Setup to look at the HV current on side walls of HPD (diode is isolated from this current path). ??. Cable capacitance. HV. 1G. PC. HPD HV gnd. diode. 10 M. B. Voltmeter 20 Mohm OR. V. RBX gnd. Scope on 50 ohm. HV current to Ground through ~5Mohm Previous measured at CERN. - PowerPoint PPT Presentation
Citation preview
HV
10 M Voltmeter20 Mohm
OR
HPD HV gnd
RBX gnd
PC
diode
1G
VB
Scope on 50 ohm
Setup to look at the HV current on side walls of HPD (diode is isolated from this current path)
Cable capacitance
??
HV vs current (B=0) HPD =~10Tohm
-1.8
-1.6
-1.4
-1.2
-1
-0.8
-0.6
-0.4
-0.2
0
0 2 4 6 8 10 12
HV (kV)
Cu
rre
nt
(nA
)HV current to Ground through ~5Mohm Previous measured at CERN
DC measured current to ground and noise counts measured in pedestal vs B ( B, E = 20 degree, HV=6 kV)
-1.4
-1.35
-1.3
-1.25
-1.2
-1.15
-1.1
-1.05
-1
0.0 0.2 0.3 0.4 0.6 0.8 1.0 1.2 1.4
B (Tesla)
Cu
rre
nt
in n
A
0
10
20
30
40
50
60
No
ise
ev
en
ts a
bo
ve
10
fC
DC current[nA]
noise counts
Correlation between DC measured current through the side-walls of HPD and Noise (CERN B157)
-1
-0.5
0
0.5
1
1.5
-5.00E-07 5.00E-07(s)
(V)
current to ground (offset 1V and x10)
Pixel 1 (triggered)
Correlation between AC measured current through the side-walls of HPD and Large noise pulses(Pulse in the HV current is first) (negative so it is opposite from photo current)
HPD 291 no Angle
0
10
20
30
40
50
60
70
80
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6
B (T)
Bia
s cu
rren
t (n
A)
Bias Curr
Big Pulses
We can see the small pulses with keithly bias supplyWithout amplifier we see also large pulses in bursts
HPD 503 at 0.7 Tesla
05
101520253035404550
0 20 40 60 80 100 120 140
Angle (deg)
Bia
s C
urr
ent
(nA
)
bias curr
Big Pulses
When getting on an angle small pulses increase but large ones disappear
Lifting the HPD windowWe cut the HV cableWe milled of this the plastic and rubber on the top on a lay After that you should heat the Indium Tin seal to 80CWe had to poke with a needle to break vacuum before we could lift theWindow.
What also could be tried is don't pop the window but just heat it ones until the solder is melted and liquid (at that piont the vacuum is not broken ) and then let cooldown and re-measure noise. It could be that there are some variation from HPD to HPD in the seal ?????