APPLICATIONS
FLS1000
2
STEADY STATE AND TIME-RESOLVED PHOTOLUMINESCENCE
The FLS1000 is a state-of-the-art, modular photoluminescence spectrometer for the most
demanding applications in Photophysics, Photochemistry, Material Sciences and Life Sciences.
The instrument sets the standards in both steady state and time-resolved spectroscopy:
The system demonstrates unmatched sensitivity and can be configured for spectral measurements
from the ultraviolet to the mid-infrared spectral range, and for lifetime measurements spanning
time resolutions over 12 orders of magnitude from picoseconds to seconds.
A N E W E R A I N P H OTO LU M I N E SC E N C E
Higher Resolution
New SoftwareFaster Measurements
Increased Automation
FLS1000
Modular construction for maximum flexibility and upgradability
Industry leading sensitivity specification >35,000:1 (SQRT Method)
Unrivalled spectral coverage from the deep UV to the MIR,
185 nm up to 5,500 nm
Unmatched monochromator performance with "plug and play"
triple-grating turrets, integrated filter wheel, 325 mm focal length
and excellent stray light rejection
Multiple light sources, detectors, single or double monochromators available
Intuitive Fluoracle® software for all steady state and time-resolved
measurements with standard and advanced data analysis options
3
AP
PLI
CA
TIO
NS
OV
ER
VIE
W
Time-Resolved Anisotropy
TRES
Upconversion materials
Nanospheres
Nanowires
Nanotubes
Quantum Wells
Defects
Phonons
ExcitonsSemiconductor
Band-Gap
Natural Dyes
Quantum DotsQuantum Rods
Fluorescent Beads
EEM
FRET
Chromaticity
Pressure Dependence
Temperature Dependence
Synthetic Dyes
Titration
Circularly-Polarised Luminescence
Spectral Anisotropy
Material Sciences• Phosphors for the Lighting and Display Industry • Doped Glasses and Ceramics • Laser Crystals and Light-Emitting Materials • Semiconductors Manufacture and Diagnostics • Materials for Light Harvest, Energy Conversion
and Energy Storage• Fluorescent Inks
Life Sciences• Fluorescence Based Assays• Drug Discovery • Detection of Toxins • DNA Sequencing• Research of Molecular Processes
and Mechanisms • Photosensitisers Research • Singlet Oxygen Detection
Environmental Sciences• Photo-oxidation of Environmental Pollutants• Detection of Total Organic Contents• Detection of Toxic Materials
Forensic Sciences and Security• Authentication and Identification of Counterfeits• Detection and Identification of Explosives
Geology• Identification of Materials by Fluorescence
and Phosphorescence• Age Determination• Trace Analysis
Request application notes or test measurements by contacting [email protected]
Laser Materials
Photovoltaic Materials
KEY:
Sample
Technique
Lanthanides
AP
PLI
CA
TIO
NS
IN M
ATE
RIA
L SC
IEN
CE
S
4
YTa7O19: Er3+-Yb3+ phosphor powder, excited with 976 nm continuous laser.Top insert: Integrated intensities versus excitation irradiance.
NaYF4:Er3+-Yb3+ phosphor powder, excited with a 976 nm microsecond pulsed laser and measured with MCS.
Insert: emission spectra of the three bands chosen for lifetime measurements.
100
101
102
103
Cou
nts
0 1000 3000 5000 7000 9000
Time / µs
4S3/2 -
4I15/2
Wavelength / nm
500 520 540 560 580 600 620 640 660 680 700
Cou
nts
/ 10
5
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
Upconversion materials
4F9/2 -
4I15/2
Wavelength / nm
Cou
nts
/ 10
2
0.0
1.0
2.0
3.0
4.0
5.0
6.0
350 400 450 500 550 600 650 700 750
Inte
nsity
/ a.
u.
Wavelength / nm350 400 450 500 550 600 650 700 750
Wavelength / nm
6.0
5.0
4.0
3.0
2.0
1.0
0.0
Inte
nsity
/ a
.u.
Inte
gra
ted
inte
nsity
Irradiance (W/cm2)
0 10 20 30 40 50 60103
104
105
106
107
108
PHOTOLUMINESCENCE OF LANTHANIDES - UPCONVERSION
Lanthanide materials have wide applications in the lighting industry, as laser gain media, in solar cells, as fluorescent labels, and in the life science industry as sensor materials.
The electronic structure of lanthanide materials causes bright emission that often can have narrow-band structure and can favour non-linear upconversion processes.
The FLS1000 offers exceptional high spectral resolution, provides stray light suppression to discriminate between weak emission and scatter from the white powder sample, and has a high dynamic range to detect spectral bands that can have dramatically different amplitude. Upgraded with a 980 nm continuous laser and thanks to a built-in computer- controlled continuous attenuator, the FLS1000 is ideal for upconversion power density experiments.
The lifetimes of the excited lanthanide states can vary from nanoseconds to seconds. The decay kinetics may be of simple exponential nature or can exhibit complex lifetime distributions. The individual emission bands in upconversion materials often show rising components too.
The FLS1000 can operate diode and solid state lasers in pulsed mode.
The graph contains lifetime measurements of the three main emission bands of the upconverted signal.
5
Spectral and lifetime measurements of some popular oxides at room temperature
1.02.03.04.05.06.07.08.09.0
300 400 500 600 700 800
Fluo
resc
ence
Inte
nsity
/ a.
u.
Em Wavelength / nm
MgO CaO
ZnO
TiO2
300 400 500 600 700 800
Wavelength / nm
0.0
2.03.0
4.05.0
6.0
7.0
8.09.0
Fluo
resc
ence
Inte
nsity
/ a
.u
CuInSe2 measured with PMT-1700 and helium cryostat
6K
26K
46K
66K
86K106K
1100 1500140013001200 16000.0
1.0
2.0
3.0
Wavelength / nm
Cou
nts/
105
GaN thin film lifetime measurements aquired with liquid nitrogen cryostat in MCS mode. The graph shows the fitted curves after advanced lifetime analysis.
110
210
310
410
Cou
nts
0100 200 400 600 800 1000 1200 1400 1600
Time / µs0 200 400 600 800 1000 1200 1400 1600
Time / µs
100
101
102
103
Cou
nts
104
1000 1500 2000 2500
600 800 1000 1200 1400
100 150 200 250
0 50 100 150 200 250Time/µs
MgO
TiO2
CaO
ZnO
Time / µs
Semiconductor Band-Gap
0
10
20
30
40
50
50 150 250 350
Aver
age
Live
time
/ μs
Temperature / K
EMISSION OF SEMICONDUCTOR MATERIALS AT ROOM TEMPERATURE
A large variety of inorganic materials show semiconductor properties. The emission from these materials at ambient temperatures can be strong or extremely weak, depending on the physics of inter-band recombination, the presence of defects and impurities.
Spectral and time-resolved photoluminescence properties of these materials reveal fundamental information about their structure and can also be used for quality control and process monitoring.
Often these materials come as white powders and a clear separation of the potentially weak emission from unwanted scatter may be challenging. Double monochromators or lifetime measurements are effective in discriminating from scattering artefacts.
EMISSION OF SEMICONDUCTOR MATERIALS AT CRYOGENIC TEMPERATURES
Many semiconductor materials show very weak emission at room temperature. Different cryostat options are offered to enable measurements down to <4 K. Liquid nitrogen, liquid helium or closed cycle helium cryostats can be directly integrated into the FLS1000 sample chamber, without additional fibre optics. Many of the cryostat options are fully controlled by the Fluoracle
software, enabling automated spectral and lifetime temperature maps to be performed.
The CuInSe2 graph shows a spectral temperature map as measured with a helium cryostat.
The GaN graph shows the results of a lifetime temperature map measured with a nitrogen cryostat, with all decay measurements analysed using the stretched exponential function of the advanced lifetime analysis package.
AP
PLI
CA
TIO
NS
IN M
ATE
RIA
L SC
IEN
CE
S
6
Various luminescent phosphor powders measured with integrating sphere.
YAG:Ce phosphor powder in BaSO4 at various concentrations.
Raw Data
0.2
0.4
0.6
0.8
1.0
1.2
1.4
Cou
nts
/ 10
6
300 350 400 450 500 550 600
Wavelength / nm
w/w%20
50
60
70
85
100
x 10
350 400 450 500 550 600 650 700 750 800
Wavelength / nm
4.0
3.5
3.0
2.5
2.0
1.5
1.0
0.5
Cou
nts
/ 10
4
1.02.03.04.05.06.07.08.09.0
300 400 500 600 700 800
Coun
ts/1
0-1
Wavelength/nmWavelength / nm
300 500 400 600 700 800
0.2
0.4
0.6
0.8
Inte
nsity
(a.u
.)
0.00.2 0.3 0.4 0.5 0.6 0.7
x
y
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.00.0 0.1
Lanthanides
104
100
95
90
85
800 20
Concentration (w/w)
Qua
ntum
Yie
ld (%
)
40 60 80 100 120
0.2
0.4
0.6
0.8
1.0
Reflectance
Nor
mal
ised
Refl
ecta
nce
300 350 400 450 500 550 600
Wavelength / nm
Reflectance
Absorbance
0.2
0.6
0.8
1.0
Nor
mal
ised
Ab
sorb
ance
0.4
300 350 400 450 500 550 600
Wavelength / nm
LIGHTING INDUSTRY PHOSPHORS
Light emitting powders can have strong emission; therefore, accurate determination of their reflection and absorption properties can be challenging. Using the integrating sphere and specially configured synchronous scans, spectral reflectance and absorbance can be reliably obtained.
Measurements of absolute photoluminescence quantum yield require high precision in the measurement of both weak and strong signals, clear and established measurements procedures, and accurate correction factors for the whole light path including the integrating sphere, all provided by the FLS1000.
The high dynamic range of our cooled detectors, our easy-to-operate measurement wizard, and the FLS1000 sphere as an integrated part of the sample chamber, will enable you to perform reliable measurements of absolute photoluminescence quantum yields.
The Fluoracle software’s chromaticity wizard provides colour coordinates and luminous intensity values. Fluoracle can display CIE 1931 and CIE 1976 colour coordinates. You can set reference values and compare a large number of spectral scans in those coordinates.
7
2500 3500 3000 4000 4500
0
200
400
600
800
1000
Tim
e /
µs
0.5
1.0
1.5
2.0
2.5
600 700 800 900 1000 1100 1200 1300 1400 1500 1600
Fluo
resc
ence
Inte
nsity
/ a.
u.
Wavelength / nm 600 800 1000 1200 1400 1600
Wavelength / nm
Inte
nsity
/ a
.u.
0.5
1.0
1.5
2.0
2.5 Photoluminescence
Electroluminescence
120 mA
80 mA
40 mA
Photoluminescence and electroluminescence for three different currents of a P3HT:PCBM thin film device, λexc = 550 nm for photoluminescence.
Co2+
doped ZnSe crystal, measured with Nd:YAG pumped OPO and
InSb detector, λexc= 808 nm.
Cr2+
Fe2+
doped ZnSe crystal, measured with Nd:YAG laser and InSb detector, λexc= 532 nm.
900 1000 1100 1200 1300 1400 1500 1600Wavelength / nm
550
600
650
700
750
800
Fix
ed W
avel
eng
th /
nm
1.0
2.0
3.0
4.0
5.0
Co
un
ts/1
02
900 1000 1100 1200 1300 1400 1500 1600Wavelength / nm
550
600
650
700
750
800
Fix
ed W
avel
eng
th /
nm
1.0
2.0
3.0
4.0
5.0
Co
un
ts/1
02
900 1000 1100 1200 1300 1400 1500 1600Wavelength / nm
550
600
650
700
750
800
Fix
ed W
avel
eng
th /
nm
1.0
2.0
3.0
4.0
5.0
Co
un
ts/1
02
900 1000 1100 1200 1300 1400 1500 1600Wavelength / nm
550
600
650
700
750
800
Fix
ed W
avel
eng
th /
nm
1.0
2.0
3.0
4.0
5.0
Co
un
ts/1
02
Wavelength / nm
550
600
650
700
750
800
Fixe
d W
avel
eng
th /
nm
1.0
2.0
3.0
4.0
5.0
Cou
nts
/ 10
2
900 1000 1100 1200 1300 1400 1500 1600
2500 3000 4000 5000
10
20
30
40
50
Tim
e /
µs
Single-walled carbon nanotubes detected with NIR-PMT
Nanotubes
Photovoltaic Materials
Laser Materials
Nanoscale materials such as quantum dots, carbon nanotubes, nanoparticles and nanostructures exhibit strong spatial confinement upon photo-excitation and tuneable emission. This makes them excellent candidates in a multitude of applications from photonic devices to sensing and biology. The structural characteristics of single-walled and multi-walled carbon nanotubes can be identified from excitation-emission mapping, as shown in the graph.
The mapping function in the Fluoracle software, together with the instrument's unsurpassed resolution and NIR spectral coverage, make the FLS1000 an excellent tool to study these materials.
CARBON NANOTUBES
POLYMER SOLAR CELL MATERIALS
As well as their use in organic light emitting devices (OLEDs), organic polymer materials are also potential candidates as easy to operate and inexpensive photovoltaic solar cell materials.
The FLS1000 features an optional electroluminescence sample holder that can be easily exchanged with the standard photoluminescence holder. Both photoluminescence and electroluminescence can be used as diagnostic tools for the characterisation of the electronic band-gap and the charge-transfer states in these organic photovoltaic cells.
MID-IR LASER MATERIALS
The emission of novel laser materials up to 5.5 µm can be measured using InSb detectors. To discriminate effectively from black body radiation time-resolved
measurements are advantageous, although phase sensitive (lock-in) technology is also available.
AP
PLI
CA
TIO
NS
IN P
HO
TOC
HE
MIS
TRY
AN
D L
IFE
SC
IEN
CE
S
Integrating sphere accessory for the FLS1000.
8
Quinine bisulphate in perchloric acid, measured using the integrating sphere.
Lifetime measurements of pyrene monomers (λem = 395 nm) and excimers (λem = 465 nm), excited at 335 nm using the nF920 nanosecond flashlamp.
Pyrene in cyclohexane at different sample concentrations, monomer and excimer spectra excited at 335 nm.
Cou
nts
Time (ns)500
102
103
1
10
0
-2.9
2.9
100 150χ2 = 1.0302-exponential
Resi
dua
ls
104
Monomer
Time (ns)500
102
103
1
10
0
-4.7
4.7
100 150χ2 = 1.0592-exponential
104
Excimer
600500450400350 5500.0
0.6
1.2
0.8
1.0
Wavelength / nm
Cou
nts
/ 1
04
650
0.4
0.2
Monomer Excimer
Natural Dyes
300 5004504003500.0
0.2
0.8
0.4
0.6
Wavelength / nm
Cou
nts
x104
1.0
x 5
550
Excitation wavelength (nm)
Qua
ntum
Yie
ld
280 300 320 340
80
60
40
20
0
The absolute method for fluorescence quantum yield (QY) measurements is now more widely used than the relative method. It does not require a QY standard and readily applicable to liquids, films and powders and can be extended into the near-infrared spectral range.
The figure shows the independence of the fluorescence QY from the wavelength of excitation for a standard organic dye. The areas of absorption for eight different excitation wavelengths and the corresponding emission spectra, scaled by a factor of 5, are shown. The inset shows the calculated quantum yields.
ABSOLUTE QUANTUM YIELD MEASUREMENTS OF ORGANIC DYES
MONOMER-EXCIMER EQUILIBRIUM AND KINETICS
The monomer or excimer states of organic molecules can be easily distinguished by the balance of their fluorescence. Care should be taken to the inner filter effect due to their high extinction coefficient. The emission spectrum of pyrene in cyclohexane is shown as a function of pyrene concentration. At the lowest concentrations only the monomer bands in the wavelength range 370 nm – 400 nm are observed, while at the highest concentrations the excimer peak at ~480 nm dominates.
While many fluorescence lifetime measurements can be described by single or multiple exponential decays, often formation or growth kinetics can be observed too. The example below shows monomer-excimer kinetics of pyrene, with the monomer showing a double exponential decay, while the excimer displays an exponential rise, followed by an exponential decay. Rise and decay times are often linked. Global analysis of the lifetime data with linked parameters is possible using the advanced data analysis package. The global fitting, results in a double exponential with τmon = 9.3 ns and τexcimer = 15.4 ns.
9
3-amino-N-methylphthalimide in glycerol, measured with EPL-375 and TCSPC. inset: spectral shift of the calculated emission spectrum.
0.0
1.0
2.0
3.0
4.0
5.0
300 350 400 450 500 550 600
Inte
nsity
/ a
.u.
Emission Wavelength / nmEmission Wavelength / nm
300 400 350 450 500 550 0.0
1.0
2.0
3.0
4.0
5.0
Inte
nsity
/ a
.u.
600
Absorption Emission
010
110
210
310
150 200 250
Cou
nts
0 50 100Time / ns
100
101
102
103
Cou
nts
0 50 100 150 200 250
Time / ns
Fluorescence decays at 520 nm, 540 nm and 560 nm
0
0.1
0.2
0.3
0.4
0 20 40 60 80 100
Lifeti
me Am
plitu
de
Lifetime / ns
0 20 40 60 80 100
Lifetime / ns
0
0.1
0.2
0.3
0.4
Life
time
Am
plit
ude Fluorescence lifetime distributions
at 520 nm, 540 nm and 560 nm
CdSe quantum rods spectra after TRES data slicing (left). Lifetime measurements and results of distribution analysis (right). Fluorescence decay measurements were excited with EPL-405.
Synthetic Dyes
Quantum Rods
0.0
0.8
0.6
0.4
0.2
Time / ns
200 40 60
Cou
nts
/ 10
3
Time / ns200 10
0
10
20
Spec
tral
Shi
ft /
nm
SOLVENT RELAXATION DYNAMICS OF ORGANIC DYES
Fluorescence kinetics can be complex for homogeneous dye solutions in polar solvents, even if rotational diffusion is eliminated with magic angle conditions.
The measurement shows 16 fluorescence decays in a polar-viscous solvent of glycerol containing one fluorophore species, with multi-exponential decays at shorter wavelengths and rise-decay kinetics at longer wavelengths.
The TRES mapping option in Fluoracle enables quick data acquisition, while the FAST package allows performing complex analysis. Using globally linked 4-exponential decay analysis, decay associated spectra (DAS) can be generated and the peak position of the spectra be determined. The shift in the peak position is shown in the insert graph and clearly shows a 2-exponential reorientation dynamic of the fluorophore.
QUANTUM DOTS AND QUANTUM RODS
Quantum dots are nanometer sized semiconductor material structures that exhibit strong and specific fluorescence emission. The emission wavelength can be tuned by altering the size or the structure. These materials have widespread applications in modern displays, light emitting diodes, diode lasers, solar cell materials, as fluorescent inks and as fluorescent labels in biochemistry. Coated with biofunctional groups, quantum dots are being used as fluorescent labels in biomedical imaging and drug screening.
Compared to their spherical counterparts, quantum rods display less overlap between absorption and emission band. This reduces reabsorption and self-quenching and has therefore benefits over spherical structures.
Switching between spectral and lifetime measurements is easy in the FLS1000, enabling the user to study all aspects of their sample in a short period of time. The graphs show spectral data and TRES lifetime measurements obtained for quantum rods. Three lifetime measurements were analysed with lifetime distribution analysis to correlate distributions of lifetimes to distributions of particle size and shape parameters.
AP
PLI
CA
TIO
NS
IN P
HO
TOC
HE
MIS
TRY
AN
D L
IFE
SC
IEN
CE
S
Fluorescence Based Assays
10
Europium cryptate lifetime measurements in an enhanced sensitivity immunoassay, measured on a 96-well plate within the plate reader. Excitation by μF2 and emission detected by gated photomultiplier (PMT-900GT).
Singlet oxygen generated by chlorine e6 in water/ethanol, pure water, and instrumental response function. Samples were excited with DPSS laser at λex = 355 nm, detection by PMT-1700.
Singlet oxygen generated by [Ru(bpy)3]Cl2 in oxygen saturated D2O, excited with μF2 and detected with PMT-1700. Inset: Background removed 1O2 emission after TRES data slicing.
Cou
nts
Time / µs1000
103
0-4.3
4.3
200 300
Resi
dua
ls
Cou
nts
Time / µs
105
106
104
100 20 305 15 25
1O2 in water/ethanol
1O2 in water
Cou
nts
Time / µs
10000102
103
2000 3000
104
25001500500
105
Photo-sensitisers
Plate reader accessory for the FLS1000.
Cou
nts
/ 10
3
Emission Wavelength / nm12401160
0.6
0.8
0.4
1280 1320
1.0
0
0.2
1200
FRET
Act
ivity
[AFP] / ng/ml10.0
0.4
0.6
0.7
100
0.5
0.1 1.0
0.3
0.2
EMISSION OF SINGLET OXYGEN
Time-resolved singlet oxygen measurements are challenging as the emission at ~1270 nm is very weak. The emission lifetime is solvent dependent and becomes comparatively short in aqueous solutions.
The figures below shows that time-resolved singlet oxygen measurements are possible with the standard microsecond flashlamp as excitation source.
By measuring time-resolved spectra the singlet oxygen emission can be discriminated from the background.
The figure below shows a time-resolved singlet oxygen emission measurement with exceptionally high temporal resolution. An instrument configured with a nanosecond laser source and MCS detection is the best choice for this type of measurement. A wide variety of ns pulsed lasers can be integrated with the FLS1000.
FRET-BASED IMMUNOASSAYS
The example shows the change in the fluorescence decay kinetics of the donor europium cryptate, quenched by a nanosecond emitting acceptor. The fraction of FRET quenched donors is proportional to the concentration of the antigen, in this case Human Alfa Fetoprotein.
The high sensitivity of this time-resolved FRET measurement in this homogeneous immunoassay allows for antigen concentrations that are significantly lower than the cancer marker threshold. Fast measurements on multiple samples are possible thanks to the plate reader accessory for the FLS1000.
Time-resolved measurements of FRET in the millisecond time scale can provide more accurate data than other FRET techniques as background emission from the assay occurring in nanoseconds can be removed by detector gating. Other residual background in the microsecond time scale is eliminated during data analysis.
Fluorescence Based Assays
Food Samples
11
Milk sample, diluted in water, Δλexc = Δλem = 3 nm.
Excitation spectrum of tryptophan and fluorescence emission spectral shift functions of a crystalline eye lens protein dissolved in phosphate buffer saline (folded protein - red) and 7M urea (denatured protein - blue).
Calibration curve for the OctSG binding assay, using Dansyl as fluorescent label, λexc= 340 nm, λem= 550 nm.
0
0.2
0.4
0.6
0.8
Inte
snity
/ a
.u.
240 260 280 300 320
350
340
330
320
Emis
sion
pea
k w
avel
eng
th /
nm
Wavelength / nm
3101.0
0
0.02
0.04
0.06
0.08
0.12
0.14
0.16
0.18
0 500 1000 1500 2000 2500 3000 3500
Concentration OctSG / μM
Ani
sotr
opy
OctSG (tested ligand)
bound Dansyl-GSH free Dansyl-GSH
0.10
0.20
0.0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
350 400 450 500 550 600 650 700 750 800300
320
340
360
380
400
420
440
460
480
500
Emission Wavelength / nm
Cou
nts
/ 10
5
Exci
tatio
n W
avel
eng
th /
nm
Biomolecules
Em Wavelength / nm320 330 340320 330 340
Em Wavelength / nm
Em Wavelength / nm340 350 360340 350 360
Em Wavelength / nm
RED-EDGE EXCITATION OF PROTEINS
Red-edge excitation fluorescence spectroscopy is a powerful method to study structure and dynamics of biological membranes, micelles and proteins. In protein research, this method takes advantage of the possibility to selectively excite a small fraction of specifically located tryptophan species at the red edge of the tryptophan absorption. The shift in the positions of the emission maximum as a function of excitation wavelength yields information about polarity and dynamics of tryptophan microenvironment and thus can reveal valuable information about the status of the protein under investigation.
The FLS1000 with its outstanding sensitivity and its automated measurement functionality is an excellent tool for red-edge excitation fluorescence spectroscopy; even for very low concentrations of the protein under investigation.
ANISOTROPY BASED LIGAND-BINDING ASSAYS
A fluorophore bound to a ligand shows higher rotational mobility when free in solution, compared to when the ligand-fluorophore pair is bound to larger proteins. This change in rotational mobility can be monitored by fluorescence anisotropy measurements.
Some ligand binding assays use this anisotropy method. The example competition assay shown here, probes the concentration of the unknown ligand OctSG, using a potassium efflux system protein as a binding side and the fluorophore Dansyl as the fluorescent probe.
The FLS1000 can be configured in T-geometry for simultaneous data acquisition from two emission channels. In this experiment, the fluorescence anisotropy was measured using a T-geometry system to simultaneously record the vertical and horizontal component of the fluorescence emission.
FOOD SAMPLE MAPPING
Excitation-Emission Maps (EEMs) provide a “fingerprint” of complex fluorescent samples, containing many components. This fingerprinting technique is used in the food and drug industry to monitor content and identify potential toxic contaminants.
The map to the right shows an EEM of a sample of diluted milk. Although milk is a highly scattering sample, the double monochromators in the FLS1000 provide reliable EEMs with the absence of scattering effects.
Edinburgh Instruments Ltd. 2 Bain Square, Kirkton Campus, Livingston, EH54 7DQ Registered in England and Wales No: 962331 VAT No:GB 271 7379 37©Edinburgh Instruments Ltd 2020
Pri
nted
Dec
emb
er -
202
0
Edinburgh Instruments
2 Bain Square, Kirkton Campus, Livingston, EH54 7DQUnited Kingdom.
For more information, contact:
T: +44 (0)1506 425 300
F: +44 (0)1506 425 320
W: www.edinst.com
T: +1-800-323-6115
F: +44 (0)1506 425 320
W: www.edinst.com
For our US office, contact:
WORLDWIDE CUSTOMER SUPPORT
Edinburgh Instruments has been at the forefront of
research, development and manufacture of state-of-the-art
photoluminescence instrumentation for over 50 years.
During this time a worldwide reputation for quality
and innovation has been established. Edinburgh
Instruments continue to push product development
of affordable, reliable and high quality instrumentation
that can be tailored and upgraded to evolve with the
science and technology.