36
4000 Series Articulator and QuickMount Face-Bow Instruction Manual

4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

4000 SeriesArticulatorandQuickMountFace-BowInstructionManual

Page 2: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

2

WHIP MIX 4000 SERIES ARTICULATORS

Model 4640

Model 4641

Page 3: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

3

CONTENTS

Introduction .......................................................................................................................................... 4

Operation Instructions for the Locking Mechanism............................................................................... 6

Obtaining Face-Bow Registration .......................................................................................................... 7

Direct Mounting the Maxillary Cast on the 4000 Series Articulator ...................................................... 9

Mounting the Maxillary Cast using the Indirect Mounting Method ....................................................... 11

Mounting the Mandibular Cast ............................................................................................................. 12

Setting the Condylar Guidance of the Model 4640 and 4641 Articulators ............................................. 13

Setting the Condylar Inclination using a Protrusive Record .................................................................. 15

Complete Denture Technigue ................................................................................................................ 16

Whip Mix Recording System ................................................................................................................ 19

Interchangeability with the 4000 Series Articulator ............................................................................... 20

Reminders and Suggestions ................................................................................................................. 21

Maintenance ......................................................................................................................................... 22

Articulator Accessories ......................................................................................................................... 23

Face-Bow Accessories .......................................................................................................................... 24

4000 Series Articulator Parts Lists and Diagrams

Model 4640 Articulator ......................................................................................................... 26

Model 4640Q Articulator ...................................................................................................... 27

Model 4641 Articulator ......................................................................................................... 28

Face-Bow Parts Lists and Diagrams

Model 8645 Face-Bow (for Direct Mounting) ....................................................................... 29

Model 9600 Face-Bow (for Direct Mounting) ....................................................................... 30

Models 9175/9185/9195 Face-Bows (for Indirect Mounting) ............................................... 31

Models 9275/9285/9295 Face-Bows (for Indirect Mounting, Plastic Ear Bow) .................... 32

Bibliography ......................................................................................................................................... 33

Page 4: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

4

INTRODUCTION

The WHIP MIX Series 4000 Articulator and “QUICKMOUNT” Face-Bow are designed to enable the user to

quickly and easily mount casts of a patient’s jaws on a mechanical likeness that will reproduce their natural

relationship and movements with an acceptable degree of accuracy. The simplicity and speed with which

the necessary registrations are obtained and transferred to the WHIP MIX 4000 Series Articulator enable the

operator to accomplish corrective and restorative dentistry with much greater precision than has ever before

been possible.

The WHIP MIX 4000 Series Articulator serves as an excellent auxiliary instrument for diagnostic and

patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is

ideal for the study of occlusion and the movements of the temporomandibular joint. With the condyle located on

the lower frame and the guidance on the upper frame (arcon design), WHIP MIX semi-adjustable articulators

have become the preferred choice of many teaching institutions. Advancing to a fully adjustable articulator

becomes a much easier process after having initial training on an arcon semi-adjustable articulator.

As with all mechanical devices, the degree of perfection of the finishesd product depends to a great

extent on the operator’s comprehension of the results desired and on a complete understanding of the

limitations of the instrument being used. The instructions that follow in this manual make no attempt to

teach concepts of occlusion and articulation. They simply outline a method to obtain a face-bow registration,

itemize the other registrations needed, and detail their in setting the articulator.

These instructions will deliver the minimum amount of guidance that is considered necessary and do

not exhaust the capabilities of the articulator. Therefore, it is suggested that the operator refresh

his/her knowledge by reviewing one or more of the several excellent texts on subject of articulator

usage (see bibliography).

The 4000 Series Articulator has the following features:

• Ergonomic design.

• Tracking Type of Condylar Guidance with Progressive Side Shift capability.

• A permanent intercondylar width of 110 mm, which is the industry standard for Semi-Adjustable

Articulators with a fixed setting.

• Fully open access from the lingual position for easy access during denture setup.

• Easy to operate Locking Mechanism, which provides the following actions: In the locked position

it provides a quick way to return to centric position and locks the articulator for added stability. In

the mid-range position the Locking Mechanism allows the condylar ball to move freely down the

track of the condylar guides during excursive movements.

Page 5: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

5

• Capability of removing the upper member for greater access for cleaning and servicing the

articulator. To remove the upper member, simply push the spring-loaded release button located

on the side of the condylar element and move the locking mechanism back to release the

condylar ball.

• A Support Pin to support the upper frame of the articulator when open.

• A built-in storage compartment for storing the Support Pin when not being used.

• Ergonomically designed Condyle Locking Knobs for greater ease in locking the condyle in

the desired position.

• Screw type mounting plates standard, but can be upgraded to include the QuickMount

Magnet System.

• Innovative “ACCUMOUNT” Mounting System of interchangeability. This will make it possible to

interchange mounted casts between any 4000 Series Articulator as found in the 2000 and

3000 Series Articulators.

The main difference between the 4000 Series Articulator and traditional WHIP MIX Articulators is the

elimination of the crossbar between the uprights on the lower member. The Locking Mechanism is

used as the primary source for securing the articulator in place.

Page 6: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

6

OPERATION INSTRUCTIONS FOR THE LOCKING MECHANISM

a. When the tabs are in the “UP” position, the condyles are locked in centric. In this locked position, theoperator is assured that the articulator is locked in centric and ensures the stability of the articulator.

b. Moving the tabs back one position places the Locking Mechanism in the mid-range position. In themid-range position the condyles can be moved down the tracking surface of the fossa element duringexcursive movements. The design of the Locking Mechanism prevents the user from accidentallyoverextending the lateral movements, which could destroy work in progress.

The upper member of the articulator can be easily removed from the lower member if necessary forgreater access during cleaning and servicing of the articulator. To accomplish this, push the spring-loaded release button located on the outside of the fossa element while simultaneously rotating thelocking mechanism back. After unlocking one side, the upper member can be moved medially toseparate it from the lower frame.

Page 7: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

7

OBTAINING FACE-BOW REGISTRATION

Fig. 2

Fig. 1

Items needed for a Face-Bow Registration:

Face-bow including nasion relator andface-bow fork (bite fork)

Elastomeric registration material

Adhesive for elastomeric

Fig. 3

I. Preparing Face-BowClean and properly disinfect the plastic earpieces before each use. If replacing them, makesure that the hole on the flat side of each isabove the side arm and the plastic is seateduntil it touches the shoulder of the earbow arm(Fig. 2). Note the ability of the nasion relatorassembly to move left and right along the crossbar and the face-bow caliper design allowingthe side arms to move equal distances duringthe opening and closing motions.

Center the nasion relator assembly on the crossbar between the two centering indications.Loosen the #8604 thumb screws on the face-bow and the #8643 thumb screws on thetransfer assembly.

II. Preparing Face-Bow ForkElastomeric registration material may be usedas a bite registration medium for the bite fork.Make sure bite fork has been sterilized andapply sufficient registration material to the topsurface of the bite fork. Position the bite fork sothat the midline of the bite fork aligns with facialmidline and support the fork until the materialhas set (Fig. 3). The record should make ashallow registration of the maxillary cusp tips,and not perforate through the bite fork. Inspectregistration for tissue interferences and trimaway any that is found with a Bard-Parker Knife.

III. Positioning Face-Bow onthe Patient and SecuringRegistrationBefore attempting to place the face-bow ontothe bite fork, it is suggested that the operatorrehearse the procedure with the patient. Cautionthe patient that the plastic ear pieces in theauditory canal will greatly amplify noises duringthe procedure. Support bite fork using cottonrolls placed underneath. A finger cot placed overeach ear piece will help facilitate properdisinfection of the face-bow.

Page 8: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

8

Fig. 4

Fig. 5

Completed Face-Bow Registration

With the bite fork positioned in the patient’s mouth,start the toggle onto bite fork shaft while the patientholds the side arms close to the ears. Note: Makesure the horizontal cross bar is above the bitefork shaft (Fig. 4). Instruct the patient to place eachear piece into the external auditory meatus and holdin place with a firm forward pressure. Center the nosepiece on the patient’s nasion and exert firm pressureon the nose piece shaft while tightening the nasionthumbscrew. This action will help to firmly seat theear pieces into the external auditory meatus.

With the patient continuing to hold onto the face-bow,position the toggle assembly on the bite fork shaftand slide until it is near, but not touching the lips andtighten the thumb screw. To prevent torquing of theface-bow and discomfort to the patient, support thefork and the horizontal bar with one hand whiletightening. Next tighten the thumb screw on thevertical bar while supporting the face-bow, againtaking care not to tilt the face-bow.

IV. Removing Face-Bow from the PatientLoosen the nasion thumb screw and slide the nasion relator away from the patient’s nose. Loosen thethumb screws on the face-bow and have the patient pull the ear pieces out of their ears. The entire face-bow assembly is removed by advising the patient to pull the ear pieces out of their ears while helping toremove the bite fork from the mouth.

V. Obtaining Interocclusal RecordsNow is a convenient time to make the interocclusal records and set the condylar guidance of thearticulator. There are several materials available that may be used to make interocclusal records. Thematerial selected should complement the particular technique used. The following interocclusal recordsshould be utilized to relate the mandibular cast to the maxillary mounting and to program the condylarguidance of the articulator:

• Centric Relation and/or Maximum Intercuspation

• Right Lateral

• Left Lateral

• Protrusive (optional)

Page 9: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

9

Fig. 6 Fig. 7

DIRECT MOUNTING THE MAXILLARY CASTON THE 4000 SERIES ARTICULATOR

Fig. 8

The Direct Face-Bow attached to the Model4640 Articulator.

I. Preparing the 4000 Series ArticulatorEnsure the mounting surface between the articulator and the mounting plate is free of any debris. Anydebris between these surfaces will affect the interchangeability accuracy of the instrument. Tighten theblack condylar locking knob while pressing the condylar guide assembly medially against the upper frameof the articulator. Set each progressive side shift setting to the 0° mark and tighten the side shift lockingscrew. Place a clean mounting plate on the upper and lower frame. Center the plastic incisal guide tableso that the rounded end of the incisal guide pin rests in the center of the table, then raise the incisal guidepin so that the upper member of the articulator is able to rest on the horizontal crossbar of the face-bow.Attach the #4611 Support Bar to the upper member.

II. Placing the Direct Mounting Face-Bow Using the Metal Face-BowSlide the Nasion Relator Assembly over until it contacts the left earbow half (Fig. 6). Loosen the thumbscrews on top of the face-bow. While holding the face-bow in one hand and the articulator in the other,position the brass mounting pin, located on the outer edge of the upright post on the lower frame, into thehole on the medial side of each plastic ear piece of the face-bow (Fig. 7). Make sure the brass-mountingpin fits securely into the hole and tighten the thumb screw knobs. Allow the anterior end of the upperframe of the articulator to rest on the horizontal cross bar of the face-bow. Note: Some clinicians liketo press the face-bow arms against one’s body while tightening the thumb screw knobs. Makesure the articulator with attached face-bow is stable. Bring the locking mechanism tabs to the “up”position to lock the articulator in position and to provide added stability.

III. Mounting the MaxillaryCast with a Direct MountingFace-BowPrepare the cast by placing retention groovesinto the base of the cast and soak the castbefore mounting. Seat the maxillary cast in theface-bow fork registration made previously andmake sure it is stable with no rocking. Someclinicians prefer to use some type of deviceplaced around the cast and bite fork forincreased stability. A face-bow fork support may

Page 10: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

10

Fig. 9

be used to support the bite fork during the mountingof the maxillary cast. The face-bow fork supportattaches to the lower frame in place of the lowermounting plate. The cross arms can be raised tocontact the under-surface of the face-bow fork toprevent flexing or movement of the fork (Fig. 9). Placemounting stone on the base of the cast and mountingplate. An inaccurate mounting will result if the bitefork flexes during mounting.

Close the upper frame to contact the cross bar,bringing the mounting stone on the two surfacestogether. Do not use too thick of a mix of mounting

stone or attempt to apply force when the stone has already begun to set. Keep the upper frame in thisclosed position until the mounting stone has set. If a device was used to secure the cast to the bite fork,carefully remove it and then remove the face-bow from the articulator. A second mix of mounting stonemay be used to fill in any voids left from the first mounting. Take care to avoid smearing excess mountingstone on the articulator parts.

Page 11: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

11

Fig. 10

Indirect Mounting Face-Bow

Fig. 11 Fig. 12

Fig. 13

MOUNTING THE MAXILLARY CAST USINGTHE INDIRECT MOUNTING METHOD

I. Preparing the Face-Bow and Articulator for Mounting CastsObtain the face-bow registration in the usual manner.Remove the face-bow from the patient and unscrewthe locking screw on the cross bar holding thevertical rod. This releases the transfer assemblywhich holds the face-bow fork (Fig. 11). Attach the#8686 Support Bar to the top of the Vertical Rod andsecure in place with the same locking screw (Fig. 12).With the #4611 Support Rod attached to the uppermember, rotate the upper frame of the articulatoruntil it is resting on the tabletop. This will make iteasier to install the transfer base assembly. The#9176Q QuickMount Transfer Base Assembly is usedfor demonstration in this manual. If present, remove

the mounting plate from the lower frame and install the #9176Q QuickMount Transfer Base Assembly.

II. Placing the Face-Bow Transfer Assembly on the ArticulatorInsert the vertical rod of the transfer assembly intothe transfer base assembly and lower it until thebottom of the vertical rod contacts the transfer base.Tighten the Clamp Screw so that the vertical rod fitssecurely in the transfer base (Fig. 13). Remove theincisal guide pin from the upper member. Place theupper frame of the articulator onto the lower frame sothe front of the upper frame now rests on the supportbar. Rotate the Locking Mechanism until the lever isin the “up” position. This locks the condyles againstthe posterior and superior walls of the fossa element.

III. Mounting the Maxillary Cast(Refer to Direct Mounting the Maxillary Cast on the 4000 Series Articulator on page 9, part III.)

Page 12: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

12

Fig. 14 Fig. 15

Fig. 16

MOUNTING THE MANDIBULAR CAST

I. Preparing the ArticulatorThe upper and lower frames are made parallel by aligning the top of the pin boss with the dark line, whichcompletely encircles the pin. The pin should then be adjusted 3–5 mm above the zero mark to compensatefor the thickness of the Centric Relation (CR) or Maximum Intercuspation (MI) registration used to mountthe lower cast. Make sure the Locking Mechanism tabs are in the “up” position. Set both progressive sideshift guides to the “0” position.

Place the articulator upside down. This positions the mounted cast with its occlusal surfaces upward.Place an interocclusal CR or MI registration record on the maxillary cast, making sure the record iscompletely seated – no rocking should occur (Fig. 14). If you haven’t done so, cut retention grooves intothe base of the lower cast and properly wet it prior to mounting. Position the mandibular cast on the CRor MI registration and check for stability.

For added stability, some clinicians use an alternate source to hold the upper and lower casts together.Hinge the lower frame into the open position and apply mounting stone to the base of the lower cast andthe lower mounting plate. Hinge the lower frame closed until the incisal guide pin meets the incisal guideblock (Fig. 15). Make sure the condylar elements are seated flush against the posterior and superior wallsof the condylar guides. Carefully hold the articulator in this position until the mounting stone has set. Asecond mix of mounting stone may be used to fill in any voids left from the first mounting to produce apleasing aesthetic result.

Finally, loosen the incisal guide pin screw and makesure the upper and lower casts contact. Retightenthe incisal guide pin screw and make sure the incisalguide pin is positioned in the center of the incisalguide block (Fig. 16).

Page 13: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

13

Fig. 17 Fig. 18

Fig. 19

SETTING THE CONDYLAR GUIDANCE OFTHE MODEL 4640 AND 4641 ARTICULATORS

After removing the CR or MI registration, release the centric lock by moving the tabs on the Multi-Functional Spring Mechanism to the mid-range position (Fig. 17). This allows the condyle mechanism tomove freely down the tracking surface of the fossa element. Loosen both condylar guides and set them tothe 0° setting indicated on the condylar inclination scale (Fig. 18). Firmly tighten the left condylar guidelocking knob and lightly secure the right condylar guide locking knob. Note: Set the progressive sideshift guides to their most open position (Fig. 19). Loosen the incisal guide pin screw and raise theincisal guide pin to prevent any interference. Retighten the incisal guide pin screw. Carefully seat the leftlateral excursion interocclusal record on the upper cast. It may be easier to perform this procedure withthe articulator inverted. While lightly holding the articulator together, make sure that the left condyle isseated “flush” against the rear wall. Gently seat the lower cast into the left lateral record and lightly holdthe articulator and casts in position. Note: The teeth of the cast should seat evenly into the leftlateral record. If the posterior teeth are raised out of the record the condylar inclination needs to bedecreased until the teeth are seated evenly in the record.

To set the inclination of the right condylar guide,carefully loosen the locking knob and rotate theguide until the teeth of the upper cast seat evenlyin the left lateral record. It is advisable whenmaking these adjustments that the contact shouldalso be judged by sight, rather than dependingsolely on the sense of touch. This helps ensure thatthe casts are not forced out of position. Tighten thecondylar guide locking knob to clamp the guide inposition. DO NOT USE EXCESSIVE PRESSUREwhen tightening the locking knob.

Page 14: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

14

Fig. 20

To set the right progressive side shift, loosen the rightside shift holding screw and move the rightside shift guide until it touches the side of thecondyle element (Fig. 20). Retighten the right sideshift guide locking screw. Record the amount ofcondylar inclination and progressive side shift foundon the left side on a patient registration card. Afterthe right condylar guidance has been recorded,the condylar guidance should be neutralized aspreviously described. The left condylar guidance isadjusted using the right lateral excursion recordand repeating steps listed above.

Page 15: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

15

SETTING THE CONDYLAR INCLINATIONUSING A PROTRUSIVE RECORD

Fig. 21 Fig. 22

Fig. 23

Custom Incisal Guide Table

Neutralize the condylar inclination and progressive side shift settings. Place the protrusive record on themaxillary cast and gently seat the mandibular cast into the protrusive record. It may easier to perform thisprocedure with the upper frame and its cast inverted. Both condylar elements will have moved away fromthe posterior surfaces of their respective condylar guides. Using sight and touch, rotate the left condylarguide until the teeth on the left side of the upper cast seat evenly into the protrusive record, then tightenthe condylar locking knob. Record the reading and repeat the procedure on the right side. Note: Thelateral records are used to determine the progressive side shift settings.

I. Fabricating a Custom Incisal Guide TableTo prevent possible wearing away of the stone casts during manipulation of the articulator, or to make apermanent record of a specific case, the natural incisal guidance may be recorded. Fabrication of a customincisal guide is accomplished by adding a layer of self-curing resin to the Plastic Incisal Guide Block or theDovetail Incisal Guide Block. Raise the incisal guide pin 1–2 mm for fabricating the custom incisal guide.Mix enough acrylic to place approximately 1/4" on top of the incisal guide block.

Carefully close the articulator in centric position and move the upper frame of the articulator back tosimulate a straight protrusive movement (end to end) (Fig. 21). From centric, move the upper member ofthe articulator to give a straight right lateral movement. Then, move the upper member of the articulator togive a straight left lateral movement. Move through all intermediate excursions between the lateral andprotrusive positions (Fig. 22). Allow the acrylic to harden and trim off excess acrylic.

Page 16: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

16

I. Constructing Occlusion RimsContour the upper and lower occlusion rims, establish occlusal vertical dimension, apply the customarymarkings to the wax rims and establish a centric jaw relation record. Create wedge-shaped V indices in theocclusion rims.

II. Preparing the Face-Bow ForkSome clinicians prefer to utilize a pronged bite forksuch as the #8631 Offset Bite Fork (Fig. 24). Cleanand properly disinfect the plastic ear pieces beforeeach use. Center the nasion relator assembly onthe cross bar. Loosen the #8604 thumb screws onthe face-bow and the #8643 thumb screws on thetransfer assembly.

III. Preparing Face-BowArmamentarium

Follow the procedures found on page 7.

IV. Positioning the Face-BowMake sure the midline of the palate and the stem of the bite fork is properly aligned. Position theface-bow onto the patient as described earlier on page 7. Once the face-bow record has been made,remove the wax rim and record base from the bite fork registration.

V. Obtaining Interocclusal RecordsTo make jaw relation records, an elastomericregistration material or other appropriate recordmedium is placed onto the V indices of the wax rim(Fig. 25). The patient is guided into centric relationand is allowed to close until the wax rims come intocontact. The registration material is allowed to setand is then removed.

VI. Preparing the Articulator for Mounting Casts

Many operators prefer to replace the plastic incisalblock with an adjustable metal table, such as the

#2460 Adjustable Incisal Guide Table (Fig. 26). Remove the plastic incisal guide block from the articulatorand install the adjustable incisal guide table. With the chiseled end pointed down, set the pin to the “0”mark, and adjust the position of the adjustable incisal guide table until the chiseled end of the incisalguide pin lies directly over the scribed line on the guide table (Fig. 27). Note: Guide pin must be setat “0” on the upper member.

COMPLETE DENTURE TECHNIQUE

Fig. 24

Fig. 25

Page 17: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

17

VII. Mounting the Maxillary CastFollow the same procedures as described earlier, Direct Mounting the Maxillary Cast on the 4000 SeriesArticulator, Mounting the Maxillary Cast using the Indirect Mounting Method.

VIII. Mounting the Mandibular CastPlace the jaw relation record between the wax occlusion rims. Invert the articulator with casts attached andapply mounting stone. Additional jaw relation records can verify accuracy. Accurate fit of the verificationrecords into the V-shaped indices will denote a precise mounting.

IX. Setting the Adjustable Incisal Guide TableAfter the anterior denture teeth are positioned, theadjustable guide table may be set.

Release the locking mechanisms on the articulator.Loosen the #2472 Indicator Clamp Knob of theadjustable incisal guide table and bring the anteriorteeth into edge-to-edge contact. Adjust the inclinationof the guide table until it contacts the incisal pin thenretighten the thumb screw (Fig. 28). Move the teethin a left lateral relation and set the right wing of thetable. Repeat this same operation for the left wingin a right lateral position.

X. Obtaining Protrusive RecordThe aesthetic try-in appointment affords the clinician the ability to verify tooth setup and obtain a well-indexed protrusive record. The protrusive record should be created at an extended protrusive position toallow for bilateral condylar movement. Set condylar inclination on the articulator as described previously.

Fig. 28

Fig. 26 Fig. 27

Page 18: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

18

Fig. 29

Fig. 30

XI. Fabricating a Remount IndexAfter denture processing, replace the mounting,cast, and denture onto the articulator. Attach theQuickMount Remount Jig on the lower frame(Fig. 29). Add sufficient stone to the remountjig to form an index of the cusp tips of thedenture teeth.

XII. Remounting CastsAfter polishing, the maxillary denture can be mountedto the articulator for a clinical remount to evaluatethe denture occlusion. Using the remount index,mount the maxillary denture and remount cast tothe articulator (Fig. 30). Using new centric relationrecords, mount the mandibular denture and remountcast. The denture occlusion may be evaluated andadjusted as needed.

Page 19: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

19

WHIP MIX RECORDING SYSTEM

Fig. 31

For the clinician who desires a recording system, Whip Mix offers the #8450 QUICK SET RECORDER.

The #8450 QUICK SET recording systems offers a time saving method for recording the protrusivecondylar path and measuring the amount of immediate side shift. The #8450 QUICK SET RECORDERprovides this information from an approximate hinge axis location.

Listed below are other potential uses of the #8450 QUICK SET RECORDER.1

1. Measurement of the timing or absence of a click during splint therapy.

2. Comparison of click patterns in vertical opening and lateral and protrusive excursions.

3. Setting an articulator for analysis or treatment.

4. Comparison of eminence angles before side-to-side restorative dentistry.

5. Estimates of eminence angles before surgical eminectomies.

6. Measurement of condylar movement after TMJ surgery.

1 Bates, Robert E., Welsch, Boyd B., and Stewart, Carol M., “Temporo Mandibular Joint Disk Position as Determined by aSimple Recorder,” Journal of Prosthetic Dentistry, Vol. 56, No. 2, pp. 221-224, 1986.

Page 20: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

20

INTERCHANGEABILITY WITH THE4000 SERIES ARTICULATOR

All 4000 Series Articulators feature the “ACCUMOUNT” Mounting System. The “ACCUMOUNT” Systemmakes it possible for the dental practitioner and the dental laboratory to interchange casts withoutexchanging articulators.

During manufacture, each 4000 Series Articulator has a special table firmly and precisely attached to thelower frame using a special fixture and low-fusing alloy.

The relationship between the upper and lower mounting plate is checked to verify precise alignment. Thisassures that casts can be interchanged between any 4000 Series Articulator without loss of accuracy.

Features and Advantages1. When clinicians and dental students use a dental laboratory that has a Series 4000 Articulator,

casts need no longer be mounted on an articulator when shipped to such laboratory.

2. Clinicians can purchase fewer instruments while maintaining the same level of care for

their patients.

3. Instrument damage and lost parts caused by shipping the articulator to and from the dental

laboratory is eliminated.

4. Precise alignment of the upper and lower frames is checked at the factory prior to shipping.

Page 21: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

21

1. When securing interocclusal records to be used in mounting casts and setting the articulator,

never allow the teeth to penetrate the recording material (impression compound, wax, gypsum,

impression paste, etc.) too deeply. They should never contact the opposing teeth, the metal

face-bow fork or any firm material that may be used as a carrying tray or handle. Any record

showing evidence of penetration should be discarded and remade.

2. The more stable a recording material is, the more it will resist distortion during its later use.

Any such material should be in a very soft state, however, during the initial recording procedure.

3. The following technique may be used to secure interocclusal records of partially

endentulous patients.

If natural dentition opposes the edentulous space, the partial occlusion rim is built up to nearly

touch the opposing teeth. Zinc oxide and eugenol impression paste is then added to the surface

of the occlusion rim of sufficient depth to register the tips of the opposing teeth when brought

into the desired relationship.

When the opposing spaces are both endentulous, one occlusion rim is built up in the customary

manner to near the occlusal plane, while the opposing rim is built to near this plane with small

cones of hard wax (or plastic) to indicate the registration in the impression paste.

When absence of teeth makes it necessary to obtain interocclusal records on partial occlusion

rims, these records must be made with the supporting soft tissue in as near a static condition

as is possible; some combination of these ideas can be planned to accomplish this with

acceptable accuracy.

4. With casts of unusually thin vertical dimensions, which would necessitate the use of a great bulk

of mounting stone, it is suggested that the mounting plate be built up to near the correct thickness

with a mix of mounting stone. This mix should be allowed to set for twenty minutes or longer before

the actual mounting procedure is performed. This will minimize any effect of the gypsum setting

expansion on the mounting accuracy.

5. It is suggested that a split cast mounting be made on the articulator before mounting the first case.

This will be used for reference purposes should the articulator ever be dropped or mishandled.

If the split cast mounting ever shows any discrepancy, return the articulator to the dealer for

factory recalibration at a nominal charge.

REMINDERS AND SUGGESTIONS

Page 22: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

22

The Whip Mix Model 4640 Articulator is a sturdily constructed instrument that will provide many

years of service with reasonable care. Both the upper and lower frames are made of cast aluminum.

All aluminum parts are anodized to prevent corrosion or staining. Each condyle element consists of a

stainless steel shaft and durable plastic condyle ball. The condylar guide assemblies are made of

anodized aluminum.

• Do not attempt to dislodge or remove the condyle ball from the condyle element shaft.

• The sealing compound placed over the #8548 Set Screws should not be disturbed.

• Do not attempt to remove the #2428 Screws holding each condyle release mechanism.

• Avoid getting wax or stone in the screw holes which may damage threads.

• Tighten screws snugly, but not too tightly. Overtightening the retaining screws can strip

the threads.

• It is a good idea to use a carrying case when the articulator is transported. Dropping the

articulator may result in bent or broken parts which may affect the articulator’s ability to

accurately reproduce a patient’s mandibular movements.

• A thin film of lubricant (Whip Mix LUBRIPLATE) or silicone spray applied to the surface upon

which the condylar elements move will provide a smooth action of these parts.

• Failing to remove excess stone, or not keeping the articulator clean, may result in corrosion

of the articulator surfaces.

• Apply silicone spray to articulator frame to prevent plaster or stone from sticking to surfaces.

• If you notice any bent parts on the articulator, have it serviced immediately to ensure it

maintains its interchangeability.

MAINTENANCE

Page 23: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

23

ARTICULATOR ACCESSORIES

#2460 Adjustable Incisal Guide #8533 Dovetail Incisal Block

#8579Q QuickMount Remounting Jig #8575 Remounting Jig

#8580B Plastic Mounting Plate #18156 Universal Tote Bag

Page 24: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

24

#9186 Face-Bow Adaptors

#8585 Face-Bow Fork Support#8587 Face-Bow Fork Support

#8631 Offset Face-Bow Fork

FACE-BOW ACCESSORIES

Page 25: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

25

QuickMount Magnetic Transfer Base Assemblyfor Hanau 2 Pin Articulators

#9176A Transfer Base Assembly

#9176Q QuickMount Magnetic TransferBase Assembly

QuickMount Face-Bow Fork Support

#8609 Bite Fork

Page 26: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

26

MODEL 4640ARTICULATOR

# PART# QTY. DESCRIPTION1 MA8543 2 FOOT BUMPER2 MA8527 1 INCISAL BLOCK SCREW3 MA4603 1 BOTTOM PLATE4 MA8526F 1 INCISAL BLOCK5 MA8344 1 MOUNTING TABLE (SHORT)6 MA2407 1 GUIDE PIN7 MA8509 2 PIN8 MA2444 1 BOSS9 MA8511 1 INCISAL PIN SCREW10 MA4601 1 TOP PLATE11 MA2410 1 BOSS12 MA2409 1 O-RING13 MA4611 1 TOP PLATE SUPPORT ROD14 MA4613 1 END BUMPER15 MA4605A 1 CONDYLAR GUIDE ASSEMBLY

(RIGHT)16 MA2432 2 CLAMP KNOB17 MA2424 2 NYLON WASHER18 MA4606A 1 CONDYLAR GUIDE ASSEMBLY

(LEFT)19 MA8519 2 FACE-BOW MOUNTING PIN20 MA2438 2 SET SCREW21 MA2408 2 CONDYLAR ELEMENT22 MA2422 2 HEX HEAD SCREW23 MA8810 2 PIN24 MA8545 2 #10 WASHER, STAINLESS STEEL25 MA8507 1 MOUNTING PLATE SCREW26 MA4602 1 UPRIGHT27 MA8345 1 MOUNTING PLATE SCREW28 MA8546 2 WASHER, BLACK FIBER, LARGE29 MA2042 2 ARTICULATOR FOOT30 MA8505 2 LEG31 MA8506 2 SCREW, SOCKET CAP32 MA8548 2 SCREW, SOCKET SET33 MA8508 2 MOUNTING PLATE KNOB34 MA8580 2 MOUNTING PLATE, METAL35 MA8580B 2 MOUNTING PLATE, PLASTIC

4000 SERIES ARTICULATOR PARTS LISTS

14

13

12

11

10

15 16

17

18

19

34

21

20

22

2324

25 1

8

9

7

6

4

3

2

1

5

26

30

31

3233

27

28

29

35

Page 27: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

27

MODEL 4640QARTICULATOR

# PART# QTY. DESCRIPTION1 MA8543 2 FOOT BUMPER2 MA8527 1 INCISAL BLOCK SCREW3 MA4603 1 BOTTOM PLATE4 MA8526F 1 INCISAL BLOCK5 MA8344 1 MOUNTING TABLE (SHORT)6 MA2407 1 GUIDE PIN7 MA8509 2 PIN8 MA2444 1 BOSS9 MA8511 1 INCISAL PIN SCREW10 MA4601 1 TOP PLATE11 MA2410 1 BOSS12 MA2409 1 O-RING13 MA4611 1 TOP PLATE SUPPORT ROD14 MA4613 1 END BUMPER15 MA4605A 1 CONDYLAR GUIDE ASSEMBLY

(RIGHT)16 MA2432 2 CLAMP KNOB17 MA2424 2 NYLON WASHER18 MA4606A 1 CONDYLAR GUIDE ASSEMBLY

(LEFT)19 MA8519 2 FACE-BOW MOUNTING PIN20 MA2438 2 SET SCREW21 MA2408 2 CONDYLAR ELEMENT22 MA2422 2 HEX HEAD SCREW23 MA8810 2 PIN24 MA8724 1 13 mm HOLDER WITH NUT25 MA8726 2 MAGNET26 MA4602 1 UPRIGHT27 MA2042 2 ARTICULATOR FOOT28 MA8505 2 LEG29 MA8506 2 SCREW, SOCKET CAP30 MA8725 1 25 mm HOLDER WITH NUT31 MA8727 1 WRENCH32 MA8728 4 QUICKMOUNT PLATE WITH DISK

14

13

12

11

10

15 16

17

18

19

32

21

20

22

23

24

251

8

9

7

6

4

3

2

1

5

26

30 27

28

29

31

Page 28: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

28

14

13

12

11

10

15 16

17

18

19

35

34

3233

21

20

22

2324

251

8

9

7

6

4

3

2

1

5

26

30

27

28

29

31

MODEL 4641ARTICULATOR

# PART# QTY. DESCRIPTION1 MA8543 2 FOOT BUMPER2 MA8527 1 INCISAL BLOCK SCREW3 MA4603 1 BOTTOM PLATE4 MA8526F 1 INCISAL BLOCK5 MA8344 1 MOUNTING TABLE (SHORT)6 MA2407 1 GUIDE PIN7 MA8509 2 PIN8 MA2444 1 BOSS9 MA8511 1 INCISAL PIN SCREW10 MA4601 1 TOP PLATE11 MA2410 1 BOSS12 MA2409 1 O-RING13 MA4611 1 TOP PLATE SUPPORT ROD14 MA4613 1 END BUMPER15 MA4621A 1 CONDYLAR GUIDE ASSEMBLY

(RIGHT)16 MA2432 2 CLAMP KNOB17 MA2424 2 NYLON WASHER18 MA4622A 1 CONDYLAR GUIDE ASSEMBLY

(LEFT)19 MA8519 2 FACE-BOW MOUNTING PIN20 MA2438 2 SET SCREW21 MA2408 2 CONDYLAR ELEMENT22 MA2422 2 HEX HEAD SCREW23 MA8810 2 PIN24 MA8545 2 #10 WASHER, STAINLESS STEEL25 MA8507 1 MOUNTING PLATE SCREW26 MA4602 1 UPRIGHT27 MA8345 1 MOUNTING PLATE SCREW28 MA8546 2 WASHER, BLACK FIBER, LARGE29 MA2042 2 ARTICULATOR FOOT30 MA8505 2 LEG31 MA8506 2 SCREW, SOCKET CAP32 MA8548 2 SCREW, SOCKET SET33 MA8508 2 MOUNTING PLATE KNOB34 MA8580 2 MOUNTING PLATE, METAL35 MA8580B 2 MOUNTING PLATE, PLASTIC

Page 29: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

29

WHIP MIX FACE-BOW PARTS LISTS

MODEL 8645FACE-BOW

FOR DIRECT MOUNTING

# PART# QTY. DESCRlPTlON1 MA8640 1 THUMB SCREW2 MA8643 1 THUMB SCREW3 MA2424 1 WASHER NYLON4 MA8617 1 SCREW5 MA8644 1 HORIZONTAL CLAMP ROD6 MA8608 1 SLIDE BAR ASSEMBLY7 MA8601 1 FACE-BOW (RIGHT)8 MA8603R 1 EAR PIECE (RIGHT)9 MA8603L 1 EAR PIECE (LEFT)10 MA8602 1 FACE-BOW (LEFT)11 MA8604 3 LOCKING SCREW12 MA8607 1 NOSE BLOCK13 MA8622 1 SCREW FOR UPRIGHT POST14 MA8606 1 NOSE PIECE SHAFT15 MA8605 1 UPRIGHT POST16 MA8609 1 BITE FORK17 MA8619 2 RETAINING RING18 MA8641 1 TOGGLE19 MA8616 1 LOCK WASHER20 MA8642 1 TOGGLENOT MA8549 PKG. OF 6 RUBBER-SHOWN BANDS

Page 30: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

30

MODEL 9600FACE-BOW

FOR DIRECT MOUNTING

# PART# QTY. DESCRlPTlON1 MA8640 1 THUMB SCREW2 MA8643 1 THUMB SCREW3 MA8617 1 SCREW4 MA2424 1 WASHER NYLON5 MA8644 1 HORIZONTAL CLAMP ROD6 MA8609 1 BITE FORK7 MA8608 1 SLIDE BAR ASSEMBLY8 MA9605 1 NUT9 MA9601-1 1 FACE-BOW (RIGHT)10 MA8603R 1 EAR PIECE (RIGHT)11 MA8603L 1 EAR PIECE (LEFT)12 MA9606A 3 SCREW13 MA9602-1 1 FACE-BOW (LEFT)14 MA8607 1 NOSE BLOCK15 MA8622 1 SCREW FOR UPRIGHT POST16 MA8606 1 NOSE PIECE SHAFT17 MA8619 2 RETAINING RING18 MA8605 1 UPRIGHT POST19 MA8641 1 TOGGLE20 MA8616 1 LOCK WASHER21 MA8642 1 TOGGLE22 MA8603A 1 EAR PIECE SOFT SETNOT MA8549 PKG. OF 6 RUBBER-SHOWN BANDS

Page 31: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

31

MODELS 9175/9185/9195

FACE-BOWSFOR INDIRECT MOUNTINGWITH DB2000/2200/9000/9800

ARTICULATORS

# PART# QTY. DESCRlPTlON1 MA9180 1 SCREW, RD HEAD2 MA9183 1 CLAMP3 MA9182 1 FIXED CLAMP4 MA2424 1 WASHER NYLON5 MA9176 1 TRANSFER BASE6 MA8644 1 HORIZONTAL

CLAMP ROD7 MA8609 1 BITE FORK8 MA9177 1 VERTICAL ROD FOR

#MA9175 FACE-BOWMA9187 1 VERTICAL ROD FOR

#MA9185 FACE-BOWMA9196 1 VERTICAL ROD FOR

#MA9195 FACE-BOW9 MA8686 1 SUPPORT BAR10 MA8679 1 CROSS BAR11 MA8601 1 FACE-BOW (RIGHT)12 MA8603R 1 EAR PIECE (RIGHT)13 MA8603L 1 EAR PIECE (LEFT)14 MA8604 4 LOCKING SCREW15 MA8602 1 FACE-BOW (LEFT)16 MA8622 1 SCREW FOR

UPRIGHT POST17 MA8607 1 NOSE BLOCK18 MA8606 1 NOSE PIECE SHAFT19 MA8605 1 UPRIGHT POST20 MA8619 3 RETAINING RING21 MA8641 1 TOGGLE CLAMP22 MA8616 1 LOCK WASHER23 MA8642 1 TOGGLE CLAMP24 MA8643 1 THUMB SCREW25 MA8640 1 THUMB SCREW26 MA9184 1 CLAMP SCREWNOT MA8549 PKG. OF 6 RUBBER-SHOWN BANDS

Page 32: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

32

MODELS 9275/9285/9295

FACE-BOWSFOR INDIRECT MOUNTING

WITH 2240/2340/8340ARTICULATORS

(PLASTIC EAR BOW)

# PART# QTY. DESCRlPTlON1 MA9180 1 SCREW, RD HEAD2 MA9183 1 CLAMP3 MA9182 1 FIXED CLAMP4 MA9176 1 TRANSFER BASE5 MA2424 1 WASHER NYLON6 MA8644 1 HORIZONTAL

CLAMP ROD7 MA8609 1 BITE FORK8 MA9610 1 CROSS BAR ASSEMBLY9 MA8604 1 LOCKING SCREW10 MA8606 1 NOSE PIECE SHAFT11 MA8607 1 NOSE BLOCK12 MA9608 4 WASHER NYLON13 MA9607 2 SCREW BUTTON HEAD14 MA9601-1 1 FACE-BOW (RIGHT)15 MA8903R 1 EAR PIECE (RIGHT)16 MA8603L 1 EAR PIECE (LEFT)17 MA9602-1 1 FACE-BOW (LEFT)18 MA9606A 1 SCREW19 MA9605 1 NUT20 MA8622 1 SCREW FOR

UPRIGHT POST21 MA8619 3 RETAINING RING22 MA8686 1 SUPPORT BAR23 MA8641 1 TOGGLE24 MA8616 1 LOCK WASHER25 MA8642 1 TOGGLE26 MA8643 1 THUMB SCREW27 MA9177 1 VERTICAL ROD FOR

#MA9275 FACE-BOWMA9187 1 VERTICAL ROD FOR

#MA9285 FACE-BOWMA9196 1 VERTICAL ROD FOR

#MA9295 FACE-BOW28 MA8640 1 THUMB SCREW29 MA9184 1 CLAMP SCREW30 MA8603A 1 EAR PIECE SOFT SETNOT MA8549 PKG. OF 6 RUBBER-SHOWN BANDS

Page 33: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

33

BIBLIOGRAPHY

The following bibliography gives more background on this instrument system.

Bates, Robert E., Welsch, Boyd B., and Stewart, Carol M.:Temporo Mandibular Joint Disk Position as Determined by a Simple Recorder.J. Pros. Dent., Vol. 56 No. 2, 221-224, 1986.

Cowan, Robert D., Sanchez, R.A., Chappell, R.P., Glaros, A.G., Hayden, W.J.:Verifying the Reliability of Interchanging Casts with Semi-Adjustable Articulators.Inter. J. Pros. Dent., Vol. 4, No. 3, 260-264, 1989.

Lee, Robert L.:Law Movements Engraved in Solid Plastic for Articulator Controls. Part 1,Recording Apparatus, J. Pros. Dent., 22:209, 1969.

Lee, Robert L.:Jaw Movements Engraved in Solid Plastic for Articulator Controls. Part II,Transfer Apparatus, J. Pros. Dent., 22:513, 1969.

Loos, Larry:Clinical Criteria Used to Select an Articulator, Compendium, Vol. XIV,No. 1, 80-82, 1993.

Lundeen, Harry C., Wirth, Carl G.:Condylar Movement Patterns Engraved in Plastic Blocks. J. Pros. Dent.,30:866, 1973.

Lundeen, H.C.:An Evaluation of Mandibular Border Movements: Their Character & Significance.J. Pros. Dent., 40:4424-452, 1978.

Lupkiewicz, S.M., Ariet, M., Fujimoto, J., Gibbs, C.H., Lundeen, H.C., & Mahan, R.E.:Reproductibility of Border Movements, Part 1, 2 & 3. IADR Progr. & Abst. 57:No. 367 & 368, 1978.

McCoy, R.B., Shyrock, E.F., & Lundeen, H.C.:A Method of Transferring Mandibular-Movement Data to Computer Storage.J. Pros. Dent., 36:510, 1976.

Sokolow, Stanley M.:Interchangeable Quick-Mounted Study Models. J. Clinical Orthodontics,Vol. XX, No. 11:779-781, 1986.

Welsch, Boyd B.:The Distribution of the Radius of the Curve Scribed During Protrusion.J. Pros. Dent., Vol. 51, No. 4:518, 1984.

Special thanks to the Postdoctoral Prosthodontics Program at theUniversity of Texas Health Science Center at San Antonio.

Page 34: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

34

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

NOTES

Page 35: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

35

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

________________________________________________________________

NOTES

Page 36: 4000 Series Articulator and QuickMount Face-Bow Instruction · patient-education purposes. Being an “ARCON” type instrument, the WHIP MIX 4000 Series Articulator is ideal for

For additional information, contactour Technical Department at:

MPL 30117 6/03

Whip Mix Corporation361 Farmington Ave.P.O. Box 17183Louisville, KY USA 40217-0183502-637-1451800-626-5651Fax 502-634-4512www.whipmix.com