AdV- VAC CTRL SYS/Cryo Interface 06Feb2013

AdV- VAC CTRL SYS/Cryo Interface 06Feb2013. Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature,

Embed Size (px)

Citation preview

Page 1: AdV- VAC CTRL SYS/Cryo Interface 06Feb2013. Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature,

AdV- VAC CTRL SYS/Cryo Interface

06Feb2013

Page 2: AdV- VAC CTRL SYS/Cryo Interface 06Feb2013. Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature,

Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature, controller of feeding valve, phase

separator electronics...

Page 3: AdV- VAC CTRL SYS/Cryo Interface 06Feb2013. Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature,

‘Control System’

• The general system that collect data from all vacuum equipment and operate them also by remote trough the supervisor SW. A ‘Vac Rack’ is foreseen for each crytrap, it will allow the operation of vacuum pumps and valves , and interface also with cryogenic equipment

• (And also interface with DAQ)

Page 4: AdV- VAC CTRL SYS/Cryo Interface 06Feb2013. Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature,

Some tasks related to cryo equipment

• Need to operate cryogenic devices (adjust the set point of LN2 level, feeding valves closing, start / stop temp sensors...) from Local .

• And also by remote (a few of them..., normally by the general vacuum SW)

• Need to provide some interlocks from the ‘cryo equipments (alarms) to the ‘rest’ of the system

• Need also to monitor ‘most’ signals from cryo equipment and record them (normally by the general vacuum SW)

Page 5: AdV- VAC CTRL SYS/Cryo Interface 06Feb2013. Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature,

Example of interlocks• Alarm of cryotrap over temperature • Set point temperature reached (<80K)• Alarm of high LN2 level • Alarm of low LN2 level• Set point LN2 reached • Phase separator alarm• ...

Page 6: AdV- VAC CTRL SYS/Cryo Interface 06Feb2013. Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature,

Signals monitored by remote

• Most of available ones, exampleItem quantity description Pc1 1 Pressure of cryostat, 4-20mA, TBC Pc2 1 Pressure of LN2 storage tank, 4-20mA T1-T8 1 8 pt100 temp sensors already converted to 4-20mA, TBC Vcryo 1 %opening of the refill valve, directly acquired by the

valve itself (Normally AI 4-20mA). Note: not the output signal transmitted by the PID regulator but the real valve status, for redundancy. Normally Closed, TBC

CL2 2 Input signal to PID regulator LN2 level inside cryostat (capacitive sensor), retransmitted by PID regulator or directly by its controller (rs232 or better I/O standard), TBC

CL1, TL1 4 LN2 level inside cryostat and storage tank (diff pressure sensors), normally AI 4-20mA

A1-A4 4 Alarms of LN2 system, DI, relays status to be acquired Spares 4 Additional DO, relays Spares 8 Additional AI, 4-20mA or 0-10V Spares 8 Additional DI, relays Spares 1 Additional rs232 device TBD

Page 7: AdV- VAC CTRL SYS/Cryo Interface 06Feb2013. Cryotrap cryogenics (simplified sketch) = the ‘electronic’ equipment needed to read LN2 level, temperature,

‘’Interface definition & task sharing’’

Discussion about the task sharing and to prepare the exchange of information