ACSM UpdatePhil Croteau, Wen Xu, Mike Cubison, John Jayne

9/23/2015 AMS Users Meeting 2015 1

Q-ACSM

ToF-ACSM

Q-ACSM and ToF-ACSM Systems

15-30 min data typical 10 min data typical

9/23/2015 AMS Users Meeting 2015 2

ToF-ACSM Sensitivity is ~5x Q-ACSM in same time

Particle Inlet (1 atm)

Thermal

Vaporization

&

Electron

Impact

Ionization

Aerodynamic Lens

40-1000 nm

Particle Beam

Generation

Particle

Composition

Quadrupole Aerosol Chemical Speciation Monitor

Laptop

Computer

RG

A

Pumps (x3)

No Sizing

Commercial grade

Mass SpectrometerPfeiffer Prisma

Slow scan rates, 0.2 sec/amu

9/23/2015 AMS Users Meeting 2015 3

ETOF

Mass Spec

Particle Inlet (1 atm)

Thermal

Vaporization

&

Electron

Impact

Ionization

Aerodynamic Lens

40-1000 nm

Particle Beam

Generation

Particle

Composition

ADQ114

4-Stage Split-Flow

No Sizing

Time-of-Flight Aerosol Chemical Speciation Monitor

9/23/2015 AMS Users Meeting 2015 4

RG

A

Sample

modeFilterFilter

mode

In ACSM, the difference airbeam signal is zero.

3-way Valve

Difference mass spectrum measured by automated valve switching

9/23/2015 AMS Users Meeting 2015 5

ACSM Around the World

Q-ACSM – 81 Systems

TOF-ACSM – 12 Systems

9/23/2015 AMS Users Meeting 2015 6

Requirements for PM2.5 ACSM

9/23/2015 AMS Users Meeting 2015 7

Issue Solution

Deliver particles to the lens Redesigned inlet system

Deliver particles to the vaporizer PM2.5 Lens

Detect particle mass Capture Vaporizer

ACSM Inlet for PM2.5

• Switch to straight path for filter valve• Eliminate bends in 1/8” tube• Minimize length of 1/8” Tube

Loss in 1/8” Tube

9/23/2015 AMS Users Meeting 2015 8

Standard Lens

9/23/2015 AMS Users Meeting 2015 9

PM2.5 comparison with standard lens

9/23/2015 AMS Users Meeting 2015 10

High pressure lens/PM2.5 lens/Standard lens comparison

9/23/2015 AMS Users Meeting 2015 11

Size-dependence of collection efficiency

9/23/2015 AMS Users Meeting 2015 12

Testing On-going in Nanjing, China

9/23/2015 AMS Users Meeting 2015 13

Coming soon…

• Q-ACSM to TOF-ACSM Upgrade• Q-ACSM Vacuum System with CTOF Mass spectrometer

• Currently putting together test/development system at ARI• Electronics and Software work in process

9/23/2015 AMS Users Meeting 2015 14

Thanks!