30
WorkCover. Watching out for you. Put your Co. Logo Here PREVENTION OF FALLS LEANING LADDERS AND STEPLADDERS PRESENTED BY: (type your name here) 1

1 FAST-CAL Last Updated 10/11/2004. 2 FAST-CAL 3 Introducing FAST-CALs A New Range of four Field Block Calibrators FAST-CALs work from -35°C to 650°C

Embed Size (px)

Citation preview

1

FAST-CAL

FAST-CALLast U

pdated 10/11/2004

2

FAST-CAL

FAST-CAL

3

Introducing FAST-CALs

A New Range of four Field Block Calibrators

FAST-CALs work from -35°C to 650°C in three ranges -35°C to 140°C,

30°C to 350°C and 35° to 650°C

4

Why are we introducing it?

During 2004 20 experienced engineers from many parts of the world specified their ideal products for Industrial Calibration

FAST-CAL realises their top ten requirements of…

FAST-CAL The product range you helped specify and design

5

FAST-CAL Top Ten

1. Rugged 2. Light weight 3. Easy to use onsite 4. Low cost /high benefit ratio, 5. Fast response, high stability 6. Time saving features 7. Multiple sensor testing 8. Software 9. Modern design 10. Compliant with latest regulations.

6

Membranes?

The team told us “Whatever you do don’t use membrane keypads. They look great new, but after a year of being poked with ball point pens and screwdrivers the results have to be seen to be believed.”

We were going to include a picture but we might upset the competition.

7

Result of Input? The FAST-CALs

Fast-Cal Features: Rugged, hard wearing stainless steel case. Hygienic with no paint to chip or flake. No membrane keypads. Lightweight, approx 6 kg depending upon the model. Easy to use on site, we used the latest computer technology to

design the handle at the exact balance point for each Fast-Cal model. A soft, light weight shoulder case makes transporting the Fast-Cal effortless.

Fast response, high stability. Isotech has put 24 years of manufacturing and design experience into optimising the performance of the Fast-Cal. Compare the specifications overleaf. The thermal gradients are minimised by non-linear heating and cooling profiles to compensate for end losses to ambient temperature.

8

More FAST-CAL Features

The Fast-Cals are so easy to use you hardly need the manual. The High and Medium Fast-Cals incorporate a built-in Insert holder for

ultra fast changes and re-starts of calibration cycles. The Fast-Cals (excepting the 2010) all have the same sized removable

inserts which can be drilled to your exact requirements. Communications are fitted as standard and a number of programs are

available depending on your needs. The Fast-Cal design is modern and complies with all the relevant

requirements and directives which are applicable. The Fast-Cals come with a certificate traceable to National Standards.

Five point UKAS Certification is available as an alternative. Mains power variance immunity has been standard on all Isotech dry

block calibrators since their conception. This feature resolves problems with fluctuating power supplies which if not acknowledged causes instability in the bath.

9

Summary of Features

10

Secrets of Success

11

Basic or Complete?

12

FAST-CAL Inserts

13

Software

CalNotePad: Included as standard. This permits you to connect both an indicator and the calibration bath to the PC. You can view data on clear configurable chart displays, log data to file and control the calibration bath.

I-cal: A software application that can automatically calibrate up to 16 temperature sensors, it provides an expandable low cost route to automatic calibration – with I-cal Capture you can store pictures (along with time and temperature data) to enable, for example, dial thermometers to be calibrated automatically; return to a set of images of the dial captured at each calibration temperature!

14

FAST-CAL HTM 2010

Validation of Washer Disinfectors, Steam Sterilisers, Autoclaves to HTM 2010 and HTM 2030

For accurate calibration of sensors choose the FAST-CAL HTM 2010

In place of a removable insert it has a fixed block Block has room for reference PRT and test

sensors Place the probes directly in the block and gain

better accuracy (Avoids thermal interface between insert and block)

15

FAST-CAL Low

Unique Advantages Wider Temperature Range

-35°C to 140°C

Better Accuracy +/-0.2 to +/-0.15°C

Deeper Calibration Well No Membranes Smallest in its Class Lightweight Soft Shoulder Case

16

FAST-CAL Low

Unique Advantages continued

Ambient is the most difficult temperature to control at

The Eurotherm controllers designed for Isotech are “seamless”

Ask the competition to control at ambient and see the instabilities

17

FAST-CAL Low

Unique Advantages continued

Insensitive to ambient variationsOne competitors bath changes its

calibration by 0.8°C for a 16°C change in ambient

Our equivalent models changed by 0.03°C for a 20°C ambient change!

This is important, these calibrators are for field use

18

Accuracy here is what we mean by accuracy

For BASIC: The agreement between the controller indication compared to a close fitting standard located at the bottom of the dry well insert

For COMPLETE: The agreement between the Indicator and a close fitting standard located at the bottom of the dry well insert

19

Accuracy International Standard

EA10-13 is the only published document relating to the full performance of dry block calibratorshttp://www.european-accreditation.org/

FAST-CAL meet the requirements of EA10-13

FAST-CAL is the most accurate in its class!

20

Integrity in Temperature Calibration

ErrorsThe largest error source will probably be

due to heat conduction along the stem of the unit under test, (UUT)

Most users do not consider this, and because it is not part of the manufacturers specification they ignore it

21

Integrity in Temperature Calibration

We Do Not Ignore Stem Conduction!E.g

A 6mm diameter Pt100 with a 30mm sensing length will have an error of 0.1% (0.1°C @ 120°C) in a bath with an immersion of 120mm

In a FAST-CAL with immersion of 148mm the error is 0.005% or 5mK, i.e. 20 times better!

For more information on stem conduction look at our website and e-learning

www.isotech.co.uk

22

Integrity in Temperature Calibration

FAST-CAL – Calibrate larger diameter sensors for the same accuracyE.g

What is the maximum diameter of thermocouple that can be calibrated for an accuracy of 0.5%?

If the immersion is 120mm = 7.5mm diameter In a FAST-CAL with 148mm immersion = 9.2mm

The extra immersion lets you calibrate larger sensors

23

There is more

An insert in a Dry Block Calibrator has an air gap between it and the heater or cooled block, the insert too has stem conduction and it can give rise to surprisingly large error

24

There is more

E.g. A block with an insert 32mm diameter and 120mm deep has a depth of 3.75 x its diameter giving a 2% stem conduction offset Equivalent to 2°C at 120°C, 4°C at 220°C etc

The FAST-CAL has an insert 25mm diameter by 148mm deep or 6 x diameter equivalent to 0.5°C at 120°C, Four times better than that of a competitors bath

25

There is more

That is why we recommend the “Complete” option because in it we measure the temperature of the insert – where the actual calibration takes place

Isotech products are better by design

26

Choice of Standards

Use our recommended Platinum Resistance Thermometers for high accuracy, low uncertainty measurements

27

Choice of Standards

For less demanding applications we can offer high quality MI thermocouples Still allows you to measure

the insert temperature where the calibration takes place

Easily angled to sit under head mounted enclosures

Low Cost

28

There is more

Mains Variation Immunity All Isotech block baths have built in immunity

to mains variations giving you the benefit of better stability

Isotech products are better by design

29

Certification and Traceability

All FAST-CALs are tested and certified at Isotech

Isotech’s NTPL laboratory has the smallest uncertainties down to +/-0.025°C for some metal block baths and +/-0.00007°C at 2 Sigma for Water Triple Point and Gallium Cells

Isotech – confidence comes as standard

30

Software

FAST-CALs ship with Cal Notepad included We put benefits like thermostat testing,

setpoint programming and monitoring into the software which is readily upgradeable and expandable not locked into firmware that is not accessible to you

31

Software FAST-CAL is compatible with

I-cal Easy I-cal Easy Enables you to doAutomatic Calibration, Certificate Printing,

Coefficient Calculator…Try the free evaluation version

32

Cost of Ownership

FAST-CALs have world class designed and built instrumentation (Designed by Eurotherm)

If anything does go wrong it takes just a few seconds to unplug and exchanged

Compare to competition where the whole unit has to be returned to the manufacturer

Eurotherm – are one of Isotech’s “Partners in Calibration”

33

Widest Product Range Worldwide

Isotech can supply all your calibration needs www.isotech.co.uk Isothermal Technology Limited

Pine Grove,Southport,Merseyside.PR9 9AGEngland Telephone 01704 543830 International +441704 543830 Fax 01704 544799 International +441704 544799 [email protected]

34

Recalibration Service

From one of our four UKAS accredited labsKeep your FAST-CAL in perfect condition.

Why not use our recalibration service? Need to know more about calibration?

E-learning, www.isotech.co.uk/elearning

35

Worldwide Service

Isotech has 90 Agencies World Wide Contact us for your nearest Isotech Outlet