1
Determination of the stabilizers antimony, bismuth, and iodate in electroless nickel baths as replacements for lead Antimony(III) and bismuth Iodate Summary M. Kubicsko 1 and B. Zumbrägel 2 Voltammetry Principle Antimony and bismuth are determined by anodic stripping voltammetry in 0.6 mol/L KCl electrolyte at pH < 2. For preconcentration, antimony(III) and bismuth are reduced at a potential of –0.4 V at the hanging mercury drop electrode (HMDE). The two metals dissolve and deposit in the mercury. In the subsequent determination step, the dissolved metals are anodically stripped off the electrode which results in two distinguished signals at 0 16 V for Principle Iodate is determined by polarography in citrate buffer at a pH value around 3. The analytical signal at –0.34 V results from a direct reduction of iodate to iodide on the dropping mercury electrode. This reaction is irreversible. Therefore, it is not possible to increase the sensitivity of the determination by deposition. Besides nickel as main component, an electroless nickel bath contains further major constituents such as electrolyte, buffer as well as reducing and complexing agents. A minor component is the so-called stabilizer. The concentration of this additive ranges from the millimolar to the micromolar level and determines the stability and performance of the bath, for which reason it has to be accurately monitored. This poster describes simple and fast, but sensitive and robust methods for the determination of the stabilizers antimony, bismuth, and iodate in Sample volume 0.200.25 mL sample the electrode, which results in two distinguished signals, at 0.16 V for antimony(III) and –0.04 V for bismuth. electroless nickel baths. Antimony(III) and bismuth are determined in parallel by anodic stripping voltammetry (ASV) in 0.6 mol/L KCl electrolyte at pH < 2, whereas the determination of iodate is carried out by polarography in 0.1 mol/L sodium citrate electrolyte at a pH value between 2.8 and 3.5. Under the tested conditions, the linear working range in the electroless nickel bath is for antimony β(Sb III )= 0.02…7 mg/L, for bismuth β(Bi) = 0.02…10 mg/L, and for iodate β(IO 3 - )= Measuring mode DP (differential pulse) Pulse amplitude: 0.02 V Deposition potential –0.4 mV time 30 s Potential scan start –0.3 V end 0.05 V Peak potential Sb(III) –0.16 ± 0.05 V Bi –0.04 ± 0.05 V Calibration technique Standard addition Sample volume 0.5 mL electroless nickel bath Electrolyte 10 mL supporting electrolyte c (KCl) = 0.6 mol/L c (HNO 3 ) = 0.07 mol/L pH value in the measuring solution 1.3…2 Working electrode Hanging Mercury Drop Electrode Reference electrode Ag/AgCl/KCl (3 mol/L) Auxiliary electrode Platinum Measuring mode DP (differential pulse) Pulse amplitude: 0.05 V Deposition potential time Potential scan start –0.1 V end –0.7 V Peak potential IO 3 –0.34 ± 0.05 V Calibration technique Standard addition Sample volume 1.0 mL electroless nickel bath Electrolyte 10 mL citrate buffer, pH 2.4 c(trisodium citrate) = 0.1 mol/L c(HCl) = 0.3 mol/L pH value in the measuring solution 2.8…3.5 Working electrode Dropping Mercury Electrode (DME) Reference electrode Ag/AgCl/KCl (3 mol/L) Auxiliary electrode Platinum pH value 8.5 ± 0.5 (adjustment with NaOH) Working electrode Hanging Mercury Drop Electrode (HMDE) Reference electrode 3 0.3…110 mg/L. Introduction Electroless nickel plating is an important and well-established process in the surface finishing industry. One of the major users of this plating process is the electronics industry in the manufacture of printed circuit boards. In the past, the addition of small amounts of lead has widely been used to stabilize the plating Calibration technique Standard addition Auxiliary electrode Platinum Calibration technique Standard addition Auxiliary electrode Platinum Reference electrode Ag/AgCl/KCl (3 mol/L) Auxiliary electrode Pt Rod Electrode 797 VA Computrace 800 Dosino Special care has to be taken for the pH value if both elements have to be determined simultaneously. At pH > 2, the antimony determination does not work anymore and at pH < 1.3, the antimony signal shifts towards the bismuth signal which hinders the separation of the analytical currents for the two elements. addition of small amounts of lead has widely been used to stabilize the plating bath. The lead concentration in the bath was mostly controlled by voltammetry. Not only because voltammetry provides a simple and robust way of determining the additive, but rather because voltammetry only detects the electrochemical active part of the additive. This is a much better measure for the performance of the additive than the total metal concentration. In recent years, a number of countries have imposed restrictions on the sale of electronics containing environmentally hazardous substances, namely lead. In this which hinders the separation of the analytical currents for the two elements. Working range Working range context, alternative stabilizers were developed and introduced. Important stabilizers of this new generation are antimony, bismuth, and iodate. They can be used as a single additive or in combination with one or both of the others. Voltammetric analysis, with its capacity to quantify the electrochemical active species, is an excellent tool for controlling the concentrations of these additives. System setup The calibration curve is at least linear up to a concentration of β(IO 3 - ) 11 mg/L. Higher concentrations have not been tested. The limit of quantification is approximately β(IO 3 - ) 30 μg/L refer- ring to the total volume of the With a deposition time of 30 s, the calibration curve is linear up to a concentration of β(Sb III ) 350 μg/L and β(Bi) 500 μg/L. The limit of quantification is approximately β(Sb III ) 4 μg/L and β(Bi) 2 μg/L referring to the total volume of References ring to the total volume of the measuring solution. Using a sample size of 1 mL for the analysis, the linear working range for iodate is β(IO 3 - )= 0.3 … 110 mg/L in the sample. Voltammetric system for the semi-automated determination of antimony, bismuth or iodate in electroless nickel baths. 1 Metrohm USA, Riverview, FL 33578/USA, [email protected], phone +1 800 727 6768 2 Metrohm International Headquarters, CH-9101 Herisau/Switzerland, [email protected] approximately β(Sb III ) 4 μg/L and β(Bi) 2 μg/L referring to the total volume of the measuring solution. Using a sample size of 0.5 mL for the analysis, the linear working range for antimony(III) is β(Sb III ) = 0.08 … 7 mg/L and β(Bi) = 0.04 … 10 mg/L in the sample. By adjusting sample size and deposition time, the sensitivity of the determination can be adapted to other sample requirements. (1) Metrohm Application Bulletin AB-074, Polarographic and stripping voltammetric analysis methods for thallium, antimony, bismuth, and iron (copper, vanadium), (downloadable under http://www.metrohm.com/com/Applications). (2) I.M. Kolthoff and J.J. Lingane, Polarography, 2nd ed., Interscience Publishers, New York (1952). Download a copy of this poster from http://www.metrohm.com/com/Applications (search for 8.000.6068EN).

700709_80006068EN

  • Upload
    abasak

  • View
    7

  • Download
    0

Embed Size (px)

DESCRIPTION

700709_80006068EN

Citation preview

Page 1: 700709_80006068EN

Determination of the stabilizers antimony, bismuth, and iodate in electroless nickel baths as replacements for lead

Antimony(III) and bismuth IodateSummary

M. Kubicsko1 and B. Zumbrägel2

Vo

ltam

met

ry

y( )

PrincipleAntimony and bismuth are determined by anodic stripping voltammetry in0.6 mol/L KCl electrolyte at pH < 2. For preconcentration, antimony(III) andbismuth are reduced at a potential of –0.4 V at the hanging mercury dropelectrode (HMDE). The two metals dissolve and deposit in the mercury. In thesubsequent determination step, the dissolved metals are anodically stripped offthe electrode which results in two distinguished signals at 0 16 V for

PrincipleIodate is determined by polarography in citrate buffer at a pH value around 3. Theanalytical signal at –0.34 V results from a direct reduction of iodate to iodide onthe dropping mercury electrode. This reaction is irreversible. Therefore, it is notpossible to increase the sensitivity of the determination by deposition.

y

Besides nickel as main component, an electroless nickel bath contains furthermajor constituents such as electrolyte, buffer as well as reducing and complexingagents. A minor component is the so-called stabilizer. The concentration of thisadditive ranges from the millimolar to the micromolar level and determines thestability and performance of the bath, for which reason it has to be accuratelymonitored. This poster describes simple and fast, but sensitive and robust methodsfor the determination of the stabilizers antimony, bismuth, and iodate in

Sample volume0.20…0.25 mL sample

the electrode, which results in two distinguished signals, at –0.16 V forantimony(III) and –0.04 V for bismuth.

y, ,electroless nickel baths.

Antimony(III) and bismuth are determined in parallel by anodic strippingvoltammetry (ASV) in 0.6 mol/L KCl electrolyte at pH < 2, whereas thedetermination of iodate is carried out by polarography in 0.1 mol/L sodium citrateelectrolyte at a pH value between 2.8 and 3.5. Under the tested conditions, thelinear working range in the electroless nickel bath is for antimony β(SbIII) =0.02…7 mg/L, for bismuth β(Bi) = 0.02…10 mg/L, and for iodate β(IO3

-) =

Measuring mode DP (differential pulse) Pulse amplitude: 0.02 V

Deposition potential –0.4 mV

time 30 s

Potential scan start –0.3 V end 0.05 V

Peak potential Sb(III) –0.16 ± 0.05 V

Bi –0.04 ± 0.05 V

Calibration technique Standard addition

Sample volume 0.5 mL electroless nickel bath

Electrolyte 10 mL supporting electrolyte

c (KCl) = 0.6 mol/L c (HNO3) = 0.07 mol/L

pH value in the measuring solution

1.3…2

Working electrode Hanging Mercury Drop Electrode Reference electrode Ag/AgCl/KCl (3 mol/L) Auxiliary electrode Platinum

Measuring mode DP (differential pulse) Pulse amplitude: 0.05 V

Deposition potential –

time –

Potential scan start –0.1 V end –0.7 V

Peak potential IO3– –0.34 ± 0.05 V

Calibration technique Standard addition

Sample volume 1.0 mL electroless nickel bath

Electrolyte 10 mL citrate buffer, pH 2.4

c(trisodium citrate) = 0.1 mol/L c(HCl) = 0.3 mol/L

pH value in the measuring solution

2.8…3.5

Working electrode Dropping Mercury Electrode (DME) Reference electrode Ag/AgCl/KCl (3 mol/L) Auxiliary electrode Platinum

pH value8.5 ± 0.5 (adjustment with NaOH)

Working electrodeHanging Mercury Drop Electrode (HMDE)

Reference electrode

3

0.3…110 mg/L.

IntroductionElectroless nickel plating is an important and well-established process in thesurface finishing industry. One of the major users of this plating process is theelectronics industry in the manufacture of printed circuit boards. In the past, theaddition of small amounts of lead has widely been used to stabilize the plating

Calibration technique Standard addition

Auxiliary electrode Platinum

Calibration technique Standard addition

Auxiliary electrode Platinum

Reference electrodeAg/AgCl/KCl (3 mol/L)Auxiliary electrode

Pt Rod Electrode 797 VA Computrace 800 Dosino

Special care has to be taken for the pH value if both elements have to bedetermined simultaneously. At pH > 2, the antimony determination does not workanymore and at pH < 1.3, the antimony signal shifts towards the bismuth signalwhich hinders the separation of the analytical currents for the two elements.

addition of small amounts of lead has widely been used to stabilize the platingbath. The lead concentration in the bath was mostly controlled by voltammetry.Not only because voltammetry provides a simple and robust way of determiningthe additive, but rather because voltammetry only detects the electrochemicalactive part of the additive. This is a much better measure for the performance ofthe additive than the total metal concentration.

In recent years, a number of countries have imposed restrictions on the sale ofelectronics containing environmentally hazardous substances, namely lead. In this which hinders the separation of the analytical currents for the two elements.

Working range Working range

g y , ycontext, alternative stabilizers were developed and introduced. Importantstabilizers of this new generation are antimony, bismuth, and iodate. They can beused as a single additive or in combination with one or both of the others.Voltammetric analysis, with its capacity to quantify the electrochemical activespecies, is an excellent tool for controlling the concentrations of these additives.

System setup

The calibration curve is at least linear upto a concentration of β(IO3

-) ≈ 11 mg/L.Higher concentrations have not beentested. The limit of quantification isapproximately β(IO3

-) ≈ 30 μg/L refer-ring to the total volume of the

With a deposition time of 30 s, the calibration curve is linear up to a concentrationof β(SbIII) ≈ 350 μg/L and β(Bi) ≈ 500 μg/L. The limit of quantification isapproximately β(SbIII) 4 μg/L and β(Bi) 2 μg/L referring to the total volume of

References

ring to the total volume of themeasuring solution. Using a sample sizeof 1 mL for the analysis, the linearworking range for iodate is β(IO3

-) =0.3 … 110 mg/L in the sample.Voltammetric system for the

semi-automated determinationof antimony, bismuth or iodatein electroless nickel baths.

1Metrohm USA, Riverview, FL 33578/USA, [email protected], phone +1 800 727 67682Metrohm International Headquarters, CH-9101 Herisau/Switzerland, [email protected]

approximately β(SbIII) ≈ 4 μg/L and β(Bi) ≈ 2 μg/L referring to the total volume ofthe measuring solution. Using a sample size of 0.5 mL for the analysis, the linearworking range for antimony(III) is β(SbIII) = 0.08 … 7 mg/L and β(Bi) = 0.04 … 10mg/L in the sample. By adjusting sample size and deposition time, the sensitivity ofthe determination can be adapted to other sample requirements.

(1) Metrohm Application Bulletin AB-074, Polarographic and stripping voltammetric analysismethods for thallium, antimony, bismuth, and iron (copper, vanadium), (downloadableunder http://www.metrohm.com/com/Applications).

(2) I.M. Kolthoff and J.J. Lingane, Polarography, 2nd ed., Interscience Publishers, New York(1952).

Download a copy of this poster from http://www.metrohm.com/com/Applications (search for 8.000.6068EN).