Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
1
Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC2/MS)Michael D. Jones, Andrew Aubin, Peter J. Lee, and Tim JenkinsWaters Corporation, Milford, MA, USA
IN T RO DU C T IO N
Organic light emitting diodes (OLEDs) are thin films that exhibit
electroluminescence when an electric current is applied. They are used
in a variety of everyday electronics such as televisions, mobile phones,
computer monitors, watches, and display screens. The continued development
and commercialization of OLEDs is of major interest to companies developing
next-generation display technology and their suppliers.
The raw materials used to construct OLED-based devices require high purity to
extend the life of the luminescence and quality of the final product, specifically
blue phosphorescent emitters. There is much intellectual property associated with
OLED materials, so manufacturing a high quality OLED raw material can also be
highly lucrative.
Intellectual property is a strong driver for the OLED chemical materials business,
which limits commoditization and drives high manufacturing costs for OLED-based
devices. Market analysis reports point to OLEDs as a key niche market over the
next five years, predicted to grow to a $5 billion market by 2016.1
Typically, OLED materials are analyzed by various microscopy techniques.
Interestingly, a limited amount of LC-based methods have been cited in
publications analyzing the stability of phosphorescent emitters used in OLED
devices.2,3,4 Many of the published methods require an analysis time greater
than 30 minutes.
In this application note, a rapid and selective method was developed utilizing
UltraPerformance Convergence Chromatography™ (UPC2TM) coupled with
photodiode array detection and mass spectrometry to analyze the purity of an
iridium complex dye, Ir(Fppy)3, shown in Figure 1. Ir(Fppy)3 is a phosphorescent
emitter material that is important for providing blue electroluminescence
longevity for use in OLED devices. The method specificity was also evaluated in
the presence of additional materials used in the construction of OLED devices.
The developed UPC2 method utilizes supercritical fluid chromatography (SFC)
carbon dioxide as the primary mobile phase, and methanol as a modifier
acting as the strong solvent to elute the planar organo-metallic based
constituents of interest.
WAT E R S SO LU T IO NS
Waters® ACQUITY UPC2TM System
ACQUITY UPC2 Column Kit
ACQUITY UPC2 PDA Detector
ACQUITY® SQD
Empower™ 3 CDS Software
K E Y W O R D S
ACQUITY UPC2, supercritical fluid
chromatography, fine chemical purity,
patent protection, stability indicating
methods, degradation profiling,
PhOLED, OLED, Ir(Fppy)3, impurities,
phosphorescent emitter
A P P L I C AT IO N B E N E F I T S■■ Provides an approach to improved profiling
of material purity
■■ 10X reduction in run time compared
to previously published liquid
chromatography techniques
■■ Significant cost reduction in mobile phase
consumption and waste
■■ Approximately $1.91 for HPLC
vs. $0.05 for UPC2
■■ Compatibility with normal phase diluents
and extraction solvents associated with
organic light emitting diode (OLED) analysis
2UPC2 Analysis of Ir(Fppy)3 Degradants in OLED Material
E X P E R IM E N TA L
Sample Description
Stock standards were prepared by diluting
Ir(Fppy)3 reference material in tetrahydrofuran
to a concentration of approximately 1.0 mg/mL.
Working standards were prepared by dilution of
the Ir(Fppy)3 stock standard with tetrahydrofuran
to a concentration of 0.1 mg/mL.
Method Conditions
System: Waters ACQUITY UPC2
Column: ACQUITY UPC2 BEH 2-EP
3.0 x 100 mm, 1.7 µm
Mobile phase A: CO2
Mobile phase B: 2 g/L Ammonium formate
in methanol
Wash solvents: 70:30 Methanol/isopropanol
Separation mode: Gradient: 10% to
25% B over 5 minutes;
hold at 25% for 1 minute
Flow rate: 2.0 mL/min
CCM back pressure: 1885 psi
Column temp.: 60 °C
Sample temp.: Ambient
Injection volume: 2.0 µL
Run time: 5.0 minutes
Detection: ACQUITY UPC2 PDA
3D Channel:
200 to 410 nm; 20 Hz
2D Channel:
258 nm @ 4.8 nm
resolution (Compensated
500 to 600 nm)
Mass spectrometer: ACQUITY SQD
MS settings: 200 to 1200 Da;
10,000 Da/sec; ES PosNeg
Make-up flow: None
Data management: Empower 3 CDS Software
R E SU LT S A N D D IS C U S S IO N
Method development
Typical SFC system configurations utilize a secondary pump to provide post
column addition of proton-rich solvent, such as formic acid, to improve the
ionization efficiency for MS analysis. The lack of ionization effectiveness of
legacy SFC configurations can be attributed to the high flow of high-percentage
liquid CO2 that lacks the protons required for ionization. The need for make-up
flow was evaluated for use with UPC2.
The ACQUITY UPC2 System configuration and methodology used for this analysis
did not require a secondary pump for make-up flow to increase ionization. The
flow to the mass spectrometer was split post-UV detection with an Upchurch zero
volume PEEK tee. However, it was determined that using 30 cm of 0.0025-inch
I.D. PEEK tubing to the inlet of the MS probe provided enough back pressure after
the split to maintain a supercritical CO2 phase state.
A method development scheme was initially designed to investigate three
ACQUITY UPC2 Columns providing differences in chromatographic selectivity.
The columns utilized were UPC2 BEH, UPC2 BEH 2-EthylPyridine (2-EP), and
UPC2 CSH Fluoro-Phenyl stationary phases, specifically designed for use with
the UPC2 instrumentation. Injections of the Ir(Fppy)3 degraded sample were
performed using a rapid 5-minute gradient screening method.
Based on this method development scheme and system configuration, it was
observed that the ACQUITY UPC2 BEH 2-EP stationary phase provided the most
optimal selectivity potential and best peak shape attributes to separate seven
unknown impurities, as shown in Figures 2 and 3. The spectra inset in Figures 2
and 3 were used to determine relationship to the Ir(Fppy)3 phosphorescent emitter.
The MS data provided added sensitivity to detect trace-level impurities such
as the smaller peaks detected between 1 and 2 minutes, shown in Figure 3. By
detecting these trace impurities by MS, the spectral information facilitates the
determination of origin or possible relationship to the main species of interest.
N
FF
N
F
F
Ir
N
F
F
Figure 1. The chemical structure of iridium complex Tris[2-(2,4-difluorophenyl)pyridine]iridium(III), referred to as Ir(Fppy)3.
3UPC2 Analysis of Ir(Fppy)3 Degradants in OLED Material
Impurity Identification
The peaks found in the UV and the MS data were integrated with spectral information extracted from the 3D
data channels in Empower 3 CDS Software. The software provided the ability to extract MS spectra from UV
peaks of interest within a single window for efficient data review.
A total of seven impurity peaks were found after subtraction of the blank injection peaks. Retention time,
UV spectra, and MS spectra are displayed in Table 1. Three of the seven impurities (RT 1.360 min,
RT 1.463 min, and RT 1.619 min) were observed to have similar UV spectral profiles to that of the Ir(Fppy)3
phosphorescent emitter. The MS spectral analysis of the seven peaks revealed a distinct relationship to the
Ir(Fppy)3 phosphorescent emitter.
The Ir(Fppy)3 MS spectra have a unique isotopic ratio, as seen in the spectra found in Figure 3. All seven
impurity peaks possessed isotopic ion ratios similar to the Ir(Fppy)3 phosphorescent emitter. Electrospray
negative ion results were observed for three of the four later-eluting peaks. The negative ion information will
help further investigations focused on impurity characterization by accurate mass measurements.
1.01
9
AU
0.00
0.03
0.06
0.09
0.12
Minutes
0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00
1.019 Peak 1
225.4
258.4
373.4
AU
0.018
0.036
0.054
0.072
nm220.00 260.00 300.00 340.00 380.00
UV 258 nm
Ir(Fppy)3
Figure 2. UV 258 nm chromatogram (compensated 500 to 600 nm) of Ir(Fppy)3 and impurity peaks found in a degraded sample preparation. The integrated peak was determined as the Ir(Fppy)3 analyte. The inset figure is the UV spectal profile of Ir(Fppy)3 with an observed λmax of 258.4 nm.
Figure 3. MS ES+ total ion chromatogram (TIC) of Ir(Fppy)3 and impurity peaks found in a degraded sample preparation. The integrated peak was determined as the Ir(Fppy)3 analyte. The adjacent figure is the MS spectral isotopic profile of Ir(Fppy)3 with an observed m/z base peak of 763.9 Da.
1.00
3
Inte
nsity
1.8x107
3.6x107
5.4x107
7.2x107
9.0x107
Minutes
0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00
1.004 Peak 1 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
761.95
763.92
Inte
nsity
0.0
7.0x106
1.4x107
2.1x107
2.8x107
m/z
744.80 748.60 752.40 756.20 760.00 763.80 767.60 771.40 775.20 779.00
Ir(Fppy)3
MS ES+
4UPC2 Analysis of Ir(Fppy)3 Degradants in OLED Material
Unknown impurity characterization using nominal mass measurements is typically undesirable; however,
the impurities observed during this experiment resemble the same isotopic patterns as the parent Ir(Fppy)3
compound. Based on the MS results, it is hypothesized that the unknown peak at retention time 1.94 minutes
is an Ir(Fppy)3 isomer. The unknown peaks at retention times 1.330 min, 1.463 min, 1.619 min, and 2.75 min
each represents a reduction of a fluorine atom from one of the fluoro-phenyl rings. The unknown peak at
3.622 min represents a reduction of two fluorine atoms from one of the fluoro-phenyl rings. The peak eluting
at 4.12 min requires further investigation by MS/MS and accurate mass to better characterize a proposed
structure, as shown in Figure 4.
1.360 Peak 2
213.7
221.9227.8239.5
262.0
317.8322.5
330.9
336.8
353.0360.2
367.4
381.8386.6392.6
AU
0.00024
0.00048
0.00072
0.00096
0.00120
nm220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
1.463 Peak 3
213.7
228.9232.5
260.8
301.1
322.5
348.2351.8355.4362.6
377.0385.4
396.2
AU
0.00025
0.00050
0.00075
0.00100
0.00125
nm220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
1.619 Peak 4
216.0
230.1233.6241.9
260.8
292.8
315.4322.5
330.9336.8348.2 360.2 375.8
393.8398.6
AU
0.00000
0.00025
0.00050
0.00075
0.00100
nm220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
1.019 Peak 1
225.4
258.4
373.4
AU
0.018
0.036
0.054
0.072
nm220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
1.345 Peak 2 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
743.93
745.90
Inte
nsity
0.0
9.0x105
1.8x106
2.7x106
3.6x106
m/z734.40 737.10 739.80 742.50 745.20 747.90 750.60 753.30 756.00 758.70
1.446 Peak 3 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
743.97
745.91
Inte
nsity
0.0
1.2x106
2.4x106
3.6x106
4.8x106
m/z738.00 740.00 742.00 744.00 746.00 748.00 750.00 752.00 754.00
1.601 Peak 4 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
743.90
745.90
746.97
Inte
nsity
0
1x106
2x106
3x106
4x106
m/z739.20 741.30 743.40 745.50 747.60 749.70 751.80 753.90 756.00
1.004 Peak 1 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
761.95
763.92
765.01
Inte
nsity
0.0
7.0x106
1.4x107
2.1x107
2.8x107
m/z759.60 760.80 762.00 763.20 764.40 765.60 766.80 768.00 769.20
Retention Time UV Spectra MS ES+ Spectra MS ES- Spectra
N/A
N/A
N/A
N/A
Ir(Fppy)3 RT λmax ES+ m/z ES- m/z
1.11 min 258.4 nm 763.9 N/A
RT λmax ES+ m/z ES- m/z
1.36 min 262.0 nm 745.9 N/A
RT λmax ES+ m/z ES- m/z
1.46 min 260.8 nm 745.9 N/A
RT λmax ES+ m/z ES- m/z
1.62 min 260.8 nm 763.9 N/A
4.124 Peak 8
212.5221.9
238.4
264.3272.6
314.2320.1323.7332.1345.9
363.8367.4
379.4383.0386.6
AU
0.00075
0.00150
0.00225
0.00300
0.00375
nm220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
3.622 Peak 7
212.5
237.2
275.0
341.3 354.2 368.6
383.0
AU
0.0006
0.0012
0.0018
0.0024
0.0030
nm220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
2.746 Peak 6
236.0
273.8
345.9
371.0384.2
AU
0.0051
0.0102
0.0153
0.0204
0.0255
nm220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
1.935 Peak 5
234.8
270.3
343.6
371.0380.6
AU
0.013
0.026
0.039
0.052
nm220.00 240.00 260.00 280.00 300.00 320.00 340.00 360.00 380.00 400.00
1.922 Peak 6 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
761.97
763.93
765.08
Inte
nsity
0.0
7.0x106
1.4x107
2.1x107
2.8x107
m/z757.80 759.60 761.40 763.20 765.00 766.80 768.60 770.40 772.20
2.731 Peak 7 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
743.96
745.95
747.05
Inte
nsity
0.0
6.0x106
1.2x107
1.8x107
2.4x107
m/z738.00 740.00 742.00 744.00 746.00 748.00 750.00 752.00 754.00 756.00
3.605 Peak 8 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
725.97
727.94
729.01Inte
nsity
0.0
1.6x106
3.2x106
4.8x106
6.4x106
m/z722.40 723.80 725.20 726.60 728.00 729.40 730.80 732.20 733.60 735.00
4.101 Peak 9 - SQ 1: MS Scan 1: 300.00-1000.00 ES+, Centroid, CV=Tune
804.01
805.99
806.96
Inte
nsity
0.0
1.2x106
2.4x106
3.6x106
4.8x106
m/z802.80 803.70 804.60 805.50 806.40 807.30 808.20 809.10 810.00
Retention Time UV Spectra MS ES+ Spectra MS ES- Spectra 1.920 Peak 1 - SQ 2: MS Scan 2: 300.00-1000.00 ES-, Centroid, CV=Tune
806.0
808.0
809.0
Inte
nsity
0.00
2.50x106
5.00x106
7.50x106
1.00x107
m/z798.60 800.80 803.00 805.20 807.40 809.60 811.80 814.00 816.20 818.40
2.730 Peak 2 - SQ 2: MS Scan 2: 300.00-1000.00 ES-, Centroid, CV=Tune
787.9
789.9
790.9
Inte
nsity
0.0
6.0x105
1.2x106
1.8x106
2.4x106
m/z785.40 786.50 787.60 788.70 789.80 790.90 792.00 793.10 794.20 795.30
4.103 Peak 3 - SQ 2: MS Scan 2: 300.00-1000.00 ES-, Centroid, CV=Tune
848.0
849.0
850.0
851.0Inte
nsity
0.0
5.0x105
1.0x106
1.5x106
2.0x106
m/z846.00 847.00 848.00 849.00 850.00 851.00 852.00 853.00 854.00 855.00
N/A
RT λmax ES+ m/z ES- m/z
1.94 min 234.8 nm 763.9 808.0
RT λmax ES+ m/z ES- m/z
2.75 min 236.0 nm 746.0 789.9
RT λmax ES+ m/z ES- m/z
3.62 min 237.2 nm 727.9 N/A
RT λmax ES+ m/z ES- m/z
4.12 min 238.4 nm 806.0 850.0
Table 1. Table of retention times, UV observed spectral profile, MS ES+ observed spectral profile, MS ES- observed spectral profile (if present). The MS spectral profiles verify similar isotopic patterns. Combined with the UV spectra, the unknown peaks were determined to be related to Ir(Fppy)3.
5UPC2 Analysis of Ir(Fppy)3 Degradants in OLED Material
Investigation of Specificity
The method was investigated in terms of specificity using a mixture of other main ingredients used in the
construction of OLED devices. The constituents within the mixture represent the phosphorescent emitter
[Ir(Fppy)3], the hole transport material (α-NHP), the host material (TCTA), and the hole blocking material/
electron transport material Alq3. The mixture was injected, and the assay method was determined to be
specific to analyze all three doped compounds without interference with the Ir[Fppy]3 phosphorescent emmitter.
System precision was determined over n=5 injections to assess method reproducibility, displayed in Figure 5.
1.01
9
1.36
01.
463
1.61
9
1.93
5
2.74
6
3.62
2
4.12
4
AU
0.00
0.03
0.06
0.09
0.12
Mi t
0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00
Ir (Fppy )3
N
FF
N
F
F
Ir
N
F
F
N
F
F
Ir
N
F
F
N
F
F
Ir (Fppy )3Isomer
N
F
Ir
N
F
F
N
F
F
Ir (Fppy )3 –FIsomers
N
F
Ir
N
F
N
F
F
Ir (Fppy
Relatedunknown
)3 -2FIr (Fppy )3 -F
Minutes
Figure 4. Proposed structure assignments for the unknown impurities found in the Ir(Fppy)3 phosphorescent emitter sample.
1.0
09
1.4
21
1.9
20
2.0
28
2.4
71
AU
0.000
0.025
0.050
0.075
0.100
Minutes0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00
Ir(Fppy)3Area = 0.4%Rt = 0.1%
TCTAArea = 1.1%Rt = 0.1% Alq3
Area = 1.1%Rt = 0.1%
α-NHPArea = 0.4%Rt = 0.1%
Figure 5. Method tested for specificity and system precision. Overlay of a mixture of commonly used components in OLED production (n=5). %RSD of retention time and area are reported above each major compound in the mixture. The additional impurity peaks are those impurities related to Ir(Fppy)3 .
Waters Corporation34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
References
1. Bcc Research Market Report SMC069B, “Organic Light Emitting Diodes (OLEDs): Technologies and Global Markets” July 2011
2. Sivasubramaniam et al, “Investigation of FIrpic in PhOLEDs via LC/MS Technique”, Cent. Eur. J. Chemistry, 7(4), (2009), 836-845
3. Baranoff et al, “Sublimation Not an Innocent Technique: A Case of Bis-Cyclometalated Iridium Emoitter for OLED”, Inorg. Chem, 47, (2008), 6575-6577
4. Kondakov, D., Lenhart, W., and Nichols, W., “Operational degradation of organic light-emitting diodes: Mechanism and identification of chemical products”, J. Appl. Physics, 101, 024512, (2007)
CO N C LU S IO NS
A rapid 5-minute UPC2/MS method was developed to analyze
Ir(Fppy)3 phosphorescent emitter purity, and an approach to monitor
long-term stability. The MS data provided by the ACQUITY SQD
facilitated characterization of three of the unknown impurity peaks.
The UPC2 approach resulted in up to 10X improvements in run time,
solvent consumption, and considerably reduced amounts of solvent
waste when compared with previously reported LC/MS techniques
that used 100% organic solvents during the analysis. The highly
selective and unmatched specificity of this UPC2 approach will help
chemical material manufacturers better understand and control the
performance of these unstable blue phospho-organic light emitting
diodes, ultimately leading to better quality OLED-based products
and a means to improve protection of intellectual property.
Waters and ACQUITY are registered trademarks of Waters Corporation. UltraPerformance Convergence Chromatography, UPC2, ACQUITY UPC2, Empower, and T he Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2012 Waters Corporation. Produced in the U.S.A.April 2012 720004305EN AG-PDF