7/28/2019 1-Standard Resistance Set
1/5
HCMC PC SPECIFICATION
STANDARD RESISTANCE SET
I. SCOPE :
This specification covers the supply of standard resistance set.
II. STANDARD OF MANUFACTURE AND TEST :
The equipment shall conform to the latest edition of theappropriate IEC standards and/or other international standards.
III. DESCRIPTION:
1. Function of equipment :The standard resistance set is a instrument for checkingresistance measuring equipment such as ohmmeter,megohmmeter, wheastone bridge, ...
2. Test method : conform to the specification of IEC standard
3. Technical specification :
- Value : from 0.000001 to 100000M-
Power : min. 1 Watt- Error : Max. 0.05% at 20C- Insulation :
100M (with resistance value from 0.000001 to 1M)
20000M (with resistance value from 1M to 1000M)
1G (with resistance value from 1000M to 100000M)
- Operating Temperature : - 15 to 55oC
-Storage Temperature : -30 to 75
o
C
IV. REQUIREMENT OF TECHNICAL DOCUMENT
SUBMITTED WITH THE QUOTATION :
-
Technical description, data sheet, catalogues and othermaterial submitted with the quotation must enable the
/var/www/apps/conversion/tmp/scratch_1/138239111.doc 1/5
7/28/2019 1-Standard Resistance Set
2/5
HCMC PC SPECIFICATION
STANDARD RESISTANCE SET
Purchaser to thoroughly evaluate the proposals as to itscompliance with the inquiry specification.
- Minimum requirements :+ Detail summary of deviations to the inquiry
specification.+ Completed inquiry datasheet expressed at section V.+ Catalogues containing outline dimension, technical
data+ List of main equipment and completed accessorieswhich can perform all function described in section III
have the following requirements :. Individual cost of main equipment and each accessory. Total cost of main equipment and completed accessory. Manufactures code of main equipment and eachaccessory
+ Connection diagram.+ Certificate of quality management.
V. REQUIREMENT OF DOCUMENT INCASE OF BID
WINNING :
1. Documents required to be provide for commission :In order to ensure the quality of supplied goods, in case of
bid winning, the Tenderer shall supply the following documentsfor the Purchaser :- Manufacturers Certificate of Quality.- Certificate of Origin issued by Chamber of Commerce of the
manufacturing country.- Certificate of calibration of equipment approved by an
independent lawful organization.
2. Documents required to be provide in accompany withsupplied products :
The Supplier shall enclose the following documents for suppliedproducts :
- 01 catalog- 01 instruction manual of installation, operation and
maintenance.
/var/www/apps/conversion/tmp/scratch_1/138239111.doc 2/5
7/28/2019 1-Standard Resistance Set
3/5
HCMC PC SPECIFICATION
STANDARD RESISTANCE SET
VI. REQUIREMENT OF INSTALLATION AND INSTRUCTION
FOR USE :
The Supplier shall ensure to supply the following service :- Installation the equipment completely.- Instruction to Purchasers staff how to use this equipment on
site.- Trial run this equipment.
VII. DATA SHEET :
N DESCRIPTION REQUIREMENT QUOTED
1. Manufacturer2. Country of origin3. Manufacture s code
4. Standard of manufactureand test
5. Standard of qualitymanagement
6. Supply of the certificate of
quality management7. In case of bid winning, the
Tenderer shall supply thefollowing documents for thePurchaser :- Manufacturers Certificateof Quality.- Certificate of Origin issuedby Chamber of Commerce of
the manufacturing country.- Certificate of calibration ofequipment approved by anindependent lawfulorganization.
Yes
Yes
Yes
8. List of main equipment andcompleted accessories metall requirements of sectionIV is supplied in tender
proposal.
Yes
9. Quoted main equipment and Yes
/var/www/apps/conversion/tmp/scratch_1/138239111.doc 3/5
7/28/2019 1-Standard Resistance Set
4/5
HCMC PC SPECIFICATION
STANDARD RESISTANCE SET
completed accessoriesdescribed in above list can
perform all functiondescribed in section IIIwithout addition of othernecessary materials.
10. The Supplier shall ensure tosupply the following service :- Installation the equipment
completely.- Instruction to Purchasers
staff how to use thisequipment on site.- Trial run this equipment.
Yes
Yes
Yes11. Function of equipment The standard
resistance set is ainstrument forchecking resistancemeasuringequipment such as
ohmmeter,megohmmeter,wheastonebridge, ...
12. Test method conform to thespecification of IECstandard
Technical specification13. Value from 0.000001
to 100000M14. Power min. 1 Watt15. Error Max. 0.05% at 20C
16. Insulation :- resistance value from
0.000001 to 1M- resistance value from
1M to 1000M- resistance value from
1000M to 100000M
100M
20000M
1G
/var/www/apps/conversion/tmp/scratch_1/138239111.doc 4/5
7/28/2019 1-Standard Resistance Set
5/5
HCMC PC SPECIFICATION
STANDARD RESISTANCE SET
17. Operating Temperature - 15 to 55oC18. Storage Temperature -30 to 75 oC
19.-
Dimensions :+ Width+ Height+ Depth
- Weight
The tenderer has toquote the
dimensions andweight of the
standard resistanceset
/var/www/apps/conversion/tmp/scratch_1/138239111.doc 5/5