SYSM-660-1 Operation Manual for Dental Unit MODEL: L1-660C & 670C Shanghai LinkGold Medical Instrument Co., Ltd Release Date:October ,2007

50625256_USER_MANUAL_FOR_L1-660C__670C

Embed Size (px)

Citation preview

Page 1: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

Operation Manual

for

Dental Unit

MODEL: L1-660C & 670C

Shanghai LinkGold Medical Instrument Co., Ltd

Release Date:October ,2007

Page 2: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

Comments and Feedback

Thank you for taking the time to use the Operation Manual for Dental

Unit L1-660C and L1-670C . We would appreciate any feedback or

comments you have about this document. Please mail, email, or

phone us with your comments. You can reach us at:

Shanghai Linkgold Medical Instrument Co., Ltd.

Address: No. 26, Chngkang Road, Shanghai, China

Zip Code: 200443

Tel: +86-21-29422163

Fax: +86-21-51561729

E-mail: [email protected]

European representative:

Lotus Global Co., Ltd.

47 Spenlow House Bermondsey London SE16 4SJ

Page 3: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

C O N T E N T S

Identification of Symbols 1Classification of Equipment (EN-60601-1) 1Environmental 1Electromagnetic Compatibility 2Flammable Gasses 2About This Document 21.About L1-660 and L1-670 dental unit 31.1. External dimension of equipment 31.2 Equipment installation 31.3 parameters 41.4 Power On/Off Button 41.5 Activate Handpiece 51.6 Handpiece Lockout 51.7 Air/Water Coolant 51.8 Water System 61.9 Foot Control Touchpad 61.10 Touchpads 62. Operation 72.1 Status Icon 72.1.1 Water supply setting of rinsing water 72.1.2 Time setting of spittoon washin 72.1.3 Heating of rinsing water 82.1.4 Water supply of spittoon washing 83. Installation and Maintenance 83.1 Pipeline layout 83.2 Connection of air pipe and water pipe 83.3 Maintenance Cleaning and Asepsis 93.3.1 Handpiece 93.3.2 Water filter 93.3.3 Relief valve of air filtration 93.3.4 Saliva ejector and suction 93.3.5 Other 104. Disposal of equipmentSuction filter 105. Failure diagnosis 116. Warranty 127. Customer Service Handpiece 128. Water/Air schematicsOther 149. Electric wiring 1510. Packing List for L1-660/L1-670 16

Page 4: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

Identification of Symbols Table1 symbols

Symbols Description

Type B applied part

Attention, consult accompanying document

~ Alternating current

Protective earth (ground)

SN Serial Number

IPX7 Protected against the effects of temporary

immersion in water

compliance with the Medical device directive

93/42/EEC

Manufacturer Name and address

European representative

Manufacture date

Classification of Equipment (EN-60601-1)

Table2 Classification of Equipment

Type/Mode Classification

Types of shock protection CLASS I EQUIPMENT, TYPE B APPLIED PART

Degree of protection against

Water ingress ORDINARY EQUIPMENT except the foot switch(IPX7)

Mode of operation CONTINUOUS OPERATION except the dental chair

(CONTINUOUS OPERATION WITH INTERMITTENT

LOADING: 5% duty cycle)

Environmental

Table 3 Environmental Specifications

Page 5: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

Temperature/Humidity Specification

Storage and Transportation :

Temperature: -40℃ to 55℃

Relative Humidity: 93% maximum

Range of the atmospheric pressure 500 hPa~1060 hPa

Working conditions:

Temperature: 5℃ to 40℃

Relative Humidity: 80% maximum

Range of the atmospheric pressure 860 hPa~1060 hPa

Electromagnetic Compatibility

This equipment has been tested and found to comply with the limits for medical devices in

EN60601-1-2. These limits are designed to provide reasonable protection against harmful

interference in a typical medical installation. Please contact Shanghai Linkgold Medical

Instrument Co., Ltd. Customer Service if you have any questions.

Flammable Gasses

Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or

nitrous oxide, where such gasses may accumulate in concentration (closed space).

About This Document

Welcome to read the Operation Manual for the operation of Dental Unit (L1-660B/C and

L1-670B/C). This guide provides an easy-to-use source of technical information for operating

and maintaining your L1-660B/C and L1-670B/C Dental Unit .

Before using the L1-660B/C and L1-670B/C Dental Unit , make note of all cautions and

warnings. Read the component descriptions, and become familiar with the locations of

features and controls.

This document contains the following information:

•Module descriptions

• Feature descriptions

• Adjustments

• Maintenance tasks

Intended Audience This guide is intended for doctors and dental staff.

Page 6: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

1. About L1-660C and L1-670C dental unit

Model: L1-660C and Model: L1-670C dental unit are suitable for the diagnosis,

treatment and operation on oral disease. Model: L1-660C/L1-670C are both fully computer

controlled system treatment equipments designed by our company according to

ergonomics principle with complete function. As for its operation of complete machine, its

fully automatic operation is all controlled by the function key of microcomputer. It is

composed of multifunction dental complex treatment unit, electric dental chair and dentist

chair. It adopts low voltage DC motor and features with simple structure and convenient

operation. Water input hose (ID): 10mm – hose fitting 1/2"

Input water pressure: 200 – 400Kpa (Tap water)

Air input hose (ID): 10mm – hose fitting 1/2" Input water pressure: 550 – 800Kpa (Oil-free Anhydrous and Clean)

Drain pipe (ID): 4mm

Input power: 230v – 50Hz 800VA

1.1 External dimension of equipment To ensure its normal operation and safe movement, please refer to Figure.1 and keep

enough space for clinical use.

Figure.1

1.2 Equipment installation Preparation before installation ( Figure.2)

According to the layout of consulting room, the lighting conditions and the convenience for

use, confirm the position of the ground box and the position of the whole equipment in

accordance with Figure.4and the form of ground box.

Page 7: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

Figure.2

1.3 Parameters

Table 4 Feature descriptions Minimum height between seat cushion and floor ≤460mm

Maximum height between seat cushion and floor ≥760mm

Backward angle of backrest ≥60°

Up and down sliding stroke of Head rest 120mm

Dental

chair

Maximum load 135kg

Operating light Illumination

Color temperature

8000Lux~15000Lux

3000K~6500K

Negative pressure of Saliva ejector ≥27kPa

Negative pressure of high vacuum aspirator ≥35kPa

High handpiece speed, torque ≥300000r/min, ≥0.06N﹒cm

Low handpiece speed, torque ≥14000r/min, ≥1N﹒cm

Treatment

unit

Delivery tray maximum ≤10Kg

High handpiece air 023Mpa~0.26Mpa handpiece

Low handpiece air 0.30Mpa~0.32Mpa

For recommendation, please use handpiece made by W&H or NSK. Which are in compliance with standards ISO 2001 or CE.

1.4 Power On/Off Button (Figure.3) The delivery system have the power on/off button, on the chair.

Press the power on/off button once to turn the power on. Press the button again to turn the

power off.

NOTE Please shut off the power before replacing the safety fuse: F5Ax250V

Φ5mmx20mm

Page 8: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

Figure.3

NOTE The power on/off button must be on for the dental unit to function.

To prolong the life of your equipment, turn the power off when it is not in use.

ELECTRICAL WARNING Switch the branch circuit breaker off prior to the servicing of

the

dental unit. 1.5 Activate Handpiece

The handpieces automatically activate when you lift a handpiece from its holder or pull the

Continental whip forward. Press the foot control to operate the activated handpiece.

1.6 Handpiece Lockout

When a handpiece is removed from the holder and the foot control is pressed, the chair

does not operate.

1.7 Air/Water Coolant Each handpiece is set for independent on/off operation. When you withdraw a handpiece

from its resting position, air and water coolant will be activated. An icon displays on the

screen to indicate ctivation.(Figure 4)

Figure 4

The water coolant button operates as a three-way switch with the toggle on the foot control.

The system defaults to the last setting selected for each individual handpiece position.

Figure 5

(A) turn drive air (B) drive air (C) barometer (D

The L1-660 and L1-670 dental unit and regulate air a

A

C

B

) water (E)

nd water to

D

air

opera

E

te dental handpieces,

Page 9: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

syringes and ancillary devices. You can adjust the delivery system water coolant flow, air

flow and drive air pressure (Figure 5).

1.8 Water System

The L1-660 or L1-670 handpiece water system sed of distilled water.

1.9 Foot Control

C A

B

Figure 6

(A) Chip Blower/Accessory Button; (B) Wet/Dry Toggle; (C) Foot Control Disc

• To turn the water coolant off, use your foot to flip the toggle.

• To turn the water coolant on, flip the toggle again.

NOTE On a deluxe touchpad the wet/dry toggle operates as a three-way switch with the

air

and water coolant buttons.

1.10 Touchpads

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

KGOLD上 海 林 戈

复 位

LIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

复 位

上 海 林 戈KGOLDLIN

The touchpad centralizethe controls of the dental unit ( Figure 7). Some touchpad buttons

have indicators to alert you if the operation is functioning.

Figure 7

Touchpad Controls

Page 10: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

The following table describes the function of each button:

Table 5 Touchpad Button Identification.

Icon Description

Chair Rising

Chair Descending

Bend Backward

Bend Forward

Reset

Setting

Heating

Mouth Rinsing

Spittoon

2. Operation

2.1 Status Icon

The logo"shanghai linkgold" on the touchpad indicates the status of the system:

• A solid blue logo indicates normal operation and power is on.

• A blinking blue logo indicates the chair stop plate or the cuspidor limit switch has been

activated. Remove any obstructing objects. The icon should return to solid blue.

2.1.1 Water supply setting of rinsing water

a. Press down set button , gently press down “SET”, the setting state starts when

the indicating light is on.

b. Press rinsing water button , starts to supply water, stop pressing when the water

in measuring cup reaches anticipated water level, stops supplying water.

c. Press set button again, setting is over, computer will memorize and store, therefore,

once press the rinsing water button, it will automatically supply water with such amount

when it is used in the future.

2.1.2 Time setting of spittoon washing

a. Press down set button , gently press down “SET” and the indicating light is on.

Page 11: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

b. Press spittoon washing button , it starts to fill water into spittoon. When it reaches

the required and set time, press , again and the setting is over. Computer will

memorize and store, therefore, once spittoon washing button is pressed tenderly, it will

automatically wash till the set time when it is used in the future.

2.1.3 Heating of rinsing water

The unit has the automatic isothermal water heater. Please operate as the following to

get hot water. Press the heating button of water heater, the indicating light beside

button will be on, and then the water heater will heat automatically till the

isothermal state. To close this system, please press this button again, and the indicating

light beside button will be off.

2.1.4 Water supply of rinsing

Press rinsing water supply button to supply water and it will be closed automatically

when it reaches preset water amount.

2.1.5 Water supply of spittoon washing

Press water supply button of spittoon washing to supply water automatically and it

will be closed automatically when ti reaches preset time of water supply.

3. Installation and Maintenance

The installation of officers for professionals.

3.1 Pipeline layout (Figure.8

Power supply

Drainage pipe1-1/2”

water(air)pipe

Figure.8

3.2 Connection of air pipe and water pipe (Figure.9)

Page 12: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

Water input(Φ8mm×1) Blue) Air input (Φ8mm×1) transparent

Figure.9

3.3 Maintenance Cleaning and Asepsis 3.3.1 Handpiece

Handpiece is a delicate device, so it shall not be pounded or dropped. It shall be cleaned

and lubricated with special detergent and lubricant every day. As for the specific method,

pleace refer to the instruction manual of handpiece provided by they manufacturer.

Clean the surface of dental drill with alcohol cotton.

Blow it with clean lubricating oil for 2-3s.

Wrap it with sterilization package.

High-temperature steam sterilization(135,15min,220KPa)

3.3.2 Water filter The unit has water filter to ensure its normal use.After serving a certain time,the dirt will block

the filter core and affect the water flow,therefore,it is necessary to clean and replace the filter

core. The filter core shall be cleaned and replaced under any of the following conditions:

It has been used for over half a year.(Depend on the local water quality).

The extracted water becomes turbid suddenly.

The specific cleaning method is as follows: open the ground box cover first, counterclockwise

rotate and dismount water filter cover to remove filter core. Clean it with the ultrasonic or

neutral detergent, wash it with tap water thoroughly after cleaning and then mount water filter

cover. The water filter can stop the inpurity of water, clean the filter core for every three months.

3.3.3 Relief valve of air filtration Drain the water when:

It is used for over one week.

The water in filter cup accounts for three-quarter of volume.

The color of water in filter cup changes turbid .

3.3.4 Saliva ejector and suction Besides the maintenance of their exterior cleanness, the cleanness of interior pipeline

system of saliva ejector and suction still need to be maintained. The method of cleaning

pipe system is as follows: regularly pump clear water or special sterilized water about 30s

Page 13: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

with recommended 1 or 2 times every day.

3.3.5 Other a) 3-way syringe tip: high-temperature steam sterilization(135°C,15min,220KPa) .

b) Dental chair: clean and sterilize it with leather detergent and medical alcohol.

4. Disposal of equipment

Please follow the local environmental protection laws and regulations for the dispose of this

unit.

5. Failure diagnosis

NO Symptom Cause Solution

1

When it is used frequently, the

dental chair doesn’t operate

suddenly,while the control system

under normal operation.

The motor is overheated

and stops operating. After stopping for a while, It could

continue to operate properly.

2

The saliva ejector or high

vacuum aspirator leak

automatically.

The mounting position

of rack valve changes.

Open the rear cover of rack,adjust

the front and back mounting

position of rack valve until it stops

When putting down the instrument

and operates when taking up the

instrument.

3 The suction of saliva

ejector and high

vacuum aspirator is small

or without suction.

1.The pressure of tap

water is too low.

2.The water filter is

blocked.

3.Pressure of supply gas

Is too low.

4.Negative pressure

Generator is blocked.

5.The pipeline and joint

leak.

1. Install supercharge water

pump if the water pressure

is too low.

. Clean the water filter and

Replace filter core.

. Adjust supply gas pressure

To 0.5MPa.

. Dismount and wash

negative pressure

generator ,remove the

blocking.

5.Check the pipeline and joint.

4 No water in handpiece. 1.The jet hole of

handpiece is blocked.

2.The water filling switch

of foot control air

switch isn’t open.

3.The water regulating

valve of handpiece

isn’t open.

1.Dredge the jet hole of

handpiece.

2.Shift the water filling switch

of foot control air switch to

the right.

3.Open the water regulating

Valve of handpiece and

adjust it properly.

5 The handpiece isn’t

powerful or doesn’t operate.

1.The operation air

pressure of handpiece

Is too low.

1.Adjust the operation air

pressure of handpiece

according to instruction

Page 14: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

2.The bearing of

handpiece is damaged.

3.The dental bur is worn

or isn’t clamped tightly.

4.The pipeline of

handpiece is blocked.

5.The air pipe inside

handpiece leaks.

manual.

2.Replace the bearing of

Handpiece.

3.Repiace with a new dental

bur and then clamp the

dental bur tightly.

4.Dismount bearing assembly

and dredge the pipeline of

handpiece.

5.Replace the housing of

handpiece.

6 The button of three-way

syringe leaks.

O-ring is worn or aging. Replace O-ring and add

Silicon grease oil to lubricate.

7 The three-way syringe

doesn’t sufficiently spray

vapor.

1.The water-pipe and

air-pipe of three-way

ssyringe is connected

oppositely.

.The water of three-way

syringe is too strong or

the air is too weak.

1. Exchange the water-pipe

with the air-pipe of three-way

syringe.

. Adjiust the water and air of

three-way syringe.

8 Rinsing water and spittoon

washing water are too few

or without water.

1.The water pressure is

too low.

2.The water filter is

blocked.

3.Pressure of supply gas

is too low.

4.Electromagnetic valve

doesn’t work.

5.The connecting wire

and connector don’t

contact properly.

6.The circuit board

breaks down.

7.The control keyboard

doesn’t contact properly.

1.Install supercharge water

pump if the water pressure

is too low.

2.Clean the water filter and

replace filter core.

3.Adjust supply gas pressure

to 0.5MPa.

4.Replace the damaged

electromagnetic valve.

5.Find out the contact failure

position of connecting wire

and reconnect it.

6.Replace defective circuit

board.

7.Replace control keyboard.

9 The treatment unit doesn’t

operate.

1. Thepower plug isn’t

properly plugged or

doesn’t properly contact.

. The power fuse is

blow-out.

. The transformer of

ground box is damaged.

. The power switch is

broken.

1.Replace power plug or

outlet.

2.Find out the reason of

blow-out first and replace

power fuse.

3.Replace the transformer of

Ground box.

4.Replace power switch.

Page 15: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

10 No water in treatment unit The water filter is blocked.

Water pipe is pinched.

Clean filter core.

Remove the pinching source.

11 There is water spraying out

from the air.

1.The joint of handpiece

Isn’t screwed tightly.

2.The filter relief valve is

full of water.

1.Screw the joint of

handpiece tightly.

2.Discharge the accumulated

water inside filter relief valve.

12 The operating light isn’t on. 1.The bulb of operating

Light is broken.

2.The connecting wire of

operating light doesn’t

properly contact.

3.The switch of operating

light doesn’t properly

contact.

4.The lead wire of ground

box transformer doesn’t

properly contact.

1.Replace a new bulb.

2.Find out the defective

connection position.

3.Replace the switch of

operating light.

4.Reconnect the lead wire of

ground box transformer or

replace the transformer of

ground box.

13 The chair doesn’t operate,

while its sound indication is

normal and other buttons

are under normal operation.

1.The relay contact

doesn’t contact properly

or the coil is broken.

2.The motor is broken or

the connecting wire is

disconnected.

1.Clean the relay contact or

replace a new relay.

2.Find out the disconnect

position,reconnect or replace

the motor.

6. Warranty Shanghai linkgold all products against defects in materials or workmanship for one year

from time of delivery. shanghai linkgold’s sole obligation under the warranty is to provide

parts for the repair, or at its option, to provide the replacement product (excluding labor).

The buyer shall have no other remedy. All special, incidental, and coincidental damages are

excluded.

Written notice of breach of warranty must be given to shanghai linkgold within the warranty

period. The warranty does not cover damage resulting from improper installation or

maintenance, accident or misuse. The warranty does not cover damage resulting from the

use of cleaning,

disinfecting or sterilization chemicals and processes. The warranty also does not cover light

bulbs. Failure to follow instructions provided in the shanghai linkgold Instructions For Use

(operation and maintenance instructions) may void the warranty.

No other warranties as to merchantability or otherwise are made. For service information,

contact your local authorized shanghai linkgold dealer.

CAUTION Restricts this device to sale by or on the order of a dentist, physician, or

any other practitioner licensed by the law of the state in which he or she practices to use or

order the use of this device

Page 16: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

7. Customer Service

For service information, contact your local authorized dealer or manufacturer.

Any optional non shanghai linkgold attachments and/or accessories must comply with EN

60601-1 and EN 60601-1-2.

Specifications are subject to change without notice.

Page 17: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

8. Water/Air schematics

Page 18: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

9. Electric wiring

Page 19: 50625256_USER_MANUAL_FOR_L1-660C__670C

SYSM-660-1

10. Packing List for L1-660C/L1-670C

No. Description Quantity Remarks

01 Dental unit 1

02 3-way syringe 2

03 High Sunction 1

04 Saliva ejector 1

05 Dental light 1

06 Doctor’s chair 1

07 1/2*φ8mm connector 2

08 Fusion 2 F5Ax250V

Φ5mmx20mm