Upload
others
View
10
Download
0
Embed Size (px)
Citation preview
AN EVALUATION OF REFLECTANCE-ABSORPTION INFRA RED SPECTROSCOPY FOR IN SITU INVESTIGATION OF ORGANIC COATINGS ON CORRODED METALS
S. C. BOYATZIS1*, V. ARGYROPOULOS1, A. SIATOU1, K. POLIKRETI1, D. CHARALAMBOUS1 and A. M. DOUVAS2
1 TEI of Athens, Dept. of Conservation of Antiquities & Works of Art, Egaleo 12210, Greece2 Institute of Microelectronics, NCSR “Demokritos”, 15310 Aghia Paraskevi, Greece
Kramers-Kronig Transformation (KKT)RAIR spectra carry both the transmittance and the specular reflectance component. When specular reflectance component dominates the spectrum, spectral lines appear with a derivative-like distortion
1. Typical application of protective coatings on metal surfaces generally involves non-ideal films of high, non-uniform thickness on non-ideal metal surfaces. RAIR spectra in these cases suffer from band distortions due to anomalous dispersion of refractive index and/or saturation due to high absorbances that fall out of the Beer law regime.
2. In films of low-medium thickness, both the transmitted and reflected beam reach the detector. Optimization involves use of the suitable polarizing angle and angle of incidence, which can eliminate the above problems. Use of p-polarized beam at Brewster’s angle, and film thicknesses where linearity in absorbance is obeyed, results in good spectra quality, in which all features carry entirely chemical information.
3. In cases of films so thick, that the transmitted beam is almost zero (because it is fully absorbed in the medium), or with metal surface is so rough that the transmitted beam is poorly reflected, the specular reflectance beam becomes important. In these cases the Kramers-Kronig transformation renders a spectral line that corresponds to an absorption spectrum.
Reflectance spectroscopy: Thin or thick films?
The Reflection-Absorption FTIR Technique
references
This question illustrates the geometrical aspects of a coating, basically controlled by the film application practice. The effectiveness of metal protection as well as the chemical information acquired spectroscopically is dependent on the film thickness, as this controls reactive species migration, light penetration, etc. The chemical spectroscopic information is averaged through the IR beam path in thin or thick films and reflects the behavior of polymeric coating chains and their functional groups with the metal and/or, if present, corrosion products.
Reflection-Absorption Infra Red Spectrometry (RAIRS) along with Attenuated Total Reflectance spectroscopy (ATR) have been widely used as characterization and monitoring tools that can be used to evaluate the effectiveness of coatings on metal[1]. Increasingly, these techniques are being applied to test coatings for cultural property made of metals, such as outdoor bronze monuments [2]. The advantage is that the RAIRS technique can carry out an in-depth characterization of layers from the bulk coating to the metal surface analyzing changes in the chemical bonds.
This work evaluates the application of RAIRS to investigate Paraloid®B-72 and Poligen® ES 91009® coatings of variable thickness on corroded iron and copper alloy coupons under the auspices of the EC project PROMET. Furthermore, it highlights how a simple FTIR system (PE Spectrum GX) can be adapted to carry out reflectance measurements coupling with a reflectance accessory (PE, 12.5˚ fixed angle) at a relatively low cost. Such adaptability makes it attractive for conservation research, since most Conservation facilities, such as ours have a simple FTIR system.
For an accurate interpretation of the spectra throughout the coating layers, the following properties are significant: thickness and uniformity of the coating, glossiness of its surface, and the nature of the metal surface. In conservation research, the effectiveness of coatings often needs to be evaluated on corroded metals which are far from being considered as ideal cases.
In thin films the transmitted beam carries basically chemical information through the small path of a few nanometers, as it records orientationalphenomena of polymeric chains towards the metal surface. Orientation is lost in thicker films, as chains tend to be more “random”, away from the metal surface.
Paraloid® B-72 on bronze and iron
μFTIR on corrosion products on iron coated with Paraloid +benzotriazole-based corrosion inhibitor
Poligen® ES910009 on bronze and iron
1. J. Umemura, Reflection–Absorption Spectroscopy of Thin Films on Metallic Substrates, Handbook Of Vibrational Spectroscopy, vol. 2, John Wiley & Sons Ltd, 2002,1016-1032.2. L. B. Brostoff, Coating strategies for the protection of outdoor bronze art and ornamentation, Ph.D. dissertation. Netherlands, 2003.3. V. Argyropoulos et al, Testing of a new wax coating Poligen ES91009 and corrosion inhibitor additives used for improving coatings for historic iron alloys, in: eds., C. Degrigny et al, METAL2007, (2007), 10-15.
Significant changes were observed in reflectance mode of Poligen films, most notably the sharp decrease of the original 1535 cm-1 carboxylate band , the increase of the 1700 cm-1 carboxylic acid band and the formation of a new band at 1605 cm-1 (only on bronze), possibly due to a copper-coordinated carboxylate phase. These features were correlated with those in model coupons incorporating the same coating materials with controlled film thicknesses and surface glossiness. Thus, our examination of infra-red RAIRS bands for the coated corroded coupons is entirely attributed to chemical changes.
Oriented acrylate
chains
random acrylate
chains
‘thin film’, (a few monolayers)
0 – 50 nm
Thick film: >50 nm
Surface roughness
Physical information can be dominant in IR reflectance in thick films, where anomalous dispersion of refractive index is important. This results in numerous distortions of the recorded chemical peaks, depending on the nature of coating medium (i.e. refractive index), the film thickness, surface smoothness, IR beam polarization and the angle of incidence. Predicting these distortions can lead to a spectrum with “pure” chemical information.
IR beam geometry during the reflectance-
absorption technique
Beam polarization and
reflectionPolarization of the
electrical IR component can be either parallel
(p-polarized) or perpendicular
(s-polarized) to the plane normal.
When angle of incidence equals the Brewster’sangle of the coating material, the p-polarized beam is totally transmitted through the material.
Three-layered system:
layer 1 (air, refr. index n1)layer 2 (coating, refr. index n2)layer 3 (metal, refr. index n3)
coating
metal
mirrors
3996,3 3000 2000 1500 1000 399,2
cm-1
%T
1699
1558
1466
1399 1171 720
7619239951085
1234600
539642
2915 28501695
1467
1253720
3083
1605
15481405
1381
1171
1083
927780
642598
543
Poligen/ bronze coupon / 0 min bakingPoligen/ bronze coupon / 5 min baking 50degPoligen/ bronze coupon / +5 min baking 60degPoligen/ bronze coupon / +5 min baking 70deg
Carboxylate peak diminished with baking/ water evaporation
Carboxylic acid peak increasing with baking /
water evaporation
new peak at 1605 cm-1
formed upon baking/ water evaporation
Poligen ES91009 on bronze
RAIR spectrum of Poligen ES91009 on Si wafer (13⁰ incidence angle). 3400(br) OH (carboxylic acid)2942(s), 2860(s) vC-H; 2649 (sh) dimeric carboxyls1706(s) C=O (acidic) 1556(s) C=O (carboxylate) 1468(m-s) δCH2, 1383(w) δCH3, 1407(w) δC-O-H1259 v αsC-O-H1179 v sC-O-C940(m, br) o.o.p. δ(O-H … O) hydrogen bonded 721 ρCH2 (long alkyl)
3996,3 3600 3200 2800 2400 2000 1800 1600 1400 1200 1000 800
cm-1
1. Original spectrum exhibiting specular derivative-like distortion2. Baseline-corrected spectrum3. KKT on original spectrum (in Transmittance mode)4. KKT on baseline-corrected spectrum (in Transmittance mode)
Effect of KKT on reflectance spectra (%T mode): carbonyl band of Paraloid B72 on Si wafer
Paraloid® B-72 Transmittance and RAIR spectra BASF Poligen® ES910009 RAIR spectra
KKT transformation on RAIR spectrum (Absorbance mode) of Paraloid on corroded bronze coupon. The result is a mostly distortionless absorbance spectrum.Comparison in IR peaks of coating and corrosion products may be done in the 1750 – 400 cm-1 region.
1800.6 1600 1400 1200 1000 800 600 395 .50 .49
0 .6
0 .8
1 .0
1 .2
1 .4
1 .6
1 .8
2 .0
2 .2
2 .4
2 .6
2 .8
3 .0
3 .19
cm-1
%T
1733
1473
1448
1386
1236
1166
1147
1025
964 858 749643
515 465
398
KKT-transformedPMTR_318
Original spectrum 74.C1PMTR_318
Comparison of the KKT- derived spectra (absorbance mode) from Paraloid on bronze coupons, brown area (artificially aged and reference). Major changes in Copper oxides and hydroxides (increase in 642, 515 and 465 cm-1 peaks.
1801.5 1600 1400 1200 1000 800 600 399 .4cm-1
A
598
1733
1448 1386
1236
1147
1025
858 749 643515
4651195
1169
600
515
465
642
1196
Artificially aged, KKT spectrum 74.C1 PMTR_318
Reference, KKT spectrum, 73.C1 PMTR_315
Difference spectrum
Introduction
Poligen ES91009 is sold commercially by BASF and reported to be a “water emulsion prepared from polyethylene wax”. FTIR spectra actually show hydrocarbon-like structure as well as carboxylic acid functionality. RAIR spectra of a film cast from water emulsion on a smooth surface (Si wafer, or metal: bronze, iron) further show carboxylate ion functionality depending upon the water content of the film.
4000,0 3000 2000 1500 1000 401,6
cm-1
%T
3738
29222853
2363
1700
1469
14081382
1259
939 721610
1173
2650
1556
761884
995
1082
(a)
(b)
CH2
CH2CH
CH2CH2
CH2CH2
CH2CH
CH2CH2
CH2CH2
C CO OH O O
-
Polyethylene-type backbone
carboxylatecarboxylic acid
Spectrum (a), recorded after evaporation of most water on cast film, shows both carboxylic acid (1700 cm-1) and carboxylate (1560 cm-1) functionalities as the structure is partly ionized. Spectrum (b) recorded one month after film casting ensures “complete” dryness of the film as carboxylate is mostly removed and the polymer is in ‘non-ionized’ state. This behavior is also confirmed by transmittance spectra of Poligen films on Si wafer.
y = 0,0197x + 3,7259R² = 0,9942
0,00
20,00
40,00
60,00
80,00
100,00
120,00
0 1000 2000 3000 4000 5000 6000
Abs
area
s (C
-H)
thickness, nm
Transmittance spectraPoligen/Si wafer: Abs(C-H) vs thickness before baking
Σειρά1
Γραμμική (Σειρά1)
The linear relationship of absorbances (measured as areas) of v(C-H) vs. film thickness assures that quantitative spectra for thick films (at least up to 5 μm) can lead to reliable Transmittance measurements for Poligen . This is crucial esp. for the carbonyl band of acrylics like Paraloid B72, which is a generally accepted marker for chemical alterations.
Reflectance spectra: angle of incidenceLinearity fails when reflectance IR bands exhibit distorted lineshapes and saturation (very high absorbance that does not obey Beer’s law), as is commonly the case in thick films. Band lineshapes can be optimized choosing Brewster’s angle (55⁰ for Paraloid) as incidence angle.
angle of incidence
4000.0 3000 2000 1500 1000 370.0cm-1
%T 2985
2954
1732
14751448
1387
1368
1236
1145
1025
970 860
76⁰ p-polarized
55⁰ p-polarized
13⁰
spin coated films on bronze; thickness 5.92 1 μm
Application of the Kramers –Kronig algorithm transforms the distorted spectral line (1, 2) line to an absorption spectrum of the film (3, 4).
12
3
4
Specular reflectance
Transmitted beam
In a typical reflection-absorption FTIR experiment the infra red beam impinges in the coating at a specified angle of incidence. A part of this beam is reflected onthe coating surface (specular reflectance). Another part is transmitted inside the film, accordingly reflects on the metal surface, re-transmits and finally, exits the film travelling to the detector (see figure). This is the absorption part of the RAIRS technique, recording the average information throughout the entire beam path. Both beam paths are detected within the same experiment: different physical information is recorded in each case, along with the chemical information of the involved materials.
conclusions
x
x
γ-FeOOH
3999,2 3000 2000 1500 1000 500 399,2
cm-1
%T
2921
2851
1698
1540
1466
1405
13811255
937 7202641
11712922
28511698
16041541
1466
1406
1381
1257938 720
1171
2918
2850
1698
14671259 11711405
1379719
1. RAIR spectrum of Poligen on iron2. RAIR spectrum of Poligen on bronze3. Transmission spectrum of Poligen on Si wafer (film thickness 1.48 μm)
film thickness 2.1 μm
2
1
3
Poligen ES91009. All spectra: incidence angle 55⁰
Paraloid B72 spectra are here recorded in micro-scale resolution. The quality of spectra varies between points. Corrosion products are also detected.
0,00
1,00
2,00
3,00
4,00
5,00
6,00
7,00
8,00
0 50 C, 5 min 50 C, 5 min+ 60 C 5 min
50 C 5 min+ 60 C 5 min+70C 5min
Abs
area
v(C
=O)
baking conditions
acidic COanionic COacidic+anionic CO
Poligen on polished bronze coupon. Film thickness 1.480 μm
TRANSMITTANCE on Si wafer: effect of baking on Abs of acidic and anionic carbonyls.
y = 0,0063x + 0,9781R² = 0,9995
y = 0,0092x + 5,791R² = 0,9962
0,00
20,00
40,00
60,00
80,00
100,00
0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10000
abso
rban
ce a
reas
thickness (nm)
Absorbances of v(C-H) and vC=O vs. film thickness
vC-H vCO Transmittance spectraThe carbonyl band and the methyl/methylene stretch are markers for chemical alterations of organic coatings.
Reliable absorbance measurements of bands v(C-H) and v(C=O) for Paraloid B72 can be taken for films up to 8-9 μm where linearity in absorbance vs. thickness is obeyed.