Perma Pure LLCPerma Pure LLC

LowLow--NOx Fallout:NOx Fallout:

Ammonia in Your Ammonia in Your

Perma Pure, 2009

Ammonia in Your Ammonia in Your

Sample SystemSample System

EPRI CEMS User Conference

May 15, 2009

NOx Fallout:NOx Fallout:

Ammonia in Your Ammonia in Your

Page 1

Ammonia in Your Ammonia in Your

Sample SystemSample System

EPRI CEMS User Conference

Page 2: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Sample Handling Sample Handling ––

Perma Pure, 2009

PROBE CONDITIONING SYSTEM ANALYZERS DATA AQUISITION

Critical Path for CEMSCritical Path for CEMS

Page 2

CONDITIONING SYSTEM ANALYZERS DATA AQUISITION

Page 3: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Perma Pure Perma Pure –– Gas Sample HandlingGas Sample Handling

3 Technologies:

• Dilution Probes

– Wet measurement

• Baldwin™ brand Thermoelectric Coolers

Perma Pure, 2009

• Baldwin™ brand Thermoelectric Coolers

– Water removal through flash condensation

• Nafion® permeation dryers and systems

– Water removal at the stack through unique membrane dryer technology

• Plus probes, filters, scrubbers, accessories

Gas Sample HandlingGas Sample Handling

Baldwin™ brand Thermoelectric Coolers

Page 3

Baldwin™ brand Thermoelectric Coolers

Water removal through flash condensation

Nafion® permeation dryers and systems

Water removal at the stack through unique membrane dryer technology

Plus probes, filters, scrubbers, accessories

Page 4: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Selective Catalytic Reduction (SCR)Selective Catalytic Reduction (SCR)

• Reduction of NOx by injection of ammonia

– NH3 and O2 react with NO and NO

H2O

Perma Pure, 2009

• Or, by injection of urea

– (NH2)2CO and O2 react with NO to form N

CO2

• Similar Processes hold for SNCR

Selective Catalytic Reduction (SCR)Selective Catalytic Reduction (SCR)

Reduction of NOx by injection of ammonia

react with NO and NO2 to form N2 and

Page 4

Or, by injection of urea

react with NO to form N2, H2O and

Similar Processes hold for SNCR

Page 5: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

SCR System for NOx Control in a BoilerSCR System for NOx Control in a Boiler

Perma Pure, 2009

SCR System for NOx Control in a BoilerSCR System for NOx Control in a Boiler

Page 5

Page 6: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

The Fallout: Ammonia SlipThe Fallout: Ammonia Slip

• When unreacted ammonia is released into the

flue gas stream. Causes:

– Added ammonia is never entirely consumed

– Catalyst temperatures are not in the optimum range

Perma Pure, 2009

– Catalyst temperatures are not in the optimum range

– Too much ammonia is used

• Can be mitigated by “slip catalyst”

The Fallout: Ammonia SlipThe Fallout: Ammonia Slip

When unreacted ammonia is released into the

flue gas stream. Causes:

Added ammonia is never entirely consumed

Catalyst temperatures are not in the optimum range

Page 6

Catalyst temperatures are not in the optimum range

Too much ammonia is used

Can be mitigated by “slip catalyst”

Page 7: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Ammonia SlipAmmonia Slip

• Range: 1 ppm up to 200+ ppm

• 5-20 ppm is a typical measurement

• In many cases, release of NH

even reported

Perma Pure, 2009

even reported

• Unfortunately, Analyzer Shop does not control

the rate of ammonia injection

• Process operators are often not concerned

about ammonia slip

Range: 1 ppm up to 200+ ppm

20 ppm is a typical measurement

In many cases, release of NH3 is not regulated or

Page 7

Unfortunately, Analyzer Shop does not control

the rate of ammonia injection

Process operators are often not concerned

Page 8: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Unwanted Consequence #1Unwanted Consequence #1

• NH3 is a catalyst in the formation of SO

SO2 and O2

• As vapors condense, you also get:

– SO + H O → H SO (Sulfuric acid)

Perma Pure, 2009

– SO3 + H2O → H2SO4 (Sulfuric acid)

– NH3 + H2SO4 → NH4HSO

(ammonium bisulfate

Unwanted Consequence #1Unwanted Consequence #1

is a catalyst in the formation of SO3 from

As vapors condense, you also get:

(Sulfuric acid)

Page 8

(Sulfuric acid)

HSO4

(ammonium bisulfate – ammonia salts)

Page 9: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Unwanted Consequence #2Unwanted Consequence #2

• Downstream formation of ammonium

bisulfate – “ammonia salts”

– due to sulfur content of the fuel source

– white and powdery when pure, it forms “large

Perma Pure, 2009

– white and powdery when pure, it forms “large

rhombic prisms” as it condenses

– collects and fouls solid surfaces in flues, probes,

filters, sample lines and inside analyzers

– forms in tight spaces, eg. O

mirror

Unwanted Consequence #2Unwanted Consequence #2

Downstream formation of ammonium

“ammonia salts”

due to sulfur content of the fuel source

white and powdery when pure, it forms “large

Page 9

white and powdery when pure, it forms “large

rhombic prisms” as it condenses

collects and fouls solid surfaces in flues, probes,

filters, sample lines and inside analyzers

forms in tight spaces, eg. O2 analyzer paramagnetic

Page 10: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Unwanted Consequence #3Unwanted Consequence #3

• Conversion of NH3 back to NO

– can happen even at lower temperatures

– can artificially inflate NOx readings

– when every PPM counts, can cause a failed RATA

Perma Pure, 2009

– when every PPM counts, can cause a failed RATA

– are you paying for NOx, but not NH

Unwanted Consequence #3Unwanted Consequence #3

back to NO

can happen even at lower temperatures

can artificially inflate NOx readings

when every PPM counts, can cause a failed RATA

Page 10

when every PPM counts, can cause a failed RATA

are you paying for NOx, but not NH3, emissions?

Page 11: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Ammonia ScrubberAmmonia Scrubber

• Protects analyzers and downstream components

– Eliminates ammonia salt deposition

• Very selective reagent

– Phosphoric acid removes only bases

Perma Pure, 2009

– Phosphoric acid removes only bases

• Very high corrosion resistance

– Housing: stainless steel, polysulfone

– Element: H3PO4 on inert ceramic

• Reliable and low maintenance

– No moving parts, only periodic refills

Ammonia ScrubberAmmonia Scrubber

Protects analyzers and downstream components

Eliminates ammonia salt deposition

Phosphoric acid removes only bases

Page 11

Phosphoric acid removes only bases

Very high corrosion resistance

stainless steel, polysulfone

on inert ceramic

Reliable and low maintenance

No moving parts, only periodic refills

Page 12: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Scrubber Flow SchematicScrubber Flow Schematic

Sample Inlet

Reagent:

Perma Pure, 2009

Reagent:

Phosphoric Acid

Anhydride

Inert

Ceramic

Burls

Scrubber Flow SchematicScrubber Flow Schematic

Filter Disk

Sample Outlet

Page 12

Flow

Diffuser

Page 13: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Hot, Wet ChemistryHot, Wet Chemistry

• Target Temperature 80

• Ammonia reacts in vapor form

• Water remains in vapor form

Perma Pure, 2009

– Otherwise, the bottom fills with liquid

– Heater blanket maintains 90

• Or, maintain 80-90°C in heated

enclosure

Hot, Wet ChemistryHot, Wet Chemistry

Target Temperature 80-90°C

Ammonia reacts in vapor form

Water remains in vapor form

Page 13

Otherwise, the bottom fills with liquid

Heater blanket maintains 90°C

C in heated

Page 14: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Hot Wet ChemistryHot Wet Chemistry

• 1st Stage Reaction: Phosphoric acid anhydride must be wetted

– Install upstream of cooler or Nafion® sample conditioning system

Perma Pure, 2009

– If downstream of cooler, spray with dewhen refilling reagent

• 2nd Stage Reaction: Formation of ammonium phosphate – salts

– Collect on the inert ceramic burls at bottom of scrubber

Hot Wet ChemistryHot Wet Chemistry

1st Stage Reaction: Phosphoric acid anhydride

Install upstream of cooler or Nafion® sample

Page 14

If downstream of cooler, spray with de-ionized water when refilling reagent

2nd Stage Reaction: Formation of ammonium

Collect on the inert ceramic burls at bottom of

Page 15: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Sample InstallationSample Installation

• Sample line feeds

first to ammonia

scrubber, then to

cooler

Perma Pure, 2009

cooler

Sample InstallationSample Installation

Page 15

Page 16: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

Spent Ammonia ScrubberSpent Ammonia Scrubber

Perma Pure, 2009

Note: Scrubber should NOT be mounted horizontally, as this one was

Spent Ammonia ScrubberSpent Ammonia Scrubber

Page 16

Note: Scrubber should NOT be mounted horizontally, as

Page 17: Ammonia in Sample System - Perma Pure · Ammonia Slip •Range: 1 ppm up to 200+ ppm •5-20 ppm is a typical measurement •In many cases, release of NH even reported Perma Pure,

MaintenanceMaintenance

Refill media every 40,000

liter/ppm/hours

Hand tighten mounting screw

Example:

Perma Pure, 2009

Example:

• Flow rate = 10 lpm

• Slip = 5 ppm

• Change Media every 33 days

Refill media every 40,000

liter/ppm/hours

Hand tighten mounting screw

Example:

Page 17

Example:

Flow rate = 10 lpm

Slip = 5 ppm

Change Media every 33 days