View
98
Download
0
Category
Preview:
Citation preview
SMSFixed Syngas Monitoring System
Professional online gas analysis systems for continuous monitoring of pyrolysis and gasification processes.
Measures All Primary Constituents of Syngas In modern gasification where syngas composition is paramount for
determining the calorific value, the a1-cbiss SMS provides an
accurate, cost-effective solution to monitoring all major constituents
of the gasification process
Highly Durable to Harsh Process Conditions The SMS is well equipped to work in environments where there are
high levels of particulates and tar present. Sample preparation and
conditioning is a key element of the system, upon which substantially
depends on the correctness of the consequent measurement, as well
as the lifetime of the measurement sensors.
The a1-cbiss SMS utilises varying degrees of high temperature probes
ranging from stainless steel through to kanthal.
KEYBENEFITS
• Proprietary infrared dual beam NDIR
detectors for CH4, CnHm, CO and CO2
• On-board Calculation and display of
the BTU index (gas calorific value, only
with syngas analysers)
• Keyboard interface for configuration
and calibration
• Stainless steel connectors for gas inlet/
outlet and zero air inlet ports
• RS232 serial COM port (for real time
data download to external PC or
laptop as text file, software included)
The SMS provides extraction, conditioning & analysis
all at a single point “ “
GASESANALYSEDCO CH4 H2CO2
O2CnHm
11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HTT: +44(0)151 666 8300 F: +44(0)151 666 8329 E: sales@a1-cbiss.com W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com
Cert No. 996ISO 9001
TECHNICALSPECIFICATION
PNEUMATIC SPECIFICATION• Gas sampling pump: 2L/min @ -300 mbar
• Pressure reducer on gas sample/calibration inlets set @
+20 mbar
TEMPERATURE & PRESSURE • Relative humidity: 0 – 95 % non condensing
• Pressure: 800 – 1200 hPa
• Temperature:+5°C to + 45°C
POWER • Power supply: 220 VAC / 110 VAC
• System power supply: 220 VAC / 110 VAC
PHYSICAL • Dimensions: 1.200mm x 600mm x 600 mm (H x W x D)
• Weight: approx. 100 kg (with instrumentation)
• Metal plate thickness: 2 mm
• Glass thickness front door: 5 mm
• Ingress protection: IP54
Complete Project Solutions
• Expert consultancy • Project management • In-house project design & build • Installation & commissioning
• After product service & support
Our complete project solutions are developed to assist you in IED compliance. We achieve this through;
Find a1-cbiss on:
MEASURINGRANGESGas name Measuring Principle Range Resolution Accuracy T90
Carbon monoxide NDIR 0-40% 0.01% ±2% FS < 10 s
Carbon dioxide NDIR 0-30% 0.01% ±2% FS < 10 s
Methane NDIR 0-10% 0.01% ±2% FS < 10 s
Hydrogen Thermal Connectivity 0-50% 0.01% ±2% FS < 10 s
Hydrocarbons (CnHm) NDIR 0-10% 0.01% ±2% FS < 10 s
Oxygen Fuel Cell 0-25% 0.01% ±2% FS < 15 s
Recommended