ÄKTA system Training Guide

Embed Size (px)

Citation preview

Page 1: ÄKTA system Training Guide

ÄKTAexplorer

amersham bioscience

amersham bioscience

amersham bioscience

amersham bioscience

ÄKTA system Training GuideExplorer 100 / 10

Page 2: ÄKTA system Training Guide

ÄKTAexplorer

amersham bioscience

amersham bioscience

amersham bioscience

amersham bioscienceBox

MonitorConductivitypH

UV Detector3 Wavelengths

Pumps

ColumnValves

Injector

SampleValve

Buffers

pH/cell

electrode

Sampleholder

Pump A

Buffer switchingvalve

OutletValve

ÄKTAexplorer

Page 3: ÄKTA system Training Guide

23 4 5

6781

B

4

A

Systempumps

Mixer

FractioncollectorFrac-950

Conductivitymonitor

pH detector

In-linefilter

Samplepump

Buffer selectvalve (V6)

Injectionvalve (V1)

Reverseflow valve (V7)

Outletvalve (V4)

Sample selectionvalve (V5)

Columnselectionvalve (V2)

Columnselectionvalve (V3)

3 Wavelengthdetector

23 4 5

6781

23 4 5

6781

23 4 5

6781

23 4 5

6781

24

7

3

165

24

7

3

165

UV

Cond

pH

ÄKTAexplorer 100

Airsensor

23 4 5

6781

B

4

A

Systempumps

Mixer

FractioncollectorFrac-950

Conductivitymonitor

pH detector

In-linefilter

Samplepump

Buffer selectvalve (V6)

Injectionvalve (V1)

Reverseflow valve (V7)

Outletvalve (V4)

Sample selectionvalve (V5)

Columnselectionvalve (V2)

Columnselectionvalve (V3)

3 Wavelengthdetector

23 4 5

6781

23 4 5

6781

23 4 5

6781

23 4 5

6781

24

7

3

165

24

7

3

165

UV

Cond

pH

ÄKTAexplorer 100 Air

Page 4: ÄKTA system Training Guide

23 4 5

6781

B

4

A

Systempumps

Mixer

FractioncollectorFrac-950

Conductivitymonitor

pH detector

In-linefilter

Buffer selectvalve (V6)

Injectionvalve (V1)

Outletvalve (V4)

Columnselectionvalve (V2)

Columnselectionvalve (V3)

3 Wavelengthdetector

23 4 5

6781

23 4 5

6781

23 4 5

6781

UV

Cond

pH

ÄKTAexplorer 10

23 4 5

6781

B

4

A

Systempumps

Mixer

FractioncollectorFrac-950

Conductivitymonitor

pH detector

In-linefilter

Samplepump

Buffer selectvalve (V6)

Injectionvalve (V1)

Outletvalve (V4)

Sample selectionvalve (V5)

Columnselectionvalve (V2)

Columnselectionvalve (V3)

3 Wavelengthdetector

23 4 5

6781

23 4 5

6781

23 4 5

6781

23 4 5

6781

UV

Cond

pH

ÄKTAexplorer 10 S

24

7

3

165

24

7

3

165

Page 5: ÄKTA system Training Guide

ÄKTAexplorer 10 XT

23 4 5

6781

B

4

A

Systempumps

Mixer

FractioncollectorFrac-950

Conductivitymonitor

pH detector

In-linefilter

Buffer selectvalve (V6)

Injectionvalve (V1)

Outletvalve (V4)

Columnselectionvalve (V2)

Columnselectionvalve (V3)

3 Wavelengthdetector

23 4 5

6781

23 4 5

6781

23 4 5

6781

UV

Cond

pH

24

7

3

165

Page 6: ÄKTA system Training Guide

PUMP P-900

Buffer Acid /Alkali WaterSalt

Proportion Valve

Pump A

Mixer

A2 A1 B2 B1

Pump B

GRADIENT MIXERÄ KTA 10 mls/min <6 mls/min

6 -14 mls/min0.6 ml mixer2 ml mixer

Ä KTA 100 mls/min <30 mls/min30 - 50 mls/min50 - 100 mls/min

2 mls mixer5 mls mixer12 mls mixer

Page 7: ÄKTA system Training Guide

PUMP P-900 Inlets

Water Salt

Buffer Select Valve IV-908

Pump A Pump B

Mixer

A2 A1B1 B2

SwitchingValves

12

34 5

67

8

A11A12

A13A14

A15A16

A17A18

Page 8: ÄKTA system Training Guide

PUMP P-900 Manually Priming Buffer Inlet Lines

Open

Draw Air out ofBuffer Line

Manual Flowpath

A1 or A2 BufferValveA

Manual Flowpath

B1 or B2 BufferValveB

Pump A Pump B

Common inlet A2 A1 B2 B1Common inlet

Page 9: ÄKTA system Training Guide

PUMP P-900 Back - Wash Lines

20% Ethanol : Water

Pump Back-WashLines

Wash OUT

WashIN

BufferIN

BufferOUT

High pressurePiston seal

Ball outletcheck valve

Ball inletcheck valve

Piston

Backwash pump diaphragm

IN-LINE one-wayvalve

Syringe fitting

Page 10: ÄKTA system Training Guide

LOAD (sample) INJECT (sample) LOAD (wash loop)

1

7

6

5

2

3

4

Pump

Column

Waste

PV-9081

2

34 5

6

7

8

Sample

1

7

6

5

2

3

4

Pump

Column

Waste

PV-9081

2

34 5

6

7

8

Sample

1

7

6

5

2

3

4

Pump

Column

Waste

PV-9081

2

34 5

6

7

8

Sample

ON-LINE FILTERS

P-901 ( 100 ml/min Pump )High Flow Fibre Filter

Polymer Support

P-903 ( 10 ml/min Pump )Low Flow HPLC Filter

Page 11: ÄKTA system Training Guide

1

7

6

5

2

3

4

PUMPLOADSAMPLE

Waste

COLUMN

Waste

LOAD

1

7

6

5

2

3

4

PUMP

Waste

COLUMN

Waste

INJECT

1

7

6

5

2

3

4

PUMP

Waste

COLUMN

Waste

WASTE

INJECTION VALVE INV- 907

Page 12: ÄKTA system Training Guide

SAMPLE LOADING METHOD

1

7

6

5

2

3

4

PUMPLOADSAMPLE

Waste

COLUMN

Waste Superloop

Luer LockAdapter

1

7

6

5

2

3

4

PUMPLOADSAMPLE

Waste

COLUMN

Waste

Needle Port

Always tighten by hand with needle inserted

Do Not Overtighten

Page 13: ÄKTA system Training Guide

Volume Max. allowed column pressure

Code No.

10 ml 4 Mpa 18-1113-8150 ml 4 Mpa 18-1113-82150 ml 2 Mpa 18-1023-85

SUPERLOOPS

10 ml

150 ml

50 ml

BUFFERFLUSH

SAMPLELOAD

SAMPLEINJECT

SAMPLEFLUSH

WASTEPUMP orSYRINGE

COLUMN COLUMN

Plunger

Buffer

Sample

EtchedGroove

Page 14: ÄKTA system Training Guide

Needle

NeedleAdapter

INV-907 Injection Valve - Injection Port

PEEK Valve Body

Valve Seal

Injection Needle Port

NeedleTube

Ferrule

Page 15: ÄKTA system Training Guide

INV-907 Injection Valve - Luer Adapter

PEEK Injection Valve BodyPEEK Injection Valve Body

PEEK Injection Luer AdaptorPEEK Injection Luer Adaptor

Valve SealValve Seal

Valve Channel Plate Valve Channel Plate

Page 16: ÄKTA system Training Guide

Flow

End Beginning

PEEK Tubing High Resistanceto Flow

Low Resistanceto Flow

Fluid Dynamics

Page 17: ÄKTA system Training Guide

Flow - Injecting a Loop

Fluid DynamicsGeneral

The way we imagine it Flows

The way it Actually Flows

Bubble Segmentation

• Partial Loading Loop = No more than 1/2 Loop Volume• Total Loading Loop = Inject 2X Loop Volume

50ul Load

60ul Load

100ul Load

200ul Load

Waste

100ul Loop

Page 18: ÄKTA system Training Guide

LOADING A SAMPLE LOOP

1

7

6

5

2

3

4

Sample

Pump

Column

Inject

Load

SamplepumpP-950

Injection valveINV-907

Waste

Page 19: ÄKTA system Training Guide

LOADING A SUPER LOOP

1

7

6

5

2

3

4

Sample

Pump

Column

Inject

Load

Waste

Sample

Buffer

Injection valveINV-907

Page 20: ÄKTA system Training Guide

MULTI-LOADING A SAMPLE LOOP

1

7

6

5

2

3

4

Pump

Inject

Load

Column

PV-908

Samples

1

2

3 4 56

7

8

I/O

Injection valveINV-907

WasteWash

Page 21: ÄKTA system Training Guide

MULTI-LOADING A SUPER LOOP

Waste

1

7

6

5

2

3

4

PumpColumn

Inject

Load

PV-9081

2

3 4 56

7

8

I/O

Wash

Samples

S3 S4 S5 S6 S8

Injection valveINV-907

Page 22: ÄKTA system Training Guide

DIRECT LOADING ON TO COLUMN

PumpColumn

Extra Tube

Inject

Load

1/16” or 1/8”Tubing

Sample

1

7

6

5

2

3

4

Waste

Page 23: ÄKTA system Training Guide

REVERSE FLOW VALVE ( INV-907 )

Column

Pump

Downward FlowPosition 1

Extra Tube

MONITOR

StopPlug

17

65

2

34

1

Column Extra Tube

MONITOR

StopPlug

2

Upward FlowPosition 2

Pump7

65

34

Page 24: ÄKTA system Training Guide

OUTLET VALVE PV-908

Default

To Fraction Collector

DETECTOR

F1

F2

F3F4

F5F6

F7

F8

I/O

Waste

F3 F5 F6

F7

F8F4

Page 25: ÄKTA system Training Guide

0

%B 10

0

50

Elu

ate

Fra

cti

on

atio

nP

eak

Fra

ctio

na

tio

nF

low

thro

ug

hF

ract

ion

ati

on

0%

B10

0%

B

Start

End

A1

A2

A3

……

……

……

….

Injectm

AU

0

500

250

Waste

Fra

ctio

n C

olle

ctio

n

Page 26: ÄKTA system Training Guide

HiTrap Connections

M6 Male

1/16” Female

1/16” Male

1/16” Female

M6 Male

M6 Female

1/16” Male

M6 male / 1/16” Female

18-1112-57

18-1112-55

Flow Cell Inlet

HiTrap Column

18-1112-58

M6 female / 1/16” male

1/16” Fingertight Fitting

Pheny l Sepharos e

Page 27: ÄKTA system Training Guide

Detector

Multiple HiTrap Arrangment ( Series )

Phenyl S

epharose

Detector

Phenyl S

epharoseP

henyl Sepharose

Detector

Phenyl S

epharoseP

henyl Sepharose

Phenyl S

epharose