VIDAS BLUE VIDAS BLUE PRESENTATION VIDAS BLUE VIDAS BLUE: New Vidas and mini Vidas look

Preview:

Citation preview

VIDAS BLUE

VIDAS BLUEVIDAS BLUE PRESENTATIONPRESENTATION

VIDAS BLUE

VIDAS BLUE:

New Vidas and mini Vidas look

VIDAS BLUE

The main changes

VIDAS BLUE

VIDAS BLUE

Strip preparation tray

Blue tray covers

VIDAS BLUE

Removingstrip preparation tray

Mini VIDAS BLUE

Mini VIDAS BLUE

New screen / keyboard Holder

New printer paperholder

Internal changes

VIDAS BLUE

VIDAS BLUE

On mini VIDAS:Addition of a long screw to fix the right trim

VIDAS BLUE

On mini VIDAS:

Screen/keyboard holder 3 fixing points:

VIDAS BLUE

New Power Supply Assembly: - Only 4 external points of fixation (4 screw)- Central chimney (V30)- No railing(bar)

VIDAS BLUE

New Power Supply Assembly:- Only 1 Power Supply Board (V12/V30)- 1 air fan- Self switching Power Supply (110/220V)

VIDAS BLUE

New Power Supply Board:-Dedicated mini Vidas and Vidas connectors.

VIDAS

Mini VIDAS

Be carefullRespect the correct connector!

VIDAS BLUE

New Power Supply Board:

-Air Fan•Fixed by 4 plastic retainersFixed by 4 plastic retainers•Spare partSpare part•If Air Fan out of service:If Air Fan out of service: Software T° error will be Software T° error will be generatedgenerated Automatic Power SupplyAutomatic Power Supplycut offcut off

VIDAS BLUE

New tray ball bearings:

Excentric ball b

earing

replaced by a normal

one fixed on a sprin

g

mechanism.

No more lateral play to adjustNo more lateral play to adjust

VIDAS BLUE

Tray motor assembly:

Addition of a pretensionning spring Addition of a pretensionning spring

CAUTION!CAUTION!

Lock the motor Assy

Lock the motor Assy

when it iswhen it is

in the good position

in the good position

VIDAS BLUE

Tower screw:

Addition of a spacerAddition of a spacer

To reduce the tower misalignmentsTo reduce the tower misalignments

VIDAS BLUE

Incubator door:

- New door plunger ball nut- New door plunger ball nut

- Addition of a safety latch- Addition of a safety latch

VIDAS BLUE

New electronic boards:

Scanner BoardScanner Board

Incubator BoardsIncubator Boards

VIDAS BLUE

New Scanner Board:

Flash ROM U5 ( V2.51)Flash ROM U5 ( V2.51)(replace the EEPROM)(replace the EEPROM)

DIP Switch SW3DIP Switch SW3(replace JP7/JP8/JP9) (replace JP7/JP8/JP9)

Power Supply LEDsPower Supply LEDs(+5V/+10V/+12V/-(+5V/+10V/+12V/-12V)12V)

New Power Supply 3.3VNew Power Supply 3.3VLEDLED

Sensors LEDsSensors LEDs

Scanner cable Scanner cable breakdown detectionbreakdown detectionLEDsLEDs

Communication LEDsCommunication LEDs

Mini Vidas Blue BacklightMini Vidas Blue BacklightPower Supply connectorPower Supply connector

J15: Scanner Cable new J15: Scanner Cable new Connector for failure Connector for failure detectiondetection(replace coax. cables ) (replace coax. cables )

VIDAS BLUE

New Scanner Board:

•100 % compatible with the old boards100 % compatible with the old boards

•One Flash ROM One Flash ROM V 2.51V 2.51 instead of 3 EPROM V2.41 instead of 3 EPROM V2.41

•SW3 replaces Jumpers JP7 to JP9SW3 replaces Jumpers JP7 to JP9

•New power supply 3,3VNew power supply 3,3V

•New connector (J18) dedicated to the mini New connector (J18) dedicated to the mini Vidas screen backlight power supply.Vidas screen backlight power supply.

•New connector (J15) for scanner failure New connector (J15) for scanner failure detection ( replaces coax cables) detection ( replaces coax cables)

REF. SCANNER BOARDS : VD8201-1REF. SCANNER BOARDS : VD8201-1

VIDAS BLUE

VIDAS BLUE

VIDAS BLUE

New Incubator Board:New Incubator Board:Double support Flash ROM U5Double support Flash ROM U5or EPROM V 3.10or EPROM V 3.10

DIP switch SW2DIP switch SW2(replace adress/(replace adress/reset jumpers)reset jumpers)

Power Supply LEDsPower Supply LEDs

Heaters LEDsHeaters LEDs

Serial communicationSerial communicationwith scanner boardwith scanner boardLEDsLEDs

Sensor LEDsSensor LEDs

VIDAS BLUE

New Incubator Board:New Incubator Board:

• Available with EPROM 3.10 Available with EPROM 3.10 oror Flash ROM Flash ROM (same version)(same version)

• Dip Switch SW2 replace all Adress and Dip Switch SW2 replace all Adress and Reset jumpers ( Reset jumpers ( silkscreensilkscreen directly on the directly on the board).board).• LEDS for Power Supply, communication, LEDS for Power Supply, communication, sensors and heaters.sensors and heaters.

• Fuses (SMD) F1 to F4Fuses (SMD) F1 to F4

REF. INCUB.Board : VD8202-2REF. INCUB.Board : VD8202-2

• 100% compatible100% compatible with the old boards with the old boards

VIDAS BLUE

VIDAS BLUE

VIDAS BLUE

SCANNERSCANNER: new parts

• New mini VIDAS/ VIDAS scanner cablesNew mini VIDAS/ VIDAS scanner cables

New scanner cables with failure detectionNew scanner cables with failure detection(replace coaxial cables)(replace coaxial cables)

The blue flat cable is still the The blue flat cable is still the samesame

VIDAS BLUE

SCANNERSCANNER: new parts• New Scanner PRE-AMP boardNew Scanner PRE-AMP board

• J1 connector unchangedJ1 connector unchanged, use with , use with both old scanner câble kits and newboth old scanner câble kits and newScanner cables.Scanner cables.

• New J2 connectorNew J2 connector, is only used with , is only used with new scanner cables and new scannernew scanner cables and new scannerBoard.Board.(Allows scanner cable failure detection)(Allows scanner cable failure detection)

VIDAS BLUE : reminder

The scanner cable failure detection system is available onlyThe scanner cable failure detection system is available only with a new Scanner Cable, new Scanner Board, with a new Scanner Cable, new Scanner Board, and new Scanner Head.and new Scanner Head.

The LEDs (DS9 to DS12 from the new scanner board) are onThe LEDs (DS9 to DS12 from the new scanner board) are onWhen a cable is broken.When a cable is broken.( ( off = cables OKoff = cables OK))

If a scanner cable failure is detected, a blocking error 161 If a scanner cable failure is detected, a blocking error 161 will be generated.will be generated.

The new Scanner head is compatible with the old Scanner headThe new Scanner head is compatible with the old Scanner headBut But don’t forget to replace the new scanner cable by an old don’t forget to replace the new scanner cable by an old scanner cable kit in order to be compatible with the old scanner scanner cable kit in order to be compatible with the old scanner boardboard

VIDAS BLUE

Changes impact

Decontamination: Decontamination: a new procedure has been defined to take a new procedure has been defined to take Into account the « Strip preparation tray ».Into account the « Strip preparation tray ».

Error 161: Error 161: If err. 161, check the integrity of the scanner If err. 161, check the integrity of the scanner cables. ( LEDs off on scanner board = good)cables. ( LEDs off on scanner board = good)

PM : Scanner cable replacement every 2 yearsPM : Scanner cable replacement every 2 years Non ApplicableNon Applicable

PM : PM : tray lateral play not necessary to set tray lateral play not necessary to set

PM : Power Supply PM : Power Supply not necessary to open it during a PM not necessary to open it during a PM(If dead, replace the complete block.)(If dead, replace the complete block.)

VIDAS BLUE

Compatibility table

(*) Scanner Head Blue : put an old scanner cable kit(*) Scanner Head Blue : put an old scanner cable kit

(**) Scanner Cable Blue: only used with a new (**) Scanner Cable Blue: only used with a new Scanner Head and new Scanner Board.Scanner Head and new Scanner Board.

VIDAS BLUE

You will receive a specific Mandatory Action Required (MAR)You will receive a specific Mandatory Action Required (MAR) from GCS soon.from GCS soon.

Feel free to contact us for many information :Feel free to contact us for many information :

E-mail : E-mail : gcsimmunomarcy@eu.biomerieux.com

More details will be added in this MAR .More details will be added in this MAR .