Application Guide

PA 800 plus Pharmaceutical Analysis System

Carbohydrate Labeling and Analysis

January 2013

Beckman Coulter, Inc.250 S. Kraemer Blvd.Brea, CA 92821 U.S.A.

B25802AA

Page 2: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and AnalysisPA 800 plus Pharmaceutical Analysis System

PN B25802AA (January 2013)

© 2012 Beckman Coulter, Inc.All rights reserved.

Beckman Coulter, Inc. grants a limited non-exclusive license to the owner or operator of a PA 800 plus instrument to make a copy, solely for laboratory use, of any portion or all of the online help and electronic documents shipped with the PA 800 plus instrument.

Beckman Coulter and the stylized logo are trademarks of Beckman Coulter, Inc., are registered in the USPTO.

All other trademarks are the property of their respective owners.

Find us on the World Wide Web at: www.beckmancoulter.com

Made in USAPrinted in USA

Page 3: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Revision History

Initial Issue, B25802AA, January 2013PA 800 plus Software Version 10.1 (includes 32 Karat software)

PA 800 plus Firmware version 10.1 (incompatible with 9.x)

B25802AA iii

Page 4: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Revision History

B25802AAiv

Page 5: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Safety Notices

Symbols and Labels

Introduction

The following is a description of symbols and labels used on the Beckman Coulter PA 800 plus Pharmaceutical Analysis System or shown in this manual.

WARNING

If the equipment is used in a manner not specified by Beckman Coulter, Inc., the protection provided by the instrument may be impaired.

General Biohazard Symbol

This caution symbol indicates a possible biohazard risk from patient specimen contamination.

Caution, Biohazard Label

This caution symbol indicates a caution to operate only with all covers in position to decrease risk of personal injury or biohazard.

B25802AA v

Page 6: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Safety NoticesSymbols and Labels

Caution, Moving Parts Label

This caution symbol warns the user of moving parts that can pinch or crush.

High Voltage Electric Shock Risk Symbol

This symbol indicates that there is high voltage and there is a risk of electric shock when the user works in this area.

Class 1 Laser Caution Label

A label reading “THIS PRODUCT CONFORMS TO APPLICABLE REQUIREMENTS OF 21 CFR 1040 AT THE DATE OF MANUFACTURE” is found near the Name Rating tag. The laser light beam is not visible.

Sharp Object Label

A label reading “CAUTION SHARP OBJECTS” is found on the PA 800 plus.

CAUTIONPARTS MOVEAUTOMATICALLY

A015081L.EPS

144557-AB

A016352L.EPS

A016350L.EPS

CLASS 1 LASER PRODUCTTHIS PRODUCT CONFORMS TO APPLICABLE REQUIREMENTS OF 21 CFR 1040 AT THE DATE OF MANUFACTURE.

MANUFACTURED:

726024-C

CAUTIONSHARP OBJECTS

A16558-AA

A016351L.EPS

B25802AAvi

Page 7: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Safety NoticesSymbols and Labels

Recycling Label

This symbol is required in accordance with the Waste Electrical and Electronic Equipment (WEEE) Directive of the European Union. The presence of this marking on the product indicates:

1. The device was put on the European Market after August 13, 2005.

2. The device is not to be disposed of via the municipal waste collection system of any member state of the European Union.

It is very important that customers understand and follow all laws regarding the proper decontamination and safe disposal of electrical equipment. For Beckman Coulter products bearing this label, please contact your dealer or local Beckman Coulter office for details on the take back program that facilitates the proper collection, treatment, recovery, recycling, and safe disposal of this device.

Disposal of Devices Containing Mercury Components

This product contains a mercury-added part. Recycle or dispose of according to local, state, or federal laws. It is very important that you understand and comply with the safe and proper disposal of devices containing mercury components (switch, lamp, battery, relay, or electrode). The mercury component indicator label can vary depending on the type of device.

B25802AA vii

Page 8: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Safety NoticesSymbols and Labels

Restriction of Hazardous Substances (RoHS) Labels

These labels and materials declaration table (the Table of Hazardous Substance's Name and Concentration) are to meet People's Republic of China Electronic Industry Standard SJ/T11364-2006 "Marking for Control of Pollution Caused by Electronic Information Products" requirements.

RoHS Caution Label

This logo indicates that this electronic information product contains certain toxic or hazardous elements, and can be used safely during its environmental protection use period. The number in the middle of the logo indicates the environmental protection use period for the product. The outer circle indicates that the product can be recycled. The logo also signifies that the product should be recycled immediately after its environmental protection use period has expired. The date on the label indicates the date of manufacture.

RoHS Environmental Label

This logo indicates that the product does not contain any toxic or hazardous substances or elements. The "e" stands for electrical, electronic, and environmental electronic information products. This logo indicates that this electronic information product does not contain any toxic or hazardous substances or elements, and is green and is environmental. The outer circle indicates that the product can be recycled. The logo also signifies that the product can be recycled after being discarded, and should not be casually discarded.

B25802AAviii

Page 9: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Safety NoticesSymbols and Labels

Alerts for Warning, Caution, Important, and Note

WARNING

WARNING indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. The warning can be used to indicate the possibility of erroneous data that could result in an incorrect diagnosis (does not apply to all products).

CAUTION

CAUTION indicates a potentially hazardous situation, which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. The caution can be used to indicate the possibility of erroneous data that could result in an incorrect diagnosis (does not apply to all products).

IMPORTANT IMPORTANT is used for comments that add value to the step or procedure being performed. Following the advice in the IMPORTANT notice adds benefit to the performance of a piece of equipment or to a process.

NOTE NOTE is used to call attention to notable information that should be followed during installation, use, or servicing of this equipment.

B25802AA ix

Page 10: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Safety NoticesSymbols and Labels

B25802AAx

Page 11: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Contents

Revision History, iii

Safety Notices, v

Symbols and Labels, vIntroduction, vGeneral Biohazard Symbol, vCaution, Biohazard Label, vCaution, Moving Parts Label, viHigh Voltage Electric Shock Risk Symbol, viClass 1 Laser Caution Label, viSharp Object Label, viRecycling Label, viiDisposal of Devices Containing Mercury Components, viiRestriction of Hazardous Substances (RoHS) Labels, viii

RoHS Caution Label, viiiRoHS Environmental Label, viii

Alerts for Warning, Caution, Important, and Note, ix

Carbohydrate Labeling and Analysis Kit, 1

Overview, 1Safety Information, 1

General Precautions, 1Intended Use, 2

Materials and Reagents, 2Contents of this Kit, 2Materials Needed but Not Provided in Kit, 2Storage Conditions, 3

Overview of the Procedure, 3

Releasing Oligosaccharides From Glycoproteins, 4Enzymatic Release of the N-Linked Oligosaccharides, 4Working with the Supernatant, 4Principle of the Labeling Method, 5

Preparing the Labeling Reagents, 5Choosing the Proper Fluorophore, 5Preparing the Reagents, 5

xi

Page 12: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Contents

Preparing the Standards, 6

Performing the Labeling Reaction, 6

Preparing the PA 800 plus Instrument, 9

Installing and Pre-conditioning the N-CHO Capillary, 10Clean the Interface Block, 11Insert the Cartridge, 11

Preparing the Buffer Trays, 11

Sample Vial Setup, 12

Running the Samples, 13Initial Conditions, 13LIF Detector Initial Conditions, 14Time Program, 15

Evaluation of Results, 17

System Shutdown and Capillary Storage, 18Short Term Storage (<24 hours) of Capillary, 18Long Term Storage (>24 hours) of Capillary, 18

Troubleshooting Procedures, 19

Appendix, 20LIF Detector Calibration Procedure, 20

Setting the Calibration Corrector Factors (CCFs), 20Automatic Calibration, 21

CCF Troubleshooting Procedures, 24CCF <1, 24CCF >1, 24No Step Change Detected, 25

xii

Page 13: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

xiii

Illustrations

Illustrations

1 Labeling Reaction of an Oligosaccharide with APTS, 5

2 Labeling Reaction Scheme, 8

3 Sample Preparation for Injection, 9

4 Capillary Performance Screen, 10

5 Universal Vials and Caps, 11

6 Buffer Tray Configuration, 12

7 Micro Vial Inside a Universal Vial, 13

8 Initial Condition Tab, 14

9 LIF Detector Initial Conditions Tab, 15

10 Time Program Tab, 16

11 Typical Current Profile, 17

12 Electropherogram - Glucose Ladder, 18

13 32 Karat Software Configuration Dialog Box, 21

14 Calibration Wizard Screen - Auto Select, 22

15 Calibration Wizard Screen - Step 2, 22

16 Calibration Wizard Screen - Step 3, 23

17 Calibration Wizard Screen - Step 4, 24

Page 14: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

xiv

Tables

Tables

1 Kit Contents Reorder Number (477600), 2

2 Materials Not Provided, 2

3 Test Run: Initial Conditions, 13

4 Test Run: LIF Detector Initial Conditions, 14

5 Carbohydrate Labeling Troubleshooting Chart, 19

6 Reagents Needed to Perform the Procedure, 20

7 CCF <1, 24

8 CCF >1, 24

Page 15: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis Kit

Overview

Beckman Coulter’s Carbohydrate Labeling & Analysis Kit uses capillary electrophoresis to separate and quantify oligosaccharides released from glycoproteins. This kit has been developed to work with the PA 800 plus Pharmaceutical Analysis System, using the 488 nm laser and the 520 nm emission filter.

This kit contains the reagents, buffer, and coated capillaries required to label, separate, and quantify oligosaccharides. This kit also provides a glucose size marker for relative size determination and a maltose standard for quantitation and mobility characterization of the released oligosaccharides.

For further information, please refer to the following publication:

Quantitative Analysis of Sugar Constituents of Glycoproteins by Capillary Electrophoresis, Fu Tai, A. Chen, Thomas S. Dobashi and Ramon A. Evangelista, “Glycobiology”, Vol. 8, No. 11, pp. 1045-1052, 1998.

Safety Information

NOTE This application guide has been validated for use in the PA 800 Enhanced Protein Characterization and the PA 800 plus Pharmaceutical Analysis Systems.

IMPORTANT In this procedure 1 M sodium cyanoborohydride in THF is used, which is a very reactive compound. Please consult the MSDS before using or disposing of this reagent.

CAUTION

When installing or removing capillaries from the cartridge, use safety goggles.

General Precautions

• Prior to using the instrument, turn on the laser and let it stabilize for 15 minutes. Turn it off when not in use for longer periods of time.

• Do not interchange the capillary (i.e. if a capillary is being used for carbohydrate analysis, do not use the same capillary in another application).

B25802AA 1

Page 16: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitMaterials and Reagents

• When the capillary is not in use or after finishing with all experiments, it is recommended to run a shutdown method to rinse the capillary and leave the ends immersed in water to avoid capillary blockage and to keep the capillary from drying out.

• Prior to using the capillary for the first time, it is recommended to rinse the capillary for 10 minutes at 30 psi pressure with DDI water and then rinse for 10 minutes at 30 psi with the Carbohydrate Separation Gel Buffer.

Intended Use

The Carbohydrate Labeling and Analysis Kit is for laboratory use only.

CAUTION

Refer to the Material Safety Data Sheets (MSDS) information, available at www.BeckmanCoulter.com, regarding the proper handling of materials and reagents. Always follow standard laboratory safety guidelines.

Materials and Reagents

Contents of this Kit

Materials Needed but Not Provided in Kit

Table 1 Kit Contents Reorder Number (477600)

Component Quantity Reorder # Code

Carbohydrate Separation Buffer 56 mL 477623

N-CHO Coated Capillary 2 477601

Labeling Dye (APTS) 4x5 mg 501309 L6

Labeling Dye Solvent 1 mL L3

Glucose Ladder Standard 50 mg G20

Quantitation/Mobility Marker (Maltose) 0.18 mg G22

APTS-M (monosaccharide grade) 20 mg 725898

Table 2 Materials Not Provided

Description Part Number

LIF Performance Test Mix 726022

Universal Plastic Vials (pack 100) A62251

Universal Rubber Vial Caps - blue (pack 100) A62250

Micro Vials (pack 50) 144709

B25802AA2

Page 17: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitOverview of the Procedure

Storage Conditions

The kit components are stable for one year when stored under the following conditions:

• Upon receipt, the kit should be stored at 2 to 8°C.

• Reconstituted Quantitation Control (G22) and labeled glucose ladder standard (G20) should be stored at -35 to -15°C.

• Reconstituted labeling reagent (APTS) shall be stored at -35 to -15°C. It is stable for 2 weeks after reconstitution.

Overview of the Procedure

The carbohydrate analysis involves the following three tasks:

• Enzymatic release of the N-linked oligosaccharides from the glycoprotein.

• Labeling of the mixture of released oligosaccharides.

• Analysis of the flouorophore-labeled oligosaccharides by capillary electrophoresis.

Assorted pipettors and pipette tips

Spatula

Analytical scale

Vortexer

Bench top micro-centrifuge

Water bath

Centrifugal vacuum evaporator

Sodium Cyanoborohydride, 1 M/THF Sigma-Aldrich 296813

2-mercaptoethanol Sigma-Aldrich M7154

Double Deionized (DDI) water filtered through 0.2 μm filter

N-glycosidase F (PNGase F) New England Biolabs P0704S

Nonidet NP-40 non ionic detergent

Table 2 Materials Not Provided

Description Part Number

B25802AA 3

Page 18: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitReleasing Oligosaccharides From Glycoproteins

Releasing Oligosaccharides From Glycoproteins

IMPORTANT This labeling kit does not contain releasing enzymes.

There are multiple enzymatic and chemical procedures for releasing oligosaccharides from proteins. In order to successfully label the released oligosaccharide, it is essential that the reducing termini of the oligosaccharide is not destroyed by the deglycosylation method.

The following is a suggested protocol for N-deglycosylation, using N-Glycosidase F (PNGase F):

Enzymatic Release of the N-Linked Oligosaccharides

1. Aliquot the sample volume in order to have between 25-300 μg of glycoprotein.

2. Dry the glycoprotein solution completely in a centrifugal vacuum evaporator.

3. Add 45 μL of 1X PBS buffer.

4. Add 1.0 μL of 5% SDS (or a volume that gives final concentration of 0.1% SDS).

5. Add 1.5 μL of 1:10 dilution of 2-mercaptoethanol in DDI water (or a volume that gives final concentration of 50 mM 2-mercaptoethanol).

6. Heat the sample for 5 minutes at 100°C.

IMPORTANT If the denatured protein precipitates, discard the sample and restart this process, repeating steps 1 through 5. For step 6, denature the protein by incubating it at 37°C for 10 minutes.

7. Cool the sample to room temperature.

8. Add 5 μL of Nonidet NP40 non-ionic detergent (or a sufficient volume that gives a final concentration of 0.75%).

9. Add PNGase F enzyme, according to its activity.

10. React for 15 hours at 37°C.

11. Add approximately 150 μL of cold ethanol (or 3 times the actual reaction mixture volume).

12. Vortex the mixture and place it on ice for complete protein precipitation.

13. Centrifuge the samples at 15,000 rpm for 5 minutes.

14. Withdraw and save the supernatant.

IMPORTANT This solution contains the RELEASED N-LINKED OLIGOSACCHARIDES.

15. Discard the solid. This is the deglycosylated protein.

Working with the Supernatant

There are two options for working with the supernatant:

For quantitative analysis, add an internal standard to the supernatant. See Preparing the Standards beginning with Step 1a.

B25802AA4

Page 19: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitPreparing the Labeling Reagents

For qualitative analysis, bring the supernatant to dryness in the centrifugal vacuum evaporator. These oligosaccharides are ready to be labeled with APTS. See Performing the Labeling Reaction in this chapter.

Principle of the Labeling Method

After release (enzymatic or chemical), the oligosaccharides can be labeled with a fluorophore called 8-aminopyrene-1,3,6, trisulfonate. The stoichiometry of labeling reaction is one APTS molecule per molecule of oligosaccharide. Figure 1 illustrates the labeling reaction of an N-linked oligosaccharide with APTS.

Figure 1 Labeling Reaction of an Oligosaccharide with APTS

The efficiency of the labeling reaction is dependent on temperature and the amount of oligosaccharides. This protocol has been optimized for labeling 5 nmoles or less of total oligosaccharides.

Samples with amounts higher than 5 nmoles may give a lower reaction yield. Use maltose (G22) as an internal labeling control and/or as an internal mobility marker.

Preparing the Labeling Reagents

Choosing the Proper Fluorophore

APTS-M is a high-purity fluorophore intended for labeling monosaccharides. APTS-M contains citric acid as a catalyst. APTS (L6) can be used for oligosaccharides.

Preparing the Reagents

1 Prepare the APTS labeling reagent.

To a vial of APTS labeling dye (L6)

a. Add 48 μL of Labeling dye solvent (15% acetic acid – L3).

b. Vortex for 5 seconds until complete dissolution.

c. Store at -35 to -15°C for up to 2 weeks when not in use.

B25802AA 5

Page 20: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitPerforming the Labeling Reaction

2 Prepare the APTS-M (Monosaccharide grade) Labeling Dye

a. Add 400 μL of DDI water to the APTS-M vial.

b. Vortex the mixture for 5 seconds until all of the solid is dissolved.

c. Store the reconstituted APTS-M at -35 to -15°C for up to two weeks.

Preparing the Standards

1 Prepare the Maltose Quantitative Control/Mobility Marker (G22).

a. Add 500 μL of DDI water to the Maltose quantitative control labeled G22. This makes a solution containing 500 nmoles of maltose, or concentration of 1 nmol/μL.

b. For quantitation purposes, add 5 μL of G22 reconstituted to your released oligosaccharides.

c. Dry down the sample in a centrifugal evaporator and proceed to labeling protocol, as described in the next section.

d. Store at -35 to -15°C when not in use.

NOTE When using the maltose solution for quantitation purposes, make sure that the solution is prepared fresh and used right away. Bacterial contamination can digest the sugars, leading to a miscalculation of your quantitation.

2 Prepare the Glucose Ladder Standard (G20).

a. Weigh and dissolve 5 mg of Glucose Ladder Standard (G20) in 80 μL DDI water in 1.5 mL microcentrifuge tube. Sonicate if necessary.

b. Aliquote at least ten 2 μL portions of the Glucose ladder standard solution to 0.5 mL microcentrifuge vials. Add 5 μL of G22 reconstituted, prepared in step 1a to the 2 μL portions of the Glucose ladder standard solution, and dry them in a centrifugal vacuum evaporator. The dried glucose ladder with Maltose (G22) can be stored at room temperature or used immediately.

Performing the Labeling Reaction

1 Label the Sample and/or Standard (G20).

a. Add 2 μL of 1 M sodium cyanoborohydride/THF to the dried oligosaccharide sample.

b. Add 2 μL of APTS Labeling Reagent to the sample.

B25802AA6

Page 21: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitPerforming the Labeling Reaction

IMPORTANT Due to the reaction of sodium cyanoborohydride with water, avoid moisture and store this product under dry conditions. Use a dry needle to dispense this chemical. Pass dry argon into the chemical bottle through the septum while dispensing 2 μL from the bottle.

c. Incubate the reaction mixture at 37°C for 4 hours, or at 60°C for 90 minutes, or at room temperature overnight (mildest reaction).

NOTE A mild labeling condition is necessary to avoid losing the sialic acid residues from your oligosaccharide chain. You can leave the reaction mixture at room temperature overnight and in the dark to do this. For Glucose Ladder incubation, select any of the temperature/ time options.

Figure 2 illustrates a summary of the overall labeling protocol.

2 Prepare the Sample and/or Standard for capillary electrophoresis.

a. To the 4 μL reaction mixture prepared above add:

• For Glucose Ladder Standard: add 96 μL of DDI water

• For all other samples: add 46 μL DDI water

This solution can be stored at 2 to 8°C.

b. Take an aliquot of 5 μL of the solution above and add 195 μL of DDI water.

c. Mix the contents well and transfer the solution into a micro vial for analysis in the PA 800 plus system see Sample Vial Setup.

Figure 3 illustrates a summary of the sample preparation for injection.

B25802AA 7

Page 22: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitPerforming the Labeling Reaction

Figure 2 Labeling Reaction Scheme

B25802AA8

Page 23: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitPreparing the PA 800 plus Instrument

Figure 3 Sample Preparation for Injection

Preparing the PA 800 plus Instrument

Before proceeding, review the following procedures in the PA 800 plus System Maintenance User’s Guide (B25799).

• Install the LIF Detector

• Calibrate the LIF Detector

• Capillary Cartridge Procedures

• Cleaning the Interface Block and the Opening Levers

B25802AA 9

Page 24: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitInstalling and Pre-conditioning the N-CHO Capillary

Installing and Pre-conditioning the N-CHO Capillary

• Prior to using the PA 800 plus system for carbohydrate analysis, the N-CHO capillary must be installed in a capillary cartridge. To properly install the capillary, refer to the Capillary Cartridge sections in the PA 800 plus System Maintenance manual (PN B25799).

IMPORTANT Be sure to pre-rinse a new capillary for 10 minutes at 30 psi pressure with DDI water and then rinse for 10 minutes at 30 psi with Carbohydrate Separation Buffer prior to the first run.

For this analysis, you will be using a 50 μm I.D. N-CHO capillary. The length from injection inlet to detector should be 40 cm, with the total length being 50.2 cm. Be sure to measure the capillary dimensions accurately and record the dimensions in the capillary/performance tab of the Advanced Method Options. See Figure 4. This will be necessary for accurate mobility determinations.

Figure 4 Capillary Performance Screen

IMPORTANT To get good reproducibility from capillary to capillary and accurate mobility assignments, it is important to adhere to the capillary pre-measurement procedure.

IMPORTANT The cut ends of capillaries should be inspected carefully under magnification. The cut must be clean (not jagged) and perpendicular to the capillary length (not angled). Poor cuts will result in poor resolution and poor sample loading.

NOTE When not in use, submerge the capillary ends in DDI water to prevent the capillary from drying out.

• Turn off the PA 800 plus system and install the LIF detection components.

• Turn on the system and permit the laser to warm up for at least 30 minutes prior to calibration.

B25802AA10

Page 25: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitPreparing the Buffer Trays

Clean the Interface Block

Clean the electrodes, opening levers, capillary tips, and interface block carefully following the cleaning procedure as described in the PA 800 plus System Maintenance manual (PN B25799). Do this on a weekly basis or when changing chemistries on the PA 800 plus system. Accumulation on the interface block can lead to current leakage errors.

Insert the Cartridge

Insert the cartridge into the system. Close the cartridge and sample covers.

Preparing the Buffer Trays

1 Fill the vials with the appropriate volumes of each reagent:

• 1.5 mL of DDI water per H20 vial (4 vials)

• 1.5 mL of Carbohydrate Separation Buffer per Gel-R vial (1 vial)

• 1.3 mL of Carbohydrate Separation Buffer per Gel-S vial (2 vials)

• 0.8 mL of DDI water per waste vial (1 vial)

Figure 5 Universal Vials and Caps

1. Universal Vial Cap2. Maximum Fill Level3. Universal Vials

B25802AA 11

Page 26: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitSample Vial Setup

2 Cover each vial with a cap as shown in Figure 5. Caps and vials are not intended for re-use and

should be disposed of after the end of the assay. Re-using vials or caps can result in pressure leakage and capillary breakage.

3 Place the reagent vials on the inlet and outlet buffer trays as indicated in Figure 6.

NOTE During the separation, salts migrate from one vial of gel buffer to the other, changing the composition of the buffer. Therefore, it is recommended that the vials of buffer be replaced with fresh buffer after 20 runs. For unattended operation of more than 20 runs, simply program the increment function to increment the water and buffer vials after 20 runs. Vials will increment in a forward direction. Ensure that the tray is filled with the appropriate number of vials to manage the number of runs that are to be performed.

NOTE The H20 vials are used to clean the capillary tips after the sample introduction step to prevent sample carryover. Replace the vials with fresh water if rinse water becomes contaminated.

A schematic of the buffer tray setup is illustrated in Figure 6.

Figure 6 Buffer Tray Configuration

Sample Vial Setup

Before placing the micro vial containing the sample into the universal vials, ensure that no bubbles are at the bottom of the sample vials. If bubbles exist, centrifuge the sample vials for two minutes at 1,000 g and repeat if necessary. Place a cap on the universal vial and ensure a good seal. See Figure 7.

Place the universal vials into the 48-position inlet sample tray positions A1 through C8.

B25802AA12

Page 27: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitRunning the Samples

Figure 7 Micro Vial Inside a Universal Vial

Running the Samples

Initial Conditions

1. Universal Cap2. Micro Vial

3. Universal Vial4. Micro Vial inside Universal Vial

Table 3 Test Run: Initial Conditions

Capillary Temperature: 20°

Sample Storage Temperature: 10°

Auxiliary Data Channel: Current

B25802AA 13

Page 28: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitRunning the Samples

Figure 8 Initial Condition Tab

LIF Detector Initial Conditions

Table 4 Test Run: LIF Detector Initial Conditions

Detection Laser Induced Fluorescence

Wavelength Excitation - 488 nm, Emission - 520 nm

Data Rate 4 Hz

Dynamic Range 100 RFU (relative fluorescence units)

Filter Setting Normal

Peak Width 16 - 25

B25802AA14

Page 29: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitRunning the Samples

Figure 9 LIF Detector Initial Conditions Tab

NOTE Prior to use, calibrate the LIF detector using the procedure under LIF Detector Calibration Procedure. For comparable results, the detection system should be calibrated at system start-up or whenever a capillary is replaced.

Time Program

1 Rinse the capillary with buffer for three minutes at 30 psi from BI:B1 to waste at vial BO:B1.

2 Inject the sample at 0.5 psi for three seconds from sample vial to buffer vial BO:C1.

NOTE Either pressure or electrokinetic sample introduction can be used for labeled carbohydrate samples. A 3 to 20 second 0.5 psi pressure injection is recommended. Longer injection times may cause peak shape and migration times to vary. For low concentration samples, electrokinetic injections may be the method of choice. The concentration and amount of sample will determine the conditions of injection - typically 0.1 to 1.0 Ws (for example,10 kV, 5 microamps, 2-20 seconds) should be used.

NOTE Salts will interfere with electrokinetic injections. Samples should first be desalted before attempting this method.

B25802AA 15

Page 30: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitRunning the Samples

3 Wait 0.2 minutes with vials BI:A4 and BO:A4. See Figure 10. This step dips the capillary in water

to protect against sample carry over. Change the rinse water vials if they are contaminated.

4 Separate Step - 20 minutes from vial BI:C1 to vial BO:C1 (buffer). See Figure 10. The constant

voltage should be at 30 kV, with reverse polarity and a 0.17 ramp time.

5 Autozero at 1.0 minute.

6 End at 20.0 minutes.

Figure 10 Time Program Tab

B25802AA16

Page 31: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitEvaluation of Results

Figure 11 Typical Current Profile

NOTE Prior to use, calibrate the LIF detector using the procedure under LIF Detector Calibration

Procedure. For comparable results, the detection system should be calibrated at system start-up or whenever a capillary is replaced.

Evaluation of Results

The test mixture contains the APTS-labeled glucose oligomers consisting of at least 20 individual oligomers. An example of this test mix is highlighted in Figure 12.

Field Strength Generated 598 V/cm

Typical Current. See Figure 11 under 20 μA

B25802AA 17

Page 32: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitSystem Shutdown and Capillary Storage

Figure 12 Electropherogram - Glucose Ladder

System Shutdown and Capillary Storage

Short Term Storage (<24 hours) of Capillary

Perform a three minute, 30 psi rinse with water. The capillary may be stored on the instrument with the capillary ends immersed in water.

Whenever the capillary has not been used for three hours or longer, rinse the capillary by performing a three minute, 30 psi rinse with water before performing a separation.

Long Term Storage (>24 hours) of Capillary

1 Perform a 3 minute, 30 psi rinse with water. Repeat the rinse with Carbohydrate Separation

Buffer for 3 minutes.

2 Remove the capillary from the instrument and place in a cassette box with the capillary ends

placed in vials of DDI water.

3 Store the cartridge box at 2°C to 8°C in an upright position.

B25802AA18

Page 33: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitTroubleshooting Procedures

Whenever the capillary has not been used for three hours or longer, rinse the capillary by performing a 3 minute, 30 psi rinse with water and then a 10 minute rinse with Carbohydrate Separation Buffer before performing a separation.

Troubleshooting Procedures

Table 5 Carbohydrate Labeling Troubleshooting Chart

Symptom Corrective Action

No peaks • Check if your LIF detector has been calibrated. If not, please refer to the LIF calibration section of this system manual to perform the LIF Calibration.

• Make sure that your laser is ON. • Make sure the probe stabilizer is properly connected to the lock down

bar (white bar that locks down your LIF cartridge). • Make sure that you set up voltage in the method properly. This

separation should be run using the reverse polarity. Check if the proper polarity is being applied by confirming that the orange LED that indicates reverse polarity is on during the separation.

• Make sure that there is no bubble trapped at the bottom of the sample vial preventing the sample from being injected.

• Make sure the window of your capillary is intact. If the window is broken, change the capillary right away. Most importantly, clean the probe aligner with a cotton swab moistened in water.

Low Intensity Peaks • Labeling reaction was not carried out properly. For a successful labeling reaction, make sure that:— the protein is in a non-tris buffer. Tris will compete with your

carbohydrate for the reductive amination.— the carbohydrate pellets were fully dried. Any residue of water will

interfere in the yield of your labeling reaction.

• Enzyme activity may differ from lot to lot. You may need to adjust the suggested amount, if the reaction gives a yield lower than what you usually expect.

Low Current • Confirm N-CHO capillary with 50 cm length is in place. Confirm temperature. Replace reagent vials with new reagent vials.

Shift in Migration time in between runs, within the same day

• Migration time shifts, with the use of this application, are associated with the length of injection. Make sure that injections are between 3 - 20 s at 0.5 psi. The injection time depends on the concentration.

Carryover • Change water dip vials or change buffer vials and increment these vials as necessary

Spikes in Electropherogram • Check for air that is dissolved in the gel buffer. Sonicate the gel for 5 minutes.

Extra Peaks • It is very important to use extra clean 0.5 mL micro-centrifuge vials, especially during the labeling step. Contaminants left over from the manufacturing process of these vials may react with APTS, giving extra peaks on separation.

B25802AA 19

Page 34: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitAppendix

Appendix

LIF Detector Calibration Procedure

The PA 800 plus LIF detector calibration kit procedure is used to calibrate the detection system for carbohydrate analysis. This will ensure consistent results from day to day and from capillary to capillary, as detection is measured in relative fluorescence units (RFU). This calibration procedure allows you to normalize the signal generated from your analyte, relative to a known concentration of sodium fluorescein. Perform the calibration procedure when installing a new capillary or after detector and cartridge changes.

IMPORTANT The LIF calibration procedure described in this application guide has changed from previous versions. The result is different RFU values than what were obtained from assays standardized on the previous LIF calibration procedure. The calibration correction factor (CCF) is a scalar multiple of all data points, so there is no impact on signal to noise performance. This change was made to reduce the risk of saturating the detector from excessively high fluorescence signals.

Setting the Calibration Corrector Factors (CCFs)

IMPORTANT The user must have instrument administration privileges.

1 Launch 32 Karat.

2 Go to Tools > Enterprise Login.

3 Enter the user name PA 800 and password plus and select Log in.

4 Select the CHO instrument icon and then Configure Instrument.

5 Select Configure on the instrument configuration dialog box.

6 Select the LIF Detector icon in the right pane of the PA 800 plus System Configuration dialog box.

7 The 32 Karat Software Configuration dialog box appears as shown in Figure 13.

Table 6 Reagents Needed to Perform the Procedure

Materials Part Number

LIF Performance Test Mix 726022

DDI water N/A

B25802AA20

Page 35: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitAppendix

Figure 13 32 Karat Software Configuration Dialog Box

8 Select LIF Calibration Wizard from the Instrument Configuration dialog box to display the

Calibration Wizard. See Figure 14.

Automatic Calibration

1 Choose Auto as the calibration option, as shown in Figure 14 and click Next.

B25802AA 21

Page 36: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitAppendix

Figure 14 Calibration Wizard Screen - Auto Select

2 Set up the target RFU value. The target value for N-CHO-coated capillary is 7, as shown in

Figure 15, and click Next.

Figure 15 Calibration Wizard Screen - Step 2

B25802AA22

Page 37: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitAppendix

3 Fill one buffer vial with 1.5 mL DDI water and another with 200 μL of diluted LIF Performance

Test Mix (1:2 with DDI water) in a micro vial. Place them in the buffer vial positions as indicated in Figure 16.

4 Place an empty vial in the waste position.

IMPORTANT The system performance test kit (713360) contains instructions for performing detector noise tests on a bare-fused silica capillary, which uses borate buffer. Do not use the borate buffer for calibration of the N-CHO capillary. Use a 50% LIF Performance Test mix solution (100 μL of LIF Test Mix with 100 μL of DDI water) in a micro vial.

5 Click Next to perform the automatic calibration, as shown below.

Figure 16 Calibration Wizard Screen - Step 3

6 After the calibration is completed, a dialog box displays as shown in Figure 17.

A number will show in the Calibration Correction Factor field.

If it is below 10, click Accept to complete the calibration procedure.

If it is above 10. See CCF Troubleshooting Procedures.

B25802AA 23

Page 38: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitAppendix

Figure 17 Calibration Wizard Screen - Step 4

CCF Troubleshooting Procedures

CCF <1

CCF >1

Table 7 CCF <1

Symptom Corrective Action

CCF is still below 1, but not below 0.1 • There is no problem with the system. Run a standard and verify acceptable performance of the system.

System performance is not acceptable or the CCF is below 0.1

• Verify that the capillary for the carbohydrate analysis was used, and it is not broken.

• Verify laser output for the laser in use on the PA 800 plus system.

• Verify that the correct filters are installed in the LIF detector. • Contact a Beckman Coulter service representative.

Table 8 CCF >1

Symptom Corrective Action

CCF is more than 1.0 but less than 10 • Run a standard to determine if the system performance is still acceptable.

CCF is more than 10 or if the system performance is not acceptable.

• Verify laser output for the laser in use on the PA 800 plus system.

• Verify that the correct filters are installed in the LIF detector. • Replace the chemistries and capillary and repeat the

calibration. • Contact a Beckman Coulter service representative.

B25802AA24

Page 39: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitAppendix

No Step Change Detected

The LIF Calibration method is basically a comparison of detector signals between a non-fluorescent solution to a known fluorescent solution. When rinsed with the non-fluorescent solution followed by a fluorescent solution rinse, you should see the first part of the detector signal near zero and the second part of the detector signal near the target fluorescent value. This detector output is step shaped in appearance.

If no step change is observed, the appropriate solutions are not passing the detector or the detector is not able to detect them.

1 Verify that the solution is rinsing through the capillary by performing a direct control 5 minute

20 psi pressure rinse of water, from buffer inlet A1 to an empty buffer vial in outlet B1.

2 Once the rinse has begun, lift the sample cover.

3 Observe the outlet end of the capillary in B1. Droplets should be formed on the outlet end of the

capillary.

• If no droplets are formed, either the capillary is plugged or the instrument has a pressure failure.

• Replace the capillary and repeat the rinse. If there is still no droplet observed, contact a Beckman Coulter service representative.

Once flow through the capillary is verified, the detection system is the only other possible cause, when no step is detected.

4 Verify that the correct filters are installed in the LIF detector.

5 Verify that the laser provided with the PA 800 plus system is connected and on.

6 Run the calibration test with a new capillary and chemistries.

If there is still no step change detected, contact a Beckman Coulter service representative.

B25802AA 25

Page 40: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

Carbohydrate Labeling and Analysis KitAppendix

B25802AA26

Page 41: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250
Page 42: Application Guide · 2014-07-24 · B25802AA January 2013 Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis Beckman Coulter, Inc. 250

© 2012 Beckman Coulter, Inc.

All Rights Reserved

www.beckmancoulter.com

Related Documents

PA 800 plus System OverviewPN B25798 • System Overview • Instrument • Sample Handling System • Capillary Cartridge

• UV Detector Optics

• The Photo Diode Array (PDA) Detector

• Laser Induced Fluorescence (LIF) Detector

• How to Run a Method • Analysis • Peak Identity • Viewing Reports

PA 800 plus System MaintenancePN B25799 • Installation Procedures • Maintenance Procedures • Interfacing PA 800 plus with

Mass Spectrometry • LIF Test Mix Calibration and

Performance Tests • Evaluating PA 800 plus UV

Filters

PA 800 plus Methods DevelopmentPN B25800 • Introduction to CE • System Overview • 32 Karat • Direct Control • Creating and Editing a

Method • Running the System • Integration • Calibration • Qualitative Analysis • Mobility • Creating Reports • Additional Resources • PA 800 plus Calculations • Understanding PDA Data • System Administration • Vial Incrementing • Sequence Vial Report

PA 800 plus IgG Purity/Heterogeneity AssayPN B25801 • Kit Contents • IgG Sample Preparation • Buffer Vial Preparation and

Loading • Running the Assay • Running IgG Controls and

Test Samples

• Checking the Separation Results

• Troubleshooting Guide • Instrument and Detector

Initial Conditions • High-Speed Separation

Methods

PA 800 plus SDS-MW Analysis

PN B25803

• Storing the Assay Components

• Sample Preparation • Sample Vial • Buffer Vial Preparation and

Loading • Tray Configuration • Running the Assay • Running SDS-MW Size

Standard and Test Samples • Checking the Separation

Results • Troubleshooting Guide • Instrument and Detector

Initial Conditions

• SDS-MW Separation Method

PA 800 plus Capillary Isoelectric Focusing (cIEF) AnalysisPN B25804 • Overview • Capillary Isoelectric Focusing

(cIEF) Analysis • Principle of cIEF Separation • Materials and Reagents • Activation of System

Suitability

• Making a cIEF System Suitability Method

• Generating a System Suitability Report

• Buffer Exchange Ptrocedure

PA 800 plus System

Administration

PN B25805 • 32 Karat Enterprise • Getting Started • System Administration

Wizard • Using 32 Karat Under System

Administration

• Additional Features

• Example Administration Setup

• User Privileges Worksheet

PA 800 plus Pre-Installation

GuidePN B25806 • Pre-Installation Check List • Site Requirements • Specifications • PA 800 plus System Training

PA 800 plus Software Validation

PackPN B25807 • Software Validation • System Overview • Product History • Technical Description

• Additional Information

• Support Policies • Supplier Information • Development Environment • Supporting Documents

PA 800 plus User Manuals Binder

PN B00208

Notebook binder that contains all of the PA 800 plus manuals.

Available in hard copy by request.