam, 2009).
An observation is made that although CV could be applied to the analysis of lab grade
solutions, it did not seem to show promise for the analysis of the natural lake water
sample where the sulfide is present at very low concentrations. Even after the reduced
lake water sample was analyzed in terms of the presence of sulfide in the sample there
92
was no detectable sulfide observed during CV analyses. This was the case for both BGCE
and MGCE. Sulfide concentration in the environmental samples, determined using some
electrochemical techniques, is reported to be higher than in the samples examined in this
study. For example, Dutta et al. (2008 and 2010) reported dissolved sulfide levels in
sewage and paper mill waste water to have been 47 mg L1- and 51 mg L1-, respectively.
Also, Font et al. (1996) reported sulfide levels in waste water downstream from a tannery
to be 20 mg L1-. For future work it is possible that different electrochemical techniques
may hold more promise for the investigation of sulfide species in water samples at low
concentrations.
93
General Conclusions
In a recent review of the value of voltammetry, Batchelor-McAuley et al. (2015)
suggested that CV could be used along with other electrochemical-based techniques to
perform the quantitative analysis of sulfide. However, based on the present study, CV
using glassy carbon electrodes appears to be inefficient as a technique for the
quantification of sulfide at environmentally relevant concentrations using BGCE and
vanadium oxide MGCE. The addition of an electrode modifier, such as vanadium oxide,
enhances the catalytic response at lower sulfide concentrations when compared to that of
the concentrations analyzed using BGCE. However, when analyzing solutions with the
vanadium oxide modifier, the supporting electrolyte solution directly overlaps with the
current response related to electrooxidation of sulfide. This causes a concern, since the
interference from the supporting electrolyte solution does not allow for the detection limit
of sulfide to be lower than 0.13 mM using the CV parameters used in this study in
combination with the vanadium oxide modifier. However, the observed standard error
results (Table 8) for the concentration of 0.13 mM at MGCE, would indicate that levels
this low would be unquantifiable leading to further limitations using CV at MGCE.
Although, the SEM results, for the MGCE, indicated that there was a homogenous
coating on the electrode surface, the EDX results did indicate the presence of any
vanadium elements for that film covering. The BGCE and MGCE analyses, performed in
this study, are limited in their application towards environmentally relevant
concentrations of sulfide. Also, the considerable variation shown in measurements made
using both the BGCE and MGCE electrodes suggest that sulfide is not easily quantified
using this approach.
Methods using hanging mercury drop electrodes with polarographic, square-wave or
stripping voltammetric techniques have been shown to be more reliable for the
quantification of sulfide in solution (Dilgin et al. 2012; Huang et al. 2012; Lawrence
2006; Cheng et al. 2005; García-Calzada et al. 1999). Mercury drop electrodes have a
unique advantage over glassy carbon electrodes, as they provide a fresh working surface
with every drop of mercury. This guarantees a clean working surface with no surface
imperfections. The issue with using mercury is that it is toxic, so other electrode
94
materials are being examined in order to circumvent concerns over the use of mercury
electrodes. Other electrode materials such as gold (Waite et al., 2006), platinum
(Kapusta et al. 1983) and boron doped diamond (Lawrence et al., 2002) electrodes have
been shown to detect sulfide in solution. In contrast, the use of glassy carbon electrodes
requires that the working surface be manually polished to remove deposits from previous
analyses. These repeated manual manipulations of the glassy carbon surface will leave
grooves and uneven surfaces, which in turn may differentially accumulate particles
during subsequent runs accounting for the variation observed among the replicates.
Refinement of polishing procedures, combined with electrochemical pre-treatment prior
to analyses may help identify electrodes that may not have been properly polished
between analyses.
Any attempt to use CV to measure individual substances in mixtures of various sulfur-
based compounds, such as polysulfides and polythionates in solution, was also found to
have been challenging. Although some qualitative information for each group of
compounds was gained when they were analyzed separately, trisulfide and tetrasulfide
showed very similar CV responses using BGCE and the disulfide showed no detectable
signal under the conditions that the CV was run. When analyzed in a mixture, the sulfide,
polysulfides and polythionates are not readily distinguished from each other with the CV
protocols used in this study. This also seems to be true for the DEPD experiments. These
experiments showed good percent recoveries of simple sulfide and detectable levels of
chromium-labile sulfides, but the DEPD method could not distinguish combinations of
these sulfur-based compounds in solution. The DEPD-colorimetric method used in this
study is very similar to the methylene blue method commonly used in other studies. In
this case, the total sulfide available in solution in the form of acid-labile, metal sulfides,
polysulfides and polythionates in a water sample can be quantified when the sample has
been reduced with chromium under the conditions described earlier (Kamyshny et al.
2008).
Clearly, sulfide was not readily detected in the lake water sample. Even after reducing
the water sample with chromium according to the procedures outlined earlier, no sulfide
was detected using either the DEPD or the CV methods. When the sample was spiked
95
with a known concentration of sulfide, sulfide was detected and measurable. However,
the water sample must first be filtered prior to analyses in order to reduce any matrix
interferences that may occur. Reduction of the sulfide-containing water samples at a
temperature of 60ºC, with chromium metal as a reducing agent, allowed for the total
recovery of sulfide. Cyclic voltammetry of the lake water sample and the chromium-
reduced lake water sample provided no qualitative or quantitative information on sulfide
concentrations using either BGCE or MGCE electrodes. A further investigation of the
techniques presented in this study, as well as others not presented may cast more light
into how sulfur compounds in a sample can be detected.
This study has provided a useful assessment of the potential of CV, using BGCE and
vanadium oxide MGCE as a method for the qualification and quantification of lab
generated sulfur based compounds in solutions. It is now clear that when these sulfur
based compounds are combined as a mixture, there is a lack of speciation between the
compounds, thereby making them not easily distinguished from one another. When
examining environmental samples, being able to have speciation of the various sulfur
compounds, contained as a mixture in the natural samples would be a great contribution
to the community at large. The chemistry of sulfur in natural systems is complex, and
highly dependent on oxidation states and the availability of molecular oxygen. In
addition, there is some likelihood that chemical changes may easily develop during
transportation and storage, especially in raw unfiltered water samples that still contain
numerous active microorganisms. Other factors that may have affected the results of the
field analyses would include the concentrations of metal ions in the water sample, the
sediment, and the time of year the sample was taken. A thorough study of chemistry of
the vanadium oxide electrode modifiers may provide the real key to the usefulness of the
CV technique for the analysis of sulfur species in natural water and other environmental
samples. Such a study could include reducing the forward switching potential to a value
less than where water oxidizes to oxygen. Exploration of other metal oxide modifiers
besides vanadium films, such as bismuth (Huang et al., 2012), and an investigation of
other electrochemical techniques for the analysis of sulfur compounds in natural samples
could turn out to be very informative.
96
Appendices
Appendix A
Supporting Electrolyte Solution:
0.1 M Potassium Phosphate Buffer
A potassium phosphate buffer solution was prepared by combining solutions of 0.075 M
dibasic potassium phosphate with 0.025 M monobasic potassium phosphate. The buffer
solution was brought to a pH of 10.3 using sodium hydroxide (NaOH) (RICCA
Chemicals). The phosphate buffer solution was filtered through a 47.00 mm, 0.45 µm
mixed cellulose membrane (Millipore Corp.) using a glass vacuum filtration apparatus,
and water aspirator. The mono and dibasic phosphate salts were reagent grade, and
required filtration to remove any particulates that would otherwise be unwanted. After
filtration, the solution was purged for 1 hour using nitrogen gas, by inserting a piece of
Teflon tubing (ID 3.2 mm x OD 6.4 mm x wall1.6 mm) to the bottom of the glass storage
bottle that was connected to the regulator supplying the gas. Before the tube was fully
removed, once the purging was complete, the tube was brought above the liquid level in
the bottle to insert a blanket of N2 gas in the head space of the storage bottle before the
buffer solution was capped for storage until use.
Solutions for Iodometric Titrations:
6 M Hydrochloric Acid Solution
Using a 100-mL volumetric flask, 50-mL of concentrated hydrochloric acid (trace metal
grade, Fisher Scientific) was added to 30-mL of ultra-pure water; then filled to the mark,
on the flask, to make a total volume of 100-mL. This solution was used for standardizing
the sulfide stock solution during the titration procedure with 25 mM iodine and
thiosulfate.
97
Standard Thiosulfate Solution 25 mM
Using a 1000-mL graduated cylinder, approximately 600-mL of the ultrapure water was
transferred into 1000-mL Erlenmeyer flask. Sodium thiosulfate pentahydrate
(Na2S2O3•5H2O) (≥ 98.5 %, Anachemia, ACS grade) weighing 6.21 ± 0.05 g was
dissolved into the ultrapure water, contained in the Erlenmeyer flask. Next, 0.4 ± 0.01 g
of NaOH was added to the Na2S2O3•5H2O solution and the contents of the Erlenmeyer
flask were brought to a total volume of 1000-mL. The NaOH was added as a
preservative. Once prepared, the thiosulfate solution was poured into an amber storage
bottle. The solution was expected to be stable for 6 months (Clesceri, Greenberg, and
Eaton, 1998). The thiosulfate solution was standardized against potassium bi-iodate
solution.
Potassium Bi-Iodate 2 mM solution
A sample of potassium bi-iodate (KIO3) (SIGMA, analytical reagent ≥ 99.8 %), weighing
0.814 ± 0.005 g was transferred to a 1000-mL Erlenmeyer flask. The KIO3 was dissolved
in about 600-mL of ultrapure water. Once dissolved the KIO3 solution was brought to a
total volume of 1000-mL, mixed well, and then the solution was transferred into an
amber storage bottle. The solution was expected to be stable for 6 months (Clesceri,
Greenberg, and Eaton, 1998).
Iodine Solution 25 mM
Using a 1000-mL graduated cylinder, approximately 600-mL of ultrapure water was
added to a 1000-mL Erlenmeyer flask. A sample of potassium iodide (KI) (Fisher, ≥ 99.0
%) weighing 24.1 ± 0.05 g was and added to the flask containing the ultrapure water.
Once the KI was dissolved completely, 3.2 ± 0.02 g of resublimed iodine solid (I2)
(Fisher, ACS grade) was weighed, added to the KI solution, and dissolved. Once the
iodine solid was completely dissolved, the remaining ultra-pure water was added to reach
a total volume of 1000-mL. The iodine solution was transferred to an amber bottle for
storage, and was expected to be stable for 6 months (Clesceri, Greenberg, and Eaton,
1998). The iodine solution was standardized against 25 mM thiosulfate solution.
98
Starch Solution 2 % (w/v)
Using a 250-mL beaker, 200-mL of ultra-pure water was heated thoroughly on a hotplate.
Some of hot water from the beaker was poured into a 100-mL volumetric flask, and the
flask was placed back on the hotplate. A sample of 2.00 g soluble starch (Lab Grade,
Fisher Scientific), was weighed, added to the hot water, and dissolved. The flask was
capped, and mixed well. The remaining hot water was poured to the mark of the
volumetric flask to reach total volume of 100-mL. The starch solution was expected to
keep stable for a period of 1 month (Clesceri, Greenberg, and Eaton, 1998). This
solution is used to check for the presence of iodine during iodometric titrations.
Solutions for DEPD Procedure
0.1 M Sodium Hydroxide (NaOH) Solution
Using a 1000-mL glass storage bottle, ultra-pure water (18.2 MΩcm) was added to within
a few inches of the top. Approximately 4.00 g of NaOH pellets were weighed and placed
in the ultra-pure water, a stir bar was added, and the solution was placed on a hot-plate
stirrer, with no heat applied, and set to 800 rpm. While the solution was being stirred, it
was also purged for 1 hour using N2 gas (99.998 %) in order to remove any presence of
oxygen from the solution. The pH was confirmed using a pH meter with an ATC pH
combination electrode (Accumet AB150 pH/mV meter and pH/ATC electrode, Fisher
Scientific).
Colour Developing Reagent
Colour developing reagent was made by weighing out 4.3 g of N,N- Diethyl-p-
phenylenediamine oxalate salt 96 % (DEPD) (Fisher Scientific), and 6.4 g iron (III)
chloride hexahydrate 97 – 102 % (FeCl3·6H2O) (ACS grade, Sigma). These two
compounds were added to 100-mL of 50% (v/v) HCl solution, placed in a glass storage
bottle, and stored at 4°C.
99
Appendix B
3𝐻2𝑆𝑂4(𝑎𝑞) + 𝐾𝐼𝑂3(𝑎𝑞) + 5𝐾𝐼(𝑎𝑞) ↔ 3𝐼2(𝑎𝑞) + 3𝐻2𝑂(𝑙) + 3𝐾2𝑆𝑂4(𝑎𝑞) (R − 1.1)
2𝑁𝑎2𝑆2𝑂3(𝑎𝑞) + 𝐼2(𝑎𝑞) → 2𝑁𝑎𝐼(𝑎𝑞) + 𝑁𝑎2𝑆4𝑂6(𝑎𝑞) (𝑅 − 1.2)
ɳ𝐾𝐼𝑂3 = 𝐶 × 𝑉(𝐿) (𝐸𝑞. 1.3)
ɳ𝐼2 = 3 × ɳ𝐾𝐼𝑂3
(𝐸𝑞. 1.4)
ɳ𝑁𝑎2𝑆2𝑂3 = 2 × ɳ𝐼2
(𝐸𝑞. 1.5)
𝑀𝑜𝑙𝑎𝑟𝑖𝑡𝑦 𝑆2𝑂3 = (ɳ𝑁𝑎2𝑆2𝑂3
𝑉𝑜𝑙. 𝑁𝑎2𝑆2𝑂3(𝐿)
) × 2 (𝐸𝑞. 1.6)
2𝑆2𝑂32− + 𝐼2 → 𝑆4𝑂6
2− + 2𝐼− (R − 1.3)
𝑀𝑜𝑙𝑎𝑟𝑖𝑡𝑦 𝐼2 = (𝑀𝑜𝑙𝑎𝑟𝑖𝑡𝑦 𝑜𝑓 𝑆2𝑂3) ∙(𝑉𝑜𝑙𝑁𝑎2𝑆2𝑂3 (𝐿))
𝑉𝑜𝑙. 𝐼2 𝑡𝑖𝑡𝑟𝑎𝑡𝑒𝑑 (𝐿) (𝐸𝑞. 1.7)
𝐼2 + 𝑆2− → 𝑆 + 2𝐼− (𝑅 − 1. 4)
𝐼2 + 2𝑆2𝑂3 → 𝑆4𝑂62− + 2𝐼− (𝑅 − 1.5)
The following equation was used to calculate the concentration of sulfide in mg/L (Eq.
1.8):
𝑚𝑔
𝐿 𝑆2− =
[(𝐴 × 𝐵) − (𝐶 × 𝐷)] × 32000
𝑉𝑜𝑙. 𝑜𝑓 𝑆2− 𝑠𝑜𝑙𝑢𝑡𝑖𝑜𝑛 (𝑚𝐿) (𝐸𝑞. 1. 8)
Where:
A = mL iodine solution
B = molarity of iodine solution
C = mL thiosulfate solution
D = molarity of thiosulfate solution
100
𝐻𝑆− ↔ 𝐻+ + 𝑆2− (𝑅 − 1.5)
𝑆2− ↔ 𝑆 + 2𝑒− (𝑅 − 1.6)
Oxidation states of various sulfur compounds.
Species S2- HS- S S2O32- S4O6
2- SO32- SO4
2-
Oxidation No. -2 -1 0 +2 +2.5 +4 +6
Stoichiometric Equations that show the amount of Na2S and S required for
making Na2S2, Na2S3, and Na2S4
𝑵𝒂𝟐𝑺 + 𝑺 → 𝑵𝒂𝟐𝑺𝟐 (𝑅 − 2.1)
The amount of Na2S and S0 required making 1.00 g Na2S2 are given in equations (2.1 a)
and (2.1 b):
1.00 𝑔 𝑁𝑎2𝑆2 × 1 𝑚𝑜𝑙 𝑁𝑎2𝑆2
110.11 𝑔 ×
1 𝑚𝑜𝑙 𝑁𝑎2𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆2 ×
78.04 𝑔 𝑁𝑎2𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆= 0.7087 𝑔 𝑁𝑎2𝑆 (2.1 𝑎)
1.00 𝑔 𝑁𝑎2𝑆2 × 1 𝑚𝑜𝑙 𝑁𝑎2𝑆2
110.11 𝑔 ×
1 𝑚𝑜𝑙 𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆2 ×
32.06 𝑔 𝑆
1 𝑚𝑜𝑙 𝑆= 0.2912 𝑔 𝑆 (2.1 𝑏)
𝑵𝒂𝟐𝑺 + 𝟐𝑺 → 𝑵𝒂𝟐𝑺𝟑 (𝑅 − 2.2)
The amount of Na2S and S0 required making 1.00 g Na2S3 are given in equations (2.2 c)
and (2.2 d):
1.00 𝑔 𝑁𝑎2𝑆3 × 1 𝑚𝑜𝑙 𝑁𝑎2𝑆3
142.17 𝑔 ×
1 𝑚𝑜𝑙 𝑁𝑎2𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆3 ×
78.04 𝑔 𝑁𝑎2𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆= 0.5489 𝑔 𝑁𝑎2𝑆 (2.2 𝑐)
1.00 𝑔 𝑁𝑎2𝑆3 × 1 𝑚𝑜𝑙 𝑁𝑎2𝑆3
142.17 𝑔 ×
2 𝑚𝑜𝑙 𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆3 ×
32.06 𝑔 𝑆
1 𝑚𝑜𝑙 𝑆= 0.4510 𝑔 𝑆 (2.2 𝑑)
𝑵𝒂𝟐𝑺 + 𝟑𝑺 → 𝑵𝒂𝟐𝑺𝟒 (𝑅 − 2.3)
The amount of Na2S and S0 required making 1.00 g Na2S4 are given in equations (2.3 e)
and (2.3 f):
101
1.00 𝑔 𝑁𝑎2𝑆4 × 1 𝑚𝑜𝑙 𝑁𝑎2𝑆4
174.24 𝑔 ×
1 𝑚𝑜𝑙 𝑁𝑎2𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆4 ×
78.04 𝑔 𝑁𝑎2𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆= 0.4479 𝑔 𝑁𝑎2𝑆 (2.3 𝑒)
1.00 𝑔 𝑁𝑎2𝑆4 × 1 𝑚𝑜𝑙 𝑁𝑎2𝑆4
174.24 𝑔 ×
3 𝑚𝑜𝑙 𝑆
1 𝑚𝑜𝑙 𝑁𝑎2𝑆4 ×
32.06 𝑔 𝑆
1 𝑚𝑜𝑙 𝑆= 0.5520 𝑔 𝑆 (2.3 𝑓)
102
References
Al-Farawati R., and van den Berg C.M.G. 1997. The determination of sulfide in
seawater by flow-analysis with voltammetric detection. Marine Chemistry. 57(3-4):
227-286.
Al-Farawati R., and van den Berg C.M.G. 1999. Metal-sulfide complexation in
seawater. Marine Chemistry. 63(3): 331-352.
Al-Farawati R., and van den Berg C.M.G. 2001. Thiols in coastal waters of the
western North Sea and English Channel. Environmental Science and Technology.
35(10): 1902-1911.
Andreae M.O. 1990. Ocean-atmosphere interactions in the global biogeochemical sulfur
cycle. Marine Chemistry. 30: 1-29.
Anklet G.T. 1996. Evaluation of metal/acid-volatile sulfide relationships in the
prediction of metal bioaccumulation by benthic macroinvertebrates. Environmental
Toxicology and Chemistry. 15(12): 2138-2146.
Ardelean M., Manea F., Vaszilcsin N., and Pode R. 2014. Electrochemical detection of
sulphide in water/seawater using nanostructured carbon–epoxy composite electrodes.
Analytical Methods. 6(13): 4775-4782.
Ateya B.G., Alkharafi F.M., Alazab A.S., and Abdullah A.M. 2007. Kinetics of the
electrochemical deposition of sulfur from sulfide polluted brines. Journal of Applied
Electrochemistry. 37(3): 395-404.
Avrahami, M., and Golding, R. M. (1968). The oxidation of the sulphide ion at very
low concentrations in aqueous solutions. Journal of the Chemical Society A:
Inorganic, Physical, Theoretical, 647-651.
103
Azizi A., Petre C.F., Olsen C., and Larachi F. 2010. Electrochemical behavior of gold
cyanidation in the presence of a sulfide-rich industrial ore versus its major
constitutive sulfide minerals. Hydrometallurgy. 101(3): 108-119.
Baciu A., Ardelean M., Pop A., Pode R., and Manea F. 2015. Simultaneous
voltammetric/amperometric determination of sulfide and nitrite in water at BDD
electrode. Sensors. 15(6): 14526-14538.
Bagarinao T. 1992. Sulfide as an environmental factor and toxicant: tolerance and
adaptations in aquatic organisms. Aquatic Toxicology. 24(1-2): 21-62.
Balasubramanian S. and Pugalenthi V. 2000. A comparative study of the
determination of sulphide in tannery waste water by ion selective electrode (ISE) and
iodimetry. Water Research. 34(17): 4201-4206.
Barrado E., Pardo R., Castrillejo Y., and Vega M. 1997. Electrochemical behavior of
vanadium compounds at a carbon paste electrode. Journal of Electroanalytical
Chemistry. 427(1-2): 35-42.
Bard A.J., and Faulkner. 2001. Electrochemical Methods: fundamentals and
applications. Wiley.
Batchelor-McAuley C., Kätelhön E., Barnes E. O., Compton R. G., Laborda E., and
Molina A. 2015. Recent Advances in Voltammetry. Chemistry Open. 4(3): 224–260
[http://doi.org/10.1002/open.201500042].
Berg J.S., Schwedt A., Kreutzmann A-C., Kuypers M.M.M., and Milucka J. 2014.
Polysulfides as intermediates in the oxidation of sulfide to sulfate by Beggiatoa spp.
Applied and Environmental Microbiology. 80(2): 629-636.
104
Berglen T.F., Berntsen T.K., Isaksen I.S.A., and Sundet J.K. 2004. A global model of
the coupled sulfur/oxidant chemistry in the troposphere: the sulfur cycle. Journal of
Geophysical Research: Atmospheres. 109(D19): 1-27.
Boulegue J., Lord C.J., and Church T.M. 1982. Sulfur speciation and associated trace
metals (Fe, Cu) in the pore waters of Great Marsh, Delaware. Geochimica et
Cosmochimica Acta. 46(3): 453-464.
Bourke A., Miller M.A., Lynch R.P., Wainright J.S., Savinell R.F., and Buckley
D.N. 2015. Effect of cathodic and anodic treatments of carbon on the electrode
kinetics of ViV/Vv oxidation-reduction. Journal of the Electrochemical Society.
162(8): A1547-A1555.
Bouroushain M. 2010. Electrochemistry of metal chalcogenides. Springer Science &
Business Media.
Bowles K.C., Russell,B.A., Ernste M.J., Kramer J.R., Manolopoulos H., and Ogden
N. 2002. Synthesis and characterization of metal sulfide clusters for toxicological
studies Environmental Toxicology and Chemistry. 21(4): 693-699.
Bowles K.C., Ernste M.J., Kramer J.R. 2003. Trace Sulfide Determination in oxic
freshwaters. Analytica Chimica Acta. 477(1): 113-124.
Brouwer H., and Murphy T. 1994. Diffusion method for the determination of acid-
volatile sulfides (AVS) in sediment. Environmental Toxicology and Chemistry.
13(8): 1273-1275.
Brouwer H., and Murphy T. 1995. Volatile sulfides and their toxicity in freshwater
sediments. Environmental Toxicology and Chemistry. 14(2): 203-208.
105
Burton E.D., Bush R.T., and Sullivan L.A. 2006. Acid-volatile sulfide oxidation in
coastal flood plain drains: iron-sulfur cycling and effects on water quality.
Environmental Science and Technology. 40(4): 1217-1222.
Cadena F., and Peters R.W. 1988. Evaluation of chemical oxidizers for hydrogen
sulfide control. Journal Water Pollution Control Federation. 60(7): 1259-1263.
Canadian Council of Ministers of the Environment (CCME) water quality
guidelines for the protection of aquatic life Accessed January 2017.
(http://www.ccme.ca/en/resources/canadian_environmental_quality_guidelines/)
Canales C., Gidi L., and Ramírez G. 2015. Electrochemical activity of modified glassy
carbon electrodes with covalent bonds towards molecular oxygen reduction.
International Journal of Electrochemical Science. 10: 1684-1695.
Canfield D.E., Raiswell R., Westrich J.T., Reaves C.M., and Berner R.A. 1986. The
use of chromium reduction in the analysis of reduced inorganic sulfur in sediments
and shales. Chemical Geology. 54(1-2): 149-155.
Canfield D.E., and Raiswell R. 1999. The evolution of the sulfur cycle. American
Journal of Science. 299(7-9): 697-723.
Casella I.G., Guascito M.R., Desimoni E. 2000. Sulfide measurements by flow
injection analysis and ion chromatography with electrochemical detection. Analytica
Chimica Acta. 409(1): 27-34.
Casella I.G., and Gatta M. 2000. Anodic electrodeposition of copper oxide/hydroxide
films by alkaline solutions containing cuprous cyanide ions. Journal of
Electroanalytical Chemistry. 494(1): 12-20.
106
Casella I.G., Contursi M., and Desimoni E. 2002. Amperometric detection of sulfur-
containing compounds in alkaline media. The Analyst. 127(5): 647-652.
Chadwell S.J., Rickard D., and Luther III G.W. 1999. Electrochemical Evidence for
Pentasulfide Complexes with Mn2+, Fe2+, Co2+, Ni2+, Cu2+ and Zn2+. Aquatic
Geochemistry. 5(1): 29-57.
Chadwell S.J., Rickard D., and Luther III G.W. 2001. Electrochemical evidence for
metal polysulfide complexes: tetrasulfide (S42-)n reactions with Mn2+, Fe2+, Co2+,
Ni2+, Cu2+, and Zn2+. Electroanalysis. 13(1): 21-29.
Chang J-L., Wei G-T., Chen T-Y., and Zen J-M. 2013. Highly stable polymeric ionic
liquid modified electrode to immobilize ferricyanide for eletroanalysis of sulfide.
Electroanalysis. 25(4): 845-849.
Chase, D., B. Horwood, and L. Lovett-Doust. 2016. Contours of Metal Contamination
in Lake Nipissing, in the Great Lakes Basin. Submitted to Ecological Applications.
Chen S-M. 2002. Preparation, characterization, and electrocatalytic oxidation properties
of iron, cobalt, nickel, and indium hexacyanoferrate. Journal of Electroanalytical
Chemistry. 521(1): 29-52.
Cheng J., Jandik P., and Avdalovic N. 2005. Pulsed amperometric detection of sulfide,
cyanide, iodide, thiosulfate, bromide and thiocyanate with microfabricated
disposable silver working electrodes in ion chromatography. Analytica Chimica
Acta. 536(1): 267-274.
Chung Y-C., Huang C., and Tseng C-P. 2001. Biological elimination of H2S and NH3
from waste gases by biofilter packed with immobilized heterotrophic bacteria.
Chemosphere. 43(8): 1043-1050.
107
Chun S.W., Jang J.Y., Park D.W., Woo H.C., and Chung J.S. 1998. Selective
oxidation of H2S to elemental sulfur over TiO2/SiO2 catalysts. Applied Catalysis B:
Environmental. 16(3): 235-243.
Chu, X., Ohmoto, H., and Cole, D.R. 2004. Kinetics of sulfur isotope exchange
between aqueous sulfide and thiosulfate involving intra- and intermolecular reactions
at hydrothermal conditions. Chemical Geology. 211(3-4): 217-235.
Ciesielski W., and Zakrzewski R. 2006. Iodimetric titration of sulfur compounds in
alkaline medium. Chemia analityczna. 51(5): 653-678.
Ciglenečki I., and Ćosović B. 1997. Electrochemical determination of thiosulfate in
seawater in the presence of elemental sulfur and sulfide. Electroanalysis. 9(10): 775-
780.
Clesceri L.S., Greenberg A.E., and Eaton A.D. 1998. Standard methods for the
examination of water and wastewater, 20th ed. American Public Health Association,
DC, pp. 4-131 and 4-167.
Cline J.D. 1969. Spectrophotometric Determination of hydrogen sulfide in natural
waters. Limnology and Oceanography. 14(3): 454-458.
Corkhill C.L., Vaughan D.J. 2009. Arsenopyrite oxidation – a review. Applied
Geochemistry. 24(12): 2342-2361.
Compton R.G., Laborda E., and Ward K.R. 2014. Understanding voltammetry:
simulation of electrode processes. Imperial College Press.
Cox H.H.J., and Deshusses M.A. 2002. Co-treatment of H2S and toluene in biotrickling
filter. Chemical Engineering Journal. 87(1): 101-110.
108
Cummings T.E., and Cox J.A. 1979. Formal potential estimation for a quasi-reversible
charge transfer followed by a first-order irreversible chemical reaction.
Electrochimica Acta. 24(10):1101-1104.
D’Elia L.F., Rincόn L., and Ortíz R. 2004. Test of vanadium pentoxide as anode for
the electrooxidation of toluene a theoretical approach of the electrode process.
Electrochimica Acta. 50(1): 217-224.
Delgado S., Alvarez M., Rodriguez-Gomez L.E., and Aguiar E. 1999. H2S generation
in a reclaimed urban wastewater pipe. Case study: Tenerife (Spain). Water Research.
33(2): 539-547.
Dikshitulu L. S. A., and Rao G. G. 1962. Titrimetric determination of Vanadium (IV)
with potassium permanganate at the room temperature, using phosphoric acid as
catalyst and ferroin as internal indicator. Fresenius' Zeitschrift für analytische
Chemie. 189(5): 421-426.
Dilgin Y., Kızılkaya B., Ertek B., Eren N., and Dilgin D.G. 2012. Amperometric
determination of sulfide based on its electrocatalytic oxidation at a pencil graphite
electrode modified with quercetin. Talanta. 89: 490-495.
Druschel G.K., Hamers R.J., and Banfield J.F. 2003. Kinetics and mechanism of
polythionate oxidation to sulfate at low pH by O2 and Fe3+. Geochimica et
Cosmochimica Acta. 67(23): 4457-4469.
Dutta P.K., Rabaey K., Yuan Z., and Keller J. 2008. Spontaneous electrochemical
removal of aqueous sulfide. Water Research. 42(20): 4965-4975.
Dutta P.K., Rozendal R.A., Yuan Z., Rabaey K., and Keller J. 2009. Electrochemical
regeneration of sulfur loaded electrodes. Electrochemistry Communications. 11(7):
1437-1440.
109
Dutta P.K., Rabaey K., Yuan Z. Rozendal R.A., and Keller J. 2010. Electrochemical
sulfide removal and recovery from paper mill anaerobic treatment effluent. Water
Research. 44(8): 2563-2571.
Dyrssen D. 1988. Sulfide complexation in surface seawater. Marine Chemistry. 24(2):
143-153.
Economou A., Bolis S.D., Efstathiou C.E., and Volikakis G.J. 2002. A “virtual”
electroanalytical instrument for square wave voltammetry. Analytica Chimica Acta.
467(1): 179-188.
Edwards S., Alharthi R., and Ghaly A. 2011. Removal of hydrogen sulfide from water.
American Journal of Environmental Sciences. 7(4): 295-305.
Farquhar J., Bao H., and Thiemens M. 2000. Atmospheric influence of Earth’s earliest
sulfur cycle. Science. 289(5480): 756-758.
Fazzini R.A.B., Cortés M.P., Padilla L., Maturana D., Budinich M., Maass A., and
Parada P. 2013. Stoichiometric modeling of oxidation of reduced inorganic sulfur
compounds (Riscs) in Acidithiobacillus thiooxidans. Biotechnology and
Bioengineering. 110(8): 2242-2251.
Feng J., Gao Q., Xu L., and Wang J. 2005. Nonlinear phenomena in the
electrochemical oxidation of sulfide. Electrochemistry Communications. 7(12):
1471-1476.
Fleet B., and Gunasingham H. 1992. Electrochemical sensors for monitoring
environmental pollutants. Talanta. 39(11): 1449-1457.
110
Font J., Gutiérrez J., Lalueza J., and Pérez X. 1996. Determination of sulfide in the
leather industry by capillary electrophoresis. Journal of Chromatography A. 740(1):
125-132.
Fossing H., and Jørgensen B. 1989. Measurement of bacterial sulfate reduction in
sediments: Evaluation of a single-step chromium reduction method.
Biogeochemistry. 8(3): 205-222.
Friedrich C.G., Rother D., Bardischewsky F., Quentmeier A., and Fischer J. 2001.
Oxidation of reduced inorganic sulfur compounds by bacteria: emergence of a
common mechanism. Applied and Environmental Microbiology. 67(7): 2873-2882.
Friedrich C.G., Bardischewsky F., Rother D., Quentmeier A., and Fischer J. 2005.
Prokaryotic sulfur oxidation. Current Opinion in Microbiology. 8(3): 253-259.
García-Calzada M., Marbán G., and Fuertes A.B. 1999. Potentiometric determination
of sulfur in solid samples with a sulphide selective electrode. Analytica Chimica
Acta. 380(1): 39-45.
Garrett R.L., Clark R.K., Carney L.L., and Grantham Sr. C.K. 1979. Chemical
scavengers for sulfides in water-base drilling fluids. Journal of Petroleum
Technology. 31(6): 787-796.
Ghosh, W., and Dam, B. 2009. Biochemistry and molecular biology of lithotrophic
sulfur oxidation by taxonomically and ecologically diverse bacteria and archaea.
FEMS microbiology reviews. 33(6): 999-1043.
Giovanelli D., Lawrence N. S., Jiang L., Jones, T. G., and Compton R. G. 2003.
Electrochemical determination of sulphide at nickel electrodes in alkaline media: a
new electrochemical sensor. Sensors and Actuators B: Chemical, 88(3): 320-328.
111
González A.G., and Herandor A. 2007. A practical guide to analytical method
validation, including measurement uncertainty and accuracy profiles. Trends in
Analytical Chemistry. 26(3): 227-238.
Grubbs F.E. 1969. Procedures for detecting outlying observations in samples.
Technometrics. 11(1): 1-21.
Gun J., Goifman A., Shkrob I., Kamyshny A., Ginzburg B., Hadas O., Dor I.,
Modetov A.D., and Lev O. 2000. Formation of polysulfides in an oxygen rich
freshwater lake and their role in the production of volatile sulfur compounds in
aquatic systems. Journal of Environmental Science and Technology. 34(22): 4741-
4746.
Gupta V.K., Ganjali M.R., Norouzi P., Khani H., Nayak A., and Agarwal S. 2011.
Electrochemical analysis of some toxic metals by ion-selective electrodes. Critical
Reviews in Analytical Chemistry. 41(4): 282-313.
Hage D.S., and Carr J.D. 2011. Analytical Chemistry and Quantitative Analysis. Upper
Saddle River New Jersey. Pearson Prentice Hall, Pearson Education Inc.
Halfyard J.E., and Hawboldt K. 2011. Separation of elemental sulfur from
hydrometallurgical residue: a review. Hydrometallurgy. 109(1): 80-89.
Harris D.C. 2010. Quantitative chemical analysis 8th edition. New York, NY. Freeman
and Company, pp. AP26-27.
Hart J.P., Crew A., Crouch E., Honeychurch K.C., Pemberton R.M. 2004. Some
recent designs and developments of screen-printed carbon electrochemical
sensors/biosensors for biomedical, environmental, and industrial analyses. Analytical
Letters. 37(5): 789-830.
112
Hassan S.S., Marzouk S.A.M., Sayour H.E.M. 2002. Methylene blue potentiometric
sensor for selective determination of sulfide ions. Analytica Chimica Acta. 466(1):
47-55.
Herlihy A.T., and Mills A.L. 1985. Sulfate reduction in freshwater sediments receiving
acid mine drainage. Applied and Environmental Microbiology. 49(1): 179-186.
Hernández-Munoz L.S., and González F.J. 2011. One-step modification of carbon
surfaces with ferrocene groups through a self-mediated oxidation of ferrocene acetate
ions. Electrochemistry Communications. 13(7): 701-703.
Holmer M., and Storkholm P. 2001. Sulphate reduction and sulfur cycling in lake
sediments: a review. Freshwater Biology. 46(4): 431-451.
Howard D.E., and Evans R.D. 1993. Acid-volatile sulfide (AVS) in a seasonally anoxic
mesotrophic lake: seasonal and spatial changes in sediment AVS. Environmental
Toxicology and Chemistry. 12(6): 1051-1057.
Howarth R.W., Giblin A., Gale J., Peterson B.J., and Luther G.W. III. 1983.
Reduced sulfur compounds in the pore waters of a New England salt marsh.
Ecological Bulletins, 135-152.
Hrbac J., Halouzka V., Trnkova L., and Vacek J. 2014. eL-Chem viewer: a freeware
package for the analysis of electroanalytical data and their post-acquisition
processing. Sensors. 14(8): 13943-13754.
Huang D-Q., Xu B-L., Tang J., Yang L-L., Yang Z-B., and Bi S-P. 2012. Bismuth
film electrodes for indirect determination of sulfide ion in water samples at trace
level by anodic stripping voltammetry. International Journal of Electrochemical
Science. 7: 2860-2873.
113
Hughes M.N., Centelles M.N., and Moore K.P. 2009. Making and working with
hydrogen sulfide the chemistry and generation of hydrogen sulfide in vitro and its
measurement in vivo: a review. Free Radical Biology and Medicine. 47(10): 1346-
1353.
Hu X., and Mutus B. 2013. Electrochemical detection of sulfide. Reviews in Analytical
Chemistry. 32(3): 247-256.
Janssen A.J.H., Lettinga G., and de Keizer A. 1999. Removal of hydrogen sulfide
from wastewater and waste gases by biological conversion to elemental sulfur
colloidal and interfacial aspects of biologically produced sulfur particles. Colloids
and Surfaces A: Physicochemical and Engineering Aspects. 151(1): 389-397.
Jeroschewski P., Haase K., Trommer A., Gründler P. 1993. Galvanic sensor for
determination of hydrogen sulfide/sulphide in aqueous media. Fresenius’ Journal of
Analytical Chemistry. 346: 930-933.
Jeroschewski P., Haase K., Trommer A., Gründler P. 1994. Galvanic sensor for
determination of hydrogen sulfide. Electroanalysis. 6(9): 769-772.
Johnston D.T. 2011. Multiple sulfur isotopes and the evolution of Earth’s surface sulfur
cycle. Earth-Science Reviews. 106(1): 161-183.
Kaasalainen H., and Stefánsson A. 2011. Sulfur speciation in natural hydrothermal
water, Iceland. Geochimica et Cosmochimica Acta. 75(10): 2777-2791.
Kalal H.S., Ghadiri M., Beigi A.A.M., and Sadjadi S.A.S. 2004. Simultaneous
determination of trace amounts of sulfite and thiosulfate in petroleum and its
distillates by extraction and differential pulse polarography. Analytical Chimica
Acta. 502(1): 133-139.
114
Kapusta S., Viehbeck A., Wilhelm S.M., and Hackerman N. 1983. The anodic
oxidation of sulfide on platinum electrodes. Journal of Electroanalytical Chemistry.
153(1-2): 157-174.
Kamau G.N. 1988. Surface preparation of glassy carbon electrodes. Analytica Chimica
Acta. 207: 1-16.
Kamyshny Jr. A., Goifman A., Gun J., Rizkov D., and Lev O. 2004. Equilibrium
distribution of polysulfide ions in aqueous solutions at 25 °C: a new approach for the
study of polysulfides’ equilibria. Journal of Environmental Science and Technology.
38(24): 6633-6644.
Kamyshny Jr. A., Ekeltchik I., Gun J., and Lev O. 2006. Method for the determination
of inorganic polysulfide distribution in aquatic systems. Analytical Chemistry. 78(8):
2631-2639.
Kamyshny Jr. A., Zilberbrand M., Ekeltchik I., Voitsekovski T., Gun J., and Lev O.
2008. Speciation of polysulfides and zerovalent sulfur in sulfide-rich water wells in
southern and central Isreal. Aquatic Geochemistry. 14(2): 171-192.
Kano K., Torimura M., Esaka Y., Goto M., and Ueda T. 1994. Electrocatalytic
oxidation of carbohydrates at copper(II)-modified electrodes and its application to
flow-through detection. Journal of Electroanalytical Chemistry. 372(1-2): 137-143.
Kariuki S., Morra M.J., Umiker K.J., and Cheng I.F. 2001. Determination of total
ionic polysulfides by differential pulse polarography. Analytica Chemica Acta.
442(2): 277-285.
Kariuki S., Babady-Bila P., and Duquette B. 2008. N,N-diethyl-p-phenylenediamine
effectiveness in analysis of polysulfides and polythionates in water. Environmental
Chemistry. 5(3): 226-230.
115
Karst-Riddoch, T. 2011. Callander Bay Subwatershed Phosphorus Budget. Prepared by
Hutchinson Environmental Sciences Ltd., for North Bay-Mattawa Conservation
Authority. 51 pp.
Kellogg W.W., Cadle R.D., Allen E.R., Lazarus A.L. and Martell E.A. 1972. The
sulfur cycle. Science. 175(4022): 587-596.
Keller-Lehmann B., Corrie S., Ravin R., Yuan Z., and Keller J. 2006. Preservation
and simultaneous analysis of relevant soluble sulfur species in sewage samples. In
Proceedings of the Second International IWA Conference on Sewer Operation and
Maintenance. Vol 26: p. 28.
Kelsall G.H. and Thompson I. 1993. Redox chemistry of H2S oxidation in the British
Gas Stretford Process part I: thermodynamics of sulfur-water systems at 298 K.
Journal of Applied Electrochemistry. 23(4): 279-286.
Khalifa H., and El-Strafy A. 1966. A new potentiometric method for the estimation of
vanadium. Fresenius' Journal of Analytical Chemistry. 227(2): 109-115.
Khudaish E.A., and Al-Hinai A.T. 2006. The catalytic activity of vanadium pentoxide
film modified electrode on the electrochemical oxidation of hydrogen sulfide in
alkaline solutions. Journal of Electroanalytical Chemistry. 587(1): 108-114
Kiema G.K., Fitzpatrick G., and McDermott M.T. 1999. Probing morphological and
compositional variations of anodized carbon electrodes with tapping-mode scanning
force microscopy. Analytical Chemistry. 71(19): 4306-4312.
Kim C., Zhou Q., Deng B., Thornton E.C., and Xu H. 2001. Chromium(VI) reduction
by hydrogen sulfide in aqueous media: stoichiometry and kinetics. Environmental
Science and Technology. 35(11): 2219-2225.
116
Kissinger P.T., and Heineman W.R. 1996. Laboratory techniques in Electroanalytical
chemistry, 2nd edition revised and expanded. Marcel Dekker Inc.
Kleinjan W.E., de Keizer A., and Janssen A.J.H. 2005. Kinetics of the chemical
oxidation of polysulfide anions in aqueous solution. Water Research. 39(17): 4093-
4100.
Knöller K., and Schubert M. 2010. Interaction of dissolved and sedimentary sulfur
compounds in contaminated aquifers. Chemical Geology. 276(3): 284-293.
Koch M.S., Mendelssohn I.A., and McKee K.L. 1990. Mechanism for the hydrogen
sulfide-induced growth limitation in wetland macrophytes. Limnology and
Oceanography. 35(2): 399-408.
Koh T. 1990. Analytical chemistry of polythionates and thiosulfate a review. Analytical
Sciences. 6(1): 3-14.
Komarnisky L.A., Christopherson R.J., and Basu T.K. 2003. Sulfur: its clinical and
toxicologic aspects. Nutrition. 19(1): 54-61.
Komorsky-Lovrić S., Lovrić M., and Scholz F. 1997. Sulfide ion electroxidation
catalysed by cobalt phthalocyanine microcrystals. Microchimica Acta. 127(1-2): 95-
99.
Kotronarou, A., & Hoffmann, M. R. (1991). Catalytic autooxidation of hydrogen
sulfide in wastewater. Environmental Science and Technology;(United States), 25(6):
1153-1160.
Kristiana I., Heitz A., Joll C., and Sathasivan A. 2010. Analysis of polysulfides in
drinking water distribution systems using headspace solid-phase microextraction and
117
gas chromatography-mass spectrometry. Journal of Chromatography A. 1217(38):
5995-6001.
Kubáň V., Dasgupta P.K., and Marx J.N. 1992. Nitroprusside and methylene blue
methods for silicone membrane differentiated flow injection determination of sulfide
in water and wastewater. Analytical Chemistry. 64(1): 36-43.
Kuhn A.T., Kelsall G.H., and Chana M.S. 1983. A review of the air oxidation of
aqueous sulfide solutions. Journal of Chemical Technology and Biotechnology.
33(8): 406-414.
Lawrence N.S., Davis J., and Compton R.G. 2000(a). Analytical strategies for the
detection of sulfide: a review. Talanta. 52(5): 771-784.
Lawrence N.S., Davis J., Marken F., Jiang L., Jones T.G.J., Davies S.N., and
Compton R. 2000(b). Electrochemical detection of sulfide: a novel dual flow cell.
Sensors and Actuators B: Chemical. 69(1): 189-192.
Lawrence N.S., Davis J., Jiang L., Jones T.G.J., Davies S.N., and Compton R.
2000(c). The electrochemical analog of the methylene blue reaction: a novel
amperometric approach to the determination of hydrogen sulfide. Electroanalysis.
12(18): 1453-1460.
Lawrence N.S, Thompson M., Prado C., Jiang L., Jones T.G.J., and Compton R.G.
2002. Amperometric detection of sulfide at a boron doped diamond electrode: The
electrocatalytic reaction of sulfide with ferricyanide in aqueous solution.
Electroanalysis. 14(7-8): 499-504.
Lawrence N.S., Deo R.P., and Wang J. 2004. Electrochemical determination of
hydrogen sulfide at carbon nanotube modified electrodes. Analytica Chimica Acta.
517(1): 131-137.
118
Lawrence N.S. 2006. Amperometric Detection of Sulfide: An Electrocatalytic Reaction
with Ferrocene Carboxylate. Electroanalysis. 18(17): 1658-1663.
Lawrence N.S., Tustin G.J., Faulkner M., and Jones T.G.J. 2006. Ferrocene
sulfonates as electrocatalysts for sulfide detection. Electrochemica Acta. 52(2): 499-
503.
Lawrence J., Robinson K.L., and Lawrence N.S. 2007. Electrochemical determination
of sulfide at various carbon substrates: a comparative study. Analytical Sciences.
23(6): 673-676.
Le Faou A., Rajgopal B.S., Daniels L., and Fauque. 1990. Thiosulfate, polythionates
and elemental sulfur assimilation and reduction in the bacterial world. FEMS
microbiology reviews. 6(4): 351-381.
Legator, M.S., Singleton, C.R., Morris, D.L., and Philips, D.L. 2001. Health effects
from chronic low-level exposure to hydrogen sulfide. Archives of Environmental
Health: An International Journal. 56(2): 123-131.
Lessner P.M., McLarnon F.R., Winnick J., and Cairns E.J. 1993. The dependence of
aqueous sulfur-polysulfide redox potential on electrolyte composition and
temperature. Journal of the Electrochemical Society. 140(7):1847-1849.
Li G., and Miao P. 2013. Electrochemical Analysis of Proteins and Cells. Springer
Briefs in Molecular Science. Springer.
Licht S., and Davis J. 1997. Disproportionation of aqueous sulfur and sulfide: kinetics
of polysulfide decomposition. Journal of Physical Chemistry B. 101(14): 2540-
2545.
119
Lorvić M., and Pižeta I. 1992. A square-wave voltammetry in a cathodic stripping
mode. Electroanalysis. 4(3): 327-337.
Luther G.W. III., Giblin A.E., and Varsolona R. 1985. Polarographic analysis of
sulfur species in marine porewaters. Limnology Oceanography. 30(4): 727-736.
Luther G.W. III., Church T.M., and Powell D. 1991. Sulfur speciation and sulfide
oxidation in the water column of the Black Sea. Deep Sea Research. 38(S2): S1121-
S1137.
Ma L., Wang L., Tan Q., Yu H., Huo J., Ma Z., Hu H., and Chen Z. 2009. Study on
synthesis and electrochemical properties of novel ferrocene-based compounds and
their applications in anion recognition. Electrochimica Acta. 54(23): 5413-5420.
MacLeod A.J. 1993. A note on the randles-sevcik function from electrochemistry.
Applied mathematics and computation. 57(2-3): 305-310.
Maleki A., and Nematollahi D. 2009. An efficient electrochemical method for the
synthesis of methylene blue. Electrochemistry Communications. 11(12): 2261-2264.
Manan N.S.A., Aldous L., Alias Y., Murray P., Yellowlees L.J., Lagunas M.C., and
Hardacre C. 2011. Electrochemistry of sulfur and polysulfides in ionic liquids. The
Journal of Physical Chemistry B. 115(47): 13873-13879.
Manova A., Strelec M., Cacho F., Lehotay J., and Beinrohr E. 2007. Determination of
dissolved sulphides in waste water samples by flow-through stripping
chronopotentiometry with a macroporous mercury-film electrode. Analytica Chimica
Acta. 588(1): 16-19.
McCreery R.L. 2008. Advanced carbon electrode materials for molecular
electrochemistry. Chemistry Reviews. 108(7): 2646-2687.
120
Millero F. J. (1986). The thermodynamics and kinetics of the hydrogen sulfide system in
natural waters. Marine Chemistry, 18(2-4), 121-147.
Millero F.J., Hubinger S., Fernandez M., and Garnett S. 1987. Oxidation of H2S in
seawater as a function of temperature, pH, and ionic strength. Journal of
Environmental Science and Technology. 21(5): 439-443.
Miura Y., and Kawaoi A. 2000. Determination of thiosulfate, thiocyanate and
polythionates in a mixture by ion-pair chromatography with ultraviolet detection.
Journal of Chromatography A. 884(1): 81-87.
Miura Y., Matsushita Y., and Haddad P.R. 2005. Stabilization of sulfide and sulfite
and ion-pair chromatography of mixtures of sulfide, sulfite, sulfate, and thiosulfate.
Journal of Chromatography A. 1085(1): 47-53.
Mohapatra B.R., Gould W.D., Dinardo O., and Koren D.W. 2008. An overview of the
biochemical and molecular aspects of microbial oxidation of inorganic sulfur
compounds. CLEAN. 36(10-11): 823-829.
Mohtadi R., Lee W.K., and Van Zee J.W. 2005. The effect of temperature on the
adsorption rate of hydrogen sulfide on Pt anodes in a PEMFC. Applied Catalysis B:
Environmental. 56(1): 37-42.
Morse J.W., and Luther III G.W. 1999. Chemical influences on trace metal-sulfide
interactions in anoxic sediments. Geochimica et Cosmochimica Acta. 63(19/20):
3373-3378.
Morse J.W., Millero F.J., Cornwell J.C., and Rickard D. 1987. The chemistry of the
hydrogen sulfide and iron sulfide systems in natural waters. Earth-Science Reviews.
24(1): 1-42.
121
Mousty C., Forano C., Fleutot S., and Dupin J-C. 2009. Electrochemical study of
anionic ferrocene derivatives intercalated in layered double hydroxides: application
to glucose amperometric biosensors. Electroanalysis. 21(3-5): 399-408.
Mund K., Weidlich R.E., and Fahlström U. 1986. Electrochemical properties of
platinum, glassy carbon, and polygraphite as stimulating electrodes. Pacing and
Clinical Electrophysiology. 9(6): 1225-1229.
Murphy R., and Strongin D.R. 2009. Surface reactivity of pyrite and related sulfides.
Surface Science Reports. 64(1): 1-45.
Mylon S.E., Hu H., and Benoit G. 2002. Unsuitability of Cr (II) reduction for the
measurement of sulfides in oxic water samples. Analytical Chemistry. 74(3): 661-
663.
Nagaoka T., Yoshino T. 1986. Surface properties of electrochemically pretreated glassy
carbon. Analytical Chemistry. 58(6): 1037-1042.
Noel M., Anantharman P.N. 1986. Voltammetric studies on glassy carbon electrodes I:
electrochemical behavior of glassy carbon electrodes in H2SO4, Na2SO4 and NaOH
media. Surface and Coatings Technology. 28(2): 161-179.
O'Brien, D. J., & Birkner, F. B. (1977). Kinetics of oxygenation of reduced sulfur
species in aqueous solution. Environmental Science and Technology. 11(12): 1114-
1120.
O`Brien J.A., Hinkley J.T., Donne S.W., and Lindquist S-E. 2010. The
electrochemical oxidation of aqueous sulfur dioxide: a critical review of work with
respect to the hybrid sulfur cycle. Electrochimica Acta. 55(3): 573-591.
122
Ohira S-I., and Toda K. 2006. Ion chromatographic measurement of sulfide,
methanethiolate, sulfite and sulfate in aqueous and air samples. Journal of
Chromatography A. 1121(2): 280-284.
Olivetti E.A., Avery K.C., Taniguchi I., Sadoway D.R., and Mayes A.M. 2008.
Electrochemical characterization of vanadium oxide nanostructured electrode.
Journal of the Electrochemcial Society. 155(7): A488-A493.
O’Reilly J.W., Dicinoski G.W., Shaw M.J., and Haddad P.R. 2001. Chromatographic
and electrophoretic separation of inorganic sulfur and sulfur-oxygen species.
Analytica Chemica Acta. 432(2): 165-192
Oriji G., Katayama Y., and Miura T. 2005. Investigations on V(IV)/V(V) and
V(II)/V(III) redox reactions by various electrochemical methods. Journal of Power
Sources. 139(1): 321-324.
Ortiz J.A., Carbonell R.G., Camargo J.A., Nieto F., Reoyo M.J., and Tarazona J.V.
1993. Acute toxicity of sulfide and lower pH in cultured rainbow trout, Atlantic
salmon, and Coho salmon. Bulletins of Environmental Contamination and
Toxicology. 50(1): 164-170.
Ostapczuk P., Valenta P., and Nürnberg H.W. 1986. Square wave voltammetry – a
rapid and reliable determination method of Zn, Cd, Pb, Cu, Ni, and Co in biological
and environmental samples. Journal of Electroanalytical Chemistry. 214(1-2): 51-
64.
Paim L.L., and Stradiotto N.R. 2010. Electrooxidation of sulfide by cobalt
pentacyanonitrosylferrate film on glassy carbon electrode by cyclic voltammetry.
Electrochemica Acta. 55(13): 4144-4147.
123
Paim L.L., Hammer P., and Stradiotto N.R. 2011. Electrochemical behavior of a
glassy carbon electrode chemically modified with nickel pentacyanonitrosylferrate in
presence of sulfur compounds. Electroanalysis. 23(6): 1488-1496.
Pan C., Wang W., Horváth A.K., Xie J., Liu Y., Wang Z., Ji C., and Gao Q. 2011.
Kinetics and mechanism of alkaline decomposition of the pentathionate ion by
simultaneous tracking of different sulfur species by high-performance liquid
chromatography. Inorganic Chemistry. 50(19): 9670-9677.
Pan C., Liu Y., Horváth A.K., Wang Z., Ji C., Hu Y., Ji C., Zhao Y., and Gao Q.
2013. Kinetics and mechanism of the alkaline decomposition of hexathionate ion.
The Journal of Physical Chemistry A. 117(14): 2924-2931.
Pandey S.K., Kim K-H., and Tang K-T. 2012. A review of sensor-based methods for
monitoring hydrogen sulfide. TrAC Trends in Analytical Chemistry. 32: 87-99.
Park D.W., Park B.K, Park D.K., and Woo H.C. 2002. Vanadium-antimony mixed
oxide catalysts for the selective oxidation of H2S containing excess water and
ammonia. Applied Catalysis A: General. 223(1): 215-224.
Pawlak Z., and Pawlak A.S. 1999. Modification of iodometric determination of total
and reactive sulfide in environmental samples. Talanta. 48(2): 347-353.
Pecsok R.L., Shields D.L., Cairns T., and McWilliam I.G. 1976. Modern methods of
chemical analysis, 2nd edition. John Wiley, New York.
Peiffer S., Klemm O., Pecher K., and Hollerung R. 1992. Redox measurements in
aqueous solutions – a theoretical approach to data interpretation, based on electrode
kinetics. Journal of Contaminant Hydrology. 10(1): 1-18.
124
Peterson B.J., Steudler P.A., Howarth R.W., Friedlander A.I., Juers D., and Bowles
F.P. 1983. Tidal export of reduced sulfur from a salt marsh ecosystem. Ecological
Bulletins, 153-165.
Peterson G.S., Ankley G.T., and Leonard E.N. 1996. Effect of bioturbation on metal-
sulfide oxidation in surficial freshwater sediments. Environmental Toxicology and
Chemistry. 15(12): 2147-2155.
Petre C.F., and Larachi F. 2006. Capillary electrophoretic separation of inorganic
sulfur-sulfide, polysulfides, and sulfur-oxygen species. Journal of Separation
Science. 29(1): 144-152.
Pikaar I., Rozendal R.A., Yuan Z., Keller J. and Rabaey K. 2011. Electrochemical
sulfide removal from synthetic and real domestic wastewater at high current
densities. Water Research. 45(6): 2281-2289.
Pocard N.L., Alsmeyer D.C., McCreery R.L., Neenan T.X., and Callstrom M.R.
1992. Doped glassy carbon: a new material for electrocatalysis. Journal of Material
Chemistry. 2(8): 771-784.
Polack R., Chen Y-W., and Belzile N. 2009. Behavior of Sb(V) in the presence of
dissolved sulfide under controlled anoxic aqueous conditions. Chemical Geology.
262(3-4): 179-185.
Puacz W., Szahun W., and Linke K. 1995. Catalytic determination of sulfide in blood.
Analyst. 120(3): 939-941.
Qunitar S.E., Santafata J.P., and Cortinez V.A. 2005. Determination of vanadium (V)
by direct automatic potentiometric titration with EDTA using a chemically modified
electrode as a potentiometric sensor. Talanta. 67(4): 843-847.
125
Rajeshwar K., Ibanez J.G., and Swain G.M. 1994. Electrochemistry and the
environment. Journal of Applied Electrochemistry. 24(11): 1077-1091.
Rajeshwar K. 1995. Photoelectrochemistry and the environment. Journal of Applied
Electrochemistry. 25(12): 1067-1082.
Raposo F. 2016. Evaluation of analytical calibration based on least-squares linear
regression for instrumental techniques: a tutorial review. Trends in Analytical
Chemistry. 77: 167-185.
Redinha J.S., Paliteiro C., and Pereira J.L.C. 1997. Determination of sulfide by
square-wave polarography. Analytica Chimica Acta. 351(1): 115-125.
Reese B.K., Finneran D.W., Mills H.J., Zhu M-X., and Morse J.W. 2011.
Examination and refinement of the determination of aqueous hydrogen sulfide by the
methylene blue method. Aquatic Geochemistry. 17(4-5): 567-582.
Reiffenstein R. J., Hulbert W. C., and Roth S. H. 1992. Toxicology of hydrogen
sulfide. Annual Review of Pharmacology and Toxicology. 32(1): 109-134.
Rohwerder T. and Sand W. 2007. Oxidation of inorganic sulfur compounds in
acidophilic prokaryotes. Engineering Life Science. 7(4): 301-309.
Roman P., Bijmans M.F.M., and Janssen A.J.H. 2014. Quantification of individual
polysulfides in lab-scale and full-scale desulfurization bioreactors. Environmental
Chemistry. 11(6): 702-708.
Rorabacher D.B. 1991. Statistical treatment for rejection of deviant values: critical
values of Dixon’s “Q” parameter and related subrange ratios at the 95% confidence
level. Analytical Chemistry. 63(2): 139-146.
126
Rosén E., and Tegman R. 1971. A preparative and X-ray powder diffraction study of
the polysulfides Na2S2, Na2S4 and Na2S5. Acta Chemica Scandinavica. 25: 3329-
3336.
Rozan T.F., Benoit G., and Luther III G.W. 1999. Measuring metal sulfide complexes
in oxic river waters with square wave voltammetry. Environmental Science and
Technology. 33(17): 3021-3026.
Rozan T.F., Lassman M.E., Ridge D.P., and Luther III G.W. 2000(a). Evidence for
iron, copper and zinc complexation as multinuclear sulphide clusters in oxic rivers.
Nature. 406(6798): 879-882.
Rozan T.F., Theberge S.M., and Luther G. III. 2000(b). Quantifying elemental sulfur,
bisulfide and polysulfides using a voltammetric method. Analytica Chemica Acta.
415(1-2): 175-184.
Safavi A., and Ramezani Z. 1997. Kinetic spectrophotometric determination of traces of
sulfide. Talanta. 44(7): 1225-1230.
Sangster, J., and Pelton, A. D. 1997. The Na-S (sodium-sulfur) system. Journal of
phase equilibria, 18(1): 89-96.
Scholz, F. (2010). Electroanalytical methods (2nd Ed). Berlin-Heidelberg: Springer.
Serafim D.M., and Stradiotto N.R. 2008. Determination of sulfur compounds in
gasoline using mercury film electrode by square wave voltammetry. Fuel. 87(7):
1007-1013.
Sharma V.K. 2002. Potassium ferrate (VI): an environmentally friendly oxidant.
Advances in Environmental Research. 6(2): 143-156.
127
Shin M.Y., Nam C.M., Park D.W., and Chung J.S. 2001. Selective oxidation of H2S to
elemental sulfur over VOx/SiO2 and V2O5 catalysts. Applied Catalysis A: General.
211(2): 213-225.
Sievert S.M., Kiene R.P. and Schulz-Vogt H.N. 2007. The sulfur cycle. Oceanography.
20(2): 117-123.
Sifuna F.W., Orata F., Okello V., and Jemutai-Kimosop S. 2016. Comparative studies
in electrochemical degradation of sulfamethoxazole and diclofenac in water by using
various electrodes and phosphate and sulfate supporting electrolytes. Journal of
Environmental Science and Health, Part A. 51(11): 954-961.
Sonne K., and Dasgupta P.K. 1991. Simultaneous photometric flow injection
determination of sulfide, polysulfide, sulfite, thiosulfate and sulfate. Analytical
Chemistry. 63(5): 427-432.
Sorokin Y.I. 2011. On the rates of sulfide formation and oxidation in the black sea
during the cold season. Oceanology. 51(6): 969-977.
Stefansson A., Arnorsson S., and Sveinbjornsdottir A.E. 2005. Redox reactions and
potentials in natural waters ay disequilibrium. Chemical Geology. 221(3-4): 289-311.
Stefansson A., Gunnarsson I., and Giroud N. 2007. New methods for the direct
determination of dissolved inorganic, organic and total carbon in natural waters by
Reagent-Free(TM) Ion Chromatography and inductively coupled plasma atomic
emission spectrometry. Analytica Chimica Acta. 582(1): 69-74.
Steudel R. 2003. Inorganic Polysulfides Sn2- and radical anions Sn
·-. In Elemental Sulfur
und Sulfur-Rich Compounds II. pp. 127-152. Springer Berlin Heidelberg.
128
Steudel R., and Holt G. 1988. Solubilization of elemental sulfur in water by cationic and
anionic surfactants. Angewandte Chemie International Edition in English. 27(10):
1358-1359.
Steudel R., Göbel T., and Holt G. 1989. The molecular nature of the hydrophilic sulfur
prepared from aqueous sulfide and sulfite (selmi sulfur sol). Zeitschrift für
Naturforschung B. 44(5): 526-530.
Syed M., Soreanu G., Falletta P., and Béland M. 2006. Removal of hydrogen sulfide
from gas streams using biological processes – a review. Canadian Biosystems
Engineering 48: 2-14.
Szpyrkowicz L., Juzzolino C., and Kaul S.N. 2001. A comparative study on oxidation
of disperse dye by electrochemical process, ozone, hypochlorite and fenton reagent.
Water Research. 35(9): 2129-2136.
Tabrizivand G., Sabzi R.E., and Farhadi K. 2007. Preparation and characterization of
a new carbon paste electrode based on ketotifen-hexacyanoferrate. Journal of Solid
State Electrochemistry. 11(1): 103-108.
Tang D., and Santschi P.H. 2000. Sensitive determination of dissolved sulfide in
estuarine water by solid-phase extraction and high-performance liquid
chromatography of methylene blue. Journal of Chromatography A. 883(1): 305-309.
Tang K., Baskaran V., and Nemati M. 2009. Bacteria of the sulfur cycle: an overview
of microbiology, biokinetics and their role in petroleum and mining industries.
Biochemical Engineering Journal. 44(1): 73-94.
Taverniers I., De Loose M., and Van Bockstaele E. 2004. Trends in quality in the
analytical laboratory. II. Analytical method validation and quality assurance. Trends
in Analytical Chemistry. 23(8): 535-552.
129
Taylor R.J. and Humffray A.A. 1973. Electrochemical studies on glassy carbon
electrodes I. Electron transfer kinetics. Electroanalytical Chemistry and Interfacial
Electrochemistry. 42(3): 347-354.
Tegman R. 1976. Thermodynamic studies of high temperature equilibria. Chemica
Scripta. 9(4): 158-166.
Thiyagarajan, N., Chang, J. L., Senthilkumar, K., and Zen, J. M. 2014. Disposable
electrochemical sensors: A mini review. Electrochemistry Communications. 38: 86-
90.
Titova T.V., Borisova N.S. and Zakharchuk N.F. 2009. Determination of sub-
micromolar amounts of sulfide by standard free anodic stripping voltammetry and
anodic stripping voltammetric titration. Analytica Chemica Acta. 653(2): 154-160.
Tolstikov G.A., Shul’ts E.E., and Tolstikov A.G. 1997. Natural Polysulfides. Russian
Chemical Reviews. 66(9): 813-826.
Tomar M., and Abdullah T. H.A. 1994. Evaluation of chemicals to control the
generation of malodorous hydrogen sulfide in waste water. Water Research. 28(12):
2545-2552.
Tse Y-H., Janda P., and Lever A.B.P. 1994. Electrode with electrochemically deposited
N, N’, N”, N’”-tetramethyltetra-3,4-pyridinoporphyrazino-cobalt(I) for detection of
sulfide ion. Analytical Chemistry. 66(3): 384-390.
Umiker K.J., Morra M.J., and Cheng F.I. 2002. Aqueous sulfur species determination
using differential pulse polarography. Microchemical Journal. 73(3):287-297.
Van der Linden W.E. and Dieker J.W. 1980. Glassy Carbon as electrode material in
electroanalytical chemistry. Analytica Chimica Acta. 119(1): 1-24.
130
von Wandruska R., and Yuan X. 1993. Determination of sulfur species by cathodic
square wave stripping voltammetry; compounds relevant to natural sulfur
mineralization. Talanta. 40(1): 37-42.
Waite T.J., Kraiya C., Trouwborst R.E., Ma S., and Luther III G.W. 2006. An
investigation into the suitability of bismuth as an alternative gold-amalgam as a
working electrode for the in situ determination of chemical redox species in the
natural environment. Electoanalysis. 18(12): 1167-1172.
Wang J., and Hutchins L.D. 1985. Activation of glassy carbon electrodes by alternating
current electrochemical treatment. Analytica Chimica Acta. 167(): 325-334.
Wang J. 2005. Stripping analysis at bismuth electrodes: a review. Electroanalysis.
17(15-16): 1341-1346.
Wang H., and Pilon L. 2012. Physical interpretation of cyclic voltammetry for
measuring electric double layer capacitances. Electrochimica Acta. 64: 130-139.
Wardencki W. 1998. Problems with the determination of environmental sulfur
compounds by gas chromatography. Journal of Chromatography A. 793(1): 1-19.
Weckhuysen B.M., and Keller D.E. 2003. Chemistry, spectroscopy and the role of
supported vanadium oxides in heterogeneous catalysis. Catalysis Today. 78(1): 25-
46.
Wetchakun K., Samerjai T., Tamaekong N., Liewhiran C., Siriwong C., Kruefu V.,
Wisitsoraat A., Tuantranont A., and Phanichphant S. 2011. Semiconducting
metal oxides as sensors for environmentally hazardous gases. Sensors and Actuators
B: Chemical. 160(1): 580-591.
131
Whitson P.E., VandenBorn H.W., and Evans D.H. 1973. Acquisition and analysis of
cyclic voltammetric data. Analytical Chemistry. 45(8): 1298-1306.
Witter A.E. and Jones D.A. 1998. Comparison of methods for inorganic sulfur
speciation in a petroleum production effluent. Environmental Toxicology and
Chemistry. 17(11): 2176-2184.
Wring S.A., and Hart J.P. 1992. Chemically modified, carbon-based electrodes and
their application as electrochemical sensors for the analysis of biologically important
compounds. Analyst. 117(8): 1215-1229.
Xie Y., and Huber C.O. 1991. Electrocatalysis and amperometric detection using an
electrode made of copper oxide and carbon paste. Analytical Chemistry. 63(17):
1714-1719.
Xu Y., Schoonen M.A.A., Nordstrom D.K., Cunningham K.M., and Ball J.W. 1998.
Sulfur geochemistry of hydrothermal waters in Yellowstone National Park: I. the
origin of thiosulfate in hot spring waters. Geochimica et Cosmochimica Acta. 62(23-
24): 3729-3743.
Xu Y., Schoonen M.A.A., Nordstrom D.K., Cunningham K.M., and Ball J.W. 2000.
Sulfur geochemistry of hydrothermal waters in Yellowstone National Park,
Wyoming, USA. II. Formation and decomposition of thiosulfate and polythionate in
Cinder Pool. Journal of Volcanology and Geothermal Research. 97(1-4): 407-423.
Yakushev E. V., and Newton A. 2012. Introduction: redox interfaces in marine waters.
In Chemical Structure of Pelagic Redox Interfaces pp. 1-12. Springer Berlin
Heidelberg.
132
Yang B., Wang S., Tian S., and Liu L. 2009. Determination of hydrogen sulfide in
gasoline by Au nanoclusters modified glassy carbon electrode. Electrochemistry
Communication. 11(6): 1230-1233.
Yao W., and Millero F.J. Oxidation of hydrogen sulfide by hydrous Fe(III) oxides in
seawater. Marine Chemistry. 52(1): 1-16.
Zen J-M., Kumar A.S., and Tsai D-M. 2003. Recent updates of chemically modified
electrodes in analytical chemistry. Electroanalysis. 15(13): 1073-1087.
Zen J-M., Chang J-L., Chen P-Y., Ohara R., and Pan K-C. 2005. Flow injection
analysis of sulfide using a cinder/tetracyano nikelate modified screen-printed
electrode. Electroanalysis. 17(9): 739-743.
Zhang, J. Z., and Millero, F. J. (1991). The rate of sulfite oxidation in
seawater. Geochimica et Cosmochimica Acta. 55(3): 677-685.
Zhang J., and Millero F.J. 1993. The products from the oxidation of H2S in seawater.
Geochimica et Cosmochimica Acta. 57(8): 1705-1718.
Zhang J., and Millero F.J. 1994. Investigation of metal sulfide complexes in sea water
using cathodic stripping square wave voltammetry. Analytica Chemica Acta. 284(3):
497-504.
Zhang L., De Schryver P., De Gusseme B., De Muynck W., Boon N., and Verstraete
W. 2008. Chemical and biological technologies for hydrogen sulfide emission
control in sewer systems: a review. Water Research. 42(1): 1-12.
Zhao L., Zhu X., Feng., and Wang B. 2006. Speciation analysis of inorganic vanadium
(V(IV)/V(V)) by graphite furnace atomic absorption spectrometry following ion-
exchange separation. International Journal of Analytical Chemistry. 86(12): 931-939.
133
Curriculum Vitae: Ashley Ryan (nee Marcellus)
SUMMARY OF QUALIFICATIONS
Central Analytical Facility (CAF) Technologist – Nipissing University
The CAF is a unique part of Nipissing University’s Research Facility. A dedicated
technologist supervises, coordinates, and performs its day-to-day operations, providing
training and support to students and faculty researchers, external clients in academe, and
government or industry clients. Six pieces of equipment requiring specialized technical
support include: High Pressure Liquid Chromatograph (HPLC), Gas Chromatograph
Mass Spectrometer (GC-MS), Fourier-Transformation Infra-Red Spectrometer (FTIR),
Nuclear Magnetic Resonance Spectrometer (NMR), Atomic Absorption Spectrometer
(AAS), Electrochemical Analyzer, and a Microplate Reader. Two additional pieces of
equipment are the Confocal Microscope and Electron Microscope. These microscopes
are also maintained and operated by an individual with specialized technical training.
This suite of research instruments is integral to the research and scientific training
missions of the University and represents a significant component of the infrastructure
and technical capacity for scientific research at NU.
The CAF technologist ensures the functioning and integrity of all equipment in the
research facility. Each instrument involves a unique set of tasks and functions that need
to be performed on a regular basis. The skills required to maintain the equipment in the
CAF must be constantly updated, and they continue to evolve as the equipment ages.
Among the many duties of the CAF technologist, one important role is as a liaison,
facilitating interactions among students, faculty, and administrators in support of the
success of the research programs at this institution. The CAF technologist is involved in
the collection of data, but also is responsible for the efficient scheduling and optimal use
of equipment. It is critical that research deadlines (especially for thesis student
researchers) are met with all parties working together smoothly.
Over the past seven years as the CAF technologist I have worked diligently to support
research initiatives at Nipissing University. I am a versatile, hard-working, innovative
individual with a practical hands-on approach; I always strive for excellence. I have a
strong work ethic, and dedicate myself to ensuring that a job is done correctly. Applying
my analytical and problem solving skills, I routinely identify challenges and develop
effective solutions so that researchers will be able to collect meaningful, high quality
results.
PROFILE SUMMARY
A versatile, hard-working, innovative individual with a practical hands-on approach who
always strives toward excellence. Ability to collect, analyze and interpret data and
quickly grasp complex issues. Excellent interpersonal, analytical and problem solving
134
skills, promptly identifying problems and developing effective solutions. Proven ability
to complete projects to highest standard and meet deadlines with meticulous attention to
detail.
EDUCATION Masters of Environmental Science
Nipissing University, North Bay
2012-Present
Research Focus: Evaluating the efficiency of cyclic voltammetry using modified and
unmodified glassy carbon electrodes for the analysis of sulfur containing compound in
water.
Expected graduation date: October 2017
Honors Bachelor of Science Biology
Nipissing University, North Bay
2003-2008
Thesis Project: Characterization of senescence patterns of Eriophorum vaginatum
Related course work: General and Analytical Chemistry, Microbiology, Plant
Physiology and Fresh Water Biology.
Equipment: Olympus FV1000 Laser Scanning Confocal Microscope
ON THE JOB TRAINING
Bruker Fourier NMR 300
Bruker Ltd. Dec 2012
Supervised installation of NMR equipment into the CAF
Onsite training provided by Bruker NMR Specialists for Operation
and Maintenance
Maintenance: Filling of liquid Nitrogen (every week); Filling of
liquid Helium (every 4-5 months) to ensure stability of the supercoil
magnet. Updating shim files, field and phase values for optimal
spectral performance
Operation: Acquisition of proton and carbon NMR spectra, high
temperature NMR parameters, recognizing bad spectra and what can
be done to correct it. Troubleshooting issues with shimming the
magnetic pull to optimize the spectra. Inputting parameters for new
solvents. How to prepare samples for NMR analysis.
Since the installation of the NMR in CAF the equipment has yielded
several peer-reviewed journals from 2013 to present. These are listed
on Dr. Mukund Jha’s webpage. Data collected from NMR have been
collected by myself, BSc Honours students and research interns
135
Solid Phase Extraction (SPE) Seminar
Waters Ltd. Mississauga, ON
Sept 2012
Introduction to importance to SPE to analytical analyses
Sample preparation
Isolation of analyte to match choice of bedding in SPE
Optimization for maximum analyte recovery
Troubleshooting issues with SPE
SFR: Electron Microscope Commissioning
Nipissing University, North Bay May 2010
On Site training of Philips CM 10 Transmission Electron Microscope
provided by Peter Maloney from SFR
Operation: alignment of electron beam, optimizing viewing
conditions based on specific sample, alignment of column for
optimum performance, etc.
Maintenance and care: Removing parts of the column and cleaning
the appropriate parts, water cooling maintenance, filter changes, water
changes to reduce bacterial growth within the smaller pipes of the
microscope, etc.
Training received allowed me to train other students in basic operation
of the EM scope; this has supported the research of graduate students,
honors students and faculty
John Dolan HPLC Troubleshooting/Diagnostics Work Shop
Marriot Hotel, Vaughan June 2010
Troubleshooting principles
o Measurements and basic practices
Performance Qualification
o Quality/validity of results as well as accurate reproducibility
Column Physics/Chemistry
o Why and how columns die; problems with the stationary
phase; how and when to avoid contamination (samples/mobile
phase)
Pumps and Autosamplers
o Maintenance to pump heads and check valves; degassing
solvents for mobile phase; what type of tubing is suitable for
each type of analyses; how to recognize failure in frits;
blockages and leaks.
Detectors
o UV detectors noise and drift in baseline; wavelength selection
and other types of detectors available (PDA-Photo Diode
Array)
Quantification
o Measuring peak area/height; issues with calibration curves and
reproducibility issues.
136
Olympus Confocal FV1000 Laser Scanning Confocal Microscope
Nipissing University, North Bay August 2007
Received onsite training from Vince Varallo (Olympus Canada)
Operation: Image acquisition control for laser optimization and
operation, acquisition setting for scan rate function and position of
specimen for optimum performance, recognizing spectral bleed and
how to correct for it, etc.
Maintenance: Recognizing signs of degradation in signal, alignments
and replacements of bulbs/diodes; proper cleaning of parts for
operation i.e. objective pieces, condensers, etc.
Web Seminars: Analytical Content
One Hour in Length, North Bay
Started viewing in March 2010 to Present (2016) Wide variety of topic are covered over these presentations
Topics include but are not limited to: troubleshooting equipment issues, samples prep,
appropriate column choices, new techniques/technologies, etc.
EXPERIENCE Central Analytical Facility Technologist
Nipissing University, North Bay
March 2009-Present
Recognizing, troubleshooting, fixing issues with all equipment in CAF, and other labs Preparing reagents and solutions on a regular basis for ongoing projects Analyzing samples for researchers, professors and collaborators Writing test reports for professors, and collaborators Setting up instruments based on methods and protocols for CAF users Maintaining all equipment in CAF, Confocal Microscope, Electron Microscope, water
systems, equipment in teaching labs Keeping records pertaining to replacement of parts either for maintenance or repair Keeping track of financials for the CAF, maintenance, and supplies budget Training students to operate instruments Writing and updating Standard Operation Procedures for each instrument Developing new analytical methods or protocols to meet research requests Developing quality control procedures for specific analytical methods
Plant Growth Facilities Intern
Nipissing University, North Bay
June 2008-March 2009
Maintained cleanliness of Greenhouse
Contained pest outbreak, using appropriate measures
Assisted with ongoing research projects conducted in the Greenhouse
Assisted students on their current projects
Learned to operate mechanical components as well as the Argus operation program that
provides climate and irrigation control
137
Research Assistant Chemistry
Biology Department
Nipissing University, North Bay
May 2005-Aug 2005
Followed existing research protocols
Validate quality of data, elements and editing
Collaborate with senior researchers in order to ensure accurate research procedures and results
Conducted assays using mine tailings and acidic solutions to isolate individual elements
PUBLISHED WORK , AWARDS, COMMUNTIY SERVICE
Rao V.K., Kaswan P, Shelke G.M., Ryan A., Jha M and Kumar A. 2015. Iodine-Mediated, Microwave-
Assisted Synthesis of 1-Arylnaph-thofurans via Cyclization of 1-(1′-Arylvinyl)-2-naphthols. Synthesis.
Vol 47, pp. 3990 – 3996
Jha, M., Edmunds M., Lund K., and Ryan A. 2014. A new route to the versatile synthesis of thiopyrano
[2,3-b:6,5-b’] diindoles via 2-(alkylthio)-indole-3-carbaldehydes. Tetrahedron Letters. Vol. 55; pp.
5691-5694.
Mirza, R. S., Laraby, C. A., & Marcellus, A. 2013. Knowing Your Behaviour: The importance of
Behavioural Assays in the Characterisation of Chemical Alarm Cues in Fishes and Amphibians.
In Chemical Signals in Vertebrates 12 (pp. 295-308). Springer New York.
Jha, M., Enaohwo, O. & Marcellus, A. 2009. Chemoselective S-benzylation of indoline-2-thiones using
benzyl alcohols. Tetrahedron Letters. Vol. 50(51); pp. 7184-7187.
Certificate of appreciation for Science Day (Feb 22, 2014); volunteered knowledge about
importance of water that engaged the youths of Scouts Canada and Girl Guides of Canada in
fun hands-on lab activities.
Image of Distinction Category for Nikon Small Worlds 2009 Competition; two images of the
four were placed in this category
Judge in North Bay Regional Science Fair (April 2015, April 2016)
Judge for North Bay Regional Robotics Competition (March 2016)