Upload
others
View
4
Download
0
Embed Size (px)
Citation preview
The ANOTEST® YMP30-S measures the admittance according to standards, and due to its design is ideal-ly suited for on-site testing.
The weather resistance of a finish is a function of seal-ing quality. According to DIN EN ISO 12373-5 and ASTM B 457-67, the admittance (Y) of a capacitor in which the anodic oxide film forms the dielectric is a good yard-stick of sealing quality.
Handheld instrument for on-site testing
Coating Thickness Material Analysis Microhardness Material Testing
ANOTEST® YMP30-S Sealing quality tester for anodic coatings on aluminum
Coating Thickness Material Analysis Microhardness Material Testing
FeaturesLarge LCDBattery or AC line operatedAutomatic measurement start after insertion of the probe in the cell
Automatic computation of the admittance refer- enced to standard sizesSignaling for poor sealing (e.g., Y > 15 μS)Acoustic signal for start and end of measurementAutomatic shut-offStoring of 10,000 measurements in up to 100 applications
Statistical evaluationDisplay selectable in 8 languagesRS232 interface for printer or PC
Order information Order no.
Instrument incl. standard accessoriesANOTEST® YMP30-S 603-800Optional AccessoriesElectrical reference 600-772Placement support 600-025Interface set MP 602-341Adapter USB – RS232 603-650Carrying box MP 602-891Carrying case MP0D/30/40 602-120Spare PartsMeasuring cable YMP30-S 603-855Grounding cable ANOTEST® 600-767Grounding clamp, stainless steel 600-766Bottle testing liquid 600-768VE meas. cell ø 6 mm 600-769VE meas. cell ø 13 mm 600-770VE meas. cell ø 26 mm 600-771
Practical and sturdy carrying box for on-site measurements
ANOTEST® YMP30-S with grounding cable and grounding clamp, meas. electrode and meas. cell
Technical dataMeas. frequency: 1 kHz (DIN EN ISO 12373-5) Measuring Test cell ø 6 mm(1): 14 - 1400 μSranges: Test cell ø 13 mm: 3 - 200 μS
Test cell ø 26 mm(1): 0,75 - 75 μSPower supply: battery 9V6LR61Instrument dimensions (L x W x H): 160 mm x 80 mm x 30 mmMass: 230 g(1) The meas. cells with a ø of 6 mm and a ø of 26 mm do not permit standardized measurements because the test areas are too small or too large
Measurement procedureA self-adhesive rubber ring is placed on the surface to be tested and the space so enclosed filled with a few drops of a test fluid. The aluminum substrate is then connected to the earthing socket of the unit and the test electrode dipped into the cell fluid, thereby form-ing a capacitor in which the anodic film acts as the dielectric. It is the admittance of this capacitor which is measured and directly indicated in μS (microsie-mens).
For further information contact:
Official Distributor Dyne Testing Ltd Newton House, 5 Parkside Court Greenhough Road Lichfield Staffordshire WS13 7FE United Kingdom
Tel: +44 (0)1543 411460 Fax: +44(0)1543 415140