102

ACCU-GAGE - Honeywell Test and Measurement Sensors

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

PressureTransducers

Load & TorqueCells

Accelerometers

DisplacementTransducers

Instrumentation

ACCU-GAGEAG Series Digital Pressure IndicatorsModels AG400 & AG401

Sensotec, Inc.2080 Arlingate LaneColumbus, Ohio 43228 USACopyright © 2002 by Sensotec, Inc. all rights reservedTelephone: (614)850-5000FAX: (614)850-1111Toll Free: 1-800-848-6564E-mail: [email protected]: http://www.sensotec.com

ACCU-GAGE Instruction ManualModels AG400, AG401

Sensotec Document Number: 008-0629-00Rev. -: March, 2002

IMPORTANT It is recommended that you read “Getting Started Quickly” on page 13 thor-oughly before applying power to this unit.

CAUTION: The operator of this instrument is advised that if the equipmentis not used in a manner not specified in this manual, the protection pro-vided by the equipment may be impaired.

CAUTION: Only qualified, service-trained personnel who are aware of thehazards involved should remove the cover from the instrument or connectexternal wiring to the instrument.

Sensotec and the Sensotec logo are registered trademarks of Sensotec,Inc.

Printed in U.S.A.

Sensotec continually improves its products, and thus the information hereinis subject to change without notice.

CO

NT

EN

TS

Contents

Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Chapter 1Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

1.1 About This Manual . . . . . . . . . . . . . . . . . . . . . . . 91.1.1 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91.1.2 Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91.1.3 Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

1.2 Related Documents . . . . . . . . . . . . . . . . . . . . . . 111.3 What is the ACCU-GAGE? . . . . . . . . . . . . . . . . . 12

1.3.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121.3.2 Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121.3.3 Optional Virtual Channels . . . . . . . . . . . . . . . . . . . . . . 12

Chapter 2Getting Started Quickly . . . . . . . . . . . . . . . . . . . . . . 13

2.1 Pressure Connection . . . . . . . . . . . . . . . . . . . . . 132.2 Turn on the Instrument . . . . . . . . . . . . . . . . . . . . 132.3 The ACCU-GAGE Instrument is Ready for Use . 14

Chapter 3Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

3.1 Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . 153.2 INITIALIZE Mode . . . . . . . . . . . . . . . . . . . . . . . . 153.3 RUN Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

3.3.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163.3.2 [VALUE] button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163.3.3 [CLEAR] Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163.3.4 [CHANNEL] button . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173.3.5 [TARE] button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173.3.6 Indicator Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

3.4 ERROR mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 183.5 SETUP Menu mode . . . . . . . . . . . . . . . . . . . . . . 19

3.5.1 Available Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193.5.2 Entering and Exiting the SETUP Menu mode . . . . . . . 193.5.3 Moving Through SETUP Menus and Menu Items . . . . 193.5.4 Exiting the SETUP Menu mode . . . . . . . . . . . . . . . . . . 19

Chapter 4Chassis Models . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214.2 Chassis Specifications . . . . . . . . . . . . . . . . . . . . 224.3 Chassis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

4.3.1 External Arrangement . . . . . . . . . . . . . . . . . . . . . . . . . 234.3.2 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234.3.3 Panel Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

ACCU-GAGE Instruction Manual page 3

4.3.4 Rack Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244.3.5 Bench Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244.3.6 Case Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244.3.7 Internal Arrangement . . . . . . . . . . . . . . . . . . . . . . . . . . 254.3.8 Pressure Transducer Mounting . . . . . . . . . . . . . . . . . . 254.3.9 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254.3.10 Vehicle Power Option . . . . . . . . . . . . . . . . . . . . . . . . . 264.3.11 Fuse Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . 264.3.12 SETUP Protection Jumper . . . . . . . . . . . . . . . . . . . . . 26

Chapter 5System Connector . . . . . . . . . . . . . . . . . . . . . . . . . . 29

5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 295.2 System Connector Pinout . . . . . . . . . . . . . . . . . 305.3 Function Input Pins . . . . . . . . . . . . . . . . . . . . . . . 31

5.3.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 315.3.2 Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

5.4 Limit Output Pins . . . . . . . . . . . . . . . . . . . . . . . . 325.4.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 325.4.2 Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Chapter 6System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

6.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 356.2 Menu Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

6.2.1 SOFSOFSOFSOFTTTTWARE RWARE RWARE RWARE REEEEVISIONVISIONVISIONVISION Menu Item . . . . . . . . . . . . . . . . . . 356.2.2 CONCONCONCONFFFFIGURATIGURATIGURATIGURATIIIIONONONON Sub-Menu . . . . . . . . . . . . . . . . . . . . . . 356.2.3 DIADIADIADIAGGGGNOSTICNOSTICNOSTICNOSTICSSSS Sub-Menu . . . . . . . . . . . . . . . . . . . . . . . . 356.2.4 INSINSINSINSTTTTALL CHALL CHALL CHALL CHAAAANNELNNELNNELNNEL Menu Item . . . . . . . . . . . . . . . . . . . . 366.2.5 DELDELDELDELEEEETE CHATE CHATE CHATE CHANNNNNELNELNELNEL Menu Item . . . . . . . . . . . . . . . . . . . . . 376.2.6 DEFDEFDEFDEFAAAAULT CHULT CHULT CHULT CHAAAANNELNNELNNELNNEL Menu Item . . . . . . . . . . . . . . . . . . . . 37

Chapter 7Serial Communications . . . . . . . . . . . . . . . . . . . . . . 39

7.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 397.2 Rear Panel Connector . . . . . . . . . . . . . . . . . . . . 397.3 Communications Protocol . . . . . . . . . . . . . . . . . 40

7.3.1 RS-232 vs. RS-485 . . . . . . . . . . . . . . . . . . . . . . . . . . . 407.3.2 Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

7.4 Serial Com. Menu . . . . . . . . . . . . . . . . . . . . . . . . 417.4.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417.4.2 INTINTINTINTEEEERFACERFACERFACERFACE Menu Item . . . . . . . . . . . . . . . . . . . . . . . . . 417.4.3 ADDADDADDADDRRRRESSESSESSESS Menu Item . . . . . . . . . . . . . . . . . . . . . . . . . . . 417.4.4 BAUBAUBAUBAUDDDD RATERATERATERATE Menu Item . . . . . . . . . . . . . . . . . . . . . . . . . 417.4.5 AUTAUTAUTAUTOOOO LINE-LINE-LINE-LINE-FFFFEEDEEDEEDEED Menu Item . . . . . . . . . . . . . . . . . . . . . 417.4.6 TRATRATRATRANNNNSMIT TSMIT TSMIT TSMIT TEEEESTSTSTST Menu Item . . . . . . . . . . . . . . . . . . . . . . 417.4.7 LEALEALEALEAVVVVE MENUE MENUE MENUE MENU Menu Item . . . . . . . . . . . . . . . . . . . . . . . . 41

7.5 Pressure Input Channel Commands . . . . . . . . . 427.5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 427.5.2 Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

page 4 008-0629-00

CO

NT

EN

TS

Chapter 8Display Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

8.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 538.2 Menu Items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53

8.2.1 UPPUPPUPPUPPEEEER CHANR CHANR CHANR CHANNNNNELELELEL Menu Item . . . . . . . . . . . . . . . . . . . . . 538.2.2 LOWLOWLOWLOWEEEER CHANR CHANR CHANR CHANNNNNELELELEL Menu Item . . . . . . . . . . . . . . . . . . . . 538.2.3 LOWLOWLOWLOWEEEER MODER MODER MODER MODE Menu Item . . . . . . . . . . . . . . . . . . . . . . . . 538.2.4 DISDISDISDISPPPPLAY DILAY DILAY DILAY DISSSSABLEABLEABLEABLE Menu Item . . . . . . . . . . . . . . . . . . . 53

Chapter 9Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

9.1 Understanding Limits, Set Points and Return Points 559.2 Limit Operation . . . . . . . . . . . . . . . . . . . . . . . . . . 56

9.2.1 Actions When Activated . . . . . . . . . . . . . . . . . . . . . . . . 569.2.2 Scan Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

9.3 Limit Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 579.3.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 579.3.2 LIMLIMLIMLIMIIIIT.ENABT.ENABT.ENABT.ENABLLLLEEEE Menu Item . . . . . . . . . . . . . . . . . . . . . . 579.3.3 LIMLIMLIMLIMIIIIT.SETPT.SETPT.SETPT.SETPOOOOINTINTINTINT Menu Item . . . . . . . . . . . . . . . . . . . . 579.3.4 LIMLIMLIMLIMIIIIT.RETUT.RETUT.RETUT.RETURRRRN PNTN PNTN PNTN PNT Menu Item . . . . . . . . . . . . . . . . . . . 579.3.5 LIMLIMLIMLIMIIIIT.ENERT.ENERT.ENERT.ENERGGGGIZEIZEIZEIZE Menu Item . . . . . . . . . . . . . . . . . . . . 579.3.6 LIMLIMLIMLIMIIIIT.LATCT.LATCT.LATCT.LATCHHHHINGINGINGING Menu Item . . . . . . . . . . . . . . . . . . . . 589.3.7 LIMLIMLIMLIMIIIIT.CHANT.CHANT.CHANT.CHANNNNNELELELEL Menu Item . . . . . . . . . . . . . . . . . . . . . 599.3.8 LIMLIMLIMLIMIIIIT.SOURT.SOURT.SOURT.SOURCCCCEEEE Menu Item . . . . . . . . . . . . . . . . . . . . . . 599.3.9 LEALEALEALEAVVVVE MENUE MENUE MENUE MENU Menu Item . . . . . . . . . . . . . . . . . . . . . . . . 59

Chapter 10Pressure Input Channel . . . . . . . . . . . . . . . . . . . . . . 61

10.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6110.2 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6210.3 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

10.3.1 Calibration Considerations . . . . . . . . . . . . . . . . . . . . . 6310.3.2 Required Pressures . . . . . . . . . . . . . . . . . . . . . . . . . . 6310.3.3 Calibration Units of Measure . . . . . . . . . . . . . . . . . . . 6310.3.4 Traceability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6310.3.5 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 63

10.4 Pressure Input Channel Specifications . . . . . . . 6410.5 Channel Menu . . . . . . . . . . . . . . . . . . . . . . . . . . 65

10.5.1 OPEROPEROPEROPERAAAATIONTIONTIONTION Sub-Menu . . . . . . . . . . . . . . . . . . . . . . . . 6510.5.2 DISPDISPDISPDISPLLLLAY UNIAY UNIAY UNIAY UNITTTTSSSS Menu Item . . . . . . . . . . . . . . . . . . . . 6610.5.3 AUXnAUXnAUXnAUXn FUNCTIOFUNCTIOFUNCTIOFUNCTIONNNN Menu Items . . . . . . . . . . . . . . . . . . . 6610.5.4 CALICALICALICALIBBBBRATIONRATIONRATIONRATION TYPETYPETYPETYPE Menu Item . . . . . . . . . . . . . . . . . . 6710.5.5 CALICALICALICALIBBBBRATIONRATIONRATIONRATION DATADATADATADATA Sub-Menu . . . . . . . . . . . . . . . . . . 6810.5.6 CALICALICALICALIBBBBRATERATERATERATE Menu Item . . . . . . . . . . . . . . . . . . . . . . . . 6810.5.7 DACDACDACDAC SSSSETUPETUPETUPETUP Sub-Menu . . . . . . . . . . . . . . . . . . . . . . . . 7010.5.8 DIAGDIAGDIAGDIAGNNNNOSTICSOSTICSOSTICSOSTICS Sub-Menu . . . . . . . . . . . . . . . . . . . . . . 71

10.6 Analog Output Configuration . . . . . . . . . . . . . . . 7210.6.1 Identifying the Output Type . . . . . . . . . . . . . . . . . . . . 7210.6.2 Channel Menu Items . . . . . . . . . . . . . . . . . . . . . . . . . 7210.6.3 Output Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72

10.7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . 7310.7.1 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7310.7.2 Common Problems and Solutions . . . . . . . . . . . . . . . 73

ACCU-GAGE Instruction Manual page 5

Chapter 11Split Display Virtual Channel . . . . . . . . . . . . . . . . . . 75

11.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7511.2 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7511.3 Setup Procedure . . . . . . . . . . . . . . . . . . . . . . . 7511.4 Channel Menu . . . . . . . . . . . . . . . . . . . . . . . . . 76

Chapter 12Mathematics Virtual Channel . . . . . . . . . . . . . . . . . 77

12.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7712.2 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7812.3 Setup Procedure . . . . . . . . . . . . . . . . . . . . . . . 7912.4 Channel Menu . . . . . . . . . . . . . . . . . . . . . . . . . 80

12.4.1 SENSSENSSENSSENSOOOOCODE PCODE PCODE PCODE P////NNNN Menu Item . . . . . . . . . . . . . . . . . . . . 8012.4.2 USERUSERUSERUSER VALUESVALUESVALUESVALUES Sub-Menu . . . . . . . . . . . . . . . . . . . . . . . 8012.4.3 VIEWVIEWVIEWVIEW SENSOCOSENSOCOSENSOCOSENSOCODEDEDEDE Menu Item . . . . . . . . . . . . . . . . . . . 8012.4.4 VIEWVIEWVIEWVIEW COMMANDCOMMANDCOMMANDCOMMANDSSSS Sub-Menu . . . . . . . . . . . . . . . . . . . . 8012.4.5 DISPDISPDISPDISPLLLLAY SETAY SETAY SETAY SETUUUUPPPP Sub-Menu . . . . . . . . . . . . . . . . . . . . . 8112.4.6 POWEPOWEPOWEPOWERRRR-ON SO-ON SO-ON SO-ON SOUUUURCERCERCERCE Menu Item . . . . . . . . . . . . . . . . . . . 82

12.5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . 83

Chapter 13Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

13.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8513.2 Error Message List . . . . . . . . . . . . . . . . . . . . . . 86

Chapter 14Setup Menu Reference . . . . . . . . . . . . . . . . . . . . . . . 91

14.1 Navigation instructions . . . . . . . . . . . . . . . . . . . 91

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

Warranty/Repair Policy . . . . . . . . . . . . . . . . . . . . . . 101

page 6 008-0629-00

CO

NT

EN

TS

ACCU-GAGE Instruction Manual page 7

page 8 008-0629-00

INT

RO

DU

CT

ION

1

Chapter 1Introduction

1.1 About This Manual

1.1.1 Scope This manual will explain the setup, features and operation of Sensotec’s 3rd gen-eration ACCU-GAGE digital pressure indicators. The AG Series includes themodels AG400 and AG401 instruments.

Further information about customer specific programming and setup will beexplained on the Customer Information Sheet that is provided with every instru-ment.

1.1.2 Conventions This manual uses the following conventions to present information:

1.1.3 Organization Chapter 1, “Introduction”, offers general information about the instrument andthis instruction manual.

Chapter 2, “Getting Started Quickly”, provides an overview of how to setup anduse the instrument.

Chapter 3, “Operating Modes”, discusses the significant features of the ACCU-GAGE and operation procedures when the instrument is in the INITIALIZE, RUN,ERROR or SETUP modes.

[TEXT IN BRACKETS] The label of a front panel button.

DDDDIIIISSSSPPPPLLLLAAAAYYYY Text that appears on the display, such as errormessages or menu items.

-> Indicates that what follows is an item from a sub-menu, such as SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> DDDDIIIIAAAAGGGGNNNNOOOOSSSSTTTTIIIICCCCSSSS.

DATA Serial communications commands sent to (orreplies from) the instrument

↵ The carriage-return character, ASCII code decimal13

Note

Notes contain important information set off fromthe text.

A stop sign highlights procedures or informationnecessary to avoid damage to equipment, dam-age to software, loss of data, loss of calibration, orinvalid test results.

The caution symbol alerts that a specific proce-dure or practice which, is not followed correctly,could cause serious personal injury.

ACCU-GAGE Instruction Manual page 9

Chapter 4, “Chassis Models”, gives information relating to the hardware chassissuch as panel and rack mounting.

Chapter 5, “System Connector”, contains information about wiring to the 25-pinSystem connector to access the Limit Outputs, Function Inputs and serial communica-tions.

Chapter 6, “System Menu”, discusses the System menu which allows you to exam-ine and change settings that affect the operation of the chassis.

Chapter 7, “Serial Communications”, briefly describes RS-232 and RS-485 com-munications. It also shows how to use the SSSSEEEERRRRIIIIAAAAL CL CL CL COOOOMMMM.... Menu to examine the settingsand test the RS-232 or RS-485 communications.

Chapter 8, “Display Menu”, describes the Display Menu which allows you to changewhat is displayed on the lower line of the instrument.

Chapter 9, “Limits”, discusses how the limits operate and describes how to altertheir settings via the Limit Menus.

Chapter 10, “Pressure Input Channel”, explains how to wire, configure, operate andcalibrate the Pressure Input channel.

Chapter 11, “Split Display Virtual Channel”, shows how you can display two valuesfrom any of the channels in the instrument at the same time.

Chapter 12, “Mathematics Virtual Channel”, describes the flexibility of customer-specific SensoCode programming.

Chapter 13, “Error Messages”, lists error messages that the instrument may display,describes their causes and, where possible, suggests solutions.

Chapter 14, “Setup Menu Reference”, is a list of all SETUP menus and a cross-ref-erence to related information in this instruction manual.

page 10 008-0629-00

INT

RO

DU

CT

ION

1

1.2 Related Documents

Certificate ofCalibration

Every ACCU-GAGE instrument is shipped with a Certificate of Calibration, which is aquality record that shows the date and results of the instrument’s factory calibration.

Customer InformationSheet

Every instrument is shipped with a Customer Information Sheet which documents theconfiguration of the instrument when it left the factory. Such information includes:

• Sensotec part number,• date of manufacture,• list of all installed channels and their setup information,• customer specific SensoCode programming of Mathematics Virtual channels and

operation notes.

CommunicationsGuide

The “SC Series Communications Guide”, Sensotec p/n 008-0610-00, describes indetail how to communicate with an SC Series instrument using RS-232 and RS-485.Since the ACCU-GAGE Models AG400 and AG401 are very similar to the SC Series,the wiring diagrams, sample programs, and command descriptions contained in thisdocument are applicable to ACCU-GAGE instruments.

A printed copy of this document is available for order, or you may download it fromSensotec’s web site, www.sensotec.com.

ACCU-GAGE Instruction Manual page 11

1.3 What is the ACCU-GAGE?

Sensotec ACCU-GAGE Digital Pressure Indicators combine a pressure transducerand a digital signal-conditioner into a single enclosure. These high-accuracy instru-ment are designed for use in both laboratory and industrial applications. A wide vari-ety of pressure ranges are available. The operation of an ACCU-GAGE instrument isbased on digital technology to provide improved accuracy, superior ease of setup, anda wealth of features.

1.3.1 Features The main features of ACCU-GAGE instruments are:

• Four alarm limits with open-collector outputs and versatile setup• Wide variety of pressure ranges from 15 to 60,000 PSI.• Up to 50,000 part resolution• Field selectable display and calibration engineering units• RS-232 communications standard (RS-485 optional)• Local or remote setup using the RS-232 or RS-485 port• Push-button on/off tare feature• Peak/Valley value capture

1.3.2 Models Several models are available with various accuracies and available pressure ranges.See “Chassis Specifications” on page 22 for details of each model.

1.3.3 OptionalVirtual Channels

The pressure transducer is connected to a hardware Input channel of the digital signalconditioner. This channel is referred to as “CHANNEL 01”. Virtual channels are soft-ware based devices that occupy a channel number, but not a physical slot, in theinstrument.

Several different types of optional Virtual channels are available in order to customizean ACCU-GAGE instrument for a particular application.

• Split Display Virtual channels allow the displaying of any two channel’s track, peakor valley values at the same time.

• Mathematics Virtual channels run small programs written in an interpretive lan-guage called SensoCode. This provides great flexibility which allows the ACCU-GAGE to do many jobs which otherwise requires a personal computer or PLC.

page 12 008-0629-00

GE

TT

ING

ST

AR

TE

DQ

UIC

KLY

2

Chapter 2Getting Started Quickly

2.1 Pressure Connection

Before connecting to the instrument’s pressure port, verify that the correctthreads are being used.

On 1/4” NPT pressure fittings, the use of a pipe thread sealant (such as Teflon®tape) is recommended to ensure a leak-tight seal.

Always use a backing wrench when connecting to the instrument’s pres-sure port.

2.2 Turn on the Instrument

ACCU-GAGE instrument are calibrated at the factory and need no further calibra-tion prior to initial use.

Connect the power cord between the instrument power source and the instru-ment, and turn the On/Off switch on the back of the instrument to the On position.

The instrument enters its INITIALIZE mode that lasts a few seconds per channel.

When the instrument enters its normal operating mode (RUN mode), you will seethe following format on the front panel display:

1111•••• 00000. P00000. P00000. P00000. PSISISISIGGGG """"

where:

• Channel number: “1” is the channel number.• Value type: The next character indicates the nature of the following value.

A blank character indicates the tracking value. The instrument is “tracking” thesignal, continuously updating the display in response to the input pressure.

Table 1: Pressure Port Threads

Pressure Range Pressure Reference Pressure Port Threads

< 1,500 PSI Gage, Absolute 1/4-18 NPT male

1,500 to 10,000 PSI Gage, Absolute 1/4-18 NPT female

15,000 PSI and above Gage, Absolute Autoclave AE F250-C(9/16-18 UNF female with

sealing cone)

ACCU-GAGE Instruction Manual page 13

A “••••" character indicates the peak value (highest value seen since the [CLEAR]button was last pressed).A “––––" character indicates the valley value (lowest value seen since the [CLEAR]button was last pressed).

• Data value: “00000.” displays the pressure value in engineering units.• Units label: “PSIG” indicates the engineering units being used (up to 4 charac-

ters).• Tare indicator: A “""""“ symbol indicates tare is off; a “ËËËË“ symbol indicates tare is on.

2.3 The ACCU-GAGE Instrument is R eady for Use

See “RUN Mode” on pag e16 for information on how to operate the instrument while itis in the RUN mode.

Apply some test stimulus on the transducer to observe changes in the display.

page 14 008-0629-00

OP

ER

AT

ING

MO

DE

S3

Chapter 3Operating Modes

3.1 Operating M odes

ACCU-GAGE instruments have four modes of operation:

• INITIALIZE, to test the digital signal conditioner and transducer upon powerup

• RUN, normal operation• ERROR, which indicates that an abnormal situation has occurred that

stopped the operation of the instrument• SETUP, a menu which allows setup and calibration of the chassis and its

channels.Each of these will be described in this chapter.

3.2 INITIALIZE Mode

When the instrument is powered up or otherwise reset, it enters the INITIALIZEmode. As the instrument enters this mode, all segments of the display and allfront panel indicator lights (if available) illuminate momentarily. Next, each chan-nel in the instrument is checked for proper operation.

If a problem is detected, the instrument may enter the ERROR mode.

ACCU-GAGE Instruction Manual page 15

3.3 RUN Mode

After the INITIALIZE mode finishes, the instrument enters the RUN mode, its normalmode of operation.

3.3.1 Display

Upper line The display will show a channel number on the far left, followed by the channel’s oper-ation messages.

For example, Channel 01, the Pressure Input Channel, will use the format below:

1111•••• 00000000000000000000.... PSIPSIPSIPSIGGGG """"

where:

• Channel number: “1” is the channel number.• Value type: The next character indicates the nature of the following value.

A blank character indicates the display is being continuously updated by “tracking”the input signal to the channel.A “••••" character indicates the peak value (highest value seen since the [CLEAR]button was last pressed).A “––––" character indicates the valley value (lowest value seen since the [CLEAR]button was last pressed).

• Data value: “00000.” displays the value from the transducer in engineering units.• Units label: “PSIG” indicates the engineering units being used (up to 4 charac-

ters).• Tare indicator: A “""""“ symbol indicates tare is off; a “ËËËË“ symbol indicates tare is on.

Lower Line The content of the display’s lower line is selected with the “DDDDIIIISSSSPPPPLLLLAAAAY MY MY MY MEEEENNNNUUUU -> LLLLOOOOWWWWEEEERRRR MMMMODODODODEEEE”menu item as either displaying an additional Virtual channel or indicating the status ofLimits. See “Indicator Lights” on page17.

3.3.2 [VALUE]button

After the channel number, the next characters indicate which data value for the dis-played channel is shown. There are three (possibly four) data values available fromeach channel:

• (blank), tracking data value• “••••", peak data value, (highest value since the peak/valley detector was cleared)• “––––", valley data value, (lowest value since the peak/valley detector was cleared)• “Û", percentage of the Analog-to-Digital converter’s full-scale digitizing capability.

This data value is only available when the channel’s “DDDDIIIIAAAAGGGGNNNNOOOOSSSSTTTTIIIICCCCSSSS->DDDDIIIISSSSPPPPLLLLAAAAYYYY ADADADADCCCC”menu item is set to “ON”.

Pressing and releasing the [VALUE] button cycles though the available sources fordata values for the displayed channel.

3.3.3 [CLEAR]Button

Pressing and releasing the [CLEAR] button will reset the peak and valley values of thechannel being monitored by the display to the track value. Additionally, any limits inthe instrument that are “latched” will be reset.

To clear the peak and valley values of all channels (including optionally installed Vir-tual channels) simultaneously, use the System connector’s Function Input #2 pin. See“System Connector” on page29 for details.

To clear the peak and valley values of Channel 01, the Pressure Input Channel,regardless of which channel is being monitored by the display, use the channel’sAUX1 or AUX2 control pins on its connector. See “Wiring” on p age62.

page 16 008-0629-00

OP

ER

AT

ING

MO

DE

S3

3.3.4 [CHANNEL]button

If the configuration of the instrument contains more than one channel, the leftmost characters of the display indicates which channel the display is monitoring.Pressing and releasing the [CHANNEL] button will cause the next channel to bedisplayed.

The [CHANNEL] button cannot be used to change which channel the lower line ismonitoring. That is selected with the “DDDDIIIISSSSPPPPLLLLAAAAYYYY MENMENMENMENUUUU -> LLLLOOOOWWWWEEEERRRR CHANNECHANNECHANNECHANNELLLL” menu item.

3.3.5 [TARE] button To reset the channel’s display to zero, press the [TARE] button while in RUNmode. To restore the tare value, press the [TARE] button again. The “Tare” indica-tor will illuminate when tare is on.

To tare all channels (including optionally installed Virtual channels) simulta-neously, use the System connector’s Function Input #1 pin. See “System Connec-tor” on page29.

To tare the Pressure Input Channel regardless of which channel is being moni-tored by the display, use the channel’s AUX1 or AUX2 control pins on its connec-tor. See “Wiring” on pag e62.

3.3.6 Indicator Lights A “""""“ symbol on the right side of a channel’s display indicates tare is off for thatchannel; a “ËËËË“ symbol indicates tare is on.

If the lower line of the display is configured with the “DDDDIIIISSSSPPPPLLLLAAAAY MY MY MY MEEEENNNNUUUU -> LLLLOWOWOWOWEEEERRRR MMMMODODODODEEEE”menu item as “LLLLIIIIMMMMIIIIT 01T 01T 01T 01----00004444”, then the lower line will monitor the status of Limits 1, 2,3 and 4. A “""""“ symbol indicates a limit is deactivated; a “ËËËË“ symbol indicates a limitis activated. For example:

LLLL1111"""" LLLL2222"""" LLLL3333ËËËË LLLL4444ËËËË

indicates that Limit 1 and Limit 2 are deactivated and Limit 3 and Limit 4 are acti-vated. See “LOWER MODE Menu Item” on page53.

These indicators may be overridden by the operation of a Mathematics Virtualchannel in special applications.

ACCU-GAGE Instruction Manual page 17

3.4 ERROR mode

The instrument enters the ERROR mode when a critical error occurs that prevents theinstrument from operating. The display alternates between displaying a two-digit codein the form “ERROR xx ON CH.yy” and a short description of the error. The first twodigits “xx” hold the error code. The last two digits, “yy” is channel number that causedthe error. For example, “EEEERRRRRRRROOOORRRR 66660000 OOOONNNN CCCCHHHH....00001111” indicates that error number 60 occurred onchannel 1.

While the instrument is in the ERROR mode, no other operations are taking place.See “Error Message List” on page 86 for a list of error codes and their probablecauses.

page 18 008-0629-00

OP

ER

AT

ING

MO

DE

S3

3.5 SETUP Menu mode

The SETUP Menu mode is used to display or change the settings that control theoperation of the instrument.

3.5.1 AvailableMenus

Each major function of the instrument has its own SETUP Menu. See “Setup MenuReference” on page 91.

3.5.2 Entering andExiting the SETUP

Menu mode

To enter the SETUP Menu mode, press and hold the [UP] and [DOWN] buttons at thesame time until you see “SSSSEEEETTTTUUUUPPPP” on the display. When you release the buttons, you willsee the first SETUP Menu item.

To leave SETUP mode and return to the RUN mode, press [EXIT].

3.5.3 MovingThrough SETUP

Menus and MenuItems

The following table shows how to move though menus and menu items, and how todisplay or change the setting of a particular item.

To change the setting of a menu item:Press [ENTER] to display the current setting, which is preceded by the ‘****’ character.Press [UP] or [DOWN] until you see the setting you want. To abandon changes,press [EXIT]. To accept the new setting and move to the next item, press [ENTER].

3.5.4 Exiting theSETUP Menu mode

To leave SETUP mode and return to the RUN mode, press [EXIT] at any time.

Table 3-1: Navigating the Menus

Button Action

[UP] or [DOWN] Scroll through available menus, sub-menus or menu items.

[ENTER] Select a menu, sub-menu or menu item

[EXIT] If changing a setting: abandons changes to a menu itemOtherwise: leaves the SETUP mode

ACCU-GAGE Instruction Manual page 19

page 20 008-0629-00

CH

AS

SIS

MO

DE

LS4

Chapter 4Chassis Models

4.1 Introduction

Sensotec ACCU-GAGE instruments are available in several different models. Allinclude one built-in Pressure Input channel and can be ordered with optional Vir-tual channels for application specific applications.

The Pressure Input channel is a printed circuit board that occupies a physical slotinside the instrument’s chassis. A 12-pin connector are located on the rear panelto connect to the Pressure Input. This channel is identified as channel 01.

A Virtual channel exists in software only; it does not occupy a physical slot insideof the instrument’s chassis. There is no rear-panel connector for a Virtual chan-nel. However, a Virtual channel does require a channel number.

ACCU-GAGE Instruction Manual page 21

4.2 Chassis Specifications

ModelAG400

ModelAG401

PRESSUREMEASUREMENT

Linearity and Hysteresis 0.05% max. 0.10% max.

Temperature Effect 0.004% / ºF (0.0022% / ºC)over operating range

Pressure Port see “Pressure Connection” onpage 13

PHYSICAL

# Pressure Input channels 1 1

# Virtual channels (max.) 8 8

Case Material Aluminum Aluminum

Form factor 3/8 DIN 3/8 DIN

Mounting bench, panel or rack

DISPLAY

# characters/line 16 16

# lines/display 2 2

Display type Vacuum Fluorescent

ENVIRONMENTAL

Storage temp. -30º C to 90º C -30º C to 90º C

Operating temp. 15º C to 40º C 15º C to 40º C

Relative Humidity 80% max. for temperatures up to 31ºC; decreasing linearly to 50% max. at

40º C

Other indoor use at altitudes up to 2000 m;Pollution Degree 2; Overvoltage Cat-

egory II

SPECIAL FEATURES

Limits Quantity 4 4

Digital, isolated controlinputs

4 4

POWER

Standard AC powered(automatic selection)

100 to 230 VAC,47 to 63Hz

100 to 230 VAC,47 to 63Hz

Optional Vehicle Powered 11-28 VDC, 20W 11-28VDC, 20W

page 22 008-0629-00

CH

AS

SIS

MO

DE

LS4

4.3 Chassis

4.3.1 ExternalArrangement

The external arrangement of AC powered Models AG400 and AG401 is givenbelow.

Figure 4-1: External Arrangement of AC powered Models AG400 and AG401

4.3.2 Rear Panel The pinout for the 25-pin System connector is provided later in this chapter. Thepinouts for the individual channels are located in the chapter for that channel.

4.3.3 Panel Mounting The panel cutout size conforms to the 3/8 DIN standard. Panel-Mounting Jacks(order code AA928) are available that slide into two slots at the sides of the instru-ment. Use the following procedure to mount a Model AG400 or AG401 into apanel.

0123456789ABCDEF

0123456789ABCDEF

ACCU-GAGE Instruction Manual page 23

Figure 4-2: Panel Cutout Drawing for Models AG400 and AG401 (not to scale)

• In the panel or rack, cut a hole as shown above. The panel may be up to 1/4” inthickness.

• Use an 0.062" Allen wrench to remove two set screws that hold the Panel-Mount-ing Jacks to the case. To do this, insert the wrench into the side slots at the rear.Remove the screws completely.

• Remove the Panel-Mounting Jacks by sliding them toward the rear. If the jacksdon't slide easily, tap them gently.

• Put the instrument through the hole in the panel.• Reinsert the Panel-Mounting Jacks into the slide slots. Slide them as far toward

the panel as possible.• Reinsert the setscrews and tighten them. This will force the Panel-Mounting Jacks

toward the rear side of the panel, drawing the instrument tightly into place.

4.3.4 RackMounting

A Rack Mounting Kit (order code AA934) is available for mounting a single ModelAG400 or AG401 into a 19”, 2U rack. It includes the Panel-Mounting Jacks describedabove as well as the 19” rack panel.

4.3.5 BenchMounting

If bench mounting the instrument, you may want the optional carrying handle/benchstand (order code AA926).

4.3.6 CaseRemoval

Be sure to remove the power cord from the power source and disconnect the pressureport before attempting to remove the instrument from its case.

Use a #0 Phillips screwdriver on the black screws to avoid damaging them.

Step 1: Remove the two Phillips-head screws that secure the front panel.

Step 2: Remove the front panel and disconnect its two connecting cables.

Step 3: Remove the four black Phillips-head machine screws that secure the backpanel of the instrument to the case. These screws are located on the rear of the

page 24 008-0629-00

CH

AS

SIS

MO

DE

LS4

case, one at each of the four corners. Do not remove the cable shield con-nection screw in the center or the two screws adjacent to the pressure port.

Step 4: The circuit boards and pressure transducer will slide out of the rear of thecase as a unit.

Installation is the reverse of removal.

4.3.7 InternalArrangement

The figure below shows the names and locations of the printed circuit boards.

Figure 4-3: Internal Arrangement of Models AG400 and AG401

Below is a description of each printed circuit board.

• The SensoBus Backplane Board serves as the connection between allboards in the instrument.

• The Front Panel/Display Board Assembly contains the display and all frontpanel controls.

• The Power Supply Board contains the +15V, -15V and +5V power supplies.• The Microprocessor Board contains the microprocessor, ROM software

storage chip, and the System connector.• The Pressure Input Channel Board contains the signal-conditioning circuitry

for the pressure transducer.

4.3.8 PressureTransducer Mounting

If the pressure transducer becomes loose in its mounting, you can re-secure it bytightening the set screws on the pressure transducer mounting flange.

Step 1: Remove the case according to “Case Removal” on page24.

Step 2: Locate the three set screws that secure the pressure transducer to itsmounting flange.

Step 3: Re-tighten the set screws with a 0.050” Allen wrench.

Step 4: Case installation is the reverse of removal.

4.3.9 Cleaning Turn off the instrument and unplug all connectors. Use a soft cloth or tissue and amild cleaner. Do not use liquid or aerosol cleaners. Do not allow any cleaner

Front Panel/Display BoardAssembly

SensoBus BackplaneBoard

Power SupplyBoardMicroprocessor

Board

PressureTransducer

Pressure Input

Channel Board

ACCU-GAGE Instruction Manual page 25

inside the instrument.

4.3.10 VehiclePower Option

Models AG400 and AG401 instruments are available with a vehicle power option foroperation with batteries and linear DC power supplies. See “Chassis Specifications”on page22 voltage and power requirements.

Due to the momentary startup inrush current of the instrument’s power supply,the use of switching power supplies with the instrument is not recommended.

4.3.11 FuseReplacement

The power-line fuses of AC-powered instruments are located within the instrument'spower entry module on the rear panel. Use two 2A, 250V fast-blow fuses (Sensotecp/n 029-3026-00).

4.3.12 SETUPProtection Jumper

The SETUP Protection Jumper, when installed, locks out access to the SETUP Menumode. This prevents un-authorized personnel from re-calibrating the instrument ormaking other changes in how the instrument operates. When the SETUP ProtectionJumper is installed, the display will show the message “SSSSEEEETTTTUUUUPPPP PPPPRRRROOOOTTTTEEEECCCCTTTTEEEEDDDD” when theinstrument is powered up.

As shipped from the factory, the SETUP Menu mode is disabled so as to protect thefactory-performed calibration. If a re-calibration or change in the instrument’s settingsis required, perform the following:

Step 1: Turn the instrument off.

Step 1: Remove the two Phillips-head screws that secure the front panel.

Step 2: Remove the front panel. Optionally, the two ribbon cables connecting the frontpanel/display board assembly may be disconnected to provide easier access.

Step 3: Locate the SETUP Protection Jumper on the SensoBus Backplane boardusing the figure below.

Step 4: Using a pair of tweezers, remove the SETUP Protection Jumper.

Step 5: Replace the front panel and re-install the two Phillips-head screws.

Re-installation of the SETUP Protection Jumper is the reverse of removal.

Figure 4-4: SensoBus Backplane Board

Installed SETUP Protection Jumper

page 26 008-0629-00

CH

AS

SIS

MO

DE

LS4

ACCU-GAGE Instruction Manual page 27

page 28 008-0629-00

SY

ST

EM

CO

NN

EC

TO

R5

Chapter 5System Connector

5.1 Introduction

The 25-pin D-subminiature System connector is used for the following:

• Communication by RS-232 or RS-485. RS-232 DCE standard designationshave been maintained.

• Digital Function Inputs, such as for resetting tare, peak and latched limits• Open-collector digital Limit Outputs for limits 1 through 4.

ACCU-GAGE Instruction Manual page 29

5.2 System Connector Pinout

The Limit Output pins and Function Input pins are electrically isolated from the rest ofthe instrument.

The RS-232 and RS-485 communications pins are electrically isolated from the rest ofthe instrument.

The RS-232 and RS-485 interfaces are exclusive; an instrument cannot have both.

Table 5-1: System Connector

Pin Name Function Input/Output Reference Pin

1 N/C No Connection N/A

2 RS-232 IN RS-232 Data In Input 7

3 RS-232 OUT RS-232 Data Out Output 7

4 CTS RS-232 Clear to Send (connected to pin 5) N/A N/A

5 RTS RS-232 Request to Send (connected to pin 4) N/A N/A

6 DSR RS-232 Data Set Ready Output 7

7 GND RS-232/RS-485 reference Reference -

8 DCD RS-232 Data Carrier Detect (not connected) N/A N/A

9 FUNC 2 Function Input #2:.Clear Peak/Valley & Limits Input 19

10 FUNC 1 Function Input #1: Tare Off for all channels Input 19

11 FUNC 0 Function Input #0 Input 19

12 RS-485 TB RS-485 Transmit B Output 7

13 RS-485 TA RS-485 Transmit A Output 7

14 L1 Limit 1 Output (Open Collector) Output 19

15 L2 Limit 2 Output (Open Collector) Output 19

16 L3 Limit 3 Output (Open Collector) Output 19

17 L4 Limit 4 Output (Open Collector) Output 19

18 N/C No Connection N/A

19 DGND DGND (Digital Ground) Reference

20 DTR RS-232 Data Terminal Ready (not connected) N/A N/C

21 FUNC 3 Function Input #3: Tare On for all channels Input 19

22 RI RS-232 Ring Indicator (pulled up to 5V) Output

23 N/C No Connection N/A N/C

24 RS-485 RB RS-485 Receive B Input 7

25 RS-485 RA RS-485 Receive A input 7

page 30 008-0629-00

SY

ST

EM

CO

NN

EC

TO

R5

5.3 Function Input Pins

5.3.1 Overview To use a Function Input pin (9, 10, 11 or 21), connect it to the DGND (pin 19) momen-tarily. This can be accomplished by a push button switch, relay contact closure, orPLC output.

Usually, the Function Input pins perform the default actions described in the “SystemConnector Pinout” on page30. However, a SensoCode program running on a Mathe-matics Virtual Channel may replace these default actions. Consult the CustomerInformation Sheet included with your instrument for details.

5.3.2 Example For example, assume that you wish to use Function Input #3 to tare all channels in theinstrument simultaneously. Connect a push-button switch as shown below.

Figure 5-1: Function Input Example

ACCU-GAGE Instruction Manual page 31

5.4 Limit Output Pins

5.4.1 Overview An open-collector output is a transistor logic output that can be used to control DCloads, drive opto-isolators or relays, or interface directly to logic circuitry. They actvery much like switches: low resistance when turned on and high resistance whenturned off. You can use them as you would a dry relay contact, with the followingrestrictions:

• The voltage applied must be DC• The polarity of the DC voltage must be observed• Maximum voltage: 50VDC• Maximum power: 2.0W

When Limits 1, 2, 3, or 4 are activated, the corresponding open-collector Limit Outputpin on the System connector will be connected to the DGND (Digital Ground) pin.However, a SensoCode program running on a Mathematics Virtual Channel mayoverride this behavior. Consult the Customer Information Sheet included with yourinstrument for details.

5.4.2 Example For example, assume that a remote indicator is to be lighted when Limit 1 is activated.An external 24 VDC power supply connects to the indicator.

When Limit 1 is deactivated, there is high resistance between the Limit 1 Output pinand the DGND (Digital Ground) pin; little current flows in the circuit and the light is off.When Limit 1 is activated, there is low resistance between the Limit 1 Output pin andthe DGND (Digital Ground) pin; the light turns on.

The resistor limits the current flowing in the circuit when the light is lighted to 48 mA.According to Ohm’s Law:

Lights or other indicators have voltage, current, and/or power ratings that must beobserved in order to avoid damaging them.

Figure 5-2: Open-Collector Output Example

I VR---=

48mA 24V500Ω-------------=

page 32 008-0629-00

SY

ST

EM

CO

NN

EC

TO

R5

ACCU-GAGE Instruction Manual page 33

page 34 008-0629-00

SY

ST

EM

ME

NU

6

Chapter 6System Menu

6.1 Overview

The System menu allows you to examine and change settings that affect thechassis of the instrument. You can view the internal software revision and theinstrument’s configuration (i.e. what types of cards are installed in each channel).

Detailed instructions on operating the instrument in the SETUP Menu mode canbe found in “SETUP Menu mode” on page 19. A diagram of all menus is locatedin the “Setup Menu Reference” on page91.

6.2 Menu Items

6.2.1 SOFTWSOFTWSOFTWSOFTWAAAARE REVRE REVRE REVRE REVIIIISIONSIONSIONSION

Menu ItemThis displays the software part number and revision that is resident in the Micro-processor Board of the instrument.

6.2.2 CONFICONFICONFICONFIGGGGURATIOURATIOURATIOURATIONNNN Sub-Menu

When selected, a sub-menu is displayed which lists all available channels in theinstrument. By pressing [ENTER] when a channel number is displayed, the cardtype installed in that channel is shown. If a card is not installed in that channel, themessage “NNNNOOOOTTTT IIIINNNNSSSSTTTTAAAALLLLLLLLEEEEDDDD” is shown. Press [ENTER] again to return to the sub-menu listing of all channels.

6.2.3 DIAGNODIAGNODIAGNODIAGNOSSSSTICSTICSTICSTICS Sub-Menu

This sub-menu allows exercising and monitoring of the System connector’s outputand input pins.

OOOOUUUUTPTPTPTPUUUUTTTT TESTESTESTESTTTT Sub-Menu Select an output pin to turn “on” (connected to pin 19) or “off” (disconnected frompin 19). The output pins are updated immediately.

IIIINNNNPPPPUUUUTTTT TESTESTESTESTTTT Menu Item When this item is selected, the status of all four digital inputs are continuouslyscanned and displayed. A “0” means that an input is not connected to pin 19 (notasserted), and a “1” means that it is connected (asserted). Press any button toexit this operation.

SSSSCCCCAAAANNNN TIMTIMTIMTIMEEEE Menu Item When selected, this menu item displays the time, in seconds, that it last took forthe chassis to service all of the channels. In the RUN mode, the chassis readseach channel’s track, peak and valley value sequentially. After each channel hasbeen serviced, the limits are processed.

The value displayed is obtained from the last execution of the RUN mode prior toentering the SETUP menu mode. If you enter the SETUP mode immediately afterpower up, the display will read “NNNNOOOOTTTT AVAILABLAVAILABLAVAILABLAVAILABLEEEE”.

ACCU-GAGE Instruction Manual page 35

6.2.4 IIIINNNNSTALL CSTALL CSTALL CSTALL CHHHHANNELANNELANNELANNEL

Menu ItemThis menu item will add a Virtual channel as the next highest channel number in thesystem. Since there is no room on the rear panel for any additional Input or Outputchannels, only Virtual Channels may be added.

Installing a channel will cause it to use the “default” or “empty” configurationinformation for that channel. Any calibration data, SensoCode mathematics

programs, display setup, or other information will be erased to default values!All other channels are unaffected.

Virtual ChannelInstallation Procedure

Use the following procedure to install a Virtual Channel:

1) Enter the SETUP menu mode, then select “SSSSYYYYSSSSTTTTEEEEMMMM MMMMEEEENNNNUUUU -> IIIINNNNSSSSTTTTAAAALLLLL CL CL CL CHAHAHAHANNNNNNNNEEEELLLL”. Theinstrument will then present a menu of card types available for installation.

2) Select the card type of the card you wish to install. After you select the cardtype to be installed, you are asked “AAAARRRREEEE YYYYOOOOU SU SU SU SUUUURRRREEEE????”. To cancel this operation, select“NO” or press the [EXIT] button. If “YES” is selected, “WWWWOOOORRRRKKKKIIIINNNNGGGG............” is displayed anthe installation will commence.

3) If the installation was successful, “DDDDOOOONNNNEEEE” will be displayed. If the installationfailed, you will see one of the following messages:

“CCCCAAAANNNN’’’’TTTT INSTALINSTALINSTALINSTALLLLL”: There is no more memory available for Virtual channels.

“SSSSYYYYSSSSTTTTEEEEMMMM IS FIS FIS FIS FUUUULLLLLLLL”: There are no unused channels available in the instrument.

After installing a Mathematics Virtual channel, you must load a SensoCode programinto the channel with a computer running the Sensotec “Instrument Utility Software”.See “Mathematics Virtual Channel” on page 77 for more information.

page 36 008-0629-00

SY

ST

EM

ME

NU

6

6.2.5 DEDEDEDELLLLETE CETE CETE CETE CHHHHANNELANNELANNELANNEL

Menu ItemThis menu item will delete the last channel in the instrument. Before deletionoccurs, the number of the channel to be deleted is displayed and you are asked“AAAARRRRE YE YE YE YOOOOU SU SU SU SUUUURRRREEEE”. To cancel this operation, select “NO” or press the [EXIT] button. If“YES” is selected, the last channel in the system will be deleted.

Re-installing a deleted channel will erase all of its configuration and cali-bration information!

Do not delete Channel 01, the Pressure Input channel, or the calibrationinformation will be lost!

6.2.6 DEDEDEDEFFFFAULT CAULT CAULT CAULT CHHHHANNELANNELANNELANNEL

Menu ItemThis menu item will reset all settings for the channel you select to their factorydefault values. Before the channel settings are reset, you are asked “AAAARRRRE YE YE YE YOOOOUUUU

SSSSUUUURRRREEEE????”. To cancel this operation, press the [EXIT] button. If “YES” is selected, thechannel’s settings will be reset.

Defaulting a channel is an operation that cannot be undone. All calibra-tion information and other settings will be erased.

ACCU-GAGE Instruction Manual page 37

page 38 008-0629-00

SE

RIA

LC

OM

MU

NIC

AT

ION

S7

Chapter 7Serial Communications

7.1 Overview

ACCU-GAGE Series instruments are designed to communicate with a remotecomputer system or terminal for the purpose of transferring data values from theinstrument to the remote system. The remote computer or terminal also can con-trol many of the functions performed by the instrument.

Since the Model AG400 and AG401 instruments are very similar to Sensotec’s SCSeries instruments, refer to document p/n 008-0610-00, “SC Series Communica-tions Guide” for details regarding serial communications. This document can bedownloaded from Sensotec’s web site, www.sensotec.com. The “SC SeriesCommunications Guide” assists with the hookup and wiring for data communica-tions and provides detailed information of all the available commands.

7.2 Rear Panel Connector

The System connector on the instrument’s rear panel is used, among other things,for serial communications. See “System Connector” on page 29 for wiring infor-mation.

All of the serial communications pins on the 25-pin System connector have 500Vof electrical isolation from all other pins and connectors on the instrument. Addi-tionally, all serial communications pins are protected against electrostatic dis-charge (ESD).

ACCU-GAGE Instruction Manual page 39

7.3 Communications Protocol

7.3.1 RS-232 vs.RS-485

ACCU-GAGE instruments are available with either of two communications protocols,RS-232 or RS-485. Only one of these can be installed at a time at the factory.

• RS-232 provides for only one receiver and transmitter per loop, and a loop lengthof no more than 50 feet.

• RS-485 allows up to 32 devices per loop, and a loop length of no more than 4000feet. All devices receive messages in parallel on the line, the so-called “multi-drop” system. To avoid garbled transmissions, only one device should respond toa particular message. Therefore, every device on the loop must have a uniqueaddress.

If you have an instrument with more than one channel, transmission must beaddressed to the appropriate channel within the instrument.

7.3.2 Parameters All ACCU-GAGE instruments use no parity, 8 data bits, and 1 stop bits (“N,8,1”) forserial communications. Baud rates available are 300, 600, 1200, 2400, 4800, 9600,19200 and 38400. As shipped from the factory, all instruments are set at 9600 baud.The baud rates are selected via the RS-232 or RS-485 and not though the front panel.

page 40 008-0629-00

SE

RIA

LC

OM

MU

NIC

AT

ION

S7

7.4 Serial Com. Menu

7.4.1 Overview The Serial Com menu allows you to examine the settings for serial communications aswell as test the communications link. You cannot change the settings for serial com-munications through the menus; these settings can only be changed via serial com-munications.

Detailed instructions on operating the instrument in the SETUP Menu mode can befound in “SETUP Menu mode” on page 19. A diagram of all menus is located in“Setup Menu Reference” on page 91.

7.4.2 INTINTINTINTEEEERFACERFACERFACERFACE

Menu ItemThis menu item indicates which serial communications interface is installed in theinstrument: “RRRRSSSS----222233332222” or “RRRRSSSS----444488885555”.

7.4.3 ADDRESADDRESADDRESADDRESSSSS MenuItem

This allows you to examine the two-character address that the instrument will respondto. The default address of “00” (ASCII codes decimal 30, decimal 30). Please see the“SC Series Serial Communications Guide” for information on using the “W4” com-mand to change the address.

7.4.4 BAUD RATEBAUD RATEBAUD RATEBAUD RATE MenuItem

This allows you to examine the baud rate that the instrument uses for serial communi-cations. The default baud rate is 9600. Please see the “SC Series Serial Communica-tions Guide” for information on using the “W1” command to change the baud rate.

7.4.5 AUAUAUAUTTTTO LINEO LINEO LINEO LINE----FEEDFEEDFEEDFEED

Menu ItemThis allows you to examine the setting (on/off) of the auto line-feed function. Whenset to “on”, a line-feed character (ASCII code decimal 10) is transmitted just before thecarriage return (ASCII code decimal 13) that signifies the end of the instrument’sresponse.

If the instrument is connected to a dumb terminal or printer, inserting a line feedbefore the carriage return will make each message sent by the instrument appear on aseparate line and thus be easier to read.

The default setting is “on”.

7.4.6 TTTTRRRRANSMITANSMITANSMITANSMIT TESTTESTTESTTEST

Menu ItemWhen this menu item is selected, the instrument immediately transmits the message“AAAADDDDDDDDRRRREEEESSSSSSSS nnnnn Tn Tn Tn TEEEESSSSTTTT” where “nn” is the two-character address of the instrument. Then themessage “MMMMEEEESSSSSSSSAAAAGGGGEEEE SSSSEEEENNNNTTTT” is shown on the display. This helps detect wiring or otherproblems with serial communications from the instrument to the computer or PLC.

7.4.7 LEAVELEAVELEAVELEAVE MENUMENUMENUMENU

Menu ItemWhen the menu item is displayed, press [ENTER] to leave this menu. Press [EXIT] toreturn to the RUN mode.

ACCU-GAGE Instruction Manual page 41

7.5 Pressure Input Channel Commands

7.5.1 Introduction This section only documents the commands that affect the operation of the PressureInput channel. Refer to document p/n 008-0610-00, “SC Series CommunicationsGuide” for details regarding serial communications. This document can be down-loaded from Sensotec’s web site, www.sensotec.com.

7.5.2 Descriptions

F0 Transmit Track Data

Purpose To transmit the channel’s tracking data value

Usage "#aaccF0↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001F0 ↵"

Reply " 5670.5 ” (typical)

Remarks The input channel continuously reads the transducer. Thisfunction transmits the most recent reading.

F1 Activate Tare

Purpose To reset the channel’s data values to zero.

Usage "#aaccF1↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001F1 ↵"

Reply "OK”

page 42 008-0629-00

SE

RIA

LC

OM

MU

NIC

AT

ION

S7

F2 Deactivate Tare

Purpose To remove the offset from the channel’s data values which wasapplied when tare was activated.

Usage "#aaccF2↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001F2 ↵"

Reply "OK”

F9 Transmit Peak Data

Purpose To transmit the channel’s peak data value

Usage "#aaccF9↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001F9 ↵"

Reply " 12620.5 ” (typical) or "N/A ”

FA Transmit Valley Data

Purpose To transmit the channel’s valley data value

Usage "#aaccFA↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001FA ↵"

Reply "-0012.5 ” (typical) or "N/A ”

ACCU-GAGE Instruction Manual page 43

FB Clear Peak and Valley Data

Purpose To reset the channel’s peak and valley data values to the trackvalue.

Usage "#aaccFB↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001FB ↵"

Reply "OK” or "N/A ”

Remarks This command is not available on the Model SC1000.

FF Transmit Analog-to-Digital ConverterReading

Purpose To transmit the Analog-to-Digital converter reading as a per-centage from -100% to +100% of the A/D converters’s fullscale.

Usage "#aaccFF↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001FF ↵"

Reply " 50.000 ” (typical)

Remarks none

page 44 008-0629-00

SE

RIA

LC

OM

MU

NIC

AT

ION

S7

FH Write DAC Control Value

Purpose To control the channel’s Digital-to-Analog Converter (DAC)manually or automatically.

Usage "#aaccFHn↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel numbern is the argument defined below.↵ is the ‘carriage return’ character (ASCII decimal 13).

Argument "AUTO” returns the DAC to its normal automatic operation(power-on default). That is, it is controlled by the "DDDDAAAACCCC. C. C. C. CHHHHAAAANNNNNNNNEEEELLLL”,"DDDDAAAACCCC. SO. SO. SO. SOUUUURRRRCCCCEEEE”, "DDDDAAAACCCC. Z. Z. Z. ZEEEERRRROOOO----SSSSCCCCAAAALLLLEEEE” and "DDDDAAAACCCC. F. F. F. FUUUULLLLLLLL----SSSSCCCCAAAALLLLEEEE” menu items.

If the argument is a numeric value between -1 and +1, theDAC is forced to manual control and the value is used to drivethe DAC from -100% to +100% of its output.

Example "#0001FH.5 ↵" will force channel 01’s DAC to +50% of full-scale.

Reply "OK” or "ERROR”

Remarks You might wish to force a channel’s DAC to a certain output tohelp calibrate the attached data acquisition system or otherdevice.

R6 Read Display Units

Purpose Reads the channel’s "DDDDIIIISSSSPPPPLLLLAAAAYYYY UNITUNITUNITUNITSSSS” menu item

Usage "#aaccR6↵" to read.

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001R6 ↵”

Reply When reading: a four-character string.

ACCU-GAGE Instruction Manual page 45

RK/WK Read/Write Known-Load CalibrationPoint

Purpose Reads or writes the channel’s "KKKKNNNNOOOOWWWWNNNN POINTPOINTPOINTPOINT xxxx/y/y/y/y” menu items

Usage "#aaccRKpp↵" to read, "#aaccWKppn↵" to write.

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel numberpp is the parameter defined below.n is the known-load calibration point in engineering units.↵ is the ‘carriage return’ character (ASCII decimal 13).

Parameter "00” accesses the "KKKKNNNNOOOOWWWWNNNN POINT 1/POINT 1/POINT 1/POINT 1/2222 ”, "KKKKNNNNOOOOWNWNWNWN POINT 1/POINT 1/POINT 1/POINT 1/3333 ” or "KKKKNNNNOOOOWWWWNNNNPPPPOOOOIIIINNNNTTTT 1/1/1/1/5555” menu items.

"01” accesses the "KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/5555 ” menu item."02” accesses the "KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/3333 ” and "KNKNKNKNOOOOWWWWNNNN POINT 3POINT 3POINT 3POINT 3////5555” menu

items."03” accesses the "KKKKNNNNOOOOWWWWNNNN POINT 4/POINT 4/POINT 4/POINT 4/5555 ” menu item."04” accesses the "KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/2222 ”, "KKKKNNNNOOOOWNWNWNWN POINT 3/POINT 3/POINT 3/POINT 3/3333 ” or "KKKKNNNNOOOOWWWWNNNN

PPPPOOOOIIIINNNNTTTT 5/5/5/5/5555” menu items.

Example "#0001RK01↵”

Reply When writing: "OK" or "ERROR”.

When reading: an ASCII-floating-point value.

Changing this value has no effect until the channel is re-cali-brated to the transducer with the Known-Load Calibrationmethod.

page 46 008-0629-00

SE

RIA

LC

OM

MU

NIC

AT

ION

S7

RM/WM Read/Write DAC Channel & Value

Purpose Reads or writes the channel’s "DDDDAAAACCCC.... CHANNECHANNECHANNECHANNELLLL” and "DDDDAAAACCCC.... SOURCSOURCSOURCSOURCEEEE”menu items

Usage "#aaccRM↵" to read, "#aaccWMn↵" to write.

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number.n is the argument defined below.↵ is the ‘carriage return’ character (ASCII decimal 13).

Argument A number which selects a channel’s value. This number iscreated by adding together the values of the desired optionsas shown.

<Underline>ChannelValue Source Value01 1. TRACK 0.02 2. PEAK 16.03 3. VALLEY 32.04 4.05 5.06 6.07 7.08 8.09 9.10 10.11 11.12 12.13 13.14 14.15 15.16 64.17 65.18 66.19 67.20 68.21 69.22 70.23 71.

Example "#0001RM33↵" will cause the channel 01’s DAC to monitorchannel 01’s valley value.

Reply When writing: "OK", "ERROR” or "N/A ”.

When reading: a number corresponding to a channel numberand data value as shown above.

Remarks none

ACCU-GAGE Instruction Manual page 47

RN/WN Read/Write DAC Zero-Scale Value

Purpose Reads or writes the channel’s "DDDDAAAACCCC.... ZERO-SCALZERO-SCALZERO-SCALZERO-SCALEEEE” menu item

Usage "#aaccRN↵" to read, "#aaccWNn↵" to write.

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel numbern is the zero Analog Output value in engineering units.↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001WN-8000 ↵”

Reply When writing: "OK" or "ERROR”.

When reading: an ASCII-floating-point value.

RO/WO Read/Write DAC Full-Scale Value

Purpose Reads or writes the channel’s "DDDDAAAACCCC.... FULL-SCALFULL-SCALFULL-SCALFULL-SCALEEEE” menu item

Usage "#aaccRO↵" to read, "#aaccWOn↵" to write.

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel numbern is the full Analog Output value in engineering units.↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001WO8000↵”

Reply When writing: "OK" or "ERROR”.

When reading: an ASCII-floating-point value.

page 48 008-0629-00

SE

RIA

LC

OM

MU

NIC

AT

ION

S7

RP/WP Read/Write Operation

Purpose Read or write the operation settings of the channel

Usage "#aaccRPpp↵" to read, "#aaccWPppn↵" to write.

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number.pp is one of the two-character parameters given below.n is the argument.↵ is the ‘carriage return’ character (ASCII decimal 13).

Parameters& Arguments

Using pp="00” accesses the "AAAAUUUUTTTTOOOO----ZZZZEEEERRRROOOO” and "LLLLIIIINNNNEEEEAAAARRRRIIIIZZZZAAAATTTTIIIIOOOONNNN”menu items.

When writing this parameter, the argument n is a numberwhich selects different operating features. This number iscreated by adding together the values of the desiredoptions as shown.Auto-Zero Value Linearization ValueOFF 0. OFF 0.ON 2. ON 16.

Using pp="01” accesses the "CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN TTTTYYYYPPPPEEEE” menu item.When writing this parameter:

n=3 means 3-Point Known-Load Calibration.n=5 means 5-Point Known-Load Calibration.

Using pp="02” accesses the "AAAAUUUUXXXX1111 FUNCTIOFUNCTIOFUNCTIOFUNCTIONNNN” menu item.Using pp="03” accesses the "AAAAUUUUXXXX2222 FUNCTIOFUNCTIOFUNCTIOFUNCTIONNNN” menu item.

When writing these parameters:n=0 means the Auxiliary Function pin is disabled.n=1 means track hold.n=2 means peak & valley hold.n=4 means peak & valley clear (edge triggered).n=16 means activate Tare function.n=32 means deactivate Tare function.

Using pp="04” accesses the "CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOON UN UN UN UNNNNIIIITTTT” menu item.This parameter can be read from, not written to.

Example Sending "#0001WP0216↵" will allow the assertion of channel01’s Auxiliary 1 Function pin to activate the Tare function. Theinstrument will reply with "OK".

Reply When writing: "OK" or "ERROR”

When reading: a numeric value or text string according to theinformation above.

ACCU-GAGE Instruction Manual page 49

RR Read Version Info

Purpose Read version information found on the "VVVVEEEERRRRSSSSIIIIOOOONNNN INFINFINFINFOOOO” menuitem.

Usage "#aaccRR↵"

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number↵ is the ‘carriage return’ character (ASCII decimal 13).

Example "#0001RR↵"

Reply A text string such as

084-1169-01 04

Remarks This is the part number and version of the firmware used bythe channel’s microprocessor.

RT/WT Read/Write Front Panel Switch Operation

Purpose Reads or writes the operation of the front panel switches whenthe Protection jumper is installed.

Usage "#aaccRT↵" to read, "#aaccWTn↵" to write.

# is the ‘pound’ or ‘hash’ character (ASCII decimal 35).aa is the two-character instrument address.cc is the two-character channel number.n is the argument defined below.↵ is the ‘carriage return’ character (ASCII decimal 13).

Argument The argument is created by adding together the values of thedesired options as shown.

[VALUE] Value [CLEAR] ValueENABLED 0. ENABLED 0.DISABLED 8. DISABLED 4.

[CHANNEL] Value [TARE] ValueENABLED 0. ENABLED 0.DISABLED 2. DISABLED 1.

Example It is desired to disable the [TARE] button for channel 02 whenthe Protection jumper is installed. Sending "#0002WT1↵" willaccomplish this.

Reply When writing: "OK" or "ERROR”.

When reading: an ASCII-floating-point value described above.

Remarks If a disabled front panel button is pressed, the message " PPPPRRRROOOO----TTTTEEEECCCCTTTTEEEEDDDD” will appear on the display.

page 50 008-0629-00

SE

RIA

LC

OM

MU

NIC

AT

ION

S7

ACCU-GAGE Instruction Manual page 51

page 52 008-0629-00

DIS

PLA

YM

EN

U8

Chapter 8Display Menu

8.1 Overview

The Display menu allows you to examine and change settings that affect the oper-ation of the display of the instrument. You can change which channel is displayedwhen the instrument is powered-up and what information is shown on the lowerline of the display.

Detailed instructions on operating the instrument in the SETUP Menu mode canbe found in “SETUP Menu mode” on page 19. A diagram of all menus is locatedin the “Setup Menu Reference” on page91.

8.2 Menu Items

The instrument has a dual-line display which consists of an upper line and a lowerline.

8.2.1 UUUUPPPPPER CPER CPER CPER CHHHHANNELANNELANNELANNEL

Menu ItemSpecifies the channel that is displayed on the upper line when the instrument ispowered-up.

To change whether the channel’s track, peak or valley data value is displayed onpower-up, use the channel’s “OOOOPPPPEEEERRRRAAAATTTTIIIIOOOONNNN -> PPPPOOOOWWWWEEEERRRR----OOOONNNN SOURCSOURCSOURCSOURCEEEE” menu item.

8.2.2 LLLLOOOOWER CWER CWER CWER CHHHHANNELANNELANNELANNEL

Menu ItemSpecifies the channel that is displayed on the lower line when the “LLLLOOOOWWWWEEEERRRR MMMMODODODODEEEE”menu item is set to “CCCCHHHHAAAANNNNNNNNEEEELLLL”.

To change whether the channel’s track, peak or valley data value is displayed onpower-up, use the channel’s “OOOOPPPPEEEERRRRAAAATTTTIIIIOOOONNNN -> PPPPOOOOWWWWEEEERRRR----OOOONNNN SOURCSOURCSOURCSOURCEEEE” menu item.

8.2.3 LOLOLOLOWWWWER MODER MODER MODER MODEEEE MenuItem

Determines what is displayed on the lower line. The choices are:

• “BBBBLLLLAAAANNNNKKKK” means that nothing is displayed.• “LLLLIIIIMMMMIIIITTTT 01-001-001-001-04444” means that the status of Limit 1, Limit 2, Limit 3, and Limit 4 are

displayed as described in “Indicator Lights” on page17• “CCCCHHHHAAAANNNNNNNNEEEELLLL” means that the channel specified with the “LLLLOOOOWWWWEEEERRRR CHANNECHANNECHANNECHANNELLLL ” menu item

is displayed.

8.2.4 DIDIDIDISSSSPLAY DPLAY DPLAY DPLAY DIIIISABLESABLESABLESABLE

Menu ItemDetermines if the display is enabled or not. In some applications, the display isnot used in order to reduce power consumption; the instrument is then completelycontrolled with the Serial Communications interface. The choices are:

• “OOOOFFFFFFFF” means that the display operates normally.• “OOOONNNN” means that the display is blanked after the instrument is powered up.

Pressing any button on the front panel will temporarily re-enable the displayuntil the instrument is reset.

ACCU-GAGE Instruction Manual page 53

page 54 008-0629-00

LIMIT

S9

Chapter 9Limits

9.1 Understanding Limits, Set Points and Return Points

Limits are signal levels at which some action (such as a light to come on or go out,or a switch to close) is desired to take place. The point at which this action takesplace is the SET POINT. The RETURN POINT is the point at which the action thattook place at the SET POINT returns to its original state.

If the SET POINT and RETURN POINT are identical and the signal level ischanging slowly near these points, “chatter” may result and the limit rapidly ener-gizes and de-energizes. To keep this from happening, we can specify the SETPOINT at which the action would occur as well as the RETURN POINT where wewant the action to cease. If the amount of “dead band” between these two valuesis more than the expected noise, the desired action will take place only once. TheSET POINT and the RETURN POINT should always be separated to insure thatsmall amounts of noise on the signal will not become a problem.

A good general rule for SET POINT and RETURN POINT values is that theyshould be separated by at least one percent of the full-scale value.

ACCU-GAGE Instruction Manual page 55

9.2 Limit Operation

9.2.1 Actions WhenActivated

When Limits 1, 2, 3, or 4 are activated, the following actions occur:

• The front panel light for the limit will illuminate.• The corresponding open-collector Limit Output pin on the System connector will

be connected to the DGND (Digital Ground) pin.

9.2.2 Scan Time The “scan time” or update rate of the limits depends on how many channels areinstalled in the instrument. To determine the scan time of the limits for your particularinstrument, see “SCAN TIME Menu Item” on page35.

page 56 008-0629-00

LIMIT

S9

9.3 Limit Menus

9.3.1 Overview There are four Limit menus which control the operation of the open-collector LimitOutputs on the System connector.

Detailed instructions on operating the instrument in the SETUP Menu mode canbe found in “SETUP Menu mode” on page 19. A diagram of all menus is locatedin “Setup Menu Reference” on page 91.

9.3.2 LIMIT.LIMIT.LIMIT.LIMIT.EEEENABLENABLENABLENABLE MenuItem

This enables or disables the operation of this limit. The two options for this menuitem are “OOOONNNN” and “OOOOFFFFFFFF”.

9.3.3 LILILILIMMMMIT.SETIT.SETIT.SETIT.SETPPPPOINTOINTOINTOINT

Menu ItemThis is the numeric value, in engineering units, for the signal level that activatesthe limit.

9.3.4 LIMILIMILIMILIMITTTT.RETUR.RETUR.RETUR.RETURNNNN PNTPNTPNTPNT

Menu ItemThis is the numeric value at which the limit deactivates. In general, this numbershould differ from the set point value by at least 1% of full scale.

Carefully consider what value should be entered for the RETURN POINT. If thesignal is expected to approach the SET POINT from a lower value, the RETURNPOINT value should be less (algebraically) than the SET POINT. If the signal isexpected to approach the SET POINT from a higher level, the RETURN POINTshould be higher (algebraically) than the SET POINT.

9.3.5 LILILILIMMMMIT.ENEIT.ENEIT.ENEIT.ENERRRRGIZEGIZEGIZEGIZE

Menu ItemThis specifies when to activate the limit relative to the set point and return point.

The options for this menu item are:

• “SSSSIIIIGGGGNNNNAAAALLLL > S> S> S> SEEEETTTTPPPPOOOOININININTTTT” means the limit will activate when the signal is higher thanthe set point.

• “SSSSIIIIGGGGNNNNAAAALLLL < S< S< S< SEEEETTTTPPPPOOOOININININTTTT” means the limit will activate when the signal is lower thanthe set point.

• “SSSSIIIIGGGGNNNNAAAALLLL INSIDINSIDINSIDINSIDEEEE” means the limit will activate when the signal is in between theset point and the return point.

• “SSSSIIIIGGGGNNNNAAAALLLL OUTSIDOUTSIDOUTSIDOUTSIDEEEE” means the limit will activate when the signal is not in betweenthe set point and the return point.

The figures below further illustrate the differences between these settings.

Figure 9-1: Limit Operation when LLLLIIIIMMMMIIIITTTT....EEEENNNNEEEERRRRGGGGIIIIZZZZEEEE = SSSSIIIIGGGGNNNNAAAALLLL > S> S> S> SEEEETTTTPPPPOOOOIIIINNNNTTTT

ACCU-GAGE Instruction Manual page 57

Figure 9-2: Limit Operation when LLLLIIIIMMMMIIIITTTT....EEEENNNNEEEERRRRGGGGIIIIZZZZEEEE = SSSSIIIIGGGGNNNNAAAALLLL < S< S< S< SEEEETTTTPPPPOOOOININININTTTT

Figure 9-3: Limit Operation when LLLLIIIIMMMMIIIITTTT....EEEENNNNEEEERRRRGGGGIIIIZZZZEEEE = SSSSIIIIGGGGNNNNAAAALLLL INSIDINSIDINSIDINSIDEEEE

Figure 9-4: Limit Operation when LLLLIIIIMMMMIIIITTTT....EEEENNNNEEEERRRRGIGIGIGIZZZZEEEE = SSSSIIIIGGGGNNNNAAAALLLL OUTSIDOUTSIDOUTSIDOUTSIDEEEE

9.3.6 LILILILIMMMMIT.LATIT.LATIT.LATIT.LATCCCCHINGHINGHINGHING

Menu ItemThis specifies whether to latch the activated limit so that only manually clearing thelimit will deactivate it. A latched limit can be manually cleared by pressing the[CLEAR] button, sending the “F8” serial communications command, or by using theFunction Input #2 (Clear Peak/Valley & Limits pin on the System Connector. See

page 58 008-0629-00

LIMIT

S9

“Function Input Pins” on page 31.

The options for this menu item are:

• “OOOONNNN” means that once a signal activates the limit, the limit will remain activated untilit is reset.

• “OOOOFFFFFFFF” means that the limit will deactivate immediately when the signal crosses thereturn point.

9.3.7 LLLLIIIIMIT.CMIT.CMIT.CMIT.CHHHHANNELANNELANNELANNEL

Menu ItemThis specifies the channel that is monitored by this limit.

9.3.8 LLLLIMIT.SIMIT.SIMIT.SIMIT.SOOOOURCEURCEURCEURCE

Menu ItemThis designates the data source of the channel monitored by this limit. Each channelhas three data sources: the live tracking value (TRACK), its highest value (PEAK),and its lowest value (VALLEY).

The options for this menu item are:

• “TTTTRRRRAAAACCCCKKKK” means the live tracking value of the channel.• “PPPPEEEEAAAAKKKK” means the highest value of the channel since the peak/valley detector was

last cleared.• “VVVVAAAALLLLLLLLEEEEYYYY” means the lowest value of the channel since the peak/valley detector was

last cleared.

9.3.9 LEAVELEAVELEAVELEAVE MENUMENUMENUMENU

Menu ItemWhen the menu item is displayed, press [ENTER] to leave this menu. Press [EXIT] toreturn to the RUN mode.

ACCU-GAGE Instruction Manual page 59

page 60 008-0629-00

PR

ES

SU

RE

INP

UT

CH

AN

NE

L10

Chapter 10Pressure Input Channel

10.1 Features

The Pressure Input channel provides a DC excitation voltage to and accepts milli-volt signals from the internal pressure transducer. These millivolt signals are digi-tized, converted into engineering units, and placed into the track, peak and valleydata values of the channel. Setup and calibration of the channel to the pressuretransducer is made at the factory. Re-calibration can be made manually throughthe SETUP mode with the known-load calibration method.

The analog-to-digital converter input circuit is ratiometric to the excitation voltage,features digital, low-pass filtering, and typically ±50,000 count resolution.

The engineering units used to display the pressure readings are field selectablefrom a built-in conversion table. In addition, the engineering units used for calibra-tion can be independently selected.

Two rear panel control inputs can be field-configured for such functions as remotetare, track/hold, disabling peak/valley detection and clearing the peak/valley val-ues. A voltage or current digital-to-analog output is also provided.

ACCU-GAGE Instruction Manual page 61

10.2 Wiring

Connect the Pressure Input channel by wiring it to its 12-pin connector. The pin-out forthis connector is shown on the following table.

The Analog Output and Analog Return pins are electrically isolated from all other pinson the instrument.

Do not connect anything to the pins labeled as “N/C”. Damage to the instru-ment or loss of accuracy may result.

Table 10-1: Pressure Input Channel Pin Connections

Pin Label Function Notes

1 (top) N/C Transducer (+)Excitation DO NOT CONNECT!

2 N/C Transducer (+)Sense DO NOT CONNECT!

3 N/C Transducer (-)Sense DO NOT CONNECT!

4 N/C Transducer (-)Excitation DO NOT CONNECT!

5 N/C Transducer (+)Signal DO NOT CONNECT!

6 N/C Transducer (-)Signal DO NOT CONNECT!

7 +OUT Analog Output referenced to pin 8

8 -OUT Analog Return -

9 N/C N/C DO NOT CONNECT!

10 DGND Digital Ground -

11 AUX1 Auxiliary Function 1(connect to pin 10 to activate)

referenced to pin 10

12(bottom)

AUX2 Auxiliary Function 2(connect to pin 10 to activate)

referenced to pin 10

page 62 008-0629-00

PR

ES

SU

RE

INP

UT

CH

AN

NE

L10

10.3 Calibration

10.3.1 CalibrationConsiderations

If you are not familiar with operating the instrument in the SETUP menu mode,see “SETUP Menu mode” on page 19. A listing of all menu items is given in“Setup Menu Reference” on page 91.

In order to obtain optimum performance from the instrument when testing or re-calibrating, Sensotec recommends the following:

• Allow a 5-minute warm-up period before testing or calibration.• Just prior to starting the calibration, exercise the pressure transducer by

applying momentarily and then removing full-scale pressure.• The pressure standard you use should be at least 4 times more accurate than

the specification of the instrument.

10.3.2 RequiredPressures

In order to re-calibrate the instrument, you must have a precision pressure stan-dard that can produce pressures near the zero-scale, quarter-scale, half-scale,three-quarter scale and full-scale pressures for the instrument's range. Examinethe table below to determine the five pressures needed to calibrate your instru-ment. For example, if your instrument has a range of 100 PSIG, your pressurestandard must be able to accurately produce pressures near 0 PSI, 25 PSI, 50PSI, 75 PSI and 100 PSI.

10.3.3 CalibrationUnits of Measure

As shipped from the factory, instruments are usually calibrated in PSI regardlessof the field-selected display engineering units. If you wish to re-calibrate theinstrument using a unit-of-measure other than PSI, you can use the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN

UUUUNNNNIIIITTTTSSSS menu item to change the units-of-measure that the instrument uses for cali-bration. Then, you can re-enter the KKKKNONONONOWWWWN PN PN PN POOOOIIIINNNNTTTT menu items to match the pressuresyou are able to produce.

10.3.4 Traceability To maintain NIST traceability, Sensotec can re-calibrate an ACCU-GAGE instru-ment for you. NIST certificates can be ordered as a separate accessory for anominal fee.

10.3.5 CalibrationProcedure

Step 1: Remove the SETUP protection jumper per “SETUP Protection Jumper” onpage 26.

Step 2: Enter the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN TYPTYPTYPTYPEEEE.There are two methods that can be used to calibrate the transducer to theInput Channel. To obtain maximum accuracy, calibrate using the 5-PointKnown Load Calibration method whenever possible. See “CALIBRATIONTYPE Menu Item” on page67.

Step 3: Enter the CALIBRATION DCALIBRATION DCALIBRATION DCALIBRATION DAAAATATATATA....

You must make certain that the KKKKNNNNOOOOWWWWN PON PON PON POIIIINNNNTTTT menu items match the pressuresyou are able to produce.

Step 4: Perform the calibration.Use the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTEEEE menu item to start the calibration process. You will beprompted to apply loads to the transducer as required.

Step 5: Install the SETUP protection jumper per “SETUP Protection Jumper” onpage 26.

ACCU-GAGE Instruction Manual page 63

10.4 Pressure Input Channel Specifications

For chassis specifications, see “Chassis Specifications” on page22

PRESSURE TRANSDUCER INPUT

A/D Converter 24-bit Sigma-Delta

Low-pass filter digital, 24-tap FIR

Displayed Resolution +/- 50,000 counts (not including min. 10%overrange/underrange capability)

Frequency Response 2 Hz

Step Response 40 ms (typical)

Maximum overpressure without trans-ducer damage or loss of calibration

full-scale < 200 PSI : 4x full-scale200 ≤ full-scale ≤ 15K PSI: 1.5x full-scale

15K PSI ≤ full-scale: 1.2x full-scale

Accuracy See “Chassis Specifications” on page22.

Temperature Effect See “Chassis Specifications” on page22.

Pressure Port See “Pressure Connection” on page 13.

AUXILLIARY INPUTS

Quantity 2

Type momentary contact closure

Response Time < 5ms

Field-Selectable Functions tare on, tare off,peak/valley clear, peak/valley hold,

track hold

ANALOG OUTPUT

Output voltage range 5, ±5, 10 or ±10 VDC (field selectable)

Output current range (optional currentoutput channels)

4-20 mA

Source any channel’s track, peak or valley value

Isolation 500V

Resolution 13 bits

Frequency Response same as input when driven by the samechannel’s tracking data

page 64 008-0629-00

PR

ES

SU

RE

INP

UT

CH

AN

NE

L10

10.5 Channel Menu

The Pressure Input channel is configured and calibrated via its channel menu.Detailed instructions on operating the instrument in the SETUP Menu mode canbe found in “SETUP Menu mode” on page 19. A diagram of all menus is locatedin “Setup Menu Reference” on page 91.

10.5.1 OPEOPEOPEOPERRRRATIONATIONATIONATION

Sub-MenuThis menu controls the operation of this channel when the instrument is in theRUN mode.

AAAAUUUUTTTTOOOO----ZZZZEEEERRRROOOO Menu Item This enables or disables the channel’s automatic zeroing function. The auto-zerofunction can be beneficial by removing the effect of slight temperature shifts in thetransducer’s output. The two options for this menu item are “OOOONNNN” and “OFOFOFOFFFFF”.

When enabled, the channel will reset its tracking value to zero when the trackingvalue has never been more positive than +10 or more negative than -10 displaycounts for any 10 second period.

For example, assume that the display is using two decimal points. If the trackingvalue never leaves the range of ----000000000000....11110000 to ++++000000000000....11110000 for any 10 second period, thentracking value will reset to 000000000000....00000000 automatically.

To compensate for the transducer’s initial warm-up shift, the auto-zerofunction operates every 2 seconds during the first 60 seconds of RUN

mode operation.

PPPPOOOOWWWWEEEERRRR----OOOONNNN SOURCSOURCSOURCSOURCEEEE Menu Item This menu selects which value is displayed by the channel when first entering theRUN mode.

The choices are:

• “TTTTRRRRAAAACCCCKKKK” means the live tracking value of the channel.• “PPPPEEEEAAAAKKKK” means the highest value of the channel.• “VVVVAAAALLLLLLLLEEEEYYYY” means the lowest value of the channel.

ACCU-GAGE Instruction Manual page 65

10.5.2 DDDDIIIISPLAYSPLAYSPLAYSPLAY UUUUNITSNITSNITSNITS

Menu ItemThis menu item controls how data values are scaled into engineering units, displayedby the channel, and transmitted via serial communications. The Pressure Input chan-nel has conversion factors, given below, for many standard pressure units-of-measurebuilt in. Note that if the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOON UN UN UN UNNNNIIIITTTT is set to something other than PSI, then thechoices available in the DIDIDIDISSSSPPPPLLLLAAAAYYYY UUUUNNNNIIIITTTTSSSS and the associated conversion factors willchange accordingly. For example, if the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOON UN UN UN UNNNNIIIITTTT is set to “kPa” then choicesavailable for the DDDDIIIISSSSPPPPLLLLAAAAYYYY UNITUNITUNITUNITSSSS will be “ kkkkPPPPaaaa to PSto PSto PSto PSIIII”,” kkkkPPPPaaaa to bato bato bato barrrr” , etc.

The decimal point position is set automatically based upon the display scaling in orderto provide the maximum resolution available; it cannot be changed manually.

When the DDDDIIIISSSSPPPPLLLLAAAAYYYY UUUUNNNNIIIITTTTSSSS setting is changed, all LLLLIIIIMMMMIIIITTTT.... SSSSEEEETTTTPPPPOOOOIIIINNNNTTTTs, LLLLIIIIMMMMIIIITTTT.... RRRREEEETTTTUUUURRRRNNNN PPPPNNNNTTTTs, tarevalues, DDDDAAAACCCC.... ZZZZEEEERRRROOOO----SSSSCCCCAAAALLLLEEEE and DDDDAAAACCCC.... FULL-SCALFULL-SCALFULL-SCALFULL-SCALEEEE values in the entire instrument that refer tothis channel will also be converted to the new DDDDIIIISSSSPPPPLLLLAAAAYYYY UNITUNITUNITUNITSSSS setting.

10.5.3 AAAAUUUUXn FUNXn FUNXn FUNXn FUNCCCCTIONTIONTIONTION

Menu ItemsThe AAAAUUUUXXXX1 F1 F1 F1 FUUUUNNNNCCCCTTTTIIIIOOOONNNN and AAAAUUUUXXXX2 F2 F2 F2 FUUUUNNNNCCCCTTTTIIIIOOOONNNN menu items determine what happens when theAuxiliary Function pins (labeled as “AUX1” and “AUX2”) on the channel’s connectorare activated. These pins are “activated” when they are connected to the DigitalGround (labeled as “DGND”) pin. The choices are:

• “DDDDIIIISSSSAAAABBBBLLLLEEEEDDDD” means that activating the pin does nothing.• “TTTTRRRRAAAACCCCKKKK HOLHOLHOLHOLDDDD” means that the tracking, peak and valley values will not be updated.• “HHHHIIIIGGGGHHHH////LLLLOOOOWWWW HOLHOLHOLHOLDDDD” means that the peak and valley values will not be updated.• “HHHHIIIIGGGGHHHH////LLLLOOOOWWWW CLEACLEACLEACLEARRRR” means that the peak and valley values are reset.• “TTTTAAAARRRREEEE OOOONNNN” activates the channel’s Tare function (resets it to zero).• “TTTTAAAARRRREEEE OFOFOFOFFFFF” deactivates the channel’s Tare function.As the Auxiliary Function pins are not isolated, it is recommended that a push-buttonswitch or relay is used to connect these pins to the Digital Ground pin.

Table 10-2: Pressure Conversion Factors

Unit of MeasureDDDDIIIISSSSPPPPLLLLAAAAYYYY UNITUNITUNITUNITSSSS Menu Item

(when CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN UNIUNIUNIUNITTTT is setto PSI)

ConversionFactor

PSI PPPPSSSSIIII to PSto PSto PSto PSIIII 1.000000

bar PPPPSSSSIIII to bato bato bato barrrr 0.06894757

mbar PPPPSSSSIIII to mto mto mto mbabababarrrr 68.947573

torr PPPPSSSSIIII to tto tto tto tororororrrrr 51.715075

inH2O @ 4°C PPPPSSSSIIII to ito ito ito inWnWnWnWAAAA 27.680673

ftH2O @ 4°C PPPPSSSSIIII to fto fto fto ftWtWtWtWAAAA 2.3067227

inHg @ 0°C PPPPSSSSIIII to ito ito ito inHnHnHnHgggg 2.0360259

mmHg @ 0°C PPPPSSSSIIII to mto mto mto mmHmHmHmHgggg 51.715075

kPa PPPPSSSSIIII to kPto kPto kPto kPaaaa 6.8947573

MPa PPPPSSSSIIII to MPto MPto MPto MPaaaa 0.0068947573

page 66 008-0629-00

PR

ES

SU

RE

INP

UT

CH

AN

NE

L10

10.5.4 CALCALCALCALIIIIBRATIONBRATIONBRATIONBRATION TYPETYPETYPETYPE

Menu ItemThis chooses the type of calibration technique to be used. The choices are:

• “TTTTYYYYPPPPEEEE= 3 P= 3 P= 3 P= 3 POOOOIIIINNNNT CT CT CT CAAAALLLL” means 3-Point Known Load Calibration. You are promptedto apply the loads to the transducer that were entered in the “KKKKNNNNOOOOWWWWNNNN POINT 1/POINT 1/POINT 1/POINT 1/3333 ”and “KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/3333 ” and “KKKKNNNNOOOOWWWWNNNN POINT 3POINT 3POINT 3POINT 3////3333” registers. This technique can beused to compensate for the non-linearity in the transducer. The usual loadsused are zero scale, half scale and full scale, but you are not restricted tothese loads.

• “TTTTYYYYPPPPEEEE= 5 P= 5 P= 5 P= 5 POOOOIIIINNNNT CT CT CT CAAAALLLL” means 5-Point Known Load Calibration. You are promptedto apply the loads to the transducer that were entered in the “KKKKNNNNOOOOWWWWNNNN POINT 1/POINT 1/POINT 1/POINT 1/5555 ”,“KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/5555 ”, “KKKKNNNNOOOOWWWWNNNN POINT 3/POINT 3/POINT 3/POINT 3/5555 ”, “KKKKNNNNOOOOWWWWNNNN POINT 4POINT 4POINT 4POINT 4////5555” and “KKKKNNNNOOOOWWWWNNNN POINT 5/POINT 5/POINT 5/POINT 5/5555 ” regis-ters. This technique can be used to compensate for the non-linearity in thetransducer. The usual loads used are zero scale, one-quarter scale, halfscale, three-quarters scale and full scale, but you are not restricted to theseloads.

The selection of the calibration type affects what menu items appear in the CCCCAAAALLLLIIII----

BBBBRRRRAAAATTTTIIIIOOOONNNN DATDATDATDATAAAA sub-menu.

To obtain maximum accuracy, always calibrate using the 5-Point KnownLoad Calibration method whenever possible.

ACCU-GAGE Instruction Manual page 67

10.5.5 CALIBCALIBCALIBCALIBRRRRATIONATIONATIONATION

DADADADATTTTAAAA Sub-MenuThis enters the values that will be used by the instrument when it calibrates itself tothe transducer. The menu items that are displayed will change according to what theCCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN TYPTYPTYPTYPEEEE is set for.

FFFFUUUULLLLLLLL SCALE MV/SCALE MV/SCALE MV/SCALE MV/VVVV MenuItem

The transducer output when its full-scale load is applied to it, in milliVolts-per-Volt.This menu item is used by the instrument to set the gain of its amplifier circuit.

This menu item is specific to the pressure transducer used in a particular instrumentand cannot be changed by the customer.

KKKKNNNNOOOOWWWWNNNN POINT x/POINT x/POINT x/POINT x/yyyy MenuItems

This enters the engineering units for the known-load calibration points. These pointsmust match the actual loads that you will apply to the instrument during calibration.The number of calibrations points depends on the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN TYPTYPTYPTYPEEEE.

When using the 3-Point Known Load Calibration type, the following menu items areavailable:

• “KKKKNNNNOOOOWWWWNNNN POINT 1/POINT 1/POINT 1/POINT 1/3333 ”: point 1 of 3, usually 0% of the transducer’s full-scale capacity.• “KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/3333 ”: point 2 of 3, usually 50% of the transducer’s full-scale capacity.• “KKKKNNNNOOOOWWWWNNNN POINT 3/POINT 3/POINT 3/POINT 3/3333 ”: point 3 of 3, usually 100% of the transducer’s full-scale capacity.

When using the 5-Point Known Load Calibration type, the following menu items areavailable:

• “KKKKNNNNOOOOWWWWNNNN POINT 1/POINT 1/POINT 1/POINT 1/5555 ”: point 1 of 5, usually 0% of the transducer’s full-scale capacity.• “KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/5555 ”: point 2 of 5, usually 25% of the transducer’s full-scale capacity.• “KKKKNNNNOOOOWWWWNNNN POINT 3/POINT 3/POINT 3/POINT 3/5555 ”: point 3 of 5, usually 50% of the transducer’s full-scale capacity.• “KKKKNNNNOOOOWWWWNNNN POINT 4/POINT 4/POINT 4/POINT 4/5555 ”: point 4 of 5, usually 75% of the transducer’s full-scale capacity.• “KKKKNNNNOOOOWWWWNNNN POINT 5/POINT 5/POINT 5/POINT 5/5555 ”: point 5 of 5, usually 100% of the transducer’s full-scale capacity.

To insure both correct operation of the transducer and application of the load,the instrument expects the voltage applied at each known-load point to beincreasing. For example, the load applied at Known-Load Point 2/5 must

cause the transducer to produce a more positive voltage than at Known-LoadPoint 1/5.

10.5.6 CALCALCALCALIIIIBRATEBRATEBRATEBRATE

Menu ItemThis menu item performs a calibration according to what was entered in the CCCCAAAALLLLIIIIBBBBRRRRAAAA----

TTTTIIIIOOOONNNN TYPTYPTYPTYPEEEE and CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN DATDATDATDATAAAA menu items.

Before performing a calibration, the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN TYPTYPTYPTYPEEEE must be selected (see“CALIBRATION TYPE Menu Item” on page67), and the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOON DN DN DN DAAAATTTTAAAA must

be entered (see “CALIBRATION DATA Sub-Menu” on page68).

For maximum accuracy, allow at least five minutes of warm-up before calibra-tion.

page 68 008-0629-00

PR

ES

SU

RE

INP

UT

CH

AN

NE

L10

If the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN TYPTYPTYPTYPEEEE is 3-Point Known Load Calibration ...

• The display will read DDDDOOOOIIIINNNNGGGG 3POINT CA3POINT CA3POINT CA3POINT CALLLL , and prompt you to AAAAPPPPPPPPLLLLYYYY 00000 U00000 U00000 U00000 UNNNNIIIITTTT

(where 00000 and UNIT are previously entered as KNOWN POINT 1/KNOWN POINT 1/KNOWN POINT 1/KNOWN POINT 1/3333and CCCCAAAALLLLIIIIBBBBRRRRAAAA----

TTTTIIIIOOOON UN UN UN UNNNNIIIITTTT). When you have applied this load to the transducer, press [ENTER].• The display will read WWWWOOOORRRRKKKKIIIINNNNGGGG, then AAAAPPPPPPPPLLLLY 0Y 0Y 0Y 00000555500000 U0 U0 U0 UNINININITTTT (where 00500 and UNIT are

previously entered as KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/3333 and CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN UNIUNIUNIUNITTTT). When you haveapplied this load to the transducer, press [ENTER].

• The display will read WWWWOOOORRRRKKKKIIIINNNNGGGG, then AAAAPPPPPPPPLLLLY 0Y 0Y 0Y 01111000000000 U0 U0 U0 UNINININITTTT (where 01000 and UNIT arepreviously entered as KKKKNNNNOOOOWWWWNNNN POINT 3/POINT 3/POINT 3/POINT 3/3333 and CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN UNIUNIUNIUNITTTT). When you haveapplied this load to the transducer, press [ENTER].

• Upon completion, the display will indicate DDDDOOOONNNNEEEE and the instrument will return tothe RUN mode.

If the CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN TYPTYPTYPTYPEEEE is 5-Point Known Load Calibration ...

• The display will read DDDDOOOOIIIINNNNGGGG 5POINT CA5POINT CA5POINT CA5POINT CALLLL , and prompt you to AAAAPPPPPPPPLLLLYYYY 00000 U00000 U00000 U00000 UNNNNIIIITTTT

(where 00000 and UNIT are previously entered as KNOWN POINT 1/KNOWN POINT 1/KNOWN POINT 1/KNOWN POINT 1/5555and CCCCAAAALLLLIIIIBBBBRRRRAAAA----

TTTTIIIIOOOONNNN UNIUNIUNIUNITTTT). When you have applied this load to the transducer, press[ENTER].

• The display will read WWWWOOOORRRRKKKKIIIINNNNGGGG, then AAAAPPPPPPPPLLLLY 0Y 0Y 0Y 00000222255550 U0 U0 U0 UNINININITTTT (where 00250 and UNIT arepreviously entered as KKKKNNNNOOOOWWWWNNNN POINT 2/POINT 2/POINT 2/POINT 2/5555 and CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN UNIUNIUNIUNITTTT). When you haveapplied this load to the transducer, press [ENTER].

• The display will read WWWWOOOORRRRKKKKIIIINNNNGGGG, then AAAAPPPPPPPPLLLLY 0Y 0Y 0Y 00000555500000 U0 U0 U0 UNINININITTTT (where 00500 and UNIT arepreviously entered as KKKKNNNNOOOOWWWWNNNN POINT 3/POINT 3/POINT 3/POINT 3/5555 and CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN UNIUNIUNIUNITTTT). When you haveapplied this load to the transducer, press [ENTER].

• The display will read WWWWOOOORRRRKKKKIIIINNNNGGGG, then AAAAPPPPPPPPLLLLY 0Y 0Y 0Y 00000777755550 U0 U0 U0 UNINININITTTT (where 00750 and UNIT arepreviously entered as KKKKNNNNOOOOWWWWNNNN POINT 4/POINT 4/POINT 4/POINT 4/5555 and CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN UNIUNIUNIUNITTTT). When you haveapplied this load to the transducer, press [ENTER].

• The display will read WWWWOOOORRRRKKKKIIIINNNNGGGG, then AAAAPPPPPPPPLLLLY 0Y 0Y 0Y 01111000000000 U0 U0 U0 UNINININITTTT (where 01000 and UNIT arepreviously entered as KKKKNNNNOOOOWWWWNNNN POINT 5/POINT 5/POINT 5/POINT 5/5555 and CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN UNIUNIUNIUNITTTT). When you haveapplied this load to the transducer, press [ENTER].

• The display will now read RRRREEEETTTTUUUURNRNRNRN PNTS.=NPNTS.=NPNTS.=NPNTS.=NOOOO. If you press [ENTER], the displaywill indicate DDDDOOOONNNNEEEE and the instrument will return to the RUN mode. If you usethe [UP] button and [ENTER] to select RRRREEEETTTTUUUURRRRNNNN POINTS=YEPOINTS=YEPOINTS=YEPOINTS=YESSSS , you may now applythe same loads again but in reverse order. This will allow the instrument tomeasure and partially eliminate any hysteresis that may be present in thetransducer. After the loads have been applied again, the display will indicateDDDDOOOONNNNEEEE and the instrument will return to the RUN mode.

ACCU-GAGE Instruction Manual page 69

10.5.7 DACDACDACDAC SSSSETUPETUPETUPETUP

Sub-MenuThis sub-menu contains four items that control the Digital-to-Analog Converter (DAC)output of the channel.

DDDDAAAACCCC.... CHANNECHANNECHANNECHANNELLLL Menu Item This chooses which channel will drive the DAC output. Normally, the DAC located ona particular channel will be driven by that channel, but that need not be the case. Forexample, if a Mathematics Virtual channel is installed on the instrument, the output ofthis Virtual channel could be used to drive this channel’s DAC output.

If the DAC is set-up to be driven by its own channel’s tracking value, the Analog Out-put will have the frequency response specified by “Pressure Input Channel Specifica-tions” on pag e64. Otherwise, the Analog Output will operate more slowly.

DDDDAAAACCCC.... SOURCSOURCSOURCSOURCEEEE Menu Item This designates the data source of the channel monitored by the analog output. Eachchannel has three data sources: the live tracking value (TRACK), its highest value(PEAK), and its lowest value (VALLEY).

The options for this menu item are:

• “TTTTRRRRAAAACCCCKKKK” means the live tracking value of the channel.• “PPPPEEEEAAAAKKKK” means the highest value of the channel since the peak/valley detector was

last cleared.• “VVVVAAAALLLLLLLLEEEEYYYY” means the lowest value of the channel since the peak/valley detector was

last cleared.If the DAC is set-up to be driven by its own channel’s tracking value, the Analog Out-put will have the frequency response specified by “Pressure Input Channel Specifica-tions” on pag e64. Otherwise, the Analog Output will operate more slowly.

DDDDAAAACCCC.... ZERO-SCALZERO-SCALZERO-SCALZERO-SCALEEEE MenuItem

This specifies what value, in engineering units, corresponds to zero output on theAnalog Output.

“Zero output” might be 0 Volts, 2.5 Volts, 5 Volts, 4 mA or 12 mA depending on if thechannel has a voltage or current output and how it is configured. See the “Digital-to-Analog Output” section of this chapter for details.

DDDDAAAACCCC.... FULL-SCALFULL-SCALFULL-SCALFULL-SCALEEEE MenuItem

Specifies what value, in engineering units, corresponds to full output on the AnalogOutput.

“Full output” might be 5 Volts, 10 Volts or 20 mA depending on if the channel has avoltage or current output and how it is configured. See the “Digital-to-Analog Output”section of this chapter for details.

page 70 008-0629-00

PR

ES

SU

RE

INP

UT

CH

AN

NE

L10

10.5.8 DIAGNODIAGNODIAGNODIAGNOSSSSTICSTICSTICSTICS

Sub-Menu

DDDDAAAACCCC FULL SFULL SFULL SFULL SCCCCALALALALEEEE Menu Item When this menu item is selected, the Analog Output of the channel is forced to itsfull-scale output, then DDDDAAAACCCC UUUUPPPPDDDDAAAATTTTEEEEDDDD is displayed. This is useful when calibrating ortrimming the readout connected to the Analog Output.

DDDDAAAACCCC ZERO SZERO SZERO SZERO SCCCCALALALALEEEE Menu Item When this menu item is selected, the Analog Output of the channel is forced to itszero-scale output, then DDDDAAAACCCC UUUUPPPPDDDDAAAATTTTEEEEDDDD is displayed. This is useful when calibrating ortrimming the readout connected to the Analog Output.

VVVVEEEERRRRSSSSIIIIOOOONNNN INFINFINFINFOOOO Menu Item This menu item displays the part number and revision level of the firmware usedby this channels microprocessor.

DDDDIIIISSSSPPPPLLLLAAAAYYYY ADADADADCCCC Menu Item The options for this menu item are:

• “OOOOFFFFFFFF” will allow the [VALUE] button to cycle through “TTTTKKKK” (tracking value), “HHHHIIII”(peak value) and “LLLLOOOO” (valley value). This is the recommended option.

• “OOOONNNN” will allow the “AAAADDDD” (percentage of the Analog-to-Digital converter’s full-scale digitizing capability) display source to be available along with “TTTTKKKK” (track-ing value), “HHHHIIII” (peak value) and “LLLLOOOO” (valley value) when the [VALUE] buttonis pressed on the front panel.

The Analog-to-Digital converter counts are displayed as a percentage from-100.00% to 100.00%.

This can be used to verify that the transducer’s output does not exceed the inputrange of the instrument.

LLLLIIIINNNNEEEEAAAARRRRIIIIZZZZAAAATTTTIOIOIOIONNNN Menu Item The options for this menu item are:

• “OOOONNNN” will allow linearization data obtained from either the 3-Point or 5-PointKnown Load Calibration to affect a channel’s scaled values. This is the rec-ommended option.

• “OOOOFFFFFFFF” will not allow linearization data to effect the channel’s scaled values.This is useful when diagnosing problems in a calibration procedure.

DDDDIIIISSSSAAAABBBBLLLLEEEE CHANNECHANNECHANNECHANNELLLL Menu Item The options for this menu item are:

• “OOOONNNN” will bypass the INITIALIZE and RUN modes of the channel. The track,peak and valley values of the channel are forced to 0. This option is helpfulwhen a channel is installed but not being used.

• “OOOOFFFFFFFF” will allow normal operation of the channel.

ACCU-GAGE Instruction Manual page 71

10.6 Analog Output Configuration

10.6.1 Identifyingthe Output Type

A Pressure Input channel is available with one of two types of digital-to-analog (DAC)outputs: voltage or current. You can determine which type of output a channel has byone of three ways:

• Consulting the instrument’s Customer Information Sheet• Examining the SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> CCCCOOOONNNNFFFFIIIIGGGGUUUURRRRAAAATTTTIIIIOOOONNNN -> CCCCHHHHAAAANNNNNNNNEEEELLLL nn TTTTYYYYPPPPEEEE menu item where nn is

the number of the channel. If the channel’s type is PPPPRRRREEEESSSSSSSSUUUURRRREEEE VouVouVouVoutttt, it has a voltageoutput. If the channel’s type is PPPPRRRREEEESSSSSSSSUUUURRRREEEE IIIIoooouuuutttt, it has a current output.

• Examining the channel’s circuit board as shown in the figure below.

10.6.2 ChannelMenu Items

The Analog Output can be driven by any channel’s track, peak or valley value.

See the “Channel Menu” section earlier in this chapter for a complete listing of SETUPmenu items available on the DDDDAAAACCCC.... SETUSETUSETUSETUPPPP sub-menu.

10.6.3 OutputSelection

Jumpers located on the channel’s circuit board determine what outputs are generatedwhen the value selected to drive the Analog Output (from the DDDDAAAACCCC.... CCCCHHHHAAAANNNNNNNNEEEELLLL and DDDDAAAACCCC....

SSSSOOOOUUUURRRRCCCCEEEE menu items) equals the DDDDAAAACCCC.... FULL SFULL SFULL SFULL SCCCCALALALALEEEE and DDDDAAAACCCC.... ZERO SZERO SZERO SZERO SCCCCALALALALEEEE settings.

Figure 10-1: Digital-to-Analog Output Jumper Locations

DDDDAAAACCCC.... ZERO SZERO SZERO SZERO SCCCCALALALALEEEE

OutputDDDDAAAACCCC.... FULL SFULL SFULL SFULL SCCCCALALALALEEEE

OutputJ30

jumperJ31

jumper

CHANNELS WITH VOLTAGE OUPUT

0-5V 2.5 Volts 5 Volts open closed

±5V 0 Volts 5 Volts open open

0-10V 5 Volts 10 Volts closed closed

±10V 0 Volts 10 Volts closed open

CHANNELS WITH CURRENT OUTPUT

4-20mA 4 mA 20 mA open open

4-20mA 12 mA 20 mA open closed

page 72 008-0629-00

PR

ES

SU

RE

INP

UT

CH

AN

NE

L10

10.7 Troubleshooting

10.7.1 Error Messages See “Error Messages” on page85 for information relating to error messages.

10.7.2 CommonProblems and

Solutions

Erratic Display Make sure that the pressure on the transducer is constant.

Instruments that read barometric pressure will indicate very small changes in airpressure, such as those caused by even a modest change in altitude.

SSSSEEEETTTTUUUUPPPP PROTECTEPROTECTEPROTECTEPROTECTEDDDD on Display As shipped from the factory, the SETUP Menu mode is disabled so as to protectthe setup and calibration data of the instrument. If a re-calibration or change inthe instrument’s settings is required, see “SETUP Protection Jumper” on page26.

+OVLD or -OVLD onDisplay

Check “Wiring” on page62 and make certain that nothing is connected to the pinslabeled as “N/C”.

Analog Output Incorrect Make certain of the type of Analog Output (voltage or current) that the channel isequipped with; see “Identifying the Output Type” on page72.

Use the DDDDIIIIAAAAGGGGNNNNOOOOSSSSTTTTIIIICCCCSSSS -> DDDDAAAACCCC FFFFUUUULLLLLLLL SSSSCCCCAAAALLLLEEEE and DDDDIIIIAAAAGGGGNNNNOOOOSSSSTTTTIIIICCCCSSSS -> DDDDAAAACCCC ZZZZEEEERRRRO SO SO SO SCCCCAAAALLLLEEEE menu items toforce the Analog Output to a known output. Then, adjust your readout device,panel meter, PLC or data acquisition system to match.

Check the output selection jumpers; see “Output Selection” on page72.

Auxiliary Function PinsNot Operating

Make sure that the AAAAUUUUXXXX1111 FFFFUUUUNNNNCCCCTTTTIIIIOOOONNNN or AAAAUUUUXXXX2 F2 F2 F2 FUUUUNNNNCCCCTTTTIIIIOOOONNNN SETUP menu items are set cor-rectly; if they are set to DDDDIIIISSSSAAAABBBBLLLLEEEEDDDD then they will not operate.

The Auxiliary Function (“AUX1” and “AUX2”) pins must be connected to pin 10,not pin 8, to activate them.

Pressure Port Rotates Always use a backing wrench when connecting to the pressure port. If the pres-sure transducer has become loose in its mounting, you can re-secure it. See“Pressure Transducer Mounting” on page 25.

ACCU-GAGE Instruction Manual page 73

page 74 008-0629-00

SP

LITD

ISP

LAY

VIR

TU

AL

CH

AN

NE

L11

Chapter 11Split Display Virtual Channel

11.1 Features

The Split Display Virtual Channel allows the display of two track, peak or valleyvalues from any two channels at the same time. One value is displayed on the leftside of the display, and another value is displayed on the right side.

The display setup used to format each value is inherited from the source channelof the value.

Split Displays are most useful when an instrument has many Virtual Mathematicschannels installed.

11.2 Wiring

A Virtual Channel exists in software only; it does not occupy a physical slot insideof the instrument’s chassis. However, a Virtual Channel does require a channelnumber.

There is no rear-panel connector for a Virtual Channel.

11.3 Setup Procedure

If you are not familiar with operating the instrument in the SETUP menu mode,see “SETUP Menu mode” on page19. A listing of all menu item is given in “SetupMenu Reference” on page91.

Step 1: Enter the value to be displayed on the left side with LLLLEEEEFFFFT CT CT CT CHHHHAAAANNNNNNNNEEEELLLL and LLLLEEEEFFFFTTTT

SSSSOOOOUUUURRRRCCCCEEEE.

Step 2: Enter the value to be displayed on the right side with RRRRIIIIGGGGHHHHT CT CT CT CHHHHAAAANNNNNNNNEEEELLLL andRRRRIIIIGGGGHHHHTTTT SOURCSOURCSOURCSOURCEEEE.

ACCU-GAGE Instruction Manual page 75

11.4 Channel Menu

Detailed instructions on operating the instrument in the SETUP Menu mode can befound in “SETUP Menu mode” on page 19. A diagram of all menus is located in“Setup Menu Reference” on page 91.

LLLLEEEEFFFFTTTT CHANNECHANNECHANNECHANNELLLL Menu Item This specifies the channel that is displayed on the left half of the split display.

LLLLEEEEFFFFTTTT SOURCSOURCSOURCSOURCEEEE Menu Item This lets you designate the data source of the channel monitored by the left half of thesplit display. Each channel has three data sources: the live tracking value (TRACK),its highest value (PEAK), and its lowest value (VALLEY).

The options for this menu item are:

• “TTTTRRRRAAAACCCCKKKK” means the live tracking value of the channel.• “PPPPEEEEAAAAKKKK” means the highest value of the channel since the peak detector was last

cleared.• “VVVVAAAALLLLLLLLEEEEYYYY” means the lowest value of the channel since the valley detector was last

cleared.

RRRRIIIIGGGGHHHHTTTT CHANNECHANNECHANNECHANNELLLL Menu Item This specifies the channel that is displayed on the right half of the split display.

RRRRIIIIGGGGHHHHTTTT SOURCSOURCSOURCSOURCEEEE Menu Item This lets you designate the data source of the channel monitored by the right half ofthe split display. Each channel has three data sources: the live tracking value(TRACK), its highest value (PEAK), and its lowest value (VALLEY).

The options for this menu item are:

• “TTTTRRRRAAAACCCCKKKK” means the live tracking value of the channel.• “PPPPEEEEAAAAKKKK” means the highest value of the channel since the peak detector was last

cleared.• “VVVVAAAALLLLLLLLEEEEYYYY” means the lowest value of the channel since the valley detector was last

cleared.

LLLLEEEEAVE MEAVE MEAVE MEAVE MENNNNUUUU Menu Item When the menu item is displayed, press [ENTER] to leave this menu. Press [EXIT] toreturn to the RUN mode.

page 76 008-0629-00

MA

TH

EM

AT

ICS

VIR

TU

AL

CH

AN

NE

L12

Chapter 12Mathematics Virtual Channel

12.1 Features

The Mathematics Virtual Channel is a powerful, flexible means to allow ACCU-GAGE instruments to perform simple to moderate functions that previouslyrequired the use of expensive programmable logic controllers (PLC).

A Mathematics Virtual Channel is akin to a PLC; it can execute programs writtenin an interpretive language called SensoCode. At the time of order, your specificapplication is analyzed by our sales and engineering staff and one or more Sen-soCode programs are created. These SensoCode programs are loaded into theMathematics Virtual channels of the instrument at the factory. The end result isan instrument that meets your specific needs but is built with proven hardwareand software.

SensoCode programs allow the instrument to perform a variety of functionsincluding:

• Mathematics : addition, subtraction, multiplication, division, square roots, etc.• Logical : if/then condition statements, comparison of values• Timing• Input : from any channel’s track, peak or valley values, open-collector inputs,

constant user-entered values from the SETUP menu• Output : Each Mathematics Channel generates its own track, peak, and valley

values. A SensoCode program can “take control” of the Limit Output and/orfront panel indicator lights.

ACCU-GAGE Instruction Manual page 77

12.2 Wiring

A Virtual Channel exists in software only; it does not occupy a physical slot inside ofthe instrument’s chassis. However, a Virtual Channel does require a channel number.

A SensoCode program that is executed by a Mathematics Virtual Channel can use theSystem connector’s Limit Output pins and Function Input pins for its own purposes.Consult the Customer Information Sheet included with the particular instrument fordetails.

page 78 008-0629-00

MA

TH

EM

AT

ICS

VIR

TU

AL

CH

AN

NE

L12

12.3 Setup Procedure

If you are not familiar with operating the instrument in the SETUP menu mode,see “SETUP Menu mode” on page19. A listing of all menu item is given in “SetupMenu Reference” on page91.

Step 1: Examine the Customer Information Sheet.The instrument’s Customer information Sheet contains important informa-tion about what each Mathematics Channel does and what Function Inputsand/or Limit Output pins of the System connector are used.

Step 2: Wire to the pins on the System connector.See “System Connector” on page 29. It is important to read the CustomerInformation Sheet to determine the wiring for your application.

Step 3: Verify operation.

ACCU-GAGE Instruction Manual page 79

12.4 Channel Menu

Each Mathematics Virtual Channel can have its configuration and SensoCode pro-gramming examined via its channel menu. It is not possible to alter the SensoCodeprogram with the SETUP menu. Detailed instructions on operating the instrument inthe SETUP Menu mode can be found in “SETUP Menu mode” on page19. A diagramof all menus is located in “Setup Menu Reference” on page91.

12.4.1 SSSSEEEENSOCODNSOCODNSOCODNSOCODEEEE P/NP/NP/NP/N

Menu ItemThis menu item displays the part number and revision level of the SensoCode pro-gram loaded in this Mathematics Virtual Channel.

12.4.2 USERUSERUSERUSER VVVVALUESALUESALUESALUES

Sub-MenuThis sub-menu contains menu items for each of the User Values. User Values areconstant numeric values that can be edited by the user and read by a SensoCode pro-gram. The use of each UserValue depends on the SensoCode program loaded in theMathematics Virtual Channel.

By default, the four items on this menu are UUUUSSSSEEEERRRR VALUVALUVALUVALUE 1E 1E 1E 1, UUUUSSSSEEEER VAR VAR VAR VALLLLUUUUE 2E 2E 2E 2, UUUUSSSSEEEERRRR VVVVAAAALLLLUUUUEEEE 3333 andUUUUSSSSEEEERRRR VVVVAAAALLLLUUUUE 4E 4E 4E 4. However, a SensoCode program can alter the menu item text for each ofthe User Values to make them more relevant to the operator. For example, UUUUSSSSEEEERRRR VVVVAAAALLLLUUUUEEEE

1111 might be re-titled JJJJEEEELLLLLLLLYYYY BEAN CBEAN CBEAN CBEAN COUNOUNOUNOUNTTTT.

12.4.3 VIEW SENSVIEW SENSVIEW SENSVIEW SENSOOOOCODECODECODECODE

Menu ItemThis menu item allows the display of each “step” (either an operand or an instruction)of the SensoCode program presently loaded into a Mathematics Virtual Channel.Pressing [ENTER] will display the next step of the SensoCode program until the entireprogram has been viewed.

SensoCode programs cannot be altered with the SETUP Mode menu; they can onlybe loaded via the serial port with the Instrument Utility Software.

12.4.4 VVVVIIIIEW COEW COEW COEW COMMMMMANDSMANDSMANDSMANDS

Sub-MenuThis sub-menu contains menu items for each of the Command Strings that can beused by a SensoCode program. Command Strings are text constants that can beread by a SensoCode program. They are not able to be changed with the SETUPmenu. The use of each Command String depends on the SensoCode programloaded in the Mathematics Virtual Channel.

The items on this menu are CCCCOMOMOMOMMMMMAAAANNNND 0D 0D 0D 0, CCCCOOOOMMMMMMMMAAAANNNND 1D 1D 1D 1 ,CCCCOOOOMMMMMMMMAAAANNNND 2D 2D 2D 2, CCCCOOOOMMMMMMMMAAAANNNND 3D 3D 3D 3, CCCCOOOOMMMMMMMMAAAANNNND 4D 4D 4D 4, CCCCOOOOMMMMMMMMAAAANNNND 5D 5D 5D 5,CCCCOOOOMMMMMMMMAAAANNNND 6D 6D 6D 6, and CCCCOOOOMMMMMMMMAAAANNNNDDDD 7777.

page 80 008-0629-00

MA

TH

EM

AT

ICS

VIR

TU

AL

CH

AN

NE

L12

12.4.5 DDDDIIIISPLAYSPLAYSPLAYSPLAY SSSSETUPETUPETUPETUP

Sub-MenuThis menu controls how data values are displayed by the channel and transmittedvia serial communications.

DDDDIIIISSSSPPPPLLLLAAAAYYYY.... DIGITDIGITDIGITDIGITSSSS Menu Item Selects the number of digits displayed by the channel. The choices are:

• “5555 DDDDIIIIGGGGIIIITTTT----BBBBIIIIPPPPOOOOLLLLAAAARRRR” displays both positive and negative numbers with five full dig-its (±99999).

• “6666 DIGIT-UNIPLOADIGIT-UNIPLOADIGIT-UNIPLOADIGIT-UNIPLOARRRR” displays positive numbers with six full digits (999999) andnegative numbers with five full digits (-99999).

• “7777 DIGIT UDIGIT UDIGIT UDIGIT UNNNNIIIIPPPPOOOOLLLLARARARAR”””” displays positive number with seven full digits (9999999)and negative numbers with six digits (-999999).

DDDDIIIISSSSPPPPLLLLAAAAYYYY.... DECPDECPDECPDECPTTTT Menu Item Selects the decimal point location on the channel’s display and serial communica-tions output. Use the [UP] and [DOWN] buttons to move the decimal point to theright and left.

DDDDIIIISSSSPPPPLLLLAAAAYYYY.... COUNT-BCOUNT-BCOUNT-BCOUNT-BYYYY Menu Item Determines by what value the display increments or decrements. This will makethe display less sensitive to a noisy signal. The choices are:

• “00000000000000001111”• “00000000000000002222”• “00000000000000005555”• “00000000000011110000”• “00000000000022220000”• “00000000000055550000”• “00000000111100000000”• “00000000222200000000”

DDDDIIIISSSSPPPPLLLLAAAAYYYY.... UNITUNITUNITUNITSSSS Menu Item Specifies the four character label that is displayed to the right of the channel’s val-ues. This menu item doesn’t change the mathematical scaling of the channel’svalues.

When a character position is flashing press the [UP] or [DOWN] button to changethe character. Press [ENTER] to advance to the next character.

DDDDIIIISSSSPPPPLLLLAAAAYYYY.... AVERAGAVERAGAVERAGAVERAGEEEE Menu Item Controls the speed with which the channel’s display values will update. Thechoices are:

• “OOOONNNN” means that the display will update four times each second. The channel’svalues will be averaged for 1/4 second, then displayed.

• “OOOOFFFFFFFF” means that the channel’s display will update as quickly as possible.

ACCU-GAGE Instruction Manual page 81

12.4.6 POWPOWPOWPOWEEEER-ONR-ONR-ONR-ON

SSSSOURCEOURCEOURCEOURCE Menu ItemThis menu selects which value is displayed by the channel when first entering theRUN mode.

The choices are:

• “TTTTRRRRAAAACCCCKKKK” means the live tracking value of the channel.• “PPPPEEEEAAAAKKKK” means the highest value of the channel.• “VVVVAAAALLLLLLLLEEEEYYYY” means the lowest value of the channel.

page 82 008-0629-00

MA

TH

EM

AT

ICS

VIR

TU

AL

CH

AN

NE

L12

12.5 Troubleshooting

See “Error Messages” on page85 for information relating to error messages.

The instrument’s Customer information Sheet contains important informa-tion about what indicator lights and which Function Input pins and/or Limit

Output pins of the System connector are used.

ACCU-GAGE Instruction Manual page 83

page 84 008-0629-00

ER

RO

RM

ES

SA

GE

S13

Chapter 13Error Messages

13.1 Overview

If the instrument detects an error during the RUN, INITIALIZE, or SETUP modes,it stops operation and enters its ERROR mode as described in “Operating Modes”on page15.

The most frequent causes of error messages are:

• Wiring/electrical connection problems• Misplaced address jumpers when installing or removing channel circuit

boards• Swapping channel circuit boards between instruments

ACCU-GAGE Instruction Manual page 85

13.2 Error Message List

Error 05, DDDDIIIIVVVVIIIIDDDDEEEE BY ZBY ZBY ZBY ZEEEERRRROOOO A Mathematics Virtual channel has attempted to divide by zero.

Error 07, SSSSQQQQUUUUAAAARRRREEEE ROOROOROOROOTTTT A Mathematics Virtual channel has attempted to take the square root of a negativenumber.

Error 09, LLLLOOOOGGGG////LLLLNNNN A Mathematics Virtual channel has attempted to take the logarithm of either zero or anegative number.

Error 20, CCCCHHHHAAAANNNNNNNNEEEELLLL >MAXC>MAXC>MAXC>MAXCHHHH An invalid channel number has been encountered by the instrument. Restart theinstrument. If this error persists, contact Sensotec with the information found in theSSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> SSSSOOOOFTFTFTFTWWWWAAAARRRREEEE REVREVREVREV.... menu item.

Error 21, BBBBLLLLOOOOWWWWNNNN STACSTACSTACSTACKKKK The instrument’s software stack has overflowed. Restart the instrument. If this errorpersists, contact Sensotec with the information found in the SSSSYYYYSSSSTTTTEEEEMMMM MMMMEEEENUNUNUNU -> SSSSOOOOFFFFTWTWTWTWARARARARE RE RE RE REEEEVVVV....

menu item.

Error 22, TTTTYYYYPPPPEEEE NOT FNOT FNOT FNOT FOOOOUUUUNNNNDDDD The instrument has not been programmed to understand the type of channel. ContactSensotec with the information found in the SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE REVREVREVREV.... menu item.

Error 23, CCCCAAAANNNN''''TTTT RENDER FRENDER FRENDER FRENDER FSSSS The full-scale value of the transducer cannot be rendered on the display with theselected DDDDIIIISSSSPPPPLLLLAAAAYYYY SSSSEEEETTTTUUUUPPPP menu settings. Change the DDDDIIIISSSSPPPPLLLLAAAAY SEY SEY SEY SETTTTUUUUPPPP menu settings to bettermatch the transducer’s capacity.

Error 24, SSSSIIIIGGGG....MMMMOOOODDDD CRC RCRC RCRC RCRC ROOOOMMMM Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 25, SSSSIIIIGGGG....MMMMOOOOD CD CD CD CRRRRC RC RC RC REEEEAAAADDDD Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 26, NNNNOOOO SIG.MODSIG.MODSIG.MODSIG.MOD.... Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 27, SSSS....EEEEEEEE WRITE NWRITE NWRITE NWRITE NAAAAKKKK The channel’s EEPROM could not be detected during a write operation. Check thatthe channel’s address jumpers are set properly so that each card has an uniqueaddress.

Error 28, SSSS....EEEEEEEE READ NREAD NREAD NREAD NAAAAKKKK The channel’s EEPROM could not be detected during a read operation. Check thatthe channel’s address jumpers are set properly.

Error 29, SSSS....EEEEE VE VE VE VEEEERRRRIIIIFFFFY FY FY FY FAAAAIIIILLLL A verification failure occurred during a write operation to the channel’s EEPROM.Restart the instrument.

Error 30, SSSS....EEEEEEEE ERASE NERASE NERASE NERASE NAAAAKKKK The channel’s EEPROM could not be detected during an erase operation. Check thatthe channel’s address jumpers are set properly.

Error 31, BBBBLLLLAAAANNNNKKKK IDENIDENIDENIDENTTTT The channel does not appear to have been installed properly. This occurs if youreplace Input or Output boards without using SSSSYYYYSSSSTTTTEEEEMMMM MMMMEEEENNNNUUUU -> ININININSSSSTTTTAAAALLLLL CL CL CL CHHHHAAAANNNNNNNNEEEELLLL to formatthe channel’s EEPROM.

Error 32, BBBBOOOOGGGGUUUUSSSS EE AEE AEE AEE ADDDDDDDDRRRR An invalid address has been passed to the programmer. Restart the instrument. If thisproblem persists, contact Sensotec with the information found in the SSSSYYYYSSSSTTTTEEEEMMMM MMMMEEEENNNNUUUU ->SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE REVREVREVREV.... menu item.

page 86 008-0629-00

ER

RO

RM

ES

SA

GE

S13

Error 33, SSSSIIIIGGGG....MMMMOOOODDDD WIREWIREWIREWIRESSSS Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 34, SSSSIIIIGGGG....MMMMOOOODDDD MISMATCMISMATCMISMATCMISMATCHHHH Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 35, BBBBOOOOGGGGUUUUSSSS MENU TMENU TMENU TMENU TYYYYPPPPEEEE A menu contains an invalid menu type. Contact Sensotec with the informationfound in the SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE REVREVREVREV.... menu item.

Error 36, BBBBAAAADDDD EE TEE TEE TEE TAAAABBBBLLLLEEEE The instrument doesn’t understand the EEPROM type setting for the channel.Contact Sensotec with the information found in the SSSSYYYYSSSSTTTTEEEEMMMM MMMMEEEENNNNUUUU -> SSSSOOOOFTFTFTFTWWWWAAAARRRRE RE RE RE REEEEVVVV....

menu item.

Error 37, BBBBAAAADDDD PERSONALITPERSONALITPERSONALITPERSONALITYYYY The channel doesn’t understand the given card type. Contact Sensotec with theinformation found in the SSSSYYYYSSSSTTTTEEEEMMMM MMMMEEEENNNNUUUU -> SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE REVREVREVREV.... menu item.

Error 38, BBBBAAAADDDD FORMAFORMAFORMAFORMATTTT The channel’s EEPROM memory contains a display setting which is not under-stood by the instrument. Check that every item in the channel’s DDDDIIIISSSSPPPPLLLLAAAAY SY SY SY SEEEETTTTUUUUPPPP

menu and the DDDDISISISISPPPPLLLLAAAAYYYY UNITUNITUNITUNITSSSS menu item has a valid setting.

Error 39, WWWWRRRROOOONNNNGGGG HARDWARHARDWARHARDWARHARDWAREEEE The channel type expected by the instrument for the channel number is not thechannel type that is physically installed. If you have changed the address jumperson one or more channels, replace them in their original configuration.

Error 41, SSSSHHHHUUUUNNNNTTTT CAL ECAL ECAL ECAL ERRRRRRRROOOORRRR Internal error. Contact Sensotec with the information found in the SSSSYYYYSSSSTTTTEEEEM MM MM MM MEEEENNNNUUUU ->SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE REVREVREVREV.... menu item.

Error 43, BBBBAAAADDDD IDENT CIDENT CIDENT CIDENT CRRRRCCCC A checksum failure has occurred when reading a channel’s EEPROM. Restartthe instrument.

Error 44, BBBBAAAADDDD FUNCTIOFUNCTIOFUNCTIOFUNCTIONNNN The instrument has requested that the channel perform a function that it is notcapable of. Restart the instrument.

Error 45, AAAA////DDDD NANANANAKKKK Hardware error. The channel’s Analog-to-Digital converter is not communicatingwith the channel’s microprocessor. Restart the instrument.

Error 46, SSSSEEEENNNNSSSSEEEE WIREWIREWIREWIRESSSS Do not connect anything to the top 6 pins (labeled as “N/C”) of the Pressure Inputchannel’s 12-pin connector.

Error 47, BBBBAAAADDDD EE TEE TEE TEE TYYYYPPPPEEEE RRRREEEEQQQQ When installing a channel, the EEPROM type requested by the channel is notvalid. Contact Sensotec with the information found in the SSSSYYYYSSSSTTTTEEEEMMMM MMMMEEEENNNNUUUU -> SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE

RRRREEEEVVVV.... menu item.

Error 48, NNNNUUUULLLLLLLL MENMENMENMENUUUU A menu contains a NULL item. Contact Sensotec with the information found inthe SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE REVREVREVREV.... menu item.

Error 49, NNNNUUUULLLLLLLL ACTION FACTION FACTION FACTION FCCCCNNNN A menu contains a NULL action function. Contact Sensotec with the informationfound in the SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE REVREVREVREV.... menu item.

Error 50, EEEEXXXXCCCCIIIITTTTAAAATTTTIIIIOOOONNNN The channel’s excitation voltage regulator has shut down because the protectioncircuit that limits the output current has activated. Do not connect anything to thetop 6 pins (labeled as “N/C”) of the Pressure Input channel’s 12-pin connector.

Error 51, CCCCAAAANNNN''''TTTT DO EDO EDO EDO EXXXXCCCC.... The channel’s EEPROM memory contains an excitation setting which is notunderstood by the instrument. Contact Sensotec with the information found in theSSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> SSSSOOOOFTFTFTFTWAWAWAWARERERERE REVREVREVREV.... menu item.

Error 52, SSSSIIIIGGGG....MMMMOOOODDDD LENGTHLENGTHLENGTHLENGTH RRRR Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

ACCU-GAGE Instruction Manual page 87

Error 53, SSSSIIIIGGGG....MMMMOOOODDDD RECTYPRECTYPRECTYPRECTYPEEEE Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 54, SSSSIIIIGGGG....MMMMOOOOD LD LD LD LEEEENNNNGGGGTTTTH WH WH WH W Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 55, SSSSIIIIGGGG....MMMMOOOODDDD VERIFVERIFVERIFVERIFYYYY

SSSSPPPP

Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 56, SSSSIIIIGGGG....MMMMOOOODDDD VERIFVERIFVERIFVERIFYYYY Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 57, WWWWRRRROOOONNNNGGGG SIG.MODSIG.MODSIG.MODSIG.MOD.... Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 58, WWWWRRRROOOONNNNG EG EG EG EE VE VE VE VEEEERRRRSSSSIIIIOOOONNNN The channel’s EEPROM memory data is not able to be used by this version of firm-ware. Make certain that you have not accidentally swapped channel circuit boardsbetween instruments. If you are certain that you want to use the channel circuit boardwith this instrument, you may reset the channel’s EEPROM memory data to factorydefaults with the SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU -> DDDDEEEEFFFFAAAAUUUULTLTLTLT CHANNECHANNECHANNECHANNELLLL menu item.

Error 59, CCCCAAAANNNN’’’’T DT DT DT DO SO SO SO SIIIIGGGG....MMMMOOOODDDD Do not connect anything to pin 9 (labeled as “N/C”) of the 12-pin connector.

Error 60, SSSSLLLLAAAAVVVVEEEE TIMEOUTIMEOUTIMEOUTIMEOUTTTT The channel’s microprocessor did not respond because the channel’s analog-to-digi-tal converter did not calibrate properly. Restart the instrument.

Error 61, LLLLOOOOAAAAD ND ND ND NOOOOT GT GT GT GT LT LT LT LOOOOAAAADDDD A load applied during a known-load calibration was not greater than a previouslyapplied load. To insure both correct operation of the transducer and application of theload, the instrument expects the voltage applied at each known-load point to beincreasing. For example, the load applied at Known-Load Point 2/5 must cause thetransducer to produce a more positive voltage than at Known-Load Point 1/5.

Error 62, SSSSLLLLAAAAVVVVEEEE NANANANAKKKK The channel’s microprocessor is responding to a message, but it did not acknowledgeit correctly. Restart the instrument.

Error 67, LLLLOOOOAAAADDDDSSSS NOT CNOT CNOT CNOT CLLLLOOOOSSSSEEEE During a 5-Point Known-Load calibration, you would typically apply loads at 0%, 25%,50%, 75% and 100% of the transducer’s full-scale and optionally calibrate the returnpoints of 75%, 50%, 25% and 0%. The instrument makes sure that transducer read-ings taken during the second application of the load are close to the readings takenduring the first application of the load. This insures both correct operation of the trans-ducer and application of the loads.

Error 68, CCCCAAAANNNN''''TTTT CALIBRATCALIBRATCALIBRATCALIBRATEEEE The instrument has detected that the transducer presently connected to the channel isdifferent than the transducer the channel was last calibrated with. Either re-connectthe original transducer or use the SETUP menu to perform a calibration.

Error 70, CCCCAAAANNNN''''TTTT DO GDO GDO GDO GAAAAIIIINNNN Consult Sensotec.

Error 71, SSSSLLLLAAAAVVVVEEEE CHECKSUCHECKSUCHECKSUCHECKSUMMMM The channel’s microprocessor did not send a correct checksum to the instrument.Restart the instrument.

Error 72, CCCCAAAANNNN''''TTTT DO FDO FDO FDO FIIIILLLLTTTTEEEERRRR The channel’s EEPROM memory contains a frequency response setting which is notunderstood by the instrument. Consult Sensotec.

Error 73, SSSSLLLLAAAAVVVVEEEE NOT INOT INOT INOT INNNNIIIITTTT The channel’s microprocessor has not yet been initialized by the instrument. Restartthe instrument.

Error 74, SSSSLLLLAAAAVVVVEEEE BAD RBAD RBAD RBAD REEEEPPPPLLLLYYYY The channel’s microprocessor replied with a message that was either not understoodor different than expected. Restart the instrument.

Error 75, SSSSLLLLAAAAVVVVEEEE NOT FNOT FNOT FNOT FOOOOUUUUNNNNDDDD The channel’s microprocessor did not respond. Restart the instrument.

page 88 008-0629-00

ER

RO

RM

ES

SA

GE

S13

Error 81, OOOOPPPPEEEERRRRAAAANNNNDDDD EEEEXXXXPPPPEEEECCCCTTTTEEEEDDDD The first step of a SensoCode program must be an Operand. The SensoCode pro-gram loaded in the Mathematics Virtual channel is invalid.

Error 82, BBBBAAAADDDD OPERANOPERANOPERANOPERANDDDD The SensoCode program loaded into a Mathematics Virtual channel contains a badoperand.

Error 83, BBBBAAAADDDD INSTRUCTIOINSTRUCTIOINSTRUCTIOINSTRUCTIONNNN The SensoCode program loaded into a Mathematics Virtual channel contains a badinstruction.

Error 84, BBBBAAAADDDD VALUVALUVALUVALUEEEE A Mathematics Virtual channel has attempted to get a value from a channel notinstalled in the instrument.

Error 85, UUUUNNNNRRRREEEEAAAACCCCHHHHAAAABBBBLLLLEEEE GGGGOOOOTTTTOOOO The destination step of a “Go to” Instruction is beyond the end of the SensoCode pro-gram. The SensoCode program loaded in the Mathematics Virtual channel is invalid.

Error 87, GGGGOOOOTTTTOOOO NOT JNOT JNOT JNOT JUUUUMMMMPPPP The SensoCode program contains a “Go to” Instruction that has not been convertedinto a “Jump” Instruction. Restart the instrument.

ACCU-GAGE Instruction Manual page 89

page 90 008-0629-00

SE

TU

PM

EN

UR

EF

ER

EN

CE

14

Chapter 14Setup Menu Reference

14.1 Navigation instructions

• Press [UP]+[DOWN] to enter the SETUP menu mode.• Press [UP] to move up.• Press [DOWN] to move down.• Press [ENTER] to select an item.• Press [EXIT] to re-start the instrument.

ACCU-GAGE Instruction Manual page 91

Table 14-1: SETUP Menu

CCCCHHHHAAAANNNNNNNNEEEELLLL 01 M01 M01 M01 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 02 M02 M02 M02 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 03 M03 M03 M03 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 04 M04 M04 M04 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 05 M05 M05 M05 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 06 M06 M06 M06 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 07 M07 M07 M07 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 08 M08 M08 M08 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 09 M09 M09 M09 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 10 M10 M10 M10 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 11 M11 M11 M11 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 12 M12 M12 M12 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 13 M13 M13 M13 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 14 M14 M14 M14 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 15 M15 M15 M15 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 16 M16 M16 M16 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 17 M17 M17 M17 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 18 M18 M18 M18 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 19 M19 M19 M19 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 20 M20 M20 M20 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 21 M21 M21 M21 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 22 M22 M22 M22 MENENENENUUUU

(menu varies according to channel type)CCCCHHHHAAAANNNNNNNNEEEELLLL 23 M23 M23 M23 MENENENENUUUU

(menu varies according to channel type)LLLLIIIIMMMMIIIITTTT 01 M01 M01 M01 MENENENENUUUU

(see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 02 M02 M02 M02 MENENENENUUUU

(see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 03 M03 M03 M03 MENENENENUUUU

(see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 04 M04 M04 M04 MENENENENUUUU

(see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 05 M05 M05 M05 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 06 M06 M06 M06 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 07 M07 M07 M07 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 08 M08 M08 M08 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 09 M09 M09 M09 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 10 M10 M10 M10 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 11 M11 M11 M11 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 12 M12 M12 M12 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 13 M13 M13 M13 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 14 M14 M14 M14 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 15 M15 M15 M15 MENENENENUUUU

(optional) (see LIMIT MENU)LLLLIIIIMMMMIIIITTTT 16 M16 M16 M16 MENENENENUUUU

(optional) (see LIMIT MENU)SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU

(see SYSTEM MENU)SSSSEEEERRRRIIIIAAAALLLL COM MCOM MCOM MCOM MENENENENUUUU

(see SERIAL COM MENU)DDDDIIIISSSSPPPPLLLLAAAAYYYY MENMENMENMENUUUU

(see DISPLAY MENU)

page 92 008-0629-00

SE

TU

PM

EN

UR

EF

ER

EN

CE

14

Table 14-2: System Men u

SSSSYYYYSSSSTTTTEEEEMMMM MENMENMENMENUUUU SSSSOOOOFFFFTTTTWWWWAAAARRRREEEE REVISIOREVISIOREVISIOREVISIONNNN page 35

CCCCOOOONNNNFFFFIIIIGGGGUUUURRRRAAAATTTTIIIIOOOONNNN DDDDIIIISSSSPPPPLLLLAAAAYYYY TYPTYPTYPTYPEEEE page 35

CCCCHHHHAAAANNNNNNNNEEEELLLL 01 T01 T01 T01 TYPYPYPYPEEEE

CCCCHHHHAAAANNNNNNNNEEEELLLL 02 T02 T02 T02 TYPYPYPYPEEEE

CCCCHHHHAAAANNNNNNNNEEEELLLL 03 T03 T03 T03 TYPYPYPYPEEEE

............

CCCCHHHHAAAANNNNNNNNEEEELLLL 22 T22 T22 T22 TYPYPYPYPEEEE

CCCCHHHHAAAANNNNNNNNEEEELLLL 23 T23 T23 T23 TYPYPYPYPEEEE

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

DDDDIIIIAAAAGGGGNNNNOOOOSSSSTTTTIIIICCCCSSSS OOOOUUUUTTTTPPPPUUUUTTTT 1, P1, P1, P1, PIIIIN 1N 1N 1N 14444 OOOONNNN page 35

OOOOFFFFFFFF

OOOOUUUUTTTTPPPPUUUUTTTT 2, P2, P2, P2, PIIIIN 1N 1N 1N 15555 OOOONNNN

OOOOFFFFFFFF

OOOOUUUUTTTTPPPPUUUUTTTT 3, P3, P3, P3, PIIIIN 1N 1N 1N 16666 OOOONNNN

OOOOFFFFFFFF

OOOOUUUUTTTTPPPPUUUUTTTT 4, P4, P4, P4, PIIIIN 1N 1N 1N 17777 OOOONNNN

OOOOFFFFFFFF

IIIINNNNPPPPUUUUTTTT TESTESTESTESTTTT page 35

SSSSCCCCAAAANNNN TIMTIMTIMTIMEEEE page 35

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

IIIINNNNSSSSTTTTAAAALLLLLLLL CHANNECHANNECHANNECHANNELLLL 66665555 STRAIN GSTRAIN GSTRAIN GSTRAIN GAAAAGEGEGEGE VVVV page 36

66666666 STRAIN GSTRAIN GSTRAIN GSTRAIN GAAAAGEGEGEGE IIII

66667777 HIGH-LEVEHIGH-LEVEHIGH-LEVEHIGH-LEVEL VL VL VL V

66668888 HIGH-LEVEHIGH-LEVEHIGH-LEVEHIGH-LEVEL IL IL IL I

66669999 RTD VRTD VRTD VRTD V OOOOUUUUTTTT

6666AAAA RTD IRTD IRTD IRTD I OOOOUUUUTTTT

6666BBBB LVDT V OULVDT V OULVDT V OULVDT V OUTTTT

6666CCCC LVDT I OULVDT I OULVDT I OULVDT I OUTTTT

6666DDDD PRESSURE V OUPRESSURE V OUPRESSURE V OUPRESSURE V OUTTTT

6666EEEE PRESSURE I OUPRESSURE I OUPRESSURE I OUPRESSURE I OUTTTT

6666FFFF FREQUENCY VFREQUENCY VFREQUENCY VFREQUENCY V OOOOUUUUTTTT

77770000 FREQUENCY IFREQUENCY IFREQUENCY IFREQUENCY I OOOOUUUUTTTT

99990000 LIMIT 0LIMIT 0LIMIT 0LIMIT 01111----00004444

99991111 LIMIT 0LIMIT 0LIMIT 0LIMIT 05555----00008888

99992222 LIMIT 0LIMIT 0LIMIT 0LIMIT 09999----11112222

99993333 LIMIT 1LIMIT 1LIMIT 1LIMIT 13333----11116666

AAAABBBB SPLIT DSPLIT DSPLIT DSPLIT DIIIISSSSPPPPLLLLAAAAYYYY

AAAAEEEE MATHEMATICMATHEMATICMATHEMATICMATHEMATICSSSS

DDDDEEEELLLLEEEETTTTEEEE CHANNECHANNECHANNECHANNELLLL page 37

LLLLEEEEAAAAVVVVEEEE MENMENMENMENUUUU

ACCU-GAGE Instruction Manual page 93

Table 14-3: LIMIT MENU

LLLLIIIIMMMMIIIITTTT xx Mxx Mxx Mxx MENENENENUUUU LLLLIIIIMMMMIIIITTTT.... ENABLENABLENABLENABLEEEE OOOONNNN page 57

OOOOFFFFFFFF

LLLLIIIIMMMMIIIITTTT.... SETPOINSETPOINSETPOINSETPOINTTTT page 57

LLLLIIIIMMMMIIIITTTT.... RETURN PNRETURN PNRETURN PNRETURN PNTTTT page 57

LLLLIIIIMMMMIIIITTTT.... ENERGIZENERGIZENERGIZENERGIZEEEE SSSSIIIIGGGGNNNNAAAALLLL>>>>SSSSEEEETTTTPPPPOOOOIIIINNNNTTTT page 57

SSSSIIIIGGGGNNNNAAAALLLL<<<<SSSSEEEETTTTPPPPOOOOIIIINNNNTTTT

SSSSIIIIGGGGNNNNAAAALLLL INSIDINSIDINSIDINSIDEEEE

SSSSIIIIGGGGNNNNAAAALLLL OUTSIDOUTSIDOUTSIDOUTSIDEEEE

LLLLIIIIMMMMIIIITTTT.... LATCHINLATCHINLATCHINLATCHINGGGG OOOONNNN page 58

OOOOFFFFFFFF

LLLLIIIIMMMMIIIITTTT.... CHANNECHANNECHANNECHANNELLLL page 59

LLLLIIIIMMMMIIIITTTT.... SOURCSOURCSOURCSOURCEEEE TTTTRRRRAAAACCCCKKKK page 59

PPPPEEEEAAAAKKKK

VVVVAAAALLLLLLLLEEEEYYYY

LLLLEEEEAAAAVVVVEEEE MENMENMENMENUUUU

Table 14-4: SERIAL COMMUNICATIONS Menu

SSSSEEEERRRRIIIIAAAALLLL COM MCOM MCOM MCOM MENENENENUUUU IIIINNNNTTTTEEEERRRRFFFFAAAACCCCEEEE page 41

AAAADDDDDDDDRRRREEEESSSSSSSS page 41

BBBBAAAAUUUUDDDD RATRATRATRATEEEE page 41

AAAAUUUUTTTTOOOO LINE-FEELINE-FEELINE-FEELINE-FEEDDDD page 41

TTTTRRRRAAAANNNNSSSSMMMMIIIITTTT TESTESTESTESTTTT page 41

LLLLEEEEAAAAVVVVEEEE MENMENMENMENUUUU

Table 14-5: DISPLAY Menu

DDDDIIIISSSSPPPPLLLLAAAAYYYY MENMENMENMENUUUU UUUUPPPPPPPPEEEERRRR CHANNECHANNECHANNECHANNELLLL page 53 LLLLOOOOWWWWEEEERRRR CHANNECHANNECHANNECHANNELLLL page 53

LLLLOOOOWWWWEEEERRRR MODMODMODMODEEEE BBBBLLLLAAAANNNNKKKK page 53 LLLLIIIIMMMMIIIITTTTSSSS 01-001-001-001-04444

CCCCHHHHAAAANNNNNNNNEEEELLLL

DDDDIIIISSSSPPPPLLLLAAAAYYYY DISABLDISABLDISABLDISABLEEEE OOOONNNN page 53

OOOOFFFFFFFF

LLLLEEEEAAAAVVVVEEEE MENMENMENMENUUUU

page 94 008-0629-00

SE

TU

PM

EN

UR

EF

ER

EN

CE

14

Table 14-6: Pressure Input C hannel Men u

CCCCHHHHAAAANNNNNNNNEEEELLLL xx Mxx Mxx Mxx MENENENENUUUU OOOOPPPPEEEERRRRAAAATTTTIIIIOOOONNNN AAAAUUUUTTTTOOOO----ZZZZEEEERRRROOOO OOOONNNN page 65

OOOOFFFFFFFF

PPPPOOOOWWWWEEEERRRR----OOOONNNN SOURCSOURCSOURCSOURCEEEE TTTTRRRRAAAACCCCKKKK page 65

PPPPEEEEAAAAKKKK

VVVVAAAALLLLLLLLEEEEYYYY

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

DDDDIIIISSSSPPPPLLLLAAAAYYYY UNITUNITUNITUNITSSSS PPPPSSSSIIII to Pto Pto Pto PSSSSIIII page 66

PPPPSSSSIIII to bto bto bto baaaarrrr

PPPPSSSSIIII to mto mto mto mbabababarrrr

PPPPSSSSIIII to tto tto tto tororororrrrr

PPPPSSSSIIII to ito ito ito inWnWnWnWAAAA

PPPPSSSSIIII to fto fto fto ftWtWtWtWAAAA

PPPPSSSSIIII to ito ito ito inHnHnHnHgggg

PPPPSSSSIIII to mto mto mto mmHmHmHmHgggg

PPPPSSSSIIII to kto kto kto kPPPPaaaa

PPPPSSSSIIII to Mto Mto Mto MPPPPaaaa

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

AAAAUUUUXXXX1111 FUNCTIOFUNCTIOFUNCTIOFUNCTIONNNN DDDDIIIISSSSAAAABBBBLLLLEEEEDDDD page 66

TTTTRRRRAAAACCCCKKKK HOLHOLHOLHOLDDDD

HHHHIIIIGGGGHHHH////LLLLOOOOWWWW HOLHOLHOLHOLDDDD

HHHHIIIIGGGGHHHH////LLLLOOOOWWWW CLEACLEACLEACLEARRRR

TTTTAAAARRRREEEE OOOONNNN

TTTTAAAARRRREEEE OFOFOFOFFFFF

AAAAUUUUXXXX2222 FUNCTIOFUNCTIOFUNCTIOFUNCTIONNNN DDDDIIIISSSSAAAABBBBLLLLEEEEDDDD page 66

TTTTRRRRAAAACCCCKKKK HOLHOLHOLHOLDDDD

HHHHIIIIGGGGHHHH////LLLLOOOOWWWW HOLHOLHOLHOLDDDD

HHHHIIIIGGGGHHHH////LLLLOOOOWWWW CLEACLEACLEACLEARRRR

TTTTAAAARRRREEEE OOOONNNN

TTTTAAAARRRREEEE OFOFOFOFFFFF

CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN TYPTYPTYPTYPEEEE TTTTYYYYPPPPEEEE====3333 POINT CPOINT CPOINT CPOINT CAAAALLLL page 67

TTTTYYYYPPPPEEEE====5555 POINT CPOINT CPOINT CPOINT CAAAALLLL

CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTIIIIOOOONNNN DATDATDATDATAAAA FFFFUUUULLLLLLLL----SSSSCCCCAAAALLLLEEEE MV/MV/MV/MV/VVVV page 68

KKKKNNNNOOOOWWWWNNNN LOAD PLOAD PLOAD PLOAD POOOOIIIINNNNTTTT 1111////5555

KKKKNNNNOOOOWWWWNNNN LOAD PLOAD PLOAD PLOAD POOOOIIIINNNNTTTT 2222////5555

KKKKNNNNOOOOWWWWNNNN LOAD PLOAD PLOAD PLOAD POOOOIIIINNNNTTTT 3333////5555

KKKKNNNNOOOOWWWWNNNN LOAD PLOAD PLOAD PLOAD POOOOIIIINNNNTTTT 4444////5555

KKKKNNNNOOOOWWWWNNNN LOAD PLOAD PLOAD PLOAD POOOOIIIINNNNTTTT 5555////5555

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

CCCCAAAALLLLIIIIBBBBRRRRAAAATTTTEEEE page 68

DDDDAAAACCCC SETUSETUSETUSETUPPPP DDDDAAAACCCC.... CHANNECHANNECHANNECHANNELLLL page 70

DDDDAAAACCCC.... SOURCSOURCSOURCSOURCEEEE TTTTRRRRAAAACCCCKKKK

PPPPEEEEAAAAKKKK

VVVVAAAALLLLLLLLEEEEYYYY

DDDDAAAACCCC.... ZERO-SCALZERO-SCALZERO-SCALZERO-SCALEEEE

DDDDAAAACCCC.... FULL-SCALFULL-SCALFULL-SCALFULL-SCALEEEE

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

ACCU-GAGE Instruction Manual page 95

DDDDIIIIAAAAGGGGNNNNOOOOSSSSTTTTIIIICCCCSSSS DDDDAAAACCCC FULL SFULL SFULL SFULL SCCCCAAAALLLLEEEE page 71

DDDDAAAACCCC ZERO SZERO SZERO SZERO SCCCCAAAALLLLEEEE

VVVVEEEERRRRSSSSIIIIOOOONNNN INFINFINFINFOOOO

DDDDIIIISSSSPPPPLLLLAAAAYYYY ADADADADCCCC OOOONNNN

OOOOFFFFFFFF

LLLLIIIINNNNEEEEAAAARRRRIIIIZZZZAAAATTTTIIIIOOOONNNN OOOONNNN

OOOOFFFFFFFF

DDDDIIIISSSSAAAABBBBLLLLEEEE CHANNECHANNECHANNECHANNELLLL OOOONNNN

OOOOFFFFFFFF

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

LLLLEEEEAAAAVVVVEEEE MENMENMENMENUUUU

Table 14-6: Pressure Input Channel Menu (Continued)

Table 14-7: Split Display Virtual Channel

CCCCHHHHAAAANNNNNNNNEEEELLLL xx Mxx Mxx Mxx MENENENENUUUU LLLLEEEEFFFFTTTT CHANNECHANNECHANNECHANNELLLL page 76 LLLLEEEEFFFFTTTT SOURCSOURCSOURCSOURCEEEE TTTTRRRRAAAACCCCKKKK page 76

PPPPEEEEAAAAKKKK

VVVVAAAALLLLLLLLEEEEYYYY

RRRRIIIIGGGGHHHHTTTT CHANNECHANNECHANNECHANNELLLL page 76 RRRRIIIIGGGGHHHHTTTT SOURCSOURCSOURCSOURCEEEE TTTTRRRRAAAACCCCKKKK page 76

PPPPEEEEAAAAKKKK

VVVVAAAALLLLLLLLEEEEYYYY

LLLLEEEEAAAAVVVVEEEE MENMENMENMENUUUU

page 96 008-0629-00

SE

TU

PM

EN

UR

EF

ER

EN

CE

14

Table 14-8: Mathematics Virtual Channel

CCCCHHHHAAAANNNNNNNNEEEELLLL xx Mxx Mxx Mxx MENENENENUUUU SSSSEEEENNNNSSSSOOOOCCCCOOOODDDDEEEE P/P/P/P/NNNN page 80

UUUUSSSSEEEERRRR VALUEVALUEVALUEVALUESSSS UUUUSSSSEEEERRRR VALUEVALUEVALUEVALUE 1111 (item name may vary) page 80 UUUUSSSSEEEERRRR VALUEVALUEVALUEVALUE 2222 (item name may vary)

UUUUSSSSEEEERRRR VALUEVALUEVALUEVALUE 3333 (item name may vary)

UUUUSSSSEEEERRRR VALUEVALUEVALUEVALUE 4444 (item name may vary)

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

VVVVIIIIEEEEWWWW SENSOCODSENSOCODSENSOCODSENSOCODEEEE page 80

VVVVIIIIEEEEWWWW COMMANDCOMMANDCOMMANDCOMMANDSSSS CCCCOOOOMMMMMMMMAAAANNNND 0D 0D 0D 0 page 80 CCCCOOOOMMMMMMMMAAAANNNND 1D 1D 1D 1

CCCCOOOOMMMMMMMMAAAANNNND 2D 2D 2D 2

CCCCOOOOMMMMMMMMAAAANNNND 3D 3D 3D 3

CCCCOOOOMMMMMMMMAAAANNNND 4D 4D 4D 4

CCCCOOOOMMMMMMMMAAAANNNND 5D 5D 5D 5

CCCCOOOOMMMMMMMMAAAANNNND 6D 6D 6D 6

CCCCOOOOMMMMMMMMAAAANNNND 7D 7D 7D 7

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

DDDDIIIISSSSPPPPLLLLAAAAYYYY SETUSETUSETUSETUPPPP DDDDIIIISSSSPPPPLLLLAAAAYYYY.... DIGITDIGITDIGITDIGITSSSS 5555 DIGIT-BIPOLADIGIT-BIPOLADIGIT-BIPOLADIGIT-BIPOLARRRR page 81

6666 DIGIT-UNIPOLADIGIT-UNIPOLADIGIT-UNIPOLADIGIT-UNIPOLARRRR

7777 DIGIT-UNIPOLADIGIT-UNIPOLADIGIT-UNIPOLADIGIT-UNIPOLARRRR

DDDDIIIISSSSPPPPLLLLAAAAYYYY.... DECPDECPDECPDECPTTTT 99999999999999999999.... page 81 9999999999999999....9999

999999999999....99999999

99999999....999999999999

9999....9999999999999999

....99999999999999999999

DDDDIIIISSSSPPPPLLLLAAAAYYYY.... COUNTBCOUNTBCOUNTBCOUNTBYYYY 00000000000000001111 page 81 00000000000000002222

00000000000000005555

00000000000011110000

00000000000022220000

00000000000055550000

00000000111100000000

00000000222200000000

DDDDIIIISSSSPPPPLLLLAAAAYYYY.... UNITUNITUNITUNITSSSS page 81 DDDDIIIISSSSPPPPLLLLAAAAYYYY.... AVERAGAVERAGAVERAGAVERAGEEEE OOOONNNN page 81

OOOOFFFFFFFF

LLLLEEEEAAAAVVVVEEEE SUB-MENSUB-MENSUB-MENSUB-MENUUUU

PPPPOOOOWWWWEEEERRRR----OOOONNNN SOURCSOURCSOURCSOURCEEEE TTTTRRRRAAAACCCCKKKK page 82 PPPPEEEEAAAAKKKK

VVVVAAAALLLLLLLLEEEEYYYY

LLLLEEEEAAAAVVVVEEEE MENMENMENMENUUUU

ACCU-GAGE Instruction Manual page 97

page 98 008-0629-00

IND

EX

Index

Numerics3-Point Known Load Calibration 67–695-Point Known Load Calibration 67–69

AACCU-GAGE

features 12Analog Output 70–73auto-zero function 65Auxilliary Function pins 66, 73

Bbuttons

CHANNEL 17CLEAR 16, 58DOWN 19ENTER 19EXIT 19TARE 17UP 19VALUE 16

Ccalibration 63case removal 24Certificate of Calibration 11channel

changing which is displayed 17, 53clearing peak and valley 16deleting last in instrument 37installation 36list of all in instrument 35reset to default settings 37

CHANNEL button 17Chassis 21CLEAR button 16, 58Communications Guide 11, 39connectors

System 30Customer Information Sheet 11

Ddisplay

averagingMathematics Virtual channel 81

changing channel shown on 17, 53decimal points

Mathematics Virtual channel 81Pressure Input channel 66

disabling 53lower line 16, 53OVLD message 73percentage of Analog-to-Digital converter input 16symbols 16track, peak or valley 16units of measure

Mathematics Virtual channel 81Pressure Input channel 66

DOWN button 19

EENTER button 19error messages 85ERROR mode 18EXIT button 19

Ffrequency response 64fuse replacement 26

Iindicator lights

limits 56lower line display of 53Tare 17

INITIALIZE mode 15installation of channels 36

LLEDs. See indicator lights.lights. See indicator lights.limits

clearing latched limits 16enabling 57indicator lights 17latching 58Limits 01-04 56open-collector outputs 32operation 56–57, 59scan time 56

ACCU-GAGE Instruction Manual page 99

SETUP menu 57wiring 30

MMathematics Virtual channel 77

display setup 81operation 79SensoCode 77

part number 80viewing program 80

SETUP menu 80troubleshooting 83user values 80wiring 78

menu. See SETUP menu mode.

Ooperating modes 15

ERROR 18INITIALIZE 15RUN 16SETUP menu 19

OVLD message 73

Ppanel mounting 23Peak/Valley values

displaying 16remote clearing of all 30

Pressure Input channelSETUP menu 65specifications 64

pressure port 13

Qquick start 13

Rrack mounting 24remote control of Track/Hold, Peak/Valley clear, Peak/Valley hold or Tare 66, 73remote control. See Auxilliary Function pins 66RS-232

wiring 30RS-232. See Serial CommunicationsRS-485

wiring 30RS-485. See Serial CommunicationsRUN mode 16

SSC3004 75SensoCode 77

part number 80software revision 80viewing 80

Serial Communications 39communications parameters 40–41RS-232 defined 40RS-485 defined 40

Serial Communications Guide 11SETUP menu mode

navigation instructions 91overview 19reference 92

SETUP Protection Jumper 26software revision

chassis 35Mathematics Virtual channel 80Pressure Input channel 71SensoCode program 80

specificationsChassis Models 22Pressure Input channel 64

Split Display Virtual channel 75display formatting 75SETUP menu 76

step response 64System Connector

diagnostics 35pinout 30

TTare

activating or deactivating 17remote of all channels 30See also Auxilliary Function pins.

Tare light 17troubleshooting 73

UUP button 19user values 80

VVALUE button 16

Wwiring

Mathematics Virtual channel 78Pressure Input channel 62

page 100 008-0629-00

WA

RR

AN

TY

/RE

PA

IRP

OLIC

Y

Warranty/Repair Policy

Limitation of Remedy andDisclaimer of Warranty

Any of our products which, under normal operating conditions, proves defective inmaterial or in workmanship within one year from the date of shipment by Senso-tec, Inc., will be repaired or replaced free of charge, provided that the buyer (1)promptly notifies Sensotec, Inc. of any such defect; (2) provides Sensotec, Inc.with satisfactory proof of the defect and that the product was properly installed,maintained, and operated within the limits of rated and normal usage; and (3)obtains from Sensotec, Inc. authorization to return the product. Any such productshall be returned with transportation charges prepaid. The replacement productwill be shipped F.O.B. our plant.

The remedy set forth herein does not extend to any product or part thereof which,under normal usage, has an inherently shorter useful life than one year. The rem-edy set forth herein does not apply to damage or to defects in any product causedby the buyer’s misuse or neglect, nor does it apply to any product which has beenrepaired or disassembled which, in the sole judgement of Sensotec, Inc. affectsthe performance of the product.

The remedy set forth herein is the buyer's exclusive remedy, and will satisfy allobligations of Sensotec, Inc. whether based on contract, negligence, or otherwise.Sensotec, Inc. is not responsible for any incidental or consequential loss or dam-age which might result from a failure of any Sensotec, Inc. product.

THIS EXPRESS WARRANTY IS MADE IN LIEU OF ANY AND ALL OTHERWARRANTIES, EXPRESSED OR IMPLIED, INCLUDING IMPLIED WARRANTYOF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE.

Obtaining Service UnderWarranty

Advanced authorization is required prior to the return to Sensotec, Inc. Beforereturning the items, either write to the Repair Department c/o Sensotec, Inc., 2080Arlingate Lane, Columbus, Ohio 43228, or call (800) 848-6564 with: 1) a partnumber; 2) a serial number of the defective product; 3) a technical description* ofthe defect; 4) a no-charge purchase order number (so products can be returned toyou correctly); and 5) ship and bill addresses. Shipment to Sensotec, Inc. shall beat Buyer's expense and repaired or replacement items will be shipped F.O.B. ourplant in Columbus, Ohio. Non-verified problems or defects may be subject to anevaluation charge. Please return the original calibration data with the unit.

Obtaining Non-warrantyService

Advance authorization is required prior to the return to Sensotec, Inc. Beforereturning the item, either write to the Repair Department c/o Sensotec, Inc., 2080Arlingate Lane, Columbus, Ohio 43228, or call (800) 848-6564 with: 1) a partnumber; 2) a serial number of the defective product; 3) a technical description* ofthe malfunction; 4) a purchase order number to cover Sensotec, Inc.'s repair cost;and 5) ship and bill addresses. After the product is evaluated by Sensotec, Inc.,we will contact you to provide the estimated repair costs before proceeding. Theminimum evaluation charge is $95. Shipment to Sensotec, Inc. shall be at Buyer'sexpense and repaired items will be shipped to you F.O.B., our plant in Columbus,Ohio. Please return the original calibration data with the unit.

Repair Warranty All repairs of Sensotec, Inc. products are warranted for a period of 90 days fromdate of shipment. This warranty applies only to those items which were founddefective and repaired. It does not apply to products in which no defect was foundand returned as is or merely re-calibrated. Out of warranty products may not becapable of being returned to the exact original specifications or dimensions.*Technical description of the defect: In order to properly repair a product, it is necessary forSensotec, Inc. to receive information specifying the reason the product is being returned.Specific test data, written observations on the failure and specific corrective action yourequire is needed.

ACCU-GAGE Instruction Manual page 101

(800)848-65642080 Arlingate Lane

Columbus, Ohio 43228Tel: (614)850-5000Fax: (614)850-1111

E-mail: [email protected]://www.sensotec.com