Upload
urke
View
214
Download
0
Embed Size (px)
Citation preview
7/25/2019 1745_manu
1/65
1745Power Quality Logger
Users Manual
PN 2560366April 2006 Rev.1, 7/06 2006 Fluke Corporation, All rights reserved.All product names are trademarks of their respective companies.
99 Washington Street
Melrose, MA 02176
Phone781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/http://www.testequipmentdepot.com/7/25/2019 1745_manu
2/65
i
Table of Contents
Title Page
Introduction .......................................................................................... 1
Information and PC Software CD..................................................... 1Logger Power Supply ....................................................................... 2Logging Functions ............................................................................ 2Symbols ............................................................................................ 4
Safety Instructions ................................................................................ 4Qualified Personnel........................................................................... 6Standard Equipment and Optional Accessories ................................ 6Features............................................................................................. 8Power Network Configurations ........................................................ 11Working with Logged Data............................................................... 11
Using the 1745 Power Quality Logger.................................................. 12About Logging Jobs.......................................................................... 12Preparing the Logger for Use............................................................ 13
Test Leads - Markings .................................................................. 15Connecting Current Probes ........................................................... 15Logging with Voltage Converters................................................. 15
Connecting the Logger...................................................................... 16Connections in 3-Phase 4-Wire (Wye) Systems ........................... 19Connections in 3-Phase 3-Wire (Delta) Systems .......................... 20Connections for Single-Phase Logging......................................... 21Connections for Medium Voltage Networks................................. 22
Logging............................................................................................. 24Completing the Logging Job......................................................... 24
Evaluating the Logged Data.............................................................. 25Methods of Logging.......................................................................... 25Voltage Ranges ................................................................................. 26
Signal Sampling ............................................................................ 27Resolution Accuracy......................................................................... 27Voltage Variations ............................................................................ 27Averaging Period .............................................................................. 28
Min/Max Values ........................................................................... 29
Voltage Interruptions .................................................................... 30Voltage Dips and Swells............................................................... 31Voltage Harmonics ........................................................................... 31
7/25/2019 1745_manu
3/65
1745Users Manual
ii
Current Harmonics ........................................................................... 32Mains Signaling............................................................................ 32THD V In Function A................................................................ 33Calculation of THD in Measuring Function P.............................. 34Flicker .......................................................................................... 35Unbalance..................................................................................... 36Frequency..................................................................................... 36Current Logging ........................................................................... 36Logging Function A ..................................................................... 36Crest Factor (CF).......................................................................... 36
Power ............................................................................................... 37Measurement Theory............................................................................ 38Maintenance ......................................................................................... 42
Lithium Battery ................................................................................ 42Disposal............................................................................................ 42
Technical Specifications....................................................................... 43Logging Parameters Overview ...................................................... 43
Maximum Number of Intervals for Logging Funtion P................ 44General Information ..................................................................... 44Environmental Specifications....................................................... 45EMC ............................................................................................. 45Power Supply ............................................................................... 45Measurement ................................................................................ 46Input Voltage................................................................................ 46
Current Input with Flexi Set ......................................................... 47Current Input for Clamp ............................................................... 47General Specifications...................................................................... 48
RMS Logging Slow Voltage Variations....................................... 48Current Logging Values ............................................................... 48Events Dips, Swells, Interruptions................................................ 48Flicker .......................................................................................... 48Power P, S, |P| .............................................................................. 49Harmonics (Logging Function A Only)........................................ 49Statistics ....................................................................................... 49
Logging Function Parameters........................................................... 50Logging Values ............................................................................ 50Applications ................................................................................. 51
PQ Log PC Application Software ........................................................ 52Live Reading (Online Test) .......................................................... 53ASCII Export................................................................................ 54Timeplot Diagram ........................................................................ 55UNIPEDE DISDIP Table ............................................................. 56Cumulative Frequency Harmonics ............................................ 56
Index
7/25/2019 1745_manu
4/65
iii
List of Tables
Table Title Page
1. Symbols.............................................................................................. 42. Standard Equipment ........................................................................... 7
3. Optional Accessories.......................................................................... 74. 1745 Power Quality Logger - Controls and Indicators....................... 95. Test Leads - Markings........................................................................ 156. Measuring Ranges.............................................................................. 267. Logging Parameters - Overview......................................................... 43
7/25/2019 1745_manu
5/65
1745Users Manual
iv
7/25/2019 1745_manu
6/65
v
List of Figures
Figure Title Page
1. 1745 Power Quality Logger ............................................................... 32. 1745 Power Quality Logger - Front View.......................................... 8
3. Supplying Operating Power to the Logger ......................................... 144. Logging in a 3-Phase 4-Wire System................................................. 195. Logging in a 3-Phase 3-Wire (Delta) System..................................... 206. Single-Phase Logging ........................................................................ 217. Measuring 3 Phase Voltages in a 3-Wire (Delta)
System with Three Voltage Converters.............................................. 228. Two-Element Delta Connections ....................................................... 239. Basic Logger Setup Parameters.......................................................... 2610. Measuring Voltage Variations............................................................ 2811. Logging Min and Max Values............................................................ 2912. Voltage Interruption ........................................................................... 3013. Voltage Dips and Swells .................................................................... 3114. Measuring Flicker Values .................................................................. 3515. Live Reading (Online Test)................................................................ 5316. ASCII Export ..................................................................................... 5417. Timeplot Diagram .............................................................................. 5518. UNIPEDE DISDIP Table................................................................... 5619. Cumulative Frequency - For Voltage and Current Harmonics ........... 56
7/25/2019 1745_manu
7/65
1745Users Manual
vi
7/25/2019 1745_manu
8/65
1
1745
Power Quality Logger
Introduction
The Fluke 1745 Power Quality Logger, see Figure 1, is a sophisticated, easy-
to-use, electrical power-recording device for the electrician or power-quality
specialist.
Note
This manual also refers to the 1745 Power Quality Logger simply as
the Logger.
The Logger conatins an automatically charged 8-hour UPS (uninterruptable
power supply) battery for long-term voltage interruption ride-through, and an
LCD status display to provide confidence in setup before you leave the Loggerin place to collect data.
You can power the logger in parallel with the test leads or through an outlet.
Adapter cords for both methods of power in the Logger are included as
standard equipment.
Youll prepare the Logger for use with the included PQ Log software. You can
then connect the Logger to an electrical power-distribution network to log a
variety of power parameters, recorded as sequential averaged values over an
averaging period you can define. The Logger can measure up to three voltagesand four currents simultaneously.
Information and PC Software CD
The CD included with the Logger contains the PQ Log application software for
Windows, along with users manuals in multiple languages.
The PQ Log software prepares the Logger for use, and downloads data from
the Logger to a connected PC.You can then view the logged data in graphical
and tabular form, export it to a spreadsheet, or create reports for printing. For
details and instructions, see the PQ Log Users Manual on the CD.
7/25/2019 1745_manu
9/65
1745Users Manual
2
Logger Power Supply
The Logger does not include a power switch, but turns on automatically
whenever its power supply leads are connected to a voltage in its allowed
range. You can plug the Loggers power supply leads into a standard wall
outlet (using the included adapter cord), or you can connect them directly to
the power network under test (in parallel with the test leads) if there is no
convenient wall outlet (only if the voltage at the test leads is below 660V
RMS.)
Logging Functions
The Logger monitors power quality and locates disturbances in low andmedium voltage distribution networks. It measures up to 3 voltages and
4 currents. Logged values are saved in your choice of sequential averaging
periods. You graphically or numerically evaluated measured values with PQ
Log.
The Logger has two types of logging functions: logging function A
(Advanced) and logging function P (Power). Function A is the full set of
parameters, and function P provides logging capability optimized for load
studies and basic power logging. Function P contains every parameter inFunction A except voltage and current harmonics and interharmonics.
Logging function P allows longer logging periods because it does not save
harmonics values.
7/25/2019 1745_manu
10/65
Power Quality LoggerIntroduction
3
Logging function parameters:
RMS Voltage of each phase (average, min, max)
RMS Current of each phase and neutral (average, min, max)
Voltage events (dips, swells, interruptions)
Power (kW, kVA, kVAR, Power PF, Power tangent)
Energy, total energy
Flicker (Pst, Plt)
Voltage THD
Current THD
Current CF
Voltage harmonics to the 50th
(not in P function) Voltage interharmonics (not in P function)
Mains signalling voltage
Unbalance
Frequency
L1 A
L2 B
L3 C
LOGGINGSTATUS
CURRENTINPUT
10VRMSMAX5VA45-65Hz
SUPPLYINPUT
100-350V
660V MAX88-660V
VOLTAGEINPUT
600VCAT300V CAT
T
RS232
POWERQUALITYLOGGER
1745
830V 480V MAXL-L L-N/
START/STOP
/
egc001.eps
Figure 1. 1745 Power Quality Logger
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
7/25/2019 1745_manu
11/65
1745Users Manual
4
Symbols
Table 1 lists the symbols used on the instrument and in this manual.
Table 1. Symbols
Symbol Description
W Important information. See the manual.
X Hazardous voltage.
Earth ground.
T Double insulation.
F Direct Current (DC).
P Conforms to European Union requirements
)Canadian Standards Association is the certified body used
for testing compliance to safety standards.
~ Do not dispose of this product as unsorted municipal waste.Contact Fluke or a qualified recycler for disposal.
; Conforms to relevant Australian Standards.
Safety Instructions
Please read this section carefully. It will make you familiar with the most
important safety instructions for using the Logger.
Warnings identify conditions and actions that pose safety hazards to theuser and Cautions identify conditions and actions that can damage theLogger.
7/25/2019 1745_manu
12/65
Power Quality LoggerSafety Instructions
5
XWarnings
To avoid electrical shock, do not connect any part ofthe Logger to systems that have higher voltages to
ground (earth) than are marked on the Logger.
Areas between the power company meter and thesource of the distribution system are characterized asCAT IV areas. To avoid electrical shock or equipmentdamage, never connect the Logger to power in CAT IVareas if the voltage-to-earth ground is greater than300 V.
To avoid damaging the Logger, never connect itsvoltage measuring inputs to phase-to-phase voltages
higher than 830 V.
To avoid damaging the Logger, never connect thepower supply leads to voltages higher than 660 V-RMSAC.
The Logger is to be used and handled only by qualifiedpersonnel (see page 8).
Maintenance work on the Logger must be done only byqualified service personnel.
Use only the current probes specified in this manual. Ifyou use flexible current probes, wear suitableprotective gloves or work on de-energized conductors.
Do not expose the Logger to moisture or humidity.
To prevent electrical shock, always connect powersupply and voltage test leads to the Logger beforeconnecting to the load.
All accessories must be approved for 600 V CAT III orhigher.
Use the Logger only with its original standardequipment or with approved optional accessories, as
listed in Table 2 and Table 3 in this manual.
7/25/2019 1745_manu
13/65
1745Users Manual
6
Connect clip-on current transformers and/or Flexi Setto insulated live conductors only.
If measuring sensors are to be connected to non-insulated live conductors, additional personalprotective measures must be taken as required by localgovernment agencies.
Caution
To avoid damage, use the 1745 Power Quality Logger, onlywith the following nominal voltages:
Single/3 phase, 4-wire (Wye) systems (P-N): 69 V to480 V
3-phase-3-wire(Delta) systems (P-P): 120 V to 830 V
XWarning
To avoid electrical shock, or damaging the Loggersinternal protective circuitry weatherproof seal, do not openthe Logger.
Qualified Personnel
The following qualifications are required for using the Logger safely:
Trained and authorized to switch on/off, ground (earth), and mark
power distribution circuits and devices in accordance with electrical
engineering safety standards.
Trained or instructed in safety engineering standards for maintainingand using appropriate safety equipment.
Trained in first aid.
Standard Equipment and Optional Accessories
Table 2 lists the standard equipment for the 1745 Power Quality Logger and
Table 3 lists optional accessories.
7/25/2019 1745_manu
14/65
Power Quality LoggerSafety Instructions
7
Table 2. Standard Equipment
Equipment Model/Part
Number
Power Quality Logger 1745
International IEC Power Plug Adapter Set 2441372
RS232 Cable, Red, Null-Modem 2540511
Shielded 4-Phase Flexi Set (15 A/150 A/1500 A/3000 A) FS17XX
Dolphin Clip, Black (4x) 2540726
Color Coding Wire Clips WC17XX
Soft Case 2715509
English Users Manual 2560366
CD with Users Manual (English, German, French, Spanish,
Portuguese, Simplified Chinese, Italian), and PQ Log
software (same languages as the manual)
2583507
Power cord adapter for parallel connection to test leads 2651702
Power cord 2715492
USB Adapter 2539565
Table 3. Optional Accessories
Description Accessory
3-Phase Flexi Set MBX 3FLEX
3-Phase 1 A/10 A micro CT EPO405A
C425 Hardcase 2654865
Permlink Software for Modem E631820090
3-Phase Voltage Leads 2645854
Inspect the contents of the shipping box for completeness and damage. Report
any damage, to the shipper.
7/25/2019 1745_manu
15/65
7/25/2019 1745_manu
16/65
Power Quality LoggerSafety Instructions
9
Table 4. 1745 Power Quality Logger - Controls and Indicators
Item Name Description
1Connector forLogger powersupply.
This is where the power cord attaches to theLogger. The power cord connects in parallelto any two test leads as long as the voltageis below 660 V absolute maximum.Whenever there is any risk that voltagecould be higher, connect the power cord to awall outlet using the appropriateinternational power plug (supplied).
Power supply voltage range: 88-660 V AC or100-350 VDC, 50 Hz / 60 Hz, 600 V CAT III.
2
START/STOP switch The START/STOP button is used to start or
end switch-operated logging sessions.
3 LCD status display Shows measured input values to provideconfidence in proper setup and test leadconnections. Every 3 seconds, the displaychanges to the next set of readings in thefollowing sequence:
1. The three voltage levels
2. The main three phase currents
3. Neutral current and real-time clock
4. Active (true) power on each phase
The cycle repeats continuously. Make surethe readings look reasonable before leavingthe Logger to collect data.
4 Power supply leadsand 3-phase plusneutral voltage testleads
Fixed installed voltage input cables for L1 orA, L2 or B, L3 or C, N.
The highest permissible nominal voltage is830 V in a 3-wire network with Deltaconnection.
In a 4-wire network with Wye connection,the highest permissible nominal voltage is480 V.
When using PTs and CTs for measuringvoltage and current in a medium-voltagenetwork, refer to the IEC 60044 internationalstandard for guidelines.
7/25/2019 1745_manu
17/65
1745Users Manual
10
Table 4. 1745 Power Quality Logger - Controls and Indicators (cont)
Item Name Description
5 Connector for FlexiSet or currentclamps
Flexi sets or current clamps are detectedautomatically at power-up. If you change thecurrent probe type, be sure to remove andrestore power so the Logger will detect thenew current probe.
Nominal ranges for the Flexi Set are 15 A,150 A, 1500 A, and 3000 A AC. Nominalinput for current clamps is 0.5 V.
6 RS232 interface port The serial RS232 interface is used tocommunicate with a PC. The Logger isconnected to the PCs serial port (or amodem for remote communication) usingthe interface cable. Use a USB adapter ifnecessary.
7 Logging Status LED Yellow blinking = Logging job not set up yet,or waiting for start time or START button.
Green blinking = logging is in progress
Steady green = Logging session is finished.Data are ready to upload to a PC.
8 Channel LEDs The logging channel LEDs indicate
whether the applied voltages are within the
nominal range set using the PQ Log
software.
Red = Overload
Green = OK
Yellow = Underload(There is no detection for the current input)
7/25/2019 1745_manu
18/65
Power Quality LoggerSafety Instructions
11
Power Network Configurations
You can set up the Logger to work with several power network configurations
(listed below). You make these settings using PQ Log software while
connected to the Logger using the interface cable. See the PQ Log UsersManual for details.
Delta system
Delta 2 element system (2 voltage/current transformer)
Wye (Star) system
Single phase
Split single phase
Working with Logged Data
See the PQ Log manual for full details. Logged data can be evaluated using the
PQ Log software to provide the following:
Amount, date/time, and duration of quick and slow voltage variations
Half-cycle: 10 ms-extreme values for 50 Hz (8.3 ms at 60 Hz) MIN
and MAX for each measuring interval
Depth and duration of voltage dips
Correlation between peak current and voltage dips
95 %-flicker values according to EN50160
Number and duration of interruptions
Compliance of harmonic levels with defined limits
Mean and peak values of phase currents
Value of neutral conductor current
Current total harmonic distortion (THD) of phase and neutral
conductor currents
Profile of active, reactive, and apparent power versus time
Monitoring of power factor (PF) and information about effectiveness
of compensation systems
Graphical representations of logging data and statistics
7/25/2019 1745_manu
19/65
1745Users Manual
12
Using the 1745 Power Quality Logger
This section explains how to operate the 1745 Power Quality Logger. You
should also refer to the PQ Log Users Manual to become familiar with thesoftware you use to prepare the Logger for use and download logged data.
A typical logging session includes four steps:
1. Preparing the Logger for use with the PQ Log software.
2. Installing the Logger at the logging site.
3. Leaving the Logger to collect data for a period.
4. Downloading and evaluating the logged data.
These steps are described in the following pages.
About Logging Jobs
Logging jobs are defined using the PQ Log software, and transferred to the
Logger over the RS-232 cable. Each job contains the following information:
Logging function P or A
Measuring period, defined by start and end times
Time activated, switch or immediate job
Nominal voltage
Power type (wye, delta, etc.)
Averaging period length
Logging time period
Interharmonics and signaling voltages Limit values for events
Memory model for events: circular (first-in/first-out, continuous), or
linear (quit logging when logging period is finished)
Logging of current - neutral wire
Optional converter ratios for current and voltage if using potential
transformers (PTs) and current transformers (CTs) at a medium-
voltage network site
7/25/2019 1745_manu
20/65
Power Quality LoggerUsing the 1745 Power Quality Logger
13
Preparing the Logger for Use
Prepare the 1745 Logger for use with PQ Log software as follows (see
Figure 3):
1. Connect the Logger to line power. Use the power supply cables to
connect to an outlet, or to the test leads phase and neutral for Wye
configurations, or any two-phase leads for delta.
Caution
If you are powering the Logger in parallel with the testleads, and the voltage under test at the Logger power
supply connections could be greater than 660 V RMS AC,plug the power supply leads into an outlet instead.Otherwise, you could damage the Logger.
2. Connect the RS232 interface cable to the serial port of your PC. Use a
USB adapter if your PC does not have a serial port.
3. Run the PQ Log software as described in the PQ Log Users Manual.
4. Set up the Logging job and transfer the settings to the Logger.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
7/25/2019 1745_manu
21/65
7/25/2019 1745_manu
22/65
7/25/2019 1745_manu
23/65
7/25/2019 1745_manu
24/65
Power Quality LoggerUsing the 1745 Power Quality Logger
17
Use the Logger only with its original standardequipment or with approved optional accessories aslisted in Table 2 and Table 3 in this manual.
Connect clip-on current transformers and/or Flexi Setto insulated live conductors only.
If measuring sensors are to be connected to non-insulated live conductiors, additional personalprotective measures must be taken as required by localgovernment agencies.
Caution
To avoid damage, use the 1745 Power Quality Logger onlywith the following nominal voltages:
Single/3-phase 4-wire (Wye) systems (P-N): 69 V to 480 V
3-phase-3-wire (Delta) systems (P-P): 120 V to 830 V
XWarning
To avoid electrical shock, or damaging the Loggersinternal protective circuitry or weatherproof seal, do notopen the Logger.
7/25/2019 1745_manu
25/65
7/25/2019 1745_manu
26/65
Power Quality LoggerUsing the 1745 Power Quality Logger
19
Connections in 3-Phase 4-Wire (Wye) Systems
Figure 4 shows the connections for logging 3-phase 4-wire (Wye) systems:
L1 A
L2 B
L3 C
LOGGING STATUS
CURRENT INPUT
10VRMSMAX5VA45-65Hz
SUPPLY INPUT
100-350V
660V MAX88-660V
VOLTAGE INPUT
600VCAT300V CAT
T
RS232
POWERQUALITYLOGGER
1745
830V 480V MAXL-L L-N/
START/ STOP
/
L1 L2 L3 N
Power Supply
egc003.eps
Figure 4. Logging in a 3-Phase 4-Wire (Wye) System
7/25/2019 1745_manu
27/65
1745Users Manual
20
Connections in 3-Phase 3-Wire (Delta) Systems
Figure 5 shows the connections for logging 3-phase 3-wire (Delta) systems.
The test lead N can be left open or connected to ground potential.
L1 A
L2 B
L3 C
LOGGINGSTATUS
CURRENTINPUT
10VRMSMAX5VA45-65Hz
SUPPLYINPUT
100-350V
660V MAX88-660V
VOLTAGEINPUT
600VCAT300V CAT
T
RS232
POWERQUALITYLOGGER
1745
830V 480V MAXL-L L-N/
START/STOP
/
L1 L2 L3 N
Power Supply
egc004.eps
Figure 5. Logging in a 3-Phase 3-Wire (Delta) System
7/25/2019 1745_manu
28/65
Power Quality LoggerUsing the 1745 Power Quality Logger
21
Connections for Single-Phase Logging
Figure 6 shows the connections for logging single-phase logging systems:
L1 A
L2 B
L3 C
LOGGING STATUS
CURRENT INPUT
10VRMSMAX5VA45-65Hz
SUPPLY INPUT
100-350V
660V MAX88-660V
VOLTAGE INPUT
600VCAT300V CAT
T
RS232
POWERQUALITYLOGGER
1745
830V 480V MAXL-L L-N/
START/ STOP
/
L1
Power Supply
N
L3
L2
egc005.eps
Figure 6. Single-Phase Logging
7/25/2019 1745_manu
29/65
1745Users Manual
22
Connections for Medium Voltage Networks
In a 3-phase 3-wire (Delta) system with three separate voltage converters and
three current transformers, the Logger can measure phase-phase (P-P, Delta) or
phase - N (P-N, Wye). See Figure 7.
L1 A
L2B
L3 C
LOGGING STATUS
CURRENT INPUT
10VRMSMAX5VA45-65Hz
SUPPLY INPUT
100-350V
660V MAX88-660V
VOLTAGE INPUT
600VCAT300V CAT
T
RS232
POWERQUALITYLOGGER
1745
830V 480V MAXL-L L-N/
START/ STOP
/
L1
Power Supply
L2
L3
N
L1
L1
L2
L2
L3
L3
egc006.eps
Figure 7. Measuring 3 Phase Voltages in a 3-Wire (Delta) System withThree Voltage Converters
7/25/2019 1745_manu
30/65
7/25/2019 1745_manu
31/65
7/25/2019 1745_manu
32/65
Power Quality LoggerUsing the 1745 Power Quality Logger
25
2. Remove the test leads of the three phases. Be sure to remove the
measuring cable of the neutral wire last.
3. Remove the current probes.
Evaluating the Logged Data
Youll use PQ Log to evaluate the logged data. Data can be read out during
logging as well as at the end.
1. Connect the Logger to line power.
2. Connect the RS232 interface cable to your PCs serial port, then to the
Logger.3. Start the PQ Log software.
4. Use PQ Log to transfer the data from the Logger to the PC.
5. Once the data is transferred, remove the RS232 interface cable and
operating power from the Logger.
6. Evaluate the data using PQ Log.
For details, refer to the PQ Log manual.
Methods of Logging
The following section describes methods of logging using the 1745 Logger.
7/25/2019 1745_manu
33/65
7/25/2019 1745_manu
34/65
7/25/2019 1745_manu
35/65
7/25/2019 1745_manu
36/65
Power Quality LoggerUsing the 1745 Power Quality Logger
29
Min/Max Values
Logging detects the highest and lowest voltage RMS values and the highest
current RMS value during the test interval, using a minimum resolution of
10 ms.
The response time can be set in PQ Log to the following:
0.5, or 1 line power period
200 ms
1, 3, or 5 seconds.
Figure 11 shows the logging min and max values of the Logger.
U
t
Max. Value
Min. Value
Measuring Interval
egb017.eps
Figure 11. Logging Min and Max Values
7/25/2019 1745_manu
37/65
7/25/2019 1745_manu
38/65
7/25/2019 1745_manu
39/65
1745Users Manual
32
Current Harmonics
Current harmonics are defined as current components that have a frequency
that is an integer multiple of the fundamental frequency of the line power
current. Logging function A records ech individual harmonic of the phase
currents and the neutral current, up to the 50thorder, and presents the
harmonics as absolute values. The values are averaged over the interval length
defined in PQ Log.
Mains Signaling
Voltage components that have frequencies hat are not integer multiples of the
fundamental frequency of the line power voltage are called Mains Signalingvoltages or ripple-control voltages (interharmonics). The Logger can be
programmed to record up to five interharmonics with a resolution of 5 Hz. This
function can also be used to monitor ripple-control signals by entering the
signal frequency of the local utility.
The Logger measures the three second RMS value of each interharmonic and
establishes statistics for EN50160 (European standard) evaluation. These
statistics are available after a minimum recording time of 24 hours, or after
regular termination of the measuring job and can be exported from PQ Log andevaluated later.
Additionally, the Logger provides a long term recording of the interharmonics.
You can select from the following special measurement methods in the PQ
Log software:
200 ms maximum value (recommended for estimating ripple-control
signal levels)
200 ms minimum value 3 seconds maximum value
Average value over interval
In PQ Log, the frequency can be entered with a resolution of 0.5 Hz, but for
evaluation, the values are corrected to a bandwidth of 5 Hz. One frequency can
be defined for each band: for example, for a ripple-control signal of 183 Hz,
the values will be corrected to 185 Hz). Interharmonics of voltages and
currents with these frequencies are recorded.See the PQ Log manual for details.
7/25/2019 1745_manu
40/65
Power Quality LoggerUsing the 1745 Power Quality Logger
33
THD V In Function A
Function A:
1
50
2
2
V
V
VTHDn n==
Vn:RMS value of harmonic frequency #n.
V1:RMS value of the fundamental frequency.
THDV: total contents of harmonics of the line power voltage as a percentage
of the fundamental.
This algorithm is according to EN 61000-4-7.THD of currents:
Function A: 1
50
2
2
I
I
THDIn
n=
=
and ( ) =
=
50
2
2
n
nIATHDI
In: RMS value of harmonic frequency #n.
I1: RMS value of the fundamental frequency.
THDI: total contents of harmonics of the current as a percentage of the
fundamental.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
7/25/2019 1745_manu
41/65
7/25/2019 1745_manu
42/65
Power Quality LoggerUsing the 1745 Power Quality Logger
35
Flicker
Flicker is the visual impression of unsteadiness in a light source whose
luminance or spectral distribution changes over time. Flicker, see Figure 14, is
logged in accordance with the IEC 61000-4-15 standard. The short-term (st)flicker Pstis logged over a default standard interval of 10 minutes, and is used
to calculate the long-term (lt) flicker Pltby taking the sliding average of 12
short term values. The interval value can be changed as needed in PQ Log.
Formula for Plt Function
312
1
3
12=
=
i
PstPlt
Time
Flicker [%]
Interval length => 1 Pst
12 x interval length => 1 flicker value Plt
egb020.eps
Figure 14. Measuring Flicker Values
7/25/2019 1745_manu
43/65
1745Users Manual
36
Unbalance
The ratio of negative-to-positive-sequence harmonics is calculated with the
angles and magnitudes of the phase voltages taken into account. These values
are averaged over the interval length defined in PQ Log.
Frequency
The line power frequency is measured and averaged over 10 seconds and the
resulting values are divided into 42 classes for establishing statistics. Values
are also averaged over the interval length defined in PQ Log.
Current Logging
The maximum values of the currents (L1 or A, L2 or B, L3 or C and N) are
measured, and the interval value of the current is calculated using the mean
value over the RMS values of the interval defined in PQ Log.
Logging Function A
If a 3-phase current sensor is connected, the neutral current is calculated on asample basis from the phase currents. If a 3-phase+N sensor is detected, you
can select between logging and calculating the neutral current in PQ Log.
Ipeak
The peak values of the current (samples, not RMS values) are averaged in the
PQ Log software over the predefined measuring interval.
Note
Short peak values do not contribute much to the average value and so Imaxcan
be higher than Ipeak.
Crest Factor (CF)
The crest factor (CF) of the currents (L1 or A, L2 or B, L3 or C, and N) is the
ratio of current peak value divided by current RMS value, and is averaged over
the interval length defined in PQ Log. For sinusoidal signals, CF = 1.41, andfor square waves, CF =1.00
7/25/2019 1745_manu
44/65
Power Quality LoggerUsing the 1745 Power Quality Logger
37
Power
The power values (L1 or A, L2 or B, L3 or C and N) are averaged over the
interval length, and the maximum value of each is recorded.
The response time can be set to 1 second or 1 minute, and is independent of the
response time for voltage and current.
In logging function P, the active power, apparent power, and reactive power of
the phases and total power of all three phases is calculated.
Logging function A also calculates the distorting power D of the phases and
Dtotal.
7/25/2019 1745_manu
45/65
7/25/2019 1745_manu
46/65
Power Quality LoggerMeasurement Theory
39
11 PPh bas =
Active power of logginginterval per phase =
=
M
jjbasP
MP
1
1
Pbas jto 200 ms value
M: Number of 200 ms intervals per logging
intervals
Total active power on all
three phases ==
3
1kktotal PP
Pk: Active power of the phase
k: Phase (k = 1, 2, 3)
Absolute value of the
active power per interval
per phase
=
=
M
jjbasbetr
PM
P1
1
Sum of the absolute
values of the active
power on all three phases
=
++=
M
j
basbasbastotalbetrPPP
MP
1321
1
Apparent power based on
RMS values of voltage
and current. Basic values
on 200 ms per phase
basbasbas IVS .=
Apparent power per
logging interval per phase
=
=
M
jjbas
SM
S1
1
Sbas j: 200 ms value
M: Number of 200 ms intervals per loggingintervals
7/25/2019 1745_manu
47/65
1745Users Manual
40
Total apparent power on
three phases
=
=
3
1K
ktotal SS
k: Phase (k = 1, 2, 3)
Distortion power. Basic
value on 200 ms per
phase
222
basbasbasbas QPSD =
Distortion power per
interval per phase ==
M
j jbas
DM
D1
1
Dbas j: 200 ms value
M: Number of 200 ms intervals per logging interval
Total distortion power on
three phases=
=
3
1k
ktotal DD
Distortion power per
phase Q
Q
S
PPF ==
Total distortion power on
three phases total
total
total
total
totaltotalQ
Q
S
PPF ==
Tangent per phase
P
Q=tan
Total tangent on three
phasestotal
totaltotal
P
Q=tan
Active power of the
fundamental per phase
Basic value for 200 ms
11 PPh bas =
7/25/2019 1745_manu
48/65
7/25/2019 1745_manu
49/65
1745Users Manual
42
Maintenance
Caution
Maintenenca work on the done only by trained andqualified personnel at a company approved service centerwithin the warranty period.
With the proper use, the Logger does not require special maintenance other
than periodic calibration at a Flukecalibration center.
If the Logger gets dirty, wipe it off carefully with a damp cloth without
cleaning agents.
Lithium BatteryThe 1745 Logger contains a vanadium pentoxide lithium rechargeable battery
and a sealed gel-type, lead-acid battery. These batteries are automatically
recharged during normal operation. Neither one is user-servicable.
Disposal
If you discard the Logger, you must recycle it at an appropriate recycling
center as required by local regulations.
http://www.fluke.com/http://www.fluke.com/http://www.fluke.com/7/25/2019 1745_manu
50/65
Power Quality LoggerTechnical Specifications
43
Technical Specifications
Logging Parameters OverviewTable 7 shows an overview of the logging parameters.
Table 7. Logging Parameters - Overview
Measuring Function P A
Voltage: mean, min, max
values
Current: Mean, max-values
Neutral current N
Voltage events
Power: P, |P|, S, D, PF,
tangent
Power total P, |P|, S, D, PF,
tangent
Energy
Flicker: Pst, Plt
Voltage harmonics
Current harmonics (L1 or A,
L2 or B, L3 or C, N,
up to 50thorder)
Interharmonics, ripple-control
signals
THDV (voltage)
THDI (current)
CF (crest factor current)
Unbalance
Frequency
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
7/25/2019 1745_manu
51/65
7/25/2019 1745_manu
52/65
7/25/2019 1745_manu
53/65
1745Users Manual
46
Memory model Linear or circular, user-selectable
Interface RS-232, 9600 to 115.000 Baud, automatic selection,
3-wire communication.
Dimensions 170 mm x 125 mm x 55 mm
Weight Approx. 0.9 kg
MeasurementA/D converter 16 bit
Sampling frequency 10.24 kHz
Anti-aliasing filter FIR-Filter, fC = 4.9 kHz
Frequency response Uncertainty < 1 % of Vm for 40 Hz to 2500 Hz
Interval length 1, 3, 5, 10, or 30 seconds,
1, 5, 10, 15, or 60 minutes
Averaging time for
min/max values
, 1 line power period, 200 ms
1, 3, 5 s
Time base Resolution: 10 ms (at 50 Hz) deviation: 2 s/day at23C.
Input Voltage
Input range VIP-N: 69, 115, 230, or 480 V AC
Input range VIP-P 120, 200, 400, or 830 V AC
Max. overload voltage 1.2 VI
Input range selection Automatically set by user-entered nominal voltagevalue.
Connections P-P or P-N, 1- or 3-phase
Nominal voltage VN 999 kV (using PTs and ratios)
Input resistance Approx. 820 keper channel. Lx-N Single phase
(L1 or A, L2 or B, L3 or C connected): app. 300 ke
Intrinsic uncertainty 0.1 % of VI
Voltage transformer Ratio : < 999 kV / VI
Ratio selection Optional: user-selectable
7/25/2019 1745_manu
54/65
Power Quality LoggerTechnical Specifications
47
Current Input with Flexi Set
Input ranges IIL1 or A, L2
or B, L3 or C, N:
15, 150, 1500, or 3000 A AC
Measuring range 0.75 A to 3000 A AC
Intrinsic uncertainty < 2 % of II
Position influence Max. 2 % of m.v. for distance conductor to
measuring head >30 mm
Stray field influence 200 mm
Temperature coeff < 0.005 % / K
Current transformer Ratio : 999 kA / II
Ratio selection Optional: user-selectable
Connection Power type selectable in PQ Log
Current Input for Clamp
Input signal: 0.5 V AC nominal (for II) 1.4 V peak
Intrinsic uncertainty < 0.3 % of II
Max. overload 10 V AC
Input resistance App. 8.2 ke
Current transformer Ratio : 999 kA / II
Ratio selection By job programming
7/25/2019 1745_manu
55/65
7/25/2019 1745_manu
56/65
Power Quality LoggerTechnical Specifications
49
Power P, S, |P|
Active power P As per EN 61036, class 2
Distorting power D As per EN 61268, class 2 (A-version only)
Max value Highest value per interval
Min value Smallest value per interval
Phase uncertainty < 0.3 degrees
Conditions Conductor centered within clamp jaws or
Flexi Set.
Harmonics (Logging Function A Only)
Vm, Im, THDV, THDI per IEC/EN 61000-4-7, class B
Voltage harmonics (function A)
intrinsic uncertainty:
For Vm < 3 % VN: < 0.15 % VN
For Vm3 % VN: < 5 % Vm
Current harmonics (function A)
Intrinsic uncertaintyi
For Im< 10 % IN: < 0.5 % IN
For Im
10 % IN: < 5 % Im
THD V (function A) intrinsic
uncertainty at VN
For THD V < 3 %: < 0.15 %
For THD V 3 %: < 5 %
THD V (function P) intrinsic
uncertainty at VN
For THD V < 3 %: < 1 %
For THD 3 %: < 5 %
THD I (functions A, P) intrinsic
uncertainty at II
For THD I < 3 %: < 2 %
For THD I 3 %: < 5 %
Statistics
Frequency 42 classes for 10 s mean values
Ripple-control signals.
Interharmonics 21 classes for 3 s mean values.
Analysis of logging dataProgramming and analysis is done by PQ Log software on PC.
7/25/2019 1745_manu
57/65
7/25/2019 1745_manu
58/65
7/25/2019 1745_manu
59/65
1745Users Manual
52
PQ Log PC Application Software
PQ Log for PCs is the application for use with the 1745 Power Quality
Logger. The data are also available in ASCII format.Programs available for setting up the Logger:
Averaging period length
Memory model
Nominal voltage
Response time for min, max values
Power type (wye, delta, etc.)
Thresholds for event detection, interruptions
Setup:
Internal clock (date/time)
Assign 1745 Power Quality Logger instrument name
Parameters for data export
Software updates
Analysis:
ASCII data export
Graphical summary of all EN50160 parameters
Live reading viewing
7/25/2019 1745_manu
60/65
Power Quality LoggerPQ Log PC Application Software
53
Live Reading (Online Test)
Figure 15 shows a typical display of online test:
egb024.bmp
Figure 15. Live Reading (Online Test)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
FAX 781.665.0780 - TestEquipmentDepot.com
7/25/2019 1745_manu
61/65
1745Users Manual
54
ASCII Export
Figure 16 shows a typical display of ASCII export.
egb025.bmp
Figure 16. ASCII Export
For special cases, additional evaluations are available:
Graphic representation of measured data
Timeplot diagrams
Application oriented analysis
Logging value list
Table of events (UNIPEDE DISDIP)
Table summary
Cumulative frequency, harmonics (Logging Function A only)
Statistical values
All exceeding table
Most critical values
7/25/2019 1745_manu
62/65
7/25/2019 1745_manu
63/65
7/25/2019 1745_manu
64/65
7/25/2019 1745_manu
65/65