1 Quality Control Quality Control Rad T 110 Rad T 110

1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

Embed Size (px)

Citation preview

Page 1: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

11

Quality ControlQuality Control

Rad T 110Rad T 110

Page 2: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

22

Aim :Aim :

The purpose of QC to achieve high The purpose of QC to achieve high quality images with the least amount of quality images with the least amount of radiation.radiation.

Quality Control is a part of Quality Quality Control is a part of Quality Assurance programAssurance program

Page 3: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

33

Test Types:Test Types:

Acceptance testAcceptance test Monitoring testMonitoring test Annual Test Annual Test After repair or tube After repair or tube

replacementreplacement

Page 4: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

44

Acceptance Test:Acceptance Test:

Performed upon newly Performed upon newly installed equipment installed equipment to verify purchase to verify purchase specification and to specification and to establish performance establish performance base line.base line.

Frequency:Frequency:• To be performed To be performed

once only when once only when new machine is new machine is installedinstalled

Page 5: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

55

Monitoring Test:Monitoring Test:

Performed for vital Performed for vital parameter on parameter on routine basis. routine basis.

Frequency:Frequency:• Every 2-3 monthsEvery 2-3 months

Page 6: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

66

Annual Test:Annual Test:

Performed for Performed for vital parameter vital parameter

Frequency:Frequency:• once a yearonce a year

Page 7: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

77

Parameters to be checkedParameters to be checked

1.1. Beam Alignment and Collimator AccuracyBeam Alignment and Collimator Accuracy

2.2. Radiation output and linearity of mR/mAs Radiation output and linearity of mR/mAs versus kVversus kV (Large and small FS(Large and small FS

3.3. Assessment of Total Beam Filtration (HVL)Assessment of Total Beam Filtration (HVL)

4.4. Assessment of Focal Spot size Assessment of Focal Spot size

5.5. Accuracy and constancy of Exposure TimerAccuracy and constancy of Exposure Timer

6.6. Measuring of Scattered Radiation (using Measuring of Scattered Radiation (using Water Phantom)Water Phantom)

7.7. Leakage Radiation from X-Ray tubeLeakage Radiation from X-Ray tube

1-5 Monitoring test, 2- Annual Test

Page 8: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

88

Parameters to be checkedParameters to be checked

1. Beam Alignment and Collimator Accuracy1. Beam Alignment and Collimator Accuracy

Beam Alignment Test Tool

Small hole

Page 9: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

99

Parameters to be checkedParameters to be checked

1. Beam Alignment and Collimator Accuracy1. Beam Alignment and Collimator Accuracy

X-ray TubeX-ray Tube

Test ToolTest Tool

Page 10: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1010

1. Beam Alignment and Collimator Accuracy1. Beam Alignment and Collimator Accuracy

Improper central ray alignment will distort a Improper central ray alignment will distort a radiographic image. radiographic image.

When used with the Collimator Test Tool, this device When used with the Collimator Test Tool, this device provides a simple means of determining if the X-ray provides a simple means of determining if the X-ray beam is perpendicular to the image receptor and beam is perpendicular to the image receptor and centered with respect to the light field.centered with respect to the light field.

Page 11: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1111

1. Beam Alignment and Collimator Accuracy1. Beam Alignment and Collimator Accuracy

Page 12: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1212

Parameters to be checkedParameters to be checked

2. Constancy of Radiation output and linearity of mR/mAs 2. Constancy of Radiation output and linearity of mR/mAs versus kVversus kV2 2 (Small & Large focus) (Small & Large focus)

Victoreen Kv Meter

Displays kVp maximum, kVp average, kVp effective, dose and time

Page 13: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1313

Parameters to be checkedParameters to be checked2.2. Constancy of Radiation output and linearity of mR/mAs versus kVConstancy of Radiation output and linearity of mR/mAs versus kV2 2 (Small & Large focus) (Small & Large focus)Set mAs=20 for LF & 10 for SFSet mAs=20 for LF & 10 for SF

X-ray TubeX-ray Tube

Table top Table top

Measure kVp & Output

Page 14: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1414

Radiation output and linearity of mR/mAs versus kVRadiation output and linearity of mR/mAs versus kV

•The mR/mAs relationship must be linear and acceptable

•kVp measured by the kV Meter must be within the acceptable range ( i.e. + 10% of kVp indicated by machine)

•The exposure dose (Output) must be within acceptable limits .

Page 15: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1515

Parameters to be checkedParameters to be checked

3. Assessment of Total Beam Filtration / (HVL) 3. Assessment of Total Beam Filtration / (HVL)

HVL(Half Value Layer): is the most frequently used factor for describing both the penetrating ability and the penetration through specific objects.

Defined: thickness of material penetrated by one half of the radiation

Unit: mm or cm

Filtration: attenuation of photon according to their Energy ( simply the ability to get rid of unnecessary radiation)

Page 16: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1616

Parameters to be checkedParameters to be checked

3. Assessment of Total Beam Filtration (HVL) 3. Assessment of Total Beam Filtration (HVL)

Page 17: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1717

Parameters to be checkedParameters to be checked

3. Assessment of Total Beam Filtration (HVL) 3. Assessment of Total Beam Filtration (HVL)

X-ray TubeX-ray Tube

Table top Table top

Measure Output only

Put Aluminum Shield with

different thickness 1,2,3,4

mm

Put Aluminum Shield with

different thickness 1,2,3,4

mm

Page 18: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1818

Parameters to be checkedParameters to be checked

3. Assessment of Total Beam Filtration (HVL) 3. Assessment of Total Beam Filtration (HVL)

AL mm Output (+ 2)Xi/X2

0 112.1

1 83

2 65.2

3 57.1

4 49.1 0.00

0.10

0.20

0.30

0.40

0.50

0.60

0.70

0.80

0.90

1.00

1.10

0 1 2 3 4 5Al mm

x/xi

HVL=3.0 mm (acceptable limit > 2.5 mm Al)

Page 19: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

1919

Parameters to be checkedParameters to be checked

4. Assessment of Focal Spot size4. Assessment of Focal Spot size

Focal Spot area: is the surface area of the anode that is bombarded by electron beam.

For Radiography units from 0.6 to 2.0 mm

For Mammo units 0.3 –to 0.6

Page 20: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2020

Parameters to be checkedParameters to be checked

4. Assessment of Focal Spot size4. Assessment of Focal Spot size

Focal Spot Test Tool

Fine Parallel line

Page 21: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2121

CouchCouch

Parameters to be checkedParameters to be checked

4. Assessment of Focal Spot size4. Assessment of Focal Spot size

X-ray TubeX-ray Tube

Low Speed film

Low Speed film

Test Tool60 cm from

the tube

Test Tool60 cm from

the tube

Focal Spot Area( level)Focal Spot Area( level)

Page 22: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2222

Parameters to be checkedParameters to be checked

4. Assessment of Focal Spot size4. Assessment of Focal Spot size

• A small focal spot size is used to obtain x-ray image with minimum blur.

•Small focal spot tend to concentrates heat and gives load on focal spot area.

•If the quantity of heat delivered during an individual exposure exceeds the track capacity, the anode surface can melt.

Page 23: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2323

Parameters to be checkedParameters to be checked

4. Assessment of Focal Spot size4. Assessment of Focal Spot size

• so … it is necessary to measure the focal spot size and compare it with the original focal spot size with the machine.

•Focal spot size of any machine is written on the top of X-ray Tube.

Page 24: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2424

Parameters to be checkedParameters to be checked

4. Assessment of Focal Spot size4. Assessment of Focal Spot size

Page 25: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2525

Parameters to be checkedParameters to be checked

5. Accuracy and constancy of Exposure Timer5. Accuracy and constancy of Exposure Timer

X-ray TubeX-ray Tube

Table top Table top

Measure the time compare it with time of exposure from the machine

Page 26: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2626

Parameters to be checkedParameters to be checked

6. 6. Measuring of Scattered Radiation (using Water Measuring of Scattered Radiation (using Water Phantom)Phantom)

When x-ray beam or gamma radiation passes through an object it can either

1. Penetrate the object without interacting

2. It can interact with matter and be completely absorbed.

3. It can interact and scattered or deflected from original direction.

Page 27: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2727

Parameters to be checkedParameters to be checked

6. 6. Measuring of Scattered Radiation (using Water Measuring of Scattered Radiation (using Water Phantom)Phantom)

X-r

ay T

ube

X-r

ay T

ube

Water or Chest

Phantom

X-ray Chest Bucky

Ionization Chamber 1 m from center

Ionization chamber

Page 28: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2828

Parameters to be checkedParameters to be checked

7. 7. Leakage Radiation from X-Ray tubeLeakage Radiation from X-Ray tube

Rarely done.Rarely done. Typically for very old units.Typically for very old units. To test if there is any leakage of radiation from the x-ray tube. To test if there is any leakage of radiation from the x-ray tube.

Page 29: 1 Quality Control Rad T 110. 2 Aim : The purpose of QC to achieve high quality images with the least amount of radiation. The purpose of QC to achieve

2929

Fluoroscopic Unhappiness and spatial ResolutionFluoroscopic Unhappiness and spatial Resolution

Parameters to be checkedParameters to be checked

•The test pattern should be placed on the input window of the II and oriented so that the bar patterns are imaged perpendicular to the raster lines of the monitor display.