2EducationSTRAUMANN
Purpose of 48-hour tool
Available for NC and RC Removal of placed implant from patient Substitute for transfer part after transfer part failure Implant needs to be discharged after removal Can be done up to 48 hours after placement,
afterwards osseointegration is too far advanced! Do not use for height correction
4EducationSTRAUMANN
Situation
Implant stuck & transfer piece breaks
Transfer part has predetermined breakage point:
RC: 100 Ncm
NC: 80 Ncm
Up until this point, the internal connection will not be damaged
The breakage point is located at the head of the transfer screw
6EducationSTRAUMANN
Removing the Implant
Remove transfer nut from implant (no unscrewing needed)
The remaining screw remains secure within the implant
CAUTION
Make sure the transfer nut is not aspirated
7EducationSTRAUMANN
Removing the Implant
Remove remaining screw with tweezers or pliers by counter clockwise rotations
CAUTION
Make sure the remaining screw is not aspirated
8EducationSTRAUMANN
Removing the Implant
Mount 48-hour tool like an abutment
Secure tool by hand-tightening the mounting screw