www.stryker.com 1
6390 Angle Position Sensor (APS) Calibration Instructions
TOOLS REQUIRED
• StrykerEMSUSBTool • Male-to-MaleUSBCable
• T25TorxDriver • TapeMeasure
PROCEDURE
1. Placethevehicleonalevelsurface.
2. MakesurethatthePower-LOADmainpoweristurnedon.
3. PressandholdthereleaseleveratthefootendofthePower-LOADsystemandpulltoremovethecotfromthe
vehiclepatientcompartment.
Note: The head end LED indicators turn solid green only when the cot is ready to unload.
4. Unloadthecotfromthevehicle:
• For cot models 6500/6506 & 6510/6516:
1. Pressandholdtheextend(+)buttononthecotcontrolswitchtoextendthecotbaseuntilthecotwheels
restontheground.
2. Releasetheextend(+)buttonafterthecotisnolongersupportedbytheliftingarms.Theliftingarmswill
continuemovinguntiltheyhavefullylowered.Do not unlock the cot from Power-LOAD.
• For cot models 6085/6086:
1. Operator1(footend)-Graspthecotframeatthefootend.Squeezeandholdthecotmanualrelease.
2. Operator2(side) -Grasp thebaseframe, liftslightly,and lower thebaseframe to its fullyextended
position.Verifythatthebaseisfullyextended.
3. Operator1(FootEnd)−Releasethecotmanualreleaseto locktheundercarriage intotheextended
position.
4. Pressthedown()buttononthePower-LOADcontrolpaneltolowertheliftingarmsandcot.Do not
unlock the cot from Power-LOAD.
5. UsingaT25Torxdriver,removethetwoscrews(A)thatsecurethecovertotheleftwingassemblytogainaccess
to theUSBport (Figure 1). Save the screwsand
coverforreinstallation.
6. UnscrewthecapthatprotectstheUSBport.
7. Set the laptop with the Stryker EMS USB tool
installedontopofthecotmattressatthefootend
ofthecot.
AFigure 1
6390 Angle Position Sensor (APS) Calibration Instructions
2 www.stryker.com
8. Plugthemale-to-maleUSBcableintotheUSBportandtoanavailableUSBportonthelaptop.
Note: Make sure that the male-to-male USB cable will not get caught on anything while the Power-LOAD
system moves in and out of the vehicle patient compartment.
9. Double-clicktheStryker EMS USB Tool icononyourdesktop.Thetoolsoftwarewillautomaticallydetectthe
Power-LOAD.
10. AftertheStrykerEMSUSBtoolandPower-LOADareconnected,clickAdvanced andthenclickCalibrate APS.
11. IntheEnter Passwordfield,enterStryker1andclickEnter.
12. Loadthecotintothevehicle:
• For cot models 6500/6506 & 6510/6516:
• Pressandholdtheretract(–)buttononthecotcontrolswitchtofullyretractthecotundercarriageuntil
thecotissupported.
• For cot models 6085/6086:
• Presstheup()buttononthePower-LOADcontrolpaneltoraisetheliftingarmstothehighestposition.
Note: The cot legs do not retract.
• Operator1(FootEnd)−Graspthecotframeatthefootend.Squeezeandholdthecotmanualrelease.
• Operator2(Side)−Stabilizethecotbyplacingonehandontheouterrail.Graspthebaseframe.After
thefootendoperatorhassqueezedthecotmanualrelease,retracttheundercarriagewithonehandand
holditinplace.
• Operator 1 (Foot End) - Release the cotmanual release to lock the undercarriage in the retracted
position.Ensurethatthecotmanualreleaseisreleased.If it isnotreleased,thecotbasewillextend
preventingthecotfromlockinginthefastener.
13. Pushthecotintothevehiclepatientcompartmentuntilthefootendcotwheelsareoverthefloorofthevehicle
patientcompartment,butnotinthefullinandlockedposition.
www.stryker.com 3
6390 Angle Position Sensor (APS) Calibration Instructions
Figure 3
14. Using a tape measure, measure the distance from the
bottomofthefootendcotwheeltothefloorwiththecot
wheelfacingthebackofthevehiclepatientcompartment
(Figure2).Makesurethatthedistancebetweenthebottom
ofthefootendcotwheeltotheflooris2in.
• Ifthemeasurementismorethan2in,clickDecrement
(-) on the Connected to Trolley calibration screen
(Figure3).
• Ifthemeasurementislessthan2in,clickIncrement
(+) on the Connected to Trolley calibration screen
(Figure3).
15. Graspthecotframeatthefootendto
pullthecotoutofthevehiclepatient
compartment.
Note: The head end LED indicators
turn solid green only when the cot is
ready to unload.
16. Unload thecot from thevehicle(see
steps3-4onpage1).
17. Load the cot into the vehicle (see
steps12-13onpage2).
18. Repeatsteps12-17until thedistance
between the bottom of the foot end
cotwheeltotheflooris2in.
19. Unplug the male-to-male USB cable
fromthePower-LOADUSBport.
20. Screw theUSB port cap back on to
theUSBport.
21. UsingaT25Torxdriver,reinstallthetwoscrews(removedinstep5)tosecurethecovertotheleftwingassembly.
Note: Make sure that the extruded “U” nut does not rotate clockwise when you tighten the bottom right T25
Torx screw on the left wing assembly cover.
22. Verifyproperoperationoftheproductbeforereturningittoservice.
3800E.CentreAve.,Portage,MI490021-800-327-0770|2014/01
Figure 2
2”