Upload
others
View
10
Download
0
Embed Size (px)
Citation preview
PN 306622565A
ARCHIMEDE AUTOMATIC BLOOD COMPONENT EXTRACTOR
ARCHIMEDE
Available in select markets only.
2 |
DELIVERING CONSISTENT, HIGH-QUALITY
BLOOD COMPONENTS
Take control of your blood component extraction process with Archimede.
Designed to deliver flexibility and efficiency, the device provides control and
management of extraction and helps you produce high-quality products.
Available in select markets only.
3 |
Exclusive Distribution
▪MOELCA SRL
▪ 35+ years of experience and expertise in medical devices
▪ Factory based in northern Italy
▪ ISO certifications: 9001, 9002 and 13485
▪ 100% focused on medical devices
Providing customers with a state-of-the-art automated blood component extractor
4 |
Archimede Features and Benefits
5 |
▪ High-quality blood components
▪ Reduction of repetitive strain injuries (RSIs)
▪ Increased productivity*
▪ Reliability
▪ Reduced noise
▪ Easy-to-load system and user-friendly interface
▪ Fast technical service
Meeting Needs for the Separation of Blood Components
*Compared to manual separation process of blood components
6 |
Archimede Overview
Automatic air removal from
plasma bag
Quickly and easily removes
remaining air from the plasma
bag to maintain the integrity of
the plasma when frozen
Saline adenine glucose
mannitol (SAG-M) press
Automates the priming of inline
red blood cell (RBC) filters with
SAG-M
Graphical display
Display of bag weights, plate
force and status of various
operating phases
Three cannula breakers
Automatically break cannulas
on lines to whole blood (WB),
RBC and SAG-M bags and are
compatible with other blood
bags on the market
Six clamps
Colored tubing pathway next to
each clamp for easy blood bag
loading and automatic
detection of correct tube
placement
Highly accurate weight
scales
Main press scale, RBC/lateral
scale and plasma scale
High-precision separation
Electric press with 18 optical
sensors that detect the blood
component interface for
accurate separation
Advanced software
Bidirectional data transmission
via LAN and WLAN,
compatible with optional RFID
and omnidirectional barcode
reader, self-diagnostic program
for remote service
7 |
▪ CE Mark
▪ Classification CND: B0180–For Blood Bags–Accessories
▪ Classified as Accessory to Blood Bag (medical device)
Medical Device Classification
8 |
Cannula Breakers
▪ Three cannula breakers to automatically break frangible connectors in blood bag tubing
▪ WB
▪ RBC
▪ SAG-M
▪ Can be adjusted to accommodate a variety of blood bags
▪ No special tools or additional inserts are required
▪ Cannula breakers automatically detect:
▪ Whether the cannula is correctly broken
▪ A mechanical obstruction preventing the breaker from moving
▪ Defective position
Reduce repetitive strain injuries (RSIs)
9 |
Electric Engine With 18 Sensors
▪ Extremely stable interface (high platelet recovery, no
need to siphon platelets after separation)
▪ Ability to adjust interface position by increments of
0.1 mm
▪ Auto-calibration of sensors to work with both lipemic
and green plasma
▪ Very quiet operation, with adjustable pressure force
from 30 kg to 60 kg
For high blood component stability*
Array with
18 sensors
T&B 30 kg T&B 40 kg
WB
Volume including CPD (mL) 526.3 ± 2.7 527.6 ± 4.3
Hct of WB 40 ± 4.4 38 ± 4.6
Hb per unit (grams) 66.3 ± 8.1 63.4 ± 8.3
PLT/unit (1 × 109) 106.1 ± 23.2 101 ± 22.4
PlasmaVolume (mL) 282.7 ± 13.1 291.5 ± 18.8
PLT in plasma (1 × 109) 1.5 ± 0.6 1.9 ± 0.7
BC
Volume (mL) 49.3 ± 0.8 44.4 ± 1.1
BC Hct (%) 39.4 ± 3.8 43.3 ± 4.6
WBC in BC (1 × 106) 2,097 ± 307 1,816 ± 357
Hb (gram per BC) 5.7 ± 0.6 5.6 ± 0.7
RCC
filtered
Volume (mL) 261.3 ± 13.5 260.8 ± 19.5
Hb per unit (gram) 49 ± 4.4 49.2 ± 6.3
Hct (%) 61 ± 2.2 61.1 ± 3.2
WBC (nrs per unit) 14,801 ± 14,890 7,649 ± 6,537
*Data on file at Terumo BCT
T&B = top and bottom (blood bags); WB = whole blood; BC = buffy coat; RCC = red blood cell
concentrate; CPD = citrate phosphate dextrose; Hct = hematocrit; Hb = hemoglobin;
PLT = platelet; WBC = white blood cell
10 |
Smart Tube Handling
▪ Indicates status with colored LED lights
▪ Facilitates tube placement
▪ Detects incorrect tube insertion
▪ Detects moisture to maintain durability
Ease of use
11 |
Air Removal From Plasma Bag
▪ For fractionators, blood banks and hospitals
▪Mechanical press to create a pressure on the plasma bag, combined with an optical sensor to
trigger clamping and sealing
▪ Small loss of plasma (2 mL to 5 mL) that is returned to the buffy coat
▪ Removable, based on process requirements
Low discard rates
Air removal
press
12 |
SAG-M Press
▪ Priming of RBC filter on the machine with SAG-M
▪ Further simplification of the separation process
▪ Foam-covered back to completely empty SAG-M
▪ Removable, based on process requirements
Process automation
SAG-M
cannula breaker
Foam-covered back
13 |
Stand-Alone Bench Equipment
▪ Can be used as stand-alone:
▪ Weighing scale
With barcode scanner
2 kg maximum, with resolution of 1 g
Any of the three scales can be used
▪ Tube sealer
Any of the four sealing heads can be used
Can also be used as a clamp in combination with the weighing scale
Increased flexibility
14 |
IT Connectivity
▪ Can work in the following modes:
▪ Stand-alone: configured via a PC
▪ Unidirectional: for full traceability with your blood banking information system (BBIS)
▪ Bidirectional: leverage donation-related information in your BBIS
Meet the needs of your customers
Archimede Computer BBIS
15 |
Remote Troubleshooting
State-of-the-art technical service
Note: Internet connection is required for remote troubleshooting.
16 |
THANKYOU
APPENDIX
18 |
Factory or Field Installable RFID Reader (Optional)
Integrated RFID reader
19 |
▪ Allows simultaneous scanning of all barcodes on a label
Factory or Field Installable Omnidirectional Barcode Reader
(Optional)
20 |
Specifications
SPECIFICATION TYPE VALUE
Product code MOE1504TER
Voltage frequency 50/60 Hz
Power 90 VAC 84 W 1A, 250 VAC 81 W 0.4A
Fuses N.2 fuses T2 AT 250 VAC
Size (W × D × H) 455 mm × 540 mm × 670 mm
Weight 48 kg
High-frequency generator Adjustable sealing time from 0.5 to 4 seconds, frequency 40.68 MHz
Standard accessories Barcode reader, WLAN antenna, filter holder, side scale support, power cable, plasma bag air removal system, side bag
holders and forked breaker, profile plates, SAG-M press
Optional accessories Omnidirectional barcode reader, RFID reader, splash guard, BlooDonorsLink
Communication LAN and WLAN for bidirectional data transfer procedure and configuration
PS2 port to connect the barcode reader
Operating conditions Temperature from +5 °C to +45 °C with < 80% RH, non-condensing
Storage conditions Temperature from −20 °C to +70 °C with 20% to 90% RH, non-condensing
European classification Class I, non-sterile accessory without measuring function to a medical device according to Council Directive 93/42/EEC, as
amended.
Regulatory compliance UNI EN ISO 13485:2012, EN60601-1 :2006, EN60601-1-2 :2007, EN60601-1-4 :1999, EN 60601-1-6 :2010, EN 60601-1-8
:2007, EN 301 489-17 V2.1.1 :2009, 2011/65/EU.