Upload
sheeron20
View
338
Download
2
Tags:
Embed Size (px)
Citation preview
Analog Electrical Devices and Measurements
2141-375 Measurement and Instrumentation
Analog Devices: Current Measurements
θsin ILBF
BLIF
=
×=rrr
Force on a conductor
F B
I A conductor is placed in a uniform magnetic field B T, at an angle of θθθθ. The current flow in the conductor is IA. Force exerted on the conductor can be calculated from
θθθθ
Analog Devices: Current Measurements
ILBkx =
Force on a conductor
B
I
With no current flowing through the conductor, the spring will be at its unstretched length. As current flows through the conductor, the spring will stretch and developed force required to balance the electromagnetic force.
IFs = kx
F = IBL
k = spring constant, x = the total distance moved by the spring and θθθθ is 90 o
xBL
kI =
D’Arsonval or PMMC Instrument
Important parts of PMMC Instrument
•Permanent magnet with two soft-iron poles• Moving coil • Controlling or restoring spring
PMMC = Permanent Magnet Moving Coil
Torque Equation and Scale
When a current I flows through a one-turn coil in a magnetic field , a force exerted on each side of the coil
L = the length of coil perpendicular to the paper
F = IBLF = IBLF = IBL
Since the force acts on each side of the coil, the total force for a coil of N turns is
D
F = NIBL
The force on each side acts at a coil diameter D, producing a deflecting torque
TD = NIBLD
Torque Equation and Scale
The controlling torque exerted by the spiral springs is proportional to the angle of deflection of the pointer:
Since all quantities except θθθθ and I are constant for any given instrument, the deflection angle is
θθθθ = CI
TC = Kθθθθ
Where K = the spring constant. For a given deflection , the controlling and deflecting torques are equal
BLIND = Kθθθθ
Therefore the pointer deflection is always proportional to the coil current. Consequently, the scale of the instrument is linear.
Galvanometer
• Galvanometer is essentially a PMMC instrument designed to be sensitive to extremely current levels.
• The simplest galvanometer is a very sensitive instrum ent with the type of center-zero scale, therefore the pointer can be deflected to either right or left of the zero position.
• The current sensitivity is stated in µµµµA/mm
• Galvanometers are often used to detect zero current or vo ltage in a circuit rather than to measure the actual level of curre nt or voltage. In this situation, the instrument is referred as a null detector.
DC Ammeter
PMMC instrument
Ammetershunt
Rs
Coil resistance
Rm
Im VmI = Is + Im Is
Rs
I
Shunt resistance
s
mms I
RIR =
m
mms II
RIR
−=
ssmm RIRI =sm VV =
• An ammeter is always connected in series with a circuit.
• The internal resistance should be very low
• The pointer can be deflected by a very small current
• Extension of ranges of ammeter can be achieved by connecting a very low shunt resistor
Example: An ammeter has a PMMC instrument with a coil resistance of Rm = 99 Ω and FSD current of 0.1 mA. Shunt resistance Rs = 1 Ω. Determine the total current passing through the ammeter at (a) FSD, (b) 0.5 FSD, and (c) 0.25 FSD.
Known: FSD of Im Rm and Rs
Solution:
DC Ammeter
0.25FSD0.5FSDFSD
2.5510I = Im + Is (mA)
2.4754.959.9Is (mA)
0.0250.050.1Im (mA)
Example: A PMMC instrument has FSD of 100 µA and a coil resistance of 1kΩ. Calculate the required shunt resistance value to convert the instrument into an ammeter with (a) FSD = 100 mAand (b) FSD = 1 A.
Known: FSD of Im Rm
Solution:
DC Ammeter
(a) FSD = 100 mA: Rs = 1.001 Ω
(b) FSD = 1 A: Rs = 0.10001 Ω
DC Ammeter: Multirange
B
AE
C
D
Rs1
Rs2
Rs3
Rs4
Rm
• A make-before-break must be used so that instrument is not left without a shunt in parallel to prevent a large current flow t hrough ammeter.
Multirange ammeter using switch shunts
Make-before break switch
B
A
E
C
D
DC Ammeter: Ayrton shunt
Rm
R1 R2 R3
VSIm
I Is Is
Im
B
C
D
AI
+
-
An Ayrton shunt used with an ammeter consists of several series-connected resistors all connected in parallel with the PMMC instrument. Range change is effected by switch between resistor junctions
Rm
R1 R2 R3
VSIm
I Is Is
Im
I
B
C
D
A
+
-
R1 + R2 + R3 in parallel with Rm
R1 + R2 in parallel with Rm + R3
Example: A PMMC instrument has a three-resistor Ayrton shunt connected across it to make an ammeter. The resistance values are R1 = 0.05 Ω, R2 = 0.45 Ω, and R3 = 4.5 Ω . The meter has Rm = 1 kΩ and FSD = 50 µA. Calculate the three ranges of the ammeter.
Known: FSD of Im Rm R1 R2 and R3
Solution:
DC Ammeter
DCBPosition
1000.05100.0510.05IFSD (mA)
Rm
R1 R2 R3
VSIm
I Is Is
Im
B
C
D
AI
+
-
DC Voltmeter
PMMC instrument
Series resistance or “multiplier”
mm
s RI
VR −=mmsm RIRIV +=
Given V = Range
V
RmRs
Im
V
Coil resistance
Multiplierresistance
mm
s RI
RangeR −=
The reciprocal of full scale current is the voltmeter sensitivity (k ΩΩΩΩ/V)
The total voltmeter resistance = Sensitivity X Range
• An ammeter is always connected across or parallel with the points in a circuit at which the voltage is to be measured.• The internal resistance should be very high
DC Voltmeter: Multirange
( )RRIV mm +=
• Multirange voltmeter using switched multiplier resistors
Meter resistance
Rm
R1
R2
R3
V
Multiplierresistors
R1 R2 R3Rm
V
• Multirange voltmeter using series-connected multiplier resistor
Where R can be R1, R2, or R3
( )RRIV mm +=
Where R can be R1, R1 + R2, or R1 + R2 + R3
Example: A PMMC instrument with FSD of 50 µA and a coil resistance of 1700 Ω is to be used as a voltmeter with ranges of 10 V, 50 V, and 100 V. Calculate the required values of multiplier resistor for the circuit (a) and (b)
Known: FSD of Im Rm
Solution:
DC Ammeter
Meter resistance
Rm
R1
R2
R3
V
Multiplierresistors
R1 R2 R3Rm
V
(a) (b)
R3R2R1
1.9983 MΩΩΩΩ998.3 kΩΩΩΩ198.3 kΩΩΩΩ
R3R2R1
1 MΩΩΩΩ800 kΩΩΩΩ198.3 kΩΩΩΩ
Ohmmeter: Voltmeter-ammeter method
V
A
VS
+
-
Rx
I
IVIx
V
+
-
VVS
+
-
Rx
I
A
+ -VA
+
-
VxV
-
Pro and con:
•Simple and theoretical oriented•Requires two meter and calculations•Subject to error: Voltage drop in ammeter (Fig. (a))
Current in voltmeter (Fig. (b))
Fig. (a) Fig. (b)
Measured Rx:
meas xR R≈
measx A A
x
V VV VR R
I I I
+= = = +
if Vx>>VA
Therefore this circuit is suitable for measure large resistance
Measured Rx: meas 1 /x
x V V x
RV VR
I I I I I= = =
+ +
meas xR R≈if Ix>>IV
Therefore this circuit is suitable for measure small resistance
Ohmmeter: Series Connection
•Voltmeter-ammeter method is rarely used in practical a pplications (mostly used in Laboratory)•Ohmmeter uses only one meter by keeping one parameter co nstant
Example: series ohmmeter
15k
0
∞
2550
75
100 µA
45k 5k
0
Infinity
Rx
Resistance tobe measured
Standardresistance
RmMeter
resistance
Meter
Battery
VS
R1
Basic series ohmmeter consisting of a PMMC and a series-connected standard resistor (R1). When the ohmmeter terminals are shorted (Rx = 0) meter full scale defection occurs. At half scale defection Rx = R1 + Rm, and at zero defection the terminals are open-circuited.
Basic series ohmmeter Ohmmeter scale
Nonlinear scale
1s
x m
VR R R
I= − −
Loading Effect: Voltage Measurement
LinearCircuit V
a
b
Vab RmVVab Rm
Rth
Vth
a
bUndisturbed condition: Rm = ∞
ab u thV V V= =
Measured condition: Rm ≠ ∞m
ab m thm th
RV V V
R R= =
+
Measurement error: 100m u
u
V Verror
V
−= ×
1
1 /m uth m
V VR R
=+General equation:
%100/1
1%100 ×
+−=×
+−=
thmthm
th
RRRR
R
Therefore, in practice, to get the acceptable results, we must have Rm ≥ 10 Rth (error ~ 9%)
Loading Effect
Circuit before measurement
10 V
R1100kΩΩΩΩ
R2100kΩΩΩΩ
5 V
5 V V
V
10 V
100kΩΩΩΩ
100kΩΩΩΩ
6.7 V
3.3 V 100kΩΩΩΩ
10 V
100kΩΩΩΩ
100kΩΩΩΩ
6 V
4 V 200kΩΩΩΩ
Circuit under measurement
V 3.3V 10100//100100
100//100=
+=measV
V 0.4V 10100//200100
100//200=
+=measV
V
10 V
100kΩΩΩΩ
100kΩΩΩΩ
5.2 V
4.8 V 1000kΩΩΩΩ
V 8.4V 10100//1000100
100//1000=
+=measV
Example Find the voltage reading and % error of each reading obtained with a voltmeter on (i) 5 V range, (ii) 10 V range and (iii) 30 V range, if the instrument has a 20 kΩ/V sensitivity, an accuracy 1% of full scale deflection and the meter is connected across Rb
Loading Effect
SOLUTION The voltage drop across Rb without the voltmeter connection
On the 5 V range
V 550k 5k 45
k 5=×
+=
+= V
RR
RV
ba
bb
kΩ 100 5V /20range =×Ω=×= VkSRm
kΩ 4.76 k 5 k 100
k 5 k 100=
+×
=+
=bm
bmeq RR
RRR
The voltmeter reading is
V 4.782 50k 76.4 k 54
k .764=
+=
+= V
RR
RV
eqa
eqb
50 V
Ra45kΩΩΩΩ
Rb5kΩΩΩΩ
45 V
5 V
Loading Effect
Error of the measurement is the combination of the loading effect and the meter error
The loading error = 4.782 - 5 = -0. 218 V
The meter error = ± 5 x = ± 0.05 V1100
∴% of error on the 5 V range:
%36.5100V 5
V 05.0V 218.0±=×
±−=
± 6.10± 0.35± 0.3-0.054.9530
± 4.40± 0.22± 0.1-0.124.8810
± 5.36± 0.27± 0.05-0.224.785
% errorTotal error (V)
Meter error (V)
Loading error (V)
Vb .
(V)Range
(V)
Loading Effect: Current Measurement
LinearCircuit A
a
b
Vab Rm
I
AVab Rm
Rth
Vth
a
b
I
Undisturbed condition: Rm = 0
Measured condition: Rm ≠ 0
Measurement error:
General equation:
Therefore, in practice, to get the acceptable results, we must have Rm ≤≤≤≤ Rth /10 (error ~ 9%)
ththu RVII /==
( )mththm RRVII +== /
( )thmum RRII /1/ +=
100×−
=u
um
I
IIerror
%100/1
1%100 ×
+−=×
+−=
mththm
m
RRRR
R
AC Voltmeter: PMMC Based
D W
Waveform Amplitude Average RMS
A
A
A
A
A
A
0
πA
πA2
AWD
D
+
0
0
AWD
D
+
A
2
A
2
A
2
A
3
A
AC Voltmeter: PMMC Based• Basic PMMC instrument is polarized, therefore its t erminals must be identified as + and -.• PMMC instrument can not response quite well with th e frequency 50 Hz or higher, So the pointer will settle at the average v alue of the current flowing through the moving coil: average-responding meter.
Full-wave Rectifier Voltmeter
RmD1
D2
D3
D4
Rs
Multiplierresistors
VpVrms
Vav
•Using 4 diodes
•On positive cycle, D1 and D4 are forward-biased, while D2 and D3 are reverse-biased
•On negative cycle, D2 and D3 are forward-biased, while D1 and D4 are reverse-biased
•The scale is calibrated for pure sine with the scale factor of 1.11 (A/ √√√√2 / 2A/ππππ)•Actually voltage to be indicated in ac
measurements is normally the rms quantity
Example: A PMMC instrument has FSD of 100 µA and a coil resistance of 1kΩ is to be employed as an ac voltmeter with FSD = 100 V (rms). Silicon diodes are used in the full-bridge rectifier circuit (a) calculate the multiplier resistance value required, (b) the position of the pointer when the rmsinput is 75 V and (c) the sensitivity of the voltmeter
Known: FSD of Im Rm
Solution:
AC Voltmeter: PMMC Based
RmD1
D2
D3
D4
Rs
Multiplierresistors
VpVrms
Vav
(a) Rs = 890.7 kΩ(b) 0.75 of FSD(c) 9 kΩ/V
AC Voltmeter: PMMC BasedHalf-wave Rectifier Voltmeter
Rs
RSH
D1
D2
RmVp
Vrms
Vav
•On positive cycle, D1 is forward-biased, while D2 i s reverse-biased
•On negative cycle, D2 is forward-biased, while D1 i s reverse-biased
•The shunt resistor R SH is connected to be able to measure the relative lar ge current.
•The scale is calibrated for pure sine with the scal e factor of 2.22 (A/ √√√√2 / A/ππππ)
Example: A PMMC instrument has FSD of 50 µA and a coil resistance of 1700 Ω is used in the half-wave rectifier voltmeter. The silicon diode (D1) must have a minimum (peak) forward current of 100 µA. When the measured voltage is 20% of FSD. The voltmeter is to indicate 50 Vrms at full scale Calculate the values of RS and RSH.
Known: FSD of Im Rm
Solution:
AC Voltmeter: PMMC Based
RS = 139.5 kΩ RSH = 778 Ω
Rs
RSH
D1
D2
RmVp
Vrms
Vav
Example The symmetrical square-wave voltage is applied to an average-responding ac voltmeter with a scale calibrated in terms of the rms value of a sine wave. If the voltmeter is the full-wave rectified configuration. Calculate the error in the meter indication. Neglect all voltage drop in all diodes.
T
Em
E
t
AC Voltmeter: PMMC Based
Solution 11%
Bridge Circuit
DC Bridge(Resistance)
AC Bridge
Inductance Capacitance Frequency
Schering Bridge Wien BridgeMaxwell BridgeHay BridgeOwen BridgeEtc.
Wheatstone BridgeKelvin BridgeMegaohm Bridge
Bridge Circuit
Bridge Circuit is a null method, operates on the princ iple of comparison. That is a known (standard) value is adjus ted until it is equal to the unknown value.
Wheatstone Bridge and Balance Condition
V
R1
R3
R2
R4
I1 I2
I3I4
The standard resistor R3 can be adjusted to null or balance the circuit.
A
B
C
D
Balance condition:
No potential difference across the galvanometer (there is no current through the galvanometer)
Under this condition: VAD = VAB
1 1 2 2I R I R=And also VDC = VBC
3 3 4 4I R I R=
where I1, I2, I3, and I4 are current in resistance arms respectively, since I1 = I3 and I2 = I4
1 2
3 4
R R
R R= or 2
4 31
x
RR R R
R= =
1 Ω
2 Ω 2 Ω
1 Ω
12 V
1 Ω
2 Ω 20 Ω
10 Ω
12 V
1 Ω
2 Ω 10 Ω
10 Ω
12 V
1 Ω
1 Ω 1 Ω
1 Ω
12 V
(a) Equal resistance (b) Proportional resistance
(c) Proportional resistance (d) 2-Volt unbalance
Example
Sensitivity of Galvanometer
A galvanometer is use to detect an unbalance conditi on in Wheatstone bridge. Its sensitivity is governed by: Current sensitivity (currents per unit defection) and internal resistance.
Thévenin Equivalent Circuit
G
A
B
C D
R3
R1 R2
R4
I1 I2
VS
Thévenin Voltage (VTH)
1 1 2 2CD AC ADV V V I R I R= − = −
where 1 21 3 2 4
and V V
I IR R R R
= =+ +
Therefore 1 2
1 3 2 4TH CD
R RV V V
R R R R
= = − + +
consider a bridge circuit under a small unbalance condition, and apply circuit analysis to solve the current through galvanometer
Sensitivity of Galvanometer (continued)
Thévenin Resistance (RTH)
R3
R1 R2
R4
A
B
C D 1 3 2 4// //THR R R R R= +
Completed Circuit
VTH
RTH
G
C
D
Ig=VTH
RTH+RgTH
gTH g
VI
R R=
+
where Ig = the galvanometer currentRg = the galvanometer resistance
Example 1 Figure below show the schematic diagram of a Wheatstone bridge with values of the bridge elements. The battery voltage is 5 V and its internal resistance negligible. The galvanometer has a current sensitivity of 10 mm/µA and an internal resistance of 100 Ω. Calculate the deflection of the galvanometer caused by the 5-Ω unbalance in arm BC
SOLUTION The bridge circuit is in the small unbalance condition since the value of resistance in arm BC is 2,005 Ω.
G
A
C
B
R3
R1 R2
R4
100 Ω Ω Ω Ω 1000 ΩΩΩΩ
2005 ΩΩΩΩ200 ΩΩΩΩ
5 V
(a)
100 ΩΩΩΩ A
B
C D200 Ω Ω Ω Ω 2005 Ω Ω Ω Ω
1000 Ω Ω Ω Ω
(b)
RTH= 734 ΩΩΩΩ
G
C
D
Ig=3.34 µµµµA
Rg= 100 ΩΩΩΩVTH
2.77 mV
(c)
D
Thévenin Voltage (VTH)
Thévenin Resistance (RTH)
100 // 200 1000 // 2005 734 THR = + = Ω
The galvanometer current2.77 mV
3.32 A734 100
THg
TH g
VI
R Rµ= = =
+ Ω + Ω
Galvanometer deflection10 mm
3.32 A 33.2 mmA
d µµ
= × =
mV 2.77
20051000
1000
200100
100 V 5
≈
+
−+
×=−= ACADTH VVV
Example 2 The galvanometer in the previous example is replaced by one with an internal resistance of 500 Ω and a current sensitivity of 1mm/µA. Assuming that a deflection of 1 mm can be observed on the galvanometer scale, determine if this new galvanometer is capable of detecting the 5-Ω unbalance in arm BC
Example 3 If all resistances in the Example 1 increase by 10 times, and we use the galvanometer in the Example 2. Assuming that a deflection of 1 mm can be observed on the galvanometer scale, determine if this new setting can be detected (the 50-Ω unbalance in arm BC)
Deflection Method
V
A
B
C D
R
R R
R+∆∆∆∆R
V
Consider a bridge circuit which have identical resistors, R in three arms, and the last arm has the resistance of R +∆R. if ∆R/R <<1
Small unbalance occur by the external environment
Thévenin Voltage (VTH)
Thévenin Resistance (RTH)
THR R≈
This kind of bridge circuit can be found in sensor applications, where the resistance in one arm is sensitive to a physical q uantity such as pressure, temperature, strain etc.
RR
RRVVV CDTH /24
/
∆+∆
==
In an unbalanced condition, the magnitude of the cu rrent or voltage drop for the meter or galvanometer portion of a bridge circuit i s a direct indication of the change in resistance in one arm.
Please correct
5 kΩΩΩΩ
Rv
6 V
Outputsignal5 kΩΩΩΩ
5 kΩΩΩΩ
(a)
0123456
0 20 40 60 80 100 120
Temp (oC)
Rv (
k ΩΩ ΩΩ)) ))
(b)
Example Circuit in Figure (a) below consists of a resistor Rv which is sensitive to the temperature change. The plot of R VS Temp. is also shown in Figure (b). Find (a) the temperature at which the bridge is balance and (b) The output signal at Temperature of 60oC.
4.5 kΩ
AC Bridge: Balance Condition
D
Z1Z2
Z4Z3
A C
D
B
I1 I2
all four arms are considered as impedance (frequency dependent components)The detector is an ac responding device: headphone, ac meterSource: an ac voltage at desired frequency
General Form of the ac Bridge
Z1, Z2, Z3 and Z4 are the impedance of bridge arms
At balance point: orBA BC 1 1 2 2E = E I Z = I Z
and1 21 3 2 4
V V I = I =
Z + Z Z + Z
V
1 4 2 3Z Z = Z Z
( ) ( )1 4 1 4 2 3 2 3Z Z =Z Zθ θ θ θ∠ + ∠ ∠ + ∠
Complex Form:
Polar Form:Magnitude balance:
Phase balance:
1 4 2 3Z Z =Z Z
1 4 2 3=θ θ θ θ∠ + ∠ ∠ + ∠
Example The impedance of the basic ac bridge are given as follows:
o1
2
100 80 (inductive impedance)
250 (pure resistance)
= Ω ∠
= Ω
Z
Z
Determine the constants of the unknown arm.
o3
4
400 30 (inductive impedance)
unknown
= ∠ Ω
=
Z
Z
Example an ac bridge is in balance with the following constants: arm AB, R = 200 Ωin series with L = 15.9 mH R; arm BC, R = 300 Ω in series with C = 0.265 µF; arm CD, unknown; arm DA, = 450 Ω. The oscillator frequency is 1 kHz. Find the constants of arm CD.
SOLUTION
The general equation for bridge balance states that
D
Z1Z2
Z4Z3
A C
D
B
I1 I2V1
2
3
4
200 100
1/ 300 600
450
unknown
R j L j
R j C j
R
ω
ω
= + = + Ω
= + = − Ω
= = Ω
=
Z
Z
Z
Z
1 4 2 3Z Z = Z Z
_
+
InvertingInput
VEE(-)
VCC(+)
Non-invertingInput
Output
_
+
Operational Amplifier: Op Amp
(a) Electrical Symbol for the op amp (b) Minimum connections to an op amp
Ideal Op Amp Rules:1. No current flows in to either input terminal2. There is no voltage difference between the two input terminals
I1
I2
V-V+
Rule 1: I1 = I2 = 0; R+/- = ∞∞∞∞Rule 2: V+ = V-; Virtually shorted
Vout
Inverting Amplifier
_
+
Rf
R1
vout
+
-vin
KCL
A
Use KCL at point A and apply Rule 1:
1
0A in A out
f
v v v v
R R
− −+ =
(no current flows into the inverting input)
Rearrange
1 1
1 10in out
Af f
v vv
R R R R
+ − + =
Apply Rule 2: (no voltage difference between inverting and non-inverting inputs)
Since V+ at zero volts, therefore V- is also at zero volts too. 0Av =
1
0in out
f
v v
R R+ =
1
fout
in
Rv
v R= −
Inverting Amplifier: another approach
Given vin = 5sin3t, R1=4.7 kΩ and Rf =47 kΩ
vout = -10vin = -50 sin 3t mV
_
+
Rf
R1
i
i
vout
+
-vin
No current flows into op amp
Since there is no current into op amp (Rule 1)
0f outV iR v−− + + =
From Rule 2: we know that V- = V+ = 0, and therefore
1
invi
R=
1 0inv iR V −− + − =0
0out fv iR= −
Combine the results, we get
1
fout
in
Rv
v R= −
1 2 3 4 5 6time
-60
-40
-20
20
40
60
mV
vin
vout
Non-inverting Amplifier
Rf
R1
vout
+
-vin
KCL
A
Use KCL at point A and apply Rule 1:
1
0A outA
f
v vv
R R
−+ =
Apply Rule 2:
_
+in Av v=
1
1 fout
in
Rv
v R= +
Given vin = 5sin3t, R1=4.7 kΩ and Rf =47 kΩ
vout = 11vin = 55 sin 3t mV 1 2 3 4 5 6time
-60
-40
-20
20
40
60
mV
vin
vout
Summing Amplifier: Mathematic Operation
Use KCL and apply Rule 1:
31 2 0A A outA A
f
v v v vv v v v
R R R R
− −− −+ + + =
_
+
Rf
Ri1
vout
+
-v1
v2
v3
i2
i3
R
R
i
vA
vB
1 2 3i i i i= + +
Since vA = 0 (Rule 2)
( )1 2 3f
out
Rv v v v
R= − + +
Sum of v1, v2 and v3
_
+
Difference Amplifier: Mathematic Operation
Use KCL and apply Rule 1:
1
1 4
0A outA v vv v
R R
−−+ =
R1
vout
+
-v1v2
R2
vA
vBSince vA = vB (Rule 2) and
32
2 3A B
Rv v v
R R
= = +
Substitute eq. (2) into eq. (1), we get
R3
R4
(1)
(2)
31 4 12
4 1 4 2 3 1
outv RR R vv
R R R R R R
+= − +
If R1 = R2 = R and R3 = R4 = Rf ( )2 1f
out
Rv v v
R= −
Difference of v1and v2
Differentiator and Integrator: Mathematic Operation
_
+
_
+
Differentiator
Integrator
R
vin
C
C
R
vin
i
vout
+
-
vout
+
-
i
i
i
vc+ -
outv iR= −
But Cdvi C
dt= in Cv v=and
inout
dvv RC
dt= −
out Cv v= −
But0
1( ) (0)
t
C Cv t idt vC
= +∫ inv iR=and
0
1(0)
t
out in Cv v dt vRC
= − +∫